| Class 2 Device Recall Zenition 50 | |
Date Initiated by Firm | April 10, 2020 |
Create Date | May 08, 2020 |
Recall Status1 |
Terminated 3 on January 11, 2021 |
Recall Number | Z-1932-2020 |
Recall Event ID |
85473 |
510(K)Number | K183101 |
Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
Product | Philips Zenition 50, System code 718096 - Product Usage: The devices are used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device. |
Code Information |
Serial Numbers 17 7 |
Recalling Firm/ Manufacturer |
Philips North America, LLC 3000 Minuteman Rd Andover MA 01810-1032
|
For Additional Information Contact | Technical Support 800-722-9377 |
Manufacturer Reason for Recall | When strain relief is lost at the stand side of the Stand MVS Trolley Cable, over time one or more electrical contacts could break due to pulling, twisting, or dropping of the connector. If this happens, live parts may be exposed (230 V) and/or the video signal can be lost. |
FDA Determined Cause 2 | Device Design |
Action | On April 10, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed of the product issue.
Customers were asked to do the following:
- Ensure that Cabling Check and Connectors Check are performed daily, as indicated in the User Routine Checks Program (section 7.4 of the Instructions for Use provided with the system).
- In case the strain relief is lost and/or damage to the cable/connector is identified; stop using the system and contact your Philips local representative so that the cable can be replaced.
- Please ensure that all staff working with the system are informed of the content of this safety notice and place a copy of it with the Instructions for Use.
Philips will be replacing the Stand MVS Trolley Cable (459800943703) in all affected
systems by a new Stand MVS Trolley Cable that has an improved strain relief design.
This action will start by end of April 2020. You will be contacted by a local Philips
representative to schedule an appointment to replace the cable.
If you need any further information or support concerning this issue, please contact your
local Philips representative:
Technical Support Line: 1-800-722-9377 |
Quantity in Commerce | 25 total |
Distribution | Worldwide distribution - US Nationwide distribution including in the states of MD and MI. The countries of Austria, Belgium, France, Germany, Netherlands, Norway, South Africa, Spain, Switzerland, and United Kingdom. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = OWB
|
|
|
|