| | Class 2 Device Recall Grafton Gel |  |
| Date Initiated by Firm | February 08, 2023 |
| Date Posted | April 04, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1333-2023 |
| Recall Event ID |
91697 |
| 510(K)Number | K051195 |
| Product Classification |
Filler, bone void, osteoinduction (w/o human growth factor) - Product Code MBP
|
| Product | Medtronic Grafton Gel:
a) DBM S41120 GRAFTON GEL 1CC, REF S41120;
b) DBM S41150 GRAFTON GEL 10CC, REF S41150;
c) DBM T41110 0.5CC GRAFTON GEL, REF T41110;
d) DBM T41110INT 0.5CC GRAFTON GEL, REF T41110INT;
e) DBM T41120 1CC GRAFTON GEL, REF T41120;
f) DBM T41120INT 1CC GRAFTON GEL, REF T41120INT;
g) DBM T41130 5CC GRAFTON GEL, REF T41130;
h) DBM T41130INT 5CC GRAFTON GEL, REF T41130INT;
i) DBM T41150 10CC GRAFTON GEL, REF T41150;
j) DBM T41150INT 10CC GRAFTON GEL, REF T41150INT;
Demineralized Bone Matrix (DBM) product |
| Code Information |
a) DBM S41120 GRAFTON GEL 1CC, REF S41120, GTIN 00613994961167, Serial Numbers: A41879
b) DBM S41150 GRAFTON GEL 10CC, REF S41150, GTIN 00613994961143, Serial Numbers: A43939, A47225
c) DBM T41110 0.5CC GRAFTON GEL, REF T41110, GTIN 00643169122956, Serial Numbers: A38528-100, A38528-101, A38528-102, A38528-103, A38528-104, A38528-105, A38528-106, A38910-058, A38910-059, A38910-060, A38910-061, A38910-062, A41224-053, A41224-054, A41224-055, A41224-056, A41224-057, A41224-058, A41224-059, A41224-060, A41224-061, A41224-062, A41224-063, A41224-064, A41224-065, A41224-066, A41224-067, A41224-068, A41224-069, A41224-070, A41224-071, A41224-072, A41935-025, A41935-026, A41935-027, A41935-028, A41935-029, A41935-030, A41935-031, A41935-032, A41935-033, A41935-034, A41935-035, A41935-036, A41935-037, A41935-038, A41935-039, A41935-040, A41935-041, A41935-042, A41935-043, A41935-044, A41935-045, A41935-046, A41935-047, A41935-048, A41935-049, A41935-050, A41935-051, A41935-052, A41935-053, A42221-076, A42221-077, A42221-078, A42221-079, A42221-080, A42221-081, A42221-082, A42221-083, A42221-084, A42221-085, A42619-011, A42619-012, A42619-013, A42619-014, A42619-015, A42619-016, A42619-017, A42619-018, A42619-019, A42619-020, A42619-021, A42619-022, A42619-023, A42619-024, A42619-025, A42619-026, A42619-027, A42619-028, A42619-029, A42619-030, A42619-031, A42619-032, A42619-033, A42619-034, A42619-035, A42619-036, A43231-127, A43231-128, A43286-011, A43286-012, A43286-013, A43286-014, A43286-015, A43286-016, A43286-017, A43286-018, A43286-019, A43286-020, A43286-021, A43286-022, A43286-023, A43286-024, A43286-025, A43286-026, A43286-027, A43286-028, A43286-029, A43286-030, A43286-031, A43286-032, A43286-033, A43286-034, A43286-035, A43286-036, A43286-037, A43286-038, A43286-039, A43286-040, A43286-041, A43286-042, A43286-043, A43286-044, A43286-045, A43286-046, A43286-047, A43286-048, A43286-049, A43286-050, A43286-051, A43286-052, A43286-053, A43286-054, A43286-055, A43286-056, A43286-057, A43286-058, A43286-059, A43286-060, A43286-061, A46158-011, A46158-012, A46158-013, A46158-014, A46158-015, A46158-016, A46158-017, A46158-018, A46158-019, A46158-020, A46158-021, A46158-022, A46158-023, A46158-024, A46158-025, A46158-026, A46158-027, A46158-028, A46158-029, A46158-030, A46158-031, A46158-032, A46158-033, A46158-034, A46158-035, A46158-036, A46158-037, A46158-038, A46158-039, A46158-040, A46158-041, A46158-042, A46158-043, A46158-044, A46158-045, A46158-046, A46158-047, A46158-048, A46158-049, A46158-050, A46158-051, A46158-052, A46158-053, A46158-054, A46158-055, A46158-056, A47115-011, A47115-012, A47115-013, A47115-014, A47115-015, A47115-016, A47115-017, A47115-018, A47115-019, A47115-020, A47115-021, A47115-022, A47115-023, A47115-024, A47115-025, A47115-026, A47115-027, A47115-028, A47115-029, A47115-030, A47115-031, A47115-032, A47115-033, A47115-034, A47115-035, A47115-036, A47115-037, A47115-038, A47115-039, A47115-040, A47293-011, A47293-012, A47293-013, A47293-014, A47293-015, A47293-016, A47293-017, A47293-018, A47293-019, A47293-020, A47293-021, A47293-022, A47293-023, A47293-024, A47293-025, A47293-026, A47293-027, A47293-028, A47293-029, A47293-030, A47293-031, A47293-032, A47293-033, A47293-034, A47293-035, A47293-036, A47431-041, A47431-042, A47431-043, A47431-044, A47431-045, A47431-046, A47431-047, A47431-048, A47431-049, A47431-050, A47431-051, A47431-052, A47431-053, A47431-054, A47431-055, A47431-056, A47431-057, A47431-058, A47431-059, A47431-060, A47431-061, A47431-062, A47431-063, A47431-064, A47431-065, A47431-066, A47431-067, A47431-068, A47431-069, A47431-070, A47431-071, A47431-072, A47431-073, A47431-074, A47431-075, A47431-076, A47431-077, A47431-078, A47431-079, A47431-080, A47431-081, A47431-082, A47573-030, A47573-031, A47573-032, A47573-033, A47573-034, A47573-035, A47573-036, A47573-037, A47573-038, A47573-039, A47573-040, A47573-041, A47573-042, A47573-043, A47573-044, A47573-045, A47573-046, A47573-047, A47573-048, A47573-049, A47573-050, A47573-051, A47573-052, A47573-053, A47573-054, A47573-055, A47573-056, A47573-057, A47573-058, A47573-059, A47573-060, A47573-061, A47573-062, A47573-063, A47573-064, A47573-065, A47573-066, A47573-067, A47573-068, A47573-069, A47573-070, A47573-071, A47573-072, A47573-073, A47573-074, A47573-075, A47573-076, A47573-077, A47573-078, A47573-079, A47573-080, A47573-081, A47573-082, A47573-083, A47573-084, A47573-085, A47573-086, A47573-087, A47573-088, A47573-089, A47573-090, A47573-091, A47573-092, A47573-093, A47573-094, A47573-095, A47672-060, A47672-061, A47672-062, A47672-063, A47672-064, A47672-065, A47672-066, A47672-067, A47672-068, A47672-069, A47672-070, A47672-071, A47672-072, A47672-073, A47672-074, A47672-075, A47672-076, A47672-077, A47672-078, A47672-079, A47672-080, A47672-081, A47672-082, A47672-083, A47672-084, A47672-085, A47672-086, A47672-087, A47672-088, A47672-089, A47672-090, A47672-091, A47672-092, A47672-093, A47672-094, A48441-046, A48441-047, A48441-048, A48441-049, A48441-050, A48441-051, A48441-052, A48441-053, A48441-054, A48441-055, A48441-056, A48441-057, A48441-058, A48441-059, A48441-060, A48441-061, A48441-062, A48441-063, A48441-064, A48441-065, A48441-066, A48441-067, A48441-068, A48441-069, A48441-070, A48441-071, A48441-072, A48441-073, A48441-074, A48441-075, A48441-076, A48441-077, A48441-078, A48441-079, A48441-080, A48441-081, A48441-082, A48441-083, A48441-084, A48441-085, A48441-086, A48441-087, A48441-088, A48441-089, A48441-090, A48441-091, A48538-115, A48538-116, A48538-117, A48538-118, A48538-119, A48538-120, A48538-121, A48538-122, A48538-123, A48538-124, A48538-125, A48538-126, A48538-127, A48538-128, A48538-129, A48538-130, A48538-131, A48538-132, A48538-133, A48538-134, A48538-135, A48538-136, A48538-137, A48538-138, A48538-139, A48538-140, A48538-141, A48538-142, A48538-143, A48538-144, A48538-145, A48538-146, A48538-147, A48538-148, A48538-149, A48538-150, A49912-056, A49912-057, A49912-058, A49912-059, A49912-060, A49912-061, A49912-062, A49912-063, A49912-064, A49912-065, A49912-066, A49912-067, A49912-068, A49912-069, A49912-070, A49912-071, A49912-072, A49912-073, A49912-074, A49912-075, A49912-076, A49912-077, A49912-078, A49912-079, A49912-080, A49912-081, A49912-082, A49912-083, A49912-084, A49912-085, A49912-086, A49912-087, A49912-088, A49912-089, A49912-090, A49912-091, A49912-092, A49912-093, A49975-051, A49975-052, A49975-053, A49975-054, A49975-055, A49975-056, A49975-057, A49975-058, A49975-059, A49975-060, A49975-061, A49975-062, A49975-063, A49975-064, A49975-065, A49975-066, A49975-067, A49975-068, A49975-069, A49975-070, A49975-071, A49975-072, A49975-073, A49975-074, A49975-075, A49975-076, A49975-077, A49975-078, A49975-079, A52779-011, A52779-012, A52779-013, A52779-014, A52779-015, A52779-016, A52779-017, A52779-018, A52779-019, A52779-020, A52779-021, A52779-022, A52779-023, A53118-021, A53118-022, A53118-023, A53118-024, A53118-025, A53118-026, A53118-027, A53118-028, A53118-029, A53118-030, A53118-031, A53118-032, A53118-033, A53118-034, A53118-035, A53118-036, A53118-037, A53118-038, A53118-039, A53118-040, A53118-041, A53118-042, A53118-043, A53118-044, A53118-045, A53118-046, A53118-047, A53118-048, A53118-049, A53118-050, A53118-051, A53118-052, A53118-053, A53118-054, A53118-055, A53118-056, A53118-057, A53118-058, A53118-059, A53118-060, A53118-061, A53118-062, A53118-063, A53118-064, A53118-065, A53118-066, A53118-067, A53118-068, A53118-069, A53118-070, A53118-071, A53118-072, A53118-073, A53118-074, A53118-075, A53118-076, A53118-077, A53118-078, A53118-079, A53118-080, A53118-081, A53177-011, A53177-012, A53177-013, A53177-014, A53177-015, A53177-016, A53177-017, A53177-018, A53177-019, A53177-020, A53177-021, A53177-022, A53177-023, A53177-024, A53177-025, A53177-026, A54244-064, A54244-065, A54244-066, A54244-067, A54244-068, A54244-069, A54244-070, A54244-071, A54244-072, A54244-073, A54350-034, A54350-035, A54350-036, A54350-037, A54350-038, A54350-039, A54350-040, A54350-041, A54350-042, A54350-043, A54350-044, A54350-045, A54350-067, A54350-068, A54350-069, A54350-070;
