• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall STANLEY Healthcare Arial 54315 Network Manager environmental control, powered

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall STANLEY Healthcare Arial 54315 Network Manager environmental control, poweredsee related information
Date Initiated by FirmJune 14, 2019
Create DateJuly 19, 2019
Recall Status1 Terminated 3 on May 13, 2021
Recall NumberZ-1996-2019
Recall Event ID 83266
Product Classification System, environmental control, powered - Product Code IQA
ProductSTANLEY Healthcare Arial 54315 Network Manager
Code Information Serial Numbers:  0008DC517FD2, 0008DC517FD3, 0008DC517FD4, 0008DC517FD5, 0008DC517FD6, 0008DC517FD7, 0008DC517FD8, 0008DC517FD9, 0008DC517FDA, 0008DC517FDB, 0008DC517FDC, 0008DC517FDD, 0008DC517FDE, 0008DC517FDF, 0008DC517FE0, 0008DC517FE1, 0008DC517FE2, 0008DC517FE3, 0008DC517FE4, 0008DC517FE5, 0008DC51807D, 0008DC51807E, 0008DC51807F, 0008DC518080, 0008DC518081, 0008DC518082, 0008DC518083, 0008DC518084, 0008DC518085, 0008DC518086, 0008DC56ECDA, 0008DC56ECDB, 0008DC56ECDC, 0008DC56ECDD, 0008DC56ECE1, 0008DC56ED32, 0008DC56ED33, 0008DC56ED37, 0008DC56ED38, 0008DC56ED39, 0008DC56ED3A, 0008DC56ED3B, 0008DC56ED3C, 0008DC56ED3D, 0008DC56ED56, 0008DC56ED57, 0008DC56ED58, 0008DC56ED5A, 0008DC56ED5B, 0008DC56ED5D, 0008DC56ED5F, 0008DC56ED60, 0008DC56ED61, 0008DC56ED66, 0008DC56ED67, 0008DC56ED68, 0008DC56ED69, 0008DC56ED6D, 0008DC56ED72, 0008DC56ED76, 0008DC56ED77, 0008DC56ED78, 0008DC56ED7A, 0008DC56ED7B, 0008DC56ED7C, 0008DC56ED7D, 0008DC56ED80, 0008DC56ED81, 0008DC56ED82, 0008DC56ED83, 0008DC56ED84, 0008DC56ED85, 0008DC56ED86, 0008DC56ED87, 0008DC56ED88, 0008DC56ED89, 0008DC56ED8A, 0008DC56ED8B, 0008DC56ED8C, 0008DC56ED8E, 0008DC56ED8F, 0008DC56ED90, 0008DC56ED91, 0008DC56ED92, 0008DC56ED93, 0008DC56ED94, 0008DC56ED98, 0008DC56ED99, 0008DC56ED9A, 0008DC56ED9E, 0008DC56EDA0, 0008DC56EDA1, 0008DC56EE26, 0008DC56EE2A, 0008DC56EE42, 0008DC56EE43, 0008DC56EE44, 0008DC56EE45, 0008DC56EE49, 0008DC56EE4A, 0008DC56EE4B, 0008DC56EE4E, 0008DC56EE4F, 0008DC56EE50, 0008DC56EE51, 0008DC56EE52, 0008DC56EE53, 0008DC56EE54, 0008DC56EE55, 0008DC56EE5A, 0008DC56EE5B, 0008DC56EE5C, 0008DC56EE60, 0008DC56EE61, 0008DC56EE6A, 0008DC56EE6B, 0008DC56EE6C, 0008DC56EE6E, 0008DC56EE6F, 0008DC56EE70, 0008DC56EE71, 0008DC56EE72, 0008DC56EE73, 0008DC56EE74, 0008DC56EE75, 0008DC56EE76, 0008DC56EE77, 0008DC56EE78, 0008DC56EE79, 0008DC56EE7A, 0008DC56EE7C, 0008DC56EE7D, 0008DC56EE7E, 0008DC56EE7F, 0008DC56EE80, 0008DC56EE81, 0008DC56EE82
FEI Number 1929691
Recalling Firm/
Manufacturer
Stanley Security Solutions Inc
4600 Vine St
Lincoln NE 68503-2823
For Additional Information ContactMr. Paul Hall
402-742-9419
Manufacturer Reason
for Recall
STANLEY Healthcare has received reports of outages on the Arial nurse call system. These outages have been traced back to the Arial 54315 Serial-to-IP Network Gateway.
FDA Determined
Cause 2
Under Investigation by firm
ActionThe firm initiated the recall by email on 06/14/2019. The letter requests that all users that receive a Warning flag on an Arial computer immediately contact the Technical Support Department to provide assistance in troubleshooting. The firm will be distributing replacement units
Quantity in Commerce137 units
DistributionNationwide and Canada
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-