• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Capnostream 35 Portable Respiratory Monitor

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Capnostream 35 Portable Respiratory Monitorsee related information
Date Initiated by FirmNovember 16, 2020
Create DateDecember 01, 2020
Recall Status1 Open3, Classified
Recall NumberZ-0515-2021
Recall Event ID 86829
510(K)NumberK150272 
Product Classification Analyzer, gas, carbon-dioxide, gaseous-phase - Product Code CCK
ProductCapnostream 35 Portable Respiratory Monitor, Product numbers PM35MN02, PM35MN01, PM35MN05, DLPM35MN02 - Product Usage: use during both no motion and motion conditions and for patients who are well or poorly perfused.
Code Information Serial numbers:  SF16030134 SF16240121 SF16360313 SF16450067 SF17010227 SF17010247 SF17030358 SF17050642 SF17090981 SF17110474 SF17120491 SF17140419 SF17150846 SF17180407 SF17200722 SF17240406 SF18020067 SF18020101 SF18020103 SF18030281 SF18090439 SF18090443 SF18110356 SF18271901 SF18271902 SF18271903 SF18271904 SF18271905 SF18271907 SF18271908 SF18271909 SF18271910 SF18271911 SF18271912 SF18271913 SF18271914 SF18271915 SF18271916 SF18271917 SF18271918 SF18271919 SF18271920 SF18271921 SF18271922 SF18271923 SF18271924 SF18271938 SF18271941 SF18280055 SF18301062 SF18420402 SF18430101 SF18430150 SF18430151 SF18430154 SF18430155 SF18430156 SF18430158 SF18430159 SF18430161 SF18430163 SF18430164 SF18430169 SF18430171 SF18430172 SF18430173 SF18430174 SF18430176 SF18430177 SF18430179 SF18430181 SF18430182 SF18430183 SF18430187 SF18430188 SF18430193 SF18430194 SF18430195 SF18430196 SF18430197 SF19250492 SF19260529 SF19330724 SF19470574 SF19480478 SF19480479 SF19480480 SF19480484 SF19480485 SF19480486 SF19480487 SF19480488 SF19480489 SF19480490 SF19480491 SF19480492 SF19480493 SF19480494 SF19480495 SF19480496 SF19480497 SF19480498 SF19480499 SF19480500 SF19480621 SF19480622 SF19480623 SF19480624 SF19480625 SF19480626 SF19480627 SF19480628 SF19480629 SF19480630 SF19480631 SF19480632 SF19480633 SF19480634 SF19480635 SF19480636 SF19480637 SF19480638 SF19480639 SF19480640 SF19480641 SF19480642 SF19480643 SF19480644 SF19480645 SF19480646 SF19480647 SF19480648 SF19480649 SF19480650 SF19480651 SF19480652 SF19480653 SF19480654 SF19480655 SF19480656 SF19480658 SF19480659 SF19480660 SF19480661 SF19480662 SF19480663 SF19480664 SF19480665 SF19480666 SF19480667 SF19480668 SF19480669 SF19480670 SF19480671 SF19480672 SF19480673 SF19480674 SF19480675 SF19480677 SF19480678 SF19480679 SF19480680 SF19480681 SF19480682 SF19480683 SF19480684 SF19480685 SF19480686 SF19480687 SF19480688 SF19480689 SF19480690 SF19480692 SF19480693 SF19480694 SF19480695 SF19480696 SF19480697 SF19480698 SF19480699 SF19480700 SF19480701 SF19480702 SF19480703 SF19480706 SF19480707 SF19480708 SF19480710 SF19480711 SF19480713 SF19480714 SF19480715 SF19480716 SF19480718 SF19480719 SF19480720 SF19490002 SF19490004 SF19490005 SF19490006 SF19490009 SF19490010 SF19490011 SF19490012 SF19490013 SF19490018 SF19490019 SF19490023 SF19490026 SF19490027 SF19490029 SF19490030 SF19490031 SF19490032 SF19490034 SF19490035 SF19490036 SF19490037 SF19490038 SF19490039 SF19490040 SF19490043 SF19490045 SF19490047 SF19490048 SF19490050 SF19490052 SF19490053 SF19490054 SF19490055 SF19490057 SF19490060 SF19490061 SF19490062 SF19490063 SF19490064 SF19490066 SF19490067 SF19490068 SF19490069 SF19490070 SF19490071 SF19490072 SF19490073 SF19490074 SF19490075 SF19490076 SF19490077 SF19490078 SF19490079 SF19490080 SF19490081 SF19490082 SF19490083 SF19490084 SF19490085 SF19490086 SF19490087 SF19490088 SF19490089 SF19490090 SF19490091 SF19490092 SF19490093 SF19490094 SF19490095 SF19490096 SF19490097 SF19490098 SF19490099 SF19490100 SF19500001 SF19500002 SF19500003 SF19500004 SF19500005 SF19500006 SF19500007 SF19500008 SF19500009 SF19500010 SF19500011 SF19500012 SF19500013 SF19500014 SF19500015 SF19500016 SF19500017 SF19500018 SF19500019 SF19500020 SF19500021 SF19500022 SF19500023 SF19500024 SF19500025 SF19500026 SF19500027 SF19500028 SF19500029 SF19500030 SF19500031 SF19500032 SF19500033 SF19500034 SF19500035 SF19500036 SF19500037 SF19500038 SF19500039 SF19500040 SF19500041 SF19500042 SF19500043 SF19500044 SF19500045 SF19500046 SF19500047 SF19500048 SF19500049 SF19500050 SF19500051 SF19500052 SF19500053 SF19500054 SF19500055 SF19500056 SF19500057 SF19500058 SF19500059 SF19500060 SF19500061 SF19500062 SF19500063 SF19500064 SF19500065 SF19500066 SF19500067 SF19500068 SF19500069 SF19500070 SF19500071 SF19500072 SF19500073 SF19500074 SF19500075 SF19500076 SF19500077 SF19500078 SF19500079 SF19500080 SF19500081 SF19500082 SF19500083 SF19500084 SF19500085 SF19500086 SF19500087 SF19500088 SF19500089 SF19500090 SF19500091 SF19500092 SF19500093 SF19500094 SF19500095 SF19500096 SF19500097 SF19500098 SF19500099 SF19500100 SF19501201 SF19501202 SF19501203 SF19501204 SF19501205 SF19501206 SF19501207 SF19501208 SF19501209 SF19501210 SF19501211 SF19501212 SF19501213 SF19501214 SF19501215 SF19501216 SF19501217 SF19501218 SF19501219 SF19501220 SF19501221 SF19501222 SF19501223 SF19501224 SF19501225 SF19501226 SF19501227 SF19501228 SF19501229 SF19501230 SF19501231 SF19501232 SF19501233 SF19501234 SF19501235 SF19501236 SF19501237 SF19501238 SF19501239 SF19501240 SF19501241 SF19501243 SF19501244 SF19501245 SF19501246 SF19501247 SF19501248 SF19501249 SF19501250 SF19501251 SF19501252 SF19501253 SF19501254 SF19501255 SF19501256 SF19501257 SF19501258 SF19501259 SF19501261 SF19501262 SF19501263 SF19501264 SF19501265 SF19501266 SF19501267 SF19501268 SF19501269 SF19501270 SF19501271 SF19501272 SF19501274 SF19501275 SF19501276 SF19501277 SF19501278 SF19501279 SF19501280 SF19501282 SF19501284 SF19501285 SF19501287 SF19501288 SF19501289 SF19501290 SF19501291 SF19501293 SF19501294 SF19501295 SF19501298 SF19501299 SF19501300 SF19510501 SF19510502 SF19510503 SF19510504 SF19510505 SF19510506 SF19510507 SF19510508 