• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Arterial Line Kit

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Arterial Line Kitsee related information
Date Initiated by FirmApril 08, 2020
Create DateMay 08, 2020
Recall Status1 Terminated 3 on December 14, 2022
Recall NumberZ-1936-2020
Recall Event ID 85488
510(K)NumberK780871 
Product Classification Gauze/sponge, internal, x-ray detectable - Product Code GDY
ProductArterial Line Kit 20ga X 3", Model 498411B, Rx Only, Sterile EO, UDI: 00886333215652 - Product Usage: Arterial Catheters are indicated for use when in instances where continuous invasive blood pressure readings and/or frequent blood gas measurements are required. Case Label: Arterial Line Kit 20ga x 3", Ref 498411B, Qty: 10/CS
Code Information Lot: 11299236
Recalling Firm/
Manufacturer
Argon Medical Devices, Inc
1445 Flat Creek Rd
Athens TX 75751-5002
For Additional Information ContactRebecca Ellis
903-675-9321
Manufacturer Reason
for Recall
Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual straight-cutting needle with a 3-0 silk suture. The thicker suture may be more difficult to tie. The curved needle requires the use of a needle driver, there is a risk of needle stick to the surgeon tries to do it with fingers instead.
FDA Determined
Cause 2
Under Investigation by firm
ActionArgon Medical Devices, Inc. sent Urgent Medical Device Recall letters, dated April 7, 2020 by mail on April 8th to the affected customer/hospital/distributors. The letter states they shipped Arterial Line Kits that include a different suture set than what is normally packaged in the kits the customers order. The kits customers normally order include a straight-cutting needle with a 3-0 silk suture. The kits customers recently received from them include a curved reverse-cutting needle with a 0-silk suture. A list of the affected product numbers is provided in the letter. They immediately corrected the production error that created this situation and reestablished their supply of Arterial Line Kits with the correct suture set. If customers would like to exchange the Arterial Line Kits they received for kits containing the correct suture set, they will gladly exchange these kits at no cost. Conversely, they understand the rise of the recent COVID-19 pandemic put Arterial Line Kits in high demand in the United States. Therefore, they do not require customers to return the affected Arterial Line Kits if they would prefer to use them as-is. To facilitate an exchange of the product received customers can contact Argon Medical Customer Service at 1-800-927-4669 or customer.service@argonmedical.com. If customers would prefer to use the product as-is, no further action is necessary. The letter included a form for customers to acknowledge receipt and account for product inventory and submit for exchange if they wish.
Quantity in Commerce150
DistributionUS Nationwide distribution including the states of CA, IL, ND, SC, TX, VA. OUS: None
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = GDY
-
-