| | Class 1 Device Recall GE |  |
| Date Initiated by Firm | December 19, 2022 |
| Date Posted | February 08, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1054-2023 |
| Recall Event ID |
91456 |
| 510(K)Number | K111445 |
| Product Classification |
System, tomography, computed, emission - Product Code KPS
|
| Product | GE Nuclear Medicine 600/800 series systems, specifically:
a) 830,
b) 830 / Dismount,
c) 830 /,
d) 830 / 5842894-20,
e) 830 / Dismount / 50HZ,
f) 830 / Dismount / 60HZ,
g) 830 3/8 inch Detector,
h) 830 BJ,
i) 830 BJ / 5144407,
j) 830 BJ / Dismount,
k) 830 ES DEM,
l) 830 NM Cyber Defense Configuration Gantry 3/8 inch,
m) 830 NM Gantry 3/8 inch,
n) 830 NM Gantry 5/8 inch,
o) 870 DR NM GANTRY FINAL - NEW ROTOR,
p) B615 to D630 UPG,
q) D630 NM GANTRY,
r) D630 NM GANTRY 3/8 inch WITH NEW ROTOR - MEA,
s) D630 NM Gantry 3/8 inch,
t) D630 NM GANTRY FINAL ASSEMBLY_,
u) D630 with new rotor,
v) D630 with new rotor/ 5144407 / Dismount,
w) D630 with new rotor/ Dismount,
x) DISCOVERY D630 X 3/8 inch,
y) DISCOVERY D630 X 3/8 inch,
z) DISCOVERY D630 X 3/8 inch,
aa) DISCOVERY D630 x 3/8 inch BHW,
bb) DISCOVERY D630 X 3/8 inch BJG,
cc) DISCOVERY D630 X 3/8 inch MEA,
dd) DISCOVERY NM 830 3/8 inch,
ee) DISCOVERY NM 830 3/8 inch,
ff) NM 830 3/8 inch,
gg) NM 830 3/8 inch BJ,
hh) NM 830 ES,
ii) NM DISCOVERY D630 3/8 inch,
jj) NUC GOLDSEAL D630 X 3/8 inch.
Emission Computed Tomography System |
| Code Information |
a) 830: Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3AG2200003, NT3AG2000015BH, NT3AG2100015BH, NT3AG2200006BH;
b) 830 / Dismount: Catalog number To be provided, UDI/DI Not Applicable, Serial numbers: NT3AG2100016BH;
c) 830 /: Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3AG2100009BH;
d) 830 / 5842894-20: Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3AG2200013;
e) 830 / Dismount / 50HZ: Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3AG2000017BH;
f) 830 / Dismount / 60HZ: Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3AG2100001;
g) 830 3/8 inch Detector: Catalog number H3910AC, UDI/DI Not Applicable, Serial Numbers: 830Y60105;
h) 830 BJ: Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3AG1900009BH; Catalog number 5376204-32-1, UDI/DI Not Applicable, Serial Numbers: NT3AG1900015BH;
i) 830 BJ / 5144407: Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3AG1900008BH;
j) 830 BJ / Dismount: Catalog number 5376204-32-1, UDI/DI To be provided, Serial Numbers: NT3AG1900016BH;
k) 830 ES DEM: Catalog number H3910AC, UDI/DI Not Applicable, Serial Numbers: 83EZ88002, 830Z60256;
l) 830 NM Cyber Defense Configuration Gantry 3/8 inch: Catalog number H3910AG, UDI/DI Not Applicable, Serial Numbers: 830Z60185, 830Z60224, 830Z60225;