SF19510509 SF19510510 SF19510511 SF19510512 SF19510513 SF19510514 SF19510515 SF19510516 SF19510517 SF19510518 SF19510519 SF19510520 SF19510521 SF19510522 SF19510523 SF19510524 SF19510525 SF19510526 SF19510527 SF19510528 SF19510529 SF19510530 SF19510531 SF19510532 SF19510533 SF19510534 SF19510535 SF19510536 SF19510538 SF19510539 SF19510540 SF19510541 SF19510542 SF19510543 SF19510544 SF19510545 SF19510546 SF19510547 SF19510548 SF19510549 SF19510550 SF19510551 SF19510552 SF19510553 SF19510554 SF19510555 SF19510556 SF19510557 SF19510558 SF19510559 SF19510560 SF19510561 SF19510562 SF19510564 SF19510565 SF19510566 SF19510567 SF19510568 SF19510569 SF19510570 SF19510572 SF19510573 SF19510574 SF19510575 SF19510576 SF19510578 SF19510579 SF19510580 SF19510581 SF19510583 SF19510584 SF19510585 SF19510587 SF19510588 SF19510589 SF19510591 SF19510592 SF19510593 SF19510594 SF19510595 SF19510596 SF19510597 SF19510598 SF19510599 SF19510600 SF19520601 SF20040001 SF20040002 SF20040003 SF20040004 SF20040005 SF20040006 SF20040007 SF20040008 SF20040009 SF20040010 SF20040011 SF20040012 SF20040013 SF20040014 SF20040015 SF20040016 SF20040017 SF20040018 SF20040019 SF20040020 SF20040021 SF20040022 SF20040023 SF20040024 SF20040025 SF20040026 SF20040027 SF20040028 SF20040029 SF20040030 SF20040031 SF20040032 SF20040033 SF20040034 SF20040035 SF20040036 SF20040037 SF20040038 SF20040039 SF20040040 SF20040041 SF20040042 SF20040043 SF20040044 SF20040045 SF20040046 SF20040047 SF20040048 SF20040049 SF20040050 SF20040451 SF20040452 SF20040453 SF20040454 SF20040455 SF20040456 SF20040457 SF20040458 SF20040459 SF20040460 SF20040461 SF20040462 SF20040463 SF20040464 SF20040465 SF20040466 SF20040467 SF20040468 SF20040469 SF20040470 SF20040471 SF20040472 SF20040473 SF20040474 SF20040475 SF20040476 SF20040477 SF20040478 SF20040479 SF20040480 SF20040481 SF20040482 SF20040483 SF20040484 SF20040485 SF20040486 SF20040487 SF20040488 SF20040489 SF20040490 SF20040491 SF20040492 SF20040493 SF20040494 SF20040495 SF20040496 SF20040497 SF20040498 SF20040499 SF20040500 SF20040501 SF20040502 SF20040503 SF20040504 SF20040505 SF20040506 SF20040507 SF20040508 SF20040509 SF20040510 SF20040511 SF20040512 SF20040513 SF20040514 SF20040515 SF20040516 SF20040517 SF20040518 SF20040519 SF20040520 SF20040521 SF20040522 SF20040523 SF20040524 SF20040525 SF20040526 SF20040527 SF20040528 SF20040529 SF20040530 SF20040531 SF20040532 SF20040533 SF20040534 SF20040535 SF20040536 SF20040537 SF20040538 SF20040539 SF20040540 SF20040541 SF20040542 SF20040543 SF20040544 SF20040545 SF20040546 SF20040547 SF20040548 SF20040549 SF20040550 SF20040551 SF20040552 SF20040553 SF20040554 SF20040555 SF20040556 SF20040557 SF20040558 SF20040559 SF20040560 SF20040561 SF20040562 SF20040563 SF20040564 SF20040565 SF20040566 SF20040567 SF20040568 SF20040569 SF20040570 SF20040571 SF20040572 SF20040573 SF20040574 SF20040575 SF20040576 SF20040577 SF20040578 SF20040579 SF20040580 SF20040581 SF20040582 SF20040583 SF20040584 SF20040585 SF20040586 SF20040587 SF20040588 SF20040589 SF20040590 SF20040591 SF20040592 SF20040593 SF20040594 SF20040595 SF20040596 SF20040597 SF20040598 SF20040599 SF20040600 SF20050651 SF20050652 SF20050653 SF20050654 SF20050655 SF20050656 SF20050657 SF20050658 SF20050659 SF20050660 SF20050661 SF20050662 SF20050663 SF20050664 SF20050665 SF20050666 SF20050667 SF20050668 SF20050669 SF20050670 SF20050671 SF20050672 SF20050673 SF20050674 SF20050675 SF20050676 SF20050677 SF20050678 SF20050679 SF20050680 SF20050681 SF20050682 SF20050683 SF20050684 SF20050685 SF20050686 SF20050687 SF20050688 SF20050689 SF20050690 SF20050691 SF20050692 SF20050693 SF20050694 SF20050695 SF20050696 SF20050697 SF20050698 SF20050699 SF20050700 SF20050701 SF20050702 SF20050703 SF20050704 SF20050705 SF20050706 SF20050707 SF20050708 SF20050709 SF20050710 SF20050711 SF20050712 SF20050713 SF20050714 SF20050715 SF20050716 SF20050717 SF20050718 SF20050719 SF20050720 SF20050721 SF20050722 SF20050723 SF20050724 SF20050725 SF20050726 SF20050727 SF20050728 SF20050729 SF20050730 SF20050731 SF20050732 SF20050733 SF20050734 SF20050735 SF20050736 SF20050737 SF20050738 SF20050739 SF20050740 SF20050741 SF20050742 SF20050743 SF20050744 SF20050745 SF20050747 SF20050748 SF20050749 SF20050750 SF20060301 SF20060302 SF20060303 SF20060304 SF20060305 SF20060306 SF20060307 SF20060308 SF20060309 SF20060310 SF20060311 SF20060312 SF20060313 SF20060314 SF20060315 SF20060316 SF20060317 SF20060318 SF20060319 SF20060320 SF20060321 SF20060322 SF20060323 SF20060324 SF20060325 SF20060326 SF20060327 SF20060328 SF20060329 SF20060330 SF20060331 SF20060332 SF20060333 SF20060334 SF20060335 SF20060336 SF20060337 SF20060338 SF20060339 SF20060340 SF20060341 SF20060342 SF20060343 SF20060344 SF20060345 SF20060346 SF20060347 SF20060348 SF20060349 SF20060350 SF20060351 SF20060352 SF20060353 SF20060354 SF20060355 SF20060356 SF20060357 SF20060358 SF20060359 SF20060360 SF20060361 SF20060362 SF20060363 SF20060364 SF20060365 SF20060366 SF20060367 SF20060368 SF20060369 SF20060370 SF20060371 SF20060372 SF20060373 SF20060374 SF20060375 SF20060376 SF20060377 SF20060378 SF20060379 SF20060380 SF20060381 SF20060382 SF20060383 SF20060384 SF20060385 SF20060386 SF20060387 SF20060388 SF20060389 SF20060390 SF20060391 SF20060392 SF20060393 SF20060394 SF20060395 SF20060396 SF20060397 SF20060398 SF20060399 SF20060400 SF20061201 SF20061202 SF20061203 SF20061204 SF20061205 SF20061206 SF20061207 SF20061208 SF20061209 SF20061210 SF20061211 SF20061212 SF20061213 SF20061214 SF20061215 SF20061216 SF20061217 SF20061218 SF20061219 SF20061220 SF20061221 SF20061222 SF20061223 SF20061224 SF20061225 SF20061226 SF20061227 SF20061228 SF20061229 SF20061230 SF20061231 SF20061232 SF20061233 SF20061234 SF20061235 SF20061236 SF20061237 SF20061238 SF20061239 SF20061240 SF20061241 SF20061242 SF20061243 SF20061245 SF20061246 SF20061247 SF20061248 SF20061249 SF20061250 SF20061251 SF20061252 SF20061253 