| | Class 2 Device Recall Discovery MI |  |
| Date Initiated by Firm | May 24, 2018 |
| Create Date | July 31, 2018 |
| Recall Status1 |
Terminated 3 on February 08, 2019 |
| Recall Number | Z-2580-2018 |
| Recall Event ID |
80484 |
| 510(K)Number | K161574 |
| Product Classification |
System, tomography, computed, emission - Product Code KPS
|
| Product | Discovery MI, labeled as Discovery DMI 15 cm 3 ring, and Discovery DMI 20cm 4 ring |
| Code Information |
Model Number 5454001-170. Serial Number (System ID Number): CJRPX1800016CN ( 4550966PT11), CJRPX1800017CN (561422MI), CJRPX1800018CN (617726PET1EB), CJRPX1800021CN (404616DMI) |
| FEI Number |
2126677
|
Recalling Firm/ Manufacturer |
GE Healthcare, LLC
3000 N Grandview Blvd
Waukesha
WI
53188-1615
|
| For Additional Information Contact | 800-437-1171 |
Manufacturer Reason for Recall | Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply). |
FDA Determined Cause 2 | Component change control |
| Action | GE Healthcare notified customers on about 05/24/2018 via URGENT MEDICAL DEVICE CORRECTION letter. The letter outlined the reason for recall and notified customers that a GE Healthcare representative would be contacting them to arrange for the correction of the devices. Any questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or a local Service Representative. |
| Quantity in Commerce | 42 units total |
| Distribution | Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = KPS
|
|
|
|