|
Class 2 Device Recall Philips Respironics V60 Ventilator |
|
Date Initiated by Firm |
October 30, 2020 |
Create Date |
January 22, 2021 |
Recall Status1 |
Completed |
Recall Number |
Z-0886-2021 |
Recall Event ID |
86646 |
510(K)Number |
K082660
|
Product Classification |
Ventilator, continuous, minimal ventilatory support, facility use - Product Code MNT
|
Product |
Philips Respironics V60 Ventilator Part Number 1053614 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories. |
Code Information |
Model: V60 Ventilator, Part Number 1053614 V60 Ventilator, Intl Opt: CFLEX,AVAPS Serial Numbers: 100007217 100021204 100021286 100021400 100029451 100042750 100042939 100043085 100055703 100056018 100062163 100062178 100063153 100063155 100063162 100063293 100063300 100069697 100069744 100069860 100070577 100070772 100132058 100133634 100144871 100144874 100144879 100144961 100144982 100162731 100162792 100169641 100195675 100219970 100219972 100219975 100219999 100220001 100221150 100239619 100243227 100246517 100247269 100252596 100252597 100252600 100252764 100252768 100252771 100252772 100252773 100252797 100253477 100253484 100253492 100253493 100253494 100253500 100254739 100286208 100286300 100311170 100317381 100317392 100318233 100318405 100318406 100318408 100318413 100318415 100318416 100318426 100318615 100318623 100336755 100336849 100336855 100336889 100336895 100336900 100337085 100337258 100337260 100337261 100343560 100344202 100344235 100350181 100350684 100350711 100350714 100350914 100351156 100351157 100352118 100352139 100352147 100357135 100357137 100357139 100357454 100357456 100357458 100357459 100357733 100357797 100358125 100358129 100358131 100358134 100358149 100358151 100358154 100358155 100358162 100358164 100358198 100358302 100358304 100358319 100358337 100362568 100362701 100363015 100363053 100364545 100364554 100364602 100365186 100370897 100373551 100375010 100375148 100375226 100375243 100375267 100375280 100375282 100375283 201002610 201002611 201002976 201023090 201025859 201025954 201025966 201028012 201028033 201028528 201028889 201028898 201028988 201028990 201029030 201029045 201029138 201029937 201032702 201032705 201032707 201032713 201032718 201032733 201032831 201032863 201032871 100350289 100350298 100350670 100351154 100351541 100351555 100351712 100352113 100352824 100352946 100358347 100361419 100362450 100362575 100362703 100362721 100362745 100362746 100363013 100363030 100364252 100364503 100364521 100364532 100364539 100364562 100364571 100364611 100364621 100364624 100365161 100365162 100365164 100365166 100365177 100365183 100368067 100368558 100368607 100368610 100368723 100368883 100369414 100369417 100369424 100369425 100369436 100373445 100375009 100375149 100375167 100375174 100375178 100375233 100375241 100375262 100375268 100375284 100375290 100375293 100375298 100375385 100375396 100376391 100377940 100378200 100378203 100378210 100378214 100378216 100378221 100378283 100378294 100378442 100378443 100378444 100378880 100378892 100378939 100380812 100027294 100061625 100061627 100061638 100061640 100061641 100084756 100120124 100120850 201003948 201009965 201009974 201032704 201032710 201032725 201032729 201032732 100003563 100004666 100004675 100004678 100004680 100004682 100004794 100004925 100006184 100007337 100007347 100010849 100010903 100010970 100011091 100011194 100011195 100011196 100011198 100020662 100022039 100022051 100022143 100022149 100022156 100022242 100029333 100036292 100036295 100042755 100042758 100042940 100044361 100044523 100046904 100050526 100052770 100055815 100055827 100055949 100061645 100061646 100061800 100063157 100063159 100068212 100069748 100069869 100069956 100075310 100082432 100082437 100082464 100092493 100095598 100095651 100101751 100101759 100101760 100110676 100110939 100122334 100122348 100132026 100132038 100132040 100132045 100132051 100133642 100148954 100162739 100162898 100176025 100184198 100189658 100195894 100219977 100230168 100233962 100252963 100253497 100254783 100260028 100260973 100261125 100261128 100261133 100261274 100261276 100266756 100266759 100266760 100266761 100266762 100266763 100266764 100266765 100266769 100266770 100267525 100267526 100267530 100267560 100277907 100277915 100284354 100285014 100286315 100378291 100378296 201000741 201000886 201001542 201003492 201004663 201009219 201009463 201009470 201022987 201022991 201025947 201025972 201025996 201026002 201029036 201029659 201029876 201029952 201029953 201031680 201033647 100003256 100003311 100004748 100005812 100005814 100005902 100005983 100006103 100006113 100006172 100006173 100006174 100006176 100006185 100007117 100007119 100007191 100007197 100007209 100007334 100007341 100007414 100007418 100007422 100007536 100007539 100007543 100007768 100007769 100007770 100007771 100007776 100007780 100007781 100007782 100007786 100007879 100007882 100007888 100010702 100010703 100010707 100010708 100010709 100010710 100010712 100010720 100010851 100010853 100010902 100010904 100010905 100010911 100011043 100013961 100013965 100014010 100014469 100014489 100014516 100014772 100014773 100014775 100014776 100014778 100014822 100014823 100014824 100014825 100014826 100014827 100014828 100014830 100014896 100014897 100014899 100015329 100015330 100015331 100015334 100015337 100017355 100019389 100019394 100019397 100019398 100019722 100019828 100019832 100019835 100019974 100020049 100020056 100020058 100020062 100020064 100020114 100020125 100020169 100020172 100020173 100020651 100020653 100021172 100021402 100022243 100024705 100024710 100024712 100024878 100024879 100024880 100024883 100024885 100024888 100024889 100024894 100024896 100024897 100024900 100024901 100026665 100026666 100026673 100026674 100026680 100026684 100026758 100026759 100026763 100029455 100029461 100029470 100030106 100030107 100030108 100030109 100030110 100030111 100030112 100030113 100030114 100030115 100030116 100030117 100030118 100030120 100030121 100030123 100030125 100030128 100030129 100030130 100030132 100030135 100053050 100053788 100053793 100053794 100053795 100053796 100053797 100053855 100053856 100053859 100053860 100053865 100054152 100054153 100054154 100054156 100054157 100054158 100054159 100054160 100054400 100054401 100054402 100054404 100054405 100057189 100057190 100057191 100057192 100057193 100057194 100057196 100057197 100057198 100057199 100057200 100057201 100057202 100057205 100057206 100059520 100059522 100059525 100059526 100059527 100059530 100059532 100059533 100059534 100059535 100060107 100060108 100060113 100060114 100060116 100060119 100060122 100060212 100072494 100072564 100072565 100072571 100078121 100078123 100078126 100078128 100078129 100081404 100081409 100081487 100081490 100081494 100081501 100081502 100081683 100081684 100081685 100081693 100081699 100081702 100081703 100081705 100081706 100081707 100081708 100081709 100081710 100081712 100081714 100081715 100081717 100081748 100081749 100081750 100081751 100081752 100084347 100084348 100084349 100084350 100084352 100084353 100084385 100084387 100084388 100084389 100084390 100084391 100084392 100084393 100084396 100084397 100084398 100084560 100084562 100084564 100084565 100084567 100084573 100089664 100097993 100098080 100098081 100098082 100098083 100098084 100098086 100098088 100098091 100102164 100110656 100110671 100110672 100110677 100110678 100110680 100110843 100110844 100110846 100110847 100110848 100110928 100110929 100110930 100110937 100110941 100110942 100110944 100110946 100110949 100110950 100110952 100110953 100110990 100110991 100131918 100132100 100132533 100133556 100133559 100133620 100133621 100133622 100133623 100133625 100133628 100133629 100133631 100142323 100142326 100142327 100142329 100142330 100142331 100142332 100142336 100142338 100142362 100142363 100142364 100142365 100142370 100142378 100142382 100142383 100142384 100147571 100148958 100148962 100149203 100149208 100149439 100149700 100149701 100149702 100149703 100149704 100149705 100149706 100149707 100149708 100149709 100149710 100149711 100149712 100149713 100149715 100149717 100149718 100149824 100149829 100149830 100149832 100149834 100149840 100158085 100159227 100159229 100159230 100159231 100159232 100159233 100159237 100159303 100159331 100159333 100159336 100159338 100159339 100159569 100159570 100159571 100159572 100159573 100159574 100159597 100159598 100159599 100159640 100159643 100159644 100159645 100159647 100159650 100159679 100169064 100169066 100169068 100169069 100169070 100169072 100169073 100169080 100169084 100169085 100169090 100169094 100169096 100169099 100169230 100176325 100176888 100178915 100178917 100179381 100179387 100179389 100179390 100179391 100179392 100179394 100179395 100179396 100179397 100179399 100179400 100179401 100179402 100179403 100179404 100179405 100179406 100179409 100179413 100179415 100179416 100179417 100179418 100179419 100179464 100179465 100179468 100180772 100180944 100180960 100180962 100180965 100180969 100181203 100181218 100181219 100181230 100181234 100181238 100181239 100181240 100181242 100181585 100181588 100181589 100181595 100181596 100181598 100181602 100181603 100181607 100181617 100181735 100181736 100181737 100181742 100181744 100181746 100181883 100181884 100181885 100181889 100181891 100181892 100181895 100181897 100181899 100182456 100184158 100184183 100184186 100184187 100184188 100184189 100184190 100184192 100184637 100186769 100186776 100186778 100186780 100186786 100186788 100186791 100186793 100186795 100190822 100193926 100193928 100193929 100193931 100193932 100193933 100193934 100193936 100193937 100193938 100193939 100194160 100194161 100194162 100194163 100194165 100194166 100194167 100194168 100228613 100228923 100228992 100229109 100229113 100237239 100237310 100237627 100239253 100239297 100239403 100239406 100239412 100239529 100239530 100239531 100247487 100248619 100252576 100252582 100252583 100252584 100252586 100252587 100252589 100252590 100252621 100252623 100252624 100252625 100252626 100252798 100256005 100256842 100256847 100256851 100256855 100256857 100256981 100257206 100257210 100257214 100257217 100257218 100257266 100257310 100257317 100257389 100257390 100257391 100257392 100257393 100257394 100257396 100257399 100257401 100257405 100257406 100257582 100257586 100257590 100257593 100257692 100257816 100257922 100270002 100270278 100270287 100270290 100270291 100270293 100270295 100271022 100271027 100271030 100271032 100271034 100271035 100271036 100271037 100271038 100271180 100271182 100271184 100271185 100277906 100278781 100278792 100278797 100278804 100278805 100278806 100278810 100278811 100278812 100278813 100278996 100278997 100278998 100278999 100279000 100279001 100279002 100279005 100279006 100279007 100279008 100279009 100279463 100279464 100279466 100279470 100279471 100282834 100282838 100284345 100284346 100284350 100284351 100284352 100284355 100284357 100284358 100284361 100286709 100286712 100286940 100286943 100286949 100286950 100286952 100292344 100341392 100346921 100346925 100346927 100346932 100350137 100350164 100350166 201001440 201001441 201001442 201001443 201001447 201001449 201001450 201001451 201001452 201001524 201001526 201001527 201001528 201001530 201001532 201001533 201001535 201001536 201001538 201001865 201002218 201002696 201004105 201007508 201007529 201007531 201007532 201007533 201007534 201007583 201007584 201007586 201007587 201007593 201007598 201007599 201007602 201007603 201007608 201007621 201007623 201015416 201015417 201015427 201015432 201015434 201015482 201015520 201015521 201015524 201015544 201015555 201028883 201028890 201028893 201028899 201029023 201029035 201029040 201029049 201029051 201029052 201029102 201029105 201029108 201029115 201029120 201029124 201029130 201029149 201029154 201029155 201029163 201029165 201029878 201029901 201029910 201029914 201029921 201029939 201029947 201029949 201029960 201029961 201029967 201029974 201032692 201032696 201032698 201032701 201032716 201032720 201032721 201032723 201032731 201032829 201032832 201032861 201032864 201032875 201034074 100005823 100005825 100006177 100007340 100007533 100007540 100007541 100007774 100007885 100007886 100010713 100010714 100010852 100010973 100013343 100013962 100014014 100019392 100021292 100021293 100021318 100021405 100029332 100037553 100039161 100039771 100039782 100039841 100039968 100039969 100039970 100039971 100043948 100053780 100070760 100070761 100070762 100070763 100087485 100092872 100092873 100092877 100092878 100092883 100101731 100101748 100101750 100101851 100105073 100105241 100105249 100105568 100105570 100105573 