• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 3 Device Recall Theralase

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 3 Device Recall Theralasesee related information
Date Initiated by FirmSeptember 26, 2019
Date PostedFebruary 13, 2020
Recall Status1 Terminated 3 on June 04, 2020
Recall NumberZ-1354-2020
Recall Event ID 84843
510(K)NumberK151816 
Product Classification Powered Light Based Laser Non-Thermal Instrument with Non-Heating Effect For Adjunctive Use in Pain Therapy - Product Code NHN
ProductTLC-2000 Therapeutic Medical Laser System
Code Information Serial Numbers: 3245 to 3336 inclusive
FEI Number 3003614490
Recalling Firm/
Manufacturer
Theralase Inc.
41 Hollinger Rd
East York Canada
For Additional Information ContactRishi Mehta
866-843-5273 Ext. 306
Manufacturer Reason
for Recall
905 nm laser diodes may be below rated power output.
FDA Determined
Cause 2
Device Design
ActionEmail was sent out to healthcare practitioners regarding the recall on September 26, 2019. .Healthcare practitioners are instructed to courier their TLC-2000 to Theralase for analysis. Their 905 nm laser diodes will be analyzed and recalibrated, as required, to meet rated power output.
Quantity in Commerce92 units
DistributionUSA; Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = NHN
-
-