SF20061254 SF20061255 SF20061256 SF20061257 SF20061258 SF20061259 SF20061260 SF20061261 SF20061262 SF20061263 SF20061264 SF20061265 SF20061266 SF20061267 SF20061268 SF20061269 SF20061270 SF20061271 SF20061272 SF20061273 SF20061274 SF20061275 SF20061276 SF20061277 SF20061278 SF20061279 SF20061280 SF20061281 SF20061282 SF20061283 SF20061284 SF20061285 SF20061286 SF20061287 SF20061288 SF20061289 SF20061290 SF20061291 SF20061292 SF20061293 SF20061294 SF20061295 SF20061296 SF20061297 SF20061298 SF20061299 SF20061300 SF20062061 SF20062062 SF20062063 SF20062064 SF20062065 SF20062066 SF20062067 SF20062068 SF20062069 SF20062070 SF20062071 SF20062072 SF20062073 SF20062074 SF20062075 SF20062076 SF20062077 SF20062078 SF20062079 SF20062080 SF20062081 SF20062082 SF20062083 SF20062084 SF20062085 SF20062086 SF20062087 SF20062088 SF20062089 SF20062090 SF20062091 SF20062092 SF20062093 SF20062094 SF20062095 SF20062096 SF20062097 SF20062098 SF20062099 SF20062100 SF20062101 SF20062102 SF20062103 SF20062104 SF20062105 SF20062106 SF20062107 SF20062108 SF20062109 SF20062110 SF20062111 SF20062112 SF20062113 SF20062114 SF20062115 SF20062116 SF20062117 SF20062118 SF20062119 SF20062120 SF20062121 SF20062122 SF20062123 SF20062124 SF20062125 SF20062126 SF20062127 SF20062128 SF20062129 SF20062130 SF20062131 SF20062132 SF20062133 SF20062134 SF20062135 SF20062136 SF20062137 SF20062138 SF20062139 SF20062140 SF20062141 SF20062142 SF20062143 SF20062144 SF20062145 SF20062146 SF20062147 SF20062148 SF20062149 SF20062150 SF20062151 SF20062152 SF20062153 SF20062154 SF20062155 SF20062156 SF20062157 SF20062158 SF20062159 SF20062160 SF20070601 SF20070602 SF20070603 SF20070604 SF20070605 SF20070606 SF20070607 SF20070608 SF20070609 SF20070610 SF20070611 SF20070612 SF20070613 SF20070614 SF20070615 SF20070616 SF20070617 SF20070618 SF20070619 SF20070620 SF20070621 SF20070622 SF20070623 SF20070624 SF20070625 SF20070626 SF20070627 SF20070628 SF20070629 SF20070630 SF20070631 SF20070632 SF20070633 SF20070634 SF20070635 SF20070636 SF20070637 SF20070638 SF20070639 SF20070640 SF20070641 SF20070642 SF20070643 SF20070644 SF20070645 SF20070646 SF20070647 SF20070648 SF20070649 SF20070650 SF20070651 SF20070652 SF20070653 SF20070654 SF20070655 SF20070656 SF20070657 SF20070658 SF20070659 SF20070660 SF20070661 SF20070662 SF20070663 SF20070664 SF20070665 SF20070666 SF20070667 SF20070668 SF20070669 SF20070670 SF20070671 SF20070672 SF20070673 SF20070674 SF20070675 SF20070676 SF20070677 SF20070678 SF20070679 SF20070680 SF20070681 SF20070682 SF20070683 SF20070684 SF20070685 SF20070686 SF20070687 SF20070688 SF20070689 SF20070690 SF20070691 SF20070692 SF20070693 SF20070694 SF20070695 SF20070696 SF20070697 SF20070698 SF20070699 SF20070700 SF20080401 SF20080402 SF20080403 SF20080404 SF20080405 SF20080406 SF20080407 SF20080408 SF20080409 SF20080410 SF20080411 SF20080412 SF20080413 SF20080414 SF20080415 SF20080416 SF20080417 SF20080418 SF20080419 SF20080420 SF20080421 SF20080422 SF20080423 SF20080424 SF20080425 SF20080426 SF20080427 SF20080428 SF20080429 SF20080430 SF20080431 SF20080432 SF20080433 SF20080434 SF20080435 SF20080436 SF20080437 SF20080438 SF20080439 SF20080440 SF20080441 SF20080442 SF20080443 SF20080444 SF20080445 SF20080446 SF20080447 SF20080448 SF20080449 SF20080450 SF20080451 SF20080452 SF20080453 SF20080454 SF20080455 SF20080456 SF20080457 SF20080458 SF20080459 SF20080460 SF20080461 SF20080462 SF20080463 SF20080464 SF20080465 SF20080466 SF20080467 SF20080468 SF20080469 SF20080470 SF20080471 SF20080472 SF20080473 SF20080474 SF20080475 SF20080476 SF20080477 SF20080478 SF20080479 SF20080480 SF20080481 SF20080482 SF20080483 SF20080484 SF20080485 SF20080486 SF20080487 SF20080488 SF20080489 SF20080490 SF20080491 SF20080492 SF20080493 SF20080494 SF20080495 SF20080496 SF20080497 SF20080498 SF20080499 SF20080500 SF20081509 SF20081510 SF20081511 SF20081512 SF20081513 SF20081514 SF20081515 SF20081516 SF20081517 SF20081518 SF20081519 SF20081520 SF20081521 SF20081522 SF20081523 SF20081524 SF20081525 SF20081526 SF20081527 SF20081528 SF20081529 SF20081530 SF20081531 SF20081532 SF20081533 SF20081534 SF20081535 SF20081536 SF20081537 SF20081538 SF20081540 SF20081541 SF20081542 SF20081543 SF20081544 SF20081545 SF20081546 SF20081547 SF20081548 SF20081549 SF20081550 SF20081551 SF20081552 SF20081553 SF20081554 SF20081555 SF20081556 SF20081557 SF20081558 SF20081559 SF20081560 SF20081561 SF20081562 SF20081563 SF20081564 SF20081565 SF20081566 SF20081567 SF20081568 SF20081569 SF20081570 SF20081571 SF20081572 SF20081573 SF20081574 SF20081575 SF20081576 SF20081577 SF20081578 SF20081579 SF20081580 SF20081581 SF20081582 SF20081583 SF20081584 SF20081585 SF20081586 SF20160651 SF20160652 SF20160655 SF20160656 SF20160657 SF20160658 SF20160659 SF20160660 SF20160663 SF20160664 SF20160665 SF20160666 SF20160667 SF20160668 SF20160669 SF20160670 SF20160671 SF20160672 SF20160673 SF20160674 SF20160675 SF20160676 SF20160678 SF20160680 SF20160681 SF20160682 SF20160683 SF20160684 SF20160685 SF20160686 SF20160687 SF20160688 SF20160689 SF20160690 SF20160691 SF20160692 SF20160693 SF20160694 SF20160695 SF20160696 SF20160697 SF20160698 SF20160699 SF20160700 SF20160701 SF20160702 SF20160703 SF20160704 SF20160705 SF20160706 SF20160707 SF20160708 SF20160709 SF20160710 SF20160711 SF20160712 SF20160713 SF20160714 SF20160715 SF20160716 SF20160717 SF20160718 SF20160719 SF20160720 SF20160721 SF20160722 SF20160723 SF20160724 SF20160725 SF20160726 SF20160727 SF20160728 SF20160729 SF20160730 SF20160731 SF20160732 SF20160733 SF20160734 SF20160735 SF20160736 SF20160737 SF20160738 SF20160739 SF20160740 SF20160741 SF20160742 SF20160743 SF20160744 SF20160745 SF20160746 SF20160747 SF20160748 SF20160749 SF20160750 SF20170940 SF20170941 SF20170942 SF20170943 SF20170944 SF20170945 SF20170946 SF20170947 SF20170948 SF20170949 SF20170950 SF20170951 SF20170952 SF20170953 SF20170954 SF20170955 SF20170956 SF20170957 SF20170958 