|
Class 2 Device Recall Model B352000 Percept PC BrainSense Implantable Neurostimulator |
![see related information](../images/related-info.gif) |
Date Initiated by Firm |
October 19, 2021 |
Create Date |
December 27, 2021 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-0418-2022 |
Recall Event ID |
88889 |
PMA Number |
P960009 |
Product Classification |
Stimulator, electrical, implanted, for parkinsonian tremor - Product Code MHY
|
Product |
Model B35200 - Percept PC BrainSense Implantable Neurostimulator |
Code Information |
GTIN 00763000253363, Serial Numbers: NPI714835H, NPI714811H, NPI714812H, NPI714814H, NPI714816H, NPI714817H, NPI714819H, NPI714820H, NPI714750H, NPI714764H, NPI714771H, NPI714794H, NPI714682H, NPI714709H, NPI714710H, NPI714594H, NPI714598H, NPI714599H, NPI714602H, NPI714603H, NPI714606H, NPI714564H, NPI714572H, NPI714574H, NPI714587H, NPI714588H, NPI714589H, NPI714373H, NPI714379H, NPI714331H, NPI714346H, NPI714358H, NPI714378H, NPI714128H, NPI714160H, NPI714060H, NPI713947H, NPI713956H, NPI713957H, NPI714001H, NPI714017H, NPI713948H, NPI713949H, NPI713951H, NPI713952H, NPI713953H, NPI713954H, NPI713955H, NPI713829H, NPI713371H, NPI713372H, NPI713422H, NPI713242H, NPI713188H, NPI712940H, NPI712785H, NPI712637H, NPI712405H, NPI712411H, NPI712412H, NPI712341H, NPI712302H, NPI712256H, NPI711876H, NPI711841H, NPI711475H, NPI711476H, NPI710876H, NPI710765H, NPI710561H, NPI710071H, NPI709964H, NPI709680H, NPI709654H, NPI709655H, NPI709575H, NPI708692H, NPI708699H, NPI708702H, NPI708709H, NPI708675H, NPI708653H, NPI708635H, NPI708457H, NPI708415H, NPI708417H, NPI708425H, NPI708367H, NPI708376H, NPI708358H, NPI708316H, NPI708288H, NPI708289H, NPI708292H, NPI708295H, NPI708299H, NPI708300H, NPI708305H, NPI708306H, NPI708309H, NPI708258H, NPI708262H, NPI708273H, NPI708127H, NPI708128H, NPI708153H, NPI708101H, NPI708102H, NPI708103H, NPI708106H, NPI708107H, NPI708110H, NPI708112H, NPI708113H, NPI708114H, NPI708117H, NPI708118H, NPI708120H, NPI708121H, NPI708123H, NPI708124H, NPI708125H, NPI708079H, NPI708080H, NPI708082H, NPI708083H, NPI708084H, NPI708085H, NPI708086H, NPI708087H, NPI708088H, NPI708089H, NPI708090H, NPI708091H, NPI708092H, NPI708093H, NPI708094H, NPI708095H, NPI708096H, NPI708097H, NPI708100H, NPI708105H, NPI708108H, NPI708109H, NPI708116H, NPI708126H, NPI708056H, NPI708059H, NPI708060H, NPI708062H, NPI708063H, NPI708064H, NPI708065H, NPI708066H, NPI708067H, NPI708068H, NPI708070H, NPI708071H, NPI708073H, NPI708074H, NPI708076H, NPI708077H, NPI708078H, NPI708081H, NPI708014H, NPI708037H, NPI708038H, NPI708039H, NPI708040H, NPI708041H, NPI708042H, NPI708043H, NPI708044H, NPI708045H, NPI708046H, NPI708048H, NPI708049H, NPI708050H, NPI708055H, NPI708057H, NPI707998H, NPI707999H, NPI708001H, NPI708002H, NPI708003H, NPI708004H, NPI708005H, NPI708007H, NPI708008H, NPI708009H, NPI708011H, NPI708012H, NPI708013H, NPI708015H, NPI708016H, NPI708017H, NPI708018H, NPI708020H, NPI708021H, NPI708022H, NPI708023H, NPI708024H, NPI708025H, NPI708026H, NPI708027H, NPI708028H, NPI708030H, NPI708031H, NPI708033H, NPI708034H, NPI708036H, NPI707947H, NPI707955H, NPI707956H, NPI707961H, NPI707964H, NPI707970H, NPI707972H, NPI707975H, NPI707978H, NPI707979H, NPI707982H, NPI707983H, NPI707985H, NPI707986H, NPI707987H, NPI707989H, NPI707990H, NPI707991H, NPI707992H, NPI707993H, NPI707995H, NPI707996H, NPI707997H, NPI708000H, NPI707942H, NPI707943H, NPI707944H, NPI707945H, NPI707946H, NPI707948H, NPI707949H, NPI707950H, NPI707951H, NPI707952H, NPI707953H, NPI707954H, NPI707957H, NPI707958H, NPI707959H, NPI707962H, NPI707963H, NPI707966H, NPI707968H, NPI707969H, NPI707971H, NPI707974H, NPI707976H, NPI707977H, NPI707918H, NPI707919H, NPI707920H, NPI707922H, NPI707923H, NPI707924H, NPI707925H, NPI707927H, NPI707928H, NPI707929H, NPI707930H, NPI707931H, NPI707932H, NPI707934H, NPI707935H, NPI707937H, NPI707938H, NPI707939H, NPI707880H, NPI707881H, NPI707883H, NPI707884H, NPI707885H, NPI707886H, NPI707888H, NPI707890H, NPI707905H, NPI707907H, NPI707914H, NPI707916H, NPI707870H, NPI707871H, NPI707872H, NPI707873H, NPI707874H, NPI707875H, NPI707877H, NPI707879H, NPI707889H, NPI707893H, NPI707894H, NPI707895H, NPI707896H, NPI707897H, NPI707898H, NPI707899H, NPI707900H, NPI707901H, NPI707902H, NPI707846H, NPI707847H, NPI707848H, NPI707849H, NPI707856H, NPI707857H, NPI707858H, NPI707860H, NPI707861H, NPI707862H, NPI707864H, NPI707865H, NPI707867H, NPI707868H, NPI707869H, NPI707828H, NPI707830H, NPI707831H, NPI707832H, NPI707833H, NPI707834H, NPI707835H, NPI707836H, NPI707837H, NPI707838H, NPI707839H, NPI707840H, NPI707841H, NPI707842H, NPI707843H, NPI707844H, NPI707845H, NPI707851H, NPI707852H, NPI707855H, NPI707811H, NPI707812H, NPI707813H, NPI707814H, NPI707816H, NPI707818H, NPI707819H, NPI707820H, NPI707821H, NPI707822H, NPI707824H, NPI707791H, NPI707792H, NPI707794H, NPI707795H, NPI707797H, NPI707798H, NPI707799H, NPI707800H, NPI707802H, NPI707804H, NPI707805H, NPI707806H, NPI707807H, NPI707808H, NPI707809H, NPI707774H, NPI707776H, NPI707777H, NPI707779H, NPI707780H, NPI707782H, NPI707783H, NPI707784H, NPI707786H, NPI707788H, NPI707789H, NPI707790H, NPI707749H, NPI707750H, NPI707752H, NPI707753H, NPI707754H, NPI707755H, NPI707756H, NPI707757H, NPI707758H, NPI707762H, NPI707763H, NPI707764H, NPI707765H, NPI707766H, NPI707767H, NPI707768H, NPI707769H, NPI707770H, NPI707771H, NPI707772H, NPI707720H, NPI707721H, NPI707722H, NPI707723H, NPI707724H, NPI707725H, NPI707727H, NPI707728H, NPI707730H, NPI707731H, NPI707732H, NPI707733H, NPI707734H, NPI707735H, NPI707737H, NPI707738H, NPI707739H, NPI707740H, NPI707741H, NPI707743H, NPI707744H, NPI707745H, NPI707712H, NPI707713H, NPI707714H, NPI707684H, NPI707689H, NPI707690H, NPI707691H, NPI707696H, NPI707698H, NPI707699H, NPI707700H, NPI707702H, NPI707704H, NPI707705H, NPI707707H, NPI707708H, NPI707664H, NPI707666H, NPI707668H, NPI707672H, NPI707673H, NPI707674H, NPI707675H, NPI707676H, NPI707677H, NPI707679H, NPI707681H, NPI707682H, NPI707683H, NPI707686H, NPI707688H, NPI707648H, NPI707649H, NPI707650H, NPI707651H, NPI707652H, NPI707654H, NPI707655H, NPI707656H, NPI707658H, NPI707659H, NPI707660H, NPI707661H, NPI707662H, NPI707663H, NPI707670H, NPI707671H, NPI707592H, NPI707597H, NPI707598H, NPI707599H, NPI707617H, NPI707621H, NPI707622H, NPI707623H, NPI707624H, NPI707625H, NPI707626H, NPI707627H, NPI707628H, NPI707629H, NPI707630H, NPI707631H, NPI707632H, NPI707633H, NPI707635H, NPI707637H, NPI707639H, NPI707640H, NPI707641H, NPI707642H, NPI707643H, NPI707644H, NPI707645H, NPI707646H, NPI707647H, NPI707653H, NPI707657H, NPI707581H, NPI707582H, NPI707583H, NPI707585H, NPI707586H, NPI707587H, NPI707588H, NPI707589H, NPI707590H, NPI707591H, NPI707593H, NPI707594H, NPI707595H, NPI707596H, NPI707600H, NPI707601H, NPI707602H, NPI707603H, NPI707605H, NPI707606H, NPI707607H, NPI707610H, NPI707611H, NPI707612H, NPI707613H, NPI707614H, NPI707615H, NPI707616H, NPI707618H, NPI707569H, NPI707570H, NPI707571H, NPI707572H, NPI707573H, NPI707574H, NPI707575H, NPI707576H, NPI707577H, NPI707578H, NPI707579H, NPI707580H, NPI707536H, NPI707537H, NPI707538H, NPI707539H, NPI707541H, NPI707542H, NPI707543H, NPI707544H, NPI707545H, NPI707546H, NPI707547H, NPI707548H, NPI707549H, NPI707550H, NPI707551H, NPI707552H, NPI707553H, NPI707554H, NPI707555H, NPI707556H, NPI707557H, NPI707558H, NPI707559H, NPI707560H, NPI707561H, NPI707562H, NPI707563H, NPI707564H, NPI707565H, NPI707566H, NPI707567H, NPI707568H, NPI707513H, NPI707514H, NPI707515H, NPI707516H, NPI707517H, NPI707518H, NPI707519H, NPI707520H, NPI707521H, NPI707522H, NPI707523H, NPI707524H, NPI707525H, NPI707526H, NPI707530H, NPI707531H, NPI707533H, NPI707534H, NPI707535H, NPI707484H, NPI707485H, NPI707490H, NPI707491H, NPI707492H, NPI707493H, NPI707494H, NPI707495H, NPI707496H, NPI707497H, NPI707498H, NPI707500H, NPI707501H, NPI707502H, NPI707506H, NPI707507H, NPI707508H, NPI707509H, NPI707439H, NPI707440H, NPI707441H, NPI707443H, NPI707444H, NPI707445H, NPI707446H, NPI707447H, NPI707449H, NPI707450H, NPI707452H, NPI707453H, NPI707454H, NPI707455H, NPI707456H, NPI707457H, NPI707458H, NPI707459H, NPI707460H, NPI707461H, NPI707462H, NPI707463H, NPI707464H, NPI707465H, NPI707466H, NPI707467H, NPI707468H, NPI707469H, NPI707470H, NPI707471H, NPI707472H, NPI707473H, NPI707474H, NPI707476H, NPI707477H, NPI707478H, NPI707479H, NPI707480H, NPI707482H, NPI707483H, NPI707486H, NPI707487H, NPI707488H, NPI707402H, NPI707406H, NPI707407H, NPI707408H, NPI707410H, NPI707412H, NPI707414H, NPI707415H, NPI707416H, NPI707417H, NPI707418H, NPI707420H, NPI707421H, NPI707422H, NPI707423H, NPI707424H, NPI707425H, NPI707426H, NPI707427H, NPI707428H, NPI707429H, NPI707430H, NPI707431H, NPI707432H, NPI707433H, NPI707434H, NPI707435H, NPI707436H, NPI707437H, NPI707438H, NPI707351H, NPI707352H, NPI707353H, NPI707354H, NPI707355H, NPI707357H, NPI707358H, NPI707359H, NPI707374H, NPI707375H, NPI707376H, NPI707377H, NPI707378H, NPI707379H, NPI707380H, NPI707381H, NPI707382H, NPI707383H, NPI707386H, NPI707394H, NPI707396H, NPI707397H, NPI707398H, NPI707399H, NPI707400H, NPI707401H, NPI707403H, NPI707404H, NPI707336H, NPI707338H, NPI707339H, NPI707341H, NPI707342H, NPI707343H, NPI707364H, NPI707367H, NPI707368H, NPI707369H, NPI707373H, NPI707385H, NPI707387H, NPI707388H, NPI707389H, NPI707390H, NPI707392H, NPI707393H, NPI707395H, NPI707306H, NPI707308H, NPI707309H, NPI707310H, NPI707311H, NPI707312H, NPI707313H, NPI707315H, NPI707316H, NPI707317H, NPI707318H, NPI707319H, NPI707320H, NPI707321H, NPI707322H, NPI707323H, NPI707326H, NPI707327H, NPI707328H, NPI707329H, NPI707331H, NPI707332H, NPI707334H, NPI707335H, NPI707277H, NPI707278H, NPI707279H, NPI707280H, NPI707282H, NPI707283H, NPI707284H, NPI707285H, NPI707286H, NPI707287H, NPI707288H, NPI707289H, NPI707290H, NPI707291H, NPI707292H, NPI707293H, NPI707294H, NPI707296H, NPI707297H, NPI707299H, NPI707300H, NPI707301H, NPI707302H, NPI707303H, NPI707304H, NPI707307H, NPI707314H, NPI707249H, NPI707250H, NPI707252H, NPI707253H, NPI707255H, NPI707256H, NPI707257H, NPI707258H, NPI707259H, NPI707261H, NPI707262H, NPI707263H, NPI707268H, NPI707269H, NPI707270H, NPI707271H, NPI707272H, NPI707273H, NPI707274H, NPI707275H, NPI707276H, NPI707208H, NPI707214H, NPI707215H, NPI707216H, NPI707218H, NPI707219H, NPI707220H, NPI707221H, NPI707222H, NPI707223H, NPI707224H, NPI707225H, NPI707226H, NPI707227H, NPI707229H, NPI707230H, NPI707232H, NPI707233H, NPI707235H, NPI707236H, NPI707237H, NPI707238H, NPI707239H, NPI707240H, NPI707241H, NPI707243H, NPI707244H, NPI707245H, NPI707246H, NPI707184H, NPI707185H, NPI707187H, NPI707189H, NPI707190H, NPI707192H, NPI707193H, NPI707196H, NPI707199H, NPI707200H, NPI707201H, NPI707202H, NPI707203H, NPI707204H, NPI707205H, NPI707206H, NPI707207H, NPI707209H, NPI707210H, NPI707211H, NPI707212H, NPI707213H, NPI707161H, NPI707162H, NPI707163H, NPI707165H, NPI707166H, NPI707167H, NPI707168H, NPI707170H, NPI707171H, NPI707172H, NPI707175H, NPI707177H, NPI707178H, NPI707181H, NPI707131H, NPI707134H, NPI707137H, NPI707138H, NPI707139H, NPI707140H, NPI707141H, NPI707142H, NPI707143H, NPI707144H, NPI707145H, NPI707146H, NPI707148H, NPI707149H, NPI707150H, NPI707151H, NPI707156H, NPI707157H, NPI707159H, NPI707160H, NPI707103H, NPI707104H, NPI707105H, NPI707106H, NPI707107H, NPI707109H, NPI707110H, NPI707111H, NPI707112H, NPI707114H, NPI707115H, NPI707116H, NPI707117H, NPI707119H, NPI707121H, NPI707122H, NPI707123H, NPI707124H, NPI707126H, NPI707127H, NPI707129H, NPI707130H, NPI707135H, NPI707136H, NPI707084H, NPI707085H, NPI707086H, NPI707089H, NPI707092H, NPI707094H, NPI707095H, NPI707096H, NPI707097H, NPI707098H, NPI707099H, NPI707100H, NPI707102H, NPI707045H, NPI707050H, NPI707053H, NPI707054H, NPI707055H, NPI707056H, NPI707057H, NPI707058H, NPI707059H, NPI707061H, NPI707062H, NPI707063H, NPI707065H, NPI707066H, NPI707069H, NPI707070H, NPI707072H, NPI707073H, NPI707074H, NPI707075H, NPI707076H, NPI707077H, NPI707078H, NPI707080H, NPI707081H, NPI707082H, NPI707083H, NPI707020H, NPI707025H, NPI707026H, NPI707027H, NPI707028H, NPI707029H, NPI707030H, NPI707031H, NPI707032H, NPI707034H, NPI707035H, NPI707036H, NPI707037H, NPI707038H, NPI707039H, NPI707041H, NPI707042H, NPI707043H, NPI707046H, NPI707047H, NPI707048H, NPI707051H, NPI707052H, NPI706993H, NPI706994H, NPI706995H, NPI706996H, NPI706997H, NPI706998H, NPI706999H, NPI707002H, NPI707003H, NPI707004H, NPI707005H, NPI707006H, NPI707007H, NPI707008H, NPI707011H, NPI707012H, NPI707013H, NPI707014H, NPI707015H, NPI707016H, NPI707017H, NPI707018H, NPI707019H, NPI707021H, NPI707023H, NPI706956H, NPI706959H, NPI706961H, NPI706962H, NPI706967H, NPI706968H, NPI706969H, NPI706970H, NPI706972H, NPI706973H, NPI706974H, NPI706975H, NPI706976H, NPI706977H, NPI706978H, NPI706979H, NPI706980H, NPI706981H, NPI706983H, NPI706985H, NPI706986H, NPI706989H, NPI706990H, NPI706991H, NPI706928H, NPI706929H, NPI706930H, NPI706931H, NPI706932H, NPI706933H, NPI706934H, NPI706935H, NPI706938H, NPI706939H, NPI706940H, NPI706941H, NPI706942H, NPI706943H, NPI706944H, NPI706946H, NPI706947H, NPI706948H, NPI706949H, NPI706951H, NPI706953H, NPI706964H, NPI706898H, NPI706899H, NPI706900H, NPI706901H, NPI706902H, NPI706903H, NPI706904H, NPI706905H, NPI706906H, NPI706907H, NPI706908H, NPI706909H, NPI706910H, NPI706911H, NPI706912H, NPI706913H, NPI706914H, NPI706915H, NPI706916H, NPI706917H, NPI706919H, NPI706920H, NPI706921H, NPI706922H, NPI706923H, NPI706925H, NPI706926H, NPI706879H, NPI706880H, NPI706881H, NPI706882H, NPI706883H, NPI706884H, NPI706885H, NPI706886H, NPI706887H, NPI706888H, NPI706889H, NPI706890H, NPI706891H, NPI706892H, NPI706893H, NPI706894H, NPI706895H, NPI706897H, NPI706876H, NPI706878H, NPI706856H, NPI706858H, NPI706861H, NPI706864H, NPI706867H, NPI706870H, NPI706871H, NPI706872H, NPI706822H, NPI706823H, NPI706824H, NPI706825H, NPI706826H, NPI706828H, NPI706829H, NPI706830H, NPI706831H, NPI706832H, NPI706833H, NPI706834H, NPI706835H, NPI706836H, NPI706837H, NPI706838H, NPI706839H, NPI706840H, NPI706841H, NPI706842H, NPI706843H, NPI706844H, NPI706845H, NPI706846H, NPI706847H, NPI706848H, NPI706849H, NPI706850H, NPI706851H, NPI706852H, NPI706853H, NPI706854H, NPI706855H, NPI706857H, NPI706863H, NPI706865H, NPI706866H, NPI706793H, NPI706794H, NPI706795H, NPI706798H, NPI706800H, NPI706801H, NPI706802H, NPI706803H, NPI706804H, NPI706805H, NPI706806H, NPI706807H, NPI706808H, NPI706809H, NPI706810H, NPI706811H, NPI706812H, NPI706813H, NPI706814H, NPI706815H, NPI706816H, NPI706817H, NPI706818H, NPI706819H, NPI706820H, NPI706827H, NPI706764H, NPI706766H, NPI706767H, NPI706768H, NPI706769H, NPI706770H, NPI706771H, NPI706772H, NPI706773H, NPI706776H, NPI706777H, NPI706778H, NPI706779H, NPI706780H, NPI706781H, NPI706782H, NPI706783H, NPI706784H, NPI706785H, NPI706786H, NPI706787H, NPI706788H, NPI706789H, NPI706790H, NPI706791H, NPI706792H, NPI706799H, NPI706721H, NPI706722H, NPI706723H, NPI706724H, NPI706726H, NPI706731H, NPI706732H, NPI706733H, NPI706734H, NPI706735H, NPI706736H, NPI706737H, NPI706738H, NPI706739H, NPI706740H, NPI706741H, NPI706742H, NPI706743H, NPI706746H, NPI706747H, NPI706749H, NPI706750H, NPI706751H, NPI706752H, NPI706754H, NPI706755H, NPI706756H, NPI706757H, NPI706759H, NPI706760H, NPI706763H, NPI706692H, NPI706702H, NPI706703H, NPI706704H, NPI706705H, NPI706706H, NPI706708H, NPI706709H, NPI706710H, NPI706711H, NPI706712H, NPI706714H, NPI706715H, NPI