m) 830 NM Gantry 3/8 inch: Catalog number H3910AC, UDI/DI Not Applicable, Serial Numbers: 830Y60050, 830Z60199, 830A60284, 830A60290, 830X60001, 830X60004, 830Y60056, 830Y60069, 830Z60158, 830Z60182, 83EA88010, 830Z60183, 830A60282, 830Z60192, 830A60271, 830Y60122, 830A60298, 830A60289, 830Y60048, 830Y60036, 830Z60239, 830A60294, 830Y60046, 830X60012, 830Y60054, 830Z60269, 830X60018, 830A60283, 830Y60089, 830Y60096, 830Z60165, 830Z60169, 830Y60051, 830Z60244, 830A60288, 830X60026, 830Y60030, 830Z60197, 830Z60227, 830A60277, 830A60334, 830Y60079, 830Y60131, 830Y60141, 83EA88008, 830Z60238, 830Z60207, 830Z60248, 830Y60137, 830Z60149, 830Z60175, 830Y60076, 830Y60092, 830Y60090, 830A60301, 830A60279, 830Z60178, 830Z60153, 830Y60132, 830X60011, 830X60016, 830Y60102, 830Z60253, 830Z60264, 830Y60049, 830Z60265, 830Z60254, 830Y60081, 830Z60263, 830Z60200, 830A60308, 830A60300, 830Z60171, 830Y60057, 830Y60029, 830Y60106, 830Y60059, 830Z60177, 830A60291, 830Z60267, 830Z60174, 830Z60245, 830Y60074, 830Z60167, 830Y60071, 830Z60246, 830Z60223, 830Z60201, 830Y60045, 830X60019, 83EZ88001, 830Y60124, 830A60306, 830A60297, 830Z60213, 830Y60111, 830Y60127, 830Y60094, 830Z60258, 830Y60058, 830Z60218, 830Y60138, 830A60280, 830A60274, 830Z60211, 830Y60101, 830Z60210, 830Z60260, 830Z60234, 830Y60118, 830X60027, 830Z60228, 830Z60168, 830Z60214, 830Y60044, 830Y60053, 830Z60173, 830Z60195, 830Y60123, 830Y60117, 830Y60129, 830Z60148, 830X60023, 830X60006, 830Y60104, 830X60015, 830Y60121, 830A60281, 830Y60133, 830Y60098, 830Z60268, 830Z60194, 830Z60198, 830X60014, 830Y60040, 830X60003, 830Z60159, 830Z60237, 830Y60060, 830A60278, 83EA88006, 830Z60222, 830Z60250, 830Z60181, 830Y60134, 830Z60152, 830Z60202, 830Z60220, 830Z60230, 830Y60062, 830Y60097, 830Z60176, 830Z60196, 830Y60125, 830X60025, 830A60293, 830A60307, 830Z60249, 830Z60180, 830Y60064, 830Y60078, 830Y60115, 830Y60114, 830Y60043, 830Y60108, 830A60275, 830Y60130, 830Z60229, 830A60292, 830Z60163, 830Y60128, 830X60021, 830Z60266, 830Y60100, 830A60304, 830Y60099, 830Y60034, 830Y60087, 830X60028, 830Y60084, 830Y60055, 830Y60136, 830Z60188, 830Z60206, 830Z60151, 830X60017, 830Y60140, 83EA88009, 830Z60154, 830Y60112, 830A60276, 830Y60082, 830Y60093, 830A60305, 830Z60232, 830Y60066, 830Z60191, 830Z60236, 830A60285, 830Z60235, 830Z60157, 830Y60126, 830Y60142, 830Y60035, 83EA88007, 830A60286, 830Y60107, 830Y60080, 830X60002, 830Z60270, 830Y60067, 830Z60193, 830Z60172, 830X60022, 830Y60095, 830Y60119, 830Y60042, 830Z60156, 830X60024, 830B60457, 830Y60047, 830Z60205, 830X60013, 830Y60073, 830Y60032, 830A60310, 830Z60219, NT3AG1900017BH, 830Z60257, 830Z60247, 830Z60240, 830Z60203, 830Y60085, 830Y60077, 830Z60255, 830A60311, 830Y60033, 830X60020, 830A60287, 830Y60038, 83EZ88004, 83EZ88005, 830X60010, 830Z60252, 830Z60166, 830Y60061, 830Y60091, 830Z60184, 830A60309, 830Z60217, 830Z60155, 830X60005, 830X60008, 830X60009, 830Y60052, 830Y60063, 830Y60143, 830Y60144, 830Y60146, 830Z60231, 830Z60187, 830A60296, 830Y60041, 830Y60088, 830Z60226, 830Z60262, 830X60007, 830Z60259, 830Z60212, 830Y60065, 830Y60135, 830Z60204, 830Y60072, 830Z60160; UDI/DI UDI_Not_Required_2, Serial Numbers: 830Y60113; , UDI/DI UDI_Not_Required_6, Serial Numbers: 830A60299; UDI/DI 010084068214082921830A6032811210700, Serial Numbers: 