| Date Initiated by Firm | August 03, 2020 |
| Create Date | October 15, 2020 |
| Recall Status1 |
Terminated 3 on February 03, 2021 |
| Recall Number | Z-0108-2021 |
| Recall Event ID |
86250 |
| 510(K)Number | K181407 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | Artis zee or Artis Q systems (listed below) - Product Usage: for single and biplane diagnostic imaging and interventional procedures.
System Name/Material Number:
Artis Q BIPLANE 10848282
ARTIS Q CEILING 10848281
Artis Q floor 10848280
ARTIS Q ZEEGO 10848283
ARTIS Q.ZEN BIPLANE 10848355
ARTIS Q.zen ceiling 10848354
Artis Q.zen floor 10848353
Artis zee Biplane 10094141
Artis zee ceiling 10094137
Artis zee floor 10094135
Artis zee floor 10094135
Artis zee multi-purpose 10094139
Artis zeego 10280959 |
| Code Information |
Software VD11C Patch distributed earlier via Update Instructions AX030/19/S or AX051/19/P. Serial Numbers: 103222 103285 103286 105016 109016 109021 109069 109105 109107 109158 109163 109189 109221 109300 109331 109367 109371 109393 109416 109451 109460 109504 109505 109520 111005 111104 111106 111117 111120 111123 111125 117014 117030 117117 117123 117129 117134 117303 117309 117310 117314 117325 117329 117330 121008 121058 121062 121071 121375 121377 121378 121383 121405 121417 121442 123010 123014 123023 123049 123208 123209 123233 124286 125201 125233 137245 137421 137438 137479 137729 138102 138104 138147 138156 138201 146323 147516 147612 147652 147670 147676 147724 147745 147880 147905 147923 147925 147926 147928 147944 148207 148282 148349 154532 154612 154633 154663 154739 155151 157704 158032 158458 158474 158552 160431 160449 160494 160509 160845 160850 160861 160885 160913 160942 160953 160959 160978 160999 161004 161008 161021 161050 161224 161226 |
| FEI Number |
2240869
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc
40 Liberty Blvd
Malvern
PA
19355
|
| For Additional Information Contact | SAME 610-219-4834 |
Manufacturer Reason for Recall | Potential exists for the collision supervision not to work properly following loss of the individual room configuration data settings causing danger of collision with fix mounted room equipment, walls or floor may occur. This might cause components of the system to fall or tilt which could result in crushing of patients, operators or staff, collision of system parts with patients, as well as result in delay or interruption of the clinical procedure |
FDA Determined Cause 2 | Software Manufacturing/Software Deployment |
| Action | Siemens iMedical Solutions USA Inc. issued A Customer Safety Advisory Notice dated 8/03/20 affected customers via AX047/20/S. Letter states reason for recall, health risk and action to take:Siemens strongly recommends paying close attention during system movements to avoid any collision and to establish appropriate emergency procedures until the corrective action has been performed. In case of collision, it is recommended to check the system for any damages before continuing with treatment. If any device part continues to move although that movement was not released, a potential malfunction may be present. In this case, the system must be shut down and Siemens Healthineers service organization must be informed.
Our service organization will contact you to schedule an appointment to perform inspection of the affected system by Siemens service engineer, and, where necessary, recover the individual room configuration data setings
This correction will be implemented via Update Instruction AX046/20/S. Following the inspection by a service engineer the individual room configuration data settings will be available again. Please feel free
to contact our service organization at 1-800-888-7436 for an earlier appointment. Following the correction, the cause will be eliminated, and any recurrence of this potential fault is prevented.
Acknowledge Receipt of this Safety Advisory Notice
Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein. |
| Quantity in Commerce | 124 U.S |
| Distribution | US Natonwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = OWB
|