Date Initiated by Firm |
June 14, 2021 |
Create Date |
July 30, 2021 |
Recall Status1 |
Terminated 3 on March 08, 2023 |
Recall Number |
Z-2168-2021 |
Recall Event ID |
88210 |
Product Classification |
Labor and Delivery Kit - Product Code MLS
|
Product |
Regard NEWBORN KIT, LD00201K, Item Number: 830016011. Medical convenience kit which includes an infant heel warmer. |
Code Information |
Lot Numbers: 77906K, 78673K, 79327K, 80825K
|
Recalling Firm/ Manufacturer |
ROi CPS LLC 3000 E Sawyer Rd Republic MO 65738-2758
|
For Additional Information Contact |
Lacy Stewart 417-647-3205
|
Manufacturer Reason for Recall |
Instructions for use updated to include the following cautionary statement for the heel warmer, "Activate away from the user's face and away from the infant and others."
|
FDA Determined Cause 2 |
Process design |
Action |
The firm notified their consignee on 06/24/2021 by email. The notice provided updated product labeling with a cautionary statement to remind the user to "activate away from the user's face and away from the infant and others." The distributor was directed to notify their customer. |
Quantity in Commerce |
688 units |
Distribution |
Missouri |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|