 Class 2 Device Recall Azurion R1.2see related information
Date Initiated by FirmMay 26, 2021
Date PostedMay 03, 2022
Recall Status1 Open3, Classified
Recall NumberZ-0886-2022
Recall Event ID 89964
510(K)NumberK172822 
Product Classification Interventional fluoroscopic x-ray system - Product Code OWB
ProductAzurion systems with software release R1.x
Code Information software release R1.x
Recalling Firm/
Manufacturer
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Veenpluis 4-6
Best Netherlands
For Additional Information ContactDusty Leppert, CHP
978-2280190
Manufacturer Reason
for Recall
In the Azurion system, the user can add a new study to a patient by selecting the option Add Study . The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a software defect, when the study is initiated by pressing Start Procedure , the Patient Type changes inadvertently to a Patient Type different than the one selected as shown in the Table below. Patient type is one of the factors involved in the dose control process. The incorrect patient type changes the technique factors to be used by the system without notification to the user.
FDA Determined
Cause 2
Software design
ActionPhilips notified customers of the issue via Electronic Product Radiation Defect notification letter dated March 4, 2022. The letter identified the affected product, problem and actions to be taken. Customers were informed that they will be contacted by a Philips Field Service Engineers to schedule the software update at no cost. For questions contact your local Philips representative: 1-800-722-9377.
Quantity in Commerce722 systems
DistributionUS Nationwide and Worldwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.