d) DBM T41110INT 0.5CC GRAFTON GEL, REF T41110INT, GTIN 00643169964839, Serial Numbers: A37724-094, A37724-095, A37724-096, A37724-097, A37724-098, A37724-099, A37724-100, A37724-101, A37724-102, A37724-103, A37724-104, A37724-105, A37724-106, A37724-107, A37724-108, A37724-109, A37724-110, A37724-111, A37724-112, A37724-113, A37724-114, A37724-115, A37724-116, A37724-117, A37724-118, A37724-119, A37724-120, A37724-121, A37724-122, A37724-123, A37724-124, A37724-125, A37724-126, A39537-031, A39537-032, A39537-033, A39537-034, A39537-035, A39537-036, A39537-037, A39537-038, A39537-039, A39537-040, A39537-041, A39537-042, A39537-043, A39537-044, A39537-045, A39537-046, A39537-047, A39537-048, A39537-049, A39537-050, A39537-051, A39537-052, A39537-053, A42444-041, A42444-042, A42444-043, A42444-044, A42444-045, A42444-046, A42444-047, A42444-048, A42444-049, A42444-050, A42444-051, A42444-052, A42571-102, A42571-103, A42571-104, A42571-105, A42571-106, A42571-107, A42571-108, A42571-109, A42780-068, A42780-069, A42780-070, A42780-071, A42780-072, A42780-073, A42780-074, A42780-075, A42780-076, A42780-077, A42780-078, A42780-079, A42780-080, A42780-081, A42780-082, A42780-083, A42780-084, A42780-085, A42780-086, A42780-087, A42780-088, A42780-089, A42780-090, A42780-091, A42780-092, A42780-093, A42780-094, A42780-095, A42780-096, A42780-097, A42780-098, A42780-099, A42780-100, A42780-101, A43098-042, A43098-043, A43098-044, A43098-045, A43098-046, A43098-047, A43098-048, A43098-049, A43098-050, A43098-051, A43098-052, A43098-053, A43098-054, A43098-055, A43098-056, A43098-057, A43098-058, A43098-059, A43098-060, A43098-061, A43098-062, A43098-063, A43098-064, A43284-036, A43284-037, A43284-038, A43284-039, A43284-040, A43284-041, A43284-042, A43284-043, A43284-044, A43284-045, A43286-067, A43286-068, A43286-069, A43286-070, A43286-071, A43286-072, A43286-073, A43286-074, A43286-075, A43286-076, A43286-077, A43286-078, A43286-079, A43286-080, A43286-081, A43286-082, A43858-011, A43858-012, A43858-013, A43858-014, A43858-015, A43858-016, A43858-017, A43858-018, A43858-019, A43858-020, A43858-021, A43858-022, A43858-023, A43858-024, A43858-025, A43858-026, A43858-027, A43858-028, A43858-029, A43858-030, A43858-031, A43858-032, A43858-033, A43858-034, A43858-035, A43858-036, A43858-037, A43858-038, A43858-039, A43858-040, A43858-041, A43858-042, A43858-043, A43858-044, A43858-045, A43858-046, A43858-047, A43858-048, A43858-049, A43858-050, A43858-051, A43858-052, A43858-053, A43858-054, A43858-055, A43858-056, A43858-057, A43858-058, A43858-059, A43858-060, A43858-061, A44260-011, A44260-012, A44260-013, A44260-014, A44260-015, A44260-016, A44260-017, A44260-018, A44260-019, A44260-020, A44260-021, A44260-022, A44260-023, A44260-024, A44260-025, A46158-057, A46158-058, A46158-059, A46158-060, A46158-061, A46158-062, A46158-063, A46158-064, A46158-065, A46158-066, A46333-031, A46333-032, A46333-033, A46333-034, A46333-035, A46333-036, A46333-037, A46333-038, A46333-039, A46333-040, A46333-041, A46333-042, A46333-043, A46333-044, A46333-045, A46333-046, A46333-047, A46333-048, A46333-049, A46333-050, A46333-051, A46333-052, A46333-053, A46791-011, A46791-012, A46791-013, A46791-014, A46791-015, A46791-016, A46791-017, A46791-018, A46791-019, A46791-020, A46791-021, A46791-022, A46791-023, A46791-024, A46791-025, A46791-026, A46791-027, A47034-041, A47034-042, A47034-043, A47034-044, A47034-045, A47034-046, A47034-047, A47034-048, A47034-049, A47034-050, A47034-051, A47034-052, A47034-053, A47034-054, A47034-055, A47034-056, A47034-057, A47034-058, A47034-059, A47034-060, A47034-061, A47034-062, A48114-084, A48114-085, A48114-086, A48114-087, A48114-088, A48114-089, A48114-090, A48114-091, A48114-092, A48114-093, A48114-094, A48114-095, A48114-096, A48114-097, A48114-098, A48114-099, A48114-100, A48114-101, A48699-011, A48699-012, A48699-013, A48699-014, A48699-015, A48699-016, A48699-017, A48699-018, A48699-019, A48699-020, A48699-021, A48699-022, A48699-023, A48699-024, A48699-025, A48699-026, A48699-027, A48699-028, A48699-029, A48699-030, A48699-031, A48699-032, A49529-011, A49529-012, A49529-013, A49529-014, A49529-015, A49529-016, A49529-017, A49529-018, A49529-019, A49529-020, A49529-021, A49529-022, A49529-023, A49529-024, A49529-025, A49529-026, A49529-027, A49529-028, A49529-029, A51547-041, A51547-042, A51547-043, A51547-044, A51547-045, A51547-046, A51547-047, A51547-048, A53185-011, A53185-012, A53185-013, A53185-014, A53185-015;
e) DBM T41120 1CC GRAFTON GEL, REF T41120, GTIN 00643169122963, Serial Numbers: A38528-107, A38528-108, A38528-109, A38528-110, A38528-111, A38528-112, A38528-113, A38528-114, A38528-115, A38528-116, A38528-117, A38528-118, A38528-119, A38528-120, A38528-121, A38528-122, A38528-123, A38528-124, A38528-125, A38528-126, A38528-127, A38528-128, A38528-129, A38528-130, A38528-132, A38528-133, A38528-134, A38528-135, A38528-136, A38528-137, A38528-138, A38528-139, A38528-140, A38528-141, A38528-142, A38528-143, A38528-144, A38528-145, A38528-146, A38528-147, A38528-148, A38528-149, A38528-150, A38528-151, A38528-152, A38528-153, A38528-154, A38528-155, A38528-156, A38528-157, A38528-158, A38528-159, A38528-160, A38528-161, A38528-162, A38528-163, A38528-165, A38528-166, A38528-167, A38528-168, A38528-169, A38528-170, A38528-171, A38528-172, A38528-173, A38528-174, A38528-175, A38528-176, A38528-177, A38528-178, A38528-179, A38528-180, A38528-181, A38528-182, A38528-183, A38528-184, A38528-185, A38528-186, A38528-187, A38528-188, A38528-189, A38528-190, A38528-191, A38528-192, A38528-194, A38528-195, A38528-196, A38528-197, A38528-198, A38528-199, A38528-200, A38528-201, A38528-202, A38528-203, A38528-204, A38528-205, A38528-206, A38528-207, A39248-129, A39248-130, A39248-131, A39248-132, A39248-133, A39248-134, A39248-135, A39248-136, A39248-137, A39248-138, A39248-139, A39248-140, A39248-141, A39248-142, A39248-143, A39248-144, A39248-145, A39248-146, A39248-147, A39248-148, A39248-149, A39248-150, A39248-152, A39248-153, A39248-154, A39248-155, A39248-157, A39248-158, A39248-159, A39248-160, A39248-161, A39248-162, A39248-163, A39248-164, A39248-165, A39248-166, A39248-167, A39248-168, A39248-169, A39248-170, A39248-171, A39248-172, A39248-173, A39248-174, A39248-175, A39248-176, A39248-177, A39248-178, A39248-179, A39248-180, A39248-181, A39248-182, A39248-183, A39248-184, A39248-185, A39248-186, A39248-187, A39248-188, A39248-189, A39248-190, A39248-191, A39248-192, A39248-193, A39248-194, A39248-195, A39248-196, A39248-197, A39248-198, A39248-199, A39248-200, A39248-201, A39248-202, A39248-203, A39248-204, A39248-205, A39248-206, A39248-207, A39248-208, A39248-209, A39248-210, A39248-211, A39248-212, A39248-213, A39248-214, A39248-215, A39248-216, A39248-217, A39248-219, A39248-220, A39248-221, A39248-222, A39248-223, A39248-224, A39248-225, A39248-226, A39248-227, A39248-228, A39248-229, A39248-230, A39248-231, A39248-232, A39248-233, A39248-234, A39248-235, A39248-236, A40738-117, A40738-118, A40738-119, A40738-120, A40738-121, A40738-122, A40738-123, A40738-124, A40738-125, A40738-126, A40738-127, A40738-128, A40738-129, A40738-130, A40738-131, A40738-132, A40738-133, A40738-134, A40738-135, A40738-136, A40738-137, A40738-138, A40738-139, A40738-140, A40738-141, A40738-142, A40738-143, A40738-144, A40738-145, A40738-146, A40738-147, A40738-148, A40738-149, A40738-150, A40738-151, A40738-152, A40738-153, A40738-154, A42221-086, A42221-087, A42221-088, A42221-089, A42221-090, A42221-091, A42221-092, A42221-093, A42221-094, A42221-095, A42221-096, A42221-097, A42221-098, A42221-099, A42221-100, A42221-101, A42221-102, A42221-103, A42221-104, A42221-105, A42221-106, A42221-107, A42221-108, A42221-109, A42221-110, A42221-111, A42221-112, A42221-113, A42221-114, A42221-115, A42221-116, A42221-117, A42221-118, A42221-119, A42221-120, A42221-121, A42221-122, A42221-123, A42221-124, A42221-125, A42221-126, A42221-127, A42221-128, A42221-129, A42221-130, A42221-131, A42221-132, A42221-133, A42221-134, A42221-135, A42221-136, A42221-137, A42221-138, A42221-139, A42221-140, A42221-141, A42221-142, A42221-143, A42221-144, A42221-145, A42221-146, A42221-147, A42221-148, A42221-149, A42221-150, A42221-151, A42221-152, A42221-153, A42221-154, A42221-155, A42221-156, A42221-157, A42221-158, A42221-159, A42221-160, A42221-161, A42221-162, A42221-163, A42221-164, A42221-165, A42221-166, A42221-167, A42221-168, A42221-169, A42221-170, A42221-171, A42221-172, A42221-173, A42221-174, A42221-175, A42221-176, A42221-177, A42221-178, A42221-179, A42221-180, A42221-181, A42221-182, A42221-183, A42221-184, A42221-185, A42221-186, A42221-187, A42221-188, A42221-189, A42221-190, A42221-191, A42221-192, A42545-014, A42545-015, A42545-016, A42545-017, A42545-018, A42545-019, A42545-020, A42545-021, A42545-022, A42545-023, A42545-024, A42545-026, A42545-027, A42545-028, A42545-029, A42545-030, A42545-031, A42545-032, A42545-033, A42545-034, A42545-035, A42545-036, A42545-037, A42545-038, A42545-039, A42545-040, A42545-041, A42545-042, A42545-043, A42545-044, A42545-045, A42545-046, A42545-047, A42545-048, A42545-049, A42545-050, A42545-051, A42545-052, A42545-053, A42545-054, A42545-055, A42545-056, A42545-057, A42545-058, A42545-059, A42545-060, A42545-061, A42545-062, A42545-063, A42545-064, A42545-065, A42545-066, A42545-067, A42545-068, A42545-069, A42545-070, A42545-071, A42545-072, A42545-073, A42545-074, A42545-075, A42545-076, A42545-077, A42545-078, A42545-079, A42545-080, A42545-081, A42545-082, A42545-083, A42545-084, A42545-085, A42545-086, A42545-087, A42545-088, A42545-089, A42545-090, A42545-091, A42545-092, A42545-093, A42545-094, A42545-095, A42982-039, A42982-040, A42982-041, A42982-042, A42982-043, A42982-044, A42982-045, A42982-046, A42982-047, A42982-048, A42982-049, A42982-050, A42982-051, A42982-052, A42982-053, A42982-054, A42982-055, A42982-056, A42982-057, A42982-058, A42982-059, A42982-060, A42982-061, A42982-062, A42982-063, A42982-064, A42982-065, A42982-067, A42982-068, A42982-069, A42982-070, A42982-071, A42982-072, A42982-073, A42982-074, A42982-075, A42982-076, A42982-077, A42982-078, A42982-079, A42982-080, A42982-081, A42982