 Class 2 Device Recallsee related information
Date Initiated by FirmJanuary 15, 2020
Date PostedJanuary 24, 2020
Recall Status1 Terminated 3 on March 17, 2023
Recall NumberZ-1173-2020
Recall Event ID 84726
Product Classification Anesthesia Conduction Kit - Product Code CAZ
ProductEPIDURAL CATHETERIZATION KIT, Product Code ALZANO-05400-B
Code Information Lots: 71F19B0681 71F19C0035 71F19D2827 71F19F2028 71F19H1633 71F19K1574
FEI Number 3015859709
Recalling Firm/
Manufacturer
Arrow International Inc
2400 Bernville Rd
Reading PA 19605-9607
For Additional Information ContactCustomer Service
866-396-2111
Manufacturer Reason
for Recall
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
FDA Determined
Cause 2
Other
ActionAn Urgent Medical Device Recall notification letter dated 1/15/20 was sent to customers. Our records indicate you have received products that are subject to this field action. We are now notifying our customers to take the following actions: 1. If you have affected stock, immediately identify product, discontinue use and quarantine any products with the product codes and lot numbers listed above. 2. To acknowledge receipt of this Urgent Medical Device Recall Notice or to return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document your receipt of this letter.
Quantity in Commerce343,105 total
DistributionThe products were distributed to the following US states: AK, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-