Class 2 Device Recall Canon Aquilion One, Model TSX305A/3 V8.3 with FIRST 2.1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Canon Aquilion One, Model TSX305A/3 V8.3 with FIRST 2.1see related information
Date Initiated by FirmMarch 25, 2019
Date PostedApril 10, 2020
Recall Status1 Open3, Classified
Recall NumberZ-1726-2020
Recall Event ID 84954
510(K)NumberK170177 
Product Classification System, x-ray, tomography, computed - Product Code JAK
ProductCanon Aquilion One, Model TSX-305A/3 V8.3 with FIRST 2.1
Code Information Canon Aquilion ONE GENESIS Edition, Whole-Body X-Ray CT Scanner, Model TSX-305A - Serial number 3LA1652001, 3LA17Y2020, 3LB1872025, 3LA1762010, 3LA1772012, 3LA1762009, 3LB1882026, 3LA1752007, 3LA1762008, 3LA1772013, 3LA1732006, 3LA1772015, 3LA1772014, 3LA16Y2003, 3LA1792017, 3LA1782016, 3LA1712005, 3LB1832021, 3LA16Z2004, 3LA17Y2019, 3LB1842022, 3LA1772011, 3LB1852024, 3LA17X2018, 3OB1852023, 6LB18Y2013, 6LB1892008, 6LB1862002, 6LB1882007, 6LB1852001, 6LB1892009, 6LB18Y2011, 6LB1872005, 6LB1872006, and 6OB1862003.
Recalling Firm/
Manufacturer
Canon Medical System, USA, INC.
2441 Michelle Dr
Tustin CA 92780-7047
For Additional Information ContactMs. Terry Schultz
714-669-1421
Manufacturer Reason
for Recall
Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.
FDA Determined
Cause 2
Software design
ActionRecall letters dated 3/25/2019 were issued on 3/25/2019 via USPS return receipt mail.
Quantity in Commerce35 devices
DistributionDistribution was made to AK, AR, CA, FL, IA, ID, IL, KS, MA, MD, MI, MN, MO, MT, NC, ND, NH, NY, OH, OK, OR, PA, TN, TX, VT, WA, WI, and WV. There was government and military distribution but no foreign distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = JAK
-
-