-082, A42982-083, A42982-084, A42982-085, A42982-086, A42982-087, A42982-088, A42982-089, A42982-090, A42982-091, A42982-092, A42982-093, A42982-094, A43231-011, A43231-012, A43231-013, A43231-015, A43231-016, A43231-017, A43231-018, A43231-019, A43231-020, A43231-021, A43231-022, A43231-023, A43231-024, A43231-025, A43231-026, A43231-027, A43231-028, A43231-029, A43231-030, A43231-031, A43231-032, A43231-033, A43231-034, A43231-035, A43231-036, A43231-037, A43231-038, A43231-039, A43231-040, A43231-041, A43231-042, A43231-043, A43231-044, A43231-045, A43231-046, A43231-047, A43231-048, A43231-049, A43231-050, A43231-051, A43231-052, A43231-053, A43231-054, A43231-055, A43231-056, A43231-057, A43231-058, A43231-059, A43231-060, A43231-061, A43231-062, A43231-063, A43231-064, A43231-065, A43231-066, A43231-067, A43231-068, A43231-069, A43231-070, A43231-071, A43231-072, A43231-073, A43231-074, A43231-075, A43231-076, A43231-077, A43231-078, A43231-079, A43231-080, A43231-081, A43231-082, A43231-083, A43231-084, A43231-085, A43231-086, A43231-087, A43231-088, A43231-089, A43231-090, A43231-091, A43231-092, A43231-093, A43231-094, A43231-095, A43231-096, A43231-097, A43231-098, A43231-099, A43231-100, A43231-101, A43231-102, A43231-103, A43231-104, A43231-105, A43231-106, A43231-107, A43231-108, A43231-109, A43231-110, A43231-111, A43231-112, A43231-113, A43231-114, A43231-115, A43231-116, A43231-117, A43231-118, A43231-119, A43231-120, A43231-121, A43231-122, A43231-123, A43231-124, A43231-125, A43231-126, A43336-040, A43336-041, A43336-042, A43336-043, A43336-044, A43336-045, A43336-046, A43336-047, A43336-048, A43336-049, A43336-050, A43336-051, A43336-052, A43336-053, A43336-054, A43336-055, A43336-056, A43336-057, A43336-058, A43336-059, A43336-060, A43336-061, A43336-062, A43336-063, A43336-064, A43336-065, A43336-066, A43336-067, A43336-068, A43336-069, A43336-070, A43336-071, A43336-072, A43336-073, A43336-074, A43336-075, A43336-076, A43336-077, A43336-078, A43336-079, A43336-080, A43336-081, A43336-082, A43336-083, A43336-084, A43336-085, A43336-086, A43336-087, A43336-088, A43336-089, A43336-090, A43336-091, A43336-092, A43336-093, A43336-094, A43336-095, A43336-096, A43336-097, A43336-098, A43336-099, A43336-100, A43336-101, A43336-102, A43336-103, A43336-104, A43336-105, A43336-106, A43336-107, A43336-108, A43336-109, A43336-110, A43336-111, A43336-112, A43336-113, A43336-114, A43336-115, A43336-116, A43336-117, A43344-248, A43344-249, A43344-250, A43344-251, A43344-252, A43344-253, A43344-254, A43344-255, A43344-256, A43344-257, A43344-258, A43344-259, A43344-260, A43344-261, A43344-262, A43344-263, A43344-264, A43344-265, A43344-266, A43344-267, A43344-268, A43344-269, A43344-270, A43344-271, A43344-272, A43344-273, A43344-274, A43344-275, A43344-276, A43344-277, A43344-278, A43344-279, A43344-280, A43344-281, A43344-282, A43344-283, A43344-284, A43344-285, A43344-286, A43344-287, A43344-288, A43344-289, A43344-290, A43344-291, A43344-292, A43344-293, A43344-294, A43344-295, A43344-296, A43344-297, A43344-298, A43344-299, A43344-300, A43344-301, A43344-302, A43344-303, A43344-304, A43344-305, A43344-306, A43344-308, A43344-309, A43344-310, A43344-311, A43344-312, A43344-313, A43344-314, A43344-315, A43344-316, A43344-317, A43345-096, A43345-097, A43345-098, A43345-099, A43345-100, A43345-101, A43345-102, A43345-103, A43345-104, A43345-105, A43345-106, A43345-107, A43345-108, A43345-109, A43345-110, A43345-111, A43345-112, A43345-113, A43345-114, A43345-115, A43345-116, A43345-117, A43345-118, A43345-119, A43345-120, A43345-121, A43345-122, A43345-123, A43345-124, A43345-125, A43345-126, A43345-127, A43345-128, A43345-129, A43345-130, A43345-131, A43345-132, A43345-133, A43345-134, A43345-135, A43345-136, A43345-137, A43345-138, A43345-139, A43345-140, A43345-141, A43345-142, A43345-143, A43345-144, A43345-145, A43345-146, A43345-147, A43345-148, A43345-149, A43345-150, A43345-151, A43345-152, A43345-153, A43345-154, A43345-155, A44405-011, A44405-012, A44405-013, A44405-014, A44405-015, A44405-016, A44405-017, A44405-018, A44405-019, A44405-020, A44405-021, A44405-022, A44405-023, A44405-024, A44405-025, A44405-026, A44405-027, A44405-028, A44405-029, A44405-030, A44405-031, A44405-032, A44405-033, A44405-034, A44405-035, A44405-036, A44405-037, A44405-038, A44405-039, A44405-040, A44405-041, A44405-042, A44405-043, A44405-044, A44405-045, A44405-046, A44405-047, A44405-048, A44405-049, A44405-050, A44405-051, A44405-052, A44405-053, A44405-054, A44413-051, A44413-052, A44413-053, A44413-054, A44413-055, A44413-056, A44413-059, A44413-060, A44413-061, A44413-062, A44413-063, A44413-064, A44413-065, A44413-066, A44413-067, A44413-068, A44413-069, A44413-070, A44413-071, A44413-072, A44413-073, A44413-074, A44413-075, A44413-076, A44413-077, A44413-078, A44413-079, A44413-080, A44413-081, A44413-082, A44413-083, A44413-084, A44413-085, A44413-086, A44413-087, A44413-088, A44413-089, A44413-090, A44413-091, A44413-092, A44413-093, A44413-094, A44413-095, A44413-097, A44413-098, A44413-099, A44413-100, A44413-101, A44413-102, A44413-103, A44413-104, A44413-105, A44413-106, A44413-108, A44413-109, A44413-110, A44413-111, A44413-112, A44413-113, A44413-114, A44413-115, A44413-116, A44413-117, A44413-118, A44413-119, A44413-120, A44413-121, A44413-122, A44413-123, A44413-124, A44413-125, A44413-126, A44413-127, A44413-128, A44413-129, A44413-130, A44413-131, A44413-132, A44413-133, A44413-134, A44413-135, A44413-136, A44413-137, A44413-138, A44413-139, A44413-140, A44413-141, A44413-142, A44413-143, A44413-144, A44413-145, A44413-146, A44413-147, A44413-148, A44413-149, A44413-150, A44413-151, A44413-152, A44413-153, A44413-154, A44413-155, A44413-156, A44413-157, A44413-158, A44413-159, A44413-160, A44413-161, A44413-162, A44413-163, A44413-164, A44413-165, A44413-166, A44413-167, A44413-168, A44413-169, A44413-170, A44413-171, A44413-172, A44413-173, A44413-174, A44413-175, A44413-176, A44413-177, A44413-178, A44413-179, A44413-180, A44413-181, A44413-182, A44413-183, A44413-184, A44413-185, A44413-186, A44413-187, A44413-188, A45217-080, A45217-081, A45217-082, A45217-083, A45217-084, A45217-085, A45217-086, A45217-087, A45217-088, A45217-089, A45217-090, A45217-091, A45217-092, A45217-093, A45217-094, A45217-095, A45217-096, A45217-097, A45217-098, A45217-099, A45217-100, A45217-101, A45217-102, A45217-103, A45217-104, A45217-105, A45217-106, A45217-107, A45217-108, A45217-109, A45217-110, A45217-111, A45217-112, A45217-113, A45217-114, A45217-115, A45217-116, A45217-117, A45217-118, A45217-119, A45217-120, A45217-121, A45217-122, A45217-123, A45217-124, A45217-125, A45217-126, A45217-127, A45217-129, A45217-130, A45217-131, A45217-132, A45217-133, A45217-135, A45217-136, A45217-137, A45217-138, A45217-139, A45217-140, A45217-141, A45217-142, A45217-143, A45217-144, A45217-145, A45217-146, A45217-147, A45217-148, A45217-149, A45217-150, A45217-151, A45217-152, A45217-153, A45217-154, A45217-155, A45217-156, A45217-157, A45217-158, A45217-159, A45217-160, A45217-161, A45217-162, A45217-163, A45217-164, A45217-165, A45217-166, A45217-167, A45217-168, A45613-046, A45613-047, A45613-048, A45613-049, A45613-050, A45613-051, A45613-052, A45613-053, A45613-054, A45613-055, A45613-056, A45613-057, A45613-058, A45613-059, A45613-060, A45613-061, A45613-062, A45613-063, A45613-064, A45613-065, A45613-066, A45613-067, A45613-068, A45613-069, A45613-070, A45613-071, A45613-072, A45613-073, A45613-074, A45613-075, A45613-076, A45613-077, A45613-078, A45613-079, A45613-080, A45613-081, A45613-082, A45613-083, A45613-084, A45613-085, A45613-086, A45613-087, A45613-088, A45613-089, A45613-090, A45613-091, A45613-092, A45613-093, A45613-094, A45613-095, A45613-096, A45613-097, A45613-098, A45613-099, A45613-100, A45613-101, A45613-102, A45613-103, A45613-104, A45613-105, A45613-106, A45613-107, A45613-108, A45613-109, A45613-110, A45613-111, A45613-112, A45613-113, A45613-114, A45613-115, A45613-116, A45613-117, A45613-118, A45613-119, A45613-120, A45613-121, A45613-122, A45613-123, A45613-124, A45613-125, A45613-126, A45613-127, A45613-128, A45613-129, A45613-130, A45613-131, A45613-132, A45613-133, A45613-134, A45613-135, A45613-136, A45613-137, A46849-026, A46849-027, A46849-028, A46849-029, A46849-030, A46849-031, A46849-032, A46849-033, A46849-034, A46849-035, A46849-036, A46849-037, A46849-038, A46849-039, A46849-040, A46849-041, A46849-042, A46849-043, A46849-044, A46849-045, A46849-046, A46849-047, A46849-048, A46849-049, A46849-050, A46849-051, A46849-052, A46849-053, A46849-054, A46849-055, A46849-056, A46849-057, A46849-058, A46849-059, A46849-060, A46849-061, A46849-062, A46849-063, A46849-064, A46849-065, A46849-066, A46849-067, A46849-068, A46849-069, A46849-070, A46849-071, A46849-072, A46849-073, A46849-074, A46849-075, A46849-076, A46849-077, A46849-078, A46849-079, A46849-080, A46849-081, A46849-082, A46849-083, A46849-084, A46849-085, A46849-086, A46849-087, A46849-088, A46849-089, A46849-090, A46849-091, A46849-092, A46849-093, A46849-094, A46849-095, A47127-103, A47127-104, A47127-105, A47127-106, A47127-107, A47127-108, A47127-109, A47127-110, A47127-112, A47127-113, A47127-114, A47127-115, A47127-116, A47127-117, A47127-118, A47127-119, A47127-120, A47127-121, A47127-122, A47127-123, A47127-124, A47127-125, A47127-126, A47127-127, A47127-128, A47127-129, A47127-130, A47127-131, A47127-132, A47127-133, A47127-134, A47127-135, A47127-136, A47127-137, A47127-138, A47127-139, A47127-140, A47127-141, A47127-142, A47127-143, A47127-144, A47127-145, A47127-146, A47127-147, A