 Class 2 Device Recall NA201SX4022C ASPIRATION NEEDLE ViziShot 22G w/o styletsee related information
Date Initiated by FirmJanuary 04, 2021
Create DateFebruary 05, 2021
Recall Status1 Terminated 3 on March 11, 2024
Recall NumberZ-1062-2021
Recall Event ID 87074
510(K)NumberK050503 
Product Classification Biopsy needle - Product Code FCG
ProductNA-201SX-4022-C ASPIRATION NEEDLE ViziShot 22G w/o stylet, model no. NA-201SX-4022-C - Product Usage: Used with ultrasound endoscopes for ultrasound guided fine needle aspiration of submucosal and extramural lesions of the tracheobronchial tree and esophagus.
Code Information 03K, 04K, 95V, 96V, 97V, 98V, 99V, 9XV, 9YV, 9ZV, 01V, 02V, 03V, 04V, 05V, 06V, 07V, 08V, 09V, 0XV
FEI Number 3003995201
Recalling Firm/
Manufacturer
Aomori Olympus Co., Ltd.
2 Chome 248-1
Okkonoki
Kuroishi Japan
For Additional Information ContactLaura Storms
484-8965688
Manufacturer Reason
for Recall
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
FDA Determined
Cause 2
Under Investigation by firm
ActionOn January 4, 2021, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were informed that the identified list of Endo-therapy devices may have a defective seal which may compromise the sterility of the product. Customers were asked to take the following actions: 1. Immediately assess any product you have in stock to identify endotherapy products with affected lot numbers listed in the recall letter. Cease use of product and quarantine any affected product. 2. Call your Olympus customer service representative at 1-888-524-7266 to obtain a Returned Goods Authorization so that you may return the product with no charge to you. Olympus will issue a credit or replacement to your facility for any returned product. 3. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number Recall-0391 and provide your contact information as indicated in the portal. 4. If you have distributed these devices outside your facility, please notify your customers of this corrective action immediately. Your notification to your customers may be enhanced by including a copy of this recall notification letter.
Quantity in Commerce7,400,000 (globally); 260,395 (US)
DistributionUS Nationwide distribution.
Total Product Life CycleTPLC Device Report