100105681 100110664 100110927 100110932 100110933 100110934 100110936 100114033 100114043 100114044 100114050 100114450 100114464 100114518 100114523 100114530 100135299 100136187 100139918 100139919 100160694 100160830 100163733 100166087 100166088 100166096 100166103 100166106 100166107 100166116 100166117 100166120 100169562 100170683 100170693 100170699 100171031 100171033 100175837 100176851 100176874 100176880 100176881 100176885 100182399 100182421 100182430 100182444 100195878 100219989 100219992 100226363 100227090 100227098 100232756 100232767 100237196 100237306 100239536 100243239 100250369 100254738 100254749 100255548 100255556 100256843 100256854 100260548 100261111 100261122 100261176 100261177 100261278 100267694 100267702 100274591 100274693 100274699 100278786 100285140 100286957 100292305 100317499 100317502 100329962 100329964 100348150 201000890 201000910 201002110 201002392 201002422 201002614 201009206 201009210 201009212 201010799 201022978 201023858 201026449 201026531 201026645 201026658 201026668 201029133 201029753 201029755 201029803 201029954 201031688 201032009 201032256 201032840 100031152 100050512 100050515 100096203 100096205 100100857 100101940 100135542 100261169 100267535 100267562 100267563 100267565 100267571 100274698 100282877 100291854 100292299 100292303 100292304 100292309 100292348 100292353 100292355 201001522 201002615 201002616 201002681 201002692 201002695 201002697 201002702 201002703 201002704 201002706 201002992 201013651 201014003 201014004 201014014 201014015 201014017 201014043 201014049 100046912 100047064 201000659 100064511 100080826 100080830 100080831 100080834 100080842 100094147 100098003 100098463 100098466 100098469 100099138 100104052 100104062 100104064 100105064 100105066 100105227 100105243 100118531 100118539 100118541 100118543 100118560 100118668 100118671 100118673 100133633 100133635 100133638 100133641 100138564 100138583 201003338 201003339 201003362 201003365 201003366 201003367 201003368 201003370 201003371 201003374 201003375 201003380 201003382 201003383 201003385 201003387 201003400 201003402 201003403 201003404 201003406 201003407 201003409 201003416 201003420 201004389 201004390 201008000 201008121 100005892 100005981 100005984 100006111 100006170 100007126 100007194 100007411 100007535 100011041 100011126 100011188 100013344 100013937 100014016 100014021 100014023 100014025 100014029 100014478 100014485 100014488 100014520 100014522 100014525 100014771 100015327 100015345 100015381 100015385 100015386 100015388 100015794 100015800 100016616 100016925 100016926 100017357 100017566 100017571 100019642 100019926 100020059 100020119 100020120 100020165 100020654 100020657 100021206 100021303 100021305 100021319 100021391 100021684 100021688 100022041 100022046 100022244 100022246 100024699 100024899 100026761 100027959 100027960 100027961 100027970 100027971 100027973 100028689 100029060 100029318 100032694 100032696 100032703 100032705 100033895 100037549 100037552 100038185 100038197 100038198 100038293 100038298 100039845 100039847 100042938 100042944 100042946 100042961 100043106 100045543 100045547 100045550 100045556 100045560 100045561 100045564 100045565 100045567 100045667 100045669 100045672 100045674 100045679 100045680 100046424 100046559 100046740 100046901 100046919 100046925 100046929 100046930 100046931 100047055 100047056 100047061 100047062 100047066 100050369 100050382 100050527 100051958 100051963 100052390 100053070 100054533 100054824 100054832 100054835 100055350 100055704 100055955 100056021 100057213 100057363 100057375 100057377 100057378 100060497 100061924 100062173 100062283 100063177 100063180 100063295 100063296 100064708 100064829 100064834 100064963 100065496 100065500 100065503 100065506 100065510 100065596 100065599 100065604 100065606 100065607 100065608 100068367 100068370 100069512 100069513 100069514 100069515 100069610 100070139 100070240 100070243 100070295 100070569 100070923 100070974 100071393 100071396 100071397 100071399 100071867 100072030 100072115 100072120 100072127 100072297 100072298 100072299 100072485 100072490 100072491 100072562 100072678 100072681 100072686 100072696 100073247 100074197 100074199 100074201 100074209 100075408 100078124 100078125 100078269 100078360 100078364 100078368 100078370 100078374 100078967 100079727 100080426 100080559 100080567 100080649 100080825 100080828 100080833 100080853 100081491 100081687 100081690 100081692 100081696 100081704 100082430 100082465 100083573 100084337 100084340 100084341 100084561 100084571 100084574 100084822 100084824 100086894 100089139 100089449 100089539 100089643 100089806 100089809 100091009 100091926 100092186 100092505 100092507 100092646 100092870 100092887 100093331 100094392 100095659 100095661 100095797 100096209 100096548 100097134 100097296 100097986 100098089 100098090 100098470 100099031 100099142 100099147 100099150 100099152 100099153 100099154 100099155 100099156 100099157 100101732 100101737 100101738 100101740 100101741 100101744 100101747 100101749 100101753 100101755 100101758 100101958 100103674 100104060 100104061 100104063 100104066 100104067 100104069 100104077 100104079 100104634 100104636 100105075 100105238 100105245 100105554 100105562 100105569 100105572 100105682 100105685 100105876 100106518 100106519 100107010 100107035 100107133 100107139 100107141 100107143 100107146 100107150 100107151 100107295 100107309 100107313 100107314 100107438 100107442 100107457 100109497 100115853 100116626 100116633 100116699 100118262 100118263 100118353 100118355 100120772 100120773 100120849 100120859 100120862 100122335 100122336 100122338 100122340 100122344 100122574 100122576 100122578 100122579 100124871 100125174 100125728 100125744 100129774 100129777 100129788 100129790 100129793 100129794 100132214 100132408 100133553 100133557 100133624 100133627 100133637 100135541 100135543 100135545 100135546 100136164 100136165
100136170 100136181 100136621 100136625 100138044 100138051 100138232 100138281 100138285 100139901 100139908 100139932 100140621 100140623 100144403 100144951 100144966 100144968 100144977 100145934 100145962 100147115 100147203 100148960 100149826 100149837 100150842 100152327 100152551 100152559 100152660 100152662 100153192 100153198 100153465 100153466 100153471 100153473 100153474 100153475 100153482 100157751 100157752 100157754 100157756 100157758 100157759 100158068 100158075 100159831 100159832 100159931 100160003 100160834 100160838 100162292 100162414 100162783 100164095 100164426 100164427 100164430 100164435 100164436 100166098 100166114 100169077 100169078 100170701 100171028 100172275 100173137 100173210 100173458 100174292 100174297 100174318 100174453 100174459 100174463 100174464 100176019 100176031 100176034 100176035 100176036 100176042 100176043 100176319 100176323 100176331 100176844 100176848 100176861 100176862 100176863 100176864 100176865 100176867 100176869 100176870 100176871 100176878 100176879 100176883 100179382 100179388 100179393 100179407 100180423 100180427 100181224 100181226 100181245 100182152 100182432 100186807 100187427 100189488 100189981 100190085 100193927 100193930 100194164 100237069 100237377 100237560 100238419 100239247 100239254 100239539 100239604 100239618 100239811 100240173 100240179 100247082 100247091 100247096 100248608 100248618 100251178 100251524 100252562 100252565 100252566 100252567 100252568 100252569 100252571 100252573 100252577 100252593 100252601 100252602 100252942 100253488 100253720 100254412 100254746 100254788 100257179 100257270 100259853 100259864 100260711 100260971 100261113 100261320 100272772 100274764 100277896 100278748 100280583 100284367 100286682 100298935 100300527 100300540 100300663 100301393 100304091 100312444 100312448 100313063 100316916 100323022 100333561 100333578 100337098 100337236 100343540 100345544 100345585 100345589 100345592 100345618 100345623 100346885 100346888 100346897 100346939 100346952 100346956 100346958 100346960 100346964 100346965 100346973 100346978 100346981 100348092 100348109 100348116 100348120 100348152 100348154 100348155 100348156 100348157 100348159 100348174 100348441 100350136 100350140 100350141 100350149 100350173 100350177 100350188 100350189 100350190 100350191 100350285 100350286 100350288 100350290 100350294 100350295 100350296 100350660 100350661 100350671 100350673 100350676 100350698 100350722 100350724 100350727 100350728 100350737 100350738 100350740 100350752 100350756 100350880 100351556 100352790 100354015 100354017 100354019 100354021 100354024 100354027 100355170 100355852 100356016 100356200 100356214 100356247 100356297 100356307 100356772 100356773 100357750 100357771 100357772 100357773 100357781 100357783 100357789 100357801 100358138 100358142 100358167 100359506 100359512 100359516 100359517 100359611 100359616 100359628 100359640 100359668 100360081 100360087 100360143 100360214 100360286 100360347 100360374 100361201 100361202 100361234 100361392 100361400 100361507 100361967 100361968 100362556 100362709 100362713 100363025 100363033 100363034 100363760 100363767 100363777 100363786 100363800 100364525 100364528 100364548 100364574 100364604 100364608 100364620 100364622 100365112 100365644 100365645 100365738 100366933 100367275 100367303 100367309 100368012 100368040 100368182 100368531 100368532 100368536 100368547 100368548 100368553 100368600 100368616 100369430 100369432 100369433 100369434 100369447 100370797 100370835 100370859 100370885 100370898 100370962 100370983 100370984 100370989 100371103 100371124 100371133 100371136 100371137 100371138 100371145 100371214 100371375 100371410 100371412 100371418 100371650 100371661 100371665 100371667 100371668 100371669 100371670 100371702 100371705 100372897 100373205 100373207 100373220 100373221 100373223 100373233 100373453 100373543 100373545 100373546 100373547 100373548 100373550 100373555 100373607 100375008 100375141 100375147 100381925 100383360 100384227 100387882 100388081 100395078 201000658 201000743 201000883 201000901 201000905 201001113 201001148 201001446 201001448 201001548 201002044 201002108 201002215 201002261 201002395 201002995 201002999 201003000 201003002 201003051 201003479 201003490 201003493 201004340 201004364 201004585 201004646 201007460 201007461 201007473 201007482 201007506 201007518 201007519 201007522 201007525 201007835 201008120 201008125 201008138 201008169 201008186 201008190 201008200 201008201 201008209 201008228 201009869 201009971 201010041 201010048 201010056 201010314 201010373 201010567 201010809 201010825 201011247 201012777 201013081 201013087 201013168 201013652 201014959 201014961 201014964 201014965 201014984 201014985 201014994 201014995 201015000 201015001 201015060 201015062 201015063 201015545 201017070 201025266 201026252 201026254 201026431 201026433 201026434 201026435 201026436 201026439 201026569 201026577 201026579 201026598 201026601 201026605 201026636 201026639 201026640 201026642 201026646 201026649 201026651 201026652 201026653 201026655 201026659 201026662 201026665 201026667 201026672 201026673 201026680 201026685 201026686 201026687 201026689 201026690 201026691 201026692 201026693 201026694 201026695 201026699 201026700 201026701 201026702 201026705 201026706 201026708 201026709 201026710 201026711 201026712 201026715 201026716 201026717 201026741 201026742 201026744 201026748 201026749 201026753 201026755 201026765 201026782 201033688 201033759 201033788 201033854 201033856 201033858 201033859 201034009 201034023 201034036 201034038 201034039 201034069 201034070 201034071 201034075 201034080 201034081 201034082 201034083 201034085 201034086 201034087 201034088 201034089 100006171 100024887 100125971 100138565 100138566 100138567 201000755 201000909 100017346 100017562 100017564 100039848 100044210 100044521 100044528 100044534 100044542 100045539 100045540 100045541 100045542 100045548 100050522 100050534 100050536 100064702 100064716 100064839 100064840 100064841 100064842 100064843 100064844 100064845 100064847 100064848 100064849 100064850 100064851 100064852 100089137 100089452 100089454 100089455 100089456 100089457 100089458 100089536 100089544 100089638 100089639 100089640 100089642 100089645 100089647 100089654 100089658 100089659 100089660 100089661 100089662 100089663 100089665 100089666 100089807 100089808 100089811 100089816 100107310 100107443 100115407 100115408 100115411 100115413 100115414 100115415 100115416 100115804 100115840 100115847 100116371 100116373 100116382 100116384 100116385 100116386 100116387 100116388 100116389 100116390 100116607 100116629 100116632 100116694 100116695 100129369 100129789 100137674 100137675 100137678 100138043 100138045 100138047 100138048 100138057 100138227 100138228 