| | Class 2 Device Recall VITROS 250 Chemistry System Refurbished |  |
| Date Initiated by Firm | April 08, 2019 |
| Create Date | July 03, 2019 |
| Recall Status1 |
Terminated 3 on November 12, 2020 |
| Recall Number | Z-1935-2019 |
| Recall Event ID |
82784 |
| 510(K)Number | K922072 |
| Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
| Product | VITROS 250 Chemistry System Refurbished, Product Code 6801759,
(UDI # 10758750001330)
Product Usage:
Performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides. |
| Code Information |
All lots |
| FEI Number |
1000136573
|
Recalling Firm/ Manufacturer |
Ortho-Clinical Diagnostics 100 Indigo Creek Dr Rochester NY 14626-5101
|
Manufacturer Reason for Recall | Potential negative drift in Creatine Kinase (CK) results when used on the VITROS 250 and 350 Systems. |
FDA Determined Cause 2 | Device Design |
| Action | On 08 April 2019, a customer letter (Ref. CL2019-090) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and and/or US Postal Service Priority Mail (for PO Boxes only) to either all customers who were shipped affected VITROS 250 or 350 Systems. This communication provided the following required actions:
- As per the VITROS CK Slide Instructions for Use, perform Quality Control (QC) testing daily to help ensure the slides are performing within acceptable limits.
- If QC results are outside of the established ranges, discard the cartridge and repeat QC testing using a fresh cartridge of VITROS CK Slides.
- If multiple cartridges are loaded on the VITROS 250/350 System, daily QC testing must be performed for each cartridge prior to use.
- Consider loading one cartridge of VITROS CK Slides on board your system at a time.
- Complete and return the request form via fax to 1-888-557-3759.
Foreign affiliates were informed by email on 08 April 2019 of the issue and instructed to notify affected consignees of the issue and appropriate actions to take. |
| Quantity in Commerce | 2411 units |
| Distribution | US Nationwide Distribution |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = JJE
|
|
|
|