| Date Initiated by Firm | May 05, 2022 |
| Create Date | June 06, 2022 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1241-2022 |
| Recall Event ID |
90103 |
| 510(K)Number | K153778 |
| Product Classification |
Device, Vascular, for Promoting Embolization - Product Code KRD
|
| Product | Nester Embolization Microcoil
RPN GPN
MWCE-18-5-2-NESTER G52733
MWCE-18-5-3-NESTER G52734
MWCE-18-7-3-NESTER-01 G47338
MWCE-18-7-2-NESTER-01 G47337
MWCE-18-7-5-NESTER-01 G47340
MWCE-18-3-2-NESTER G52731
MWCE-18-7-8-NESTER-01 G47342
MWCE-18-7-8-NESTER G52739
MWCE-18-7-5-NESTER G23521
MWCE-18-3-3-NESTER G52732
MWCE-18-5-3-NESTER-01 G47334
MWCE-18-7-6-NESTER G52738
MWCE-18-7-6-NESTER-01 G47341
MWCE-18-3-3-NESTER-01 G47332
MWCE-18-7-4-NESTER G52737
MWCE-18-7-3-NESTER G52736
MWCE-18-3-2-NESTER-01 G47331
MWCE-18-5-2-NESTER-01 G47333
MWCE-18-2-2-NESTER-01 G47344
MWCE-18-5-5-NESTER-01 G47335
MWCE-18-7-2-NESTER G52735
MWCE-18-7-4-NESTER-01 G47339
MWCE-18-7-10-NESTER G52741
MWCE-18-5-5-NESTER G23520 |
| Code Information |
Lots:
14445809
NS14440413
14438720
NS14533529
NS14531987
NS14512935
NS14531991
NS14531985
NS14531989
NS14531990
NS14531993
NS14531994
14230667
14219518
14228061
14232591
14233388
14233393
14280197
14299971
14239561
14239563
14240096
14227837
14231033
14231052
14232846
14318778
14343805
NS14318613
NS14345341
14325852
14366728
NS14341560
NS14345338
NS14345796
14317075
14366729
14395251
14446172
14452759
14461634
14542803
14542804
14542805
NS14533531
14502814
14233390
14223991
14280196
14299968
14299969
14228357
14240093
14240095
14240097
14242960
14280193
14280194
NS14242956
NS14267436
NS14285103
NS14285105
NS14285113
NS14285116
NS14285118
14343806
14343808
14401670
14560261
14242333
14242961
14244909
14230666
14231051
14231480
14236762
14239564
14240094
14242899
14280195
14244951
14340480
14343008
14363296
NS14329321
14363792
14371377
14392159
14557484
14558691
14560262
14244907
14239562
14303167
14303168
14233392
14244908
14387092
14391580
14368792
14394030
14404724
14411527
14411533
NS14393837
14392161
14310711
14392479
NS14401834
14393836
14394139
14407670
NS14411530
14392162
14395248
14395250
14401669
14404664
14458554
14469137
14461546
14557483
14560260
14568246
14324499
14324561
14329743
NS14357214
14368791
14405012
14404722
14407669
14407731
14444295
14445814
14459064
14239565
14248731
14270214
14230668
14231034
14232592
14299970
NS14280198
14343807
14329319
14363297
14329744
14329746
14356323
NS14354206
NS14356996
NS14356998
NS14357000
NS14357208
NS14357211
14329738
14334099
NS14458938
14458553
14438764
14459660
14460713
NS14445808
14117167
NS14204660
NS14204661
14264718
14317112
14322914
14314095
14354105
14354308
NS14345792
NS14354207
NS14356995
NS14357209
NS14357216
14331801
14345342
14395165
14329353
14461977
14461477
14461828
14427826
14443855
14461633
14464822
14492202
14493151
14493152
14248725
14248726
14259109
14264230
14261712
14275595
14341563
NS14357210
14329745
14333908
14396081
14399139
14338855
14343042
14345340
14329742
14438721
14450442
14461547
14440430
14459065
14568247
14568247
14295067
NS14285117
NS14329737
NS14331626
14329320
14333886
14377708
NS14357207
14381297
14399140
14466958
14472466
14460714
14492201
14531047
14531984
14568248
14571878
14571880
14574880
14574881
14513458
NS14219519
14204662
14204670
14204671
NS14204659
NS14221709
NS14222394
NS14222395
NS14224135
14244953
NS14331628
14278732
14408758
NS14411445
NS14411451
NS14411458
NS14418475
NS14420111
14411430
14204669
NS14221710
NS14221711
NS14222393
NS14224732
NS14224733
NS14227161
14204831
NS14222396
14244906
14270212
NS14331627
14275593
14295066
14336480
14354205
14392480
14406553
14407581
14408760
14410615
14411443
14414399
14410985
14423203
14410614
14411462
NS14227159
14368790
14427825
14501828
14501829
14427592
14427824
14466959
14470990
14472464
14472465
14468751
14492203
14571879
14574882
14530352
14541815
14513459
14515399
14516002
14516654
14526502
14530353
14531046
14541814
14542035
14532062
14221708
14323043
14327107
14341564
14345339
14336479
14343007
14473168
14469795
14471408
14515400
14516655
14515996
14533527
14533528
14537554
14539480
14531986
14537553
14542040
14545259
14276060
14275594
14278731
14285684
14290348
14411526
14343009
14333907
14356324
NS14411535
NS14411547
NS14411552
NS14411554
14571883
14551461
NS14581367
NS14589845
NS14589846
NS14589849
NS14589850
NS14589852
NS14589853
NS14589854
14402618
14418585
14418586
14345791
14345793
14345788
14350205
14345789
14352728
14356322
14345787
14348032
14359695
14354310
14356327
NS14379763
14401830
14414400
14413821
14420112
14407730
14409288
14501831
14539894
14438763
14516001
NS14571882
NS14574879
NS14577031
NS14577032
14602397
14542065
14545258
14545260
NS14589847
14345786
NS14411553
14360272
14363791
14368403
14395249
14404900
14413823
14411426
14426440
14426612
14427591
14427841
14434086
14433107
14408759
14418479
14433108
NS14531992
NS14533530
14467172
14473216
14539190
NS14572219
14539479
14545463
14545607
14411425
14427842
14432995
14404721
14431270