100138234 100138235 100138237 100138239 100138280 100138282 100138380 100138586 100142328 100142333 100142335 100142337 100142339 100142360 100142361 100142366 100142367 100142369 100142371 100142372 100142373 100142374 100142375 100142376 100142377 100142379 100142380 100142381 100144387 100144496 100144499 100144529 100144530 100176854 100179649 100180189 100180197 100180201 100180202 100180206 100180207 100180209 100180210 100180212 100180213 100180214 100180215 100180217 100180218 100180219 100180222 100180223 100180224 100180226 100180229 100180411 100181582 100181597 100181613 100181620 100181747 100181748 100181894 100182389 100182397 100182422 100182423 100182428 100182437 100182438 100182458 100182459 100182462 100194170 100195464 100195467 100195639 100195642 100195873 100248580 100251173 100251176 100252572 100252574 100252603 100252605 100252607 100252608 100252610 100252611 100252612 100252613 100252618 100252619 100259784 100259792 100259794 100259858 100259862 100260180 100260715 100261168 100261171 100291730 100291852 100292315 100292317 100292319 100292341 100292343 100292346 100292349 100292350 100293517 100293518 100293519 100293600 100293607 100293613 100293614 100298641 100300517 100300553 100309411 100309413 100309724 100309739 100310995 100310996 100310998 100310999 100311160 100311165 100314715 100315158 100315186 100315362 100315364 100316915 100317268 100317394 100318226 100318234 100318306 100318316 100318427 100318594 100318600 100318601 100318602 100319756 100319766 100319768 100319798 100320044 100320047 100320049 100320050 100320051 100320052 100320142 100320361 100350145 100350170 100350175 100350176 100350668 100350672 100350678 100350679 100350685 100350689 100350694 100350695 100350703 100350709 100350759 100361425 100361438 100361450 100361454 100361480 100361538 100361547 100361560 100361582 100361583 100361584 100361756 100361776 100362056 100362064 100362398 100391878 100391897 100391961 100391982 100392139 100392279 201001541 201001546 201001549 201001550 201001551 201001552 201001553 201001554 201001555 201001556 201002275 201002278 201002381 201002382 201002383 201002384 201002385 201002387 201002388 201002389 201002390 201002391 201002393 201002396 201002398 201002937 201002980 201002987 201002997 201003001 201003003 201003004 201003005 201003006 201003007 201003418 201004087 201004088 201004089 201004091 201004094 201007442 201007447 201007449 201007451 201007538 201007582 201007597 201007620 201007622 201007624 201007647 201007648 201007649 201007651 201007652 201007663 201007669 201010739 201013643 201013644 201013646 201013653 201014001 201014002 201014010 201014011 201014012 201014013 201014016 201014040 201014045 201014046 201014048 201022966 201022974 201022979 201022984 201022993 201022995 201022998 201023010 201023011 201023042 201023043 201023045 201023048 201023089 201023098 201023102 201038681 201038684 201038705 201038720 201038729 201038730 201038736 201038740 100025548 100026667 100026668 100026675 100026676 100026677 100026685 100026757 100026762 100026765 100026772 100027093 100027096 100027549 100027553 100027554 100027555 100027556 100027557 100028664 100028666 100029072 100029073 100029210 100029211 100029218 100029220 100029221 100029223 100029241 100029322 100029325 100029337 100029338 100029341 100029342 100029453 100029462 100029474 201002338 201002420 201002423 201002425 100259881 100260021 100036160 100068202 100070576 100070766 100084829 100084830 100100996 100148859 100170696 100190091 100219982 100219983 100226366 100226379 100226381 100234783 100234789 100240152 100243233 100274696 100274980 100277816 100277914 100311779 100314693 100336896 100337240 100338952 100339201 100339233 100343535 100343570 100343576 100343609 100343623 100344363 100360287 100363076 100363740 100369618 100369703 201024190 201031686 201032734 100003075 100003245 100003251 100003312 100003314 100003315 100003316 100003317 100003318 100003319 100003320 100003323 100003324 100003325 100003329 100003330 100003332 100003333 100003335 100003336 100003337 100003364 100003365 100003366 100003367 100003369 100003370 100003371 100003372 100003373 100003374 100003375 100003376 100003377 100003405 100003406 100003407 100003408 100003409 100003410 100003412 100003413 100003414 100003416 100003417 100003418 100003419 100003480 100003482 100003483 100003484 100003609 100004677 100004791 100005816 100005901 100007113 100007195 100007205 100007286 100007335 100007339 100007413 100007416 100007419 100007424 100007538 100007542 100007544 100007783 100007870 100007873 100007878 100007884 100010718 100010719 100010844 100010907 100010910 100010975 100011087 100011092 100011095 100011120 100011189 100011190 100011193 100011197 100011199 100011200 100013933 100013935 100013936 100013938 100013939 100013972 100014017 100014027 100014028 100015324 100015325 100015326 100015328 100015335 100015339 100015344 100016920 100016923 100016931 100016932 100017399 100017468 100017470 100017472 100017473 100017567 100017568 100017672 100017676 100017733 100019925 100020051 100020069 100020070 100020113 100020171 100020650 100020652 100020655 100020659 100020660 100020667 100020749 100021170 100021171 100021175 100021177 100021178 100021200 100021288 100021298 100021301 100021389 100021398 100021403 100036287 100036299 100037557 100039244 100039268 100039762 100039764 100039767 100039769 100039772 100039774 100039778 100045544 100046414 100046417 100046558 100046905 100047059 100047069 100050513 100050524 100050529 100050538 100053785 100053857 100053858 100053861 100053863 100053864 100053866 100053867 100053868 100053869 100055829 100055937 100062053 100062055 100062056 100062057 100062059 100062060 100062061 100062162 100062164 100062236 100069863 100079566 100079710 100079711 100079722 100079725 100079726 100079728 100079729 100079800 100079801 100080645 100080656 100080824 100080832 100080838 100080839 100081120 100081121 100081122 100081123 100092834 100092835 100092838 100092879 100092880 100092881 100092882 100095580 100095589 100095590 100095656 100095747 100095752 100095753 100095754 100095758 100095759 100107019 100107452 100115848 100115851 100116369 100116608 100116616 100116617 100116618 100116625 100116687 100116692 100118342 100118552 100118675 100118681 100118688 100118920 100118927 100118929 100118931 100118932 100125176 100125466 100125468 100125469 100125470 100125478 100125481 100125485 100125486 100125488 100125489 100125490 100125493 100125494 100125714 100125716 100125720 100125723 100125726 100125727 100125729 100125731 100125732 100125736 100125743 100125749 100125751 100125754 100125755 100125761 100125840 100125843 100125952 100132011 100132013 100132053 100132106 100134678 100135544 100135549 100136174 100136184 100136617 100136618 100136620 100136622 100136638 100136639 100136712 100136718 100136762 100136763 100136765 100139679 100139719 100139720 100139722 100139724 100139725 100139726 100139727 100139728 100139729 100139730 100139806 100139896 100139902 100139904 100139905 100139911 100139912 100139916 100139922 100139925 100139930 100139931 100139934 100140587 100140588 100140589 100140590 100140591 100140592 100140594 100140598 100140626 100140629 100140631 100143598 100143599 100143681 100143684 100143688 100143690 100143691 100143692 100143693 100143694 100143695 100143697 100143698 100143699 100143700 100143702 100143703 100143704 100143705 100143706 100143721 100143722 100143723 100143726 100143730 100143732 100143750 100158071 100158072 100158073 100158076 100158077 100158081 100158083 100158087 100158088 100158091 100159696 100159698 100159699 100159830 100159923 100159925 100159929 100160005 100160695 100160823 100162298 100162416 100162732 100162902 100163434 100163584 100163585 100163590 100163712 100163723 100163726 100169644 100169657 100170464 100170466 100170468 100170477 100170563 100170567 100170573 100170576 100170680 100170685 100171034 100172261 100172269 100172270 100174967 100176318 100176324 100176868 100181593 100181896 100182470 100187313 100189451 100189452 100189454 100189458 100189460 100189461 100189469 100189749 100189757 100189804 100189807 100189905 100190006 100190007 100190096 100195463 100195468 100195662 100195672 100195965 100195971 100196127 100196437 100219976 100219996 100223462 100223482 100226369 100226377 100228993 100228995 100228997 100228998 100229063 100229100 100229103 100229105 100229106 100229108 100230166 100231172 100232745 100232751 100232760 100232761 100232762 100232766 100232773 100233054 100233057 100233066 100237223 100237323 100237327 100237370 100237371 100237372 100237376 100237562 100237563 100237621 100239402 100240177 100243226 100243231 100247255 100247267 100247273 100247402 100247405 100247406 100248598 100248602 100248610 100248612 100252765 100252766 100252779 100252780 100252789 100253479 100254792 100257671 100259869 100260956 100261003 100261114 100261120 100261123 100261126 100261127 100261129 100261131 100261136 100261137 100261173 100261275 100270297 100270299 100271723 100271733 100271735 100271736 100271803 100271805 100271808 100271809 100271906 100271908 100271909 100271913 100274274 100274279 100274808 100274818 100274821 100274824 100282829 100282830 100282839 100282862 100282864 100282868 100282873 100282876 100282878 100282879 100284362 100285018 100285799 100286058 100286215 100286464 100286477 100286684 100286685 100286692 100286708 100286969 100286972 100286974 100286975 100290668 100292295 100292296 100292300 100292302 100292306 100292307 100292310 100292311 100292312 100292314 100292316 100292354 100298920 100298931 100329965 100329966 100336759 100339208 100339222 100339228 100339229 100339237 100361417 100361483 100361498 100362706 100362716 100362739 100363657 100364524 100364547 100364607 100368016 100368058 100375219 100378845 100378850 201000882 201000887 201000888 201000906 201000911 201001945 201001949 201001952 201001953 201001954 201001955 201001956 201001957 201001959 201002049 201002050 201002052 201002100 201002103 201002106 201002262 201002295 201002297 201002302 201002304 201002910 201002915 201002917 201002918 201002966 201002977 201002984 201002985 201002986 201002988 201002989 201002991 201002993 201002994 201003042 201003043 201003044 201003046 201004052 201004086 201004095 201004096 201004097 201004098 201004099 201004100 201006540 201006542 201007445 201007588 201007589 201007591 201007592 201007609 201007625 201007627 201007630 201007631 201007636 201007656 201007662 201007664 201007668 201007671 201007678 201007679 201010046 201010274 201010275 201010278 201010369 201010371 201010372 201010570 201010596 201010598 201010607 201012899 201012982 201013080 201013092 201013154 201013766 201013809 201013814 201013815 201013816 201013818 201013819 201013820 201013821 201013822 201013823 201013825 201013826 201013827 201013828 201013829 201022980 201023006 201023053 201023094 201023096 201023832 201023850 201023856 201023871 201024010 201024061 201026525 201026568 201026570 201026573 201026574 201026575 201026576 201026602 201026641 201026647 201029792 201029802 201029808 201029816 201029823 201029825 201029827 201029835 201029836 201029838 201029839 201029841 201029856 201029858 201029861 201029862 201029866 201029911 201029917 201029934 201029943 201029950 201029959 201029963 201029964 201029970 201029972 201031684 201031697 201031753 201031860 201031862 201031863 201031867 201031869 201031870 201031876 201031877 201031880 201032178 201032706 201032711 201032735 201032828 201032833 201032862 201032867 201032873 201038694 201038725 201038733 201038738 100006100 100007547 100007872 100010717 100010842 100010846 100010847 100010848 100010906 100010968 100011048 100022161 100052845 100052849 100052851 100052859 100032462 100032700 100065367 100097276 100097289 100007198 100084339 100094324 100094390 100005980 100005985 100006106 100006114 100006117 100007189 100007192 100007206 100007215 100007281 100007428 100007546 100010971 100011093 100014018 100014777 100014779 100014829 100022040 100022045 100022047 100022157 100024697 100024702 100024706 100024707 100024711 100024893 100027186 100036319 100039964 100043298 100043299 100043301 100043302 100044540 100044543 100046415 100046531 100050519 100051957 100052868 100055938 100055940 100055943 100056019 100061796 100061933 100062042 100062048 100062239 100081760 100082433 100082438 100084758 100092875 100092876 100096380 100105556 100114028 100114037 100114039 100114042 100114045 100114047 100114048 100114049 100114051 100114054 100114461 100114466 100114532 100116378 100116606 100116689 100125737 100125846 100125850 100125972 100132036 100132057 100136167 100136168 100136175 100136634 100136640 100136767 100144867 100144981 