Class 2 Device Recall TiLite

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall TiLitesee related information
Date Initiated by FirmSeptember 11, 2018
Create DateAugust 20, 2019
Recall Status1 Terminated 3 on July 29, 2022
Recall NumberZ-2294-2019
Recall Event ID 83360
510(K)NumberK072311 K990358 K023606 
Product Classification Wheelchair, mechanical - Product Code IOR
ProductTiLite X-Short anti-tip tubes, models TRA, AERO T, ZRA, AERO Z, 2GX, AERO X, Twist, and AERO R, a component of the TiLite wheelchair (anti-tip extension for manual wheelchairs). Product Usage: The affected component is part of a manual wheelchair to provide mobility to physically impaired individuals. The affected component is used to prevent the wheelchair from tipping backwards.
Code Information 11908818, 11914822, 11918140, 11922587, 11927950, 11928694, 11931072, 11931597, 11935222, 11936116, 11937341, 11944462, 11946693, 11946914, 11948640, 11958774, 11961303, 11963610, 11971197, 11972416, 11973295, 11973373, 11982499, 11988601, 11992794, 11994940, 11995536, 11995570, 11996713, 11996864, 12005061, M2005478, M2005586, M2019937, M2019958, M2019959, M2019961, M2019968, M2019996, M2021020, M2021023, M2021026, M2021029, M2021030, M2021038, M2021044, M2021048, M2021049, M2021063, M2021069, M2021077, M2021078, M2021097, M2021105, M2021115, M2021117, M2021120, M2021124, M2021130, M2021144, M2021146, M2021150, M2021155, M2021166, M2021170, M2021173, M2021180, M2021181, M2021186, M2021187, M2021194, M2021215, M2021219, M2021229, M2021230, M2021231, M2021233, M2021237, M2021244, M2021246, M2021259, M2021271, M2021272, M2021280, M2021281, M2021286, M2021287, M2021300, M2021301, M2021302, M2021304, M2021307, M2021328, M2021348, M2021351, M2021362, M2021366, M2021367, M2021370, M2021371, M2021394, M2021399, M2021401, M2021404, M2021410, M2021414, M2021420, M2021422, M2021423, M2021430, M2021435, M2021449, M2021453, M2021464, M2021468, M2021477, M2021480, M2021499, M2021510, M2021514, M2021516, M2021517, M2021519, M2021524, M2021538, M2021539, M2021562, M2021567, M2021571, M2021576, M2021583, M2021587, M2021606, M2021619, M2021620, M2021625, M2021640, M2021641, M2021663, M2021679, M2021695, M2021699, M2021704, M2021705, M2021706, M2021719, M2021721, M2021728, M2021731, M2021740, M2021745, M2021756, M2021757, M2021771, M2021775, M2021776, M2021794, M2021796, M2021800, M2021809, M2021812, M2021814, M2021819, M2021835, M2021837, M2021838, M2021840, M2021848, M2021854, M2021860, M2021882, M2021884, M2021888, M2021894, M2021900, M2021905, M2021911, M2025005, M2025017, M2025020, M2025025, M2025030, M2025056, M2025057, M2025063, M2025070, M2025075, M2025077, M2025083, M2025088, M2025090, M2025097, M2025101, M2025114, M2025132, M2025135, M2025136, M2025139, M2025145, M2025157, M2025170, M2025175, M2025176, M2025180, M2025195, M2025223, M2025243, M2025246, M2025255, M2025259, M2025262, M2025265, M2025271, M2025279, M2025316, M2025324, M2025326, M2025334, M2025337, M2025340, M2025342, M2025351, M2025352, M2025357, M2025365, M2025390, M2025394, M2025396, M2025401, M2025406, M2025418, M2025420, M2025427, M2025431, M2025442, M2025443, M2025452, M2025466, M2025473, M2025485, M2025487, M2025490, M2025496, M2025505, M2025519, M2025523, M2025531, M2025546, M2025560, M2025564, M2025571, M2025578, M2025579, M2025581, M2025596, M2025599, M2025607, M2025617, M2025621, M2025628, M2025633, M2025637, M2025643, M2025648, M2025663, M2025685, M2025691, M2025700, M2025706, M2025728, M2025729, M2025744, M2025749, M2025759, M2025762, M2025770, M2025787, M2025793, M2025802, M2025809, M2025814, M2025815, M2025821, M2025834, M2025835, M2025843, M2025845, M2025846, M2025848, M2025863, M2025867, M2025880, M2025886, M2025890, M2025911, M2025915, M2025917, M2025926, M2025956, M2025969, M2025972, M2025999, M2026002, M2026003, M2026004, M2026005, M2026034, M2026035, M2026037, M2026045, M2026053, M2026058, M2026109, M2026112, M2026125, M2026146, M2026156, M2026162, M2026186, M2026187, M2026208, M2026225, M2026232, M2026237, M2026246, M2026254, M2026268, M2026273, M2026274, M2026276, M2026279, M2026280, M2026284, M2026298, M2026303, M2026307, M2026312, M2026313, M2026320, M2026329, M2026330, M2026332, M2026341, M2026353, M2026362, M2026374, M2026377, M2026379, M2026397, M2026400, M2026421, M2026426, M2026431, M2026437, M2026440, M2026441, M2026445, M2026468, M2026475, M2026485, M2026487, M2026489, M2026490, M2026528, M2026530, M2026531, M2026535, M2026536, M2026537, M2026539, M2026542, M2026544, M2026545, M2026546, M2026555, M2026568, M2026570, M2026577, M2026578, M2026580, M2026586, M2026587, M2026610, M2026629, M2026630, M2026633, M2026641, M2026649, M2026658, M2026679, M2026715, M2026716, M2026720, M2026737, M2026741, M2026748, M2026749, M2026751, M2026759, M2026765.M2026785, M2026790, M2026806, M2026811, M2026812, M2026813, M2026814, M2026825, M2026837, M2026851, M2026855, M2026856, M2026859, M2026862, M2026880, M2026899, M2026903, M2026914, M2026925, M2026926, M2026937, M2026949, M2026952, M2026955, M2026965, M2026980, M2026984, M2026985, M2026991, M2026995, M2026997, M2027027, M2027040, M2027050, M2027056, M2027057, M2027058, M2027061, M2027063, M2027065, M2027066, M2027080, M2027081, M2027088, M2027092, M2027097, M2027100, M2027105, M2027113, M2027122, M2027124, M2027131, M2027132, M2027140, M2027144, M2027159, M2027161, M2027165, M2027175, M2027176, M2027178, M2027190, M2027202, M2027214, M2027218, M2027221, M2027228, M2027236, M2027240, M2027257, M2027261, M2027263, M2027283, M2027286, M2027305, M2027308, M2027359, M2027361, M2027368, M2027371, M2027380, M2027391, M2027392, M2027395, M2027396, M2027416, M2027427, M2027432, M2027439, M2027442, M2027456, M2027463, M2027465, M2027470, M2027483, M2027487, M2027503, M2027510, M2027519, M2027534, M2027539, M2027543, M2027552, M2027558, M2027563, M2027575, M2027583, M2027587, M2027595, M2027609, M2027616, M2027623, M2027626, M2027627, M2027651, M2027655, M2027661, M2027663, M2027664, M2027668, M2027672, M2027677, M2027679, M2027681, M2027683, M2027691, M2027712, M2027719, M2027720, M2027731, M2027738, M2027740, M2027743, M2027744, M2027757, M2027801, M2027812, M2027814, M2027820, M2027834, M2027839, M2027852, M2027854, M2027858, M2027875, M2027878, M2027887, M2027893, M2027895, M2027896, M2027898, M2027900, M2027903, M2027912, M2027919, M2027920, M2027923, M2027933, M2027934, M2027937, M2027945, M2027949, M2027965, M2027967, M2027969, M2027970, M2027986, M2027988, M2027995, M2027996, M2027999