706717H, NPI706718H, NPI706719H, NPI706720H, NPI706725H, NPI706727H, NPI706730H, NPI706682H, NPI706683H, NPI706684H, NPI706685H, NPI706686H, NPI706687H, NPI706688H, NPI706689H, NPI706690H, NPI706691H, NPI706693H, NPI706694H, NPI706695H, NPI706696H, NPI706697H, NPI706698H, NPI706699H, NPI706700H, NPI706701H, NPI706707H, NPI706659H, NPI706667H, NPI706668H, NPI706669H, NPI706670H, NPI706671H, NPI706673H, NPI706674H, NPI706675H, NPI706676H, NPI706677H, NPI706678H, NPI706679H, NPI706617H, NPI706618H, NPI706619H, NPI706620H, NPI706621H, NPI706622H, NPI706623H, NPI706624H, NPI706625H, NPI706626H, NPI706627H, NPI706628H, NPI706629H, NPI706630H, NPI706631H, NPI706632H, NPI706633H, NPI706634H, NPI706635H, NPI706636H, NPI706637H, NPI706639H, NPI706645H, NPI706646H, NPI706647H, NPI706648H, NPI706649H, NPI706650H, NPI706651H, NPI706652H, NPI706653H, NPI706654H, NPI706655H, NPI706656H, NPI706660H, NPI706661H, NPI706662H, NPI706663H, NPI706664H, NPI706665H, NPI706666H, NPI706577H, NPI706579H, NPI706581H, NPI706585H, NPI706585H, NPI706587H, NPI706588H, NPI706589H, NPI706591H, NPI706592H, NPI706594H, NPI706595H, NPI706596H, NPI706597H, NPI706599H, NPI706600H, NPI706601H, NPI706602H, NPI706603H, NPI706604H, NPI706605H, NPI706606H, NPI706607H, NPI706608H, NPI706609H, NPI706610H, NPI706611H, NPI706612H, NPI706613H, NPI706614H, NPI706615H, NPI706616H, NPI706545H, NPI706546H, NPI706547H, NPI706549H, NPI706550H, NPI706551H, NPI706552H, NPI706554H, NPI706555H, NPI706559H, NPI706560H, NPI706561H, NPI706562H, NPI706563H, NPI706578H, NPI706508H, NPI706511H, NPI706512H, NPI706513H, NPI706516H, NPI706517H, NPI706518H, NPI706523H, NPI706524H, NPI706525H, NPI706526H, NPI706527H, NPI706528H, NPI706529H, NPI706530H, NPI706532H, NPI706533H, NPI706534H, NPI706538H, NPI706539H, NPI706540H, NPI706541H, NPI706542H, NPI706543H, NPI706544H, NPI706558H, NPI706494H, NPI706495H, NPI706496H, NPI706497H, NPI706498H, NPI706499H, NPI706500H, NPI706501H, NPI706502H, NPI706503H, NPI706504H, NPI706505H, NPI706506H, NPI706507H, NPI706510H, NPI706519H, NPI706520H, NPI706439H, NPI706440H, NPI706453H, NPI706454H, NPI706455H, NPI706457H, NPI706458H, NPI706464H, NPI706465H, NPI706470H, NPI706475H, NPI706417H, NPI706418H, NPI706419H, NPI706421H, NPI706422H, NPI706423H, NPI706424H, NPI706425H, NPI706427H, NPI706428H, NPI706429H, NPI706431H, NPI706432H, NPI706433H, NPI706434H, NPI706435H, NPI706437H, NPI706438H, NPI706376H, NPI706377H, NPI706378H, NPI706379H, NPI706380H, NPI706382H, NPI706383H, NPI706384H, NPI706386H, NPI706389H, NPI706391H, NPI706392H, NPI706393H, NPI706394H, NPI706395H, NPI706396H, NPI706398H, NPI706399H, NPI706400H, NPI706401H, NPI706402H, NPI706403H, NPI706404H, NPI706405H, NPI706406H, NPI706407H, NPI706408H, NPI706409H, NPI706410H, NPI706411H, NPI706412H, NPI706413H, NPI706414H, NPI706415H, NPI706416H, NPI706341H, NPI706316H, NPI706274H, NPI706276H, NPI706277H, NPI706278H, NPI706279H, NPI706280H, NPI706281H, NPI706282H, NPI706283H, NPI706284H, NPI706285H, NPI706286H, NPI706287H, NPI706288H, NPI706236H, NPI706238H, NPI706239H, NPI706240H, NPI706242H, NPI706243H, NPI706244H, NPI706245H, NPI706247H, NPI706248H, NPI706250H, NPI706251H, NPI706252H, NPI706253H, NPI706254H, NPI706256H, NPI706257H, NPI706258H, NPI706259H, NPI706261H, NPI706262H, NPI706264H, NPI706265H, NPI706266H, NPI706268H, NPI706269H, NPI706270H, NPI706271H, NPI706272H, NPI706273H, NPI706225H, NPI706226H, NPI706227H, NPI706228H, NPI706229H, NPI706230H, NPI706231H, NPI706233H, NPI706234H, NPI706235H, NPI706156H, NPI706157H, NPI706158H, NPI706159H, NPI706160H, NPI706162H, NPI706163H, NPI706164H, NPI706165H, NPI706166H, NPI706167H, NPI706168H, NPI706169H, NPI706170H, NPI706171H, NPI706172H, NPI706174H, NPI706175H, NPI706176H, NPI706177H, NPI706178H, NPI706179H, NPI706180H, NPI706181H, NPI706182H, NPI706183H, NPI706185H, NPI706186H, NPI706122H, NPI706123H, NPI706124H, NPI706127H, NPI706130H, NPI706131H, NPI706132H, NPI706133H, NPI706134H, NPI706135H, NPI706137H, NPI706138H, NPI706139H, NPI706140H, NPI706141H, NPI706142H, NPI706143H, NPI706144H, NPI706146H, NPI706148H, NPI706149H, NPI706150H, NPI706151H, NPI706152H, NPI706153H, NPI706154H, NPI706155H, NPI706094H, NPI706095H, NPI706096H, NPI706097H, NPI706098H, NPI706100H, NPI706101H, NPI706102H, NPI706103H, NPI706104H, NPI706105H, NPI706106H, NPI706108H, NPI706109H, NPI706111H, NPI706112H, NPI706113H, NPI706115H, NPI706117H, NPI706118H, NPI706119H, NPI706120H, NPI706059H, NPI706064H, NPI706066H, NPI706067H, NPI706069H, NPI706070H, NPI706071H, NPI706072H, NPI706078H, NPI706079H, NPI706081H, NPI706082H, NPI706083H, NPI706084H, NPI706085H, NPI706086H, NPI706087H, NPI706089H, NPI706090H, NPI706091H, NPI706092H, NPI706093H, NPI706099H, NPI706021H, NPI706022H, NPI706023H, NPI706025H, NPI706026H, NPI706027H, NPI706028H, NPI706029H, NPI706030H, NPI706031H, NPI706032H, NPI706033H, NPI706035H, NPI706037H, NPI706039H, NPI706040H, NPI706041H, NPI706042H, NPI706044H, NPI706045H, NPI706046H, NPI706047H, NPI706048H, NPI706049H, NPI706050H, NPI706051H, NPI706052H, NPI706053H, NPI706055H, NPI706056H, NPI706057H, NPI706058H, NPI706060H, NPI706062H, NPI706065H, NPI706068H, NPI706012H, NPI706013H, NPI706014H, NPI706015H, NPI706016H, NPI706017H, NPI706018H, NPI706019H, NPI706020H, NPI705987H, NPI705996H, NPI705999H, NPI706001H, NPI706002H, NPI706004H, NPI706008H, NPI706010H, NPI706011H, NPI705976H, NPI705977H, NPI705978H, NPI705979H, NPI705980H, NPI705981H, NPI705982H, NPI705983H, NPI705984H, NPI705985H, NPI705986H, NPI705989H, NPI705990H, NPI705991H, NPI705992H, NPI705993H, NPI705994H, NPI705997H, NPI705998H, NPI706000H, NPI706006H, NPI706007H, NPI705906H, NPI705908H, NPI705909H, NPI705910H, NPI705911H, NPI705912H, NPI705914H, NPI705915H, NPI705916H, NPI705918H, NPI705919H, NPI705920H, NPI705921H, NPI705922H, NPI705924H, NPI705925H, NPI705926H, NPI705927H, NPI705932H, NPI705933H, NPI705937H, NPI705938H, NPI705939H, NPI705940H, NPI705945H, NPI705950H, NPI705863H, NPI705865H, NPI705866H, NPI705867H, NPI705869H, NPI705871H, NPI705889H, NPI705890H, NPI705892H, NPI705893H, NPI705894H, NPI705896H, NPI705897H, NPI705899H, NPI705900H, NPI705903H, NPI705904H, NPI705907H, NPI705913H, NPI705811H, NPI705812H, NPI705813H, NPI705814H, NPI705815H, NPI705816H, NPI705817H, NPI705819H, NPI705822H, NPI705834H, NPI705835H, NPI705836H, NPI705837H, NPI705838H, NPI705840H, NPI705842H, NPI705843H, NPI705845H, NPI705846H, NPI705847H, NPI705848H, NPI705849H, NPI705850H, NPI705851H, NPI705852H, NPI705853H, NPI705854H, NPI705856H, NPI705857H, NPI705858H, NPI705859H, NPI705860H, NPI705861H, NPI705806H, NPI705807H, NPI705808H, NPI705810H, NPI705818H, NPI705820H, NPI705823H, NPI705826H, NPI705827H, NPI705828H, NPI705830H, NPI705792H, NPI705793H, NPI705794H, NPI705796H, NPI705797H, NPI705798H, NPI705800H, NPI705801H, NPI705803H, NPI705804H, NPI705805H, NPI705595H, NPI705597H, NPI705579H, NPI705580H, NPI705581H, NPI705593H, NPI705539H, NPI705542H, NPI705543H, NPI705544H, NPI705545H, NPI705546H, NPI705547H, NPI705550H, NPI705551H, NPI705552H, NPI705553H, NPI705554H, NPI705555H, NPI705556H, NPI705557H, NPI705558H, NPI705559H, NPI705560H, NPI705561H, NPI705562H, NPI705563H, NPI705564H, NPI705565H, NPI705566H, NPI705567H, NPI705568H, NPI705569H, NPI705570H, NPI705574H, NPI705575H, NPI705578H, NPI705518H, NPI705519H, NPI705520H, NPI705521H, NPI705522H, NPI705523H, NPI705524H, NPI705525H, NPI705526H, NPI705527H, NPI705528H, NPI705529H, NPI705530H, NPI705531H, NPI705532H, NPI705533H, NPI705534H, NPI705535H, NPI705509H, NPI705510H, NPI705511H, NPI705512H, NPI705513H, NPI705514H, NPI705515H, NPI705516H, NPI705517H, NPI705465H, NPI705466H, NPI705467H, NPI705468H, NPI705473H, NPI705474H, NPI705475H, NPI705476H, NPI705478H, NPI705479H, NPI705480H, NPI705481H, NPI705482H, NPI705485H, NPI705486H, NPI705487H, NPI705488H, NPI705489H, NPI705490H, NPI705491H, NPI705492H, NPI705493H, NPI705494H, NPI705496H, NPI705497H, NPI705498H, NPI705499H, NPI705500H, NPI705501H, NPI705503H, NPI705504H, NPI705505H, NPI705506H, NPI705507H, NPI705508H, NPI705464H, NPI705469H, NPI705470H, NPI705471H, NPI705457H, NPI705458H, NPI705459H, NPI705460H, NPI705461H, NPI705462H, NPI705463H, NPI705426H, NPI705427H, NPI705428H, NPI705429H, NPI705430H, NPI705431H, NPI705432H, NPI705433H, NPI705434H, NPI705435H, NPI705436H, NPI705437H, NPI705438H, NPI705439H, NPI705440H, NPI705441H, NPI705442H, NPI705443H, NPI705444H, NPI705445H, NPI705446H, NPI705447H, NPI705448H, NPI705450H, NPI705451H, NPI705452H, NPI705453H, NPI705454H, NPI705407H, NPI705408H, NPI705409H, NPI705410H, NPI705411H, NPI705412H, NPI705413H, NPI705414H, NPI705418H, NPI705419H, NPI705420H, NPI705421H, NPI705422H, NPI705423H, NPI705381H, NPI705382H, NPI705384H, NPI705385H, NPI705386H, NPI705398H, NPI705351H, NPI705353H, NPI705354H, NPI705357H, NPI705358H, NPI705359H, NPI705360H, NPI705362H, NPI705363H, NPI705368H, NPI705369H, NPI705374H, NPI705375H, NPI705376H, NPI705378H, NPI705379H, NPI705380H, NPI705387H, NPI705391H, NPI705392H, NPI705395H, NPI705396H, NPI705330H, NPI705332H, NPI705333H, NPI705334H, NPI705335H, NPI705336H, NPI705292H, NPI705293H, NPI705294H, NPI705295H, NPI705296H, NPI705297H, NPI705298H, NPI705299H, NPI705300H, NPI705301H, NPI705302H, NPI705303H, NPI705304H, NPI705305H, NPI705306H, NPI705307H, NPI705308H, NPI705309H, NPI705310H, NPI705311H, NPI705312H, NPI705313H, NPI705314H, NPI705315H, NPI705316H, NPI705317H, NPI705318H, NPI705319H, NPI705320H, NPI705321H, NPI705322H, NPI705323H, NPI705324H, NPI705325H, NPI705326H, NPI705327H, NPI705328H, NPI705329H, NPI705331H, NPI705337H, NPI705339H, NPI705344H, NPI705345H, NPI705346H, NPI705348H, NPI705350H, NPI705252H, NPI705253H, NPI705254H, NPI705255H, NPI705257H, NPI705258H, NPI705260H, NPI705263H, NPI705264H, NPI705265H, NPI705266H, NPI705267H, NPI705268H, NPI705269H, NPI705270H, NPI705271H, NPI705272H, NPI705273H, NPI705274H, NPI705275H, NPI705275H, NPI705277H, NPI705278H, NPI705279H, NPI705281H, NPI705282H, NPI705283H, NPI705284H, NPI705285H, NPI705286H, NPI705288H, NPI705289H, NPI705290H, NPI705291H, NPI705231H, NPI705234H, NPI705235H, NPI705236H, NPI705237H, NPI705238H, NPI705241H, NPI705242H, NPI705244H, NPI705245H, NPI705246H, NPI705246H, NPI705247H, NPI705249H, NPI705250H, NPI705175H, NPI705176H, NPI705177H, NPI705178H, NPI705179H, NPI705185H, NPI705186H, NPI705187H, NPI705188H, NPI705189H, NPI705191H, NPI705192H, NPI705193H, NPI705194H, NPI705195H, NPI705196H, NPI705197H, NPI705198H, NPI705199H, NPI705201H, NPI705204H, NPI705206H, NPI705210H, NPI705145H, NPI705146H, NPI705147H, NPI705150H, NPI705151H, NPI705152H, NPI705153H, NPI705154H, NPI705157H, NPI705158H, NPI705159H, NPI705160H, NPI705161H, NPI705162H, NPI705163H, NPI705165H, NPI705166H, NPI705167H, NPI705168H, NPI705169H, NPI705170H, NPI705171H, NPI705172H, NPI705173H, NPI705174H, NPI705180H, NPI705183H, NPI705184H, NPI705113H, NPI705114H, NPI705120H, NPI705121H, NPI705122H, NPI705123H, NPI705124H, NPI705126H, NPI705127H, NPI705129H, NPI705130H, NPI705131H, NPI705132H, NPI705133H, NPI705134H, NPI705136H, NPI705137H, NPI705138H, NPI705139H, NPI705140H, NPI705142H, NPI705144H, NPI705155H, NPI705156H, NPI705088H, NPI705089H, NPI705090H, NPI705091H, NPI705092H, NPI705093H, NPI705094H, NPI705095H, NPI705096H, NPI705097H, NPI705098H, NPI705099H, NPI705102H, NPI705103H, NPI705104H, NPI705105H, NPI705106H, NPI705107H, NPI705108H, NPI705109H, NPI705110H, NPI705111H, NPI705112H, NPI705115H, NPI705116H, NPI705117H, NPI705118H, NPI705049H, NPI705050H, NPI705051H, NPI705052H, NPI705053H, NPI705054H, NPI705055H, NPI705056H, NPI705057H, NPI705058H, NPI705059H, NPI705060H, NPI705061H, NPI705062H, NPI705063H, NPI705064H, NPI705065H, NPI705066H, NPI705067H, NPI705068H, NPI705069H, NPI705070H, NPI705081H, NPI705082H, NPI705083H, NPI705084H, NPI705085H, NPI705086H, NPI705100H, NPI705101H, NPI705019H, NPI705021H, NPI705022H, NPI705023H, NPI705024H, NPI705025H, NPI705027H, NPI705030H, NPI705031H, NPI705032H, NPI705033H, NPI705034H, NPI705035H, NPI705036H, NPI705037H, NPI705038H, NPI705039H, NPI705040H, NPI705041H, NPI705042H, NPI705043H, NPI705044H, NPI705045H, NPI705046H, NPI705047H, NPI705048H, NPI704986H, NPI704992H, NPI704993H, NPI704994H, NPI704995H, NPI704996H, NPI704997H, NPI704998H, NPI705000H, NPI705001H, NPI705003H, NPI705004H, NPI705005H, NPI705006H, NPI705007H, NPI705008H, NPI705009H, NPI705010H, NPI705011H, NPI705012H, NPI705013H, NPI705014H, NPI705015H, NPI705016H, NPI705017H, NPI705018H, NPI704951H, NPI704953H, NPI704954H, NPI704955H, NPI704956H, NPI704957H, NPI704958H, NPI704959H, NPI704960H, NPI704961H, NPI704962H, NPI704963H, NPI704964H, NPI704965H, NPI704966H, NPI704968H, NPI704972H, NPI704973H, NPI704974H, NPI704975H, NPI704976H, NPI704977H, NPI704978H, NPI704979H, NPI704980H, NPI704981H, NPI704982H, NPI704983H, NPI704984H, NPI704985H, NPI704987H, NPI704990H, NPI704991H, NPI704913H, NPI704923H, NPI704925H, NPI704926H, NPI704927H, NPI704928H, NPI704929H, NPI704931H, NPI704932H, NPI704934H, NPI704935H, NPI704936H, NPI704937H, NPI704938H, NPI704939H, NPI704940H, NPI704941H, NPI704942H, NPI704943H, NPI704945H, NPI704946H, NPI704947H, NPI704949H, NPI704950H, NPI704952H, NPI704875H, NPI704879H, NPI704885H, NPI704886H, NPI704887H, NPI704889H, NPI704890H, NPI704891H, NPI704892H, NPI704893H, NPI704895H, NPI704896H, NPI704897H, NPI704898H, NPI704899H, NPI704900H, NPI704901H, NPI704902H, NPI704903H, NPI704904H, NPI704905H, NPI704906H, NPI704907H, NPI704908H, NPI704909H, NPI704910H, NPI704914H, NPI704915H, NPI704916H, NPI704917H, NPI704919H, NPI704920H, NPI704921H, NPI704922H, NPI704855H, NPI704857H, NPI704858H, NPI704859H, NPI704860H, NPI704862H, NPI704863H, NPI704864H, NPI704865H, NPI704866H, NPI704867H, NPI704868H, NPI704869H, NPI704871H, NPI704872H, NPI704873H, NPI704874H, NPI704876H, NPI704877H, NPI704878H, NPI704880H, NPI704881H, NPI704883H, NPI704894H, NPI704816H, NPI704817H, NPI704819H, NPI704820H, NPI704821H, NPI704823H, NPI704825H, NPI704826H, NPI704827H, NPI704828H, NPI704830H, NPI704831H, NPI704832H, NPI704833H, NPI704834H, NPI704835H, NPI704837H, NPI704838H, NPI704839H, NPI704841H, NPI704842H, NPI704843H, NPI704844H, NPI704846H, NPI704848H, NPI704850H, NPI704851H, NPI704853H, NPI704854H, NPI704796H, NPI704797H, NPI704798H, NPI704799H, NPI704800H, NPI704801H, NPI704802H, NPI704803H, NPI704804H, NPI704805H, NPI704806H, NPI704808H, NPI704809H, NPI704810H, NPI704811H, NPI704812H, NPI704813H, NPI704814H, NPI704815H, NPI704777H, NPI704779H, NPI704780H, NPI704781H, NPI704782H, NPI704783H, NPI704784H, NPI704785H, NPI704787H, NPI704788H, NPI704789H, NPI704790H, NPI704791H, NPI704792H, NPI704793H, NPI704795H, NPI704740H, NPI704743H, NPI704744H, NPI704745H, NPI704746H, NPI704747H, NPI704748H, NPI704750H, NPI704751H, NPI704752H, NPI704753H, NPI704754H, NPI704755H, NPI704756H, NPI704757H, NPI704758H, NPI704759H, NPI704760H, NPI704761H, NPI704763H, NPI704764H, NPI704765H, NPI704766H, NPI704767H, NPI704768H, NPI704771H, NPI704772H, NPI704773H, NPI704774H, NPI704775H, NPI704776H, NPI704701H, NPI704702H, NPI704709H, NPI704711H, NPI704713H, NPI704714H, NPI704716H, NPI704717H, NPI704720H, NPI704722H, NPI704723H, NPI704724H, NPI704725H, NPI704730H, NPI704731H, NPI704732H, NPI704733H, NPI704735H, NPI704737H, NPI704649H, NPI704650H, NPI704654H, NPI704656H, NPI704659H, NPI704661H, NPI704662H, NPI704663H, NPI704665H, NPI704666H, NPI704667H, NPI704668H, NPI704669H, NPI704670H, NPI704671H, NPI704672H, NPI704673H, NPI704674H, NPI704675H, NPI704676H, NPI704677H, NPI704