|
Class 2 Device Recall Philips Respironics V60 Ventilator |
![see related information](../images/related-info.gif) |
Date Initiated by Firm |
October 30, 2020 |
Create Date |
January 22, 2021 |
Recall Status1 |
Completed |
Recall Number |
Z-0899-2021 |
Recall Event ID |
86646 |
510(K)Number |
K082660
|
Product Classification |
Ventilator, continuous, minimal ventilatory support, facility use - Product Code MNT
|
Product |
Philips Respironics V60 Ventilator Part Number DU1053617 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories. |
Code Information |
Model: V60 Ventilator, Part Number DU1053617 V60 Ventilator, Us Demo Unit Serial Numbers: 100002908 100002911 100002913 100002914 100002916 100002959 100003001 100003003 100003005 100003007 100003018 100003046 100003052 100003081 100003674 100003675 100003685 100003689 100003728 100003736 100003757 100003909 100003910 100003913 100003916 100003918 100003919 100004032 100004033 100004077 100004081 100004120 100004122 100004123 100004124 100004125 100004127 100004128 100004131 100004320 100004322 100004323 100004324 100004325 100004508 100004542 100004543 100004545 100004547 100004549 100004550 100004553 100004554 100004555 100004556 100004557 100004558 100004620 100004621 100004622 100004623 100004624 100004625 100004626 100004627 100004982 100004983 100004984 100004985 100004986 100004987 100004988 100004992 100004994 100004995 100004999 100005110 100005116 100005117 100005141 100005147 100005150 100005239 100005241 100005242 100005247 100005251 100005334 100005338 100005347 100005388 100005486 100005489 100005491 100005595 100005690 100005697 100005699 100005987 100006282 100006368 100006884 100006887 100006892 100006893 100006932 100006934 100007044 100007345 100007635 100007712 100008137 100008145 100008148 100008354 100008359 100008422 100008425 100008432 100008644 100008650 100008657 100008854 100008943 100008999 100009000 100009087 100009097 100009098 100009334 100009776 100009856 100009858 100009960 100009973 100010102 100010167 100010202 100010205 100010232 100010518 100011305 100011778 100012044 100012282 100012284 100012290 100012291 100012358 100012542 100012545 100012770 100012896 100012968 100013483 100013600 100013629 100013665 100013666 100013673 100013674 100013675 100015786 100016534 100017356 100024283 100024300 100025783 100025793 100025947 100026813 100026818 100026825 100026948 100027405 100027421 100027744 100029670 100029674 100029677 100029679 100029680 100029682 100029687 100029696 100029698 100029831 100029834 100029836 100029837 100029840 100029841 100029843 100029845 100029846 100029847 100029848 100029851 100029852 100029853 100029854 100029856 100029857 100029860 100030028 100030030 100030033 100030034 100031561 100031683 100032466 100032471 100032473 100032474 100032477 100032478 100032479 100032539 100033462 100034676 100034677 100034680 100034691 100034696 100034806 100034807 100034810 100034821 100035425 100035617 100035734 100035745 100035748 100035750 100035751 100035752 100035753 100035757 100035758 100037766 100037774 100037775 100038022 100038420 100038429 100038595 100038597 100038599 100038600 100038608 100038611 100038710 100038712 100038713 100038718 100038719 100038722 100038723 100038725 100038726 100038730 100038733 100039252 100040159 100040163 100040166 100040758 100040848 100040850 100040852 100040854 100040855 100040856 100041393 100041394 100042014 100042015 100042016 100042018 100042019 100042020 100042023 100042024 100042025 100042026 100042027 100042030 100042115 100043383 100043388 100043389 100043390 100043391 100043392 100043393 100043394 100043395 100043420 100043421 100043422 100043427 100043428 100043429 100043433 100043435 100043437 100043709 100043711 100044618 100044621 100044628 100044630 100044632 100044633 100044634 100044635 100044637 100044639 100044640 100044641 100044643 100044772 100044773 100044775 100044779 100044792 100044793 100044794 100044795 100044796 100044923 100045121 100045292 100045835 100045836 100045837 100045838 100045839 100045840 100045841 100045843 100045844 100045846 100045847 100045848 100045849 100045850 100045851 100045852 