100145123 100146856 100146857 100149204 100152333 100160696 100160822 100162784 100162896 100162904 100163725 100170688 100176022 100176040 100176044 100176322 100176327 100176330 100176337 100176338 100176339 100176858 100176859 100176875 100178916 100179385 100184345 100186760 100186764 100186765 100186766 100186767 100186782 100186783 100186790 100189475 100189481 100189755 100189800 100189812 100189819 100189914 100190092 100190731 100195638 100195869 100195898 100219997 100221153 100221157 100221160 100223232 100226357 100226680 100226686 100226870 100228994 100230022 100230026 100230062 100230077 100232742 100232758 100233060 100233966 100233967 100233970 100233978 100234781 100234782 100234787 100234790 100234791 100234792 100234796 100237183 100237190 100237191 100237192 100237194 100237232 100237236 100237300 100237308 100237312 100237321 100237324 100237373 100237375 100238416 100239252 100239534 100239580 100246475 100246507 100246513 100247101 100247102 100247113 100247114 100247260 100247264 100247266 100247271 100247272 100247369 100247401 100247403 100247407 100247410 100247411 100247417 100248604 100248613 100248620 100252965 100254734 100254752 100255626 100256856 100256858 100257670 100257699 100257802 100257812 100257925 100259870 100259874 100259876 100259878 100259879 100260024 100260030 100260037 100260038 100260039 100260045 100261132 100261139 100266767 100266768 100266771 100267527 100267529 100267531 100267532 100267533 100267534 100267536 100267539 100267561 100267567 100271734 100271739 100271802 100271806 100271807 100271905 100274441 100274442 100274443 100274588 100274589 100274590 100274592 100274596 100274762 100274794 100274798 100274981 100277800 100277815 100277890 100277917 100278745 100278747 100278756 100278758 100278759 100278791 100279003 201001539 201007595 201007600 201007601 201007605 201009461 201010757 201010841 201010864 201013790 201013831 201014954 201015202 201015207 201015208 201015422 201022985 201022989 201023083 201023669 201024016 201024075 201024085 201025161 201025162 201025163 201025164 201025165 201025177 201025181 201025183 201025186 201025187 201025188 201025192 201025195 201025201 201025404 201025863 201025869 201025995 201026004 201028896 201028925 201028961 201029020 201029021 201029022 201029658 201029719 201029777 201029779 201029780 201029784 201029810 201029820 201029837 201029846 201029854 201029902 201029905 201029915 201029918 201029923 201029931 201029936 201029938 201029940 201029946 201032243 201032245 201032249 201032712 201032830 201032841 201032874 201033571 201033575 201033594 201033602 201033607 201033609 201033610 201033612 201033613 201033640 201033643 201033644 201033648 201033649 201033651 201033652 201033654 201033655 201033656 201033657 201033659 201033661 201033662 201033664 201033666 201033667 201033668 201033671 201033677 201033701 201034077 100004743 100007778 100007880 100020764 100029336 100039840 100044079 100044517 100044532 100050171 100062043 100063152 100063160 100081755 100082428 100082463 100091022 100096202 100116377 100120131 100125719 100136172 100136173 100136180 100143752 100144866 100144873 100144875 100144955 100145223 100148963 100153195 100153472 100158070 100158089 100160693 100160820 100170478 100170579 100176866 100186763 100189751 100219290 100219291 100221151 100221172 100223467 100223474 100226348 100231067 100233058 100237198 100237309 100246491 100246503 100247107 100247108 100247262 100250362 100251174 100252616 100252617 100253474 100260026 100261134 100267568 100267693 100286309 201000702 201000745 201000893 201002279 201007514 201007642 201007655 201009214 201010835 201013647 201025309 201025864 201025994 201026438 201028911 201029016 201029736 201029757 201029811 201029812 201031757 201032240 201032695 201033675 100002958 100003078 100003079 100003160 100003162 100003163 100003165 100003166 100003201 100003204 100003241 100003242 100003244 100003246 100003248 100003252 100003253 100003255 100003313 100003327 100003378 100003415 100003481 100003562 100003571 100003572 100003607 100003615 100003618 100004667 100004668 100004669 100004672 100004674 100004676 100004679 100005889 100005890 100005891 100005899 100005982 100006110 100006182 100007109 100007120 100007125 100007218 100007348 100007415 100007421 100007875 100007876 100007883 100011192 100013280 100013281 100013283 100013342 100014528 100015789 100015792 100015793 100017348 100021290 100021291 100021300 100021406 100021407 100036288 100039793 100043305 100052860 100053600 100054161 100061635 100061918 100062280 100070575 100087611 100087688 100088050 100088051 100088056 100088057 100090934 100090935 100091011 100091012 100091016 100091017 100091018 100094349 100095597 100101059 100101072 100101076 100101945 100101946 100105563 100114441 100114465 100114520 100115846 100115850 100115852 100115858 100116366 100116604 100116628 100125841 100132043 100132102 100132108 100139808 100139917 100140595 100144877 100144956 100144957 100158084 100158086 100158090 100160691 100160698 100160699 100160835 100162412 100162735 100162738 100162781 100162788 100162789 100162790 100162791 100162899 100162901 100163580 100163592 100163710 100163713 100163736 100165699 100166085 100166101 100166112 100166113 100169563 100169564 100169565 100169642 100169650 100169767 100169770 100169771 100170469 100170569 100172257 100172258 100175815 100176886 100181890 100181893 100182386 100186774 100187426 100187428 100189456 100189457 100189462 100189477 100189478 100189479 100189482 100189490 100189492 100189657 100189752 100189801 100189802 100189803 100189805 100189806 100189810 100195645 100195882 100219274 100219277 100219282 100219283 100219285 100219287 100219304 100219968 100219971 100219974 100219978 100219980 100219981 100219988 100219991 100219994 100221146 100223483 100226354 100226359 100226368 100226371 100226375 100226376 100226378 100226678 100226679 100226682 100228725 100231164 100231166 100232746 100233963 100237561 100239295 100239304 100239386 100239395 100239410 100239538 100239583 100243222 100243634 100243933 100246506 100247259 100247268 100247480 100247481 100248312 100248586 100248605 100252763 100252769 100252770 100252944 100255620 100256852 100256853 100258022 100258032 100259996 100259997 100260020 100260044 100261135 100261322 100266766 100267537 100267538 100267688 100267691 100267695 100267705 100270298 100271191 100271732 100271910 100272767 100274450 100274454 100274594 100274599 100274697 100277903 100277910 100278752 100278784 100284353 100284369 100285012 100286470 100286973 100309651 100318603 201000699 201000721 201000740 201000746 201000751 201000913 201001146 201004597 201007594 201007680 201008124 201008191 201008212 201009208 201010752 201010758 201010826 201010827 201013101 201013817 201015200 201015210 201022999 201024086 201025167 201025173 201025179 201025382 201025402 201025871 201026660 201026671 201026681 201029046 201029116 201029737 201029748 201029754 201029764 201029767 201029768 201029800 201029805 201029809 201029813 201029818 201029828 201029831 201029848 201029864 201029925 201029928 201029930 201031674 201031709 201031758 201032012 201032250 201032432 201032481 201032518 201032699 201032717 201032722 201032827 201032834 201032870 201033599 201033660 201033669 201033741 100003072 100003073 100003076 100003077 100003168 100003200 100003202 100003205 100003240 100003243 100003247 100003250 100003322 100003328 100003331 100003334 100003411 100003560 100003564 100003566 100003567 100003568 100003570 100003575 100003576 100003603 100003606 100003608 100003616 100004610 100004614 100004615 100004616 100004618 100004619 100004670 100004681 100004683 100004741 100004744 100004745 100004789 100004790 100004792 100004793 100004922 100004929 100005807 100005808 100005810 100005811 100005813 100005815 100005817 100005819 100005820 100005821 100005822 100005824 100005826 100005895 100005896 100005897 100005900 100005904 100005905 100005906 100005970 100005972 100005973 100005975 100005976 100005986 100006102 100006104 100006105 100006112 100006116 100006119 100006167 100006169 100006179 100006180 100006181 100007111 100007112 100007116 100007121 100007122 100007124 100007127 100007211 100007785 100007787 100007874 100007877 100007887 100013963 100013967 100014012 100015791 100016922 100017469 100017471 100017570 100017573 100017574 100017675 100017729 100019831 100020746 100020752 100020761 100021304 100021401 100021689 100022038 100023996 100024683 100024696 100024701 100024704 100026097 100027301 100029323 100030222 100031141 100031142 100031145 100031146 100031147 100031149 100031150 100031153 100031157 100031159 100031925 100036436 100038195 100039852 100042757 100045552 100046247 100050516 100052842 100052850 100052862 100052946 100052947 100052949 100053076 100053596 100053605 100053682 100055707 100055939 100056015 100060233 100061928 100061934 100062166 100062169 100062266 100068194 100068208 100068211 100070579 100070759 100070764 100070767 100071388 100071394 100071395 100075369 100075370 100080560 100080837 100081686 100081688 100081694 100081695 100081758 100082429 100082434 100084759 100084821 100084826 100084827 100084833 100087345 100087470 100087689 100092837 100095586 100095587 100095599 100095650 100095652 100095653 100095746 100096307 100096382 100096404 100096541 100104653 100105067 100105070 100105071 100105076 100105078 100105107 100105239 100105250 100105253 100105557 100105558 100105561 100105564 100105565 100105683 100105686 100110674 100110842 100110845 100110938 100110940 100110943 100110947 100114025 100114032 100114035 100114038 100114041 100115849 100116374 100116375 100116611 100116630 100119485 100120117 100120119 100120130 100120132 100120752 100122341 100122349 100122577 100122584 100122590 100122683 100122687 100125484 100125487 100125491 100125735 100131919 100132009 100132010 100132021 100132034 100132039 100132041 100132052 100132054 100132055 100132105 100132107 100135551 100135552 100135553 100136166 100136169 100136171 100136177 100136178 100136179 100136188 100136619 100136629 100136642 100136644 100136710 100136717 100136764 100138041 100138229 100138230 100138236 100138287 100138580 100139819 100139899 100139920 100140596 100144960 100146851 100146855 100147390 100152326 100158057 100158058 100158061 100158064 100158065 100158067 100158074 100158093 100163719 100169763 100170467 100170470 100170473 100170561 100170562 100170566 100170572 100170575 100170577 100170690 100170691 100170692 100170697 100170700 100171030 100171035 100171036 100174177 100174320 100174971 100174979 100174981 100176032 100176321 100176332 100176845 100176872 100176873 100176882 100176887 100181594 100181609 100182146 100182151 100182425 100182453 100186772 100189450 100189453 100189455 100189464 100189754 100189828 100189891 100189915 100189916 100189921 100189923 100190081 100190094 100195311 100195457 100195461 100196430 100219281 100219289 100219987 100221144 100221145 100221148 100221169 100226358 100226364 100226367 100226372 100226373 100226383 100228836 100228996 100229062 100231163 100234667 100234793 100239409 100239411 100239522 100239533 100243513 100243517 100243620 100243622 100243623 100243624 100243625 100243901 100243911 100243931 100247085 100247086 100247092 100247094 100247095 100247258 100247419 100253487 100254785 100254790 100255547 100274080 100274277 100274419 100274445 100274452 100274585 100274761 100284347 100284359 100284363 100284368 100285001 100285003 100285004 100285007 100285144 100286657 100286683 100290816 100291189 100292313 201000697 201000701 201000706 201000720 201000752 201000878 201000892 201000899 201000908 201001521 201001529 201001534 201001537 201001544 201001545 201002399 201002421 201002619 201003488 201004399 201004584 201004594 201004598 201004660 201004661 201004662 201004682 201004686 201004691 201007604 201007635 201009440 201009444 201009467 201010588 201010753 201010756 201010802 201010804 201010810 201010812 201010813 201010832 201010834 201010860 201010862 201010866 201015497 201015522 201015543 201015546 201015552 201022968 201023007 201023044 201024004 201024058 201024063 201024073 201024192 201024197 201025199 201025293 201025303 201025400 201026643 201026648 201026657 201026663 201029741 201029744 201029760 201029763 201029765 201029770 201029773 201029775 201029785 201029788 201029794 201029796 201029797 201029833 201029916 201029942 201032013 201032469 201032697 201032842 201032866 201032869 100095646 100095755 100271188 100003249 100003368 100007118 100007204 100007881 100010706 100010716 100010909 100010969 100010972 100011046 100011094 100011117 100011119 100011121 100011128 100017349 100017490 100017572 100017668 100017669 100017673 