688H, NPI704689H, NPI704690H, NPI704692H, NPI704695H, NPI704697H, NPI704699H, NPI704700H, NPI704703H, NPI704624H, NPI704625H, NPI704626H, NPI704627H, NPI704628H, NPI704631H, NPI704632H, NPI704633H, NPI704635H, NPI704638H, NPI704639H, NPI704640H, NPI704641H, NPI704643H, NPI704645H, NPI704646H, NPI704647H, NPI704648H, NPI704651H, NPI704652H, NPI704653H, NPI704658H, NPI704679H, NPI704680H, NPI704681H, NPI704683H, NPI704684H, NPI704685H, NPI704686H, NPI704687H, NPI704604H, NPI704605H, NPI704606H, NPI704607H, NPI704608H, NPI704609H, NPI704610H, NPI704611H, NPI704612H, NPI704613H, NPI704614H, NPI704615H, NPI704616H, NPI704619H, NPI704620H, NPI704622H, NPI704642H, NPI704694H, NPI704559H, NPI704560H, NPI704561H, NPI704562H, NPI704563H, NPI704564H, NPI704565H, NPI704566H, NPI704567H, NPI704568H, NPI704569H, NPI704570H, NPI704571H, NPI704572H, NPI704573H, NPI704574H, NPI704575H, NPI704576H, NPI704577H, NPI704579H, NPI704580H, NPI704581H, NPI704582H, NPI704583H, NPI704584H, NPI704585H, NPI704586H, NPI704587H, NPI704588H, NPI704589H, NPI704590H, NPI704591H, NPI704592H, NPI704593H, NPI704594H, NPI704595H, NPI704596H, NPI704597H, NPI704598H, NPI704599H, NPI704600H, NPI704601H, NPI704602H, NPI704531H, NPI704532H, NPI704534H, NPI704535H, NPI704536H, NPI704537H, NPI704540H, NPI704541H, NPI704543H, NPI704544H, NPI704545H, NPI704546H, NPI704547H, NPI704549H, NPI704550H, NPI704553H, NPI704556H, NPI704557H, NPI704558H, NPI704474H, NPI704495H, NPI704496H, NPI704497H, NPI704499H, NPI704500H, NPI704501H, NPI704502H, NPI704503H, NPI704504H, NPI704505H, NPI704506H, NPI704507H, NPI704508H, NPI704510H, NPI704511H, NPI704512H, NPI704513H, NPI704514H, NPI704515H, NPI704516H, NPI704517H, NPI704518H, NPI704520H, NPI704521H, NPI704522H, NPI704523H, NPI704524H, NPI704527H, NPI704528H, NPI704530H, NPI704538H, NPI704539H, NPI704453H, NPI704454H, NPI704456H, NPI704457H, NPI704458H, NPI704459H, NPI704460H, NPI704461H, NPI704462H, NPI704463H, NPI704464H, NPI704465H, NPI704466H, NPI704467H, NPI704468H, NPI704469H, NPI704470H, NPI704471H, NPI704472H, NPI704475H, NPI704476H, NPI704477H, NPI704478H, NPI704479H, NPI704480H, NPI704481H, NPI704482H, NPI704483H, NPI704484H, NPI704485H, NPI704486H, NPI704487H, NPI704489H, NPI704491H, NPI704492H, NPI704498H, NPI704392H, NPI704393H, NPI704394H, NPI704397H, NPI704398H, NPI704399H, NPI704402H, NPI704403H, NPI704404H, NPI704405H, NPI704406H, NPI704407H, NPI704408H, NPI704409H, NPI704410H, NPI704411H, NPI704412H, NPI704413H, NPI704414H, NPI704415H, NPI704416H, NPI704418H, NPI704419H, NPI704421H, NPI704423H, NPI704424H, NPI704425H, NPI704426H, NPI704427H, NPI704428H, NPI704429H, NPI704430H, NPI704431H, NPI704432H, NPI704433H, NPI704434H, NPI704435H, NPI704436H, NPI704437H, NPI704438H, NPI704439H, NPI704440H, NPI704441H, NPI704442H, NPI704443H, NPI704445H, NPI704446H, NPI704447H, NPI704448H, NPI704449H, NPI704450H, NPI704451H, NPI704452H, NPI704455H, NPI704378H, NPI704379H, NPI704380H, NPI704381H, NPI704382H, NPI704384H, NPI704388H, NPI704389H, NPI704390H, NPI704391H, NPI704395H, NPI704396H, NPI704400H, NPI704334H, NPI704336H, NPI704339H, NPI704340H, NPI704341H, NPI704342H, NPI704343H, NPI704344H, NPI704346H, NPI704347H, NPI704348H, NPI704349H, NPI704350H, NPI704351H, NPI704352H, NPI704353H, NPI704354H, NPI704355H, NPI704356H, NPI704357H, NPI704358H, NPI704359H, NPI704360H, NPI704361H, NPI704362H, NPI704363H, NPI704364H, NPI704365H, NPI704366H, NPI704367H, NPI704369H, NPI704370H, NPI704371H, NPI704372H, NPI704373H, NPI704374H, NPI704375H, NPI704376H, NPI704297H, NPI704298H, NPI704299H, NPI704300H, NPI704302H, NPI704303H, NPI704304H, NPI704305H, NPI704306H, NPI704307H, NPI704308H, NPI704309H, NPI704310H, NPI704312H, NPI704313H, NPI704314H, NPI704315H, NPI704316H, NPI704317H, NPI704319H, NPI704320H, NPI704321H, NPI704322H, NPI704323H, NPI704325H, NPI704326H, NPI704328H, NPI704329H, NPI704330H, NPI704331H, NPI704332H, NPI704333H, NPI704335H, NPI704337H, NPI704338H, NPI704246H, NPI704250H, NPI704251H, NPI704252H, NPI704253H, NPI704254H, NPI704255H, NPI704256H, NPI704257H, NPI704258H, NPI704259H, NPI704260H, NPI704261H, NPI704263H, NPI704264H, NPI704266H, NPI704267H, NPI704268H, NPI704269H, NPI704271H, NPI704272H, NPI704273H, NPI704275H, NPI704276H, NPI704277H, NPI704278H, NPI704279H, NPI704282H, NPI704283H, NPI704284H, NPI704287H, NPI704288H, NPI704289H, NPI704290H, NPI704291H, NPI704292H, NPI704293H, NPI704294H, NPI704295H, NPI704296H, NPI704209H, NPI704210H, NPI704211H, NPI704212H, NPI704213H, NPI704230H, NPI704240H, NPI704241H, NPI704242H, NPI704243H, NPI704244H, NPI704245H, NPI704247H, NPI704270H, NPI704181H, NPI704182H, NPI704185H, NPI704189H, NPI704194H, NPI704195H, NPI704196H, NPI704197H, NPI704198H, NPI704203H, NPI704206H, NPI704208H, NPI704217H, NPI704218H, NPI704223H, NPI704225H, NPI704226H, NPI704227H, NPI704228H, NPI704231H, NPI704232H, NPI704233H, NPI704234H, NPI704237H, NPI704183H, NPI704184H, NPI704186H, NPI704190H, NPI704192H, NPI704193H, NPI704135H, NPI704136H, NPI704137H, NPI704147H, NPI704163H, NPI704178H, NPI704117H, NPI704118H, NPI704120H, NPI704121H, NPI704122H, NPI704123H, NPI704124H, NPI704125H, NPI704126H, NPI704127H, NPI704128H, NPI704132H, NPI704133H, NPI704134H, NPI704138H, NPI704140H, NPI704145H, NPI704146H, NPI704152H, NPI704153H, NPI704156H, NPI704157H, NPI704158H, NPI704159H, NPI704160H, NPI704161H, NPI704164H, NPI704165H, NPI704166H, NPI704172H, NPI704173H, NPI704084H, NPI704097H, NPI704103H, NPI704104H, NPI704105H, NPI704106H, NPI704107H, NPI704129H, NPI704130H, NPI704131H, NPI704070H, NPI703969H, NPI703972H, NPI703635H, NPI703553H, NPI703516H, NPI703408H, NPI702551H, NPI702575H, NPI702434H, NPI702503H, NPI702504H, NPI702456H, NPI702363H
GTIN 00763000311483, Serial Numbers: NPI713609H, NPI713632H, NPI713598H, NPI713515H, NPI713587H, NPI713380H, NPI713414H, NPI713142H, NPI713164H, NPI713193H, NPI713137H, NPI713010H, NPI713011H, NPI713012H, NPI713013H, NPI713014H, NPI713015H, NPI713018H, NPI713019H, NPI713021H, NPI713023H, NPI713027H, NPI713030H, NPI713032H, NPI713067H, NPI713074H, NPI712971H, NPI712982H, NPI712987H, NPI712991H, NPI712993H, NPI712994H, NPI713001H, NPI713002H, NPI713005H, NPI713007H, NPI712962H, NPI712966H, NPI712967H, NPI712969H, NPI712670H, NPI712672H, NPI712673H, NPI712684H, NPI712686H, NPI712691H, NPI712693H, NPI712695H, NPI712699H, NPI712707H, NPI712620H, NPI712627H, NPI712628H, NPI712629H, NPI712633H, NPI712636H, NPI712642H, NPI712643H, NPI712651H, NPI712652H, NPI712653H, NPI712657H, NPI712583H, NPI712586H, NPI712587H, NPI712589H, NPI712590H, NPI712595H, NPI712598H, NPI712607H, NPI712609H, NPI712615H, NPI712579H, NPI713729H, NPI712040H, NPI712062H, NPI711992H, NPI711945H, NPI711933H, NPI711969H, NPI713624H, NPI711776H, NPI711796H, NPI711797H, NPI711798H, NPI711804H, NPI711815H, NPI711847H, NPI711744H, NPI711753H, NPI711760H, NPI711762H, NPI711763H, NPI711765H, NPI711766H, NPI711767H, NPI711785H, NPI711787H, NPI713731H, NPI711691H, NPI711692H, NPI711695H, NPI711697H, NPI711729H, NPI711736H, NPI711741H, NPI711742H, NPI713689H, NPI711658H, NPI711706H, NPI711712H, NPI711720H, NPI711726H, NPI711673H, NPI711674H, NPI711683H, NPI711685H, NPI711687H, NPI711688H, NPI713686H, NPI711682H, NPI711619H, NPI711236H, NPI711252H, NPI711255H, NPI711258H, NPI711262H, NPI711216H, NPI711225H, NPI711226H, NPI711174H, NPI711175H, NPI711182H, NPI711184H, NPI711191H, NPI711198H, NPI711202H, NPI711205H, NPI711137H, NPI711143H, NPI711146H, NPI711160H, NPI711169H, NPI711116H, NPI711126H, NPI711130H, NPI711131H, NPI711133H, NPI711100H, NPI711107H, NPI711109H, NPI711111H, NPI711113H, NPI711114H, NPI711064H, NPI711066H, NPI711067H, NPI711070H, NPI711077H, NPI710896H, NPI710842H, NPI710846H, NPI710789H, NPI710808H, NPI710809H, NPI710819H, NPI710827H, NPI710777H, NPI710781H, NPI710783H, NPI710785H, NPI710743H, NPI710686H, NPI710687H, NPI710690H, NPI710691H, NPI710696H, NPI710699H, NPI710700H, NPI710704H, NPI710705H, NPI710662H, NPI710625H, NPI710632H, NPI710634H, NPI710655H, NPI710657H, NPI710596H, NPI710600H, NPI710608H, NPI710548H, NPI710562H, NPI710563H, NPI710569H, NPI710582H, NPI710584H, NPI710593H, NPI710512H, NPI710514H, NPI710530H, NPI710542H, NPI710555H, NPI710559H, NPI710468H, NPI710472H, NPI710483H, NPI710518H, NPI710522H, NPI710457H, NPI710450H, NPI710409H, NPI710370H, NPI710336H, NPI710354H, NPI710299H, NPI710301H, NPI710316H, NPI710318H, NPI710325H, NPI710209H, NPI710214H, NPI710176H, NPI710036H, NPI709378H, NPI709324H, NPI708817H, NPI708823H, NPI708777H, NPI708788H, NPI708792H, NPI708797H, NPI708765H, NPI708766H, NPI708713H, NPI708322H, NPI706343H, NPI705397H, NPI705388H, NPI705406H, NPI704141H, NPI704144H, NPI704155H, NPI704169H, NPI704087H, NPI704096H, NPI704100H, NPI704108H, NPI704114H, NPI704048H, NPI704008H, NPI704012H, NPI704014H, NPI704018H, NPI704019H, NPI704027H, NPI704037H, NPI704039H, NPI704042H, NPI704044H, NPI703968H, NPI703987H, NPI703997H, NPI703953H, NPI703959H, NPI703971H, NPI703892H, NPI703901H, NPI703909H, NPI703922H, NPI703846H, NPI703849H, NPI703855H, NPI703878H, NPI703883H, NPI703884H, NPI703886H, NPI703887H, NPI703805H, NPI703832H, NPI703836H, NPI703837H, NPI703625H, NPI703628H, NPI703636H, NPI703640H, NPI703643H, NPI703649H, NPI703653H, NPI703655H, NPI703670H, NPI703678H, NPI703681H, NPI703684H, NPI703612H, NPI703614H, NPI703616H, NPI703618H, NPI703619H, NPI703620H, NPI703622H, NPI703624H, NPI703472H, NPI703447H, NPI703415H, NPI703416H, NPI703417H, NPI703424H, NPI703430H, NPI703376H, NPI703388H, NPI703406H, NPI703409H, NPI703333H, NPI703335H, NPI703337H, NPI703338H, NPI703351H, NPI703353H, NPI703354H, NPI703355H, NPI703356H, NPI703358H, NPI703362H, NPI703293H, NPI703295H, NPI703298H, NPI703300H, NPI703301H, NPI703315H, NPI703316H, NPI703321H, NPI703323H, NPI703329H, NPI703331H, NPI703289H, NPI703228H, NPI703230H, NPI703231H, NPI703232H, NPI703252H, NPI703166H, NPI703199H, NPI703205H, NPI703206H, NPI703207H, NPI703209H, NPI703216H, NPI703162H, NPI703163H, NPI703092H, NPI703095H, NPI703096H, NPI703098H, NPI703104H, NPI703105H, NPI703108H, NPI703109H, NPI703110H, NPI703117H, NPI703121H, NPI703123H, NPI703125H, NPI703128H, NPI703130H, NPI703132H, NPI703137H, NPI703138H, NPI703142H, NPI703145H, NPI703062H, NPI703064H, NPI703065H, NPI703069H, NPI703076H, NPI703082H, NPI703086H, NPI703010H, NPI703015H, NPI703016H, NPI703031H, NPI703032H, NPI703034H, NPI703036H, NPI703037H, NPI703038H, NPI703040H, NPI703041H, NPI703042H, NPI703043H, NPI703044H, NPI703049H, NPI703050H, NPI703051H, NPI703053H, NPI703054H, NPI703058H, NPI703059H, NPI702974H, NPI702976H, NPI702977H, NPI702978H, NPI702979H, NPI702980H, NPI702981H, NPI702985H, NPI702987H, NPI702988H, NPI702989H, NPI702990H, NPI702992H, NPI702995H, NPI702996H, NPI702998H, NPI702999H, NPI703000H, NPI703001H, NPI703002H, NPI703003H, NPI703004H, NPI703006H, NPI703008H, NPI703011H, NPI703012H, NPI703019H, NPI703020H, NPI703024H, NPI703025H, NPI703027H, NPI703029H, NPI702915H, NPI702917H, NPI702919H, NPI702921H, NPI702923H, NPI702924H, NPI702925H, NPI702928H, NPI702929H, NPI702933H, NPI702936H, NPI702937H, NPI702938H, NPI702940H, NPI702942H, NPI702943H, NPI702944H, NPI702945H, NPI702946H, NPI702951H, NPI702953H, NPI702954H, NPI702956H, NPI702959H, NPI702960H, NPI702961H, NPI702962H, NPI702968H, NPI702969H, NPI702973H, NPI702881H, NPI702882H, NPI702883H, NPI702884H, NPI702885H, NPI702886H, NPI702887H, NPI702888H, NPI702890H, NPI702891H, NPI702892H, NPI702893H, NPI702894H, NPI702895H, NPI702898H, NPI702900H, NPI702901H, NPI702903H, NPI702906H, NPI702907H, NPI702911H, NPI702913H, NPI702922H, NPI702862H, NPI702865H, NPI702866H, NPI702867H, NPI702868H, NPI702822H, NPI702824H, NPI702826H, NPI702827H, NPI702829H, NPI702830H, NPI702831H, NPI702832H, NPI702833H, NPI702835H, NPI702836H, NPI702837H, NPI702839H, NPI702840H, NPI702844H, NPI702845H, NPI702849H, NPI702856H, NPI702869H, NPI702870H, NPI702871H, NPI702872H, NPI702873H, NPI702874H, NPI702875H, NPI702876H, NPI702877H, NPI702773H, NPI702774H, NPI702776H, NPI702777H, NPI702780H, NPI702787H, NPI702789H, NPI702791H, NPI702792H, NPI702796H, NPI702799H, NPI702800H, NPI702802H, NPI702806H, NPI702807H, NPI702808H, NPI702811H, NPI702818H, NPI702704H, NPI702705H, NPI702707H, NPI702708H, NPI702713H, NPI702738H, NPI702739H, NPI702740H, NPI702743H, NPI702745H, NPI702746H, NPI702748H, NPI702756H, NPI702669H, NPI702672H, NPI702673H, NPI702674H, NPI702675H, NPI702679H, NPI702682H, NPI702685H, NPI702686H, NPI702689H, NPI702690H, NPI702691H, NPI702693H, NPI702695H, NPI702698H, NPI702701H, NPI702710H, NPI702718H, NPI702719H, NPI702611H, NPI702614H, NPI702615H, NPI702616H, NPI702618H, NPI702619H, NPI702627H, NPI702628H, NPI702631H, NPI702644H, NPI702645H, NPI702647H, NPI702648H, NPI702650H, NPI702653H, NPI702654H, NPI702656H, NPI702659H, NPI702662H, NPI702663H, NPI702665H, NPI702589H, NPI702590H, NPI702596H, NPI702599H, NPI702601H, NPI702607H, NPI702610H, NPI702620H, NPI702621H, NPI702623H, NPI702624H, NPI702549H, NPI702558H, NPI702569H, NPI702576H, NPI702579H, NPI702580H, NPI702581H, NPI702582H, NPI702586H, NPI702514H, NPI702520H, NPI702524H, NPI702527H, NPI702530H, NPI702532H, NPI702533H, NPI702535H, NPI702537H, NPI702539H, NPI702540H, NPI702542H, NPI702545H, NPI702546H, NPI702559H, NPI702564H, NPI702571H, NPI702573H, NPI702574H, NPI702583H, NPI702430H, NPI702478H, NPI702479H, NPI702480H, NPI702482H, NPI702485H, NPI702486H, NPI702487H, NPI702489H, NPI702490H, NPI702491H, NPI702492H, NPI702493H, NPI702498H, NPI702499H, NPI702500H, NPI702501H, NPI702502H, NPI702507H, NPI702508H, NPI702509H, NPI702510H, NPI702511H, NPI702517H, NPI702518H, NPI702519H, NPI702523H, NPI702526H, NPI702528H, NPI702529H, NPI702423H, NPI702449H, NPI702455H, NPI702462H, NPI702464H, NPI702477H, NPI702345H, NPI702348H, NPI702351H, NPI702354H, NPI702355H, NPI702359H, NPI702360H, NPI702364H, NPI702367H, NPI702378H, NPI702379H, NPI702383H, NPI702397H, NPI702401H, NPI702305H, NPI702310H, NPI702315H, NPI702316H, NPI702318H, NPI702320H, NPI702322H, NPI702323H, NPI702327H, NPI702328H, NPI702330H, NPI702335H, NPI702337H, NPI702339H, NPI702343H, NPI702344H, NPI703151H, NPI702271H, NPI702276H, NPI702277H, NPI702279H, NPI702280H, NPI702285H, NPI702286H, NPI702287H, NPI702292H, NPI702293H, NPI702296H, NPI702297H, NPI702298H, NPI702308H, NPI702311H, NPI702233H, NPI702234H, NPI702235H, NPI702238H, NPI702240H, NPI702241H, NPI702242H, NPI702243H, NPI702247H, NPI702248H, NPI702253H, NPI702259H, NPI702264H, NPI702266H, NPI702267H, NPI702303H, NPI702206H, NPI702207H, NPI702209H, NPI702210H, NPI702217H, NPI702220H, NPI702226H, NPI702229H, NPI702232H, NPI702152H, NPI702157H, NPI702160H, NPI702165H, NPI702167H, NPI702172H, NPI702174H, NPI702175H, NPI702176H, NPI702177H, NPI702179H, NPI702180H, NPI702181H, NPI702189H, NPI702190H, NPI702191H, NPI702113H, NPI702115H, NPI702116H, NPI702121H, NPI702122H, NPI702125H, NPI702129H, NPI702134H, NPI702139H, NPI702140H, NPI702142H, NPI702144H, NPI702146H, NPI702148H, NPI702149H, NPI702153H, NPI702154H, NPI702163H, NPI702085H, NPI702096H, NPI702101H, NPI702104H, NPI702105H, NPI702123H, NPI702050H, NPI702053H, NPI702057H, NPI702065H, NPI702066H, NPI702068H, NPI702070H, NPI702071H, NPI702072H, NPI702075H, NPI702076H, NPI702079H, NPI702082H, NPI702084H, NPI702089H, NPI702090H, NPI702023H, NPI702024H, NPI702028H, NPI702029H, NPI702035H, NPI702037H, NPI701994H, NPI701995H, NPI701996H, NPI701998H, NPI701999H, NPI702000H, NPI702001H, NPI702014H, NPI702016H