100045853 100045854 100045856 100045857 100045858 100045859 100045860 100045861 100045862 100046015 100046016 100046017 100046018 100046019 100046020 100046021 100046022 100046023 100046024 100046026 100046029 100046030 100046031 100046032 100046034 100047834 100047835 100048013 100048866 201000040 201000041 201000042 201000043 201000044 201000045 201000046 201000047 201000048 201000049 201000050 201000051 201000052 201000053 201000054 201000055 201000065 201000067 201000078 201000079 201000096 201000097 201000098 201000099 201000100 201000128 201000129 201000130 201000131 201000132 201000133 201000134 201000135 201000145 201000146 201000147 201000148 201000150 201000152 201000153 201000154 201000155 201000156 201000175 201000176 201000196 201000198 201000200 201000201 201000202 201000220 201000297 201000298 201000311 201000312 201000313 201000314 201000315 201000316 201000330 201000331 201000333 201000334 201000337 201000338 201000339 201000340 201000341 201000343 201000380 201000381 201000383 201000384 201000385 201000491 201000494 201000496 201001638 201002012 201002131 201002137 201002138 201002139 201002460 201002461 201002462 201002480 201002556 201002557 201002565 201002571 201002590 201002591 201002597 201002629 201002647 201002651 201002663 201002785 201002802 201002967 201003277 201003279 |
Recalling Firm/ Manufacturer |
Respironics California, LLC 2271 Cosmos Ct Carlsbad CA 92011-1517
|
For Additional Information Contact |
800-722-9377
|
Manufacturer Reason for Recall |
Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC power.
|
FDA Determined Cause 2 |
Device Design |
Action |
On 10/30/2020, the firm sent an "URGENT-MEDICAL DEVICE CORRECTION" Notification Letter via priority mail informing customer that the Ventilators may experience two potential issues with the battery (see table below for additional
information).
1. The ventilator may exhibit a high priority Check Vent: Battery Failed alarm if the battery is allowed to over discharge.
2. The battery may fail to charge if the ventilator is stored for an extended period of time without being plugged into AC power.
If the ventilator exhibits a high priority Battery Failed alarm while connected to
AC power, the V60 ventilator will annunciate an audible and visual alarm. The
ventilator will continue to deliver prescribed therapy; however, the audible and
visual alarm will continue until the device is powered down; therefore, the patient
should be transferred to an alternate source of ventilation as soon as one
becomes available. The hazard associated with transferring a patient to an
alternate source of ventilation is a temporary loss of therapy leading to potential
hypercarbia or hypoxemia (moderate).
If AC power fails and the backup battery is not functioning, an audible and visual
alarm will annunciate for at least two minutes. The ventilator will no longer be
operative resulting in a total loss of therapy. The patient will require an alternate
source of ventilation. The hazard associated with a ventilator shutdown is a total
loss of therapy leading to potential hypercarbia or hypoxemia (severe).
Customers are instructed:
Follow the instructions in the V60 user manual and the instructions below to
reduce any risk associated with potential battery issues.
1) Avoid allowing the ventilator battery to become completely discharged.
Otherwise, the battery may become over-discharged and require long
recharge times of up to 16 hours or more. The over-discharged condition
may permanently damage the battery so that it is unable to recharge. To
prevent the occurrence o |
Quantity in Commerce |
481 systems |
Distribution |
Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AS, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE,NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Afghanistan, Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, ST. Pier Miquel, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, Wallis Futuna, and Zambia. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
|
510(K) Database |
510(K)s with Product Code = MNT and Original Applicant = RESPIRONICS CALIFORNIA, INC.
|
|
|
|