14432995
14357212
14352729
14345794
14356325
14356964
14357215
14434547
14436671
14434548
14434595
NS14461975
14262241
14264229
14360275
14362491
14423286
14357213
14360273
14360592
14362491
NS14440408
14427849
14436087
NS14459068
NS14461976
14432180
14436670
NS14464824
NS14531988
14470988
14470989
14476959
14482117
14469136
14485533
14605139
14551465
14552979
14553253
14555750
14558470
14150466
14150467
14150468
14150469
NS14165566
NS14165567
14146661
14147375
14148063
14148259
14149471
14145634
14359694
14360276
NS14372466
14368404
NS14401831
NS14411450
NS14411453
14360277
NS14381274
NS14528436
14394029
14488157
14485718
14555470
14555472
14147340
14148258
14149405
14156620
NS14172962
14146659
14147374
14149472
14149473
14156792
14144005
14148062
14149787
14161034
14163040
NS14177890
NS14227839
NS14231035
NS14239568
NS14285100
14267434
14290347
14542802
14485734
14488321
14488329
14478826
14160864
14173697
14177891
NS14177889
14295069
14295070
14295071
14295072
14299750
14290349
14293975
14293976
14296632
14501830
14539895
14487851
NS14528437
NS14528438
NS14530354
14394565
NS14167628
14163039
14165565
14167626
14168332
14170845
14170846
14172960
14170238
14170380
14172961
14181210
NS14222392
14223990
14235116
NS14222392
NS14267441
NS14267448
NS14267449
14295068
14317113
14323042
14301594
14296631
NS14438149
NS14440407
NS14440416
14438762
NS14458935
NS14458936
NS14458939
NS14461978
14541936
14368789
14369064
14354200
14566982
14431272
NS14267447
NS14267450
14223572
14223573
14224134
NS14244954
NS14267435
NS14267438
NS14267444
NS14267446
14222391
14223993
14224137
14227334
14227836
14227160
14229459
14223634
14230669
NS14231481
NS14242334
NS14242335
NS14242335X
NS14242963
NS14248732
NS14267440
NS14267442
NS14267443
NS14267445
NS14285107
NS14285108
NS14285109
NS14285112
14362932
NS14387091
NS14392160
NS14404723
NS14411427
NS14411429
NS14411433
NS14411442
NS14411549
NS14413822
NS14418478
NS14424354
NS14411431
NS14411432
NS14411441
NS14411449
NS14411452
NS14411550
NS14411555
NS14411556
14438518
14447010
NS14531979
14181872
14223992
14228060
14229489
14229490
14221706
14227838
14228355
14221707
14227840
14228356
14228358
14231479
14271104
14271105
14276627
14262242
NS14239566
NS14285106
NS14285110
NS14285111
14223989
14228356
14231032
14372468
14373505
NS14404726
NS14411440
NS14411455
NS14411456
NS14411457
NS14411551
14368630
14372645
14373511
14373557
14373933
14379764
14381275
14382644
14383707
14383708
14411439
NS14401833
NS14404725
NS14404727
NS14411446
NS14411447
NS14411448
14387108
14387110
NS14401832
NS14375880
14228059
14229380
14230665
14231053
14232840
14233391
14233394
14236754
NS14274105
14231478
14233389
14242962
NS14239567
NS14242959
NS14244952
NS14267437
NS14267437X
14293977
14303290
14303289
14340477
14340479
14374504
14373512
14374528
14374529
14387109
14401671
14146660
14301593
14340478
NS14323044
NS14324153
14411529
NS14372467
14411459
14411528
14411531
14411532
14411534
14360274
14382641
14411444
14387089
NS14404728
14440429
14442188
NS14458937
14441794
14440414
14441795
14438151
UDI:
(01)00827002473447(17)270201(10)14445809
(01)00827002527331(17)270201(10)NS14440413
(01)00827002473386(17)270201(10)14438720
(01)00827002527348(17)270301(10)NS14533529
(01)00827002527386(17)270301(10)NS14531987
(01)00827002527317(17)270301(10)NS14512935
(01)00827002527317(17)270301(10)NS14531991
(01)00827002527379(17)270301(10)NS14531985
(01)00827002527379(17)270301(10)NS14531989
(01)00827002527331(17)270301(10)NS14531990
(01)00827002527379(17)270301(10)NS14531993
(01)00827002527331(17)270301(10)NS14531994
(01)00827002527317(17)261101(10)14230667
(01)00827002527362(17)261101(10)14219518
(01)00827002473423(17)261101(10)14228061
(01)00827002473393(17)261101(10)14232591
(01)00827002473447(17)261101(10)14233388
(01)00827002473393(17)261101(10)14233393
(01)00827002527317(17)261101(10)14280197
(01)00827002473317(17)261101(10)14299971
(01)00827002527317(17)261101(10)14239561
(01)00827002527379(17)261101(10)14239563
(01)00827002527379(17)261101(10)14240096
(01)00827002527331(17)261101(10)14227837
(01)00827002473447(17)261101(10)14231033
(01)00827002527317(17)261101(10)14231052
(01)00827002473317(17)261101(10)14232846
(01)00827002235205(17)261201(10)14318778
(01)00827002527379(17)261201(10)14343805
(01)00827002527386(17)261201(10)NS14318613
(01)00827002527317(17)261201(10)NS14345341
(01)00827002473348(17)261201(10)14325852
(01)00827002527324(17)261201(10)14366728
(01)00827002527362(17)261201(10)NS14341560
(01)00827002235212(17)261201(10)NS14345338
(01)00827002527331(17)261201(10)NS14345796
(01)00827002473447(17)261201(10)14317075
(01)00827002527362(17)261201(10)14366729
(01)00827002473317(17)270202(10)14395251
(01)00827002527331(17)270202(10)14446172
(01)00827002473386(17)270202(10)14452759
(01)00827002473348(17)270202(10)14461634
(01)00827002473386(17)270302(10)14542803
(01)00827002473386(17)270302(10)14542804
(01)00827002527362(17)270302(10)14542805
(01)00827002235212(17)270302(10)NS14533531
(01)00827002527393(17)270302(10)14502814