GTIN 00763000311490, Serial Numbers: NPI713612H, NPI713628H, NPI713629H, NPI713636H, NPI713521H, NPI713385H, NPI713162H, NPI713163H, NPI713184H, NPI713148H, NPI713154H, NPI713167H, NPI713114H, NPI712984H, NPI712990H, NPI712996H, NPI712997H, NPI712949H, NPI712950H, NPI712951H, NPI712954H, NPI712819H, NPI712830H, NPI712834H, NPI712776H, NPI712778H, NPI712779H, NPI712781H, NPI712783H, NPI712790H, NPI712794H, NPI712810H, NPI712757H, NPI712769H, NPI712772H, NPI712775H, NPI712734H, NPI712736H, NPI712730H, NPI712619H, NPI712624H, NPI712631H, NPI712634H, NPI712654H, NPI712660H, NPI712661H, NPI712663H, NPI712664H, NPI712582H, NPI712597H, NPI712599H, NPI712601H, NPI712612H, NPI712616H, NPI712617H, NPI712647H, NPI712547H, NPI712123H, NPI711999H, NPI711921H, NPI711905H, NPI711906H, NPI711937H, NPI711954H, NPI711964H, NPI711966H, NPI711970H, NPI711884H, NPI711775H, NPI711816H, NPI711756H, NPI711791H, NPI711696H, NPI711699H, NPI711728H, NPI713683H, NPI713684H, NPI711702H, NPI711715H, NPI711721H, NPI713687H, NPI713688H, NPI711664H, NPI711679H, NPI713685H, NPI711611H, NPI713723H, NPI713725H, NPI711240H, NPI711246H, NPI711256H, NPI711189H, NPI711197H, NPI711145H, NPI711172H, NPI711120H, NPI711129H, NPI711132H, NPI711097H, NPI711099H, NPI711106H, NPI711084H, NPI711089H, NPI711094H, NPI711068H, NPI711069H, NPI711071H, NPI711072H, NPI711076H, NPI711093H, NPI711096H, NPI710722H, NPI710740H, NPI710763H, NPI710626H, NPI710633H, NPI710647H, NPI710652H, NPI710598H, NPI710601H, NPI710603H, NPI710617H, NPI710618H, NPI710620H, NPI710550H, NPI710591H, NPI710513H, NPI710529H, NPI710531H, NPI710538H, NPI710539H, NPI710558H, NPI710482H, NPI710492H, NPI710493H, NPI710497H, NPI710504H, NPI710507H, NPI710509H, NPI710431H, NPI710439H, NPI710389H, NPI710378H, NPI710309H, NPI710313H, NPI709326H, NPI709333H, NPI709335H, NPI708825H, NPI708832H, NPI708812H, NPI708761H, NPI708714H, NPI708719H, NPI708720H, NPI708721H, NPI706074H, NPI706076H, NPI705415H, NPI705342H, NPI705256H, NPI704708H, NPI704142H, NPI704150H, NPI704179H, NPI704170H, NPI704171H, NPI704092H, NPI704093H, NPI704102H, NPI704111H, NPI704049H, NPI704058H, NPI704059H, NPI704063H, NPI704064H, NPI704067H, NPI704069H, NPI704073H, NPI704081H, NPI704015H, NPI704016H, NPI704024H, NPI704040H, NPI703970H, NPI703978H, NPI703981H, NPI703983H, NPI703989H, NPI703992H, NPI703994H, NPI703995H, NPI703996H, NPI704000H, NPI704007H, NPI703944H, NPI703949H, NPI703951H, NPI703914H, NPI703932H, NPI703942H, NPI703844H, NPI703853H, NPI703863H, NPI703871H, NPI703872H, NPI703888H, NPI703793H, NPI703794H, NPI703797H, NPI703801H, NPI703808H, NPI703816H, NPI703821H, NPI703834H, NPI703651H, NPI703671H, NPI703673H, NPI703674H, NPI703675H, NPI703613H, NPI703462H, NPI703475H, NPI703480H, NPI703481H, NPI703489H, NPI703491H, NPI703495H, NPI703517H, NPI703451H, NPI703455H, NPI703458H, NPI703414H, NPI703374H, NPI703379H, NPI703382H, NPI703397H, NPI703332H, NPI703340H, NPI703359H, NPI703361H, NPI703363H, NPI703364H, NPI703367H, NPI703370H, NPI703306H, NPI703308H, NPI703314H, NPI703318H, NPI703320H, NPI703326H, NPI703267H, NPI703268H, NPI703270H, NPI703223H, NPI703229H, NPI703234H, NPI703237H, NPI703242H, NPI703246H, NPI703254H, NPI703255H, NPI703256H, NPI703257H, NPI703259H, NPI703260H, NPI703262H, NPI703169H, NPI703172H, NPI703174H, NPI703202H, NPI703211H, NPI703213H, NPI703217H, NPI703218H, NPI703221H, NPI703147H, NPI703148H, NPI703149H, NPI703150H, NPI703153H, NPI703154H, NPI703155H, NPI703156H, NPI703157H, NPI703158H, NPI703159H, NPI703164H, NPI703165H, NPI703179H, NPI703182H, NPI703183H, NPI703186H, NPI703188H, NPI703091H, NPI703097H, NPI703099H, NPI703103H, NPI703106H, NPI703107H, NPI703113H, NPI703114H, NPI703115H, NPI703119H, NPI703122H, NPI703129H, NPI703133H, NPI703134H, NPI703135H, NPI703140H, NPI703146H, NPI703075H, NPI703079H, NPI703081H, NPI703009H, NPI703013H, NPI703017H, NPI703018H, NPI703033H, NPI703056H, NPI702984H, NPI702986H, NPI703021H, NPI703022H, NPI702926H, NPI702931H, NPI702939H, NPI702950H, NPI702952H, NPI702955H, NPI702958H, NPI702965H, NPI702967H, NPI702905H, NPI702861H, NPI702817H, NPI702823H, NPI702828H, NPI702834H, NPI702842H, NPI702843H, NPI702852H, NPI702854H, NPI702858H, NPI702755H, NPI702764H, NPI702765H, NPI702767H, NPI702768H, NPI702769H, NPI702770H, NPI702772H, NPI702775H, NPI702778H, NPI702779H, NPI702783H, NPI702786H, NPI702788H, NPI702790H, NPI702809H, NPI702810H, NPI702812H, NPI702814H, NPI702821H, NPI702702H, NPI702703H, NPI702706H, NPI702714H, NPI702727H, NPI702732H, NPI702741H, NPI702749H, NPI702753H, NPI702754H, NPI702758H, NPI702759H, NPI702760H, NPI702666H, NPI702667H, NPI702668H, NPI702676H, NPI702677H, NPI702680H, NPI702681H, NPI702684H, NPI702688H, NPI702692H, NPI702694H, NPI702699H, NPI702700H, NPI702709H, NPI702711H, NPI702716H, NPI702720H, NPI702723H, NPI702726H, NPI702602H, NPI702604H, NPI702633H, NPI702636H, NPI702639H, NPI702641H, NPI702642H, NPI702646H, NPI702649H, NPI702655H, NPI702658H, NPI702664H, NPI702605H, NPI702608H, NPI702556H, NPI702557H, NPI702544H, NPI702560H, NPI702433H, NPI702439H, NPI702421H, NPI702436H, NPI702441H, NPI702442H, NPI702445H, NPI702447H, NPI702450H, NPI702452H, NPI702453H, NPI702454H, NPI702457H, NPI702458H, NPI702460H, NPI702461H, NPI702463H, NPI702467H, NPI702468H, NPI702471H, NPI702476H, NPI702357H, NPI702361H, NPI702368H, NPI702369H, NPI702372H, NPI702377H, NPI702402H, NPI702409H, NPI702419H, NPI702422H, NPI702444H, NPI702317H, NPI702342H, NPI702346H, NPI702272H, NPI702278H, NPI702281H, NPI702282H, NPI702283H, NPI702295H, NPI702245H, NPI702249H, NPI702252H, NPI702258H, NPI702265H, NPI702268H, NPI702269H, NPI702270H, NPI702273H, NPI702192H, NPI702199H, NPI702205H, NPI702211H, NPI702222H, NPI702182H, NPI702185H, NPI702188H, NPI702194H, NPI702196H, NPI702213H, NPI702112H, NPI702117H, NPI702138H, NPI702150H, NPI702086H, NPI702103H, NPI702047H, NPI702048H, NPI702052H, NPI702055H, NPI702064H, NPI702009H, NPI702025H, NPI702030H, NPI702032H, NPI702034H, NPI702046H, NPI702021H
GTIN 00763000311506, Serial Numbers: NPI714342H, NPI713648H, NPI713647H, NPI713651H, NPI713592H, NPI712727H, NPI712715H, NPI712731H, NPI713730H, NPI711980H, NPI711985H, NPI711986H, NPI711991H, NPI712000H, NPI712004H, NPI712006H, NPI711948H, NPI711934H, NPI711782H, NPI711800H, NPI711788H, NPI711738H, NPI711592H, NPI711605H, NPI711537H, NPI711245H, NPI711247H, NPI711249H, NPI711257H, NPI711263H, NPI711265H, NPI711213H, NPI711224H, NPI711173H.NPI711194H, NPI711201H, NPI711206H, NPI711207H, NPI711152H, NPI711161H, NPI711163H, NPI711165H, NPI711119H, NPI711090H, NPI711095H, NPI711074H, NPI711078H, NPI711083H, NPI710787H, NPI710651H, NPI710526H, NPI709216H, NPI708885H, NPI708848H, NPI708852H, NPI708855H, NPI708806H, NPI708809H, NPI708810H, NPI708816H, NPI708798H, NPI708754H, NPI708745H, NPI708609H, NPI706370H, NPI705366H, NPI705389H, NPI704933H, NPI704870H, NPI704707H, NPI704368H, NPI704151H, NPI704168H, NPI704077H, NPI704090H, NPI704091H, NPI704099H, NPI704045H, NPI704046H, NPI704047H, NPI704056H, NPI704060H, NPI704066H, NPI704080H, NPI704010H, NPI704011H, NPI704017H, NPI704038H, NPI704041H, NPI703976H, NPI703982H, NPI703984H, NPI703985H, NPI704004H, NPI704006H, NPI703943H, NPI703950H, NPI703952H, NPI703954H, NPI703956H, NPI703957H, NPI703961H, NPI703891H, NPI703899H, NPI703900H, NPI703911H, NPI703919H, NPI703920H, NPI703921H, NPI703923H, NPI703924H, NPI703928H, NPI703931H, NPI703935H, NPI703936H, NPI703937H, NPI703938H, NPI703939H, NPI703841H, NPI703843H, NPI703847H, NPI703851H, NPI703852H, NPI703854H, NPI703856H, NPI703857H, NPI703859H, NPI703860H, NPI703864H, NPI703869H, NPI703874H, NPI703875H, NPI703876H, NPI703879H, NPI703881H, NPI703787H, NPI703788H, NPI703789H, NPI703790H, NPI703792H, NPI703796H, NPI703798H, NPI703802H, NPI703803H, NPI703804H, NPI703806H, NPI703807H, NPI703812H, NPI703814H, NPI703817H, NPI703819H, NPI703825H, NPI703826H, NPI703827H, NPI703828H, NPI703830H, NPI703838H, NPI703848H, NPI703689H, NPI703690H, NPI703694H, NPI703695H, NPI703696H, NPI703703H, NPI703704H, NPI703705H, NPI703631H, NPI703641H, NPI703644H, NPI703647H, NPI703656H, NPI703680H, NPI703683H, NPI703615H, NPI703617H, NPI703623H, NPI703467H, NPI703468H, NPI703469H, NPI703471H, NPI703474H, NPI703479H, NPI703494H, NPI703434H, NPI703435H, NPI703436H, NPI703437H, NPI703438H, NPI703439H, NPI703440H, NPI703442H, NPI703444H, NPI703445H, NPI703446H, NPI703448H, NPI703450H, NPI703452H, NPI703453H, NPI703454H, NPI703456H, NPI703418H, NPI703419H, NPI703421H, NPI703423H, NPI703426H, NPI703427H, NPI703429H, NPI703431H, NPI703432H, NPI703371H, NPI703381H, NPI703384H, NPI703385H, NPI703389H, NPI703390H, NPI703391H, NPI703393H, NPI703394H, NPI703399H, NPI703400H, NPI703401H, NPI703402H, NPI703404H, NPI703405H, NPI703413H, NPI703334H, NPI703339H, NPI703343H, NPI703346H, NPI703349H, NPI703357H, NPI703360H, NPI703365H, NPI703366H, NPI703368H, NPI703369H, NPI703294H, NPI703299H, NPI703303H, NPI703304H, NPI703305H, NPI703309H, NPI703311H, NPI703312H, NPI703319H, NPI703322H, NPI703324H, NPI703327H, NPI703328H, NPI703265H, NPI703266H, NPI703275H, NPI703276H, NPI703279H, NPI703280H, NPI703281H, NPI703282H, NPI703287H, NPI703291H, NPI703292H, NPI703227H, NPI703233H, NPI703238H, NPI703240H, NPI703245H, NPI703247H, NPI703249H, NPI703251H, NPI703258H, NPI703261H, NPI703170H, NPI703201H, NPI703203H, NPI703214H, NPI703070H, NPI703073H, NPI703074H, NPI703080H, NPI703085H, NPI702994H, NPI702927H, NPI702966H, NPI702819H, NPI702820H, NPI702848H, NPI702761H, NPI702795H, NPI702797H, NPI702801H, NPI702804H, NPI702805H, NPI702813H, NPI702815H, NPI702816H, NPI702712H, NPI702729H, NPI702733H, NPI702734H, NPI702742H, NPI702751H, NPI702752H, NPI702715H, NPI702722H, NPI702609H, NPI702612H, NPI702613H, NPI702617H, NPI702625H, NPI702626H, NPI702629H, NPI702632H, NPI702638H, NPI702640H, NPI702593H, NPI702594H, NPI702595H, NPI702598H, NPI702600H, NPI702512H, NPI702531H, NPI702536H, NPI702538H, NPI702555H, NPI702587H, NPI702428H, NPI702481H, NPI702483H, NPI702484H, NPI702494H, NPI702496H, NPI702515H, NPI702516H, NPI702521H, NPI702522H, NPI702438H, NPI702446H, NPI702466H, NPI702399H, NPI702404H, NPI702406H, NPI702412H, NPI702437H, NPI702312H, NPI702313H, NPI702324H, NPI702329H, NPI702331H, NPI702333H, NPI702334H, NPI702336H, NPI702338H, NPI702340H, NPI702341H, NPI702347H, NPI702289H, NPI702254H, NPI702198H, NPI702201H, NPI702208H, NPI702219H, NPI702221H, NPI702223H, NPI702224H, NPI702228H, NPI702186H, NPI702111H, NPI702135H, NPI702108H, NPI702120H, NPI702058H, NPI702059H, NPI702063H, NPI702067H, NPI702010H, NPI702011H, NPI702041H, NPI702020H, NPI701988H
GTIN 00763000311513, Serial Numbers: NPI711701H, NPI711703H, NPI711704H, NPI711705H, NPI711707H, NPI711709H, NPI711662H, NPI711678H, NPI711686H, NPI711690H, NPI711219H, NPI711185H, NPI711186H, NPI711200H, NPI711142H, NPI711144H, NPI711157H, NPI711170H, NPI711104H, NPI710774H, NPI710775H, NPI710776H, NPI710750H, NPI710698H, NPI710703H, NPI710709H, NPI710640H, NPI710607H, NPI710612H, NPI710613H, NPI710614H, NPI704139H, NPI704095H, NPI704101H, NPI704115H, NPI704052H, NPI704054H, NPI704009H, NPI704013H, NPI704026H, NPI703999H, NPI704001H, NPI703965H, NPI703917H, NPI703925H, NPI703940H, NPI703946H, NPI703868H, NPI703691H, NPI703697H, NPI703698H, NPI703699H, NPI703702H, NPI703706H, NPI703632H, NPI703634H, NPI703380H, NPI703348H, NPI703352H, NPI703317H, NPI703124H, NPI703136H, NPI703139H, NPI703063H, NPI702975H, NPI702982H, NPI702993H, NPI702997H, NPI703028H, NPI702947H, NPI702948H, NPI702949H, NPI702963H, NPI702964H, NPI702910H, NPI702860H, NPI702863H, NPI702864H, NPI702781H, NPI702784H, NPI702785H, NPI702794H, NPI702798H, NPI702697H, NPI702747H, NPI702757H, NPI702683H, NPI702687H, NPI702603H, NPI702591H, NPI702597H, NPI702606H, NPI702622H, NPI702550H, NPI702562H, NPI702565H, NPI702525H, NPI702534H, NPI702547H, NPI702548H, NPI702552H, NPI702553H, NPI702554H, NPI702561H, NPI702563H, NPI702566H, NPI702567H, NPI702568H, NPI702572H, NPI702578H, NPI702425H, NPI702427H, NPI702429H, NPI702431H, NPI702432H, NPI702435H, NPI702488H, NPI702495H, NPI702497H, NPI702506H, NPI702418H, NPI702440H, NPI702469H, NPI702470H, NPI702472H, NPI702473H, NPI702475H, NPI702350H, NPI702366H, NPI702370H, NPI702371H, NPI702374H, NPI702376H, NPI702380H, NPI702387H, NPI702388H, NPI702389H, NPI702392H, NPI702393H, NPI702398H, NPI702403H, NPI702408H, NPI702413H, NPI702415H, NPI702416H, NPI702307H, NPI702309H, NPI702314H, NPI702319H, NPI702325H, NPI702326H, NPI702294H, NPI702244H, NPI702251H, NPI702255H, NPI702257H, NPI702262H, NPI702300H, NPI702200H, NPI702204H, NPI702302H, NPI702126H, NPI702127H, NPI702130H, NPI702133H, NPI702136H, NPI702137H, NPI702141H, NPI702145H, NPI702093H, NPI702102H, NPI702106H, NPI702049H, NPI702083H, NPI702027H, NPI702033H, NPI702036H, NPI702038H, NPI702042H
GTIN 00763000311520, Serial Numbers: NPI715057H, NPI715071H, NPI714841H, NPI713183H, NPI713203H, NPI713157H, NPI713160H, NPI712809H, NPI712666H, NPI712646H, NPI711990H, NPI711929H, NPI711878H, NPI711848H, NPI711746H, NPI711733H, NPI711734H, NPI711680H, NPI711218H, NPI711188H, NPI711195H, NPI711196H, NPI711134H, NPI711147H, NPI711151H, NPI711154H, NPI711164H, NPI711166H, NPI711121H, NPI711124H, NPI711128H, NPI711101H, NPI710453H, NPI710454H, NPI710376H, NPI710298H, NPI710306H, NPI709797H, NPI709814H, NPI709426H, NPI709262H, NPI708841H, NPI708843H, NPI708845H, NPI708839H, NPI708010H, NPI706713H, NPI706077H, NPI705756H, NPI705417H, NPI705338H, NPI704822H, NPI704738H, NPI704637H, NPI704644H, NPI704285H, NPI704220H, NPI704221H, NPI704050H, NPI704062H, NPI704065H, NPI704071H, NPI704078H, NPI704031H, NPI704033H, NPI704034H, NPI703945H, NPI703910H, NPI703912H, NPI703927H, NPI703626H, NPI703629H, NPI703630H, NPI703639H, NPI703642H, NPI703645H, NPI703433H, NPI703425H, NPI703428H, NPI703372H, NPI703375H, NPI703378H, NPI703387H, NPI703398H, NPI703403H, NPI703410H, NPI703345H, NPI703310H, NPI703313H, NPI703239H, NPI703241H, NPI703244H, NPI703171H, NPI703173H, NPI703175H, NPI703196H, NPI703197H, NPI703212H, NPI703168H, NPI703176H, NPI703177H, NPI703178H, NPI703181H, NPI703184H, NPI703185H, NPI703187H, NPI703189H, NPI703190H, NPI703191H, NPI703192H, NPI703194H, NPI703195H, NPI703090H, NPI703093H, NPI703094H, NPI703100H, NPI703101H, NPI703102H, NPI703112H, NPI703116H, NPI703118H, NPI703120H, NPI703141H, NPI703144H, NPI703066H, NPI703067H, NPI703071H, NPI703072H, NPI703077H, NPI703078H, NPI703083H, NPI703087H, NPI703089H, NPI703035H, NPI703045H, NPI703046H, NPI703047H, NPI703052H, NPI702762H, NPI702782H, NPI702353H, NPI702356H, NPI702358H, NPI702381H, NPI702382H, NPI702395H, NPI702396H, NPI702400H, NPI702410H, NPI702411H, NPI702420H, NPI702237H, NPI702147H, NPI702056H, NPI702061H, NPI702069H, NPI702077H, NPI702080H, NPI702022H, NPI702039H, NPI702040H, NPI702002H, NPI702005H, NPI702007H, NPI702008H, NPI702012H, NPI702017H, NPI701986H, NPI701987H, NPI701989H, NPI701993H
GTIN 00763000311537, Serial Numbers: NPI712041H, NPI710737H, NPI710748H, NPI710682H, NPI710684H, NPI710718H, NPI710670H, NPI710673H, NPI709322H, NPI709329H, NPI709332H, NPI707009H, NPI703948H, NPI702384H, NPI702385H, NPI702391H, NPI702178H, NPI702109H, NPI702110H, NPI702124H, NPI702128H, NPI702132H, NPI702143H, NPI702088H, NPI702092H, NPI702107H, NPI702078H
GTIN 00763000311544, Serial Numbers: NPI714840H, NPI712977H, NPI711955H, NPI711190H, NPI703658H, NPI703672H, NPI703676H, NPI703677H, NPI703344H, NPI703350H, NPI702880H, NPI702763H, NPI702766H, NPI702736H, NPI702737H, NPI702750H, NPI702671H, NPI702678H, NPI702696H, NPI702717H, NPI702721H, NPI702724H, NPI702424H, NPI702394H, NPI702407H, NPI702151H, NPI702156H, NPI702114H, NPI702161H, NPI701990H, NPI701991H, NPI701992H
