| Date Initiated by Firm | December 11, 2019 |
| Create Date | February 05, 2020 |
| Recall Status1 |
Terminated 3 on September 04, 2020 |
| Recall Number | Z-1071-2020 |
| Recall Event ID |
84530 |
| Product Classification |
Dressing change tray - Product Code OXQ
|
| Product | Centurion- Driveline System Kit
TRIDM250 OUTPATIENT DRIVELINE MGMT SYSTEM |
| Code Information |
TRIDM250 OUTPATIENT DRIVELINE MGMT SYSTEM 2019012890 |
| FEI Number |
3008403546
|
Recalling Firm/ Manufacturer |
Centurion Medical Products Corporation 100 Centurion Way Williamston MI 48895-9086
|
| For Additional Information Contact | SAME 517-546-5400 |
Manufacturer Reason for Recall | Incomplete seals on the sterile package may compromise the sterility |
FDA Determined Cause 2 | Packaging process control |
| Action | Centurion issued a notification emailed to Medline Industries,
Centurion's parent company. Centurion issued mailing letters via First Class Mail beginning on 12/12/2019, letter states reason for recall, health risk and action to take: Please identify all inventory that you have within your possession and segregate it in a location where it cannot be accessed for use. 2) If you have transferred possession of this product to another individual, to another department or to another location in your organization, you are required by FDA regulations to notify them of this recall
communication. Please ask that any affected product be returned to you for collection. 3) Affected product is not to be returned to Centurion, and should be destroyed/discarded. Document the destruction of all product on the enclosed accountability record. 4) Please complete the enclosed accountability record, whether or not you have product, and fax or email at
your earliest convenience to the attention of Lisa Carpenter, Director of Quality Assurance Centurion Medical Products Corporation, (517) 546-5400 Ext.1122. |
| Quantity in Commerce | 100 |
| Distribution | Nationwide
Foreign: GHANA |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|