Class 2 Device Recall BriteBlade Pro

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall BriteBlade Pro see related information
Date Initiated by Firm November 27, 2019
Date Posted January 13, 2020
Recall Status1 Terminated 3 on April 27, 2021
Recall Number Z-0802-2020
Recall Event ID 84367
Product Classification Laryngoscope, rigid - Product Code CCW
Product BriteBlade Pro Single-se Fiber Optic Miller 2, CE 0120, Do not Re-use, Rx Only, REF, 040-722U, Sterile EO, Qty 10, (01) 15055788722783

The device is intended for direct vision laryngoscopy to obtain an illuminated view of the vocal cords and the glottis by insertion into the oral cavity to manipulate the tongue and/or the epiglottis.
Code Information Part Number/Lot Number  040-722U 181000115 I CCW 040-722U 181001520 I CCW 040-722U 181200006 I CCW 040-722U 181202055 I CCW 040-722U 190100006 I CCW 
Recalling Firm/
Manufacturer
Flexicare Medical Ltd.
Cynon Valley Business Park, Mountain Ash
Rhondda Cynon Taff United Kingdom
Manufacturer Reason
for Recall
The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.
FDA Determined
Cause 2
Nonconforming Material/Component
Action On 11/27/2019, the firm "Flexicare" sent a "Urgent: medical Device Recall" letter via USPS Priority mail. The letter informed consignees of the recall. In addition to informing consignees of the recall the letter asked consignees to take the following actions: Complete the acknowledgment and receipt form and return this form to Flexicare Inc. On receipt of the form you will be contacted by a representative of Flexicare Inc. to make all necessary further arrangements. The form asked the following of the consignee: -Acknowledge that you have read and understand the recall instructions provided - If you have experienced any adverse events associated with the recalled products. - Check your stock and quarantined inventory - Send completed response forms marked for the attention of Product Manager Anesthesia to email Ian.ross@flexicare.com or by fax to 949-450-9992 - Identify and notify your customers that were shipped product from the affected lot numbers If you have any questions, please call Product Manager Anesthesia, at 949-450-9999 ext. 102 or email at Ian.Ross@flexicare.com.
Quantity in Commerce 1,024,890 devices
Distribution US: Nationwide OUS: Australia, Canada, China, Czech Republic, Denmark, Finland, Ireland, Italy, Kuwait, Lebanon, Netherlands, Panama, Romania, Russia, South Korea, UAE, UK