47127-148, A47127-149, A47127-150, A47127-151, A47127-152, A47127-153, A47127-154, A47127-155, A47127-156, A47127-157, A47127-158, A47127-159, A47127-160, A47127-161, A47127-162, A47127-163, A47127-164, A47127-165, A47127-166, A47127-167, A47127-168, A47127-169, A47127-170, A47127-171, A47127-172, A47127-173, A47127-174, A47127-175, A47127-176, A47127-177, A47127-178, A47127-179, A47127-180, A47127-181, A47127-182, A47127-183, A47127-184, A47127-185, A47127-186, A47127-187, A47127-188, A47127-189, A47127-190, A47127-191, A47127-192, A47127-193, A47127-194, A47127-195, A47127-196, A47127-197, A47127-198, A47127-199, A47127-200, A47127-201, A47127-202, A47127-203, A47127-204, A47127-205, A47127-206, A47127-207, A47127-208, A47127-209, A47293-037, A47293-038, A47293-039, A47293-040, A47293-041, A47293-042, A47293-043, A47293-044, A47293-045, A47293-046, A47293-047, A47293-048, A47293-049, A47293-050, A47293-051, A47293-052, A47293-053, A47293-054, A47293-055, A47293-056, A47293-057, A47293-058, A47293-059, A47293-060, A47293-061, A47293-062, A47293-063, A47293-064, A47293-065, A47293-066, A47293-067, A47293-068, A47293-071, A47293-072, A47293-073, A47293-074, A47293-075, A47293-076, A47293-077, A47293-078, A47293-079, A47293-080, A47293-081, A47293-082, A47293-083, A47293-084, A47293-085, A47293-086, A47293-087, A47293-088, A47293-089, A47293-090, A47293-091, A47293-092, A47293-093, A47293-094, A47293-095, A47293-096, A47293-097, A47293-098, A47293-099, A47293-100, A47293-101, A47293-102, A47293-103, A47293-104, A47293-105, A47293-106, A47293-107, A47293-108, A47293-109, A47293-110, A47293-111, A47293-112, A47293-113, A47293-114, A47293-115, A47293-116, A47293-117, A47293-118, A47293-119, A47293-120, A47293-121, A47293-122, A47293-123, A47293-124, A47293-125, A47293-126, A47293-127, A47293-128, A47293-129, A47293-130, A47293-131, A47293-132, A47293-133, A47293-134, A47293-135, A47482-026, A47482-027, A47482-028, A47482-029, A47482-030, A47482-031, A47482-032, A47482-033, A47482-034, A47482-035, A47482-036, A47482-037, A47482-038, A47482-039, A47482-040, A47482-042, A47482-043, A47482-044, A47482-045, A47482-046, A47482-047, A47482-048, A47482-049, A47482-050, A47482-051, A47482-052, A47482-053, A47482-054, A47482-055, A47482-056, A47482-057, A47482-058, A47482-059, A47482-060, A47482-061, A47482-062, A47482-063, A47482-064, A47482-065, A47482-066, A47482-067, A47482-068, A47482-069, A47482-070, A47482-071, A47482-072, A47482-073, A47482-074, A47482-075, A47482-076, A47482-077, A47482-078, A47482-079, A47482-080, A47482-082, A47482-083, A47482-084, A47482-085, A47482-086, A47482-087, A47482-088, A47482-089, A47482-090, A47482-091, A48188-011, A48188-012, A48188-013, A48188-014, A48188-015, A48188-016, A48188-017, A48188-018, A48188-019, A48188-020, A48188-021, A48188-022, A48188-023, A48188-024, A48188-025, A48188-026, A48188-027, A48188-028, A48188-029, A48188-030, A48188-031, A48188-032, A48188-033, A48188-034, A48188-035, A48188-036, A48188-037, A48188-038, A48188-039, A48188-040, A48188-041, A48188-042, A48188-043, A48188-044, A48188-045, A48188-046, A48188-047, A48188-048, A48188-049, A48188-050, A48188-051, A48188-052, A48188-053, A48188-054, A48188-055, A48188-056, A48538-151, A48538-152, A48538-153, A48538-154, A48538-155, A48538-156, A48538-157, A48538-158, A48538-159, A48538-160, A48538-161, A48538-162, A48538-163, A48538-164, A48538-165, A48538-166, A48538-167, A48538-168, A48538-169, A48538-170, A48538-171, A48538-172, A48538-173, A48538-174, A48538-175, A48538-176, A48538-177, A48538-178, A48538-179, A48538-180, A48538-181, A48538-182, A48538-183, A48538-184, A48538-185, A48538-186, A48538-187, A48538-188, A48538-189, A48538-190, A48538-191, A48538-192, A48538-193, A48538-194, A48538-195, A48538-196, A48538-197, A48538-198, A48538-199, A48538-200, A48538-201, A48538-202, A48538-203, A48538-204, A48538-205, A48538-206, A48538-207, A48538-208, A48538-209, A48538-210, A48538-211, A48538-212, A48538-213, A48538-214, A48538-215, A48538-216, A48538-217, A48538-218, A48538-219, A48538-220, A48538-221, A48538-222, A48538-223, A48538-224, A48538-225, A48538-226, A48538-227, A48538-228, A48538-229, A48538-230, A48538-231, A48538-232, A48538-233, A48538-234, A48538-235, A48538-236, A48538-237, A48538-238, A48538-239, A48538-240, A48538-241, A48538-242, A48538-243, A48538-244, A48538-245, A48538-246, A48538-247, A48538-248, A48538-249, A48538-250, A48538-251, A48538-252, A48672-036, A48672-037, A48672-038, A48672-039, A48672-040, A48672-042, A48672-043, A48672-044, A48672-045, A48672-046, A48672-047, A48672-048, A48672-049, A48672-050, A48672-051, A48672-052, A48672-053, A48672-054, A48672-055, A48672-056, A48672-057, A48672-058, A48672-059, A48672-060, A48672-061, A48672-062, A48672-063, A48672-064, A48672-065, A48672-066, A48672-067, A48672-068, A48672-069, A48672-070, A48672-071, A48672-072, A48672-073, A48672-074, A48672-075, A48672-076, A48672-077, A48672-078, A48672-079, A48672-080, A48672-081, A48672-082, A48672-083, A48672-084, A48672-085, A48672-086, A48672-087, A48672-088, A48672-089, A48672-090, A48672-091, A48672-092, A48672-093, A48672-094, A48672-095, A48672-096, A48672-097, A48672-098, A48672-099, A48672-100, A48672-101, A48672-102, A48672-103, A48672-104, A48672-105, A48672-106, A48672-107, A50166-021, A50166-022, A50166-023, A50166-025, A50166-026, A50166-027, A50166-028, A50166-029, A50166-030, A50166-031, A50166-032, A50166-033, A50166-034, A50166-035, A50166-036, A50166-037, A50166-038, A50166-039, A50166-040, A50166-041, A50166-042, A50166-044, A50166-045, A50166-046, A50166-047, A50166-048, A50166-049, A50166-050, A50166-051, A50166-052, A50166-053, A50166-054, A50166-055, A50166-056, A50166-057, A50166-058, A50166-059, A50166-060, A50166-061, A50166-062, A50166-063, A50166-064, A50166-065, A50166-066, A50166-067, A50166-068, A50968-061, A50968-062, A50968-063, A50968-064, A50968-065, A50968-066, A50968-067, A50968-068, A50968-069, A50968-070, A50968-071, A50968-072, A50968-073, A50968-074, A50968-075, A50968-076, A50968-077, A50968-078, A50968-079, A50968-080, A50968-081, A50968-082, A50968-083, A50968-084, A50968-085, A50968-086, A50968-087, A50968-088, A50968-089, A50968-090, A50968-091, A50968-092, A50968-093, A50968-094, A50968-095, A50968-096, A50968-097, A50968-098, A50968-099, A50968-100, A50968-101, A50968-102, A50968-103, A50968-104, A50968-105, A50968-106, A50968-107, A50968-108, A50968-109, A50968-110, A52779-024, A52779-025, A52779-026, A52779-027, A52779-028, A52779-029, A52779-030, A52779-031, A52779-032, A52779-033, A52779-034, A52779-035, A52779-036, A52779-037, A52779-038, A52779-039, A52779-040, A52779-041, A52779-042, A52779-043, A52779-045, A52779-046, A52779-047, A52779-048, A52779-049, A52779-050, A52779-051, A52779-052, A52779-053, A52779-054, A52779-055, A52779-056, A52779-057, A52779-058, A52779-059, A52779-060, A52779-061, A52779-062, A52779-063, A52779-064, A52779-065, A52779-066, A52779-067, A52779-068, A52779-069, A52779-070, A52779-071, A52779-072, A52779-073, A52779-074, A52779-075, A52779-076, A52779-077, A52779-078, A52779-079, A52779-080, A52779-081, A52779-082, A52779-083, A52779-084, A52779-085, A52779-087, A52779-088, A52779-089, A52779-090, A52779-091, A52779-092, A52779-093, A52779-094, A52779-095, A52779-096, A52779-097, A52779-098, A52914-072, A52914-073, A52914-074, A52914-075, A52914-076, A52914-077, A52914-078, A52914-079, A52914-080, A52914-081, A52914-082, A52914-083, A52914-084, A52914-085, A52914-086, A52914-087, A52914-088, A52914-089, A52914-090, A52914-091, A52914-092, A52914-093, A52914-094, A52914-095, A52914-097, A52914-098, A52914-099, A52914-100, A52914-101, A52914-102, A52914-103, A52914-104, A52914-105, A52914-106, A52914-107, A52914-108, A52914-109, A52914-110, A52914-111, A52914-112, A53133-095, A53133-096, A53133-097, A53133-098, A53133-099, A53133-100, A53133-101, A53133-102, A53133-103, A53133-104, A53133-105, A53133-106, A53133-107, A53133-108, A53133-109, A53133-110, A53133-111, A53133-112, A53133-113, A53133-114, A53133-115, A53133-116, A53133-117, A53133-118, A53133-119, A53133-120, A53133-121, A53133-122, A53133-123, A53133-124, A53133-125, A53133-126, A53133-127, A53133-128, A53133-129, A53133-130, A53133-131, A53133-132, A53133-133, A53133-134, A53133-135, A53133-136, A53133-137, A53133-138, A53133-139, A53133-140, A53133-141, A53133-142, A53133-143, A53133-144, A53133-145, A53133-146, A53133-147, A53133-148, A53133-149, A53133-150, A53133-151, A53133-152, A53133-153, A53133-154, A53133-155, A53133-156, A53133-157, A53133-158, A53133-159, A53133-160, A53133-161, A53133-162, A53133-163, A53133-164, A53133-165, A53133-166, A53133-167, A53133-168, A53133-169, A53133-170, A53133-171, A53133-172, A53133-173, A53133-174, A53133-175, A53133-176, A53133-177, A53133-178, A53770-031, A53770-032, A53770-033, A53770-034, A53770-035, A53770-036, A53770-037, A53770-040, A53770-041, A53770-042, A53770-043, A53770-044, A53770-045, A53770-046, A53770-047, A53770-048, A53770-049, A53770-050, A53770-051, A53770-052, A53770-053, A53770-054, A53770-055, A53770-056, A53770-058, A53770-059, A53770-060, A53770-061, A53770-062, A53770-065, A53770-067, A53770-068, A53770-069, A53770-070, A53770-071, A53770-073, A53770-074, A53770-075, A53770-076, A53770-077, A53770-078, A53770-079, A53770-080, A53770-081, A53770-082, A53770-083, A53770-084, A53770-085, A53878-139, A53878-140, A53878-141, A53878-142, A53878-143, A53878-144, A53878-145, A53878-146, A53878-147, A53878-148, A53878-149, A53878-150, A53878-152, A53878-153, A53878-154, A53878-155, A53878-156, A53878-157, A53878-158, A53878-159, A53878-160, A53878-161, A53878-162, A53878-163, A53878-164, A53878-165, A53878-166, A53878-167, A53878-168, A53878-169, A53878-170, A53878-171, A53878-172, A53878-173, A53878-174, A53878-175, A53878-176, A53878-177, A53878-178, A53878-179, A53878-180, A53878-181, A53878-182, A53878-183, A53878-184, A53878-186, A53878-187, A53878-188, A53878-189, A53878-190, A53878-191, A53878-192, A53878-193, A53878-194, A53878-195, A53878-196, A53878-197, A53878-198, A53878-199, A53878-200, A53878-201, A53878-202, A53878-203, A53878-204, A53878-205, A53878-206, A53878-207, A53878-208, A53878-209, A53878-210, A53878-211, A53878-212, A53878-213, A53878-214, A53878-216, A53878-219, A53878-220, A53878-221, A53878-222, A53878-223, A53878-224, A53878-225, A53878-226, A54373-053, A54373-054, A54373-055, A54373-056, A54373-057, A54373-058, A54373-059, A54373-060, A54373-061, A54373-062, A54373-063, A54373-064, A54373-065, A54373-066, A54373-067, A54373-068, A54373-069, A54373-070, A54373-071, A54373-072, A54373-073, A54373-074, A54373-075, A54373-076, A54373-077, A54373-078, A54373-079, A54373-080, A54373-081, A54373-082, A54373-083, A54373-084, A54373-085, A54373-086, A54373-087, A54373-088, A54373-089, A54373-090, A54373-091, A54373-092, A55124-084, A55124-085, A55124-086, A55124-087, A55124-088, A55124-089, A55124-090, A55124-091, A55124-092, A55124-093, A55124-094, A55124-095, A55124-096, A55124-097, A55124-098, A55124-099, A55124-100, A55124-101, A55124-102, A55124-103, A55124-104, A55124-105, A55124-106, A55124-107, A55124-108, A55124-109, A55124-110, A55124-111, A55124-112, A55124-113, A55124-114, A55124-115, A55124-116, A55124-117, A55124-118, A55124-119, A55124-120, A55124-121, A55124-122, A55124-123, A55124-124, A55124-125, A55124-126, A55124-127, A55124-128, A55124-129, A55124-130, A55124-131, A55124-132, A55124-133, A55124-134, A55124-135, A55124-136, A55124-137, A55124-138, A55124-139, A55124-140, A55124-141, A55124-142, A55124-143, A55124-144, A55124-145, A55124-146, A55124-147, A55124-148, A55124-149, A55124-150, A55124-151, A55124-152, A55124-153, A55124-154, A55124-155, A55124-157, A55124-158, A55124-159, A55124-160;
f) DBM T41120INT 1CC GRAFTON GEL, REF T41120INT, GTIN 00643169964846, Serial Numbers: A39248-237, A39248-238, A39248-239, A39248-240, A39248-241, A39248-242, A39248-243, A39248-244, A39248-245, A39248-246, A39248-247, A39248-248, A39248-249, A39248-250, A39248-251, A39248-252, A39248-253, A39248-254, A39248-255, A39248-256, A39248-257, A39248-258, A39248-259, A39248-260, A39248-261, A39248-262, A39248-263, A39248-264, A39248-265, A39248-266, A39248-267, A39248-268, A39537-054, A39537-055, A39537-057, A39537-058, A39537-059, A39537-060, A39537-061, A39537-062, A39537-063, A39537-064, A39537-065, A39537-066, A39537-067, A39537-068, A39537-069, A39537-070, A39537-071, A39537-072, A39537-073, A39537-074, A39537-075, A39537-076, A39537-077, A39537-078, A39537-079, A39537-080, A39537-081, A39537-082, A39537-083, A39537-084, A39537-085, A39537-086, A39537-087, A39537-088, A39537-089, A39537-090, A39537-091, A39537-092, A39537-093, A39537-094, A39537-095, A39537-096, A39537-097, A39537-098, A39537-099, A39537-100, A39537-101, A39537-102, A39537-103, A39537-104, A39537-105, A39537-106, A39537-107, A39537-108, A39537-109, A39537-110, A39537-111, A39537-112, A39537-113, A39537-114, A39537-115, A39537-116, A39537-117, A39537-118, A39537-119, A39537-120, A39537-121, A39537-122, A39537-123, A39537-124, A39537-125, A39537-126, A39537-127, A39537-128, A39537-129, A39537-130, A39537-131, A39537-132, A39537-133, A39537-134, A39537-135, A39537-136, A39537-137, A39537-138, A39537-139, A39537-140, A39537-141, A39537-142, A39537-143, A39537-144, A39537-145, A39537-146, A39537-147, A39537-148, A39537-149, A39537-150, A39537-151, A39537-152, A39537-153, A39537-154, A40304-098, A40304-099, A40304-100, A40304-101, A40304-102, A40304-103, A40304-104, A40304-105, A40304-106, A40304-107, A40304-108, A40304-109, A40304-110, A40304-111, A40304-112, A40304-113, A40304-114, A40304-115, A40304-116, A40304-117, A40304-118, A40304-119, A40304-120, A40304-121, A40304-122, A40304-123, A40304-124, A40304-125, A40304-126, A40304-127, A40304-128, A40304-129, A40304-130, A40304-131, A40304-132, A40304-133, A40304-134, A40304-135, A40304-136, A40304-137, A40304-138, A40304-139, A40304-140, A40304-141, A40304-142, A40304-143, A40304-144, A40304-145, A40304-146, A40304-147, A40304-148, A40304-149, A40304-150, A40304-151, A40304-152, A40304-153, A40304-154, A40304-155, A40304-156, A40304-157, A40304-158, A40304-159, A40304-160, A40304-161, A40304-162, A40304-163, A40304-164, A40304-165, A40304-166, A40304-167, A40304-168, A40304-169, A40304-170, A40304-171, A40304-172, A40304-173, A40304-174, A40304-175, A40304-176, A40304-177, A40304-178, A40304-179, A40304-180, A40304-181, A40304-182, A40304-183, A40304-184, A40304-185, A40304-186, A40304-187, A40304-188, A40304-189, A40304-190, A40304-191, A40304-192, A40304-193, A40304-194, A40304-195, A40304-196, A40304-197, A40304-198, A40304-199, A40304-200, A40304-201, A40304-202, A40304-203, A40304-204, A40304-205, A40304-206, A40304-207, A40304-208, A40304-209, A40738-155, A40738-156, A40738-157, A40738-158, A40738-159, A40738-160, A40738-161, A40738-162, A40738-163, A40738-164, A40738-165, A40738-166, A40738-167, A40738-168, A40738-169, A40738-170, A40738-171, A40738-172, A40738-173, A40738-174, A40738-175, A40738-176, A40738-177, A40738-178, A40738-179, A40738-180, A40738-181, A40738-182, A40738-183, A40738-184, A40738-185, A40738-186, A40738-187, A40738-188, A40738-189, A40738-190, A40738-191, A40754-031, A40754-032, A40754-033, A40754-034, A40754-035, A40754-036, A40754-037, A40754-038, A40754-039, A40754-040, A40754-041, A40754-042, A40754-043, A40754-044, A40754-045, A40754-046, A40754-047, A40754-048, A40754-049, A40754-050, A40754-051, A40754-052, A40754-053, A40754-054, A40754-055, A40754-056, A40754-057, A40754-058, A40754-059, A40754-060, A40754-061, A40754-062, A40754-063, A40754-064, A40754-065, A40754-066, A40754-067, A40754-068, A40754-069, A40754-070, A40754-071, A40754-072, A40754-073, A40754-074, A40754-075, A40754-076, A40754-077, A40754-078, A40754-079, A40754-080, A40754-081, A40754-082, A40754-083, A40754-084, A40754-085, A40754-086, A40754-087, A40754-088, A40754-089, A40754-090, A40754-091, A40754-092, A40754-093, A40754-094, A40754-095, A40754-096, A40754-097, A40754-098, A40754-099, A40754-100, A40754-101, A40754-102, A40754-103, A40754-104, A40754-105, A40754-106, A40754-107, A40754-108, A41001-021, A41001-022, A41001-023, A41001-024, A41001-025, A41001-026, A41001-027, A41001-028, A41001-029, A41001-030, A41001-031, A41001-032, A41001-033, A41001-034, A41001-035, A41001-036, A41001-037, A41001-038, A41001-039, A41001-040, A41001-041, A41001-042, A41001-043, A41001-044, A41001-045, A41001-046, A41001-047, A41001-048, A41001-049, A41001-050, A41001-051, A41001-052, A41001-053, A41001-054, A41001-055, A41001-057, A41001-058, A41001-059, A41001-060, A41001-061, A41001-062, A41001-063, A41001-064, A41001-065, A41001-066, A41001-067, A41001-068, A41001-069, A41001-070, A41001-071, A41001-072, A41001-073, A41001-074, A41001-084, A41001-085, A41001-086, A41001-087, A41001-088, A41001-089, A41001-090, A41001-091, A41001-092, A41001-093, A41001-095, A41001-096, A41001-097, A41001-098, A41001-099, A41001-100, A41001-101, A41001-102, A41001-103, A41001-104, A41001-105, A41001-106, A41001-107, A41001-108, A41001-109, A41001-110, A41001-111, A41001-112, A41001-113, A41001-114, A41001-115, A41001-116, A41001-117, A41001-118, A41001-119, A41001-120, A41001-121, A41001-122, A41001-123, A41001-124, A41001-125, A41001-126, A41001-127, A41001-128, A41001-129, A41001-130, A41001-131, A41001-132, A41001-133, A41001-134, A41001-135, A41001-136, A41001-137, A41001-138, A41001-139, A41001-140, A41001-141, A41001-142, A41001-143, A41001-144, A41001-145, A41001-146, A41001-147, A41001-148, A41001-149, A41001-150, A41001-151, A41001-152, A41001-153, A41001-154, A41001-155, A41001-156, A41001-157, A41001-158, A41001-159, A41001-160, A41001-161, A41001-162, A41001-163, A41001-164, A41001-165, A41001-166, A41001-167, A41001-168, A41001-169, A41001-170, A41001-171, A41001-172, A41001-173, A41001-174, A41001-175, A41001-176, A41001-177, A41001-178, A41001-179, A41001-180, A41001-181, A41001-182, A41001-183, A41001-184, A41001-185, A41001-186, A41001-187, A41001-188, A41001-189, A41001-190, A41001-191, A41001-192, A41001-193, A41001-194, A41001-195, A41001-196, A41001-197, A41001-198, A41001-199, A41001-200, A41007-101, A41007-102, A41007-103, A41007-104, A41007-105, A41007-106, A41007-107, A41007-108, A41007-109, A41007-110, A41007-111, A41007-112, A41007-113, A41007-114, A41007-115, A41007-116, A41007-117, A41007-118, A41007-119, A41007-120, A41007-121, A41007-122, A41007-123, A41007-124, A41007-125, A41007-126, A41007-127, A41007-128, A41007-129, A41007-130, A41007-131, A41007-132, A41007-133, A41007-134, A41007-135, A41007-136, A41007-137, A41007-138, A41007-139, A41007-140, A41007-141, A41007-142, A41007-143, A41007-144, A41007-145, A41007-146, A41007-147, A41007-148, A41007-149, A41007-150, A41007-151, A41007-152, A41007-153, A41007-154, A41007-155, A41007-156, A41007-157, A41007-158, A41007-159, A41007-160, A41007-161, A41007-162, A41007-163, A41007-164, A41007-165, A41007-166, A41007-167, A41007-168, A41007-169, A41007-170, A41007-171, A41007-172, A41007-173, A41007-174, A41007-175, A41007-176, A41007-177, A41007-178, A41007-179, A41007-180, A41007-181, A41007-182, A41007-183, A41007-184, A41007-185, A41007-186, A41007-187, A41007-188, A41007-189, A41007-190, A41007-191, A41007-192, A41007-193, A41007-194, A41007-