• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Torcon NB Advantage

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Torcon NB Advantagesee related information
Date Initiated by FirmOctober 21, 2019
Create DateNovember 18, 2019
Recall Status1 Terminated 3 on June 30, 2020
Recall NumberZ-0488-2020
Recall Event ID 84125
510(K)NumberK133130 
Product Classification Catheter, intravascular, diagnostic - Product Code DQO
ProductTorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicians trained and experienced in angiographic techniques.
Code Information REF (RPN): HNB5.0-38-70-P-NS-SHK1.0 REF (GPN): G27689 Lot Number: NS9904739 UDI: (01)00827002276895(17)220801(10)NS9904739 
FEI Number 1820334
Recalling Firm/
Manufacturer
Cook Inc.
750 N Daniels Way
Bloomington IN 47404-9120
For Additional Information ContactCook Medical Customer Relations Department
812-339-7361
Manufacturer Reason
for Recall
The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches of Torcon NB Advantage Catheters.
FDA Determined
Cause 2
Process change control
ActionOn October 21, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail instructing them to: 1. Examine inventory immediately to determine if you have affected product(s), and quarantine affected product(s). Immediately cease all distribution and use of this product. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Note: Unaffected products that are returned will not be credited. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA: " Visit http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm to obtain a form to fax or mail. " Call the FDA at 800.FDA.1088. We recognize that this situation is a potential disruption to your normal operations, and we sincerely apologize. Thank you for your immediate assistance in this matter. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235. We look forward to your response.
Quantity in Commerce5
DistributionWorldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, IL, IN, KS, LA, MA, MD, Ml, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV, and PR. Countries of AU, BE, CA, CN, CZ, DK, FR, DE, GT, HK, IT, KR, LU, MX, NL, PK, PL, ZA, ES, SE, CH, TW, and GB.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = DQO
-
-