830A60328; UDI/DI 010084068214082921830Z6015011200100, Serial Numbers: 830Z60150; UDI/DI 010084068214082921830A6027211210100, Serial Numbers: 830A60272; UDI/DI 010084068214082921830B6042411220500, Serial Numbers: 830B60424; UDI/DI 010084068214082921830Z6019011200400, Serial Numbers: 830Z60190; UDI/DI 010084068214082921830A6027311210100, Serial Numbers: 830A60273; UDI/DI 010084068214082921830Z6018911200400, Serial Numbers: 830Z60189; UDI/DI 010084068214082921830Y6007511190600, Serial Numbers: 830Y60075; UDI/DI (01)00840682140829(21)830B60400(11)220200, Serial Numbers: 830B60400; UDI/DI 010084068214082921830Z6017011200200, Serial Numbers: 830Z60170; UDI/DI (01)00840682140829(21)830A60346(11)211100, Serial Numbers: 830A60346; UDI/DI 010084068214082921830Y6010311190900, Serial Numbers: 830Y60103; UDI/DI 010084068214082921830Y6014511191200, Serial Numbers: 830Y60145; UDI/DI 010084068214082921830A6030311210500, Serial Numbers: 830A60303; UDI/DI (01)00840682140829(21)830A60345(11)211000, Serial Numbers: 830A60345; UDI/DI 010084068214082921830A6029511210300, Serial Numbers: 830A60295; UDI/DI 010084068214082921830Z6023311200900, Serial Numbers: 830Z60233; UDI/DI 010084068214082921830Z6020911200600, Serial Numbers: 830Z60209; UDI/DI (01)00840682140829(21)830B60386(11)220100, Serial Numbers: 830B60386; UDI/DI (01)00840682140829(21)830B60434(11)220600, Serial Numbers: 830B60434; UDI/DI 010084068214082921830Y6010911191000, Serial Numbers: 830Y60109; UDI/DI 010084068214082921830Z6016211200200, Serial Numbers: 830Z60162; UDI/DI 010084068214082921830Z6017911200300, Serial Numbers: 830Z60179; UDI/DI 010084068214082921830Z6018611200300, Serial Numbers: 830Z60186; UDI/DI 010084068214082921830B6042711220500, Serial Numbers: 830B60427; UDI/DI 010084068214082921830A6037811211200, Serial Numbers: 830A60378; , UDI/DI To be provided, Serial Numbers: 830B60383, 830A60337, 830A60373, 830A60368, 830B60387, 830B60458, 830B60459, 830B60460, 830B60462, 830B60464, 830B60465, 830B60466, 830B60421, 830B60430, 830B60443, 830B60456, 830B60473, 830B60384, 83EB88026, 83EA88017, 830A60317, 830B60401, 830B60432, 830B60422, 830B60467, 830B60471, 830B60398, 830A60364, 83EB88027, 83EB88028, 830A60326, 830B60411, 830B60469, 830B60472, 830B60389, 830A60333, 830A60347, 830A60348, 830B60447, 830A60313, 830A60375, 830A60374, 830B60412, 830A60339, 830B60461, 830B60463, 830B60470, 830A60365, 830B60446, 830B60397, 83EB88034, 83EB88025, 83EA88011, 830B60431, 830B60428, 830B60416, 830B60441, 830A60363;
n) 830 NM Gantry 5/8 inch: Catalog number H3910AD, UDI/DI To be provided, Serial Numbers: 830B60468, 830B60379;
o) 870 DR NM GANTRY FINAL - NEW ROTOR: Catalog number 5376204-32-1, UDI/DI Not Applicable, Serial Numbers: NT3AG2000009BH;
p) B615 to D630 UPG: Catalog number To be provided, UDI/DI To be provided, Serial Numbers: NT2UG2100001; Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT2UG2100002;