195, A41007-196, A41007-197, A41007-198, A41007-199, A41007-200, A41007-201, A41007-202, A41008-036, A41008-037, A41008-038, A41008-039, A41008-040, A41008-041, A41008-042, A41008-043, A41008-044, A41008-045, A41008-046, A41008-047, A41008-048, A41008-049, A41008-050, A41008-051, A41008-052, A41008-053, A41008-054, A41008-055, A41008-056, A41008-057, A41008-058, A41008-059, A41008-060, A41008-061, A41008-062, A41008-063, A41008-064, A41008-065, A41008-066, A41008-067, A41008-068, A41008-069, A41008-070, A41008-071, A41008-072, A41008-073, A41008-074, A41008-075, A41008-076, A41008-077, A41008-078, A41008-079, A41008-080, A41008-081, A41008-082, A41008-083, A41008-084, A41008-085, A41008-086, A41008-087, A41008-088, A41008-089, A41008-090, A41008-091, A41008-092, A41008-093, A41008-094, A41008-095, A41008-096, A41008-097, A41008-098, A41008-099, A41008-100, A41008-101, A41008-102, A41008-103, A41008-104, A41008-105, A41008-106, A41008-107, A41008-108, A41008-109, A41008-110, A41008-111, A41008-112, A41008-113, A41008-114, A41008-115, A41008-116, A41008-117, A41008-118, A41008-119, A41008-120, A41008-121, A41008-122, A41008-123, A41008-124, A41008-125, A41312-042, A41312-043, A41312-044, A41312-045, A41312-046, A41312-047, A41312-048, A41312-049, A41312-050, A41312-051, A41312-052, A41312-053, A41312-054, A41312-055, A41312-056, A41312-057, A41312-058, A41312-059, A41312-060, A41312-061, A41312-062, A41312-063, A41312-064, A41312-065, A41312-066, A41312-067, A41312-068, A41312-069, A41312-070, A41312-071, A41312-072, A41312-073, A41312-074, A41312-075, A41312-076, A41312-077, A41312-078, A41312-079, A41312-080, A41312-081, A41312-082, A41312-083, A41312-084, A41312-085, A41312-086, A41312-087, A41312-088, A41312-089, A41312-090, A41312-091, A41312-092, A41312-093, A41312-094, A41312-095, A41312-096, A41312-097, A41312-098, A41312-099, A41312-100, A41312-101, A41312-102, A41312-103, A41312-104, A41312-105, A41312-106, A41312-107, A41312-108, A41312-109, A41312-110, A41312-111, A41312-112, A41312-113, A41312-114, A41312-115, A41312-116, A41312-117, A41312-118, A41312-119, A41312-120, A41312-121, A41312-122, A41312-123, A41312-124, A41312-125, A41312-126, A41312-127, A41312-128, A41312-129, A41312-130, A41312-131, A41312-132, A41312-133, A41312-134, A41312-135, A41312-136, A41312-137, A41312-138, A41312-139, A41312-140, A41312-141, A41312-142, A41312-143, A41312-144, A41312-145, A41312-146, A41312-147, A41312-148, A41312-149, A41312-150, A41312-151, A41544-036, A41544-037, A41544-038, A41544-039, A41544-040, A41544-041, A41544-042, A41544-043, A41544-044, A41544-045, A41544-046, A41544-047, A41544-048, A41544-049, A41544-050, A41544-051, A41544-052, A41544-053, A41544-054, A41544-055, A41544-056, A41544-057, A41544-058, A41544-059, A41544-060, A41544-061, A41544-062, A41544-063, A41544-064, A41544-066, A41544-067, A41544-068, A41544-069, A41544-070, A41544-071, A41544-072, A41544-073, A41544-074, A41544-075, A41544-076, A41544-077, A41544-078, A41544-079, A41544-080, A41544-081, A41544-082, A41544-083, A41544-084, A41544-085, A41544-086, A41544-087, A41544-088, A41544-089, A41544-090, A41544-091, A41544-092, A41544-093, A41544-094, A41544-095, A41544-096, A41544-097, A41544-098, A41544-099, A41544-100, A41544-101, A41544-102, A41544-103, A41544-104, A41544-105, A41544-106, A41544-107, A41544-108, A41544-109, A41544-110, A41544-111, A41544-112, A41544-113, A41544-114, A41544-115, A41544-116, A41544-117, A41544-118, A41544-119, A41544-120, A41544-121, A41544-122, A41544-123, A41544-124, A41544-125, A41544-126, A41544-127, A41544-128, A41544-129, A41544-130, A41544-131, A41544-132, A41544-133, A41544-134, A41544-135, A41544-136, A41544-137, A41627-021, A41627-022, A41627-023, A41627-024, A41627-025, A41627-026, A41627-027, A41627-028, A41627-029, A41627-030, A41627-031, A41627-032, A41627-033, A41627-034, A41627-035, A41627-036, A41627-037, A41627-039, A42095-031, A42095-032, A42095-033, A42095-034, A42095-035, A42095-036, A42095-037, A42095-038, A42095-039, A42095-040, A42095-041, A42095-042, A42095-043, A42095-044, A42095-045, A42095-046, A42095-047, A42095-048, A42095-049, A42095-050, A42095-051, A42095-052, A42095-053, A42095-054, A42095-055, A42095-056, A42095-057, A42095-058, A42095-059, A42095-060, A42095-061, A42095-062, A42095-063, A42095-064, A42095-065, A42095-066, A42095-067, A42095-068, A42095-069, A42095-070, A42095-071, A42095-072, A42095-073, A42095-074, A42095-075, A42095-076, A42095-077, A42095-078, A42095-079, A42095-080, A42095-081, A42095-082, A42095-083, A42095-084, A42095-085, A42095-086, A42095-087, A42095-088, A42095-089, A42095-090, A42095-091, A42095-092, A42095-093, A42095-094, A42095-095, A42095-096, A42095-097, A42095-098, A42095-099, A42095-100, A42095-101, A42095-102, A42095-103, A42095-104, A42484-035, A42484-036, A42484-037, A42484-038, A42484-039, A42484-040, A42484-041, A42484-042, A42484-043, A42484-044, A42484-045, A42484-046, A42484-047, A42484-048, A42484-049, A42484-050, A42484-051, A42484-052, A42484-053, A42484-054, A42484-055, A42484-056, A42484-057, A42484-058, A42484-059, A42484-060, A42484-061, A42484-062, A42484-063, A42484-064, A42484-065, A42484-066, A42484-067, A42484-068, A42484-069, A42484-070, A42484-071, A42484-072, A42484-073, A42484-074, A42484-075, A42484-076, A42484-077, A42545-011, A42545-012, A42545-013, A42929-036, A42929-037, A42929-038, A42929-039, A42929-040, A42929-041, A42929-042, A42929-043, A42929-044, A42929-045, A42929-046, A42929-047, A42929-048, A42929-049, A42929-050, A42929-051, A42929-052, A42929-053, A42929-054, A42929-055, A42929-056, A42929-057, A42929-058, A42929-059, A42929-060, A42929-061, A42929-062, A42929-063, A42929-064, A42929-065, A42929-066, A42929-067, A42929-068, A42929-069, A42929-070, A42929-071, A42929-072, A42929-073, A42929-074, A42929-075, A42929-076, A42929-077, A42929-078, A42929-079, A42929-080, A42929-081, A42929-082, A42929-083, A42929-084, A42929-085, A42929-086, A42929-087, A42929-088, A42929-089, A42929-090, A42929-091, A42929-092, A42929-093, A42929-094, A42929-095, A42929-096, A42929-097, A42929-098, A42929-099, A42929-100, A42929-101, A42929-102, A42929-103, A42929-104, A42929-105, A42929-106, A42929-107, A42929-108, A42929-109, A42929-110, A42929-112, A42929-113, A42929-114, A42929-115, A42929-116, A42929-117, A42929-118, A42929-119, A42929-120, A42929-121, A42929-122, A42929-123, A42929-124, A42929-125, A42929-126, A42929-127, A42929-128, A42929-129, A42929-130, A42929-131, A42929-132, A42929-133, A42929-134, A42929-135, A42929-136, A42929-137, A42929-138, A42982-095, A42982-096, A42982-097, A42982-098, A42982-099, A42982-100, A42982-101, A42982-102, A42982-103, A42982-104, A42982-105, A42982-106, A43284-047, A43284-048, A43284-049, A43284-050, A43284-051, A43284-052, A43284-053, A43284-054, A43284-055, A43284-056, A43284-057, A43284-058, A43284-059, A43284-060, A43284-061, A43284-062, A43284-063, A43284-064, A43284-065, A43284-066, A43284-067, A43284-068, A43284-069, A43284-070, A43284-071, A43284-072, A43284-073, A43284-074, A43284-075, A43284-076, A43284-077, A43284-078, A43284-079, A43284-080, A43284-081, A43284-082, A43284-083, A43284-084, A43284-085, A43284-086, A43284-087, A43284-088, A43284-089, A43284-090, A43284-091, A43284-092, A43284-093, A43284-094, A43284-095, A43284-096, A43284-097, A43284-098, A43284-099, A43284-100, A43284-101, A43284-102, A43284-103, A43284-104, A43284-105, A43284-106, A43284-107, A43284-108, A43284-109, A43284-110, A43284-111, A43284-112, A43284-113, A43284-114, A43284-115, A43284-116, A43284-117, A43284-118, A43284-119, A43284-120, A43284-121, A43284-122, A43284-123, A43284-124, A43284-125, A43336-118, A43336-119, A43336-120, A43336-121, A43336-122, A43336-123, A43336-124, A43336-125, A44080-047, A44080-048, A44080-049, A44080-050, A44080-051, A44080-052, A44080-053, A44080-054, A44080-055, A44080-056, A44080-057, A44080-058, A44080-059, A44080-060, A44080-061, A44080-062, A44080-063, A44080-064, A44080-065, A44080-066, A44080-067, A44080-068, A44080-069, A44080-070, A44080-071, A44080-072, A44080-073, A44080-074, A44080-075, A44080-076, A44080-077, A44080-078, A44080-079, A44080-080, A44080-081, A44080-082, A44080-083, A44080-084, A44080-085, A44080-086, A44080-087, A44080-088, A44080-089, A44080-090, A44080-091, A44080-092, A44080-093, A44080-094, A44080-095, A44080-096, A44080-097, A44080-098, A44080-099, A44080-100, A44405-066, A44405-067, A44405-068, A44405-069, A44405-070, A44405-071, A44405-072, A44405-073, A44405-074, A44405-075, A44405-076, A44405-077, A44405-078, A44405-079, A44405-080, A44405-081, A44405-082, A44405-083, A44405-084, A44405-085, A44405-086, A44405-087, A44405-088, A44405-089, A44405-090, A44405-091, A44405-092, A44405-093, A44405-094, A44405-095, A44405-096, A44405-097, A44405-098, A44405-099, A44405-100, A44405-101, A44405-102, A44405-103, A44405-104, A44405-105, A44405-106, A44405-107, A44405-108, A44413-189, A44413-190, A44413-191, A44413-192, A44413-193, A44413-194, A44413-195, A44413-196, A44413-197, A44413-198, A44413-199, A44413-200, A44413-201, A44413-202, A44413-203, A44413-204, A44413-205, A44413-206, A44413-207, A44413-208, A44413-209, A44413-210, A44413-211, A44413-212, A44413-213, A44413-214, A44413-215, A44413-216, A44413-217, A44413-218, A44413-219, A44413-220, A44413-221, A44413-222, A44413-223, A44753-031, A44753-032, A44753-033, A44753-034, A44753-035, A44753-036, A44753-037, A44753-038, A44753-039, A44753-040, A44753-041, A44753-042, A44753-043, A44753-044, A44753-045, A44753-046, A44753-047, A44753-048, A44753-049, A44753-050, A44753-051, A44753-052, A44753-053, A44753-054, A44753-055, A44753-056, A44753