| | Class 2 Device Recall Artis zee/zeego & Artis Q/Q.zen |  |
| Date Initiated by Firm | July 30, 2019 |
| Create Date | August 24, 2019 |
| Recall Status1 |
Terminated 3 on April 29, 2020 |
| Recall Number | Z-2367-2019 |
| Recall Event ID |
83565 |
| 510(K)Number | K181407 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | Artis zee and Artis Q/ Q.zen Interventional Fluoroscopic X-Ray System with patient tables as follows:
Systems Material Number
Artis zee biplane 10094141;
Artis Q floor 10848280;
Artis Q biplane 10848282;
Artis Q.zen floor 10848353;
Artis Q.zen biplane 10848355;
Artis zee ceiling 10094137;
Artis Q ceiling 10848281;
Artis Q zeego 10848283
ArtisQ.zen ceiling 10848354
Product Usage:
Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. The Artis family can also support the acquisition of position triggered imaging for spatial data synthesis. |
| Code Information |
Serial Numbers: 123245 103302 121501 148401 109589 148393 111129 148400 103310 148397 155235 121505 109563 121506 131235 121502 105125 |
| FEI Number |
2240869
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc
40 Liberty Blvd
Malvern
PA
19355-1418
|
| For Additional Information Contact | SAME 610-219-4834 |
Manufacturer Reason for Recall | Artis zee/Q systems patient tables may potentially have cracks in the table mainframe, and cause mechanical detachment of the upper part of the patient bed and lead to hazardous situations for patients and medical personnel due to falling parts |
FDA Determined Cause 2 | Process control |
| Action | Siemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice (AX011/19/S) on 8/1/19 a . The letter identifies the reason for recall, health risk and action to take: Siemens Service organization will inspect the table mainframe for possible cracks and will contact you shortly to arrange a date to perform this corrective action. Contact the service organization for an earlier appointment at 1-800-888-7436. |
| Quantity in Commerce | US 17 |
| Distribution | US Nationwide Distribution - AL, AR, KS, MI, MN, NC, ND, NY, OH, SC, TX,
WI |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|