 Class 2 Device Recall Infatherm Infant Transport Mattress; Model Number 989805616831see related information
Date Initiated by FirmNovember 26, 2021
Date PostedJanuary 18, 2022
Recall Status1 Open3, Classified
Recall NumberZ-0505-2022
Recall Event ID 89173
510(K)NumberK936084 
Product Classification Pack, hot or cold, disposable - Product Code IMD
ProductInfa-Therm Transport Mattress, Reference Number 989805616831 1015
Code Information All units are impacted
FEI Number 3016618143
Recalling Firm/
Manufacturer
Philips North America Llc
222 Jacobs St
Cambridge MA 02141-2289
Manufacturer Reason
for Recall
After the device was cleared by the FDA, Philips made changes to the labeling that do not fall within the existing FDA clearance. These devices with the modified labeling cannot be distributed or sold without new 510(k) clearance.
FDA Determined
Cause 2
No Marketing Application
ActionAn URGENT Medical Device Recall notification letter dated November 2021 was distributed to customers. The actions that should be taken by the customer / user to prevent risks to patients or users: 1. Identify all Infa-Therm Transport Mattress in your possession. 2. Destroy all Infa-Therm Transport Mattress, regardless if they are expired or not. 3. Please complete the Urgent Medical Device Recall Response Form online by using the URL provided on the form. 4. Identify and source an alternative device to meet your needs. Philips will no longer offer the Infa-Therm Transport Mattress. 5. Share this communication with your staff and other impacted departments and organizations. Actions planned by Philips to correct the problem: We understand that this may be an inconvenience to you. Philips will reimburse you for the quantity of the lnfa-Therm Transport Mattresses destroyed. If you need any further information concerning this issue, please email infantWarming@Philips.com (preferred communication path) or contact +1-629-215-7280 Monday through Friday, 8AM -3PM Central Standard Time. If you email infantWarming@Philips.com, you will receive an email response with the link to submit your Recall Acknowledgement form online, as well as a response to your inquiry.
Quantity in Commerce141,744 units (Updated to 141,750 units as of 2/15/22)
DistributionUS Nationwide and Canada
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.