q) D630 NM GANTRY: Catalog number H3100JY, UDI/DI 01008406821392742163GX5002611180700, Serial Numbers: 63GX50026; Catalog number H3100JY, UDI/DI Not Applicable, Serial Numbers: 0000063GX50023;
r) D630 NM GANTRY 3/8 inch WITH NEW ROTOR - MEA: Catalog number 5376204-31-1, UDI/DI To be provided, Serial Numbers: NT2PG2000001BH;
s) D630 NM Gantry 3/8 inch: Catalog number H3101RH, UDI/DI Not Applicable, Serial Numbers: 63GY50074, 63GX50024, 63GZ50134, 63GY50110, 63GY50119, 63GY50067, 63GY50123, 63GX50001, 63GX50037, 63GX50060, 63GY50089, 63GY50087, 63GY50098, 63GY50097, 63GX50034, 63GZ50152, 63GY50094, 63GY50128, 63GY50131, 63GX50014, 63GY50079, 63GY50130, 63GY50101, 63GZ50147, 63GZ50136, 63GX50019, 63GZ50150, 63GZ50141, 63GY50078, 63GX50022, 63GX50021, 63GX50031, 63GX50018, 63GY50082, 63GX50009, 63GY50080, 63GY50118, 63GZ50133, 63GY50077, 63GX50043, 63GY50116, 63GX50020, 63GX50054, 63GX50008, 63GX50035, 63GX50050, 63GX50061, 63GX50062, 63GY50085, 63GZ50151, 63GY50125, 63GY50102, 63GX50032, 63GX50055, 63GY50090, 63GX50013, 63GX50047, 63GX50002, 63GX50025, 63GX50033, 63GX50038, 63GY50115, 63GX50044, 63GY50129, 63GY50120, 63GY50068, 63GX50063, 63GY50104, 63GY50069, 63GX50016, 63GY50075, 63GX50017, 63GX50005, 63GY50084, 63GX50029, 63GZ50135, 63GX50012, 63GY50127, 63GY50124, 63GX50046, 63GZ50148, 63GX50039, 63GX50049, 63GY50083, 63GX50011, 63GX50030, 63GY50108, 63GY50103, 63GY50113, 63GY50114, 63GZ50146, 63GX50057, 63GZ50145, 63GA50153, 63GZ50149, 63GY50100, 63GZ50142, 63GX50045, 63GA50155, 63GA50154, 63GY50111, 63GY50112, 0000063GX50048, 63GX50040, 63GX50003, 63GX50065, 63GZ50137, 63GY50107, 63GZ50138, 63GX50042, 63GY50088, 63GY50126, 63GX50053, 63GX50004, 63GY50086, 63GZ50132, 63GX50056, 63GX50066, 63GX50059;UDI/DI 01008406821392742163GY5007611190200, Serial Numbers: 63GY50076; UDI/DI 01008406821392742163GX5005111181000, Serial Numbers: 63GX50051; UDI/DI 01008406821392742163GX5005211181000, Serial Numbers: 63GX50052; UDI/DI 01008406821392742163GY5007011190100, Serial Numbers: 63GY50070; UDI/DI 01008406821392742163GY5009511190400, Serial Numbers: 63GY50095; UDI/DI 01008406821392742163GY5012111191100, Serial Numbers: 63GY50121; UDI/DI 01008406821392742163GX5002711180700, Serial Numbers: 63GX50027; UDI/DI 01008406821392742163GX5003611180700, Serial Numbers: 0000063GX50036; UDI/DI 01008406821392742163GY5009611190400, Serial Numbers: 63GY50096; UDI/DI 01008406821392742163GX5000711180500, Serial Numbers: 63GX50007; UDI/DI 01008406821392742163GX5000611180500, Serial Numbers: 63GX50006; UDI/DI 01008406821392742163GX5001011180600, Serial Numbers: 63GX50010; UDI/DI 01008406821392742163GX5001511180600, Serial Numbers: 63GX50015; UDI/DI 01008406821392742163GY5011711190900, Serial Numbers: 63GY50117; UDI/DI 01008406821392742163GY5007111190100, Serial Numbers: 63GY50071; UDI/DI 01008406821392742163GY5007211190100, Serial Numbers: 63GY50072; UDI/DI (01)00840682139274(21)63GY50106(11)190700, Serial Numbers: 63GY50106; UDI/DI 01008406821392742163GY5007311190100, Serial Numbers: 63GY50073; UDI/DI To be provided, Serial Numbers: 63GX50041, 63GY50105, 63GA50161, 63GB50165, 63GA50162, 63GA50157, 63GA50159, 63GA50160, 63GA50156, 63GA50163, 63GB50166;