, M2028027, M2028032, M2028045, M2028087, M2028095, M2028097, M2028108, M2028109, M2028126, M2028127, M2028128, M2028135, M2028142, M2028149, M2028153, M2028156, M2028163, M2028174, M2028186, M2028187, M2028200, M2028208, M2028210, M2028211, M2028213, M2028217, M2028224, M2028231, M2028238, M2028240, M2028245, M2028249, M2028254, M2028264, M2028276, M2028279, M2028280, M2028293, M2028297, M2028299, M2028318, M2028326, M2028329, M2028347, M2028350, M2028358, M2028362, M2028369, M2028376, M2028388, M2028389, M2028403, M2028407, M2028412, M2028416, M2028422, M2028429, M2028457, M2028481, M2028482, M2028483, M2028487, M2028498, M2028501, M2028504, M2028507, M2028508, M2028509, M2028514, M2028531, M2028545, M2028547, M2028564, M2028614, M2028618, M2028625, M2028631, M2028637, M2028649, M2028651, M2028653, M2028674, M2028691, M2028703, M2028705, M2028707, M2028711, M2028712, M2028714, M2028719, M2028746, M2028747, M2028751, M2028766, M2028770, M2028779, M2028787, M2028789, M2028794, M2028810, M2028821, M2028843, M2028850, M2028851, M2028853, M2028858, M2028864, M2028867, M2028871, M2028887, M2028910, M2028912, M2028917, M2028920, M2028922, M2028931, M2028941, M2028943, M2028946, M2028947, M2028953, M2028965, M2028978, M2028984, M2028994, M2029004, M2029008, M2029012, M2029014, M2029015, M2029029, M2029042, M2029043, M2029044, M2029050, M2029054, M2029055, M2029072, M2029077, M2029083, M2029091, M2029092, M2029094, M2029105, M2029144, M2029152, M2029161, M2029171, M2029172, M2029174, M2029182, M2029188, M2029195, M2029196, M2029206, M2029209, M2029215, M2029218, M2029238, M2029246, M2029250, M2029256, M2029262, M2029281, M2029291, M2029303, M2029314, M2029326, M2029336, M2029347, M2029367, M2029372, M2029374, M2029379, M2029381, M2029385, M2029393, M2029402, M2029405, M2029408, M2029412, M2029425, M2029429, M2029432, M2029433, M2029442, M2029450, M2029452, M2029472, M2029474, M2029475, M2029479, M2029488, M2029496, M2029497, M2029499, M2029500, M2029506, M2029515, M2029517, M2029520, M2029521, M2029536, M2029540, M2029541, M2029549, M2029556, M2029569, M2029581, M2029611, M2029647, M2029669, M2029680, M2029683, M2029684, M2029702, M2029729, M2029740, M2029745, M2029753, M2029771, M2029772, M2029778, M2029801, M2029829, M2029850, M2029867, M2029877, M2029885.M2029889, M2029892, M2029900, M2029902, M2029915, M2029919, M2029922, M2029933, M2029962, M2029964, M2029971, M2029976, M2029989, M2030002, M2030003, M2030007, M2030011, M2030012, M2030027, M2030031, M2030037, M2030042, M2030052, M2030053, M2030078, M2030088, M2030100, M2030143, M2030145, M2030157, M2030160, M2030162, M2030163, M2030166, M2030168, M2030173, M2030176, M2030196, M2030201, M2030204, M2030206, M2030207, M2030209, M2030215, M2030217, M2030243, M2030244, M2030252, M2030255, M2030264, M2030284, M2030290, M2030291, M2030306, M2030308, M2030309, M2030330, M2030344, M2030346, M2030358, M2030362, M2030369, M2030371, M2030373, M2030380, M2030384, M2030387, M2030396, M2030397, M2030402, M2030403, M2030408, M2030418, M2030419, M2030427, M2030428, M2030435, M2030436, M2030443, M2030449, M2030452, M2030453, M2030461, M2030464, M2030470, M2030483, M2030489, M2030491, M2030504, M2030511, M2030514, M2030519, M2030542, M2030547, M2030560, M2030