100017677 100017678 100017730 100017731 100019616 100021173 100021174 100021176 100021312 100021680 100025545 100025547 100025550 100025552 100027539 100027974 100029219 100029464 100029471 100029472 100035929 100036158 100036300 100036301 100036312 100043096 100043098 100043099 100043100 100043105 100043107 100043108 100043110 100043111 100043112 100043114 100043115 100043292 100043296 100045554 100045559 100045668 100045671 100045673 100045675 100046537 100050198 100050210 100050520 100050523 100050530 100050531 100050532 100050533 100050537 100051079 100051080 100051081 100051082 100051083 100051084 100051085 100051088 100051089 100054816 100065375 100065502 100065504 100065507 100068188 100068190 100068191 100068199 100068200 100068222 100069597 100069599 100069601 100069602 100069606 100072497 100072697 100072734 100072735 100072737 100072738 100073200 100075305 100075311 100075371 100080827 100080829 100080835 100080836 100081411 100081414 100081484 100084563 100084566 100091929 100091931 100091933 100091934 100092968 100092991 100093324 100093338 100095581 100095583 100095592 100095594 100095595 100095645 100097132 100097284 100097285 100097286 100097979 100097988 100097990 100097995 100098085 100098087 100098092 100100190 100100751 100100753 100100754 100100755 100100756 100100757 100100833 100100834 100100835 100103670 100104073 100104628 100104633 100104638 100104645 100105069 100105560 100105566 100105677 100105678 100105680 100105684 100105687 100105688 100105689 100107312 100118074 100118076 100118078 100118082 100118260 100118261 100118265 100120749 100120754 100120767 100129367 100129368 100129371 100129372 100129373 100129374 100129375 100129376 100129377 100129378 100129379 100129440 100129441 100129442 100129443 100129444 100129445 100129446 100129447 100131707 100131708 100131709 100131710 100131711 100131712 100131713 100131714 100131715 100131716 100131717 100131718 100131719 100132012 100132014 100132015 100132016 100132017 100132018 100132019 100132037 100132049 100132059 100132101 100132103 100132104 100132200 100132202 100132203 100132204 100132206 100132207 100132209 100132210 100132211 100132212 100132213 100132215 100132216 100132217 100132218 100132260 100132261 100132262 100132263 100132264 100132265 100132266 100132267 100132268 100132269 100132270 100132271 100132272 100132273 100132274 100132275 100132276 100132277 100132278 100132279 100132280 100132281 100132388 100132389 100132390 100132391 100132392 100132393 100132394 100132395 100132396 100132397 100132398 100132399 100132400 100132401 100132402 100132403 100132404 100132405 100132406 100132407 100132409 100132410 100132411 100132412 100132413 100132414 100132415 100132416 100132417 100132418 100132419 100132420 100132421 100132422 100132423 100132424 100132425 100132426 100132427 100132534 100132535 100132536 100132537 100132538 100132539 100132540 100132541 100132542 100132543 100132544 100132545 100132546 100132547 100132548 100132549 100132550 100132551 100132552 100132553 100132554 100132555 100132556 100132557 100132559 100132560 100132561 100132562 100132563 100132564 100132565 100132566 100132567 100132568 100132643 100133554 100133555 100133558 100137536 100137781 100137838 100137921 100138046 100138049 100138052 100138054 100138059 100138288 100138290 100138291 100138415 100138416 100138419 100152548 100152549 100152550 100152552 100152553 100152554 100152556 100152557 100152558 100152661 100169062 100169063 100169065 100169067 100169075 100169079 100169082 100169083 100169086 100169088 100169089 100169091 100169092 100169093 100169095 100169097 100169098 100169100 100169176 100169178 100169222 100169224 100169225 100169226 100169227 100176023 100182426 100182429 100189742 100189753 100189910 100189913 100189982 100190088 100190090 100190095 100190735 100190738 100190809 100190810 100190811 100190813 100195313 100195319 100195649 100195864 100195896 100221161 100221166 100223234 100231165 100234670 100238418 100243245 100243645 100244189 100244199 100244200 100244208 100244209 100244504 100244583 100244584 100244586 100244587 100244588 100244592 100244593 100244594 100244596 100244600 100252962 100252964 100259854 100259866 100259868 100260550 100261170 100261174 100261279 100266748 100266750 100266751 100266754 100266755 100266757 100266758 100272761 100272763 100278744 100278746 100278755 100278780 100278782 100278788 100278789 100278795 100278798 100278800 100278802 100278803 100278807 100286688 100286689 100286690 100286693 100286694 100286695 100286696 100286698 100286699 100286701 100286702 100286704 100286705 100286706 100286707 100286710 100286711 100286956 100286976 100288311 100288319 100288321 100288323 100288338 100289087 100290663 100291851 100292318 100292347 100292351 100292352 100293603 100300549 100306463 100306465 100306469 100312533 100312539 100312540 100313079 100313117 100317127 100317389 100318315 100318604 100319799 100321723 100321727 100321786 100328615 100328909 100328913 100328914 100329180 100329182 100329316 100329783 100329785 100329786 100329790 100329791 100329792 100329794 100329795 100329796 100329797 100329798 100329954 100329955 100329956 100329957 100329976 100345502 100345607 100359510 100359610 100368041 100375168 100375172 100389231 100389387 100389418 100389461 100389495 100389993 100390088 100390130 100390160 100390480 201004647 201004648 201004649 201007528 201007530 201007537 201007590 201007611 201007640 201007643 201007650 201007653 201010276 201010365 201010374 201010562 201010602 201010609 201010617 201010732 201010733 201010734 201010735 201010736 201010737 201010738 201010755 201010759 201010773 201010798 201010822 201010823 201010824 201010828 201010830 201010836 201010837 201010842 201010843 201010844 201010861 201010863 201010865 201010881 201010896 201010928 201010939 201010940 201010941 201010942 201010943 201010944 201010945 201010946 201010949 201010950 201010951 201022731 201022737 201022738 201022761 201022768 201023092 201023097 201023100 201023101 201023845 201026524 201026567 201026592 201026688 201026704 201026719 201026746 201026752 201027789 201027794 201027798 201027808 201027870 201027871 201027873 201027874 201027876 201027893 201027896 201027906 201027910 201027911 201027917 201027932 201027965 201027972 201027979 201027994 201028000 201028006 201028009 201028013 201028016 201028017 201028022 201028025 201028029 201028034 201029656 201029879 201029882 201029922 201029957 201038689 201038697 201038718 201038719 201038723 100036289 100062165 100122684 100139935 100243253 100261004 100062179 100063171 100068361 100069955 100069958 100081489 100081495 100081498 100081499 100081500 100126164 100219292 100219293 100266742 100292340 100318227 100318230 100318238 100348444 100350147 100350154 100350155 100350162 100350167 100350299 100352831 100354822 100354836 100354857 100355144 100355147 100355159 100355160 100355843 100355845 100355846 100355849 100355850 100355853 100355854 100355855 100355857 100355858 100355860 100355863 100355865 100355874 100357723 100357724 100357728 100357729 100358132 100358303 100358315 100358354 100359631 100359778 100359779 100360091 100360146 100360208 100360254 100361580 100361581 100362453 100363018 100363054 100363080 100363761 100363794 100364256 100364541 100364616 100365105 100365119 100365174 100365194 100367144 100367149 100367153 100367158 100367159 100367163 100367164 100367171 100367172 100367261 100367277 100367302 100367304 100368007 100368018 100368044 100368180 100368187 100368190 100368620 100368637 100368884 100369416 100369440 100375004 100375019 100375142 100375170 100375179 100375260 100375261 100375265 100375273
100375274 100375381 100375382 100375384 100375388 100375391 100375393 100375394 201008070 201008073 201008135 201008222 201022727 201023091 201038683 100003601 100006175 100017732 100021317 100021687 100022037 100022048 100022049 100022050 100022052 100022053 100022147 100022151 100022153 100022154 100022155 100022245 100051953 201000885 100226677 100271028 100271187 100271190 100271192 100271193 100286963 100348122 100375173 100011122 100021314 100021682 100021686 100027963 100027965 100032693 100032699 100032704 100036291 100038192 100038299 100042943 100044531 100045546 100045549 100045551 100051950 100051951 100051952 100051955 100057354 100057359 100057360 100057367 100062281 100064711 100064714 100064846 100065490 100065505 100069521 100069523 100069524 100069525 100069526 100069527 100069528 100069592 100069593 100069594 100069595 100069596 100069604 100069605 100069607 100069608 100069609 100069872 100069988 100070567 100070971 100072488 100075095 100078127 100081496 100081700 100083153 100083154 100083216 100083219 100083220 100083225 100088052 100089530 100089532 100089534 100089537 100089538 100089540 100089541 100089542 100089543 100089546 100091287 100094341 100094342 100094344 100094389 100095749 100096305 100096381 100096383 100096546 100096553 100096918 100096932 100097025 100097139 100097283 100097290 100097291 100097292 100097293 100097295 100097297 100097380 100097383 100097387 100097392 100097393 100097394 100101733 100101739 100101850 100114036 100114528 100120748 100120755 100120760 100125717 100125733 100137676 100138233 100138582 100144385 100144868 100144959 100144965 100144970 100144978 100145139 100150738 100150841 100158079 100160000 100160004 100160689 100163582 100163591 100164089 100164092 100164093 100164096 100164425 100174281 100174283 100174286 100174294 100174295 100174306 100174314 100174315 100174317 100176853 100180415 100181604 100182140 100182431 100182454 100190087 100190093 100190729 100190732 100190733 100190737 100190800 100190808 100190812 100195641 100219979 100219993 100219995 100221152 100221163 100223207 100225263 100225265 100225267 100225271 100225273 100230028 100230071 100239394 100243223 100243241 100243255 100244590 100247093 100255623 100257165 100257168 100257169 100257170 100257177 100261130 100272766 100274269 100274278 100274418 100274444 100274446 100274447 100274448 100274449 100274455 100274790 100274800 100274801 100274802 100274805 100274811 100274813 100274816 100279004 100282485 100282851 100282852 100282859 100286051 100286052 100286053 100286209 100286691 100286697 100286703 100288306 100288328 100288333 100292308 100293606 100293609 100310991 100312449 100312451 100312453 100312454 100312522 100315195 100322883 100322900 100328917 100337091 100337221 100337224 100337235 100337238 100337245 100346971 100348083 100348102 100348110 100348111 100348126 100350192 100350896 100351353 100351551 100351566 100351567 100351583 100351586 100351680 100351684 100351685 100351686 100351692 100351694 100351695 100351696 100352073 100352130 100352795 100352820 100354007 100354012 100354016 100354022 100355809 100355851 100355861 100355864 100355867 100357136 100357721 100357722 100358318 100358322 100358342 100358343 100358351 100359614 100361218 100361569 100362067 100362451 100362513 100362579 100362704 100362719 100362725 100362737 100362738 100362931 100363040 100363072 100363073 100363746 100364527 100364534 100364544 100364551 100364557 100364564 100364601 100364610 100364623 100365109 100365110 100365180 100366958 100367173 100367177 100367183 100367267 100367278 100367307 100368043 100368046 100368048 100368049 100368077 100368502 100368634 100369411 100369428 100369435 100391860 201000753 201001112 201001115 201001118 201002990 201002996 201003056 201004601 201004627 201004632 201004645 201004650 201004653 201004692 201008132 201008168 201009216 201009217 201009357 201009441 201009442 201009443 201009460 201009466 201009468 201009469 201009852 201009988 201010053 201010054 201010057 201010058 201013738 201014019 201023886 201024070 201028917 201028932 201028937 201029013 201029119 201029973 201032480 201032690 201032691 201032693 201032694 201032727 201032730 201032860 201032865 201032872 201033663 201033672 201033676 201033716 201033740 201038419 201038699 201038701 201038713 201038714 201038715 100039961 100039962 100039963 100042519 100042962 100043102 100043104 100373229 100378276 100378455 100378458 100378460 100378461 100378462 100378641 100384285 201002742 201002745 100101064 100110675 100125495 100125730 100138418 100144872 100158078 100162953 100252761 100267690 201009213 201026256 201026257 201026532 201026572 201026578 201026644 201026650 201026661 201026714 201026740 201026757 201026780 100003164 100003167 100003169 100003338 100003339 100003459 100003561 100003600 100003602 100003604 100003605 100003610 100003612 100003613 100003614 100004747 100004795 100004798 100004920 100004921 100004926 100005893 100005894 100005974 100006101 100006108 100006109 100006166 100006168 100006178 100007212 100007283 100007417 100007420 