SF20170959 SF20170960 SF20170961 SF20170962 SF20170963 SF20170964 SF20170965 SF20170966 SF20170967 SF20170968 SF20170969 SF20170970 SF20170971 SF20170972 SF20170973 SF20170974 SF20170975 SF20170976 SF20170977 SF20170978 SF20170979 SF20170980 SF20170981 SF20170982 SF20170983 SF20170984 SF20170985 SF20170986 SF20170987 SF20170988 SF20170989 SF20170990 SF20170991 SF20170992 SF20170993 SF20170994 SF20170995 SF20170996 SF20170997 SF20170998 SF20170999 SF20171000 SF20171001 SF20171002 SF20171003 SF20171004 SF20171005 SF20171006 SF20171007 SF20171008 SF20171009 SF20171010 SF20171011 SF20171012 SF20171013 SF20171014 SF20171015 SF20171016 SF20171017 SF20171018 SF20171019 SF20171020 SF20171021 SF20171022 SF20171023 SF20171024 SF20171025 SF20171026 SF20171027 SF20171028 SF20171029 SF20171030 SF20171031 SF20171032 SF20171033 SF20171034 SF20171035 SF20171036 SF20171037 SF20171038 SF20171039 SF20180001 SF20180002 SF20180003 SF20180004 SF20180005 SF20180006 SF20180007 SF20180008 SF20180009 SF20180010 SF20180011 SF20180012 SF20180013 SF20180014 SF20180015 SF20180016 SF20180017 SF20180018 SF20180019 SF20180020 SF20180021 SF20180022 SF20180023 SF20180024 SF20180025 SF20180026 SF20180027 SF20180028 SF20180029 SF20180030 SF20180031 SF20180032 SF20180033 SF20180034 SF20180035 SF20180036 SF20180037 SF20180038 SF20180039 SF20180040 SF20180041 SF20180042 SF20180043 SF20180044 SF20180045 SF20180046 SF20180047 SF20180048 SF20180049 SF20180050 SF20180051 SF20180052 SF20180053 SF20180054 SF20180055 SF20180056 SF20180057 SF20180058 SF20180059 SF20180060 SF20180061 SF20180062 SF20180063 SF20180064 SF20180065 SF20180066 SF20180067 SF20180068 SF20180069 SF20180070 SF20180071 SF20180072 SF20180073 SF20180074 SF20180075 SF20180076 SF20180077 SF20180078 SF20180079 SF20180080 SF20180081 SF20180082 SF20180083 SF20180084 SF20180085 SF20180086 SF20180087 SF20180088 SF20180089 SF20180090 SF20180091 SF20180092 SF20180093 SF20180094 SF20180095 SF20180096 SF20180097 SF20180098 SF20180099 SF20180100 SF20180371 SF20180372 SF20180373 SF20180374 SF20180375 SF20180376 SF20180377 SF20180378 SF20180379 SF20180380 SF20180381 SF20180382 SF20180383 SF20180384 SF20180385 SF20180386 SF20180387 SF20180388 SF20180389 SF20180390 SF20180391 SF20180392 SF20180393 SF20180394 SF20180395 SF20180396 SF20180397 SF20180398 SF20180399 SF20180400 SF20180401 SF20180402 SF20180403 SF20180404 SF20180405 SF20180406 SF20180407 SF20180408 SF20180409 SF20180410 SF20180411 SF20180412 SF20180413 SF20180414 SF20180415 SF20180416 SF20180417 SF20180418 SF20180419 SF20180420 SF20180421 SF20180422 SF20180423 SF20180424 SF20180425 SF20180426 SF20180427 SF20180428 SF20180429 SF20180430 SF20180431 SF20180432 SF20180433 SF20180434 SF20180435 SF20180436 SF20180437 SF20180438 SF20180439 SF20180440 SF20180441 SF20180442 SF20180443 SF20180444 SF20180445 SF20180446 SF20180447 SF20180448 SF20180449 SF20180450 SF20180451 SF20180452 SF20180453 SF20180454 SF20180455 SF20180456 SF20180457 SF20180458 SF20180459 SF20180460 SF20180461 SF20180462 SF20180463 SF20180464 SF20180465 SF20180466 SF20180467 SF20180468 SF20180469 SF20180470 SF20180471 SF20180472 SF20180473 SF20180474 SF20180475 SF20180476 SF20180477 SF20180478 SF20180479 SF20180480 SF20180481 SF20180482 SF20180483 SF20180484 SF20180485 SF20180486 SF20180487 SF20180488 SF20180489 SF20180490 SF20180491 SF20180492 SF20180493 SF20180494 SF20180495 SF20180496 SF20180497 SF20180498 SF20180499 SF20180500 SF20180501 SF20180502 SF20180503 SF20180504 SF20180505 SF20180506 SF20180507 SF20180508 SF20180509 SF20180510 SF20180511 SF20180512 SF20180513 SF20180514 SF20180515 SF20180516 SF20180517 SF20180518 SF20180519 SF20180520 SF20180521 SF20180522 SF20180523 SF20180524 SF20180525 SF20180526 SF20180527 SF20180528 SF20180529 SF20180530 SF20180531 SF20180532 SF20180533 SF20180534 SF20180535 SF20180536 SF20180537 SF20180538 SF20180539 SF20180540 SF20180541 SF20180542 SF20180543 SF20180544 SF20180545 SF20180546 SF20180547 SF20180548 SF20180549 SF20180550 SF20180551 SF20180552 SF20180553 SF20180554 SF20180555 SF20180556 SF20180557 SF20180558 SF20180559 SF20180560 SF20180561 SF20180562 SF20180563 SF20180564 SF20180565 SF20180566 SF20180567 SF20180568 SF20180569 SF20180570 SF20190501 SF20190502 SF20190503 SF20190504 SF20190505 SF20190507 SF20190508 SF20190509 SF20190510 SF20190511 SF20190512 SF20190513 SF20190514 SF20190515 SF20190516 SF20190517 SF20190518 SF20190519 SF20190520 SF20190521 SF20190522 SF20190523 SF20190524 SF20190525 SF20190526 SF20190527 SF20190529 SF20190530 SF20190531 SF20190532 SF20190533 SF20190534 SF20190535 SF20190536 SF20190537 SF20190538 SF20190539 SF20190540 SF20190541 SF20190542 SF20190543 SF20190544 SF20190545 SF20190546 SF20190547 SF20190548 SF20190549 SF20190550 SF20190551 SF20190552 SF20190553 SF20190554 SF20190555 SF20190556 SF20190557 SF20190558 SF20190559 SF20190560 SF20190561 SF20190562 SF20190563 SF20190564 SF20190565 SF20190566 SF20190567 SF20190568 SF20190569 SF20190570 SF20190571 SF20190572 SF20190573 SF20190574 SF20190575 SF20190576 SF20190577 SF20190578 SF20190579 SF20190580 SF20190581 SF20190582 SF20190584 SF20190585 SF20190586 SF20190587 SF20190588 SF20190589 SF20190590 SF20190591 SF20190592 SF20190593 SF20190594 SF20190595 SF20190596 SF20190597 SF20190598 SF20190599 SF20190600 SF20191614 SF20191615 SF20191616 SF20191617 SF20191618 SF20191619 SF20191620 SF20191621 SF20191622 SF20191623 SF20191624 SF20191625 SF20191626 SF20191627 SF20191628 SF20191629 SF20191630 SF20191631 SF20191632 SF20191633 SF20191634 SF20191635 SF20191636 SF20191637 SF20191638 SF20191639 SF20191640 SF20191641 SF20191642 SF20191643 SF20191644 SF20191645 SF20191646 SF20191647 SF20191648 SF20191649 SF20191650 SF20191651 SF20191652 SF20191653 SF20191654 SF20191655 SF20191656 SF20191657 SF20191658 SF20191659 SF20191660 SF20191661 SF20191662 SF20191663 SF20191664 SF20191665 SF20191666 SF20191667 SF20191668 SF20191669 SF20191670 SF20191671 SF20191672 