GTIN 00763000311551, Serial Numbers: NPI713226H, NPI712973H, NPI712948H, NPI710821H, NPI710602H, NPI710589H, NPI710594H, NPI710328H, NPI703302H, NPI703264H, NPI703208H, NPI703161H, NPI703143H, NPI703060H, NPI703068H, NPI703048H, NPI703057H, NPI703023H, NPI702971H, NPI702972H, NPI702851H, NPI702853H, NPI702670H, NPI702635H, NPI702652H, NPI702231H, NPI702166H, NPI702171H, NPI702183H, NPI702187H, NPI701997H, NPI702015H
GTIN 00763000311582, Serial Numbers: NPI712956H. NPI712957H. NPI712958H. NPI712932H. NPI712939H. NPI712782H. NPI712770H. NPI712771H. NPI712774H. NPI712688H. NPI712015H. NPI711979H. NPI711983H. NPI711877H. NPI711864H. NPI711871H. NPI711873H. NPI711858H. NPI711861H. NPI711865H. NPI711844H. NPI710568H. NPI710571H. NPI710578H. NPI710587H. NPI710588H. NPI710511H. NPI710534H. NPI710491H. NPI710506H. NPI710520H. NPI710459H. NPI710470H. NPI710474H. NPI710475H. NPI710480H. NPI710448H. NPI710300H. NPI710304H. NPI710305H. NPI710311H. NPI710314H. NPI710315H. NPI710317H. NPI710319H. NPI710329H. NPI708715H. NPI708717H. NPI708718H. NPI708722H. NPI708726H. NPI708730H. NPI704176H. NPI704116H. NPI704025H. NPI703973H. NPI703974H. NPI703975H. NPI703977H. NPI703980H. NPI703988H. NPI703991H. NPI703998H. NPI704002H. NPI704003H. NPI703964H. NPI703967H. NPI703893H. NPI703896H. NPI703897H. NPI703898H. NPI703845H. NPI703862H. NPI703867H. NPI703882H. NPI703627H. NPI703633H. NPI703648H. NPI703621H. NPI703470H. NPI703482H. NPI703492H. NPI703493H. NPI703500H. NPI703519H. NPI703461H. NPI703263H. NPI703269H. NPI703271H. NPI703272H. NPI703273H. NPI703274H. NPI703277H. NPI703278H. NPI702904H. NPI702909H. NPI702912H. NPI702657H. NPI702362H. NPI702299H. NPI702288H. NPI702290H. NPI702291H. NPI702236H. NPI702246H. NPI702250H. NPI702256H. NPI702261H. NPI702263H. NPI702274H. NPI702195H. NPI702225H. NPI702162H. NPI702170H. NPI702173H. NPI702197H. NPI702155H. NPI702095H. NPI702097H. NPI702098H. NPI702099H. NPI702100H
GTIN 00763000311599, Serial Numbers: NPI713721H, NPI711232H, NPI711223H, NPI711159H, NPI710797H, NPI710711H, NPI704068H, NPI704032H, NPI703850H, NPI703833H
GTIN 00763000311605, Serial Numbers: NPI711719H, NPI710610H, NPI710611H, NPI710616H, NPI710619H, NPI710547H, NPI710560H, NPI710565H, NPI710564H, NPI710496H, NPI707340H, NPI707022H, NPI705757H, NPI704162H, NPI704167H, NPI704053H, NPI704075H, NPI704082H, NPI702193H, NPI702202H
GTIN 00763000311643, Serial Numbers: NPI711208H, NPI711141H, NPI708243H, NPI708129H, NPI705424H, NPI704824H, NPI704542H, NPI703894H, NPI703903H, NPI703904H, NPI703842H, NPI703873H, NPI703880H, NPI703800H, NPI703815H, NPI703824H, NPI703840H, NPI703420H, NPI703395H, NPI703297H, NPI703284H, NPI703285H, NPI703286H
GTIN 00763000353971, Serial Numbers: NPI714129H, NPI713776H, NPI713504H, NPI713554H, NPI712339H, NPI712343H, NPI712281H, NPI711862H, NPI711783H, NPI711789H, NPI711790H, NPI711713H, NPI711646H, NPI711621H, NPI711644H, NPI711645H, NPI713623H, NPI711150H, NPI711087H, NPI711015H, NPI710592H, NPI710382H, NPI710312H, NPI710323H, NPI709851H, NPI709850H, NPI709795H, NPI709796H, NPI709799H, NPI709762H, NPI709766H, NPI709767H, NPI709770H, NPI709771H, NPI709772H, NPI709775H, NPI709778H, NPI709780H, NPI709782H, NPI709785H, NPI709786H, NPI709787H, NPI709788H, NPI709792H, NPI709794H, NPI709677H, NPI709682H, NPI709684H, NPI709686H, NPI709687H, NPI709688H, NPI709639H, NPI709656H, NPI709657H, NPI709658H, NPI709659H, NPI709660H, NPI709662H, NPI709663H, NPI709665H, NPI709668H, NPI709669H, NPI709526H, NPI709527H, NPI709529H, NPI709530H, NPI709531H, NPI709533H, NPI709424H, NPI709430H, NPI709431H, NPI709433H, NPI709427H, NPI709376H, NPI709377H, NPI709379H, NPI709325H, NPI709327H, NPI709328H, NPI709330H, NPI709334H, NPI708601H, NPI708607H, NPI708608H, NPI708610H, NPI708597H, NPI708605H, NPI708606H, NPI708538H, NPI708548H, NPI708549H, NPI708455H, NPI708458H, NPI708430H, NPI708409H, NPI708410H, NPI708371H, NPI708378H, NPI708379H, NPI708382H, NPI708249H, NPI708257H, NPI708131H, NPI708119H, NPI707910H, NPI707344H, NPI707361H, NPI707194H, NPI706642H, NPI706582H, NPI706583H, NPI706564H, NPI706565H, NPI706567H, NPI706568H, NPI706572H, NPI706573H, NPI706575H, NPI706576H, NPI706442H, NPI706443H, NPI706444H, NPI706445H, NPI706447H, NPI706448H, NPI706449H, NPI706450H, NPI706451H, NPI706365H, NPI706371H, NPI706372H, NPI706373H, NPI706374H, NPI706375H, NPI706340H, NPI706342H, NPI706345H, NPI706346H, NPI706347H, NPI706348H, NPI706349H, NPI706352H, NPI706353H, NPI706354H, NPI706356H, NPI706357H, NPI706358H, NPI706359H, NPI706360H, NPI706361H, NPI706362H, NPI706363H, NPI706364H, NPI706366H, NPI706367H, NPI706368H, NPI706303H, NPI706312H, NPI706313H, NPI706314H, NPI706315H, NPI706317H, NPI706318H, NPI706319H, NPI706320H, NPI706321H, NPI706322H, NPI706323H, NPI706324H, NPI706325H, NPI706326H, NPI706327H, NPI706328H, NPI706330H, NPI706331H, NPI706332H, NPI706333H, NPI706334H, NPI706335H, NPI706337H, NPI706338H, NPI706339H, NPI706289H, NPI706290H, NPI706291H, NPI706292H, NPI706293H, NPI706296H, NPI706297H, NPI706298H, NPI706299H, NPI706300H, NPI706301H, NPI706305H, NPI706307H, NPI706308H, NPI706309H, NPI706310H, NPI706196H, NPI706197H, NPI706199H, NPI706200H, NPI706201H, NPI706202H, NPI706203H, NPI706204H, NPI706205H, NPI706206H, NPI706208H, NPI706210H, NPI706211H, NPI706212H, NPI706213H, NPI706214H, NPI706215H, NPI706216H, NPI706218H, NPI706219H, NPI706220H, NPI706221H, NPI706222H, NPI706223H, NPI706187H, NPI706188H, NPI706189H, NPI706190H, NPI706194H, NPI706195H, NPI706136H, NPI705961H, NPI705963H, NPI705964H, NPI705965H, NPI705969H, NPI705952H, NPI705953H, NPI705955H, NPI705957H, NPI705881H, NPI705886H, NPI705888H, NPI705767H, NPI705768H, NPI705774H, NPI705775H, NPI705786H, NPI705787H, NPI705788H, NPI705790H, NPI705791H, NPI705758H, NPI705759H, NPI705760H, NPI705763H, NPI705764H, NPI705765H, NPI705766H, NPI705769H, NPI705770H, NPI705771H, NPI705772H, NPI705776H, NPI705777H, NPI705778H, NPI705779H, NPI705780H, NPI705783H, NPI705622H, NPI705624H, NPI705625H, NPI705626H, NPI705627H, NPI705628H, NPI705629H, NPI705596H, NPI705600H, NPI705601H, NPI705602H, NPI705604H, NPI705605H, NPI705606H, NPI705607H, NPI705608H, NPI705609H, NPI705610H, NPI705611H, NPI705612H, NPI705613H, NPI705614H, NPI705615H, NPI705616H, NPI705617H, NPI705618H, NPI705620H, NPI705582H, NPI705584H, NPI705585H, NPI705586H, NPI705589H, NPI705590H, NPI705591H, NPI705592H, NPI705594H, NPI705577H, NPI705455H, NPI705456H, NPI705425H, NPI705367H, NPI705390H, NPI705393H, NPI705394H, NPI705218H, NPI705220H, NPI705224H, NPI705225H, NPI705226H, NPI705228H, NPI705229H, NPI705230H, NPI705202H, NPI705205H, NPI705207H, NPI705208H, NPI705209H, NPI705211H, NPI705212H, NPI705213H, NPI705214H, NPI705215H, NPI705221H, NPI705223H, NPI705141H, NPI705071H, NPI705072H, NPI705073H, NPI705074H, NPI705075H, NPI705076H, NPI705077H, NPI705078H, NPI705079H, NPI705080H, NPI705028H, NPI705029H, NPI704969H, NPI704970H, NPI704818H, NPI704736H, NPI704660H, NPI704705H, NPI704617H, NPI704552H, NPI704088H, NPI704076H, NPI703906H, NPI703490H, NPI703377H
GTIN 00763000420987, Serial Numbers: NPI715240H, NPI715247H, NPI715263H, NPI715269H, NPI715203H, NPI715213H, NPI715118H, NPI714908H, NPI714845H, NPI714847H, NPI714848H, NPI714849H, NPI714850H, NPI714851H, NPI714852H, NPI714854H, NPI714856H, NPI714857H, NPI714858H, NPI714859H, NPI714860H, NPI714862H, NPI714863H, NPI714827H, NPI714828H, NPI714829H, NPI714830H, NPI714831H, NPI714832H, NPI714833H, NPI714834H, NPI714837H, NPI714839H, NPI714842H, NPI714843H, NPI714807H, NPI714718H, NPI714720H, NPI714721H, NPI714723H, NPI714724H, NPI714749H, NPI714752H, NPI714753H, NPI714754H, NPI714755H, NPI714756H, NPI714758H, NPI714759H, NPI714760H, NPI714761H, NPI714768H, NPI714769H, NPI714773H, NPI714786H, NPI714791H, NPI714792H, NPI714802H, NPI714808H, NPI714660H, NPI714667H, NPI714670H, NPI714675H, NPI714685H, NPI714686H, NPI714688H, NPI714697H, NPI714701H, NPI714715H, NPI714728H, NPI714729H, NPI714607H, NPI714608H, NPI714609H, NPI714610H, NPI714619H, NPI714620H, NPI714621H, NPI714624H, NPI714626H, NPI714628H, NPI714629H, NPI714630H, NPI714631H, NPI714632H, NPI714633H, NPI714635H, NPI714636H, NPI714639H, NPI714645H, NPI714649H, NPI714651H, NPI714654H, NPI714655H, NPI714552H, NPI714553H, NPI714554H, NPI714557H, NPI714558H, NPI714562H, NPI714571H, NPI714575H, NPI714576H, NPI714578H, NPI714584H, NPI714585H, NPI714586H, NPI714534H, NPI714535H, NPI714537H, NPI714538H, NPI714539H, NPI714540H, NPI714541H, NPI714544H, NPI714548H, NPI714549H, NPI714550H, NPI714495H, NPI714496H, NPI714498H, NPI714499H, NPI714502H, NPI714505H, NPI714508H, NPI714509H, NPI714510H, NPI714513H, NPI714515H, NPI714516H, NPI714518H, NPI714519H, NPI714521H, NPI714526H, NPI714527H, NPI714532H, NPI714533H, NPI714419H, NPI714422H, NPI714431H, NPI714436H, NPI714437H, NPI714438H, NPI714439H, NPI714443H, NPI714444H, NPI714450H, NPI714451H, NPI714452H, NPI714453H, NPI714454H, NPI714455H, NPI714463H, NPI714466H, NPI714394H, NPI714395H, NPI714402H, NPI714403H, NPI714404H, NPI714405H, NPI714406H, NPI714407H, NPI714408H, NPI714410H, NPI714412H, NPI714415H, NPI714417H, NPI714432H, NPI714433H, NPI714434H, NPI714435H, NPI714475H, NPI714476H, NPI714477H, NPI714478H, NPI714485H, NPI714490H, NPI714491H, NPI714493H, NPI714327H, NPI714329H, NPI714330H, NPI714332H, NPI714335H, NPI714338H, NPI714340H, NPI714341H, NPI714343H, NPI714347H, NPI714355H, NPI714356H, NPI714360H, NPI714361H, NPI714362H, NPI714363H, NPI714364H, NPI714365H, NPI714366H, NPI714370H, NPI714385H, NPI714387H, NPI714391H, NPI714392H, NPI714393H, NPI714233H, NPI714234H, NPI714237H, NPI714238H, NPI714241H, NPI714242H, NPI714243H, NPI714246H, NPI714247H, NPI714250H, NPI714254H, NPI714255H, NPI714265H, NPI714269H, NPI714270H, NPI714273H, NPI714274H, NPI714282H, NPI714289H, NPI714294H, NPI714301H, NPI714302H, NPI714304H, NPI714305H, NPI714306H, NPI714307H, NPI714310H, NPI714311H, NPI714320H, NPI714168H, NPI714176H, NPI714177H, NPI714178H, NPI714179H, NPI714180H, NPI714184H, NPI714185H, NPI714186H, NPI714187H, NPI714188H, NPI714189H, NPI714191H, NPI714195H, NPI714196H, NPI714197H, NPI714198H, NPI714199H, NPI714200H, NPI714203H, NPI714205H, NPI714206H, NPI714208H, NPI714209H, NPI714215H, NPI714216H, NPI714221H, NPI714226H, NPI714227H, NPI714230H, NPI714105H, NPI714106H, NPI714107H, NPI714108H, NPI714113H, NPI714126H, NPI714127H, NPI714146H, NPI714148H, NPI714155H, NPI714162H, NPI714163H, NPI714166H, NPI714170H, NPI714171H, NPI714172H, NPI714173H, NPI714032H, NPI714036H, NPI714039H, NPI714040H, NPI714041H, NPI714044H, NPI714046H, NPI714047H, NPI714048H, NPI714049H, NPI714050H, NPI714053H, NPI714055H, NPI714056H, NPI714058H, NPI714061H, NPI714062H, NPI714065H, NPI714068H, NPI714071H, NPI714074H, NPI714076H, NPI714077H, NPI714083H, NPI714087H, NPI714089H, NPI714092H, NPI714095H, NPI714099H, NPI714101H, NPI714103H, NPI714109H, NPI714110H, NPI714112H, NPI714114H, NPI714116H, NPI714119H, NPI714121H, NPI714122H, NPI714123H, NPI714125H, NPI713936H, NPI713938H, NPI713940H, NPI713946H, NPI713958H, NPI713959H, NPI713960H, NPI713961H, NPI713964H, NPI713966H, NPI713969H, NPI713973H, NPI713974H, NPI713975H, NPI713977H, NPI713978H, NPI713989H, NPI713990H, NPI713991H, NPI713994H, NPI713995H, NPI713997H, NPI713998H, NPI713999H, NPI714002H, NPI714003H, NPI714004H, NPI714006H, NPI714008H, NPI714010H, NPI714011H, NPI714012H, NPI714013H, NPI714014H, NPI714015H, NPI714016H, NPI714018H, NPI714029H, NPI714030H, NPI713870H, NPI713873H, NPI713875H, NPI713880H, NPI713881H, NPI713883H, NPI713887H, NPI713888H, NPI713890H, NPI713893H, NPI713895H, NPI713898H, NPI713908H, NPI713910H, NPI713911H, NPI713916H, NPI713922H, NPI713925H, NPI713931H, NPI713932H, NPI713934H, NPI713937H, NPI713832H, NPI713834H, NPI713836H, NPI713837H, NPI713838H, NPI713847H, NPI713848H, NPI713849H, NPI713853H, NPI713857H, NPI713862H, NPI713772H, NPI713775H, NPI713783H, NPI713786H, NPI713795H, NPI713797H, NPI713798H, NPI713801H, NPI713804H, NPI713805H, NPI713808H, NPI713810H, NPI713811H, NPI713813H, NPI713816H, NPI713818H, NPI713819H, NPI713820H, NPI713823H, NPI713824H, NPI713841H, NPI713842H, NPI713735H, NPI713736H, NPI713737H, NPI713738H, NPI713739H, NPI713740H, NPI713741H, NPI713744H, NPI713745H, NPI713746H, NPI713751H, NPI713754H, NPI713761H, NPI713762H, NPI713763H, NPI713764H, NPI713765H, NPI713766H, NPI713767H, NPI713769H, NPI713770H, NPI713771H, NPI713691H, NPI713695H, NPI713696H, NPI713699H, NPI713702H, NPI713703H, NPI713705H, NPI713706H, NPI713710H, NPI713665H, NPI713669H, NPI713670H, NPI713672H, NPI713673H, NPI713675H, NPI713676H, NPI713677H, NPI713678H, NPI713682H, NPI713712H, NPI713715H, NPI713716H, NPI713718H, NPI713579H, NPI713589H, NPI713591H, NPI713593H, NPI713594H, NPI713595H, NPI713596H, NPI713597H, NPI713599H, NPI713601H, NPI713602H, NPI713604H, NPI713605H, NPI713606H, NPI713607H, NPI713511H, NPI713513H, NPI713522H, NPI713523H, NPI713524H, NPI713525H, NPI713528H, NPI713530H, NPI713531H, NPI713532H, NPI713533H, NPI713534H, NPI713535H, NPI713536H, NPI713537H, NPI713538H, NPI713539H, NPI713540H, NPI713542H, NPI713543H, NPI713544H, NPI713545H, NPI713546H, NPI713547H, NPI713551H, NPI713552H, NPI713556H, NPI713558H, NPI713561H, NPI713565H, NPI713566H, NPI713567H, NPI713569H, NPI713570H, NPI713571H, NPI713573H, NPI713574H, NPI713575H, NPI713577H, NPI713581H, NPI713584H, NPI713585H, NPI713586H, NPI713428H, NPI713429H, NPI713430H, NPI713431H, NPI713432H, NPI713433H, NPI713434H, NPI713435H, NPI713436H, NPI713441H, NPI713442H, NPI713443H, NPI713444H, NPI713447H, NPI713448H, NPI713449H, NPI713451H, NPI713452H, NPI713508H, NPI713509H, NPI713517H, NPI713518H, NPI713519H, NPI713520H, NPI713358H, NPI713374H, NPI713375H, NPI713377H, NPI713378H, NPI713379H, NPI713381H, NPI713382H, NPI713383H, NPI713384H, NPI713386H, NPI713387H, NPI713388H, NPI713389H, NPI713390H, NPI713391H, NPI713392H, NPI713393H, NPI713394H, NPI713395H, NPI713396H, NPI713398H, NPI713403H, NPI713404H, NPI713405H, NPI713408H, NPI713409H, NPI713410H, NPI713411H, NPI713412H, NPI713413H, NPI713416H, NPI713417H, NPI713418H, NPI713419H, NPI713421H, NPI713423H, NPI713424H, NPI713425H, NPI713426H, NPI713427H, NPI713439H, NPI713440H, NPI713322H, NPI713326H, NPI713328H, NPI713329H, NPI713330H, NPI713331H, NPI713332H, NPI713336H, NPI713337H, NPI713338H, NPI713339H, NPI713340H, NPI713341H, NPI713342H, NPI713344H, NPI713345H, NPI713346H, NPI713347H, NPI713350H, NPI713351H, NPI713354H, NPI713355H, NPI713359H, NPI713360H, NPI713363H, NPI713364H, NPI713370H, NPI713286H, NPI713287H, NPI713288H, NPI713289H, NPI713291H, NPI713292H, NPI713293H, NPI713293H, NPI713294H, NPI713294H, NPI713295H, NPI713296H, NPI713297H, NPI713298H, NPI713300H, NPI713301H, NPI713302H, NPI713303H, NPI713304H, NPI713305H, NPI713306H, NPI713307H, NPI713308H, NPI713311H, NPI713312H, NPI713314H, NPI713316H, NPI713317H, NPI713318H, NPI713319H, NPI713320H, NPI713321H, NPI713254H, NPI713256H, NPI713258H, NPI713262H, NPI713265H.NPI713268H, NPI713269H, NPI713270H, NPI713271H, NPI713272H, NPI713276H, NPI713277H, NPI713279H, NPI713281H, NPI713285H, NPI713212H, NPI713213H, NPI713214H, NPI713216H, NPI713217H, NPI713219H, NPI713221H, NPI713222H, NPI713224H, NPI713225H, NPI713227H, NPI713228H, NPI713231H, NPI713232H, NPI713235H, NPI713236H, NPI713237H, NPI713241H, NPI713243H, NPI713244H, NPI713246H, NPI713247H, NPI713248H, NPI713249H, NPI713250H, NPI713251H, NPI713252H, NPI713253H, NPI713260H, NPI713