100007423 100007425 100007426 100007427 100007429 100007430 100007529 100007531 100007532 100007534 100007537 100007773 100007775 100007777 100007779 100010850 100010854 100010967 100011118 100015796 100015798 100017396 100017474 100017475 100017481 100017486 100017487 100017488 100017489 100017491 100017492 100017575 100026679 100026683 100026756 100026766 100026770 100031915 100031918 100031919 100031920 100031922 100036098 100036294 100036298 100036315 100036442 100036444 100036445 100036448 100038193 100042746 100042749 100042751 100042753 100042936 100042945 100043303 100044530 100046565 100047063 100050514 100050517 100050528 100051962 100052399 100052412 100052867 100053594 100053679 100053862 100053871 100055934 100055935 100055944 100055947 100056017 100061637 100061652 100062054 100062181 100062282 100063154 100063156 100069950 100069953 100084386 100084399 100084568 100084828 100084832 100084834 100087346 100087473 100087687 100091008 100091014 100091024 100091028 100091029 100092506 100092874 100092884 100096343 100101735 100101852 100101947 100105675 100105676 100114027 100120122 100120127 100122347 100136182 100136185 100147583 100161923 100162295 100162299 100162409 100162421 100162422 100162733 100162737 100162782 100162785 100162787 100162793 100162897 100162900 100185837 100185963 100186762 100186794 100187321 100219278 100219280 100219284 100219295 100219969 100219973 100219985 100219990 100226351 100226370 100226384 100226869 100226894 100230025 100230063 100230064 100230068 100230069 100230073 100230075 100231066 100231162 100237197 100237303 100239398 100239401 100243229 100243230 100243232 100243234 100243235 100243237 100243238 100243240 100243242 100243243 100243246 100243247 100243248 100243249 100243254 100243510 100243627 100243633 100243641 100247097 100248579 100248595 100248596 100248599 100248606 100252585 100252594 100252598 100252599 100252609 100252775 100252796 100252799 100252940 100252943 100259999 100260033 100260970 100261002 100261121 100261140 100261175 100261178 100261179 100261273 100270300 100271039 100271724 100271725 100271726 100271727 100271728 100271729 100271730 100271737 100271738 100271804 100271907 100271911 100274453 100274597 100274763 100274791 100274792 100274793 100274797 100274803 100274804 100274806 100274807 100274809 100274810 100274812 100274814 100274815 100274817 100277460 100277823 100277912 100277919 100278751 100278757 100278783 100278787 100278790 100278793 100286303 100286467 100286473 100286476 100286654 100286680 100286681 100311167 100314632 100314680 100314691 100314692 100316908 100317274 100317279 100317282 100317385 100317386 100317390 100317391 100317396 100317397 100317495 100317496 100317498 100317500 100317501 100317503 100318307 100319770 100325264 100325265 100325268 100325269 100327212 100327213 100327214 100332357 100336584 100336603 100336613 100336867 100336882 100336903 100336904 100336908 100337082 201001114 201001119 201001120 201001121 201001141 201001142 201001145 201001147 201002909 201002912 201002916 201003047 201003048 201003050 201003052 201003473 201003485 201007607 201010364 201010366 201010378 201010379 201010566 201010601 201010603 201010604 201010780 201010786 201010789 201010791 201013163 201015500 201024188 201024189 201024193 201024194 201024195 201024198 201024199 201025160 201025166 201025169 201025170 201025171 201025172 201025174 201025175 201025316 201025394 201025396 201025397 201025399 201025401 201025407 201025909 201025910 201025921 201025928 201025944 201029106 201029657 201029734 201029745 201029769 201029771 201029772 201029774 201029781 201029782 201029826 201029900 201029919 201029920 201029965 201032452 201033665 100239400 100239525 100239527 100243519 100243628 100243630 100243646 100252620 100309573 100311176 100314628 100314690 100327768 100327772 100327774 100350168 100350662 100350706 100354003 100354014 100011042 100011044 100027972 100027975 100027976 100039842 100052406 201002340 201007606 201007645 201007665 201027819 201027882 201027889 100026682 100029208 100057368 100057370 100057372 100074202 100074203 100075101 100075383 100100131 100100187 100153483 100240178 100345539 201004602 201004640 201007515 201028015 100004799 100005809 100005887 100005977 100006115 100029071 100029076 100029080 100029084 100029205 100029206 100029214 100029215 100029243 100029319 100029321 100029324 100029329 100042748 100042937 100043084 100043093 100044352 100044355 100044356 100044360 100044362 100044519 100044520 100044527 100044529 100052840 100052857 100052866 100059521 100059524 100059528 100059531 100061919 100061925 100061926 100061927 100061935 100061938 100061940 100062037 100062039 100062040 100062045 100062046 100062047 100062049 100062050 100062051 100062052 100062063 100062265 100069694 100069867 100083156 100083158 100083535 100083579 100084338 100084343 100091020 100091023 100092839 100092889 100094427 100095585 100095654 100095655 100095658 100095660 100095662 100095663 100095664 100095748 100095756 100095757 100096200 100096397 100096398 100096403 100096544 100096549 100096550 100105072 100105079 100105244 100105550 100105551 100105553 100144862 100144863 100144864 100144953 100144958 100144962 100144969 100144973 100144974 100144975 100144976 100145134 100145136 100145137 100181739 100182144 100219984 100285789 100286041 100286046 100286047 100286293 100286298 100286305 100286317 100286653 100286656 100286658 100286659 100286687 100286945 100310993 100311166 100337096 100337099 100337223 100337225 100337226 100337228 100337230 100337232 100337233 100337234 100337237 100337239 100337247 100346930 100348080 100350178 100350292 100350675 100350718 100352122 100352123 100353255 100354006 100354030 100354823 100354825 100354839 100354872 100354877 100354881 100354882 100354884 100362743 100363016 100363656 100364530 100365108 201003521 201003940 201004081 201004092 201004101 201004102 201004103 201004104 201004106 201025168 201025178 201025182 201025184 201025189 201025191 201025908 201025911 201025912 201025914 201025916 201025917 201025918 201025919 201025922 201025924 201025927 201025930 201025934 201025936 201025938 201025939 201025943 201025946 201025992 201029033 201029043 201033646 201033650 201038680 201038687 201038690 201038695 201038696 201038700 201038727 201038732 100014487 100015384 100036443 100138231 100139924 100219272 100219273 100219279 100219286 100239537 100243225 100274598 100277811 100278750 100318231 100318312 100318624 100350172 100350713 100350885 100350887 201000907 201007641 201007670 100020656 100020747 100020754 100020758 100020760 100096405 100125492 100125715 100125722 100125724 100125734 201000897 201013830 100010711 100010721 100010841 100010843 100052863 100219275 100219276 100252795 100391373 100069698 201027792 201027895 201027933 100271912 100357727 100007115 100014486 100015340 100015341 100017397 100017477 100017478 100017479 100017482 100017483 100017563 100017565 100017674 100019390 100020130 100021313 100021315 100021316 100021450 100021681 100021683 100042744 100043103 100044526 100044536 100054415 100057195 100057369 100063298 100068210 100069598 100069603 100072483 100072487 100072489 100072492 100072495 100084572 100090667 100090925 100091935 100093330 100093337 100096559 100096927 100096928 100096933 100097980 100097981 100097991 100098467 100098479 100099528 100100178 100100838 100101399 100101734 100101742 100101743 100101752 100115845 100118071 100118073 100118075 100118077 100118083 100120745 100120746 100120761 100124811 100124859 100124955 100124957 100124959 100124965 100124970 100124973 100124975 100124976 100124977 100124978 100124979 100124980 100147561 100147569 100147584 100161930 100161960 100162417 100162441 100162443 100162446 100174300 100180413 100189465 100189809 100189813 100189814 100189817 100189818 100189821 100189822 100189824 100189907 100189917 100189983 100196404 100221155 100228827 100228989 100228991 100229107 100229110 100229111 100237379 100248601 100255558 100256991 100259860 100259867 100259872 100259877 100260022 100260025 100271800 100271801 100278796 100278801 100278808 100292345 100293601 100314678 100314714 100314878 100315184 100315188 100315189 100315190 100315191 100315193 100315196 100315360 100315611 100315870 100315876 100315880 100315881 100315884 100315885 100315887 100315888 100317395 100318225 100318228 100318235 100318236 100318237 100318239 100318605 100318620 100318621 100318622 100319270 100319769 100319786 100320056 100320057 100320143 100320149 100320150 100320152 100320153 100320371 100345346 100345503 100345517 100345575 100345598 100346861 100346902 100346915 100346918 100346919 100346920 100346924 100346926 100348086 100367305 100368053 100368056 100368062 100368079 100368080 201004246 201004259 201004388 201004392 201008176 201008223 201010112 201010113 201010127 201010151 201010153 201010154 201010155 201010156 201010165 201010183 201010185 201010189 201010193 201010195 201010241 201010245 201010247 201010249 201010250 201015206 201015229 201027997 201028007 201029652 201034073 201034076 201038698 201038704 201038717 201038728 100003254 100003257 100003617 100007193 100013968 100029242 100053598 100053870 100061922 100061923 100062038 100062180 100063151 100063163 100132030 100132042 100132044 100132046 100132048 100226891 100233964 100237302 100239303 100239390 100239396 100239399 100239405 100239524 100243224 100243918 100247265 100248621 100256845 100266745 100274457 100274692 100274695 100274760 100277916 100278753 100282846 100282861 100282869 100282880 100286304 100286306 100286312 100286463 100329959 100329967 100336756 100336907 201001543 201013172 201023049 201025393 201025923 201025926 201025933 201029776 201029822 201029824 201029830 201029832 201029849 201029884 201029969 201029977 201031754 201032011 100005979 100005989 100007110 100007114 100019921 100029320 100042747 100042756 100043095 100043097 100055823 100055824 100055826 100055931 100061911 100062237 100062284 100063167 100063168 100075402 100094329 100094340 100094348 100101943 100105567 100105571 100107308 100114516 100114521 100114526 100114534 100139921 100139933 100147576 100148857 100148956 100149190 100149198 100149199 100149200 100149201 100149209 100158060 100162780 100163583 100164432 100164441 100170465 100170694 100170695 100171037 100223282 100223478 100231107 100239258 100239383 100239385 100243221 100243252 100251537 100251651 100278749 100290804 201000879 201000912 201002219 201002397 201002400 201002402 201007441 201007523 201007535 201007536 201007596 201010797 201015526 201015550 201024191 201026432 201026750 201029746 201029795 201029840 201032724 100014483 100014770 100015332 100063158 100070770 100084351 100087606 100089545 100102142 100144952 100179386 100179414 100179461 100179467 100229102 100229112 100239407 100239523 100247368 100248313 100248607 100248609 100256990 100257095 100257178 100257215 100257388 100257585 100257691 100270283 201000748 201007513 201007520 201008031 201029039 201029962 201032719 100004665 100005898 100005971 100005978 100006183 100007280 100007336 100007530 100007548 100007772 100007784 100010715 100014020 100014466 100014467 100014470 100014471 100014472 100014474 100014475 100014476 100014477 100014479 100014480 100014481 100014482 100014484 100014517 100014519 100014521 100014524 100014527 100014529 100133626 100150746 100152330 100152331 100152332 100284349 100284360 100285002 100285005 100285010 100285017 100317387 100317393 100317504 100318229 100318402 100318411 100318414 100318418 100318422 100318625 100343596 100343645 100343685 201009207 201009209 201009215 201009218 201009358 201009464 201009465 201009472 201010754 201010795 201010796 201010806 201010807 201010808 201010811 201010831 201010839 201015547 201015549 201015554 201031859 100221171 100284356 100284365 100318232 100318407 201028010 201028024 201029958 201029971 201033653 201033658 201033674 201033687 100274081 100274084 100274085 100274087 100286686 100042941 100042942 100057366 100116622 100125712 100144963 100144972 100290662 201002220 201010175 100068192 100096392 100096401 100350287 100350680 100350687 100350702 100350704 100350707 100350708 100350720 100351704 100352060 100352131 100352138 100352780 100354820 100354853 100361388 100363750 100363773 100363778 100363787 100363792 100364549 100364605 201040665 201040669 201040671 201040683 201040707 100153468 100157753 201015433 201015525 100021203 100026669 100026670 100026687 100026764 100026767 100026768 100026773 100027969 100027977 100029326 100031154 100036438 100036454 100039143 100039152 100039245 100039765 100039766 100039768 100039770 100039775 100039776 100039777 100039779 100039780 100042960 100043086 100043087 100043088 100043089 