| | Class 2 Device Recall TherMax Blood Warmer Unit |  |
| Date Initiated by Firm | September 19, 2019 |
| Date Posted | October 30, 2019 |
| Recall Status1 |
Terminated 3 on December 11, 2020 |
| Recall Number | Z-0247-2020 |
| Recall Event ID |
83816 |
| 510(K)Number | K190910 |
| Product Classification |
Dialyzer, high permeability with or without sealed dialysate system - Product Code KDI
|
| Product | TherMax Blood Warmer Unit |
| Code Information |
Product Code: 955630, UDI: 00085412639505, Serial Numbers: 101332, 101335, 101358, 101369, 101402, 101403, 101404, 101406, 101415, 101416, 101429, 101430, 101437, 101439, 101440, 101441, 101443. Product Code: 955515, Serial Numbers 100261, 100263, 100280, 100281, 100282, 100283, 100284, 100285, 100286, 100287, 100288, 100289, 100290, 100291, 100292, 100294, 100296, 100298, 100299, 100306, 100307, 100308, 100309, 100310, 100312, 100313, 100314, 100315, 100321, 100322, 100323, 100324, 100325, 100326, 100327, 100328, 100330, 100341, 100342, 100343, 100344, 100345, 100346, 100347, 100348, 100349, 100356, 100357, 100358, 100359, 100360, 100361, 100362, 100363, 100364, 100365, 100366, 100367, 100368, 100369, 100370, 100391, 100392, 100393, 100395, 100396, 100397, 100398, 100399, 100400, 100401, 100402, 100403, 100404, 100405, 100416, 100417, 100418, 100419, 100420, 100421, 100422, 100423, 100424, 100427, 100428, 100429, 100446, 100450, 100453, 100454, 100455, 100456, 100458, 100459, 100460, 100486, 100487, 100489, 100490, 100496, 100497, 100498, 100499, 100500, 100501, 100502, 100503, 100504, 100505, 100532, 100533, 100535, 100537, 100538, 100539, 100540, 100541, 100545, 100548, 100550, 100551, 100552, 100555, 100581, 100582, 100583, 100584, 100585, 100586, 100587, 100590, 100591, 100595, 100596, 100599, 100601, 100602, 100603, 100605, 100631, 100632, 100633, 100639, 100640, 100641, 100642, 100643, 100644, 100645, 100646, 100649, 100650, 100652, 100653, 100654, 100655, 100756, 100757, 100758, 100759, 100761, 100762, 100763, 100764, 100765, 100766, 100767, 100768, 100769, 100770, 100771, 100772, 100773, 100774, 100775, 100776, 100777, 100778, 100779, 100780, 100781, 100782, 100783, 100784, 100785, 100786, 100787, 100788, 100789, 100790, 100791, 100793, 100794, 100795, 100796, 100798, 100799, 100800, 100801, 100802, 100803, 100804, 100805, 100806, 100807, 100808, 100810, 100811, 100812, 100813, 100814, 100815, 100816, 100817, 100818, 100819, 100820, 100821, 100822, 100823, 100825, 100826, 100827, 100829, 100830, 100856, 100857, 100858, 100859, 100860, 100861, 100863, 100864, 100865, 100868, 100869, 100871, 100872, 100874, 100875, 100878, 100879, 100880, 100881, 100882, 100883, 100884, 100885, 100886, 100887, 100888, 100889, 100890, 100891, 100892, 100893, 100894, 100895, 100896, 100897, 100898, 100899, 100900, 100931, 100932, 100933, 100934, 100936, 100938, 100939, 100940, 100941, 100942, 100943, 100945, 100946, 100947, 100948, 100949, 100950, 100952, 100954, 100955, 100981, 100983, 100984, 100985, 100986, 100987, 100988, 100989, 100990, 100991, 100992, 100997, 100998, 100999, 101000, 101001, 101002, 101003, 101004, 101005, 101006, 101007, 101008, 101009, 101010, 101011, 101012, 101013, 101014, 101015, 101016, 101017, 101018, 101019, 101020, 101021, 101022, 101024, 101026, 101027, 101030, 101081, 101082, 101083, 101084, 101085, 101086, 101088, 101089, 101090, 101091, 101092, 101094, 101095, 101096, 101097, 101098, 101099, 101101, 101103, 101104, 101105, 101106, 101107, 101108, 101109, 101110, 101111, 101112, 101115, 101121, 101122, 101123, 101124, 101126, 101127, 101172, 101173, 101174, 101175, 101176, 101177, 101179, 101180, 101181, 101182, 101183, 101184, 101185, 101186, 101187, 101188, 101189, 101190, 101192, 101229, 101230, 101231, 101233, 101235, 101236, 101237, 101241, 101246, 101247, 101248, 101249, 101250, 101251, 101252, 101253, 101254, 101255, 101256, 101257, 101258, 101259, 101271, 101272, 101273, 101274, 101276, 101277, 101278, 101458, 101460, 101469, 101470, 101471, 101472, 101473, 101474, 101475, 101476, 101477, 101478, 101479, 101480, 101481, 101482, 101483, 101484, 101485, 101486, 101487, 101488, 101489, 101490, 101491, 101492, 101508, 101509, 101510, 101511, 101512, 101513, 101514, 101515, 101516, 101517, 101519, 101520, 101521, 101522, 101538, 101539, 101540, 101541, 101542, 101543, 101544, 101545, 101546, 101547, 101548, 101549, 101550, 101551, 101552, 101553, 101554, 101555, 101556, 101557, 101558, 101559, 101560, 101561, 101562, 101593, 101594, 101595, 101596, 101597, 101598, 101599, 101600, 101601, 101602, 101603, 101604, 101605, 101607. |
| FEI Number |
1417572
|
Recalling Firm/ Manufacturer |
Baxter Healthcare Corporation 1 Baxter Pkwy Deerfield IL 60015-4625
|
Manufacturer Reason for Recall | TherMax Blood Warmers may not be in compliance with an electrical safety standard, which requires a protective earth
connection that can sustain a current of 25 amps for 10 seconds. |
FDA Determined Cause 2 | Device Design |
| Action | Baxter Healthcare notified customers on about 09/19/2019, via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that specific TherMax Blood Warmers listed below may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds. The affected units may not have an adequate ground path, which could cause the units to become damaged if used with improperly grounded outlets. The damage could cause the TherMax Blood Warmer units to be nonfunctional during clinical use.
Instructions included that customers could continue to safely use affected devices with a properly grounded outlet, that a local Baxter representative will contact each facility to arrange for the correction of affected devices, to complete and return the Baxter Customer Reply Form via fax to 224-270-5457 or scanning and e-mailing it to fca@baxter.com, and to notify customers if the product was further distributed.
For general questions regarding this communication, contact Baxter Corporate Product Surveillance at 800-437-5176, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday. |
| Quantity in Commerce | 504 units total |
| Distribution | Nationwide distribution to Georgia and Maryland.
International distribution to Belgium, Germany, Finland, France, Greece, Italy, Luxemburg, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom, Australia, and New Zealand |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|