(01)00827002473379(17)261102(10)14233390
(01)00827002235212(17)261102(10)14223991
(01)00827002527317(17)261102(10)14280196
(01)00827002473317(17)261102(10)14299968
(01)00827002527317(17)261102(10)14299969
(01)00827002527386(17)261102(10)14228357
(01)00827002527393(17)261102(10)14240093
(01)00827002527355(17)261102(10)14240095
(01)00827002527324(17)261102(10)14240097
(01)00827002527348(17)261102(10)14242960
(01)00827002527331(17)261102(10)14280193
(01)00827002473317(17)261102(10)14280194
(01)00827002527355(17)261102(10)NS14242956
(01)00827002527386(17)261102(10)NS14267436
(01)00827002527355(17)261102(10)NS14285103
(01)00827002527362(17)261102(10)NS14285105
(01)00827002235212(17)261102(10)NS14285113
(01)00827002527355(17)261102(10)NS14285116
(01)00827002235212(17)261102(10)NS14285118
(01)00827002473393(17)261202(10)14343806
(01)00827002527379(17)261202(10)14343808
(01)00827002235205(17)270103(10)14401670
(01)00827002527348(17)270303(10)14560261
(01)00827002473324(17)261103(10)14242333
(01)00827002527355(17)261103(10)14242961
(01)00827002235212(17)261103(10)14244909
(01)00827002527393(17)261103(10)14230666
(01)00827002527317(17)261103(10)14231051
(01)00827002473423(17)261103(10)14231480
(01)00827002527362(17)261103(10)14236762
(01)00827002473416(17)261103(10)14239564
(01)00827002527393(17)261103(10)14240094
(01)00827002527324(17)261103(10)14242899
(01)00827002527317(17)261103(10)14280195
(01)00827002527324(17)261103(10)14244951
(01)00827002527355(17)261203(10)14340480
(01)00827002527324(17)261203(10)14343008
(01)00827002473324(17)261203(10)14363296
(01)00827002527379(17)261203(10)NS14329321
(01)00827002527386(17)270104(10)14363792
(01)00827002473324(17)270104(10)14371377
(01)00827002527355(17)270104(10)14392159
(01)00827002473393(17)270304(10)14557484
(01)00827002527379(17)270304(10)14558691
(01)00827002473348(17)270304(10)14560262
(01)00827002473324(17)261104(10)14244907
(01)00827002527324(17)261104(10)14239562
(01)00827002473379(17)261104(10)14303167
(01)00827002473324(17)261104(10)14303168
(01)00827002473348(17)261104(10)14233392
(01)00827002527362(17)261104(10)14244908
(01)00827002527393(17)270105(10)14387092
(01)00827002527317(17)270105(10)14391580
(01)00827002527393(17)270105(10)14368792
(01)00827002527379(17)270105(10)14394030
(01)00827002473423(17)270105(10)14404724
(01)00827002527416(17)270105(10)14411527
(01)00827002527416(17)270105(10)14411533
(01)00827002235205(17)270105(10)NS14393837
(01)00827002473447(17)270105(10)14392161
(01)00827002527379(17)261105(10)14310711
(01)00827002473379(17)270106(10)14392479
(01)00827002527362(17)270106(10)NS14401834
(01)00827002235205(17)270106(10)14393836
(01)00827002473348(17)270106(10)14394139
(01)00827002473423(17)270106(10)14407670
(01)00827002527416(17)270106(10)NS14411530
(01)00827002473423(17)270106(10)14392162
(01)00827002473324(17)270106(10)14395248
(01)00827002473379(17)270106(10)14395250
(01)00827002527393(17)270106(10)14401669
(01)00827002527348(17)270106(10)14404664
(01)00827002473393(17)270206(10)14458554
(01)00827002473348(17)270206(10)14469137
(01)00827002527379(17)270206(10)14461546
(01)00827002527379(17)270306(10)14557483
(01)00827002527348(17)270306(10)14560260
(01)00827002527324(17)270306(10)14568246
(01)00827002527317(17)261206(10)14324499
(01)00827002473348(17)261206(10)14324561
(01)00827002527317(17)261206(10)14329743
(01)00827002527386(17)261206(10)NS14357214
(01)00827002473423(17)270107(10)14368791
(01)00827002527355(17)270107(10)14405012
(01)00827002527355(17)270107(10)14404722
(01)00827002527393(17)270107(10)14407669
(01)00827002527393(17)270107(10)14407731
(01)00827002473416(17)270207(10)14444295
(01)00827002473409(17)270207(10)14445814
(01)00827002527355(17)270207(10)14459064
(01)00827002473423(17)261107(10)14239565
(01)00827002473355(17)261107(10)14248731
(01)00827002473423(17)261107(10)14270214
(01)00827002527393(17)261107(10)14230668
(01)00827002527393(17)261107(10)14231034
(01)00827002527393(17)261107(10)14232592
(01)00827002473379(17)261107(10)14299970
(01)00827002527355(17)261107(10)NS14280198
(01)00827002473393(17)261207(10)14343807
(01)00827002527317(17)261207(10)14329319
(01)00827002527324(17)261207(10)14363297
(01)00827002473331(17)261207(10)14329744
(01)00827002527393(17)261207(10)14329746
(01)00827002527355(17)261207(10)14356323
(01)00827002527331(17)261207(10)NS14354206
(01)00827002527355(17)261207(10)NS14356996
(01)00827002527317(17)261207(10)NS14356998
(01)00827002235212(17)261207(10)NS14357000
(01)00827002527348(17)261207(10)NS14357208
(01)00827002527379(17)261207(10)NS14357211
(01)00827002527393(17)261207(10)14329738
(01)00827002527348(17)261207(10)14334099
(01)00827002527331(17)270208(10)NS14458938
(01)00827002527317(17)270208(10)14458553
(01)00827002527393(17)270208(10)14438764
(01)00827002473447(17)270208(10)14459660
(01)00827002527324(17)270208(10)14460713
(01)00827002527317(17)270208(10)NS14445808
(01)00827002473409(17)260908(10)14117167
(01)00827002527355(17)261008(10)NS14204660
(01)00827002527317(17)261008(10)NS14204661