SF20191673 SF20191674 SF20191675 SF20191676 SF20191677 SF20191678 SF20191679 SF20191680 SF20191681 SF20191682 SF20191683 SF20191684 SF20191685 SF20191686 SF20191687 SF20191688 SF20191689 SF20191690 SF20191691 SF20191692 SF20191693 SF20191694 SF20191695 SF20191696 SF20191697 SF20191698 SF20191699 SF20191700 SF20191701 SF20191702 SF20191703 SF20191704 SF20191705 SF20191706 SF20191707 SF20191708 SF20191709 SF20191710 SF20191711 SF20191712 SF20191713 SF20200793 SF20200794 SF20200795 SF20200796 SF20200797 SF20200798 SF20200799 SF20200800 SF20200801 SF20200802 SF20200803 SF20200804 SF20200805 SF20200806 SF20200807 SF20200808 SF20200809 SF20200810 SF20200811 SF20200812 SF20200813 SF20200814 SF20200815 SF20200816 SF20200817 SF20200818 SF20200819 SF20200820 SF20200821 SF20200822 SF20200823 SF20200824 SF20200825 SF20200826 SF20200827 SF20200828 SF20200829 SF20200830 SF20200831 SF20200832 SF20200833 SF20200834 SF20200835 SF20200836 SF20200837 SF20200838 SF20200839 SF20200840 SF20200841 SF20200842 SF20200843 SF20200844 SF20200845 SF20200846 SF20200847 SF20200848 SF20200849 SF20200850 SF20200851 SF20200852 SF20200853 SF20200854 SF20200855 SF20200856 SF20200857 SF20200858 SF20200859 SF20200860 SF20200861 SF20200862 SF20200863 SF20200864 SF20200865 SF20200866 SF20200867 SF20200868 SF20200869 SF20200870 SF20200871 SF20200872 SF20200873 SF20200874 SF20200875 SF20200876 SF20200877 SF20200878 SF20200879 SF20200880 SF20200881 SF20200882 SF20200883 SF20200884 SF20200885 SF20200886 SF20200887 SF20200888 SF20200889 SF20200890 SF20200891 SF20200892 SF20200893 SF20200894 SF20200895 SF20200896 SF20200897 SF20200898 SF20200899 SF20200900 SF20210201 SF20210202 SF20210203 SF20210204 SF20210205 SF20210206 SF20210207 SF20210208 SF20210209 SF20210210 SF20210211 SF20210212 SF20210213 SF20210214 SF20210215 SF20210216 SF20210217 SF20210218 SF20210219 SF20210220 SF20210221 SF20210222 SF20210223 SF20210224 SF20210225 SF20210226 SF20210227 SF20210228 SF20210229 SF20210230 SF20210231 SF20210232 SF20210233 SF20210234 SF20210235 SF20210236 SF20210237 SF20210238 SF20210239 SF20210240 SF20210241 SF20210242 SF20210243 SF20210244 SF20210245 SF20210246 SF20210247 SF20210248 SF20210249 SF20210250 SF20210251 SF20210252 SF20210253 SF20210254 SF20210255 SF20210256 SF20210257 SF20210258 SF20210259 SF20210260 SF20210261 SF20210262 SF20210263 SF20210264 SF20210265 SF20210266 SF20210267 SF20210268 SF20210269 SF20210270 SF20210271 SF20210272 SF20210273 SF20210274 SF20210275 SF20210276 SF20210277 SF20210278 SF20210279 SF20210280 SF20210281 SF20210282 SF20210283 SF20210284 SF20210285 SF20210286 SF20210287 SF20210288 SF20210289 SF20210290 SF20210291 SF20210292 SF20210293 SF20210294 SF20210295 SF20210296 SF20210297 SF20210298 SF20210299 SF20210300 SF20211651 SF20211652 SF20211653 SF20211654 SF20211655 SF20211656 SF20211657 SF20211658 SF20211659 SF20211660 SF20211661 SF20211662 SF20211663 SF20211664 SF20211665 SF20211666 SF20211667 SF20211668 SF20211669 SF20211670 SF20211671 SF20211672 SF20211673 SF20211674 SF20211675 SF20211676 SF20211677 SF20211678 SF20211679 SF20211680 SF20211681 SF20211682 SF20211683 SF20211684 SF20211685 SF20211686 SF20211687 SF20211688 SF20211689 SF20211690 SF20211691 SF20211692 SF20211693 SF20211694 SF20211695 SF20211696 SF20211697 SF20211698 SF20211699 SF20211700 SF20211701 SF20211702 SF20211703 SF20211704 SF20211705 SF20211706 SF20211707 SF20211708 SF20211709 SF20211710 SF20211711 SF20211712 SF20211713 SF20211714 SF20211715 SF20211716 SF20211717 SF20211718 SF20211719 SF20211720 SF20211721 SF20211722 SF20211723 SF20211724 SF20211725 SF20211726 SF20211727 SF20211728 SF20211729 SF20211730 SF20211731 SF20211732 SF20211733 SF20211734 SF20211735 SF20211736 SF20211737 SF20211738 SF20211739 SF20211740 SF20211741 SF20211742 SF20211743 SF20211744 SF20211745 SF20211746 SF20211747 SF20211748 SF20211749 SF20211750 SF20220401 SF20220402 SF20220403 SF20220404 SF20220405 SF20220406 SF20220407 SF20220408 SF20220409 SF20220410 SF20220411 SF20220412 SF20220413 SF20220414 SF20220415 SF20220416 SF20220417 SF20220418 SF20220419 SF20220420 SF20220421 SF20220422 SF20220423 SF20220424 SF20220425 SF20220426 SF20220427 SF20220428 SF20220429 SF20220430 SF20220431 SF20220432 SF20220433 SF20220434 SF20220435 SF20220436 SF20220437 SF20220438 SF20220439 SF20220440 SF20220441 SF20220442 SF20220443 SF20220444 SF20220445 SF20220446 SF20220447 SF20220448 SF20220449 SF20220450 SF20220451 SF20220452 SF20220453 SF20220454 SF20220455 SF20220456 SF20220457 SF20220458 SF20220459 SF20220460 SF20220461 SF20220462 SF20220463 SF20220464 SF20220465 SF20220466 SF20220467 SF20220468 SF20220469 SF20220470 SF20220471 SF20220472 SF20220473 SF20220474 SF20220475 SF20220476 SF20220477 SF20220478 SF20220479 SF20220480 SF20220481 SF20220482 SF20220483 SF20220484 SF20220485 SF20220486 SF20220487 SF20220488 SF20220489 SF20220490 SF20220491 SF20220492 SF20220493 SF20220494 SF20220495 SF20220496 SF20220497 SF20220498 SF20220499 SF20220500 SF20230801 SF20230802 SF20230803 SF20230804 SF20230805 SF20230806 SF20230807 SF20230808 SF20230809 SF20230810 SF20230811 SF20230812 SF20230813 SF20230814 SF20230815 SF20230816 SF20230817 SF20230818 SF20230819 SF20230820 SF20230821 SF20230822 SF20230823 SF20230824 SF20230825 SF20230826 SF20230827 SF20230828 SF20230829 SF20230830 SF20230831 SF20230832 SF20230833 SF20230834 SF20230835 SF20230836 SF20230837 SF20230838 SF20230839 SF20230840 SF20230841 SF20230842 SF20230843 SF20230844 SF20230845 SF20230846 SF20230847 SF20230848 SF20230849 SF20230850 SF20230851 SF20230852 SF20230853 SF20230854 SF20230855 SF20230856 SF20230857 SF20230858 SF20230859 SF20230860 SF20230861 SF20230862 SF20230863 SF20230864 SF20230865 SF20230866 SF20230867 SF20230868 SF20230869 SF20230870 SF20230871 SF20230872 SF20230873 SF20230874 SF20230875 SF20230876 SF20230877 SF20230878 SF20230879 SF20230880 SF20230881 SF20230882 SF20230883 SF20230884 SF20230885 SF20230886 SF20230887 SF20230888 SF20230889 