Class 1 Device Recall CARESCAPE Respiratory Module EsCOVX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 1 Device Recall CARESCAPE Respiratory Module EsCOVX see related information
Date Initiated by Firm September 27, 2019
Date Posted January 10, 2020
Recall Status1 Terminated 3 on September 26, 2023
Recall Number Z-0755-2020
Recall Event ID 84034
510(K)Number K183394  K171028  K150245  
Product Classification Analyzer, gas, carbon-dioxide, gaseous-phase - Product Code CCK
Product CARESCAPE Respiratory Module E-sCOVX
Code Information Mfg. Lot or Serial # UDI # SGV19276034HA 010084068210428921SGV19276034HA11190702 SGV19266019HA 010084068210428921SGV19266019HA11190626 SGV14404046HA Not Applicable 
Recalling Firm/
Manufacturer
GE Healthcare, LLC
3000 N Grandview Blvd
Waukesha WI 53188-1615
For Additional Information Contact GE Healthcare Service
800-437-1171
Manufacturer Reason
for Recall
Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
FDA Determined
Cause 2
Process control
Action The firm issued the Urgent Medical Device Correction Letter dated 09/27/2019. The letter describes the failure and requests replacement of the affected module with another module. If a replacement is not available, the consignee is directed to follow the instructions below to ensure that your respiratory module O2 reading is within specifications (a) before any new patient case is started and (b) for prolonged patient cases, at a minimum of once a day: 1) Connect the module to the host device; 2) Ensure the module is fully warmed up by having it powered on for minimum 20 minutes; 3) Configure the O2 reading and waveform to the host device screen as defined in the host device instructions; 4) Feed 100% O2 gas flow from the ventilator to the module and observe the O2 reading and waveform; and 5) Ensure the O2 reading is within specifications 100% +/-3% and there is no erratic behavior in the waveform. 6) If you are unable to complete the above steps successfully, remove the respiratory module from clinical use and quarantine. Contact GE Healthcare Service or your local Service Representative. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity in Commerce 3 units
Distribution US OUS: Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Korea (Republic of), Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Peru, Poland, Republic of Serbia, Romania, Russia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uganda, and Unites Kingdom
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = CCK and Original Applicant = GE HEALTHCARE FINLAND OY
-
-