t) D630 NM GANTRY FINAL ASSEMBLY, : Catalog number 5791570, UDI/DI To be provided, Serial Numbers: 830B60423, 830B60429;
u) D630 with new rotor: Catalog number H2405AC, UDI/DI To be provided, Serial Numbers: NT2PG1900006, NT2PG1900005;
v) D630 with new rotor/ 5144407 / Dismount: Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT2PG1900007BH;
w) D630 with new rotor/ Dismount: Catalog number 5376204-31-1, UDI/DI Not Applicable, Serial Numbers: NT2PG1800001BH;
x) DISCOVERY D630 X3/8 inch: Catalog number 5376204-31-1, UDI/DI Not Applicable, Serial Numbers: NT2PG1900009BH, NT2PG1900008BH;
y) DISCOVERY D630X 3/8 inch: Catalog number 5791570, UDI/DI Not Applicable, Serial Numbers: 0000063GZ50139;
z) DISCOVERY D630X3/8 inch: Catalog number H3101RH, UDI/DI 01008406821392742163GY5009111190400, Serial Numbers: 63GY50091; UDI/DI Not Applicable, Serial Numbers: 63GY50093, 63GY50099, 63GZ50144, 63GZ50140, 63GY50122, 63GY50092, 63GB50164, 63GX50064, 63GX50028, 63GX50058, 63GY50081, 63GY50109, 63GZ50143;
aa) DISCOVERY D630x3/8 inch BHW: Catalog number H2405AC, UDI/DI 01008406821393281119060021NT2PG1900010BH, Serial Numbers: NT2PG1900010BH;
bb) DISCOVERY D630X3/8 inch BJG: Catalog number 5376204-31-1, UDI/DI Not Applicable, Serial Numbers: NT2PG1900004BH, NT2PG1900001BH, NT2PG1900002BH, NT2PG1900012BH;
cc) DISCOVERY D630X3/8 inch MEA: Catalog number 5376204-31-1, UDI/DI 01008406821393281119100021NT2PG1900011BH, Serial Numbers: NT2PG1900011BH;
dd) DISCOVERY NM 830 3/8 inch: Catalog number 5376204-32-1, UDI/DI Not Applicable, Serial Numbers: NT3AG2200012BH; UDI/DI To be provided, Serial Numbers: NT3AG2100004BH, NT3AG2200007BH; UDI/DI Not Applicable, Serial Numbers: NT3AG2200012BH;
ee) DISCOVERY NM 830 3/8 inch: Catalog number H3910AC, UDI/DI Not Applicable, Serial Numbers: 830Y60110, 830Z60164;
ff) NM 830 3/8 inch: Catalog number 5791570, UDI/DI To be provided, Serial Numbers: 830A60336, 830Y60031; Catalog number 5376204-31-1, UDI/DI To be provided, Serial Numbers: NT3AG2000005BH; Catalog number 5376204-32, UDI/DI 010084068214082921830A6035511211100, Serial Numbers: 830A60355; UDI/DI 010084068214082921830A6032511210700, Serial Numbers: 830A60325; UDI/DI 010084068214082921830A6037111211200, Serial Numbers: 830A60371; UDI/DI 010084068214082921830B6040911220300, Serial Numbers: 830B60409; UDI/DI 010084068214082921830B6043911220700, Serial Numbers: 830B60439; UDI/DI 010084068214082921830B6041411220400, Serial Numbers: 830B60414; UDI/DI 010084068214082921830A6032411210700, Serial Numbers: 830A60324; UDI/DI 010084068214082921830B6041511220400, Serial Numbers: 830B60415; UDI/DI 010084068214082921830B6043311220600, Serial Numbers: 830B60433; UDI/DI 010084068214082921830A6032911210800, Serial Numbers: 830A60329; UDI/DI 010084068214082921830B6044211220700, Serial Numbers: 830B60442; UDI/DI 01008406821408292183EB8803111220700, Serial Numbers: 83EB88031; UDI/DI 