(01)00827002527317(17)261108(10)14264718
(01)00827002527324(17)261208(10)14317112
(01)00827002473386(17)261208(10)14322914
(01)00827002527362(17)261208(10)14314095
(01)00827002527355(17)261208(10)14354105
(01)00827002235212(17)261208(10)14354308
(01)00827002527393(17)261208(10)NS14345792
(01)00827002527331(17)261208(10)NS14354207
(01)00827002527317(17)261208(10)NS14356995
(01)00827002527362(17)261208(10)NS14357209
(01)00827002527331(17)261208(10)NS14357216
(01)00827002527386(17)261208(10)14331801
(01)00827002527362(17)261208(10)14345342
(01)00827002473386(17)261208(10)14395165
(01)00827002527355(17)261208(10)14329353
(01)00827002527386(17)270209(10)14461977
(01)00827002473324(17)270209(10)14461477
(01)00827002473331(17)270209(10)14461828
(01)00827002473423(17)270209(10)14427826
(01)00827002473423(17)270209(10)14443855
(01)00827002473379(17)270209(10)14461633
(01)00827002473386(17)270209(10)14464822
(01)00827002473393(17)270309(10)14492202
(01)00827002527362(17)270309(10)14493151
(01)00827002527379(17)270309(10)14493152
(01)00827002473331(17)261109(10)14248725
(01)00827002473379(17)261109(10)14248726
(01)00827002473386(17)261109(10)14259109
(01)00827002473416(17)261109(10)14264230
(01)00827002473317(17)261109(10)14261712
(01)00827002527386(17)261109(10)14275595
(01)00827002527348(17)261209(10)14341563
(01)00827002527317(17)261209(10)NS14357210
(01)00827002473355(17)261209(10)14329745
(01)00827002473317(17)261209(10)14333908
(01)00827002473386(17)261209(10)14396081
(01)00827002473393(17)261209(10)14399139
(01)00827002473386(17)261209(10)14338855
(01)00827002527362(17)261209(10)14343042
(01)00827002527317(17)261209(10)14345340
(01)00827002473447(17)261209(10)14329742
(01)00827002473393(17)270210(10)14438721
(01)00827002473379(17)270210(10)14450442
(01)00827002527362(17)270210(10)14461547
(01)00827002527393(17)270210(10)14440430
(01)00827002527393(17)270210(10)14459065
(01)00827002527324(17)270310(10)14568247
(01)00827002527324(17)270310(10)14568247
(01)00827002527386(17)261110(10)14295067
(01)00827002527386(17)261110(10)NS14285117
(01)00827002527348(17)261110(10)NS14329737
(01)00827002527386(17)261110(10)NS14331626
(01)00827002527379(17)261210(10)14329320
(01)00827002527331(17)261210(10)14333886
(01)00827002473386(17)261210(10)14377708
(01)00827002527355(17)261210(10)NS14357207
(01)00827002527324(17)261210(10)14381297
(01)00827002527379(17)261210(10)14399140
(01)00827002527355(17)270211(10)14466958
(01)00827002527379(17)270211(10)14472466
(01)00827002527331(17)270211(10)14460714
(01)00827002473386(17)270311(10)14492201
(01)00827002527379(17)270311(10)14531047
(01)00827002473393(17)270311(10)14531984
(01)00827002473324(17)270311(10)14568248
(01)00827002473393(17)270311(10)14571878
(01)00827002473393(17)270311(10)14571880
(01)00827002527362(17)270311(10)14574880
(01)00827002473386(17)270311(10)14574881
(01)00827002527331(17)270311(10)14513458
(01)00827002527379(17)261011(10)NS14219519
(01)00827002527379(17)261011(10)14204662
(01)00827002473423(17)261011(10)14204670
(01)00827002527324(17)261011(10)14204671
(01)00827002527362(17)261011(10)NS14204659
(01)00827002527331(17)261011(10)NS14221709
(01)00827002473331(17)261011(10)NS14222394
(01)00827002473348(17)261011(10)NS14222395
(01)00827002527393(17)261011(10)NS14224135
(01)00827002527362(17)261111(10)14244953
(01)00827002527386(17)261111(10)NS14331628
(01)00827002473379(17)261111(10)14278732
(01)00827002527331(17)270112(10)14408758
(01)00827002527331(17)270112(10)NS14411445
(01)00827002527331(17)270112(10)NS14411451
(01)00827002527331(17)270112(10)NS14411458
(01)00827002527386(17)270112(10)NS14418475
(01)00827002527355(17)270112(10)NS14420111
(01)00827002527348(17)270112(10)14411430
(01)00827002527317(17)261012(10)14204669
(01)00827002527379(17)261012(10)NS14221710
(01)00827002235212(17)261012(10)NS14221711
(01)00827002473324(17)261012(10)NS14222393
(01)00827002473393(17)261012(10)NS14224732
(01)00827002527355(17)261012(10)NS14224733
(01)00827002473386(17)261012(10)NS14227161
(01)00827002527348(17)261012(10)14204831
(01)00827002473317(17)261012(10)NS14222396
(01)00827002473317(17)261112(10)14244906
(01)00827002473447(17)261112(10)14270212
(01)00827002527386(17)261112(10)NS14331627
(01)00827002527386(17)261112(10)14275593
(01)00827002527393(17)261112(10)14295066
(01)00827002473324(17)261212(10)14336480
(01)00827002527317(17)261212(10)14354205
(01)00827002527324(17)270113(10)14392480
(01)00827002527324(17)270113(10)14406553
(01)00827002527331(17)270113(10)14407581
(01)00827002527348(17)270113(10)14408760
(01)00827002527355(17)270113(10)14410615
(01)00827002527324(17)270113(10)14411443
(01)00827002473447(17)270113(10)14414399
(01)00827002473379(17)270113(10)14410985
(01)00827002473331(17)270113(10)14423203
(01)00827002527386(17)270113(10)14410614
(01)00827002527362(17)270113(10)14411462
(01)00827002473379(17)261013(10)NS14227159
(01)00827002473331(17)270114(10)14368790
(01)00827002527393(17)270214(10)14427825
(01)00827002527348(17)270214(10)14501828
(01)00827002473348(17)270214(10)14501829