100043090 100043092 100043094 100051961 100051966 100069695 100069746 100069862 100069866 100069868 100069871 100069873 100069954 100070570 100070572 100070573 100070574 100070578 100070580 100070581 100070582 100084840 100088047 100088058 100096394 100096395 100118554 100120128 100120747 100120750 100120757 100120758 100125713 100125718 100125725 100125756 100125847 100125849 100139677 100139711 100139898 100139910 100139913 100139915 100139927 100139928 100139929 100144865 100144869 100144954 100144971 100144979 100144980 100145120 100152324 100152325 100152328 100152329 100152555 100153196 100153463 100153469 100153476 100153477 100158059 100158063 100172260 100172268 100174284 100174291 100174293 100174302 100174311 100174450 100174452 100174461 100174466 100181608 100181888 100182127 100182134 100182142 100182387 100182396 100186761 100186768 100186770 100186771 100186773 100186777 100186781 100186784 100186785 100186787 100189494 100189559 100189811 100189816 100230027 100230029 100230067 100232493 100232739 100232755 100232759 100237189 100237195 100237222 100239384 100239388 100239404 100239521 100247087 100252592 100252767 100252950 100252969 100252971 100253476 100253481 100253486 100253491 100253496 100253498 100253553 100259735 100259880 100260027 100260031 100260035 100260036 100260040 100260043 100260175 100260177 100260179 100267569 100267574 100267696 100267698 100271181 100271189 100274819 100274820 100274822 100274823 100274825 100274826 100282483 100282831 100282832 100282836 100282837 100282840 100282841 100282842 100282843 100282847 100282848 100282853 100282856 100282857 100282860 100282863 100282866 100282867 100282870 100282872 100282874 100282875 100282883 100286655 100286978 100292301 201000904 201002216 201002424 201013046 201013082 201013174 201029650 201029740 201029742 201029749 201029750 201029751 201029752 201029756 201029759 201029786 201029787 201029789 201029793 201029814 201029815 201029829 201029843 201029850 201029855 201029912 201029913 201033570 201033589 201033621 201033686 100049197 100049199 100180208 100180417 100181738 100181740 100031143 100031144 100068185 100068197 100068203 100068205 100068207 100068213 100068218 100068219 100068221 100068224 100068225 100069696 100069747 100069864 100071505 100071508 100071509 100071517 100071518 100083213 100083226 100083293 100158056 100346904 100346931 100348090 100350186 100350297 100350681 100350716 100350748 100351355 100352126 100352140 100352143 100357128 100358124 100358178 100358311 100358313 100358316 100361418 100361965 100361970 100362578 100362718 100362932 100363037 100363043 100363059 100363744 100363797 100364518 201004462 201004593 201004595 201004596 201004599 100003611 100004671 100087343 201029804 201033574 100043109 100090916 100090921 100090923 100090924 100090933 100091007 100091010 100091013 100091015 100091019 100091025 100091026 100091027 100091031 100091032 100095582 100095584 100003071 100003161 100005818 100007210 100007216 100007343 100007412 100010704 100010705 100013964 100014011 100014026 100016933 100017398 100017476 100017480 100017484 100017485 100017560 100017561 100020661 100020745 100020756 100021179 100021201 100021202 100021205 100021207 100021208 100021287 100021289 100021294 100021295 100021297 100021299 100021302 100021392 100021395 100021396 100021397 100021399 100022042 100022043 100022044 100022054 100022144 100022145 100022146 100022150 100022152 100022158 100022160 100022162 100022240 100024709 100024882 100024890 100029327 100029328 100030119 100030332 100030347 100030765 100030772 100030774 100030778 100030780 100030781 100030785 100031140 100031148 100031151 100031155 100031156 100031158 100038187 100038188 100038189 100038190 100038194 100038200 100038201 100038205 100038211 100038213 100038215 100038286 100038287 100038288 100038289 100038290 100038291 100038292 100038294 100038295 100038296 100038297 100039248 100039763 100039773 100039781 100039783 100039785 100039786 100039787 100039788 100039790 100039791 100039792 100039794 100039795 100039843 100039846 100039850 100039851 100039853 100039854 100039856 100039857 100039946 100039947 100039948 100039949 100039950 100039951 100039952 100039954 100039955 100039957 100039958 100039959 100039960 100039965 100039966 100039967 100043101 100043113 100043293 100043297 100043300 100043314 100043316 100043317 100043319 100044335 100044522 100044524 100044525 100044537 100044541 100044544 100044545 100044546 100045538 100045545 100045553 100045555 100045558 100045562 100045563 100045566 100050525 100051949 100051956 100051960 100052388 100052389 100052391 100052392 100052393 100052394 100052395 100052396 100052397 100052398 100052402 100052410 100052847 100052858 100052861 100052864 100052865 100052950 100053602 100053603 100053786 100060101 100060102 100060104 100060105 100060106 100060109 100060110 100060111 100060112 100060115 100061649 100061774 100061776 100061777 100061778 100061779 100061792 100061794 100061797 100061806 100094217 100094346 100094391 100094395 100095588 100095591 100095593 100095596 100095751 100095793 100095794 100095795 100095796 100095798 100095799 100096116 100096117 100096118 100096119 100096201 100096204 100096206 100096207 100096208 100096210 100096300 100096301 100096302 100096303 100096304 100096306 100096340 100096341 100096342 100096344 100096350 100096351 100096384 100096385 100096386 100096387 100096388 100096389 100096390 100096391 100096393 100096396 100096399 100096400 100096402 100096543 100096545 100096547 100096551 100096552 100096554 100096555 100096556 100096557 100096558 100096560 100096609 100096610 100096611 100096612 100097288 100098000 100098001 100098475 100099517 100099522 100099523 100099526 100099531 100100168 100100171 100100172 100100174 100100175 100100180 100100183 100115410 100115417 100115418 100115463 100115800 100115841 100115842 100115843 100115844 100115854 100115855 100115856 100115857 100115859 100115860 100116363 100116364 100116365 100116367 100116368 100116370 100116372 100116376 100116379 100116380 100116381 100116391 100116392 100116605 100116609 100116610 100116612 100116613 100116614 100116615 100116619 100116621 100116623 100116624 100116627 100116631 100116634 100116635 100116636 100116683 100116684 100116686 100116690 100116693 100116696 100148964 100158062 100158066 100158080 100158092 100158419 100159234 100159235 100159239 100159300 100159301 100159302 100159304 100159305 100159306 100159307 100159308 100159309 100159310 100159311 100159312 100159314 100159315 100159316 100159317 100159318 100159319 100159320 100159321 100159322 100159323 100159324 100159325 100159326 100159327 100159328 100159329 100159330 100159332 100159334 100159335 100159337 100159562 100159564 100159565 100159567 100159641 100159773 100159823 100159879 100159921 100163734 100163735 100164444 100164455 100165168 100165177 100165686 100165687 100165689 100165690 100165691 100165692 100165693 100165694 100165695 100165696 100165697 100165698 100165939 100165941 100166080 100166081 100166082 100166083 100166084 100166086 100166089 100166090 100166091 100166092 100166093 100166094 100166095 100166097 100166099 100166100 100166102 100166104 100166105 100166108 100166109 100166110 100166111 100166115 100166118 100166119 100350929 100351359 100351579 100351580 100351689 100351691 100351705 100352141 100352812 100352953 100353294 100353599 100354008 100354011 100354026 100354826 100354834 100354868 100354878 100362561 100362717 100363035 100363036 100364511 100364553 100364560 100365102 100365113 100368055 100368625 100373556 100373557 100375007 100375169 100375171 100375175 100375176 100375177 100375222 100375223 100375227 100375232 100375240 100375259 100375269 100375271 100375272 100375275 100375281 100375285 100375289 100375292 100375294 100375386 100375387 100375389 100375390 100375392 100375395 100375397 100375400 100375401 100375403 100375404 100377889 100378014 100378018 100378078 100378206 100378208 100378303 100378454 100378643 100378849 100378889 201000690 201000691 201000692 201000693 201000694 201000695 201000696 201000698 201000700 201000703 201000704 201000705 201000707 201000709 201000742 201000744 201000747 201000749 201000754 201000756 201000876 201000880 201000889 201000891 201000894 201000900 201000902 201000903 201001117 201001863 201001950 201001958 201001989 201002101 201002102 201002104 201002105 201002698 201002699 201002707 201002708 201002711 201002720 201003386 201003393 201003395 201003396 201003397 201003443 201003447 201003478 201003482 201003494 201017071 201017072 201017073 201017074 201026251 201026255 201026258 201026259 201026355 201026437 201026533 201026600 201026609 201026623 201026624 201026625 201026626 201026627 201026628 201026629 201026630 201026631 201026632 201026633 201026634 201026635 201026637 201026638 201026656 201026669 201026682 201026684 201026697 201026698 201026703 201026718 201026743 201026745 201026747 201026751 201026756 201026758 201026759 201026760 201026761 201026762 201026763 201026764 201026766 201026767 201026768 201026769 201026770 201026771 201026772 201026781 201026783 201027475 201027698 201027754 201027755 201027757 201027805 201027817 201027826 201027833 201027877 201027909 201027916 201027927 201027939 201027964 201027968 201027987 201027992 201027993 201027996 100005988 100007123 100007871 100013960 100013966 100013969 100014465 100036163 100036164 100042752 100043313 100050521 100052765 100052856 100052869 100052944 100052951 100053872 100056016 100061929 100068216 100068220 100070583 100070771 100075306 100082439 100082461 100089122 100092504 100097997 100099520 100099532 100100166 100100181 100101619 100101754 100110935 100110945 100110951 100114519 100125848 100139926 100149196 100175805 100176333 100195447 100219986 100226374 100226382 100232757 100248587 100248594 100259865 100259875 201000875 201000877 201000884 201000895 201001520 201001523 201001525 201001531 201013768 201022996 201023837 201023852 201029653 201029791 201029801 201029817 201033573 201033577 201033580 201033642 100014468 100027962 100027966 100039972 100045712 100059529 100062235 100063169 100063299 100064704 100064713 100064965 100072736 100073202 100073251 100081415 100084342 100097287 100101745 100104637 100104646 100114468 100114471 100114473 100114474 100114551 100114764 100114765 100115202 100115207 100115276 100125738 100126679 100133630 100133636 100144878 100145133 100145143 100149716 100149822 100164094 100174039 100174304 100176884 100180422 100180425 100181583 100181616 100186792 100187425 100187429 100189474 100189487 100189750 100189894 100190817 100190826 100223189 100223455 100225270 100247787 100248156 100248603 100248611 100248614 100248616 100248617 100248622 100254466 100254552 100254787 100254793 100261138 100261172 100261302 100261321 100267566 100270294 100270296 100271024 100271025 100271026 100271029 100271031 100271033 100271183 100271731 100290666 201002051 201002229 201002260 201002296 201002301 201002540 201002998 201008032 201008137 201008187 201009211 201009462 201013083 100004673 100036152 100036309 100039953 100039956 100045557 100045677 100045683 100049071 100049078 100057357 100057358 100057364 100057365 100057371 100057374 100058490 100058491 100058493 100058494 100064562 100064718 100064853 100067532 100067533 100070864 100070919 100070927 100073204 100073244 100073245 100073246 100083819 100083928 100084334 100084335 100084336 100084344 100084346 100084395 100098473 100099294 100099295 100099529 100099534 100100135 100100139 100100280 100100283 100129775 100129776 100129778 100129780 100129781 100129795 100138284 100138587 100149719 100149821 100149835 100161958 100161959 100161962 100162293 100162294 100162406 100162413 100162423 100162440 100162442 100174289 100174290 100174301 100174303 100174310 100174312 100174313 100174316 100174321 100174451 100174454 100174455 100174456 100174457 100174458 100174460 100174462 100174465 100174467 100174963 201002904 201004341 201010997 201010998 201011160 201011166 201011168 201011175 201011176 201011177 201011178 201015438 201015439 201015542 201015551 201015553 201031675 201031687 201031691 201031695 201032703 201032708 201032709 201032714 201032715 201032726 201032728 100027106 100027958 100027967 100029330 100029340 100043295 100044059 100063164 100063166 100063170 100081701 100081756 100083536 100084345 100092497 100092836 100096542 100100854 100101073 100101078 100101398 100116688 100116698 100122331 100122337 100122581 100122583 100122586 100122588 100122682 100122686 100122689 100122690 100122691 100125842 100125844 100125851 100133632 100133639 100134283 100134293 100134345 100134394 100134396 100139678 100139721 100139723 100139914 100139923 100140593 100140599 