| Date Initiated by Firm | March 22, 2019 |
| Create Date | May 07, 2019 |
| Recall Status1 |
Terminated 3 on May 12, 2020 |
| Recall Number | Z-1295-2019 |
| Recall Event ID |
82613 |
| 510(K)Number | K181560 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | Cios Alpha with software version VA30-mobile C-arm x-ray
Material # 11105200
Product Usage:
The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients |
| Code Information |
Cios Alpha system with SW VA30 |
| FEI Number |
2240869
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc
40 Liberty Blvd
Malvern
PA
19355-1418
|
| For Additional Information Contact | SAME 610-219-4834 |
Manufacturer Reason for Recall | Continuous operation of the Cios Alpha at high tube outputs may result in increased wear of radiation emitting components, risk of premature failure of the Monoblock and loss of imaging X-rays exists |
FDA Determined Cause 2 | Device Design |
| Action | Siemens issued Customer Safety Advisory Letter dated 3/22/19 advising of problem, health risk and action to take: recommends to avoid overwriting the organ programs, as this can lead to frequent operations. Siemens is distributing this letter and the addendum to the user manual to all potentially affected users Contact our service organization for any additional questions at 1-800-888-7436. |
| Quantity in Commerce | 2 US |
| Distribution | US Nationwide Distribution in the states of NC, MA |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|