261H, NPI713263H, NPI713273H, NPI713144H, NPI713146H, NPI713175H, NPI713177H, NPI713178H, NPI713179H, NPI713180H, NPI713181H, NPI713182H, NPI713185H, NPI713186H, NPI713187H, NPI713190H, NPI713191H, NPI713195H, NPI713196H, NPI713197H, NPI713198H, NPI713200H, NPI713204H, NPI713205H, NPI713206H, NPI713207H, NPI713208H, NPI713209H, NPI713210H, NPI713111H, NPI713115H, NPI713119H, NPI713120H, NPI713121H, NPI713122H, NPI713123H, NPI713124H, NPI713125H, NPI713127H, NPI713132H, NPI713133H, NPI713135H, NPI713151H, NPI713153H, NPI713155H, NPI713158H, NPI713161H, NPI713170H, NPI713173H, NPI713174H, NPI713046H, NPI713047H, NPI713049H, NPI713050H, NPI713051H, NPI713052H, NPI713053H, NPI713054H, NPI713055H, NPI713056H, NPI713057H, NPI713061H, NPI713075H, NPI713076H, NPI713077H, NPI713083H, NPI713085H, NPI713086H, NPI713088H, NPI713089H, NPI713090H, NPI713091H, NPI713092H, NPI713093H, NPI713095H, NPI713096H, NPI713097H, NPI713098H, NPI713101H, NPI713102H, NPI713104H, NPI713105H, NPI713106H, NPI713107H, NPI713108H, NPI713109H, NPI713110H, NPI713112H, NPI713113H, NPI713116H, NPI713117H, NPI713118H, NPI713024H, NPI713025H, NPI713026H, NPI713028H, NPI713029H, NPI713033H, NPI713034H, NPI713036H, NPI713039H, NPI713042H, NPI713044H, NPI713045H, NPI713048H, NPI713058H, NPI713060H, NPI713062H, NPI713063H, NPI713065H, NPI713066H, NPI713068H, NPI713069H, NPI713070H, NPI713072H, NPI713073H, NPI713078H, NPI713079H, NPI713080H, NPI713081H, NPI713082H, NPI712974H, NPI712978H, NPI712980H, NPI712985H, NPI712988H, NPI712992H, NPI712995H, NPI712955H, NPI712959H, NPI712961H, NPI712963H, NPI712902H, NPI712903H, NPI712905H, NPI712906H, NPI712908H, NPI712909H, NPI712910H, NPI712920H, NPI712922H, NPI712923H, NPI712925H, NPI712927H, NPI712929H, NPI712931H, NPI712933H, NPI712934H, NPI712935H, NPI712936H, NPI712937H, NPI712941H, NPI712942H, NPI712943H, NPI712944H, NPI712945H, NPI712946H, NPI712947H, NPI712858H, NPI712859H, NPI712862H, NPI712885H, NPI712886H, NPI712887H, NPI712888H, NPI712889H, NPI712890H, NPI712891H, NPI712892H, NPI712893H, NPI712894H, NPI712896H, NPI712897H, NPI712898H, NPI712899H, NPI712901H, NPI712911H, NPI712912H, NPI712914H, NPI712915H, NPI712918H, NPI712919H, NPI712850H, NPI712851H, NPI712852H, NPI712854H, NPI712855H, NPI712856H, NPI712857H, NPI712860H, NPI712863H, NPI712865H, NPI712866H, NPI712868H, NPI712869H, NPI712870H, NPI712871H, NPI712872H, NPI712874H, NPI712875H, NPI712876H, NPI712877H, NPI712878H, NPI712879H, NPI712881H, NPI712882H, NPI712884H, NPI712812H, NPI712812H, NPI712813H, NPI712814H, NPI712815H, NPI712817H, NPI712818H, NPI712820H, NPI712821H, NPI712822H, NPI712823H, NPI712825H, NPI712827H, NPI712828H, NPI712832H, NPI712833H, NPI712837H, NPI712838H, NPI712839H, NPI712841H, NPI712842H, NPI712843H, NPI712844H, NPI712845H, NPI712846H, NPI712847H, NPI712849H, NPI712777H, NPI712780H, NPI712784H, NPI712787H, NPI712788H, NPI712789H, NPI712792H, NPI712793H, NPI712796H, NPI712797H, NPI712798H, NPI712799H, NPI712801H, NPI712802H, NPI712803H, NPI712804H, NPI712805H, NPI712807H, NPI712808H, NPI712811H, NPI712740H, NPI712742H, NPI712743H, NPI712746H, NPI712747H, NPI712748H, NPI712749H, NPI712750H, NPI712751H, NPI712752H, NPI712753H, NPI712754H, NPI712755H, NPI712756H, NPI712758H, NPI712759H, NPI712760H, NPI712762H, NPI712763H, NPI712764H, NPI712766H, NPI712767H, NPI712773H, NPI712724H, NPI712725H, NPI712726H, NPI712728H, NPI712729H, NPI712735H, NPI712737H, NPI712738H, NPI712671H, NPI712676H, NPI712678H, NPI712679H, NPI712681H, NPI712690H, NPI712696H, NPI712697H, NPI712698H, NPI712700H, NPI712701H, NPI712703H, NPI712706H, NPI712709H, NPI712710H, NPI712711H, NPI712716H, NPI712718H, NPI712719H, NPI712733H, NPI712630H, NPI712632H, NPI712638H, NPI712641H, NPI712656H, NPI712659H, NPI712665H, NPI712585H, NPI712588H, NPI712592H, NPI712604H, NPI712605H, NPI712606H, NPI712608H, NPI712610H, NPI712611H, NPI712613H, NPI712645H, NPI712648H, NPI714864H, NPI712530H, NPI712531H, NPI712532H, NPI712533H, NPI712534H, NPI712535H, NPI712536H, NPI712538H, NPI712539H, NPI712540H, NPI712541H, NPI712542H, NPI712543H, NPI712544H, NPI712545H, NPI712546H, NPI712549H, NPI712551H, NPI712553H, NPI712554H, NPI712555H, NPI712556H, NPI712557H, NPI712558H, NPI712559H, NPI712485H, NPI712486H, NPI712487H, NPI712488H, NPI712489H, NPI712490H, NPI712491H, NPI712502H, NPI712504H, NPI712505H, NPI712506H, NPI712507H, NPI712509H, NPI712510H, NPI712511H, NPI712512H, NPI712513H, NPI712515H, NPI712516H, NPI712518H, NPI712519H, NPI712520H, NPI712521H, NPI712522H, NPI712523H, NPI712525H, NPI712526H, NPI712527H, NPI712434H, NPI712435H, NPI712437H, NPI712439H, NPI712440H, NPI712441H, NPI712442H, NPI712443H, NPI712445H, NPI712446H, NPI712449H, NPI712451H, NPI712452H, NPI712453H, NPI712455H, NPI712456H, NPI712457H, NPI712458H, NPI712459H, NPI712463H, NPI712477H, NPI712478H, NPI712479H, NPI712482H, NPI712483H, NPI712484H, NPI712493H, NPI712496H, NPI712498H, NPI712499H, NPI712500H, NPI712501H, NPI712409H, NPI712410H, NPI712413H, NPI712431H, NPI712433H, NPI712436H, NPI712438H, NPI712356H, NPI712357H, NPI712382H, NPI712385H, NPI712386H, NPI712388H, NPI712389H, NPI712390H, NPI712391H, NPI712392H, NPI712393H, NPI712394H, NPI712395H, NPI712396H, NPI712397H, NPI712398H, NPI712399H, NPI712400H, NPI712401H, NPI712404H, NPI712406H, NPI712407H, NPI712414H, NPI712415H, NPI712417H, NPI712418H, NPI712419H, NPI712420H, NPI712421H, NPI712422H, NPI712423H, NPI712424H, NPI712425H, NPI712426H, NPI712427H, NPI712428H, NPI712429H, NPI712430H, NPI712309H, NPI712310H, NPI712311H, NPI712331H, NPI712332H, NPI712333H, NPI712334H, NPI712335H, NPI712336H, NPI712337H, NPI712344H, NPI712349H, NPI712350H, NPI712351H, NPI712352H, NPI712353H, NPI712355H, NPI712358H, NPI712359H, NPI712360H, NPI712361H, NPI712362H, NPI712363H, NPI712364H, NPI712365H, NPI712366H, NPI712368H, NPI712369H, NPI712370H, NPI712371H, NPI712372H, NPI712373H, NPI712374H, NPI712376H, NPI712377H, NPI712378H, NPI712379H, NPI712380H, NPI712381H, NPI712284H, NPI712294H, NPI712295H, NPI712296H, NPI712298H, NPI712299H, NPI712300H, NPI712303H, NPI712304H, NPI712305H, NPI712306H, NPI712307H, NPI712312H, NPI712313H, NPI712314H, NPI712315H, NPI712317H, NPI712318H, NPI712319H, NPI712320H, NPI712322H, NPI712323H, NPI712324H, NPI712325H, NPI712326H, NPI712328H, NPI712330H, NPI712217H, NPI712218H, NPI712219H, NPI712222H, NPI712229H, NPI712230H, NPI712231H, NPI712232H, NPI712243H, NPI712244H, NPI712245H, NPI712246H, NPI712248H, NPI712249H, NPI712254H, NPI712257H, NPI712258H, NPI712259H, NPI712260H, NPI712261H, NPI712262H, NPI712263H, NPI712264H, NPI712268H, NPI712269H, NPI712270H, NPI712271H, NPI712273H, NPI712274H, NPI712275H, NPI712276H, NPI712277H, NPI712278H, NPI712280H, NPI712282H, NPI712283H, NPI712285H, NPI712286H, NPI712287H, NPI712288H, NPI712289H, NPI712290H, NPI712291H, NPI712292H, NPI712293H, NPI712215H, NPI712216H, NPI712220H, NPI712221H, NPI712223H, NPI712224H, NPI712226H, NPI712227H, NPI712233H, NPI712234H, NPI712235H, NPI712236H, NPI712237H, NPI712238H, NPI712239H, NPI712240H, NPI712242H, NPI712250H, NPI712251H, NPI712252H, NPI712160H, NPI712161H, NPI712162H, NPI712169H, NPI712204H, NPI712205H, NPI712207H, NPI712208H, NPI712209H, NPI712209H, NPI712210H, NPI712211H, NPI712212H, NPI712213H, NPI712214H, NPI712164H, NPI712166H, NPI712167H, NPI712168H, NPI712171H, NPI712172H, NPI712173H, NPI712174H, NPI712175H, NPI712176H, NPI712177H, NPI712178H, NPI712180H, NPI712181H, NPI712182H, NPI712183H, NPI712184H, NPI712185H, NPI712186H, NPI712187H, NPI712188H, NPI712189H, NPI712190H, NPI712193H, NPI712194H, NPI712195H, NPI712196H, NPI712197H, NPI712198H, NPI712199H, NPI712200H, NPI712201H, NPI712202H, NPI712118H, NPI712119H, NPI712120H, NPI712121H, NPI712122H, NPI712124H, NPI712125H, NPI712126H, NPI712127H, NPI712128H, NPI712129H, NPI712130H, NPI712131H, NPI712132H, NPI712133H, NPI712134H, NPI712135H, NPI712136H, NPI712137H, NPI712138H, NPI712139H, NPI712140H, NPI712141H, NPI712142H, NPI712145H, NPI712147H, NPI712148H, NPI712149H, NPI712150H, NPI712151H, NPI712152H, NPI713732H, NPI712068H, NPI712071H, NPI712072H, NPI712073H, NPI712074H, NPI712075H, NPI712076H, NPI712077H, NPI712078H, NPI712079H, NPI712080H, NPI712081H, NPI712083H, NPI712084H, NPI712085H, NPI712086H, NPI712087H, NPI712088H, NPI712089H, NPI712090H, NPI712091H, NPI712092H, NPI712093H, NPI712094H, NPI712095H, NPI712096H, NPI712097H, NPI712098H, NPI712099H, NPI712100H, NPI712101H, NPI712102H, NPI712103H, NPI712104H, NPI712105H, NPI712106H, NPI712107H, NPI712108H, NPI712110H, NPI712111H, NPI712112H, NPI712113H, NPI712114H, NPI712115H, NPI712117H, NPI712153H, NPI712155H, NPI712156H, NPI712157H, NPI712158H, NPI712159H, NPI712038H, NPI712039H, NPI712042H, NPI712043H, NPI712044H, NPI712048H, NPI712049H, NPI712051H, NPI712052H, NPI712054H, NPI712055H, NPI712056H, NPI712057H, NPI712058H, NPI712059H, NPI712060H, NPI712066H, NPI711972H, NPI711973H, NPI711974H, NPI711976H, NPI711977H, NPI711978H, NPI711982H, NPI711984H, NPI711987H, NPI711988H, NPI711989H, NPI711994H, NPI711995H, NPI711997H, NPI711998H, NPI712001H, NPI712003H, NPI712005H, NPI712007H, NPI712008H, NPI711920H, NPI711927H, NPI711928H, NPI711932H, NPI711942H, NPI711944H, NPI711950H, NPI711951H, NPI711891H, NPI711902H, NPI711903H, NPI711907H, NPI711908H, NPI711909H, NPI711917H, NPI711936H, NPI711938H, NPI711939H, NPI711940H, NPI711952H, NPI711958H, NPI711960H, NPI711968H, NPI711971H, NPI711868H, NPI711882H, NPI711883H, NPI711885H, NPI711886H, NPI711887H, NPI711889H, NPI711890H, NPI711892H, NPI711893H, NPI711894H, NPI711895H, NPI711896H, NPI711897H, NPI711898H, NPI711899H, NPI711899H, NPI711900H, NPI711911H, NPI713366H, NPI713369H, NPI713494H, NPI711850H, NPI711852H, NPI711853H, NPI711854H, NPI711857H, NPI711821H, NPI711823H, NPI711824H, NPI711825H, NPI711826H, NPI711828H, NPI711830H, NPI711833H, NPI711834H, NPI711835H, NPI711836H, NPI711842H, NPI711851H, NPI711772H, NPI711773H, NPI711777H, NPI711780H, NPI711794H, NPI711799H, NPI711802H, NPI711806H, NPI711810H, NPI711811H, NPI711812H, NPI711813H, NPI711817H, NPI711818H, NPI711819H, NPI711820H, NPI711822H, NPI711837H, NPI711838H, NPI711839H, NPI711843H, NPI711845H, NPI711846H, NPI711849H, NPI711743H, NPI711745H, NPI711748H, NPI711751H, NPI711755H, NPI711757H, NPI711759H, NPI711761H, NPI711769H, NPI711781H, NPI711786H, NPI713493H, NPI711693H, NPI711711H, NPI711737H, NPI711740H, NPI713367H, NPI711714H, NPI711722H, NPI711724H, NPI711659H, NPI711661H, NPI711663H, NPI711665H, NPI711666H, NPI711668H, NPI711671H, NPI711676H, NPI711684H, NPI711689H, NPI713491H, NPI711629H, NPI711630H, NPI711631H, NPI711632H, NPI711633H, NPI711634H, NPI711635H, NPI711637H, NPI711638H, NPI711640H, NPI711642H, NPI711677H, NPI711612H, NPI711616H, NPI711617H, NPI711618H, NPI711620H, NPI711622H, NPI711624H, NPI711625H, NPI711626H, NPI711628H, NPI711641H, NPI711643H, NPI711647H, NPI711648H, NPI711650H, NPI711651H, NPI711654H, NPI711655H, NPI711656H, NPI711657H, NPI711593H, NPI711594H, NPI711595H, NPI711597H, NPI711601H, NPI711606H, NPI711607H, NPI711613H, NPI711614H, NPI711615H, NPI711558H, NPI711560H, NPI711561H, NPI711562H, NPI711563H, NPI711566H, NPI711568H, NPI711569H, NPI711570H, NPI711571H, NPI711572H, NPI711573H, NPI711574H, NPI711575H, NPI711576H, NPI711577H, NPI711578H, NPI711579H, NPI711580H, NPI711581H, NPI711582H, NPI711583H, NPI711584H, NPI711585H, NPI711586H, NPI711588H, NPI711589H, NPI711591H, NPI711600H, NPI711608H, NPI711609H, NPI713726H, NPI714866H, NPI711507H, NPI711509H, NPI711510H, NPI711512H, NPI711513H, NPI711514H, NPI711515H, NPI711518H, NPI711519H, NPI711522H, NPI711523H, NPI711524H, NPI711525H.NPI711526H, NPI711528H, NPI711529H, NPI711530H, NPI711531H, NPI711532H, NPI711533H, NPI711534H, NPI711536H, NPI711540H, NPI711541H, NPI711546H, NPI711547H, NPI711549H, NPI711550H, NPI711551H, NPI711553H, NPI711555H, NPI711488H, NPI711489H, NPI711490H, NPI711491H, NPI711492H, NPI711493H, NPI711494H, NPI711495H, NPI711496H, NPI711498H, NPI711499H, NPI711500H, NPI711501H, NPI711502H, NPI711503H, NPI711504H, NPI711505H, NPI711506H, NPI713485H, NPI711450H, NPI711451H, NPI711452H, NPI711453H, NPI711454H, NPI711455H, NPI711457H, NPI711459H, NPI711460H, NPI711461H, NPI711462H, NPI711463H, NPI711464H, NPI711465H, NPI711466H, NPI711467H, NPI711469H, NPI711470H, NPI711471H, NPI711472H, NPI711473H, NPI711477H, NPI711478H, NPI711479H, NPI711480H, NPI711481H, NPI711483H, NPI711484H, NPI711485H, NPI711486H, NPI711487H, NPI711413H, NPI711414H, NPI711415H, NPI711416H, NPI711417H, NPI711418H, NPI711419H, NPI711420H, NPI711421H, NPI711423H, NPI711424H, NPI711425H, NPI711426H, NPI711428H, NPI711429H, NPI711430H, NPI711431H, NPI711432H, NPI711433H, NPI711434H, NPI711436H, NPI711437H, NPI711439H, NPI711440H, NPI711441H, NPI711443H, NPI711444H, NPI711445H, NPI711447H, NPI711448H, NPI711449H, NPI711397H, NPI711398H, NPI711399H, NPI711400H, NPI711401H, NPI711357H, NPI711358H, NPI711359H, NPI711362H, NPI711364H, NPI711365H, NPI711368H, NPI711369H, NPI711371H, NPI711377H, NPI711378H, NPI711380H, NPI711381H, NPI711384H, NPI711387H, NPI711388H, NPI711389H, NPI711390H, NPI711392H, NPI711394H, NPI711402H, NPI711403H, NPI711406H, NPI711360H, NPI711361H, NPI711363H, NPI711367H, NPI711374H, NPI711375H, NPI711376H, NPI711382H, NPI711385H, NPI711386H, NPI711303H, NPI711305H, NPI711306H, NPI711307H, NPI711308H, NPI711309H, NPI711311H, NPI711312H, NPI711314H, NPI711315H, NPI711316H, NPI711317H, NPI711318H, NPI711319H, NPI711320H, NPI711321H, NPI711322H, NPI711323H, NPI711324H, NPI711325H, NPI711326H, NPI711335H, NPI711336H, NPI711339H, NPI711343H, NPI711347H, NPI711348H, NPI711349H, NPI711350H, NPI711351H, NPI711352H, NPI711353H, NPI711354H, NPI711355H, NPI711356H, NPI711268H, NPI711270H, NPI711271H, NPI711272H, NPI711273H, NPI711274H, NPI711275H, NPI711276H, NPI711277H, NPI711278H, NPI711279H, NPI711281H, NPI711282H, NPI711283H, NPI711285H, NPI711286H, NPI711287H, NPI711288H, NPI711289H, NPI711290H, NPI711291H, NPI711294H, NPI711295H, NPI711296H, NPI711298H, NPI711299H, NPI711301H, NPI711302H, NPI711304H, NPI711327H, NPI711328H, NPI711329H, NPI711330H, NPI711332H, NPI711333H, NPI711334H, NPI711337H, NPI711338H, NPI711340H, NPI711341H, NPI711342H, NPI711344H, NPI711345H, NPI711346H, NPI711229H, NPI711230H, NPI711231H, NPI711234H, NPI711235H, NPI711237H, NPI711238H, NPI711241H, NPI711242H, NPI711248H, NPI711250H, NPI711251H, NPI711253H, NPI711254H, NPI711259H, NPI711261H, NPI711264H, NPI711266H, NPI711267H, NPI711211H, NPI711212H, NPI711220H, NPI711227H, NPI711176H, NPI711180H, NPI711192H, NPI711193H, NPI711203H, NPI711148H, NPI711153H, NPI711155H, NPI711156H, NPI711168H, NPI711115H, NPI711127H, NPI711098H, NPI711102H, NPI711103H, NPI711105H, NPI711122H, NPI711123H, NPI711088H, NPI711092H, NPI711063H, NPI711065H, NPI711073H, NPI711075H, NPI711086H, NPI711027H, NPI711028H, NPI711029H, NPI711030H, NPI711031H, NPI711032H, NPI711034H, NPI711035H, NPI711036H, NPI711038H, NPI711041H, NPI711044H, NPI711045H, NPI711048H, NPI711049H, NPI711050H, NPI711051H, NPI711052H, NPI711053H, NPI711054H, NPI711055H, NPI711056H, NPI711057H, NPI711059H, NPI711060H, NPI711008H, NPI711009H, NPI711010H, NPI711011H, NPI711012H, NPI711013H, NPI711016H, NPI711017H, NPI711018H, NPI711019H, NPI711020H, NPI711021H, NPI711022H, NPI711023H, NPI711025H, NPI711039H, NPI711040H, NPI710975H, NPI710977H, NPI710978H, NPI710979H, NPI710980H, NPI710981H, NPI710982H, NPI710983H, NPI710984H, NPI710985H, NPI710986H, NPI710987H, NPI710989H, NPI710990H, NPI710991H, NPI710992H, NPI710994