SF20230890 SF20230891 SF20230892 SF20230893 SF20230894 SF20230895 SF20230896 SF20230897 SF20230898 SF20230899 SF20230900 SF20231501 SF20231502 SF20231503 SF20231504 SF20231505 SF20231506 SF20231507 SF20231508 SF20231509 SF20231510 SF20231511 SF20231512 SF20231513 SF20231514 SF20231515 SF20231516 SF20231517 SF20231518 SF20231519 SF20231520 SF20231521 SF20231522 SF20231523 SF20231524 SF20231525 SF20231526 SF20231527 SF20231528 SF20231529 SF20231530 SF20231531 SF20231532 SF20231533 SF20231534 SF20231535 SF20231536 SF20231537 SF20231538 SF20231539 SF20231540 SF20231541 SF20231542 SF20231543 SF20231544 SF20231545 SF20231546 SF20231547 SF20231548 SF20231549 SF20231550 SF20231551 SF20231552 SF20231553 SF20231554 SF20231555 SF20231556 SF20231557 SF20231558 SF20231559 SF20231560 SF20231561 SF20231562 SF20231563 SF20231564 SF20231565 SF20231566 SF20231567 SF20231568 SF20231569 SF20231570 SF20231571 SF20231572 SF20231573 SF20231574 SF20231575 SF20231576 SF20231577 SF20231578 SF20231579 SF20231580 SF20231581 SF20231582 SF20231583 SF20231584 SF20231585 SF20231586 SF20231587 SF20231588 SF20231589 SF20231590 SF20231591 SF20231592 SF20231593 SF20231594 SF20231595 SF20231596 SF20231597 SF20231598 SF20231599 SF20231600 SF20240811 SF20240812 SF20240813 SF20240814 SF20240815 SF20240816 SF20240817 SF20240818 SF20240819 SF20240820 SF20240821 SF20240822 SF20240823 SF20240824 SF20240825 SF20240826 SF20240827 SF20240828 SF20240829 SF20240830 SF20240831 SF20240832 SF20240833 SF20240834 SF20240835 SF20240836 SF20240837 SF20240838 SF20240839 SF20240840 SF20240841 SF20240842 SF20240843 SF20240844 SF20240845 SF20240846 SF20240847 SF20240848 SF20240849 SF20240850 SF20240851 SF20240852 SF20240853 SF20240854 SF20240855 SF20240856 SF20240857 SF20240858 SF20240859 SF20240860 SF20240861 SF20240862 SF20240863 SF20240864 SF20240865 SF20240866 SF20240867 SF20240868 SF20240869 SF20240870 SF20240871 SF20240872 SF20240873 SF20240874 SF20240875 SF20240876 SF20240877 SF20240878 SF20240879 SF20240880 SF20240881 SF20240882 SF20240883 SF20240884 SF20240885 SF20240886 SF20240887 SF20240888 SF20240889 SF20240890 SF20240891 SF20240892 SF20240893 SF20240894 SF20240895 SF20240896 SF20240897 SF20240898 SF20240899 SF20240900 SF20240901 SF20240902 SF20240903 SF20240904 SF20240905 SF20240906 SF20240907 SF20240908 SF20240909 SF20240910 SF20251119 SF20251120 SF20251121 SF20251122 SF20251123 SF20251124 SF20251125 SF20251126 SF20251127 SF20251128 SF20251129 SF20251130 SF20251131 SF20251132 SF20251133 SF20251134 SF20251135 SF20251136 SF20251137 SF20251138 SF20251139 SF20251140 SF20251141 SF20251142 SF20251143 SF20251144 SF20251145 SF20251146 SF20251147 SF20251148 SF20251149 SF20251150 SF20251151 SF20251152 SF20251153 SF20251154 SF20251155 SF20251156 SF20251157 SF20251158 SF20251159 SF20251160 SF20251161 SF20251162 SF20251163 SF20251164 SF20251165 SF20251166 SF20251167 SF20251168 SF20251169 SF20251170 SF20251171 SF20251172 SF20251173 SF20251174 SF20251175 SF20251176 SF20251177 SF20251178 SF20251179 SF20251180 SF20251181 SF20251182 SF20251183 SF20251184 SF20251185 SF20251186 SF20251187 SF20251188 SF20251189 SF20251190 SF20251191 SF20251192 SF20251193 SF20251194 SF20251195 SF20251196 SF20251197 SF20251198 SF20251199 SF20251200 SF20251201 SF20251202 SF20251203 SF20251204 SF20251205 SF20251206 SF20251207 SF20251208 SF20251209 SF20251210 SF20251211 SF20251212 SF20251213 SF20251214 SF20251215 SF20251216 SF20251217 SF20251218 SF20251319 SF20251320 SF20251321 SF20251322 SF20251323 SF20251324 SF20251325 SF20251326 SF20251327 SF20251328 SF20251329 SF20251330 SF20251331 SF20251332 SF20251333 SF20251334 SF20251335 SF20251336 SF20251337 SF20251338 SF20251339 SF20251340 SF20251341 SF20251342 SF20251343 SF20251344 SF20251345 SF20251346 SF20251347 SF20251348 SF20251349 SF20251350 SF20251351 SF20251352 SF20251353 SF20251354 SF20251355 SF20251356 SF20251357 SF20251358 SF20251359 SF20251360 SF20251361 SF20251362 SF20251363 SF20251364 SF20251365 SF20251366 SF20251367 SF20251368 SF20251369 SF20251370 SF20251371 SF20251372 SF20251373 SF20251374 SF20251375 SF20251376 SF20251377 SF20251378 SF20251379 SF20251380 SF20251381 SF20251382 SF20251383 SF20251384 SF20251385 SF20251386 SF20251387 SF20251388 SF20251389 SF20251390 SF20251391 SF20251392 SF20251393 SF20251394 SF20251395 SF20251396 SF20251398 SF20251399 SF20251400 SF20251401 SF20251402 SF20251403 SF20251404 SF20251405 SF20251406 SF20251407 SF20251408 SF20251409 SF20251410 SF20251411 SF20251412 SF20251413 SF20251414 SF20251415 SF20251416 SF20251417 SF20251418 SF20261401 SF20261402 SF20261403 SF20261404 SF20261405 SF20261406 SF20261407 SF20261408 SF20261409 SF20261410 SF20261411 SF20261412 SF20261413 SF20261414 SF20261415 SF20261416 SF20261417 SF20261418 SF20261419 SF20261420 SF20261421 SF20261422 SF20261423 SF20261424 SF20261425 SF20261426 SF20261427 SF20261428 SF20261429 SF20261430 SF20261431 SF20261432 SF20261433 SF20261434 SF20261435 SF20261436 SF20261437 SF20261438 SF20261439 SF20261440 SF20261441 SF20261442 SF20261443 SF20261444 SF20261445 SF20261446 SF20261447 SF20261448 SF20261449 SF20261450 SF20261451 SF20261452 SF20261453 SF20261454 SF20261455 SF20261456 SF20261457 SF20261458 SF20261459 SF20261460 SF20261461 SF20261462 SF20261463 SF20261464 SF20261465 SF20261466 SF20261467 SF20261468 SF20261469 SF20261470 SF20261471 SF20261472 SF20261473 SF20261474 SF20261475 SF20261476 SF20261477 SF20261478 SF20261479 SF20261480 SF20261481 SF20261482 SF20261483 SF20261484 SF20261485 SF20261486 SF20261487 SF20261488 SF20261489 SF20261490 SF20261491 SF20261492 SF20261493 SF20261494 SF20261495 SF20261496 SF20261497 SF20261498 SF20261499 SF20261500 SF20261601 SF20261602 SF20261603 SF20261604 SF20261605 SF20261606 SF20261607 SF20261608 SF20261609 SF20261610 SF20261611 SF20261612 SF20261613 SF20261614 SF20261615 SF20261616 SF20261617 SF20261618 SF20261619 SF20261620 SF20261621 SF20261622 SF20261623 SF20261624 SF20261625 SF20261626 SF20261627 