010084068214082921830B6044411220700, Serial Numbers: 830B60444; UDI/DI 010084068214082921830A6034411211000, Serial Numbers: 830A60344; UDI/DI 010084068214082921830B6040411220300, Serial Numbers: 830B60404; UDI/DI 010084068214082921830A6033211210800, Serial Numbers: 830A60332; UDI/DI 010084068214082921830A6032311210700, Serial Numbers: 830A60323; UDI/DI 010084068214082921830B6040211220200, Serial Numbers: 830B60402; UDI/DI 010084068214082921830A6035211211100, Serial Numbers: 830A60352; UDI/DI 010084068214082921830B6044811221000, Serial Numbers: 830B60448; UDI/DI 010084068214082921830A6031511210600, Serial Numbers: 830A60315; UDI/DI 010084068214082921830A6031811210600, Serial Numbers: 830A60318; UDI/DI 010084068214082921830B6042511220500, Serial Numbers: 830B60425; UDI/DI 010084068214082921830A6031911210600, Serial Numbers: 830A60319; UDI/DI 010084068214082921830A6036611211200, Serial Numbers: 830A60366; Catalog number H3910AC, UDI/DI Not Applicable, Serial Numbers: 830Y60147, 830B60455, 830Z60261, 830Z60221, 830Z60215, 830Y60070, 830Y60083, 830Y60037, 830Z60161, 830Y60039, 830A60302, 830Z60216, 830Z60242, 830Z60241, 830Z60251, 830Y60086, 830Z60243, 830Z60208; UDI/DI UDI_Not_Required_2, Serial Numbers: 830Y60116; UDI/DI To be provided, Serial Numbers: 830A60356, 830B60449, 830A60350, 83EB88032, 830A60322, 830B60445, 830B60435, 830A60358, 830B60408, 830A60349, 830B60453, 830A60357, 830A60335, 830B60410, 830B60395, 830A60354, 830B60385, 830A60331, 830B60396, 830A60360, 830B60381, 830B60406, 830B60403, 830A60320, 830A60321, 830B60426, 830B60388, 830A60369, 830B60399, 830B60419, 830A60377, 830A60316, 830B60382, 830B60452, 830A60370, 830A60376, 830B60413, 830A60314, 830A60367, 830B60394, 830B60436, 830A60342, 830A60362, 830B60440, 830B60407, 830A60312, 830B60380, 830B60418, 830A60353, 830A60330, 830B60390, 830A60372, 830A60359, 830A60341, 830A60338, 830A60327, 830B60405, 830A60361, 830B60454, 830A60340, 830A60351; Catalog number To be provided, UDI/DI To be provided, Serial Numbers: 830B60437, 830B60393, 830B60420, 830B60438, 830A60343;
gg) NM 830 3/8 inch BJ: Catalog number 5376204-32-1, UDI/DI Not Applicable, Serial Numbers: NT3AG2200001BH, NT3AG2100007BH, NT3AG2100003BH, NT3AG2000008BH, NT3AG2000002BH, NT3AG2200010BH, NT3AG1900002BH , NT3AG2200009BH, NT3AG1900014BH, NT3AG1900012BH, NT3AG2000006BH, NT3AG2100008BH, NT3AG2000013BH, NT3AG2100010BH, NT3AG2200005BH, NT3AG1900011BH, NT3AG2000016BH, NT3AG2100002BH, NT3AG2200011BH, NT3AG2100011BH, NT3AG1900018BH, NT3AG2000001BH, NT3AG1900003BH, NT3AG1900001BH, NT3AG1800001BH, NT3AG2100014BH, NT3AG1900010BH, NT3AG2000003BH, NT3AG2000012BH, NT3AG1900006BH, NT3AG2100005BH, NT3AG2000011BH, NT3AG2200008BH, NT3AG1800002BH, NT3AG2000010BH; UDI/DI To be provided, Serial Numbers: NT3AG2200014BH, NT3AG1900013BH, NT3AG2200002BH, NT3AG2100012BH, NT3AG2100013BH; Catalog number H2406AA, UDI/DI Not Applicable, Serial Numbers: NT3DG1800003BH, NT3AG1900004BH, NT3AG2200016, NT3AG2200015BH; Catalog number H3910AC, UDI/DI Not Applicable, Serial Numbers: 83EZ88003; Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3AG2200004BH; Catalog number To be provided, UDI/DI To be provided, Serial Numbers: NT3AG1900005BH, NT3AG2000004BH;
