| Date Initiated by Firm | February 14, 2020 |
| Date Posted | March 13, 2020 |
| Recall Status1 |
Terminated 3 on April 20, 2021 |
| Recall Number | Z-1496-2020 |
| Recall Event ID |
85071 |
| 510(K)Number | K181407 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | Artis Q.zen Biplane, Model No. 10848355
Artis Q.zen Ceiling, Model No. 10848354
Artis Q.zen Floor, Model No. 10848353
Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. |
| Code Information |
Serial Numbers 123225 105120 123026 105028 111005 111122 111124 111125 123236 111002 123053 123050 123008 123035 123003 123032 123208 123209 123231 123206 123207 123240 105025 111121 123219 111115 111116 111117 111118 111119 111120 123229 123205 123224 123232 123233 123234 123027 111107 123055 123056 105017 123010 123046 111000 111106 123052 111123 111128 123200 123015 123016 105112 ***Added 3/30/2020*** 123031 123022 |
| FEI Number |
2240869
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc 40 Liberty Blvd Malvern PA 19355-1418
|
| For Additional Information Contact | Siemens Customer Service 800-888-7436 |
Manufacturer Reason for Recall | A software issue could potentially cause the stand and table movements to be blocked. |
FDA Determined Cause 2 | Software design |
| Action | On February 14, 2020, the firm distributed Urgent Medical Device Correction letters to customers. Customers were informed of the software error which affects the listed systems. In case of software failure, planned procedures may have to be terminated and performed on an alternative x-ray system. Customers were instructed to do the following:
Issue # 1: Potential issue with the Message "Stand battery charging: DO NOT POWER OFF" on Artis zeego systems
Customer Action: Please check the system messages before power off/shutdown of the system. If possible, do not switch off the system while the message "Stand battery charging: DO NOT POWER OFF" is displayed and wait until the battery has been charged or replaced.
Issue # 2: Potential issue of blocked stand movements on Artis zeego systems
Issue # 3: Potential issue of blocked table movements on Artis zee/ Q/ Q.zen systems with Siemens Healthineers tables
For Issues #2-3, customers may continue to use their systems, but in case the problem occurs, customers are instructed to follow standard emergency procedures in place in the event of system failures. Customers are encouraged to ensure that patient treatment can be continued in other ways if there is any possible danger for the safety of the patient.
Siemens will correct the software error via a software update. Siemens' service organization will contact customers to arrange a date to perform this corrective action. For an earlier appointment, please contact the firm's service organization at 1-800-888-7436. |
| Quantity in Commerce | 1452 worldwide (638 US) |
| Distribution | US Nationwide distribuiton. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|