-057, A44753-058, A44753-059, A44753-060, A44753-061, A44753-062, A44753-063, A44753-064, A44753-065, A44753-066, A44753-067, A44753-068, A44753-069, A44753-070, A44753-071, A44753-072, A44753-073, A44753-074, A44753-075, A44753-076, A44753-077, A44753-078, A44753-079, A44753-080, A44753-081, A44753-082, A44753-083, A44753-084, A44753-085, A44753-086, A44753-087, A44753-088, A44753-089, A44753-090, A44753-091, A44753-092, A44753-093, A44753-094, A44753-095, A44753-096, A44753-097, A44753-098, A44753-099, A44753-100, A44753-101, A44753-102, A44753-103, A44753-104, A44753-105, A44753-106, A44753-107, A44753-108, A44753-109, A44753-110, A44753-111, A44753-112, A44753-113, A44753-114, A44753-115, A44753-116, A44753-117, A44753-118, A44753-119, A44753-120, A44753-121, A44753-122, A44753-123, A44753-124, A44753-125, A44753-126, A44753-127, A44753-128, A44753-129, A44753-130, A44753-131, A44753-132, A44753-133, A44753-134, A44753-135, A45946-026, A45946-027, A45946-028, A45946-029, A45946-030, A45946-031, A45946-032, A45946-033, A45946-034, A45946-035, A45946-036, A45946-037, A45946-038, A45946-039, A45946-040, A45946-041, A45946-042, A45946-043, A45946-044, A45946-045, A45946-046, A45946-047, A45946-048, A45946-049, A45946-050, A45946-051, A45946-052, A45946-053, A45946-054, A45946-055, A45946-056, A45946-057, A45946-058, A45946-059, A45946-060, A45946-061, A45946-062, A45946-063, A45946-064, A45946-065, A45946-066, A45946-067, A45946-068, A45946-069, A45946-070, A45946-071, A45946-072, A45946-073, A45946-074, A45946-075, A45946-076, A45946-077, A45946-078, A45946-079, A45946-080, A45946-081, A45946-082, A45946-083, A45946-084, A45946-085, A45946-086, A45946-087, A45946-088, A45946-089, A45946-090, A45946-091, A45946-092, A45946-093, A45946-094, A45946-095, A45946-096, A45946-097, A45946-098, A45946-099, A45946-100, A45946-101, A45946-102, A45946-103, A45946-104, A45946-105, A45946-106, A45946-107, A45946-108, A45946-109, A45946-110, A46149-046, A46149-047, A46149-048, A46149-049, A46149-050, A46149-051, A46149-052, A46149-053, A46149-054, A46149-055, A46149-056, A46149-057, A46149-058, A46149-059, A46149-060, A46149-061, A46149-062, A46149-063, A46149-064, A46149-065, A46149-066, A46149-067, A46149-068, A46149-069, A46149-070, A46149-071, A46149-072, A46149-073, A46149-074, A46149-075, A46149-076, A46149-077, A46149-078, A46149-079, A46149-080, A46149-081, A46149-082, A46149-083, A46149-084, A46149-085, A46149-086, A46149-087, A46149-088, A46149-089, A46149-090, A46149-091, A46149-092, A46149-093, A46149-094, A46149-095, A46149-096, A46149-097, A46149-098, A46149-099, A46149-100, A46149-101, A46149-102, A46149-103, A46149-104, A46149-105, A46149-106, A46149-107, A46149-108, A46149-109, A46149-110, A46149-111, A46149-112, A46149-113, A46149-114, A46149-115, A46149-116, A46149-117, A46149-118, A46149-119, A46149-120, A46149-121, A46149-122, A46149-123, A46149-124, A46149-125, A46149-126, A46149-127, A46149-128, A46149-129, A46149-130, A46149-131, A46149-132, A46149-133, A46149-134, A46149-135, A46149-136, A46149-137, A46149-138, A46149-139, A46149-140, A46149-141, A46149-142, A46149-143, A46333-054, A46333-055, A46333-056, A46333-057, A46333-058, A46333-059, A46333-060, A46333-061, A46333-062, A46333-063, A46333-064, A46333-065, A46333-066, A46333-067, A46333-068, A46333-069, A46333-070, A46333-071, A46333-072, A46333-073, A46333-074, A46333-075, A46333-076, A46333-077, A46333-078, A46333-079, A46333-080, A46333-081, A46333-082, A46791-028, A46791-029, A46791-030, A46791-031, A46791-032, A46791-033, A46791-034, A46791-035, A46791-036, A46791-037, A46791-038, A46791-039, A46791-040, A46791-041, A46791-042, A46791-043, A46791-044, A46791-045, A46791-046, A46791-047, A46791-048, A46791-049, A46791-050, A46791-051, A46791-052, A46791-053, A46791-054, A46791-055, A46791-056, A46791-057, A46791-058, A46791-059, A46791-060, A46791-061, A46791-062, A46791-063, A46791-064, A46791-065, A46791-066, A46791-067, A46791-068, A46791-069, A46791-070, A46791-071, A46791-072, A46791-073, A46791-074, A46791-075, A46791-076, A46791-077, A46791-078, A46791-080, A46791-081, A46791-082, A46791-083, A46791-084, A46791-085, A46791-086, A46791-087, A46791-088, A46791-089, A46791-090, A46791-091, A46791-092, A46791-093, A46791-094, A46791-095, A46791-096, A46791-097, A46791-098, A46791-099, A46791-100, A46791-101, A46791-102, A46791-103, A46791-104, A46791-105, A46791-106, A46791-107, A46791-108, A46791-109, A46791-110, A46791-111, A46791-112, A46791-113, A46791-114, A46791-115, A46791-116, A46791-117, A46791-118, A46791-119, A46791-120, A46791-121, A46791-122, A46791-123, A46791-124, A46791-125, A46791-126, A46791-127, A46791-128, A46791-129, A46791-130, A46791-131, A46791-132, A46791-133, A46791-134, A46791-135, A46791-136, A46791-137, A46791-138, A46791-139, A46791-140, A46791-141, A46791-142, A46791-143, A46791-144, A46791-145, A46791-146, A46791-147, A46791-148, A46791-149, A46791-150, A46791-151, A46791-152, A46846-036, A46846-037, A46846-038, A46846-039, A46846-040, A46846-041, A46846-042, A46846-043, A46846-044, A46846-045, A46846-046, A46846-047, A46846-048, A46846-049, A46846-050, A46846-051, A46846-052, A46846-053, A46846-054, A46846-055, A46846-056, A46846-057, A46846-058, A46846-059, A46846-060, A46846-061, A46846-062, A46846-063, A46846-064, A46846-065, A46846-066, A46849-096, A46849-097, A46849-098, A46849-099, A46849-100, A46849-101, A46849-102, A46849-103, A46849-104, A46849-105, A46849-106, A46849-107, A46849-108, A46849-109, A46849-110, A46849-111, A46849-112, A46849-113, A46849-114, A46849-115, A47027-097, A47027-098, A47027-099, A47027-100, A47027-101, A47027-102, A47027-103, A47027-104, A47027-105, A47027-106, A47027-107, A47027-108, A47027-109, A47027-110, A47027-111, A47027-112, A47027-113, A47027-114, A47027-115, A47027-116, A47027-117, A47027-118, A47027-119, A47027-120, A47027-121, A47027-122, A47027-123, A47027-124, A47027-125, A47027-126, A47027-127, A47027-128, A47027-129, A47027-130, A47027-131, A47027-132, A47027-133, A47027-134, A47027-135, A47027-136, A47027-137, A47027-138, A47027-139, A47027-140, A47027-141, A47027-142, A47027-143, A47027-144, A47027-145, A47027-146, A47027-147, A47027-148, A47027-149, A47027-150, A47027-151, A47027-152, A47027-153, A47027-154, A47027-155, A47027-156, A47027-157, A47027-158, A47027-159, A47027-160, A47027-161, A47027-162, A47027-163, A47027-164, A47027-165, A47027-166, A47027-167, A47027-168, A47027-169, A47027-170, A47027-171, A47027-172, A47027-173, A47027-174, A47027-175, A47027-176, A47027-177, A47027-178, A47027-179, A47027-180, A47027-181, A47027-182, A47027-183, A47027-184, A47027-185, A47027-186, A47027-187, A47027-188, A47027-189, A47027-190, A47027-191, A47027-192, A47027-193, A47027-194, A47027-195, A47027-196, A47027-197, A47027-198, A47027-199, A47027-200, A47027-201, A47027-202, A47167-011, A47167-012, A47167-013, A47167-014, A47167-015, A47167-016, A47167-017, A47167-018, A47167-019, A47167-020, A47167-021, A47167-022, A47167-023, A47167-024, A47167-025, A47167-026, A47167-027, A47167-028, A47167-029, A47167-030, A47167-031, A47167-032, A47167-033, A47167-034, A47167-035, A47167-036, A47167-037, A47167-038, A47167-039, A47167-040, A47167-041, A47167-042, A47167-043, A47167-044, A47167-045, A47167-046, A47167-047, A47167-048, A47167-049, A47167-050, A47167-051, A47167-052, A47167-053, A47167-054, A47167-055, A47167-056, A47167-057, A47167-058, A47167-059, A47167-060, A47167-061, A47167-062, A47167-063, A47167-064, A47167-065, A47167-066, A47167-067, A47167-068, A47167-069, A47167-070, A47167-071, A47167-072, A47167-073, A47167-074, A47167-075, A47167-076, A47167-077, A47167-078, A47167-079, A47167-080, A47167-081, A47167-082, A47433-106, A47433-107, A47433-108, A47433-109, A47433-110, A47433-111, A47433-112, A47433-113, A47433-114, A47433-115, A47433-116, A47433-117, A47433-118, A47433-119, A47433-120, A47433-121, A47433-122, A47433-123, A47433-124, A47433-125, A47433-126, A47433-127, A47433-128, A47433-129, A47433-130, A47433-131, A47433-132, A47433-133, A47433-134, A47433-135, A47433-136, A47433-137, A47433-138, A47433-139, A47433-140, A47433-141, A47433-142, A47433-143, A47433-144, A47482-092, A47482-093, A47482-094, A47482-095, A47482-096, A47482-097, A47482-098, A47482-099, A47482-100, A47482-101, A47482-102, A47482-103, A47482-104, A47482-105, A47482-106, A47482-107, A47482-108, A47482-109, A47993-080, A47993-081, A47993-082, A47993-083, A47993-084, A47993-085, A47993-086, A47993-087, A47993-088, A47993-089, A47993-090, A47993-091, A47993-092, A47993-093, A47993-094, A47993-095, A47993-096, A47993-097, A47993-098, A47993-099, A47993-100, A47993-101, A47993-102, A47993-103, A47993-104, A47993-105, A47993-106, A47993-107, A47993-108, A47993-109, A47993-110, A47993-111, A47993-112, A47993-113, A47993-114, A47993-115, A47993-116, A47993-117, A47993-118, A47993-119, A47993-120, A47993-121, A47993-122, A47993-123, A47993-124, A47993-125, A47993-126, A47993-127, A47993-128, A47993-129, A47993-130, A47993-131, A47993-132, A47993-133, A47993-134, A48182-011, A48182-012, A48182-013, A48182-014, A48182-015, A48182-016, A48182-017, A48182-018, A48182-019, A48182-020, A48182-021, A48182-022, A48182-023, A48182-024, A48182-025, A48182-026, A48182-027, A48182-028, A48182-029, A48182-030, A48182-031, A48182-032, A48182-033, A48182-034, A48182-035, A48182-036, A48182-037, A48182-038, A48182-039, A48182-040, A48182-041, A48182-042, A48182-043, A48182-044, A48182-045, A48182-046, A48182-047, A48182-048, A48182-049, A48182-050, A48182-051, A48182-052, A48182-053, A48182-054, A48182-055, A48182-056, A48182-057, A48182-058, A48182-059, A48182-060, A