Class 2 Device Recall Oakworks vascular exam table

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Oakworks vascular exam tablesee related information
Date Initiated by FirmJune 29, 2021
Create DateAugust 25, 2021
Recall Status1 Terminated 3 on October 05, 2023
Recall NumberZ-2349-2021
Recall Event ID 88365
Product Classification Table, examination, medical, powered - Product Code LGX
ProductVascular Table without Fowler (USV1) and Ultrasound Vascular Table with Fowler (USV2). Intended Use: Vascular Ultrasound examination table Catalog Number: 66783 and 64219
Code Information Serial Number/Part Number VCF766794 USGNV22723723MTTZZCE2S; VCF767303 USGNV23023723MTTZZ2S; VCF771999 USGNV23023723MTTZZ2S; VCF774252 USGNV23023723MTTZZ2S; VCF776073 USGNV23023723MTTZZ2S; VCF776198 USGNV23023723MTTZZ2S; VCF778246 USGNV23023723MTTZZ2S; VCF778247 USGNV23023723MTTZZ2S; VCF779362 USGNV23023723MTTZZ2S; VCF780260 USGNV23023723MTTZZ2S; VCS781126 USGNV13023723MTTZZ2S; VCF783652 USGNV22723723MTTZZF82S; VCF783969 USGNV23023723MTTZZF82S; VCF785961 USGNV22723723MTTZZCE2S; VCS766698 USGNV12723723MTTZZF62S; VCF768278 USGNV23023723MTTZZCEDRMC2S; VCF770016 USGNV23023723MTTZZ2S; VCF770017 USGNV23023723MTTZZ2S; VCF770571 USGNV23023723MTTZZCE2S; VCF771782 USGNV23023723MTTZZF8MC2S; VCF776974 USGNV22723723MTTZZ2S; VCF778602 USGNV23023723MTTZZ2S; VCF779609 USGNV23023723MTTZZF8CE2S; VCF781198 USGNV23023723MTTZZ2S; VCF784255 USGNV23023723MTTZZCE2S; VCF785376 USGNV23023723MTTZZ2S; VCF785650 USGNV23023723MTTZZF8MC2S; VCF786137 USGNV23023723MTTZZ2S; VCF786173 USGNV23023723MTTZZCE2S; VCF771553 USGNV23023723MTTZZCEMC2S; VCF771554 USGNV23023723MTTZZCEMC2S; VCF785065 USGNV22723723MTTZZF8CEDRDL2S; VCF771992 USGNV23023723MTTZZF8DRDL2S; VCF771993 USGNV23023723MTTZZF8DRDL2S; VCF782916 USGNV23023723MTTZZF8DRDL2S; VCF772035 USGNV22723723MTTZZ2S; VCF777288 USGNV22723723MTTZZ2S; VCF779024 USGNV22723723MTTZZ2S; VCF779025 USGNV22723723MTTZZ2S; VCF779026 USGNV22723723MTTZZ2S; VCF780977 USGNV22723723MTTZZ2S; VCF785219 USGNV22723723MTTZZ2S; VCF771311 USGNV23023723MTTZZ2S; VCF785744 USGNV22723723MTTZZCE2S; VCS772698 USGNV12723723MTTZZ2S; VCF771236 USGNV22723723MTTZZF8CE2S; VCF775658 USGNV23023723MTTZZ2S; VCF776370 USGNV23023723MTTZZ2S; VCF792903 USGNV23023723MTTZZMC2S; VCF773748 USGNV23023723MTTZZF8MC2S; VCF773748 USGNV23023723MTTZZCE2S; VCF779429 USGNV23023723MTTZZF8CE2S; VCF779921 USGNV23023723MTTZZF82S; VCF783233 USGNV23023723MTTZZ2S; VCF786329 USGNV22723723MTTZZ2S;
FEI Number 1000121370
Recalling Firm/
Manufacturer
Oakworks Inc
923 E Wellspring Rd
New Freedom PA 17349-8408
For Additional Information ContactSAME
717-235-6807
Manufacturer Reason
for Recall
Weld failure- the bracket attaches the tabletop tilting actuator to the tabletop frame. If the bracket weld fails, the tabletop tilting action is non-functional
FDA Determined
Cause 2
Process control
ActionOakworks issued Urgent Medical Device Recall letter dated 6/29/21. Letter states reason for recall , health risk and action to take: Actions to be taken are to immediately examine your inventory of OAKWORKS Inc., Ultrasound Vascular Table without Fowler (V1) and Ultrasound Vascular Table with Fowler (V2) to identify and locate all affected product(s). Do not distribute these affected product(s). Please quarantine the table(s). For tables that have been in use, they remain fully functional. Implement the field action as described below for the product listed in the Product and Distribution Information. Enclosed is a Medical Device Advisory Notice Response Form for each product we believe you have. Please complete this form in its entirety and return to OAKWORKS Inc., within 48 hours of receipt of this advisory notice via any of the options listed below. If you have further distributed this product, immediately contact your customer(s) with a copy of this letter and all applicable response forms. Then, provide OAKWORKS Inc. with the contact information for your customer(s) to whom this product(s) has been distributed by using the Medical Device Advisory Notice Response Form within 48 hours. Please note that you are required by the United States Food and Drug Administration to contact your customer(s) who are currently in possession of the product and inform them of this Medical Device Advisory Notice. MAIL: Oakworks Inc. EMAIL: qualitycare@oakworks.com Attn: Quality Care MDAN Response FAX: (717) 235-6798 923 East Wellspring Rd. New Freedom, PA 17349 What OAKWORKS Inc. will do next: Upon receipt of the Medical Device Advisory Notice Response Form(s), and to minimize the downtime of your OAKWORKS Ultrasound Vascular Table (V1) or (V2), we will setup a service visit to replace the Tilt Actuator Bracket on the table(s). The correction installation takes approximately 30 minutes. The replaced bracket will be disposed of by the service provider. If you have any quest
Quantity in Commerce55 units