• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall MEDLINE Sterile Needle Counter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall MEDLINE Sterile Needle Countersee related information
Date Initiated by FirmApril 24, 2018
Create DateMay 25, 2018
Recall Status1 Terminated 3 on August 04, 2020
Recall NumberZ-1936-2018
Recall Event ID 80017
Product Classification Needle, suturing, disposable - Product Code GAB
ProductMEDLINE O.R. Necessities(R) STERILE R, Needle Counter/Compte-aiguiles/Caja de agujas, Gold Standard/Standard Or/Oro y Estander, 40 Ct. Foam Block / Magnet with Blade Remover, Coussinet en mousse pour 40 aiguilles /Magnetise avec quitte-lame Or, 40 Ct Bloque de espuma / Iman Con extractor de hoja - Oro, REF NC40FBRGS Product Usage: A needle counter is intended to be used safely and easily accounting for used, disposable surgical sharps during a surgical procedure. The product does not come in contact with the surgical site or the patient.
Code Information Lot Numbers: 67016114502 67016124502 67017014501 67017024502 67017064502 67017064503 67017074501 67017074502 67017084502 67017084503 67017104501 
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries Inc
Three Lakes Drive
Northfield IL 60093
For Additional Information ContactKassandra Cotner
866-359-1704
Manufacturer Reason
for Recall
The product packaging seal has the potential to experience seal creep. This may result in gaps to be created in the seals of the packaging and may impact the sterility of the Needle Counters.
FDA Determined
Cause 2
Package design/selection
ActionMedline sent an Urgent Recall letter dated April 24, 2018. The letter identified the affected product, problem and actions to be taken. Customer were instructed to check inventory, quarantine all affected product and return completed enclosed response form. Those consignees who further distributed the product were directed to notify their customers. For question contact 866-359-1704.
Quantity in Commerce2600 units
DistributionWorldwide Distribution - US Nationwide and the countries of: Canada, Saudi Arabia, Cayman Island, and United Arab Emirates
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-