Class 2 Device Recall Powered Environmental Control System |
|
Date Initiated by Firm |
November 17, 2021 |
Create Date |
January 14, 2022 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-0497-2022 |
Recall Event ID |
89181 |
Product Classification |
System, environmental control, powered - Product Code IQA
|
Product |
Senior Living Arial Emergency and Nurse Call Systems |
Code Information |
Model No. 0900-989; Lot No.: 7VQYZB3, G1GG8B3, 1TX2613, 7VQWZB3, GRRT243, 1TX3613, G1H58B3, HJP7N23, HJP9N23, 51SDR33, and GRRV243; UDI No.: (01)00840948100987(21)7VQYZB3, (01)00840948100987(21)1TX2613, (01)00840948100987(21)7VQZZB3, (01)00840948100987(21)GRRT243, (01)00840948100987(21)G1GF8B3, (01)00840948100987(21)1TX3613, (01)00840948100987(21)G1H58B3, (01)00840948100987(21)HJP7N23, (01)00840948100987(21)HJP9N23, (01)00840948100987(21)51SDR33, (01)00840948100987(21)GRRV243
|
Recalling Firm/ Manufacturer |
Stanley Security Solutions Inc 4600 Vine St Lincoln NE 68503-2823
|
For Additional Information Contact |
Paul Hall 402-742-9419
|
Manufacturer Reason for Recall |
Systems are not alarming due to increased memory consumption.
|
FDA Determined Cause 2 |
Component design/selection |
Action |
The firm distributed a notice titled "Voluntary Market Withdrawal of Dell Optiplex 3070/3080 Servers" informing customers of an issue with Arial Call Systems outfitted with any Dell 3070/3080 servers, resulting in failure of alarm. The firm asked that customers facilitate the replacement of the Dell 3070/3080 server components through a service technician. Customers with any questions were directed to contact Mr. Garth Jack at (402)742-9335.
A second notice from the firm titled "Urgent: Field Safety Notice Adisory Regarding Use of Dell Optiplex 3070/3080 Servers with Stanley Healthcare Arial Emergency and Nurse Call Systems," dated 12/17/2021 was sent to customers by certified mail. Customers who own an Arial Call System and did not purchase a Dell 3070/3080 server from Stanley but utilize one with their system need to discontinue use and replace the server as soon as possible. Customers who purchased a Dell 3070/3080 server used in conjunction with the Arial Call system were informed that a representative from Stanley will be reaching out to coordinate the replacement of the server at no cost. Replacement servers will be shipped to customers, but their replacement needs to be conducted by a Stanley representative, so they are to be safely stored until this can be coordinated. Dell 3070/3080 servers removed following correction need to be returned to Stanley Healthcare using a provided RMA number, along with a completed Certification of Products Returned form that will be provided. The firm asks that the Field Safety Notice be distributed within organizations with affected product. Customers with any questions are to contact Mr. Garth Jack at (402)742-9335 or by email at garth.jack@SBDinc.com. |
Quantity in Commerce |
11 units (systems) |
Distribution |
Worldwide distribution - US Nationwide distribution in the states of FL, MD, MI, MN, SD, TN, and WA. The countries of Canada and Qatar. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|