(01)00827002527393(17)270214(10)14427592
(01)00827002473423(17)270214(10)14427824
(01)00827002473393(17)270214(10)14466959
(01)00827002473416(17)270214(10)14470990
(01)00827002527317(17)270214(10)14472464
(01)00827002527317(17)270214(10)14472465
(01)00827002527393(17)270214(10)14468751
(01)00827002235212(17)270314(10)14492203
(01)00827002527379(17)270314(10)14571879
(01)00827002527362(17)270314(10)14574882
(01)00827002527324(17)270314(10)14530352
(01)00827002527393(17)270314(10)14541815
(01)00827002527386(17)270314(10)14513459
(01)00827002473324(17)270314(10)14515399
(01)00827002473355(17)270314(10)14516002
(01)00827002473447(17)270314(10)14516654
(01)00827002527317(17)270314(10)14526502
(01)00827002527393(17)270314(10)14530353
(01)00827002473379(17)270314(10)14531046
(01)00827002527355(17)270314(10)14541814
(01)00827002473317(17)270314(10)14542035
(01)00827002527324(17)270314(10)14532062
(01)00827002527331(17)261014(10)14221708
(01)00827002527348(17)261214(10)14323043
(01)00827002527317(17)261214(10)14327107
(01)00827002527386(17)261214(10)14341564
(01)00827002473317(17)261214(10)14345339
(01)00827002473317(17)261214(10)14336479
(01)00827002527317(17)261214(10)14343007
(01)00827002527331(17)270215(10)14473168
(01)00827002527317(17)270215(10)14469795
(01)00827002473379(17)270215(10)14471408
(01)00827002473386(17)270315(10)14515400
(01)00827002473409(17)270315(10)14516655
(01)00827002527355(17)270315(10)14515996
(01)00827002473423(17)270315(10)14533527
(01)00827002473423(17)270315(10)14533528
(01)00827002527393(17)270315(10)14537554
(01)00827002527393(17)270315(10)14539480
(01)00827002473416(17)270315(10)14531986
(01)00827002527331(17)270315(10)14537553
(01)00827002473317(17)270315(10)14542040
(01)00827002473331(17)270315(10)14545259
(01)00827002527317(17)261115(10)14276060
(01)00827002473409(17)261115(10)14275594
(01)00827002473331(17)261115(10)14278731
(01)00827002527317(17)261115(10)14285684
(01)00827002527317(17)261115(10)14290348
(01)00827002527416(17)261215(10)14411526
(01)00827002527324(17)261215(10)14343009
(01)00827002473317(17)261215(10)14333907
(01)00827002473386(17)261215(10)14356324
(01)00827002527416(17)261215(10)NS14411535
(01)00827002527416(17)261215(10)NS14411547
(01)00827002527416(17)261215(10)NS14411552
(01)00827002527416(17)261215(10)NS14411554
(01)00827002527379(17)270316(10)14571883
(01)00827002527393(17)270316(10)14551461
(01)00827002473447(17)270316(10)NS14581367
(01)00827002527317(17)270316(10)NS14589845
(01)00827002527331(17)270316(10)NS14589846
(01)00827002527355(17)270316(10)NS14589849
(01)00827002527386(17)270316(10)NS14589850
(01)00827002527317(17)270316(10)NS14589852
(01)00827002527355(17)270316(10)NS14589853
(01)00827002527386(17)270316(10)NS14589854
(01)00827002527362(17)261216(10)14402618
(01)00827002527379(17)261216(10)14418585
(01)00827002527379(17)261216(10)14418586
(01)00827002473423(17)261216(10)14345791
(01)00827002527393(17)261216(10)14345793
(01)00827002473423(17)261216(10)14345788
(01)00827002473386(17)261216(10)14350205
(01)00827002527393(17)261216(10)14345789
(01)00827002473447(17)261216(10)14352728
(01)00827002473324(17)261216(10)14356322
(01)00827002473379(17)261216(10)14345787
(01)00827002527317(17)261216(10)14348032
(01)00827002527317(17)261216(10)14359695
(01)00827002527348(17)261216(10)14354310
(01)00827002473393(17)261216(10)14356327
(01)00827002235205(17)261216(10)NS14379763
(01)00827002527362(17)270117(10)14401830
(01)00827002527348(17)270117(10)14414400
(01)00827002527324(17)270117(10)14413821
(01)00827002527317(17)270117(10)14420112
(01)00827002527386(17)270117(10)14407730
(01)00827002527362(17)270117(10)14409288
(01)00827002473348(17)270217(10)14501831
(01)00827002473324(17)270217(10)14539894
(01)00827002473423(17)270217(10)14438763
(01)00827002527348(17)270217(10)14516001
(01)00827002527379(17)270317(10)NS14571882
(01)00827002527362(17)270317(10)NS14574879
(01)00827002527362(17)270317(10)NS14577031
(01)00827002527362(17)270317(10)NS14577032
(01)00827002473393(17)270317(10)14602397
(01)00827002473317(17)270317(10)14542065
(01)00827002527317(17)270317(10)14545258
(01)00827002473379(17)270317(10)14545260
(01)00827002527393(17)270317(10)NS14589847
(01)00827002473331(17)261217(10)14345786
(01)00827002527416(17)261217(10)NS14411553
(01)00827002527317(17)261217(10)14360272
(01)00827002527324(17)261217(10)14363791
(01)00827002527362(17)261217(10)14368403
(01)00827002473331(17)270118(10)14395249
(01)00827002527331(17)270118(10)14404900
(01)00827002235212(17)270118(10)14413823
(01)00827002473324(17)270118(10)14411426
(01)00827002527317(17)270118(10)14426440
(01)00827002473386(17)270118(10)14426612
(01)00827002527317(17)270118(10)14427591
(01)00827002473379(17)270118(10)14427841
(01)00827002235212(17)270118(10)14434086
(01)00827002473331(17)270118(10)14433107
(01)00827002527317(17)270118(10)14408759
(01)00827002527331(17)270118(10)14418479
(01)00827002527386(17)270118(10)14433108
(01)00827002527355(17)270218(10)NS14531992
(01)00827002527348(17)270218(10)NS14533530
