• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Kerlix Bandage Roll

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Kerlix Bandage Rollsee related information
Date Initiated by FirmAugust 01, 2018
Create DateOctober 03, 2018
Recall Status1 Terminated 3 on June 03, 2020
Recall NumberZ-0016-2019
Recall Event ID 80924
510(K)NumberK990530 
Product Classification Dressing, wound, occlusive - Product Code NAD
ProductKerlix Bandage Roll, 100 Cotton, 8 Ply, Large 4-1/2 x 9.3 (11.4 cm x 2.8 m), REF 6716 Product Usage: Used as both primary and secondary dressing for bandaging heads, limbs and difficult-to-dress wounds.
Code Information Lot Numbers: 18E071762,18E072262,18E072362,18E132262,18E132362
FEI Number 1282497
Recalling Firm/
Manufacturer
COVIDIEN LLC
15 Hampshire St
Mansfield MA 02048-1113
For Additional Information Contact
508-261-8000
Manufacturer Reason
for Recall
Potential for product sterility breach due to a compromised or pinched seal defect
FDA Determined
Cause 2
Nonconforming Material/Component
ActionOn August 1, 2018 Cardinal Health issued Urgent Medical Device Recall notices to U.S. customers via Fed Ex and OUS customers were directly contacted by Quality & Regulatory Affair representatives supporting that country. Customers were advised to take the following actions: 1. Inspect stock and quarantine affected product. 2. Immediately stop using affected product. 3. Return the completed acknowledgment form via fax to 847-689-9101 or 614-652-9648, whether or not you have affected product, as Cardinal Health is required to confirm receipt of this notification from customers, and to prevent further notices. 4. Notify any customers to whom you may have distributed product affected by this recall. 5. Contact the appropriate Customer Service group to arrange for credit and return of any affected product you may have: Hospital 800-964-5227 Federal Government 800-444-1166 Distributors800-635-6021 All other customers 888-444-5440 Customers with any questions regarding this letter may call (800) 292-9332.
Quantity in Commerce2,730,821 units in total
DistributionUS Nationwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = NAD
-
-