| | Class 2 Device Recall ARTIS Pheno, Fluoroscopic XRay system |  |
| Date Initiated by Firm | May 30, 2018 |
| Create Date | July 20, 2018 |
| Recall Status1 |
Terminated 3 on February 14, 2020 |
| Recall Number | Z-2524-2018 |
| Recall Event ID |
80417 |
| 510(K)Number | K163286 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | ARTIS Pheno, Fluoroscopic X-Ray system, Model Number 10849000 |
| Code Information |
164033, 164071, 164034, 164095, 164057, 164054, 164055, 164080, 164072, 164088, 164048, 164094, 164068, 164019, 164013, 164073, 164025, 164026, 164027, 164060, 164083, 164021, 164066, 164075, 164074, 164084, 164086, 164098, 164028, 164053, 164063, 164064, 164024, 164049, 164018, 164022, 164061, 164100, 164097, 164011 |
| FEI Number |
2240869
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc
40 Liberty Blvd
Malvern
PA
19355-1418
|
| For Additional Information Contact | Meredith Adams 610-448-6461 |
Manufacturer Reason for Recall | During manual LAO/RAO rotation at maximum speed or a combination of LAO/RAO and Cran/Caud rotation at maximum speed, the possibility exists for the safety monitoring within the system control to become activated. As a result, the C-arm stops the rotation and can drop up to 40 mm (1.6 ). |
FDA Determined Cause 2 | Device Design |
| Action | An Urgent Medical Device Correction notification was distributed to customers on 5/30/18. The notification provided the following information to customers:
Siemens will correct the error by inspecting all affected ARTIS Pheno systems. As an immediate measure to prevent movement stops during a C-arm rotation followed by a C-arm drop, the maximum rotation speeds will be reduced. This correction limits the maximum manual LAO/RAO rotation from 25/sec to 20/sec and Cran/Caud rotation from 20/sec to 10/sec.
All other system parameters as well as 3D acquisition and Perivision/Peristepping remain unaffected. You may continue to use the ARTIS Pheno system while awaiting implementation of the corrective action.
Our service organization will contact you shortly to arrange a date to perform this corrective action with Update
Instruction AX004/18/S. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436.
Following the corrective action, the cause is eliminated and a recurrence of the fault is prevented. Additionally, Siemens is developing a solution to restore the system back to the original settings. This solution will be distributed to all affected customers upon completion. |
| Quantity in Commerce | 40 in the United States |
| Distribution | The devices were distributed to the following US states: AL, CA, CO, FL, IL, KY, MA, MI, MN, MS, ME, NJ, NY, OH, TN, TX, UT, and VA. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = OWB
|
|
|
|