(01)00827002527324(17)270218(10)14467172
(01)00827002527317(17)270218(10)14473216
(01)00827002527348(17)270318(10)14539190
(01)00827002527331(17)270318(10)NS14572219
(01)00827002527331(17)270318(10)14539479
(01)00827002527386(17)270318(10)14545463
(01)00827002473317(17)270318(10)14545607
(01)00827002527317(17)270119(10)14411425
(01)00827002473324(17)270119(10)14427842
(01)00827002527317(17)270119(10)14432995
(01)00827002527324(17)270119(10)14404721
(01)00827002527317(17)270119(10)14431270
(01)00827002527317(17)270119(10)14432995
(01)00827002527355(17)261219(10)14357212
(01)00827002527317(17)261219(10)14352729
(01)00827002235212(17)261219(10)14345794
(01)00827002473423(17)261219(10)14356325
(01)00827002527348(17)261219(10)14356964
(01)00827002527355(17)261219(10)14357215
(01)00827002473447(17)270120(10)14434547
(01)00827002473416(17)270120(10)14436671
(01)00827002473386(17)270120(10)14434548
(01)00827002527317(17)270120(10)14434595
(01)00827002235212(17)270120(10)NS14461975
(01)00827002527348(17)261020(10)14262241
(01)00827002473386(17)261020(10)14264229
(01)00827002527393(17)261220(10)14360275
(01)00827002527362(17)261220(10)14362491
(01)00827002473348(17)261220(10)14423286
(01)00827002527331(17)261220(10)14357213
(01)00827002527324(17)261220(10)14360273
(01)00827002527331(17)261220(10)14360592
(01)00827002527362(17)261220(10)14362491
(01)00827002527348(17)270121(10)NS14440408
(01)00827002527355(17)270121(10)14427849
(01)00827002527379(17)270121(10)14436087
(01)00827002527379(17)270121(10)NS14459068
(01)00827002235212(17)270121(10)NS14461976
(01)00827002527362(17)270121(10)14432180
(01)00827002473355(17)270121(10)14436670
(01)00827002235212(17)270121(10)NS14464824
(01)00827002527362(17)270221(10)NS14531988
(01)00827002473423(17)270221(10)14470988
(01)00827002473423(17)270221(10)14470989
(01)00827002527393(17)270221(10)14476959
(01)00827002527393(17)270221(10)14482117
(01)00827002527379(17)270221(10)14469136
(01)00827002473331(17)270221(10)14485533
(01)00827002527348(17)270321(10)14605139
(01)00827002473423(17)270321(10)14551465
(01)00827002527331(17)270321(10)14552979
(01)00827002473317(17)270321(10)14553253
(01)00827002527355(17)270321(10)14555750
(01)00827002473317(17)270321(10)14558470
(01)00827002473423(17)260921(10)14150466
(01)00827002527393(17)260921(10)14150467
(01)00827002473423(17)260921(10)14150468
(01)00827002527393(17)260921(10)14150469
(01)00827002235212(17)260921(10)NS14165566
(01)00827002235212(17)260921(10)NS14165567
(01)00827002473348(17)260921(10)14146661
(01)00827002473386(17)260921(10)14147375
(01)00827002527386(17)260921(10)14148063
(01)00827002473416(17)260921(10)14148259
(01)00827002473324(17)260921(10)14149471
(01)00827002473317(17)260921(10)14145634
(01)00827002473447(17)261221(10)14359694
(01)00827002527393(17)261221(10)14360276
(01)00827002527348(17)261221(10)NS14372466
(01)00827002527379(17)261221(10)14368404
(01)00827002527348(17)261221(10)NS14401831
(01)00827002527331(17)261221(10)NS14411450
(01)00827002527393(17)261221(10)NS14411453
(01)00827002473423(17)261221(10)14360277
(01)00827002527348(17)261221(10)NS14381274
(01)00827002527331(17)270222(10)NS14528436
(01)00827002473317(17)270222(10)14394029
(01)00827002527317(17)270222(10)14488157
(01)00827002527317(17)270222(10)14485718
(01)00827002473317(17)270322(10)14555470
(01)00827002473317(17)270322(10)14555472
(01)00827002473386(17)260922(10)14147340
(01)00827002527317(17)260922(10)14148258
(01)00827002527348(17)260922(10)14149405
(01)00827002527317(17)260922(10)14156620
(01)00827002527317(17)260922(10)NS14172962
(01)00827002527324(17)260922(10)14146659
(01)00827002473379(17)260922(10)14147374
(01)00827002527379(17)260922(10)14149472
(01)00827002473409(17)260922(10)14149473
(01)00827002527362(17)260922(10)14156792
(01)00827002527317(17)260922(10)14144005
(01)00827002473317(17)260922(10)14148062
(01)00827002527317(17)260922(10)14149787
(01)00827002473317(17)260922(10)14161034
(01)00827002473331(17)260922(10)14163040
(01)00827002235212(17)260922(10)NS14177890
(01)00827002473416(17)261022(10)NS14227839
(01)00827002527386(17)261022(10)NS14231035
(01)00827002527379(17)261022(10)NS14239568
(01)00827002527393(17)261122(10)NS14285100
(01)00827002235212(17)261122(10)14267434
(01)00827002527362(17)261122(10)14290347
(01)00827002527362(17)270223(10)14542802
(01)00827002527317(17)270223(10)14485734
(01)00827002527362(17)270223(10)14488321
(01)00827002473324(17)270223(10)14488329
(01)00827002527317(17)270223(10)14478826
(01)00827002473317(17)260923(10)14160864
(01)00827002473423(17)260923(10)14173697
(01)00827002527324(17)260923(10)14177891
(01)00827002527379(17)260923(10)NS14177889
(01)00827002527393(17)261123(10)14295069
(01)00827002473423(17)261123(10)14295070
(01)00827002527393(17)261123(10)14295071
(01)00827002473423(17)261123(10)14295072
(01)00827002527355(17)261123(10)14299750
(01)00827002527324(17)261123(10)14290349
(01)00827002473317(17)261123(10)14293975
(01)00827002473317(17)261123(10)14293976
(01)00827002473348(17)261123(10)14296632
