 Class 2 Device Recall syngo.CT Cardiac Planningsee related information
Date Initiated by FirmMarch 21, 2019
Create DateApril 09, 2019
Recall Status1 Terminated 3 on January 11, 2022
Recall NumberZ-1113-2019
Recall Event ID 82517
510(K)NumberK170221 
Product Classification System, x-ray, tomography, computed - Product Code JAK
Productsyngo.via syngo.CT Cardiac Planning, Model Number 10496180 Product Usage: Syngo.CT Cardiac Planning and syngo CT. Cardiac Function are image analysis software packages for evaluating cardiac CT angiography (CTA) volume sets. The software packages are designed to support the physician in determining the functional parameters of the left and right ventricles, confirming the presence or absence of physicianidentified myocardial enhancement defects and evaluation documentation and follow-up of any such finding.
Code Information Software versions VB20A or VB30A
FEI Number 2240869
Recalling Firm/
Manufacturer
Siemens Medical Solutions USA, Inc
40 Liberty Blvd
Malvern PA 19355-1418
For Additional Information ContactRebecca Tudor
610-219-4834
Manufacturer Reason
for Recall
There is a potential risk of a wrong measurement in the annulus plane during a TAVI planning procedure. This risk is due to a software issue found in the TAVI algorithm.
FDA Determined
Cause 2
Software design
ActionUrgent Customer Safety Advisory Notice letters dated 3/20/19 were sent to customers. The letter identified the affected product, problem and actions to be taken. For questions call 1-800-888-7436.
Quantity in Commerce112 total
DistributionUS Nationwide Distribution in the states of: AK, AL, AR, AZ, CA, CT, FL, GA, IA, IL, KS, KY, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, and WI.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = JAK
-
-