• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall BriteBlade Pro

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall BriteBlade Prosee related information
Date Initiated by FirmNovember 27, 2019
Date PostedJanuary 13, 2020
Recall Status1 Terminated 3 on April 27, 2021
Recall NumberZ-0803-2020
Recall Event ID 84367
Product Classification Laryngoscope kit - Product Code OGH
ProductBriteBlade Pro Single-Use Fiber Optic Handle and Blade Mac 3, CE 0120, Do not Re-use, Rx Only, REF, 040-333U, Sterile EO, Qty 10, (01)15055788724138(17)230801(10)180900072 The device is intended for direct vision laryngoscopy to obtain an illuminated view of the vocal cords and the glottis by insertion into the oral cavity to manipulate the tongue and/or the epiglottis.
Code Information Part Number/Lot Number  040-333U 170900328 I OGH 040-333U 180800461 I OGH 040-333U 180900072 I OGH 040-333U 180900104 I OGH 040-333U 181000107 I OGH 040-333U 181100013 I OGH 040-333U 181100097 I OGH 040-333U 181200024 I OGH  
FEI Number 3006363476
Recalling Firm/
Manufacturer
Flexicare Medical Ltd.
Cynon Valley Business Park, Mountain Ash
Rhondda Cynon Taff United Kingdom
Manufacturer Reason
for Recall
The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionOn 11/27/2019, the firm "Flexicare" sent a "Urgent: medical Device Recall" letter via USPS Priority mail. The letter informed consignees of the recall. In addition to informing consignees of the recall the letter asked consignees to take the following actions: Complete the acknowledgment and receipt form and return this form to Flexicare Inc. On receipt of the form you will be contacted by a representative of Flexicare Inc. to make all necessary further arrangements. The form asked the following of the consignee: -Acknowledge that you have read and understand the recall instructions provided - If you have experienced any adverse events associated with the recalled products. - Check your stock and quarantined inventory - Send completed response forms marked for the attention of Product Manager Anesthesia to email Ian.ross@flexicare.com or by fax to 949-450-9992 - Identify and notify your customers that were shipped product from the affected lot numbers If you have any questions, please call Product Manager Anesthesia, at 949-450-9999 ext. 102 or email at Ian.Ross@flexicare.com.
Quantity in Commerce1,024,890 devices
DistributionUS: Nationwide OUS: Australia, Canada, China, Czech Republic, Denmark, Finland, Ireland, Italy, Kuwait, Lebanon, Netherlands, Panama, Romania, Russia, South Korea, UAE, UK
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-