| | Class 2 Device Recall |  |
| Date Initiated by Firm | May 14, 2020 |
| Create Date | June 09, 2020 |
| Recall Status1 |
Terminated 3 on April 08, 2021 |
| Recall Number | Z-2310-2020 |
| Recall Event ID |
85701 |
| 510(K)Number | K083843 |
| Product Classification |
Plate, fixation, bone - Product Code HRS
|
| Product | A.L.P.S. T15 Multi-Directional Locking Screw 3.5mm x 60mm, Item Number: 110017960 - Product Usage: Intended for temporary internal fixation and stabilization of osteotomies and fractures during the normal healing process. |
| Code Information |
Lot Number: RM540H (01)00887868132490(17)260623(10)RM540H |
| FEI Number |
3003506715
|
Recalling Firm/ Manufacturer |
Biomet
6303 Blue Lagoon Dr Ste 100
Miami
FL
33126-6023
|
| For Additional Information Contact | 411 Technical Services 574-371-3071 |
Manufacturer Reason for Recall | Distributed to the field without having completed design history files or design transfer activities |
FDA Determined Cause 2 | Device Design |
| Action | Zimmmer Biomet issued Urgent Medical Device Recall letter , dated May 14, 2020 to Risk Manger, Distributors,Surgeons, stating reason for recall, health risk and action to take:
1. Review this notification and ensure that affected personnel are aware of the contents.
2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility.
3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form will be returned even if you do not have affected products at your facility.
4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation.
5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.
Surgeon Responsibilities:
1. Review this notification for awareness of the contents.
2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. |
| Quantity in Commerce | 15 OUS |
| Distribution | Worldwide distribution - US Nationwide distribution including in the states of NY, GA, FL, MO and the countries of Canada, AUSTRALIA, CHILE, CHINA, JAPAN, MALAYSIA, NETHERLANDS. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|