SF20261628 SF20261629 SF20261630 SF20261631 SF20261632 SF20261633 SF20261634 SF20261635 SF20261636 SF20261637 SF20261638 SF20261639 SF20261640 SF20261641 SF20261642 SF20261643 SF20261644 SF20261645 SF20261646 SF20261647 SF20261648 SF20261649 SF20261650 SF20261651 SF20261652 SF20261653 SF20261654 SF20261655 SF20261656 SF20261657 SF20261658 SF20261659 SF20261660 SF20261661 SF20261662 SF20261663 SF20261664 SF20261665 SF20261666 SF20261667 SF20261668 SF20261669 SF20261670 SF20261671 SF20261672 SF20261673 SF20261674 SF20261675 SF20261676 SF20261677 SF20261678 SF20261679 SF20261680 SF20261681 SF20261682 SF20261683 SF20261684 SF20261685 SF20261686 SF20261687 SF20261688 SF20261689 SF20261690 SF20261691 SF20261692 SF20261693 SF20261694 SF20261695 SF20261696 SF20261697 SF20261698 SF20261699 SF20261700 SF20262101 SF20262102 SF20262103 SF20262104 SF20262105 SF20262106 SF20262107 SF20262108 SF20262109 SF20262110 SF20262111 SF20262112 SF20262113 SF20262114 SF20262115 SF20262116 SF20262117 SF20262118 SF20262119 SF20262120 SF20262121 SF20262122 SF20262123 SF20262124 SF20262125 SF20262126 SF20262127 SF20262128 SF20262129 SF20262130 SF20262131 SF20262132 SF20262133 SF20262134 SF20262135 SF20262136 SF20262137 SF20262138 SF20262139 SF20262140 SF20262141 SF20262142 SF20262143 SF20262144 SF20262145 SF20262146 SF20262147 SF20262148 SF20262149 SF20262150 SF20262151 SF20262152 SF20262153 SF20262154 SF20262155 SF20262156 SF20262157 SF20262158 SF20262159 SF20262160 SF20262161 SF20262162 SF20262163 SF20262164 SF20262165 SF20262166 SF20262167 SF20262168 SF20262169 SF20262170 SF20262171 SF20262172 SF20262173 SF20262174 SF20262175 SF20262176 SF20262177 SF20262178 SF20262179 SF20262180 SF20262181 SF20262182 SF20262183 SF20262184 SF20262185 SF20262186 SF20262187 SF20262188 SF20262189 SF20262190 SF20262191 SF20262192 SF20262193 SF20262194 SF20262196 SF20262197 SF20262198 SF20262199 SF20262200 SF20270401 SF20270402 SF20270403 SF20270404 SF20270405 SF20270406 SF20270407 SF20270408 SF20270409 SF20270410 SF20270411 SF20270412 SF20270413 SF20270414 SF20270415 SF20270416 SF20270417 SF20270418 SF20270419 SF20270420 SF20270421 SF20270422 SF20270423 SF20270424 SF20270425 SF20270426 SF20270427 SF20270428 SF20270429 SF20270430 SF20270431 SF20270432 SF20270433 SF20270434 SF20270435 SF20270436 SF20270437 SF20270438 SF20270439 SF20270440 SF20270441 SF20270442 SF20270443 SF20270444 SF20270445 SF20270446 SF20270447 SF20270448 SF20270449 SF20270450 SF20270451 SF20270452 SF20270453 SF20270454 SF20270455 SF20270456 SF20270457 SF20270458 SF20270459 SF20270460 SF20270461 SF20270462 SF20270463 SF20270464 SF20270465 SF20270466 SF20270467 SF20270468 SF20270469 SF20270470 SF20270471 SF20270472 SF20270473 SF20270474 SF20270475 SF20270476 SF20270477 SF20270478 SF20270479 SF20270480 SF20270481 SF20270482 SF20270483 SF20270484 SF20270485 SF20270486 SF20270487 SF20270488 SF20270489 SF20270490 SF20270491 SF20270492 SF20270493 SF20270494 SF20270495 SF20270496 SF20270497 SF20270498 SF20270499 SF20270500 SF20271101 SF20271102 SF20271103 SF20271104 SF20271105 SF20271106 SF20271107 SF20271108 SF20271109 SF20271110 SF20271111 SF20271112 SF20271113 SF20271114 SF20271115 SF20271116 SF20271117 SF20271118 SF20271119 SF20271120 SF20271121 SF20271122 SF20271123 SF20271124 SF20271125 SF20271126 SF20271127 SF20271128 SF20271129 SF20271130 SF20271131 SF20271132 SF20271133 SF20271134 SF20271135 SF20271136 SF20271137 SF20271138 SF20271139 SF20271140 SF20271141 SF20271142 SF20271143 SF20271144 SF20271145 SF20271146 SF20271147 SF20271148 SF20271149 SF20271150 SF20271151 SF20271152 SF20271153 SF20271154 SF20271155 SF20271156 SF20271157 SF20271158 SF20271159 SF20271160 SF20271161 SF20271162 SF20271163 SF20271164 SF20271165 SF20271166 SF20271167 SF20271168 SF20271169 SF20271170 SF20271171 SF20271172 SF20271173 SF20271174 SF20271175 SF20271176 SF20271177 SF20271178 SF20271179 SF20271180 SF20271181 SF20271182 SF20271183 SF20271184 SF20271185 SF20271186 SF20271187 SF20271188 SF20271189 SF20271190 SF20271191 SF20271192 SF20271193 SF20271194 SF20271195 SF20271196 SF20271197 SF20271198 SF20271199 SF20271200 SF20280401 SF20280402 SF20280403 SF20280404 SF20280405 SF20280406 SF20280407 SF20280408 SF20280409 SF20280410 SF20280411 SF20280412 SF20280413 SF20280414 SF20280415 SF20280416 SF20280417 SF20280418 SF20280419 SF20280420 SF20280421 SF20280422 SF20280423 SF20280424 SF20280425 SF20280426 SF20280427 SF20280428 SF20280429 SF20280430 SF20280431 SF20280432 SF20280433 SF20280434 SF20280435 SF20280436 SF20280437 SF20280438 SF20280439 SF20280440 SF20280441 SF20280442 SF20280443 SF20280444 SF20280445 SF20280446 SF20280447 SF20280448 SF20280449 SF20280450 SF20280451 SF20280452 SF20280453 SF20280454 SF20280455 SF20280456 SF20280457 SF20280458 SF20280459 SF20280460 SF20280461 SF20280462 SF20280463 SF20280464 SF20280465 SF20280466 SF20280467 SF20280468 SF20280469 SF20280470 SF20280471 SF20280472 SF20280473 SF20280474 SF20280475 SF20280476 SF20280477 SF20280478 SF20280479 SF20280480 SF20280481 SF20280482 SF20280483 SF20280484 SF20280485 SF20280486 SF20280487 SF20280488 SF20280489 SF20280490 SF20280491 SF20280492 SF20280493 SF20280494 SF20280495 SF20280496 SF20280497 SF20280498 SF20280499 SF20280500 SF20281301 SF20281302 SF20281303 SF20281304 SF20281305 SF20281306 SF20281307 SF20281308 SF20281309 SF20281310 SF20281311 SF20281312 SF20281313 SF20281314 SF20281315 SF20281316 SF20281317 SF20281318 SF20281319 SF20281320 SF20281321 SF20281322 SF20281323 SF20281324 SF20281325 SF20281326 SF20281327 SF20281328 SF20281329 SF20281330 SF20281331 SF20281332 SF20281333 SF20281334 SF20281335 SF20281336 SF20281337 SF20281338 SF20281339 SF20281340 SF20281341 SF20281342 SF20281343 SF20281344 SF20281345 SF20281346 SF20281347 SF20281348 SF20281349 SF20281350 SF20281351 SF20281352 SF20281353 SF20281354 SF20281355 SF20281356 SF20281357 SF20281358 SF20281359 SF20281360 SF20281361 SF20281362 SF20281363 SF20281364 SF20281365 