578, M2030599, M2030610, M2030613, M2030616, M2030617, M2030629, M2030638, M2030665, M2030666, M2030671, M2030672, M2030677, M2030685, M2030694, M2030706, M2030709, M2030718, M2030759, M2030763, M2030764, M2030781, M2030783, M2030788, M2030791, M2030792, M2030804, M2030811, M2030822, M2030832, M2030836, M2030838, M2030859, M2030869, M2030897, M2030899, M2030913, M2030914, M2030924, M2030934, M2030936, M2030942, M2030949, M2030950, M2030951, M2030962, M2030975, M2030979, M2030983, M2030987, M2030991, M2030998, M2031500, M2031513, M2031519, M2031522, M2031523, M2031531, M2031532, M2031534, M2031535, M2031544, M2031548, M2031552, M2031553, M2031555, M2031557, M2031563, M2031567, M2031579, M2031587, M2031592, M2031601, M2031602, M2031609, M2031610, M2031618, M2031623, M2031629, M2031641, M2031650, M2031659, M2031665, M2031683, M2031705, M2031711, M2031732, M2031734, M2031735, M2031745, M2031748, M2031751, M2031755, M2031763, M2031769, M2031770, M2031779, M2031780, M2031805, M2031809, M2031826, M2031837, M2031838, M2031858, M2031860, M2031862, M2031875, M2031881, M2031887, M2031898, M2031908, M2031917, M2031921, M2031923, M2031930, M2031933, M2031944, M2031946, M2031952, M2031957, M2031959, M2031961, M2031973, M2032009, M2032026, M2032036, M2032039, M2032042, M2032046, M2032053, M2032059, M2032074, M2032078, M2032083, M2032086, M2032111, M2032112, M2032122, M2032124, M2032135, M2032140, M2032141, M2032144, M2032145, M2032147, M2032149, M2032156, M2032157, M2032163, M2032165, M2032181, M2032193, M2032196, M2032204, M2032235, M2032238, M2032250, M2032255, M2032259, M2032264, M2032265, M2032281, M2032286, M2032287, M2032303, M2032307, M2032309, M2032315, M2032317, M2032325, M2032327, M2032358, M2032360, M2032374, M2032385, M2032386, M2032398, M2032409, M2032421, M2032442, M2032443, M2032452, M2032453, M2032474, M2032479, M2032502, M2032547, M2032550, M2032553, M2032558, M2032560, M2032562, M2032563, M2032569, M2032570, M2032571, M2032574, M2032585, M2032632, M2032633, M2032637, M2032644, M2032645, M2032665, M2032668, M2032699, M2032701, M2032709, M2032716, M2032717, M2032723, M2032730, M2032733, M2032738, M2032744, M2032758, M2032763, M2032767, M2032770, M2032771, M2032820, M2032821, M2032839, M2032842, M2032844, M2032849, M2032854, M2032861, M2032885, M2032891, M2032893, M2032920, M2032923, M2032930, M2032931, M2032940, M2032941, M2032942, M2032944, M2032946, M2032954, M2032955, M2032977, M2032978, M2032985, M2032986, M2032989, M2032996, M2033004, M2033006, M2033007, M2033012, M2033018, M2033020, M2033021, M2033022, M2033027, M2033028, M2033034, M2033035, M2033049, M2033056, M2033062, M2033063, M2033067, M2033068, M2033072, M2033080, M2033081, M2033085, M2033087, M2033097, M2033101, M2033117, M2033119, M2033123, M2033124, M2033129, M2033135, M2033138, M2033139, M2033143, M2033149, M2033153, M2033158, M2033161, M2033162, M2033170, M2033178, M2033180, M2033182, M2033188, M2033194, M2033203, M2033204, M2033207, M2033216, M2033221, M2033227, M2033232, M2033236, M2033257, M2033258, M2033260, M2033263.M2033265, M2033266, M2033267, M2033280, M2033284, M2033294, M2033296, M2033297, M2033299, M2033302, M2033305, M2033306, M2033309, M2033314, M2033324, M2033326, M2033328, M2033333, M2033337, M2033339, M