(01)00827002527348(17)270224(10)14501830
(01)00827002473324(17)270224(10)14539895
(01)00827002527324(17)270224(10)14487851
(01)00827002527317(17)270224(10)NS14528437
(01)00827002527362(17)270224(10)NS14528438
(01)00827002527379(17)270224(10)NS14530354
(01)00827002473317(17)270324(10)14394565
(01)00827002527331(17)260924(10)NS14167628
(01)00827002473447(17)260924(10)14163039
(01)00827002527317(17)260924(10)14165565
(01)00827002527324(17)260924(10)14167626
(01)00827002473379(17)260924(10)14168332
(01)00827002473348(17)260924(10)14170845
(01)00827002473355(17)260924(10)14170846
(01)00827002473447(17)260924(10)14172960
(01)00827002527355(17)260924(10)14170238
(01)00827002473317(17)260924(10)14170380
(01)00827002527317(17)260924(10)14172961
(01)00827002473423(17)260924(10)14181210
(01)00827002527317(17)261024(10)NS14222392
(01)00827002527379(17)261024(10)14223990
(01)00827002235205(17)261024(10)14235116
(01)00827002527317(17)261024(10)NS14222392
(01)00827002235205(17)261024(10)NS14267441
(01)00827002235205(17)261024(10)NS14267448
(01)00827002235205(17)261024(10)NS14267449
(01)00827002527393(17)261124(10)14295068
(01)00827002527379(17)261124(10)14317113
(01)00827002527355(17)261124(10)14323042
(01)00827002527362(17)261124(10)14301594
(01)00827002527331(17)261124(10)14296631
(01)00827002473447(17)270125(10)NS14438149
(01)00827002527355(17)270125(10)NS14440407
(01)00827002527379(17)270125(10)NS14440416
(01)00827002473348(17)270125(10)14438762
(01)00827002527331(17)270125(10)NS14458935
(01)00827002527348(17)270125(10)NS14458936
(01)00827002527348(17)270125(10)NS14458939
(01)00827002527386(17)270125(10)NS14461978
(01)00827002527324(17)270225(10)14541936
(01)00827002473317(17)270325(10)14368789
(01)00827002473317(17)270325(10)14369064
(01)00827002473317(17)270325(10)14354200
(01)00827002473317(17)270325(10)14566982
(01)00827002473317(17)270325(10)14431272
(01)00827002235205(17)261025(10)NS14267447
(01)00827002235205(17)261025(10)NS14267450
(01)00827002527393(17)261026(10)14223572
(01)00827002473423(17)261026(10)14223573
(01)00827002527393(17)261026(10)14224134
(01)00827002235212(17)261026(10)NS14244954
(01)00827002235212(17)261026(10)NS14267435
(01)00827002235212(17)261026(10)NS14267438
(01)00827002235205(17)261026(10)NS14267444
(01)00827002235205(17)261026(10)NS14267446
(01)00827002527317(17)261027(10)14222391
(01)00827002473317(17)261027(10)14223993
(01)00827002527317(17)261027(10)14224137
(01)00827002473324(17)261027(10)14227334
(01)00827002473317(17)261027(10)14227836
(01)00827002473386(17)261027(10)14227160
(01)00827002527362(17)261027(10)14229459
(01)00827002527393(17)261027(10)14223634
(01)00827002473423(17)261027(10)14230669
(01)00827002527393(17)261027(10)NS14231481
(01)00827002235212(17)261027(10)NS14242334
(01)00827002235212(17)261027(10)NS14242335
(01)00827002235212(17)261027(10)NS14242335X
(01)00827002527393(17)261027(10)NS14242963
(01)00827002235212(17)261027(10)NS14248732
(01)00827002235212(17)261027(10)NS14267440
(01)00827002235205(17)261027(10)NS14267442
(01)00827002235205(17)261027(10)NS14267443
(01)00827002235205(17)261027(10)NS14267445
(01)00827002527416(17)261027(10)NS14285107
(01)00827002527416(17)261027(10)NS14285108
(01)00827002527416(17)261027(10)NS14285109
(01)00827002527416(17)261027(10)NS14285112
(01)00827002473379(17)261227(10)14362932
(01)00827002527393(17)261227(10)NS14387091
(01)00827002527317(17)261227(10)NS14392160
(01)00827002235212(17)261227(10)NS14404723
(01)00827002527331(17)261227(10)NS14411427
(01)00827002527355(17)261227(10)NS14411429
(01)00827002527393(17)261227(10)NS14411433
(01)00827002527317(17)261227(10)NS14411442
(01)00827002527416(17)261227(10)NS14411549
(01)00827002527379(17)261227(10)NS14413822
(01)00827002527348(17)261227(10)NS14418478
(01)00827002527416(17)261227(10)NS14424354
(01)00827002235212(17)261227(10)NS14411431
(01)00827002527386(17)261227(10)NS14411432
(01)00827002235212(17)261227(10)NS14411441
(01)00827002235205(17)261227(10)NS14411449
(01)00827002527386(17)261227(10)NS14411452
(01)00827002527416(17)261227(10)NS14411550
(01)00827002527416(17)261227(10)NS14411555
(01)00827002527416(17)261227(10)NS14411556
(01)00827002527324(17)270128(10)14438518
(01)00827002473324(17)270128(10)14447010
(01)00827002527317(17)270228(10)NS14531979
(01)00827002473393(17)260928(10)14181872
(01)00827002527317(17)261028(10)14223992
(01)00827002473416(17)261028(10)14228060
(01)00827002473317(17)261028(10)14229489
(01)00827002527317(17)261028(10)14229490
(01)00827002527331(17)261028(10)14221706
(01)00827002527348(17)261028(10)14227838
(01)00827002527348(17)261028(10)14228355
(01)00827002527386(17)261028(10)14221707
(01)00827002527386(17)261028(10)14227840
(01)00827002527324(17)261028(10)14228356
(01)00827002527386(17)261028(10)14228358
(01)00827002527355(17)261028(10)14231479
(01)00827002473393(17)261028(10)14271104
(01)00827002527379(17)261028(10)14271105
(01)00827002527379(17)261028(10)14276627
(01)00827002473386(17)261028(10)14262242
(01)00827002527393(17)261028(10)NS14239566