SF20281366 SF20281367 SF20281368 SF20281369 SF20281370 SF20281371 SF20281372 SF20281373 SF20281374 SF20281375 SF20281376 SF20281377 SF20281378 SF20281379 SF20281380 SF20281381 SF20281382 SF20281383 SF20281384 SF20281385 SF20281386 SF20281387 SF20281388 SF20281389 SF20281390 SF20281391 SF20281392 SF20281393 SF20281394 SF20281395 SF20281396 SF20281397 SF20281398 SF20281399 SF20281400 SF20290401 SF20290402 SF20290403 SF20290404 SF20290405 SF20290406 SF20290407 SF20290408 SF20290409 SF20290410 SF20290411 SF20290412 SF20290413 SF20290414 SF20290415 SF20290416 SF20290417 SF20290418 SF20290419 SF20290420 SF20290421 SF20290422 SF20290423 SF20290424 SF20290425 SF20290426 SF20290427 SF20290428 SF20290429 SF20290430 SF20290431 SF20290432 SF20290433 SF20290434 SF20290435 SF20290436 SF20290437 SF20290438 SF20290439 SF20290440 SF20290441 SF20290442 SF20290443 SF20290444 SF20290445 SF20290446 SF20290447 SF20290448 SF20290449 SF20290450 SF20290451 SF20290452 SF20290453 SF20290454 SF20290455 SF20290456 SF20290457 SF20290458 SF20290459 SF20290460 SF20290461 SF20290462 SF20290463 SF20290464 SF20290465 SF20290466 SF20290467 SF20290468 SF20290469 SF20290470 SF20290471 SF20290472 SF20290473 SF20290474 SF20290475 SF20290476 SF20290477 SF20290478 SF20290479 SF20290480 SF20290481 SF20290482 SF20290483 SF20290484 SF20290485 SF20290486 SF20290487 SF20290488 SF20290489 SF20290490 SF20290491 SF20290492 SF20290493 SF20290494 SF20290495 SF20290496 SF20290497 SF20290498 SF20290499 SF20290500 SF20300501 SF20300502 SF20300503 SF20300504 SF20300505 SF20300506 SF20300507 SF20300508 SF20300509 SF20300510 SF20300511 SF20300512 SF20300513 SF20300514 SF20300515 SF20300516 SF20300517 SF20300518 SF20300519 SF20300520 SF20300521 SF20300522 SF20300523 SF20300524 SF20300525 SF20300526 SF20300537 SF20300538 SF20300539 SF20300540 SF20300541 SF20300542 SF20300543 SF20300544 SF20300545 SF20300546 SF20300547 SF20300548 SF20300549 SF20300550 SF20300551 SF20300552 SF20300553 SF20300554 SF20300555 SF20300556 SF20300557 SF20300558 SF20300559 SF20300560 SF20300561 SF20300562 SF20300563 SF20300564 SF20300565 SF20300566 SF20300567 SF20300568 SF20300569 SF20300570 SF20300571 SF20300572 SF20300573 SF20300574 SF20300576 SF20300577 SF20300578 SF20300579 SF20300580 SF20300581 SF20300582 SF20300583 SF20300584 SF20300585 SF20300586 SF20300587 SF20300588 SF20300589 SF20300590 SF20300591 SF20300592 SF20300593 SF20300594 SF20300595 SF20300596 SF20300597 SF20300598 SF20300599 SF20300600 SF20310401 SF20310402 SF20310403 SF20310404 SF20310405 SF20310406 SF20310407 SF20310408 SF20310409 SF20310410 SF20310411 SF20310412 SF20310413 SF20310414 SF20310415 SF20310416 SF20310417 SF20310418 SF20310419 SF20310420 SF20310421 SF20310422 SF20310423 SF20310424 SF20310425 SF20310426 SF20310427 SF20310428 SF20310429 SF20310430 SF20310431 SF20310432 SF20310433 SF20310434 SF20310435 SF20310436 SF20310437 SF20310438 SF20310439 SF20310440 SF20310441 SF20310442 SF20310443 SF20310444 SF20310445 SF20310446 SF20310447 SF20310448 SF20310449 SF20310450 SF20310451 SF20310452 SF20310453 SF20310454 SF20310455 SF20310456 SF20310457 SF20310458 SF20310459 SF20310460 SF20310461 SF20310462 SF20310463 SF20310464 SF20310465 SF20310466 SF20310467 SF20310468 SF20310469 SF20310470 SF20310471 SF20310472 SF20310473 SF20310474 SF20310475 SF20310476 SF20310477 SF20310478 SF20310479 SF20310480 SF20310481 SF20310482 SF20310483 SF20310484 SF20310485 SF20310486 SF20310487 SF20310488 SF20310489 SF20310490 SF20310491 SF20310492 SF20310493 SF20310494 SF20310495 SF20310496 SF20310497 SF20310498 SF20310499 SF20310500 SF20310607 SF20310608 SF20310609 SF20310610 SF20310611 SF20310612 SF20310613 SF20310614 SF20310615 SF20310616 SF20310617 SF20310618 SF20310619 SF20310620 SF20310621 SF20310622 SF20310623 SF20310624 SF20310625 SF20310626 SF20310627 SF20310628 SF20310629 SF20310630 SF20310631 SF20310632 SF20310633 SF20310634 SF20310635 SF20310636 SF20310637 SF20310638 SF20310639 SF20310640 SF20310641 
Recalling Firm/
Manufacturer
Covidien Llc
15 Hampshire St
Mansfield MA 02048-1113
For Additional Information ContactTechnical Service
800-255-6774
Manufacturer Reason
for Recall
The firm has released software update V01.05.02.16 (also known as V1.5.2) in response to customer reports of a false display of the message, "Temperature Exceeds Limits" followed by automatic shutdown of the monitor with no accompanying alarm.
FDA Determined
Cause 2
Software design
ActionOn November 16, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed of the software issue affecting devices. Medtronic has developed a software update which includes a fix for this issue, available at http://medtronic.com/cap35software. Medtronic has also issued an addendum to the Capnostream35 Portable Respiratory Operator's Manual which can be accessed at manuals.medtronic.com. Customers are to take the following actions: - Share the notification of the recall with all care environments where the Capnostream35 portable respiratory monitors are used. If your facility has distributed these portable respiratory monitors to other persons or facilities, please promptly forward a copy of this letter to those recipients. -Access and download software version V1.5.2 from the Medtronic website via the following link which includes directions on how to download and install the software on all devices at your facility. http://medtronic.com/cap35software -Return the completed Acknowledgement and Receipt Form by fax or email even if you have no inventory. If you are aware of any incidents related to this issue or if you have any questions, please contact our Technical Service Department at 800.255.6774, option 1, or rs.hqtsweb@medtronic.com to provide information regarding those events so regulatory reporting obligations can be fulfilled.
Quantity in Commerce3217
DistributionWorldwide distribution - US Nationwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = CCK
-
-