2033351, M2033353, M2033354, M2033355, M2033358, M2033364, M2033367, M2033372, M2033373, M2033382, M2033389, M2033395, M2033397, M2033399, M2033405, M2033407, M2033415, M2033449, M2033457, M2033463, M2033466, M2033471, M2033476, M2033479, M2033480, M2033488, M2033490, M2033491, M2033503, M2033512, M2033518, M2033519, M2033520, M2033526, M2033537, M2033573, M2033578, M2033586, M2033589, M2033591, M2033594, M2033604, M2033609, M2033612, M2033613, M2033623, M2033633, M2033642, M2033644, M2033647, M2033648, M2033663, M2033668, M2033671, M2033677, M2033679, M2033680, M2033690, M2033714, M2033715, M2033716, M2033729, M2033736, M2033739, M2033746, M2033782, M2033787, M2033788, M2033793, M2033794, M2033811, M2033816, M2033818, M2033821, M2033836, M2033841, M2033854, M2033876, M2033881, M2033886, M2033889, M2033900, M2033914, M2033915, M2033916, M2033917, M2033918, M2033945, M2033959, M2033973, M2033986, M2033987, M2033992, M2033996, M2034000, M2034010, M2034025, M2034026, M2034029, M2034037, M2034045, M2034054, M2034060, M2034061, M2034064, M2034065, M2034069, M2034070, M2034073, M2034074, M2034075, M2034082, M2034096, M2034114, M2034126, M2034141, M2034155, M2034162, M2034163, M2034171, M2034174, M2034185, M2034193, M2034194, M2034202, M2034216, M2034227, M2034234, M2034248, M2034258, M2034262, M2034272, M2034273, M2034276, M2034277, M2034278, M2034281, M2034316, M2034323, M2034325, M2034328, M2034335, M2034349, M2034351, M2034365, M2034368, M2034371, M2034375, M2034386, M2034388, M2034392, M2034412, M2034416, M2034424, M2034427, M2034437, M2034446, M2034449, M2034451, M2034454, M2034456, M2034459, M2034461, M2034471, M2034477, M2034486, M2034493, M2034494, M2034497, M2034503, M2034506, M2034520, M2034521, M2034533, M2034536, M2034541, M2034550, M2034556, M2034559, M2034578, M2034584, M2034586, M2034587, M2034593, M2034602, M2034605, M2034608, M2034609, M2034615, M2034617, M2034619, M2034631, M2034639, M2034648, M2034661, M2034674, M2034686, M2034704, M2034705, M2034706, M2034707, M2034716, M2034729, M2034737, M2034745, M2034752, M2034754, M2034755, M2034759, M2034760, M2034764, M2034778, M2034779, M2034786, M2034805, M2034808, M2034813, M2034819, M2034820, M2034822, M2034823, M2034847, M2034856, M2034867, M2034870, M2034874, M2034877, M2034880, M2034881, M2034884, M2034886, and M2034907.  
FEI Number 3002653226
Recalling Firm/
Manufacturer
TiSport, Llc
2701 W Court St
Pasco WA 99301-3911
For Additional Information ContactJamie Crawford
615-975-7463
Manufacturer Reason
for Recall
The anti-tip tubes potentially fail in certain situations and the wheelchairs can tip backwards.
FDA Determined
Cause 2
Component design/selection
ActionThe recalling firm issued distributor letters dated 9/7/2018 via certified mail beginning on 9/11/2018. The letter identified the affected product, problem and actions to be taken.
Quantity in Commerce1,452 wheelchairs
DistributionDistribution was nationwide, including Puerto Rico. There was also government distribution but no military distribution. Foreign distribution was made to Canada, Australia, Brazil, China, Italy, Saudi Arabia, Switzerland, United Kingdom, and Uruguay.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = IOR
510(K)s with Product Code = IOR
-
-