hh) NM 830 ES: Catalog number 5376204-32, UDI/DI 01008406821408292183EA8801811211000, Serial Numbers: 83EA88018; UDI/DI 01008406821408292183EA8801411210800, Serial Numbers: 83EA88014; UDI/DI 01008406821408292183EB8801911220100, Serial Numbers: 83EB88019; UDI/DI 01008406821408292183EB8802311220200, Serial Numbers: 83EB88023; UDI/DI 01008406821408292183EB8803011220500, Serial Numbers: 83EB88030; UDI/DI 01008406821408292183EA8801211210600, Serial Numbers: 83EA88012; UDI/DI 01008406821408292183EA8801511210900, Serial Numbers: 83EA88015; UDI/DI 01008406821408292183EA8801311210800, Serial Numbers: 83EA88013; UDI/DI 01008406821408292183EB8802411220200, Serial Numbers: 83EB88024; Catalog number H3910AC, UDI/DI To be provided, Serial Numbers: 83EB88029, 83EB88022, 83EA88016; UDI/DI Not Applicable, Serial Numbers: 83EB88021; Catalog number To be provided, UDI/DI To be provided, Serial Numbers: 83EB88020;
ii) NM DISCOVERY D630 3/8 inch: Catalog number 5474054, UDI/DI To be provided, Serial Numbers: 22404;
jj) NUC GOLDSEAL D630X3/8 inch: Catalog number S9630NU, UDI/DI Not Applicable, Serial Numbers: NT2QG1900003BH.
|
Recalling Firm/ Manufacturer |
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING 4, Hayozma St Tirat Carmel Israel
|
| For Additional Information Contact | GE HealthCare Service 800-4371171 |
Manufacturer Reason for Recall | A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury. |
FDA Determined Cause 2 | Process control |
| Action | GE Healthcare issue an URGENT MEDICAL DEVICE CORRECTION on 12/19/2022 by mail. The notice explained the issue and the risk, and requested the following action by taken:
.GE Healthcare asks that you STOP USE of your Nuclear Medicine system, if your
system is listed in the Affected Product Details and Appendix of this letter, until an
inspection of your system is completed by GE Healthcare and you have been
provided with a letter indicating the system can be returned to clinical use.
.GE Healthcare will correct all affected products at no cost to you. GE Healthcare
will inspect your system to ensure that the appropriate mitigation is correctly
implemented to address this potential life-threatening concern. A GE Healthcare
representative will contact you to arrange for this correction.
.If you have any questions or concerns regarding this notification, please contact
GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
|
| Quantity in Commerce | 748 units |
| Distribution | Worldwide distribution. US nationwide including Puerto Rico, Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Ethiopia, Finland, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Libya, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Uzbekistan, and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = KPS
|
|
|
|