H, NPI710995H, NPI710996H, NPI710998H, NPI710999H, NPI711000H, NPI711001H, NPI711003H, NPI711004H, NPI711005H, NPI711006H, NPI711007H, NPI710945H, NPI710946H, NPI710947H, NPI710947H, NPI710948H, NPI710949H, NPI710950H, NPI710951H, NPI710952H, NPI710953H, NPI710954H, NPI710957H, NPI710966H, NPI710968H, NPI710970H, NPI710973H, NPI710928H, NPI710932H, NPI710933H, NPI710934H, NPI710935H, NPI710937H, NPI710938H, NPI710939H, NPI710940H, NPI710941H, NPI710942H, NPI710943H, NPI710944H, NPI710955H, NPI710956H, NPI710958H, NPI710959H, NPI710960H, NPI710961H, NPI710962H, NPI710963H, NPI710964H, NPI710965H, NPI710886H, NPI710887H, NPI710888H, NPI710889H, NPI710890H, NPI710891H, NPI710892H, NPI710893H, NPI710894H, NPI710897H, NPI710898H, NPI710899H, NPI710901H, NPI710903H, NPI710904H, NPI710912H, NPI710913H, NPI710914H, NPI710925H, NPI710929H, NPI710930H, NPI710931H, NPI710871H, NPI710873H, NPI710875H, NPI710877H, NPI710878H, NPI710879H, NPI710880H, NPI710881H, NPI710882H, NPI710883H, NPI710884H, NPI710885H, NPI710895H, NPI710905H, NPI710906H, NPI710908H, NPI710909H, NPI710910H, NPI710911H, NPI710915H, NPI710916H, NPI710917H, NPI710918H, NPI710919H, NPI710920H, NPI710921H, NPI710923H, NPI710924H, NPI710833H, NPI710836H, NPI710838H, NPI710839H, NPI710840H, NPI710841H, NPI710843H, NPI710844H, NPI710845H, NPI710847H, NPI710848H, NPI710849H, NPI710850H, NPI710851H, NPI710852H, NPI710853H, NPI710854H, NPI710856H, NPI710858H, NPI710859H, NPI710860H, NPI710861H, NPI710862H, NPI710863H, NPI710864H, NPI710865H, NPI710866H, NPI710867H, NPI710868H, NPI710869H, NPI710872H, NPI710874H, NPI710791H, NPI710794H, NPI710796H, NPI710799H, NPI710800H, NPI710801H, NPI710802H, NPI710804H, NPI710805H, NPI710806H, NPI710807H, NPI710811H, NPI710812H, NPI710813H, NPI710814H, NPI710815H, NPI710816H, NPI710817H, NPI710818H, NPI710822H, NPI710824H, NPI710825H, NPI710826H, NPI710828H, NPI710829H, NPI710830H, NPI710831H, NPI710832H, NPI710834H, NPI710835H, NPI710837H, NPI710753H, NPI710756H, NPI710759H, NPI710760H, NPI710761H, NPI710762H, NPI710764H, NPI710766H, NPI710767H, NPI710768H, NPI710769H, NPI710770H, NPI710771H, NPI710773H, NPI710778H, NPI710784H, NPI710786H, NPI710788H, NPI710792H, NPI710798H, NPI710720H, NPI710721H, NPI710723H, NPI710724H, NPI710725H, NPI710726H, NPI710727H, NPI710728H, NPI710729H, NPI710730H, NPI710731H, NPI710732H, NPI710733H, NPI710734H, NPI710736H, NPI710738H, NPI710739H, NPI710741H, NPI710742H, NPI710744H, NPI710745H, NPI710746H, NPI710747H, NPI710749H, NPI710754H, NPI710757H, NPI710758H, NPI710779H, NPI710681H, NPI710683H, NPI710688H, NPI710689H, NPI710692H, NPI710693H, NPI710694H, NPI710695H, NPI710701H, NPI710702H, NPI710706H, NPI710707H, NPI710712H, NPI710713H, NPI710714H, NPI710715H, NPI710716H, NPI710717H, NPI710719H, NPI710661H, NPI710663H, NPI710664H, NPI710665H, NPI710666H, NPI710667H, NPI710668H, NPI710669H, NPI710674H, NPI710675H, NPI710676H, NPI710677H, NPI710678H, NPI710679H, NPI710680H, NPI710624H, NPI710628H, NPI710629H, NPI710630H, NPI710635H, NPI710636H, NPI710637H, NPI710638H, NPI710639H, NPI710641H, NPI710643H, NPI710645H, NPI710648H, NPI710649H, NPI710650H, NPI710653H, NPI710654H, NPI710656H, NPI710658H, NPI710659H, NPI710660H, NPI710599H, NPI710541H, NPI710549H, NPI710572H, NPI710574H, NPI710519H, NPI710527H, NPI710528H, NPI710533H, NPI710535H, NPI710540H, NPI710543H, NPI710545H, NPI710556H, NPI710557H, NPI710462H, NPI710466H, NPI710476H, NPI710477H, NPI710478H, NPI710484H, NPI710485H, NPI710486H, NPI710487H, NPI710489H, NPI710494H, NPI710498H, NPI710499H, NPI710500H, NPI710502H, NPI710503H, NPI710505H, NPI710508H, NPI710510H, NPI710516H, NPI710517H, NPI710523H, NPI710524H, NPI710525H, NPI710452H, NPI710455H, NPI710458H, NPI710460H, NPI710461H, NPI710463H, NPI710464H, NPI710467H, NPI710469H, NPI710471H, NPI710428H, NPI710429H, NPI710430H, NPI710432H, NPI710433H, NPI710434H, NPI710435H, NPI710436H, NPI710437H, NPI710440H, NPI710442H, NPI710443H, NPI710444H, NPI710445H, NPI710446H, NPI710447H, NPI710449H, NPI710451H, NPI710390H, NPI710392H, NPI710408H, NPI710410H, NPI710411H, NPI710412H, NPI710413H, NPI710414H, NPI710415H, NPI710416H, NPI710417H, NPI710418H, NPI710419H, NPI710420H, NPI710421H, NPI710422H, NPI710423H, NPI710424H, NPI710425H, NPI710426H, NPI710347H, NPI710348H, NPI710349H, NPI710350H, NPI710351H, NPI710352H, NPI710356H, NPI710357H, NPI710360H, NPI710369H, NPI710371H, NPI710372H, NPI710373H, NPI710374H, NPI710375H, NPI710377H, NPI710379H, NPI710380H, NPI710381H, NPI710384H, NPI710385H, NPI710386H, NPI710387H, NPI710388H, NPI710391H, NPI710393H, NPI710394H, NPI710395H, NPI710397H, NPI710398H, NPI710399H, NPI710400H, NPI710401H, NPI710402H, NPI710403H, NPI710404H, NPI710406H, NPI710407H, NPI710291H, NPI710292H, NPI710295H, NPI710330H, NPI710331H, NPI710333H, NPI710334H, NPI710335H, NPI710337H, NPI710338H, NPI710339H, NPI710340H, NPI710341H, NPI710342H, NPI710343H, NPI710345H, NPI710346H, NPI710353H, NPI710355H, NPI710359H, NPI710361H, NPI710362H, NPI710364H, NPI710365H, NPI710366H, NPI710367H, NPI710368H, NPI710294H, NPI710296H, NPI710297H, NPI710310H, NPI710320H, NPI710321H, NPI710322H, NPI710324H, NPI710262H, NPI710263H, NPI710264H, NPI710266H, NPI710267H, NPI710268H, NPI710269H, NPI710270H, NPI710271H, NPI710272H, NPI710273H, NPI710274H, NPI710275H, NPI710276H, NPI710277H, NPI710278H, NPI710279H, NPI710280H, NPI710281H, NPI710282H, NPI710283H, NPI710284H, NPI710285H, NPI710286H, NPI710287H, NPI710288H, NPI710289H, NPI710290H, NPI710233H, NPI710235H, NPI710236H, NPI710237H, NPI710238H, NPI710239H, NPI710240H, NPI710241H, NPI710242H, NPI710243H, NPI710244H, NPI710245H, NPI710246H, NPI710247H, NPI710248H, NPI710249H, NPI710250H, NPI710251H, NPI710252H, NPI710253H, NPI710254H, NPI710255H, NPI710256H, NPI710258H, NPI710259H, NPI710260H, NPI710205H, NPI710206H, NPI710207H, NPI710210H, NPI710211H, NPI710212H, NPI710215H, NPI710216H, NPI710217H, NPI710218H, NPI710221H, NPI710222H, NPI710223H, NPI710224H, NPI710225H, NPI710226H, NPI710228H, NPI710229H, NPI710230H, NPI710231H, NPI710232H, NPI710234H, NPI710149H, NPI710162H, NPI710166H, NPI710168H, NPI710169H, NPI710170H, NPI710171H, NPI710175H, NPI710178H, NPI710179H, NPI710180H, NPI710181H, NPI710182H, NPI710184H, NPI710185H, NPI710186H, NPI710187H, NPI710188H, NPI710189H, NPI710190H, NPI710191H, NPI710193H, NPI710194H, NPI710195H, NPI710197H, NPI710198H, NPI710199H, NPI710200H, NPI710201H, NPI710202H, NPI710203H, NPI710204H, NPI710128H, NPI710129H, NPI710130H, NPI710131H, NPI710132H, NPI710133H, NPI710134H, NPI710135H, NPI710136H, NPI710137H, NPI710138H, NPI710139H, NPI710140H, NPI710141H, NPI710142H, NPI710143H, NPI710144H, NPI710145H, NPI710146H, NPI710148H, NPI710150H, NPI710151H, NPI710152H, NPI710153H, NPI710154H, NPI710156H, NPI710157H, NPI710158H, NPI710159H, NPI710160H, NPI710161H, NPI710163H, NPI710164H, NPI710165H, NPI710108H, NPI710110H, NPI710111H, NPI710112H, NPI710113H, NPI710114H, NPI710116H, NPI710117H, NPI710118H, NPI710119H, NPI710120H, NPI710121H, NPI710123H, NPI710124H, NPI710125H, NPI710126H, NPI710127H, NPI710079H, NPI710080H, NPI710081H, NPI710082H, NPI710084H, NPI710086H, NPI710087H, NPI710088H, NPI710089H, NPI710090H, NPI710091H, NPI710092H, NPI710093H, NPI710094H, NPI710095H, NPI710096H, NPI710097H, NPI710098H, NPI710099H, NPI710100H, NPI710101H, NPI710103H, NPI710104H, NPI710105H, NPI710106H, NPI710107H, NPI710055H, NPI710057H, NPI710060H, NPI710061H, NPI710062H, NPI710063H, NPI710064H, NPI710066H, NPI710067H, NPI710068H, NPI710069H, NPI710070H, NPI710072H, NPI710073H, NPI710074H, NPI710075H, NPI710076H, NPI710077H, NPI710083H, NPI710022H, NPI710023H, NPI710024H, NPI710025H, NPI710026H, NPI710028H, NPI710029H, NPI710030H, NPI710031H, NPI710032H, NPI710033H, NPI710034H, NPI710035H, NPI710038H, NPI710039H, NPI710040H, NPI710044H, NPI710045H, NPI710046H, NPI710047H, NPI710048H, NPI710050H, NPI710051H, NPI710052H, NPI710053H, NPI710054H, NPI710058H, NPI710059H, NPI709983H, NPI709984H, NPI709985H, NPI709986H, NPI709987H, NPI709988H, NPI709989H, NPI709990H, NPI709991H, NPI709992H, NPI709993H, NPI709994H, NPI709995H, NPI709996H, NPI709997H, NPI709998H, NPI709999H, NPI710000H, NPI710001H.NPI710002H, NPI710003H, NPI710004H, NPI710005H, NPI710006H, NPI710007H, NPI710008H, NPI710009H, NPI710010H, NPI710011H, NPI710013H, NPI710014H, NPI710015H, NPI710016H, NPI710017H, NPI710018H, NPI710019H, NPI710020H, NPI710021H, NPI709961H, NPI709962H, NPI709963H, NPI709967H, NPI709968H, NPI709970H, NPI709971H, NPI709972H, NPI709973H, NPI709974H, NPI709975H, NPI709981H, NPI709982H, NPI709943H, NPI709944H, NPI709945H, NPI709946H, NPI709949H, NPI709953H, NPI709956H, NPI709957H, NPI709958H, NPI709959H, NPI709960H, NPI709969H, NPI709976H, NPI709977H, NPI709978H, NPI709979H, NPI709980H, NPI709935H, NPI709947H, NPI709950H, NPI709954H, NPI709955H, NPI709916H, NPI709918H, NPI709919H, NPI709920H, NPI709921H, NPI709922H, NPI709923H, NPI709924H, NPI709925H, NPI709926H, NPI709927H, NPI709928H, NPI709929H, NPI709930H, NPI709931H, NPI709932H, NPI709933H, NPI709934H, NPI709936H, NPI709937H, NPI709938H, NPI709939H, NPI709940H, NPI709941H, NPI709942H, NPI709877H, NPI709878H, NPI709879H, NPI709880H, NPI709881H, NPI709882H, NPI709883H, NPI709884H, NPI709885H, NPI709886H, NPI709887H, NPI709888H, NPI709889H, NPI709890H, NPI709891H, NPI709892H, NPI709893H, NPI709894H, NPI709895H, NPI709897H, NPI709898H, NPI709900H, NPI709901H, NPI709902H, NPI709903H, NPI709904H, NPI709906H, NPI709907H, NPI709908H, NPI709909H, NPI709910H, NPI709911H, NPI709912H, NPI709913H, NPI709914H, NPI709915H, NPI709852H, NPI709853H, NPI709855H, NPI709857H, NPI709858H, NPI709859H, NPI709860H, NPI709862H, NPI709863H, NPI709864H, NPI709865H, NPI709866H, NPI709867H, NPI709868H, NPI709869H, NPI709870H, NPI709871H, NPI709872H, NPI709873H, NPI709874H, NPI709875H, NPI709824H, NPI709825H, NPI709826H, NPI709828H, NPI709829H, NPI709831H, NPI709832H, NPI709834H, NPI709836H, NPI709837H, NPI709838H, NPI709839H, NPI709840H, NPI709841H, NPI709843H, NPI709844H, NPI709846H, NPI709847H, NPI709848H, NPI709849H, NPI709800H, NPI709803H, NPI709805H, NPI709806H, NPI709807H, NPI709808H, NPI709809H, NPI709810H, NPI709812H, NPI709813H, NPI709815H, NPI709816H, NPI709817H, NPI709818H, NPI709819H, NPI709820H, NPI709821H, NPI709823H, NPI709827H, NPI709830H, NPI709833H, NPI709835H, NPI714869H, NPI709747H, NPI709748H, NPI709749H, NPI709750H, NPI709751H, NPI709752H, NPI709753H, NPI709754H, NPI709755H, NPI709756H, NPI709757H, NPI709758H, NPI709759H, NPI709760H, NPI709761H, NPI709763H, NPI709764H, NPI709765H, NPI709768H, NPI709769H, NPI709773H, NPI709774H, NPI709779H, NPI709781H, NPI709783H, NPI709784H, NPI709790H, NPI709791H, NPI709704H, NPI709706H, NPI709707H, NPI709708H, NPI709709H, NPI709710H, NPI709711H, NPI709712H, NPI709713H, NPI709714H, NPI709715H, NPI709716H, NPI709717H, NPI709718H, NPI709720H, NPI709722H, NPI709723H, NPI709724H, NPI709725H, NPI709727H, NPI709728H, NPI709729H, NPI709730H, NPI709731H, NPI709732H, NPI709734H, NPI709735H, NPI709736H, NPI709737H, NPI709738H, NPI709739H, NPI709740H, NPI709741H, NPI709742H, NPI709743H, NPI709744H, NPI709745H, NPI709675H, NPI709678H, NPI709679H, NPI709681H, NPI709683H, NPI709685H, NPI709690H, NPI709691H, NPI709693H, NPI709694H, NPI709695H, NPI709696H, NPI709697H, NPI709698H, NPI709699H, NPI709700H, NPI709702H, NPI709703H, NPI709638H, NPI709640H, NPI709641H, NPI709642H, NPI709643H, NPI709644H, NPI709645H, NPI709646H, NPI709647H, NPI709648H, NPI709649H, NPI709650H, NPI709651H, NPI709652H, NPI709653H, NPI709661H, NPI709664H, NPI709666H, NPI709667H, NPI709670H, NPI709671H, NPI709672H, NPI709673H, NPI709673H, NPI709674H, NPI709608H, NPI709610H, NPI709611H, NPI709615H, NPI709616H, NPI709617H, NPI709619H, NPI709620H, NPI709621H, NPI709623H, NPI709625H, NPI709626H, NPI709627H, NPI709628H, NPI709629H, NPI709630H, NPI709631H, NPI709635H, NPI709636H, NPI709566H, NPI709567H, NPI709568H, NPI709569H, NPI709570H, NPI709571H, NPI709574H, NPI709576H, NPI709588H, NPI709592H, NPI709593H, NPI709594H, NPI709595H, NPI709597H, NPI709598H, NPI709600H, NPI709601H, NPI709602H, NPI709603H, NPI709604H, NPI709605H, NPI709607H, NPI709609H, NPI709534H, NPI709560H, NPI709561H, NPI709562H, NPI709563H, NPI709564H, NPI709565H, NPI709572H, NPI709573H, NPI709577H, NPI709578H, NPI709579H, NPI709580H, NPI709581H, NPI709582H, NPI709583H, NPI709585H, NPI709586H, NPI709589H, NPI709590H, NPI709591H, NPI709596H, NPI709503H, NPI709523H, NPI709524H, NPI709525H, NPI709528H, NPI709535H, NPI709536H, NPI709537H, NPI709539H, NPI709540H, NPI709541H, NPI709542H, NPI709543H, NPI709544H, NPI709545H, NPI709546H, NPI709547H, NPI709548H, NPI709549H, NPI709550H, NPI709551H, NPI709552H, NPI709553H, NPI709554H, NPI709555H, NPI709556H, NPI709557H, NPI709558H, NPI709559H, NPI709488H, NPI709489H, NPI709492H, NPI709493H, NPI709494H, NPI709495H, NPI709496H, NPI709498H, NPI709499H, NPI709500H, NPI709501H, NPI709502H, NPI709504H, NPI709507H, NPI709508H, NPI709509H, NPI709510H, NPI709511H, NPI709512H, NPI709513H, NPI709514H, NPI709515H, NPI709516H, NPI709517H, NPI709518H, NPI709519H, NPI709520H, NPI709522H, NPI709449H, NPI709450H, NPI709451H, NPI709452H, NPI709453H, NPI709454H, NPI709455H, NPI709456H, NPI709457H, NPI709458H, NPI709459H, NPI709460H, NPI709463H, NPI709464H, NPI709465H, NPI709466H, NPI709467H, NPI709468H, NPI709469H, NPI709470H, NPI709471H, NPI709472H, NPI709473H, NPI709475H, NPI709478H, NPI709481H, NPI709482H, NPI709483H, NPI709484H, NPI709485H, NPI709486H, NPI709487H, NPI709491H, NPI709423H, NPI709434H, NPI709435H, NPI709436H, NPI709438H, NPI709439H, NPI709442H, NPI709443H, NPI709444H, NPI709445H, NPI709446H, NPI709447H, NPI709391H, NPI709392H, NPI709393H, NPI709394H, NPI709395H, NPI709396H, NPI709397H, NPI709399H, NPI709400H, NPI709401H, NPI709402H, NPI709403H, NPI709405H, NPI709406H, NPI709408H, NPI709409H, NPI709410H, NPI709411H, NPI709412H, NPI709413H, NPI709414H, NPI709415H, NPI709416H, NPI709417H, NPI709418H, NPI709419H, NPI709420H, NPI709421H, NPI709350H, NPI709351H, NPI709354H, NPI709355H, NPI709356H, NPI709357H, NPI709358H, NPI709360H, NPI709361H, NPI709362H, NPI709363H, NPI709364H, NPI709365H, NPI709366H, NPI709367H, NPI709368H, NPI709369H, NPI709370H, NPI709371H, NPI709372H, NPI709373H, NPI709374H, NPI709375H, NPI709380H, NPI709381H, NPI709382H, NPI709383H, NPI709384H, NPI709385H, NPI709386H, NPI709387H, NPI709388H, NPI709389H, NPI709297H, NPI709298H, NPI709302H, NPI709303H, NPI709304H, NPI709305H, NPI709306H, NPI709307H, NPI709308H, NPI709309H, NPI709310H, NPI709311H, NPI709312H, NPI709313H, NPI709314H, NPI709315H, NPI709316H, NPI709317H, NPI709318H, NPI709319H, NPI709320H, NPI709321H, NPI709331H, NPI709336H, NPI709338H, NPI709339H, NPI709340H, NPI709341H, NPI709342H, NPI709343H, NPI709344H, NPI709345H, NPI709346H, NPI709347H, NPI709348H, NPI709349H, NPI709352H, NPI709353H, NPI709263H, NPI709264H, NPI709265H, NPI709266H, NPI709267H, NPI709268H, NPI709270H, NPI709271H, NPI709272H, NPI709273H, NPI709274H, NPI709276H, NPI709277H, NPI709278H, NPI709279H, NPI709280H, NPI709281H, NPI709283H, NPI709284H, NPI709285H, NPI709286H, NPI709287H, NPI709288H, NPI709289H, NPI709290H, NPI709291H, NPI709292H, NPI709293H, NPI709294H, NPI709295H, NPI709238H, NPI709240H, NPI709241H, NPI709242H, NPI709243H, NPI709244H, NPI709245H, NPI709246H, NPI709247H, NPI709248H, NPI709249H, NPI709250H, NPI709251H, NPI709252H, NPI709255H, NPI709256H, NPI709257H, NPI709258H, NPI709260H, NPI709211H, NPI709212H, NPI709213H, NPI709214H, NPI709215H, NPI709217H, NPI709218H, NPI709219H, NPI709220H, NPI709221H, NPI709222H, NPI709223H, NPI709224H, NPI709225H, NPI709226H, NPI709227H, NPI709229H, NPI709230H, NPI709232H, NPI709233H, NPI709234H, NPI709235H, NPI709236H, NPI709237H, NPI709239H, NPI709186H, NPI709187H, NPI709188H, NPI709189H, NPI709190H, NPI709191H, NPI709192H, NPI709193H, NPI709196H, NPI709197H, NPI709199H, NPI709200H, NPI709201H, NPI709202H, NPI709203H, NPI709204H, NPI709205H, NPI709206H, NPI709207H, NPI709208H, NPI709209H, NPI709210H, NPI709149H, NPI709150H, NPI709151H, NPI709152H, NPI709153H, NPI709154H, NPI709155H, NPI709156H, NPI709157H, NPI709158H, NPI709159H, NPI709160H, NPI709161H, NPI709162H, NPI709163H, NPI709164H, NPI709165H, NPI709166H, NPI709167H, NPI709168H, NPI709169H, NPI709170H, NPI709172H, NPI709173H, NPI709174H, NPI709175H, NPI709176H, NPI709177H, NPI709178H, NPI709179H, NPI709180H, NPI709181H, NPI709182H, NPI709184H, NPI709185H, NPI709117H, NPI709118H, NPI709119H, NPI709120H, NPI709121H, NPI709122H, NPI709123H, NPI709124H, NPI709125H, NPI709126H, NPI709127H, NPI709128H, NPI709129H, NPI709130H, NPI709131H, NPI709132H, NPI709133H, NPI709134H, NPI709135H, NPI709136H, NPI709137H, NPI709139H, NPI709140H, NPI709141H, NPI709142H, NPI709143H, NPI709144H, NPI709145H, NPI709146H, NPI709147H, NPI709079H, NPI709080H, NPI709080H, NPI709081H, NPI709082H, NPI709083H, NPI709084H, NPI709085H, NPI709087H, NPI709088H, NPI709089H, NPI709090H, NPI709091H, NPI709092H, NPI709093H, NPI709094H, NPI709095H, NPI709096H, NPI709097H, NPI709098H, NPI709099H, NPI709100H, NPI709101H, NPI709102H, NPI709103H, NPI709105H, NPI709106H, NPI709107H, NPI709109H, NPI709110H, NPI709113H, NPI709114H, NPI709115H, NPI709116H, NPI709056H, NPI709058H, NPI709059H, NPI709060H, NPI709061H, NPI709062H, NPI709063H, NPI709064H, NPI709065H, NPI709066H, NPI709068H, NPI709069H, NPI709070H, NPI709071H, NPI709072H, NPI709073H, NPI709074H, NPI709075H, NPI709077H, NPI709078H, NPI709086H, NPI709018H, NPI709029H, NPI709030H, NPI709031H, NPI709032H, NPI709033H, NPI709034H, NPI709035H, NPI709036H, NPI709037H, NPI709038H, NPI709039H, NPI709040H, NPI709041H, NPI709042H, NPI709043H, NPI709044H, NPI709045H, NPI709046H, NPI709048H, NPI709049H, NPI709050H, NPI709051H, NPI709052H, NPI709053H, NPI709054H, NPI709055H, NPI709057H, NPI709067H, NPI708997H, NPI709000H, NPI709001H, NPI709006H, NPI709007H, NPI709008H, NPI709009H, NPI709010H, NPI709011H, NPI709012H, NPI709013H, NPI709014H, NPI709015H, NPI709016H, NPI709017H, NPI709019H, NPI709020H, NPI709021H, NPI709022H, NPI709023H, NPI709024H, NPI709026H, NPI709027H, NPI709028H, NPI708969H, NPI708970H, NPI708971H, NPI708972H, NPI708973H, NPI708974H, NPI708975H, NPI708976H, NPI708977H, NPI708978H, NPI708979H, NPI708980H, NPI708981H, NPI708984H, NPI708985H, NPI708986H, NPI708987H, NPI708988H, NPI708989H, NPI708991H, NPI708992H, NPI708993H, NPI708994H, NPI708995H, NPI708996H, NPI708998H, NPI709002H, NPI709003H, NPI709004H, NPI709005H, NPI708931H, NPI708933H, NPI708934H, NPI708935H, NPI708936H, NPI708937H, NPI708938H, NPI708939H, NPI708940H, NPI708941H, NPI708942H, NPI708943H, NPI708944H, NPI708945H, NPI708946H, NPI708947H, NPI708948H, NPI708949H, NPI708950H, NPI708951H, NPI708952H, NPI708953H, NPI708954H, NPI708956H, NPI708958H, NPI708959H, NPI708960H, NPI708961H, NPI708962H, NPI708963H, NPI708964H, NPI708965H, NPI708967H, NPI708968H, NPI708893H, NPI708894H, NPI708895H, NPI708896H, NPI708897H, NPI708898H, NPI708899H, NPI708900H, NPI708901H, NPI708902H, NPI708903H, NPI708904H, NPI708905H, NPI708906H, NPI708907H, NPI708908H, NPI708909H, NPI708910H, NPI708911H, NPI708912H, NPI708913H, NPI708914H, NPI708915H, NPI708916H, NPI708917H, NPI708918H, NPI708919H, NPI708920H, NPI708921H, NPI708922H, NPI708923H, NPI708924H, NPI708925H, NPI708926H, NPI708929H, NPI708930H, NPI708863H, NPI708864H, NPI708865H, NPI708866H, NPI708867H, NPI708868H, NPI708869H, NPI708870H, NPI708871H, NPI708872H, NPI708873H, NPI708874H, NPI708875H, NPI708876H, NPI708878H, NPI708879H, NPI708880H, NPI708881H, NPI708882H, NPI708883H, NPI708884H, NPI708886H, NPI708887H, NPI708888H, NPI708889H, NPI708890H, NPI708891H, NPI708892H, NPI708824H, NPI708826H, NPI708827H, NPI708828H, NPI708829H, NPI708830H, NPI708831H, NPI708833H, NPI708834H, NPI708835H, NPI708836H, NPI708837H, NPI708838H, NPI708840H, NPI708842H, NPI708844H, NPI708846H, NPI708847H, NPI708849H, NPI708850H, NPI708853H, NPI708854H, NPI708856H, NPI708857H, NPI708859H, NPI708860H, NPI708861H, NPI708862H, NPI708807H, NPI708808H, NPI708811H, NPI708813H, NPI708814H, NPI708815H, NPI708819H, NPI708820H, NPI708821H, NPI708822H, NPI708773H, NPI708776H, NPI708778H, NPI708779H, NPI708780H, NPI708781H, NPI708782H, NPI708783H, NPI708784H, NPI708785H, NPI708786H, NPI708787H, NPI708789H.NPI708790H, NPI708791H, NPI708794H, NPI708795H, NPI708796H, NPI708799H, NPI708800H, NPI708801H, NPI708804H, NPI708805H, NPI708747H, NPI708748H, NPI708749H, NPI708750H, NPI708751H, NPI708752H, NPI708753H, NPI708756H, NPI708757H, NPI708758H, NPI708759H, NPI708762H, NPI708763H, NPI708764H, NPI708767H, NPI708768H, NPI708770H, NPI708771H, NPI708772H, NPI708774H, NPI708775H, NPI708732H, NPI708735H, NPI708736H, NPI708737H, NPI708738H, NPI708739H, NPI708740H, NPI708741H, NPI708742H, NPI708743H, NPI708746H, NPI708694H, NPI708695H, NPI708698H, NPI708703H, NPI708704H, NPI708706H, NPI708708H, NPI708710H, NPI708711H, NPI708712H, NPI708674H, NPI708676H, NPI708677H, NPI708678H, NPI708680H, NPI708681H, NPI708683H, NPI708685H, NPI708687H, NPI708688H, NPI708689H, NPI708690H, NPI708691H, NPI708693H, NPI708647H, NPI708648H, NPI708652H, NPI708654H, NPI708655H, NPI708656H, NPI708657H, NPI708658H, NPI708659H, NPI708660H, NPI708661H, NPI708662H, NPI708663H, NPI708664H, NPI708665H, NPI708666H, NPI708667H, NPI708668H, NPI708670H, NPI708672H, NPI708673H, NPI708682H, NPI708684H, NPI708596H, NPI708598H, NPI708600H, NPI708614H, NPI708615H, NPI708616H, NPI708617H, NPI708619H, NPI708620H, NPI708621H, NPI708622H, NPI708623H, NPI708624H, NPI708625H, NPI708626H, NPI708627H, NPI708628H, NPI708629H, NPI708630H, NPI708631H, NPI708632H, NPI708633H, NPI708634H, NPI708636H, NPI708637H, NPI708638H, NPI708639H, NPI708640H, NPI708642H, NPI708643H, NPI708644H, NPI708645H, NPI708646H, NPI708649H, NPI708650H, NPI708651H, NPI713616H, NPI708565H, NPI708567H, NPI708568H, NPI708569H, NPI708570H, NPI708571H, NPI708572H, NPI708573H, NPI708574H, NPI708576H, NPI708578H, NPI708579H, NPI708580H, NPI708581H, NPI708583H, NPI708584H, NPI708585H, NPI708586H, NPI708587H, NPI708588H, NPI708589H, NPI708590H, NPI708591H, NPI708592H, NPI708593H, NPI708594H, NPI708595H, NPI708599H, NPI708602H, NPI708604H, NPI708611H, NPI708612H, NPI708613H, NPI708537H, NPI708539H, NPI708540H, NPI708542H, NPI708544H, NPI708546H, NPI708551H, NPI708552H, NPI708554H, NPI708555H, NPI708556H, NPI708557H, NPI708558H, NPI708559H, NPI708560H, NPI708561H, NPI708562H, NPI708563H, NPI708564H, NPI708566H, NPI708528H, NPI708529H, NPI708530H, NPI708531H, NPI708532H, NPI708533H, NPI708534H, NPI708535H, NPI708536H, NPI708541H, NPI708577H, NPI708509H, NPI708510H, NPI708512H, NPI708513H, NPI708514H, NPI708515H, NPI708516H, NPI708517H, NPI708519H, NPI708520H, NPI708521H, NPI708522H, NPI708523H, NPI708524H, NPI708526H, NPI708527H, NPI708487H, NPI708488H, NPI708489H, NPI708490H, NPI708491H, NPI708492H, NPI708493H, NPI708494H, NPI708495H, NPI708496H, NPI708497H, NPI708498H, NPI708499H, NPI708500H, NPI708501H, NPI708502H, NPI708504H, NPI708505H, NPI708506H, NPI708508H, NPI708454H, NPI708462H, NPI708463H, NPI708464H, NPI708465H, NPI708466H, NPI708467H, NPI708468H, NPI708469H, NPI708471H, NPI708472H, NPI708473H, NPI708474H, NPI708475H, NPI708476H, NPI708477H, NPI708479H, NPI708481H, NPI708482H, NPI708483H, NPI708484H, NPI708485H, NPI708486H, NPI708440H, NPI708441H, NPI708442H, NPI708443H, NPI708444H, NPI708445H, NPI708447H, NPI708448H, NPI708449H, NPI708450H, NPI708451H, NPI708452H, NPI708453H, NPI708456H, NPI708461H, NPI708416H, NPI708419H, NPI708420H, NPI708421H, NPI708422H, NPI708423H, NPI708424H, NPI708427H, NPI708428H, NPI708429H, NPI708431H, NPI708432H, NPI708433H, NPI708434H, NPI708435H, NPI708436H, NPI708438H, NPI708439H, NPI708388H, NPI708389H, NPI708390H, NPI708391H, NPI708401H, NPI708403H, NPI708404H, NPI708405H, NPI708407H, NPI708408H, NPI708412H, NPI708413H, NPI708414H, NPI708352H, NPI708353H, NPI708354H, NPI708355H, NPI708356H, NPI708359H, NPI708363H, NPI708364H, NPI708365H, NPI708366H, NPI708368H, NPI708369H, NPI708370H, NPI708372H, NPI708373H, NPI708374H, NPI708375H, NPI708377H, NPI708380H, NPI708381H, NPI708383H, NPI708384H, NPI708385H, NPI708386H, NPI708387H, NPI708392H, NPI708395H, NPI708396H, NPI708397H, NPI708398H, NPI708342H, NPI708343H, NPI708344H, NPI708345H, NPI708346H, NPI708347H, NPI708348H, NPI708349H, NPI708350H, NPI708351H, NPI708360H, NPI708361H, NPI708362H, NPI708327H, NPI708328H, NPI708334H, NPI708335H, NPI708336H, NPI708337H, NPI708338H, NPI708339H, NPI708340H, NPI708341H, NPI708311H, NPI708312H, NPI708313H, NPI708314H, NPI708317H, NPI708318H, NPI708320H, NPI708321H, NPI708323H, NPI708324H, NPI708325H, NPI708326H, NPI708330H, NPI708331H, NPI708265H, NPI708274H, NPI708275H, NPI708276H, NPI708277H, NPI708278H, NPI708279H, NPI708280H, NPI708281H, NPI708282H, NPI708283H, NPI708284H, NPI708286H, NPI708287H, NPI708290H, NPI708291H, NPI708293H, NPI708294H, NPI708297H, NPI708298H, NPI708301H, NPI708302H, NPI708303H, NPI708304H, NPI708307H, NPI708308H, NPI708310H, NPI708245H, NPI708246H, NPI708247H, NPI708251H, NPI708252H, NPI708253H, NPI708254H, NPI708255H, NPI708259H, NPI708260H, NPI708261H, NPI708263H, NPI708264H, NPI708266H, NPI708267H, NPI708268H, NPI708269H, NPI708270H, NPI708271H, NPI708272H, NPI708218H, NPI708220H, NPI708221H, NPI708222H, NPI708223H, NPI708224H, NPI708225H, NPI708226H, NPI708228H, NPI708229H, NPI708230H, NPI708231H, NPI708232H, NPI708233H, NPI708234H, NPI708235H, NPI708236H, NPI708237H, NPI708238H, NPI708239H, NPI708240H, NPI708241H, NPI708242H, NPI708244H, NPI708207H, NPI708209H, NPI708210H, NPI708211H, NPI708212H, NPI708213H, NPI708216H, NPI708217H, NPI708181H, NPI708182H, NPI708183H, NPI708184H, NPI708185H, NPI708186H, NPI708187H, NPI708189H, NPI708192H, NPI708193H, NPI708194H, NPI708195H, NPI708196H, NPI708197H, NPI708198H, NPI708200H, NPI708203H, NPI708205H, NPI708206H, NPI708208H, NPI708157H, NPI708158H, NPI708159H, NPI708160H, NPI708165H, NPI708166H, NPI708167H, NPI708169H, NPI708170H, NPI708171H, NPI708172H, NPI708173H, NPI708174H, NPI708176H, NPI708177H, NPI708178H, NPI708180H, NPI708132H, NPI708135H, NPI708136H, NPI708138H, NPI708139H, NPI708140H, NPI708141H, NPI708142H, NPI708143H, NPI708144H, NPI708145H, NPI708146H, NPI708147H, NPI708148H, NPI708149H, NPI708150H, NPI708151H, NPI708152H, NPI708154H, NPI708155H, NPI708156H, NPI708162H, NPI708098H, NPI708069H, NPI707984H, NPI707988H, NPI707941H, NPI707965H, NPI707936H, NPI707913H, NPI707876H, NPI707909H, NPI707859H, NPI707829H, NPI707853H, NPI707810H, NPI707796H, NPI707803H, NPI707781H, NPI707787H, NPI707747H, NPI707748H, NPI707773H, NPI707726H, NPI707710H, NPI707711H, NPI707685H, NPI707706H, NPI707693H, NPI707665H, NPI707667H, NPI707619H, NPI707620H, NPI707636H, NPI707608H, NPI707609H, NPI707512H, NPI707529H, NPI707540H, NPI707510H, NPI707511H, NPI707475H, NPI707481H, NPI707370H, NPI707372H, NPI707391H, NPI707248H, NPI707251H, NPI707254H, NPI707260H, NPI707264H, NPI707266H, NPI707267H, NPI707234H, NPI707197H, NPI707179H, NPI707188H, NPI707147H, NPI707158H, NPI707120H, NPI707128H, NPI707090H, NPI707091H, NPI707101H, NPI707064H, NPI707071H, NPI707079H, NPI706958H, NPI706960H, NPI706966H, NPI706982H, NPI706936H, NPI706965H, NPI706918H, NPI706924H, NPI706927H, NPI706896H, NPI706821H, NPI706862H, NPI706796H, NPI706765H, NPI706774H, NPI706775H, NPI706797H, NPI706728H, NPI706729H, NPI706672H, NPI706640H, NPI706598H, NPI706548H, NPI706566H, NPI706515H, NPI706535H, NPI709296H, NPI706522H, NPI706446H, NPI706381H, NPI706385H, NPI706387H, NPI706344H, NPI706355H, NPI706329H, NPI706246H, NPI706267H, NPI706217H, NPI706192H, NPI706193H, NPI706128H, NPI706121H, NPI706073H, NPI706075H, NPI706038H, NPI705960H, NPI705971H, NPI705929H, NPI705931H, NPI705941H, NPI705868H, NPI705855H, NPI705573H, NPI705416H, NPI705383H, NPI705399H, NPI705400H, NPI705401H, NPI705402H, NPI705403H, NPI705352H, NPI705355H, NPI705356H, NPI705361H, NPI705364H, NPI705365H, NPI705370H, NPI705377H, NPI705404H, NPI705405H, NPI705340H, NPI705343H, NPI705349H, NPI710027H, NPI705143H, NPI704988H, NPI704845H, NPI704852H, NPI704786H, NPI704762H, NPI704706H, NPI704718H, NPI704721H, NPI704655H, NPI704657H, NPI704634H, NPI704636H, NPI704682H, NPI704618H, NPI704621H, NPI704630H, NPI704578H, NPI704603H, NPI704548H, NPI704554H, NPI704509H, NPI704519H, NPI704494H, NPI704401H, NPI704420H, NPI704444H, NPI704377H, NPI704383H, NPI704345H, NPI704311H, NPI704324H, NPI704327H, NPI704248H, NPI704286H, NPI704214H, NPI704235H, NPI704148H, NPI704177H, NPI704143H, NPI704098H, NPI704109H, NPI704110H, NPI704055H, NPI704072H, NPI704020H, NPI703990H, NPI704005H, NPI703907H, NPI703908H, NPI703929H, NPI703930H, NPI703885H, NPI703810H, NPI703646H, NPI703565H, NPI703610H, NPI703486H, NPI703443H, NPI703307H, NPI703288H, NPI703243H, NPI703248H, NPI703210H, NPI703215H, NPI703222H, NPI703152H, NPI703180H, NPI703193H, NPI703126H, NPI703030H, NPI702793H, NPI702513H, NPI702386H, NPI702044H
GTIN 00763000524821, Serial Numbers: NPI708731H
GTIN 00763000524838, Serial Numbers: NPI710327H
GTIN 00763000524845, Serial Numbers: NPI708733H
GTIN 00763000524883, Serial Numbers: NPI710326H
|
Recalling Firm/ Manufacturer |
Medtronic Neuromodulation 7000 Central Ave Ne Minneapolis MN 55432-3568
|
For Additional Information Contact |
Lindsey Dickinson 763-514-2000
|
Manufacturer Reason for Recall |
During cardioversion events, the Model B35200 Percept PC Implantable Neurostimulator could become unresponsive and non-functional. Cardioversion is a procedure performed by a cardiologist to correct an abnormal heartbeat. Medtronic has determined that the cardioversion may damage the circuit electronics in the Percept PC stimulator, thereby, causing the Percept PC stimulator to become non-responsive.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Medtronic distributed a correction notice via mail courier service to impacted U.S. consignees, delivered 06-October-2021.
A second attempt to notify US consignees is scheduled for the week of 08-November-2021.
A third attempt to notify US consignees is scheduled for the week of 10-Jan-2022.
In addition to Medtronic s communications to Percept PC implanting and managing physicians, Medtronic will notify affected patients implanted with Percept PC INS devices of this issue in the US. US Patient notifications are scheduled to happen within a month s timeframe. Medtronic will retain delivery confirmations for all patient notification mailings.
|
Quantity in Commerce |
10,006 units |
Distribution |
Worldwide distribution - US Nationwide and Puerto Rico. The countries of Australia, Canada, Austria, Belgium, Canary Islands, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Israel, Italy, Luxembourg, Netherlands, Norway, Poland, Portugal, Slovakia, Slovenia, Spain, Sweden, Switzerland, United Kingdom, India, Japan, Korea, Republic Of, Argentina, Brazil, Colombia, Thailand, and Hong Kong.
|
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
PMA Database |
PMAs with Product Code = MHY and Original Applicant = MEDTRONIC Inc.
|
|
|
|