(01)00827002527416(17)261028(10)NS14285106
(01)00827002527416(17)261028(10)NS14285110
(01)00827002527416(17)261028(10)NS14285111
(01)00827002527362(17)261028(10)14223989
(01)00827002527324(17)261028(10)14228356
(01)00827002527379(17)261028(10)14231032
(01)00827002527324(17)261228(10)14372468
(01)00827002473416(17)261228(10)14373505
(01)00827002527379(17)261228(10)NS14404726
(01)00827002527331(17)261228(10)NS14411440
(01)00827002527324(17)261228(10)NS14411455
(01)00827002527355(17)261228(10)NS14411456
(01)00827002527379(17)261228(10)NS14411457
(01)00827002527416(17)261228(10)NS14411551
(01)00827002527317(17)261228(10)14368630
(01)00827002527379(17)261228(10)14372645
(01)00827002527317(17)261228(10)14373511
(01)00827002527317(17)261228(10)14373557
(01)00827002473355(17)261228(10)14373933
(01)00827002235205(17)261228(10)14379764
(01)00827002527317(17)261228(10)14381275
(01)00827002527386(17)261228(10)14382644
(01)00827002235205(17)261228(10)14383707
(01)00827002235205(17)261228(10)14383708
(01)00827002527362(17)261228(10)14411439
(01)00827002527362(17)261228(10)NS14401833
(01)00827002527355(17)261228(10)NS14404725
(01)00827002235212(17)261228(10)NS14404727
(01)00827002235205(17)261228(10)NS14411446
(01)00827002235205(17)261228(10)NS14411447
(01)00827002235205(17)261228(10)NS14411448
(01)00827002527317(17)261228(10)14387108
(01)00827002527379(17)261228(10)14387110
(01)00827002527324(17)261228(10)NS14401832
(01)00827002527317(17)261228(10)NS14375880
(01)00827002473379(17)261029(10)14228059
(01)00827002527355(17)261029(10)14229380
(01)00827002527324(17)261029(10)14230665
(01)00827002473348(17)261029(10)14231053
(01)00827002473386(17)261029(10)14232840
(01)00827002473393(17)261029(10)14233391
(01)00827002473409(17)261029(10)14233394
(01)00827002527379(17)261029(10)14236754
(01)00827002527355(17)261029(10)NS14274105
(01)00827002473324(17)261029(10)14231478
(01)00827002473317(17)261029(10)14233389
(01)00827002527362(17)261029(10)14242962
(01)00827002527331(17)261029(10)NS14239567
(01)00827002527331(17)261029(10)NS14242959
(01)00827002527331(17)261029(10)NS14244952
(01)00827002527355(17)261029(10)NS14267437
(01)00827002527355(17)261029(10)NS14267437X
(01)00827002473324(17)261129(10)14293977
(01)00827002527324(17)261129(10)14303290
(01)00827002527317(17)261129(10)14303289
(01)00827002527331(17)261129(10)14340477
(01)00827002473379(17)261129(10)14340479
(01)00827002527317(17)261229(10)14374504
(01)00827002527355(17)261229(10)14373512
(01)00827002473393(17)261229(10)14374528
(01)00827002473409(17)261229(10)14374529
(01)00827002527324(17)261229(10)14387109
(01)00827002235205(17)261229(10)14401671
(01)00827002473331(17)260930(10)14146660
(01)00827002527362(17)261130(10)14301593
(01)00827002527317(17)261130(10)14340478
(01)00827002527348(17)261130(10)NS14323044
(01)00827002527348(17)261130(10)NS14324153
(01)00827002527416(17)261230(10)14411529
(01)00827002527355(17)261230(10)NS14372467
(01)00827002527355(17)261230(10)14411459
(01)00827002527416(17)261230(10)14411528
(01)00827002527416(17)261230(10)14411531
(01)00827002527416(17)261230(10)14411532
(01)00827002527416(17)261230(10)14411534
(01)00827002473348(17)261230(10)14360274
(01)00827002527324(17)261230(10)14382641
(01)00827002527379(17)261230(10)14411444
(01)00827002527355(17)261230(10)14387089
(01)00827002527393(17)261230(10)NS14404728
(01)00827002527317(17)270131(10)14440429
(01)00827002527324(17)270131(10)14442188
(01)00827002527355(17)270131(10)NS14458937
(01)00827002527317(17)270131(10)14441794
(01)00827002527386(17)270131(10)14440414
(01)00827002527324(17)270131(10)14441795
(01)00827002473317(17)270131(10)14438151
|
| FEI Number |
1820334
|
Recalling Firm/ Manufacturer |
Cook Incorporated 750 N Daniels Way Bloomington IN 47404-9120
|
| For Additional Information Contact | Cook Medical Customer Relations Department 800-457-4500 |
Manufacturer Reason for Recall | The loading cartridges included with these devices may contain a small, unintended, stainless-steel cannula. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | On May 5, 2022, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were advised that the presence of a small stainless steel cannula inside the loading cartridge may go undetected by the user.
Customers were instructed to examine their inventory and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. Customers should return the product to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Please share this information with appropriate personnel, down to the user level, within your organization or with any organization where the affected devices may have been transferred.
Physicians should practice their standard patient monitoring following the procedure for an early identification of any complications to mitigate their severity.
If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235. |
| Quantity in Commerce | 106,033 |
| Distribution | Domestic distribution nationwide. Foreign distribution worldwide. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = KRD
|