| | Class 2 Device Recall Bigliani Flatow Shoulder System/Bigliani Flatow The Complete Shoulder Solution |  |
| Date Initiated by Firm | October 10, 2019 |
| Create Date | November 07, 2019 |
| Recall Status1 |
Terminated 3 on May 19, 2020 |
| Recall Number | Z-0298-2020 |
| Recall Event ID |
83962 |
| 510(K)Number | K982981 |
| Product Classification |
Prosthesis, shoulder, non-constrained, metal/polymer cemented - Product Code KWT
|
| Product | BF Humeral Stem, Various Sizes, Item Nos. 00430000606
00430000613
00430000706
00430000713
00430000813
00430000817
00430000820
00430000913
00430001013
00430001017
00430001020
00430001113
00430001213
00430001217
00430001220
00430001313
00430001413
00430001417
00430001420
00430001513
00430001613
00430001713
00430001813
47430103100 |
| Code Information |
All product manufactured prior to January 2014. |
| FEI Number |
1000220733
|
Recalling Firm/ Manufacturer |
Zimmer Biomet, Inc.
1800 W Center St
Warsaw
IN
46580-2304
|
| For Additional Information Contact | Zimmer Biomet Customer Service 574-371-3071 |
Manufacturer Reason for Recall | Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). |
FDA Determined Cause 2 | Packaging change control |
| Action | The firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags.
Customers were asked to take the following actions:
1. Review this notification and ensure that pertinent personnel are aware of the contents.
2. Ensure that backup devices for the affected products are available, as per standard orthopedic procedure planning, to reduce the likelihood of an extension of surgery time.
3. Examine devices thoroughly for adherent LDPE bag residue prior to use. Use a backup device if LDPE bag residue is detected.
4. Return any devices which display adherent LDPE bag residue to Zimmer Biomet, consistent with the standard instructions for use regarding the disposition of damaged components.
5. Complete the Certificate of Acknowledgement and return to the firm.
6. Retain a copy of the acknowledgement form with your field action records in the event of a compliance audit of your facility's documentation.
If you have further questions or concerns regarding this recall, please call the firm's customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. |
| Quantity in Commerce | 5,459,583 total devices |
| Distribution | Distributed nationwide. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = KWT
|
|
|
|