100140620 100140622 100140624 100140628 100140630 100143696 100143720 100143724 100143727 100147567 100147582 100148856 100148953 100148965 100149193 100149195 100153194 100153197 100153199 100153461 100153470 100153478 100153479 100153481 100159825 100233063 100233949 100233965 100233968 100233969 100233971 100233972 100233973 100233974 100233975 100233976 100233977 100233979 100233980 100234664 100234665 100234666 100234668 100234669 100234671 100234672 100234673 100234780 100234784 100234785 100234786 100234788 100234794 100234795 100234797 100234798 100234799 100234800 100234801 100234802 100317653 100343578 100344364 100346917 100346922 100346923 100346928 100346929 100346933 100348098 100351152 100351153 100351155 100351159 100351688 100351693 100352093 100352133 100352783 100352785 100352828 100352832 100352957 100363039 100364857 100365115 100365167 100365173 100365176 100365179 100365184 201002617 201002684 201002686 201002687 201002688 201002701 201004680 201010273 201010367 201010370 201010375 201010608 201015527 201015548 201023057 201023099 201023831 201023862 201029944 201033576 201033578 201033579 201033587 201033588 201033600 201033601 201033603 201033604 201033606 201033608 201033611 201033616 201033618 201033622 201033623 201033624 201033626 201033627 201033629 201033641 201033645 201033673 201033713 201033714 201033718 100029466 100029475 100034519 100035912 100036293 100036296 100036297 100036308 100036318 100036440 100036441 100036450 100036451 100036453 100042947 100044518 100044538 100046922 100047057 100061626 100061642 100061798 100061801 100061802 100061803 100061804 100061807 100061808 100061809 100061912 100061913 100061914 100061915 100061916 100061917 100061921 100061930 100061931 100061932 100061936 100061937 100061939 100062034 100062035 100062036 100062041 100062044 100062058 100062062 100062167 100062168 100062170 100062171 100062172 100062174 100062175 100062176 100062177 100081405 100081406 100081407 100081408 100081410 100081412 100081413 100081416 100081417 100081418 100081480 100081481 100081482 100081483 100081485 100081486 100081488 100081497 100081689 100081697 100084394 100084569 100084570 100090931 100090932 100090936 100091030 100091096 100091097 100091158 100091159 100091260 100091262 100091263 100091264 100091266 100091267 100091268 100091269 100091270 100091271 100091272 100091274 100091275 100091276 100091277 100091278 100091279 100091280 100091281 100091282 100091283 100091284 100091285 100091286 100124794 100124795 100124796 100124797 100124798 100124799 100124800 100124801 100124802 100124803 100124804 100124805 100124808 100124809 100124810 100124812 100124813 100124855 100124856 100124857 100124858 100129370 100129786 100129787 100129791 100129792 100149720 100149836 100149846 100149849 100150008 100150186 100150190 100150192 100150194 100150195 100150196 100150197 100150198 100150199 100150200 100150201 100150202 100150208 100150211 100150728 100150729 100150730 100150731 100150732 100150733 100150734 100150735 100150736 100150737 100150739 100150740 100150741 100150742 100150743 100150744 100150745 100150747 100150748 100150749 100150750 100150751 100150752 100150753 100150754 100150755 100150756 100150757 100150836 100150837 100150838 100150839 100150840 100150843 100150844 100150845 100150846 100150847 100150848 100150849 100150850 201002978 201002979 201002981 201002982 201003008 201003009 201003376 201009872 201009975 201010040 201010042 201010043 201010044 201010047 201010049 201010055 201010947 201010948 201010999 201011111 201011113 201011161 201011162 201011163 201011164 201011165 201011167 201011170 201011171 201011172 201011173 201011174 201011240 201011241 201011242 201011243 201011244 201011245 201011246 201011248 201011249 201011250 201011251 201011252 100004742 100004923 100004928 100044359 100046920 100050535 100069600 100090536 100157755 100274082 201000750 201003072 100093339 100094126 100094215 100094326 100094327 100094328 100042745 100044533 100044539 100055825 100055946 100095657 100095750 100118352 100118532 100118547 100120129 100135547 100140625 100157757 100157760 100164428 100164429 100164434 100164438 100189459 100240149 100240151 100266743 100266746 100266747 100266749 100266752 100266753 100282844 100282845 100282849 100282850 100282855 100288325 100288327 100288330 100298806 100346961 100348446 100350133 100350138 100350142 100350144 100350148 201004652 201004654 201007916 201008074 201009341 201010833 201010840 201026696 201026754 201027307 201027308 201027309 201027310 201027318 201027432 201027433 201027434 201027436 201027440 201032522 201038693 201038702 201038712 201038746 100030348 100036455 100043315 100051954 100055828 100055932 100055933 100055936 100055942 100055945 100056020 100118949 100159313 100181610 100181886 100226380 100230065 100230066 100252588 100252760 100260042 100270301 100274587 100348087 100348088 201002612 201002705 201029851 201029968 100003569 100004611 100004612 100004613 100004664 100004746 100004749 100004796 100004924 100004927 100007869 100044186 100050082 100061805 100065514 100068187 100068193 100068201 100068204 100068215 100068360 100069861 100069870 100069951 100069952 100069965 100070758 100070765 100073201 100073243 100073248 100073250 100073259 100075295 100075398 100083570 100083571 100083574 100083576 100083600 100087487 100090780 100091927 100091930 100091932 100092491 100105679 100110931 100114046 100114052 100114053 100114463 100114517 100116620 100116691 100120118 100124865 100125746 100125752 100134675 100135540 100136183 100136186 100136627 100138238 100138283 100138584 100138588 100138591 100140597 100140627 100142324 100142325 100142334 100143701 100143735 100144870 100144876 100145122 100145126 100145127 100145129 100145130 100145132 100145138 100146845 100146846 100146848 100147184 100147190 100147199 100147200 100147202 100147204 100147206 100147284 100147288 100147289 100147292 100147295 100147299 100147382 100147385 100147387 100147389 100147391 100147392 100147394 100147399 100148961 100158082 100160697 100163715 100163716 100163721 100169561 100169643 100169649 100169656 100169765 100170471 100170472 100170474 100170475 100170476 100170479 100170564 100170565 100170568 100170570 100170571 100170574 100170578 100170681 100170682 100170684 100170686 100170689 100170698 100170702 100170703 100171029 100171032 100172264 100172267 100174037 100174038 100174280 100174282 100174288 100174298 100176041 100176343 100176346 100176860 100176876 100176877 100179398 100180418 100181599 100181743 100181745 100182124 100182442 100182465 100184202 100189747 100190089 100190100 100190734 100190805 100190807 100190814 100190815 100190816 100190819 100190820 100190823 100190824 100190825 100190827 100190828 100190829 100190830 100190831 100190832 100190833 100190834 100190835 100190836 100190837 100195884 100221143 100229104 100229114 100229117 100229630 100231167 100239532 100239804 100243511 100243512 100243516 100243637 100244599 100244607 100246484 100246498 100247409 100247412 100247413 100247416 100247482 100247483 100252615 100254745 100254747 100254748 100254750 100254751 100255621 100255622 100256846 100271023 100274451 100282486 100282487 100282871 100282881 100286700 100286947 100286948 100293602 100293611 100300434 100300513 100300514 100300515 100300516 100300518 100300521 100300522 100300523 100300531 100300533 100300534 100300537 100300538 100300543 100300545 100300548 100300551 100300552 100300554 100300555 100300556 100300557 100300641 100300645 100300646 100300647 100300649 100305668 100309655 100309657 100311174 100311374 100311378 100311379 100311464 100311472 100312531 100314486 100314497 100314626 100314629 100314630 100314631 100314633 100314634 100314676 100314679 100314681 100314685 100314686 100314688 100314694 100314695 100314697 100314712 100314713 100320649 100322649 100322869 100322895 100332724 100333565 100348084 100348144 100350182 100350717 100350719 100350730 100350732 100350734 100351554 100352064 100352782 100352784 100352789 100352834 100352949 100352955 100353243 100353250 100353251 100353252 100353253 100353259 100353261 100353262 100353263 100353264 100353266 100353267 100353268 100353269 100353270 100354035 100354463 100354702 100354704 100354708 100354711 100354821 100354844 100354847 100354856 100354861 100354862 100354863 100354866 100354880 100354885 100354886 100354890 100356082 100356201 100356206 100356259 100356263 100356283 100356306 100356766 100356767 100356769 100356771 100357720 100362499 100362574 100362933 100363079 100364258 100364512 100364514 100364515 100364516 100364519 100364522 100364523 100364531 100364537 100364558 100364570 100364576 100364578 100364580 100365182 100366923 100368052 100368054 100368081 100368514 100368526 100368554 100369415 100369420 100369431 100369441 100369446 100371218 100375145 100375236 100375238 100392625 100398443 201003095 201003500 201004378 201004387 201004391 201009471 201010785 201010787 201010821 201013155 201015203 201022976 201023008 201027802 201027809 201027821 201027831 201028001 201028044 201028905 201029029 201029654 201029655 201029758 201029761 201029766 201029778 201029783 201029819 201029834 201029844 201029845 201029847 201029852 201029853 201029857 201029859 201029860 201029863 201029865 201029867 201029868 201029869 201029870 201029871 201029873 201029875 201029877 201029880
201029881 201029883 201029885 201029887 201029903 201029904 201029907 201029908 201029909 201029924 201029927 201029932 201029933 201029941 201029945 201029955 201029956 201029976 201030945 201030946 201030950 201030951 201030955 201030957 201030959 201031157 201031158 201031163 201031197 201031198 201031199 201031202 201031207 201031210 201031445 201031451 201031456 201031459 201031521 201031522 201031523 201031524 201031525 201031526 201031529 201031530 201031533 201031534 201031678 201031682 201031690 201031864 201031873 201032759 |
Recalling Firm/ Manufacturer |
Respironics California, LLC 2271 Cosmos Ct Carlsbad CA 92011-1517
|
For Additional Information Contact |
800-722-9377
|
Manufacturer Reason for Recall |
Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC power.
|
FDA Determined Cause 2 |
Device Design |
Action |
On 10/30/2020, the firm sent an "URGENT-MEDICAL DEVICE CORRECTION" Notification Letter via priority mail informing customer that the Ventilators may experience two potential issues with the battery (see table below for additional
information).
1. The ventilator may exhibit a high priority Check Vent: Battery Failed alarm if the battery is allowed to over discharge.
2. The battery may fail to charge if the ventilator is stored for an extended period of time without being plugged into AC power.
If the ventilator exhibits a high priority Battery Failed alarm while connected to
AC power, the V60 ventilator will annunciate an audible and visual alarm. The
ventilator will continue to deliver prescribed therapy; however, the audible and
visual alarm will continue until the device is powered down; therefore, the patient
should be transferred to an alternate source of ventilation as soon as one
becomes available. The hazard associated with transferring a patient to an
alternate source of ventilation is a temporary loss of therapy leading to potential
hypercarbia or hypoxemia (moderate).
If AC power fails and the backup battery is not functioning, an audible and visual
alarm will annunciate for at least two minutes. The ventilator will no longer be
operative resulting in a total loss of therapy. The patient will require an alternate
source of ventilation. The hazard associated with a ventilator shutdown is a total
loss of therapy leading to potential hypercarbia or hypoxemia (severe).
Customers are instructed:
Follow the instructions in the V60 user manual and the instructions below to
reduce any risk associated with potential battery issues.
1) Avoid allowing the ventilator battery to become completely discharged.
Otherwise, the battery may become over-discharged and require long
recharge times of up to 16 hours or more. The over-discharged condition
may permanently damage the battery so that it is unable to recharge. To
prevent the occurrence o |
Quantity in Commerce |
9,863 systems |
Distribution |
Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AS, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE,NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Afghanistan, Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, ST. Pier Miquel, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, Wallis Futuna, and Zambia. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
|
510(K) Database |
510(K)s with Product Code = MNT and Original Applicant = RESPIRONICS CALIFORNIA, INC.
|
|
|
|