48182-061, A48182-062, A48182-063, A48182-064, A48182-065, A48182-066, A48182-067, A48182-068, A48182-069, A48182-070, A48182-071, A48182-072, A48182-073, A48182-074, A48182-075, A48182-076, A48182-077, A48182-078, A48182-079, A48182-080, A48182-081, A48182-082, A48182-083, A48182-084, A48182-085, A48182-086, A48182-087, A48182-088, A48182-089, A48182-090, A48182-091, A48182-092, A48182-093, A48182-094, A48182-095, A48182-096, A48182-097, A48182-098, A48182-099, A48182-100, A48182-101, A48182-102, A48182-103, A48182-104, A48182-105, A48182-106, A48182-107, A48182-108, A48182-109, A48182-110, A48182-111, A48182-112, A48182-113, A48182-114, A48182-115, A48182-116, A48182-117, A48182-118, A48182-119, A48182-120, A48182-121, A48182-122, A48182-123, A48182-124, A48182-125, A48182-126, A48182-127, A48182-128, A48182-129, A48182-130, A48182-131, A48182-132, A48182-133, A48182-134, A48182-135, A48182-136, A48182-137, A48182-138, A48182-139, A48182-140, A48182-141, A48182-142, A48182-143, A48182-144, A48182-145, A48182-146, A48182-147, A48182-148, A48182-149, A48182-150, A48182-151, A48182-152, A48182-153, A48182-154, A48182-155, A48182-156, A48182-157, A48182-158, A48182-159, A48182-160, A48182-161, A48188-057, A48188-058, A48188-059, A48188-060, A48188-061, A48188-062, A48188-063, A48188-064, A48188-065, A48188-066, A48188-067, A48188-068, A48188-070, A48188-071, A48188-072, A48188-073, A48188-074, A48188-075, A48188-076, A48188-077, A48188-078, A48188-079, A48188-080, A48188-081, A48188-082, A48188-083, A48188-084, A48188-085, A48188-086, A48188-087, A48188-088, A48188-089, A48188-090, A48188-091, A48188-092, A48188-093, A48188-094, A48188-095, A48188-096, A48188-097, A48188-098, A48188-099, A48188-100, A48188-101, A48188-102, A48188-103, A48188-104, A48188-105, A48188-106, A48188-107, A48871-051, A48871-052, A48871-053, A48871-054, A48871-055, A48871-056, A48871-057, A48871-058, A48871-059, A48871-060, A48871-061, A48871-062, A48871-063, A48871-064, A48871-066, A48871-067, A48871-068, A48871-069, A48871-070, A48871-071, A48871-072, A48871-073, A48871-074, A48871-075, A48871-076, A48871-077, A48871-078, A48871-079, A48871-080, A48871-081, A48871-082, A48871-083, A48871-084, A48871-085, A48871-086, A48871-087, A48871-088, A48871-089, A48871-090, A48871-091, A48871-092, A48871-093, A48871-094, A48871-095, A48871-096, A48871-097, A48871-098, A48871-099, A48871-100, A48871-101, A48871-102, A48871-103, A48871-104, A48871-105, A48871-106, A48871-107, A48871-108, A48871-109, A48871-110, A48871-111, A48871-112, A48871-113, A48871-114, A48871-115, A48871-116, A48871-117, A48871-118, A48871-119, A48871-120, A48871-121, A48871-122, A48871-123, A48871-124, A48871-125, A48871-126, A48871-127, A48871-128, A48871-129, A48871-130, A48871-131, A48871-132, A48871-133, A48871-134, A48871-135, A48871-136, A48871-137, A48871-138, A48871-139, A48871-140, A49002-052, A49002-053, A49002-054, A49002-055, A49002-056, A49002-057, A49002-058, A49002-059, A49002-060, A49002-061, A49002-062, A49002-063, A49002-064, A49002-065, A49002-066, A49002-067, A49002-068, A49002-069, A49002-070, A49002-071, A49002-072, A49002-073, A49529-030, A49529-031, A49529-032, A49529-033, A49529-034, A49529-035, A49529-036, A49529-037, A49529-038, A49529-039, A49529-040, A49529-041, A49529-042, A49529-043, A49529-044, A49529-045, A49529-046, A49529-047, A49529-048, A49529-049, A49529-050, A49529-051, A49529-052, A49529-053, A49529-054, A49529-055, A49529-056, A49529-057, A49529-058, A49529-059, A49529-060, A49529-061, A49529-062, A49529-063, A49529-064, A49529-065, A49529-066, A49529-067, A49529-068, A49529-069, A49529-070, A49529-071, A49529-072, A49529-073, A49529-074, A49529-075, A49529-076, A49529-077, A49529-078, A49529-079, A49529-080, A49529-081, A49529-082, A49529-083, A49529-084, A49529-085, A49529-086, A49529-087, A49529-088, A49707-064, A49707-065, A49707-066, A49707-067, A49707-068, A49707-069, A49707-070, A49707-071, A49707-072, A49707-073, A49707-074, A49707-075, A49707-076, A49707-077, A49707-078, A49707-079, A49707-080, A49707-081, A49707-082, A49707-083, A49707-084, A49707-085, A49707-086, A49707-087, A49707-088, A49707-089, A49707-090, A49707-091, A49707-092, A49707-093, A49707-094, A49707-095, A49707-096, A49707-097, A49707-098, A49707-099, A49707-100, A49707-101, A49707-102, A49707-103, A49707-104, A49707-105, A49707-106, A49707-107, A49707-108, A49707-109, A49707-110, A49707-111, A49707-112, A49707-113, A49707-114, A49707-115, A49707-116, A49707-117, A49707-118, A49707-119, A49707-120, A49707-121, A49707-122, A49707-123, A49707-124, A49707-125, A49707-126, A49707-127, A49707-128, A49707-129, A49707-130, A49707-131, A49707-132, A49707-133, A49707-134, A49707-135, A49707-136, A49707-137, A49707-138, A49707-139, A49707-140, A49707-141, A49707-142, A49707-143, A49707-144, A49707-145, A49707-146, A49707-147, A49707-148, A49707-149, A49707-150, A49707-151, A49707-152, A49707-153, A49707-154, A49707-155, A49707-156, A49707-157, A49707-158, A49707-159, A49707-160, A49707-161, A49707-162, A49707-163, A49707-164, A49707-165, A49707-166, A49707-167, A49707-168, A49707-169, A49707-170, A49707-171, A49707-172, A49707-173, A49707-174, A49707-175, A49707-176, A49707-177, A49707-178, A49707-179, A49707-180, A49707-181, A49707-182, A49707-183, A49707-184, A50166-069, A50166-070, A50166-071, A50166-072, A50166-073, A50166-074, A50166-075, A50166-076, A50166-077, A50166-078, A50166-079, A50166-080, A50166-081, A50166-082, A50166-083, A50166-084, A50166-085, A50166-086, A50166-087, A50166-088, A50166-089, A50166-090, A50166-091, A50166-092, A50166-093, A50166-094, A50166-095, A50166-096, A50166-097, A50166-098, A50166-099, A50166-100, A50166-101, A50166-102, A50166-103, A50166-104, A50166-105, A50166-106, A50166-107, A50166-108, A50166-109, A50910-029, A50910-030, A50910-031, A50910-032, A50910-033, A50910-034, A50910-035, A50910-036, A50910-037, A50910-038, A50910-039, A50910-040, A50910-041, A50910-042, A50910-043, A50910-044, A50910-045, A50910-046, A50910-047, A50910-048, A50910-049, A50910-050, A50910-051, A50910-052, A50910-053, A50910-054, A50910-055, A50910-056, A50910-057, A50910-058, A50910-059, A50910-060, A50910-061, A50910-062, A50910-063, A50910-064, A50910-065, A50910-068, A50910-069, A50910-070, A50910-071, A50910-072, A50910-073, A50910-074, A50910-075, A50910-076, A50910-077, A50910-078, A50910-079, A50910-080, A50910-081, A50910-082, A50910-083, A50910-084, A50910-085, A50910-086, A50910-087, A50910-088, A50910-089, A50910-090, A50910-091, A50910-092, A50910-093, A50910-094, A50910-095, A50910-096, A50910-097, A50910-098, A50910-099, A50910-100, A50910-101, A50910-102, A50910-103, A50910-104, A50910-105, A50910-106, A50910-107, A50910-108, A51167-041, A51167-042, A51167-043, A51167-044, A51167-045, A51167-046, A51167-047, A51167-048, A51167-049, A51167-050, A51167-051, A51167-052, A51167-053, A51167-054, A51167-055, A51167-056, A51167-057, A51167-058, A51167-059, A51167-060, A51167-061, A51167-062, A51167-063, A51167-064, A51167-065, A51167-066, A51167-067, A51167-068, A51167-069, A51167-070, A51167-071, A51167-072, A51167-073, A51167-074, A51167-075, A51167-076, A51167-077, A51167-078, A51167-079, A51167-080, A51167-081, A51167-082, A51167-083, A51167-084, A51167-085, A51167-086, A51167-087, A51167-088, A51167-089, A51167-090, A51167-091, A51167-092, A51167-093, A51167-094, A51167-095, A51167-096, A51167-097, A51167-098, A51167-099, A51167-100, A51167-101, A51167-102, A51167-103, A51167-104, A51167-105, A51167-106, A51167-107, A51167-108, A51167-109, A51597-011, A51597-012, A51597-013, A51597-014, A51597-015, A51597-016, A51597-017, A51597-018, A51597-019, A51597-020, A51597-021, A51597-022, A51597-023, A51597-024, A51597-025, A51597-026, A51597-027, A51597-028, A51597-029, A51597-030, A51597-031, A51597-032, A51597-033, A51597-034, A51597-035, A51597-036, A51597-037, A51597-038, A51597-039, A51597-040, A51597-041, A51597-042, A51597-043, A51597-044, A51597-045, A51597-046, A51597-047, A51597-048, A51597-049, A51597-050, A51597-051, A51597-052, A51597-053, A51597-054, A51597-055, A51597-056, A51597-057, A51597-058, A51597-059, A51597-060, A51597-061, A51597-062, A51597-063, A51597-064, A51597-065, A51597-066, A51597-067, A51597-068, A51597-069, A51597-070, A51597-071, A51597-072, A51597-073, A51597-074, A51597-075, A51597-076, A51597-077, A51597-078, A51597-079, A51597-080, A51597-081, A51597-082, A51597-083, A51597-084, A51597-085, A51597-086, A51597-087, A51597-088, A51597-089, A51847-011, A51847-012, A51847-013, A51847-014, A51847-015, A51847-016, A51847-017, A51847-018, A51847-019, A51847-020, A51847-021, A51847-022, A51847-023, A51847-024, A51847-025, A51847-026, A51847-027, A51847-028, A51847-029, A51847-030, A51847-031, A51847-032, A51847-033, A51847-034, A51847-035, A51847-036, A51847-037, A51847-038, A51847-039, A51847-040, A51847-041, A51847-042, A51847-043, A51847-044, A51847-045, A51847-046, A51847-047, A51847-048, A51847-049, A51847-050, A51847-051, A51847-052, A51847-053, A51847-054, A51847-055, A51847-056, A51847-057, A51847-058, A51946-011, A51946-012, A51946-013, A51946-014, A51946-015, A51946-016, A51946-017, A51946-018, A51946-019, A51946-020, A51946-021, A51946-022, A51946-023, A51946-024, A51946-025, A51946-026, A51946-027, A51946-028, A51946-029, A51946-030, A51946-031, A51946-032, A51946-033, A51946-034, A51946-035, A51946-036, A51946-037, A51946-038, A51946-039, A51946-040, A51946-041, A51946-042, A51946-043, A51946-044, A51946-045, A51946-046, A51946-047, A51946-048, A51946-049, A51946-050, A51946-051, A51946-052, A51946-053, A51946-054, A51946-055, A51946-056, A51946-057, A51946-058, A51946-059, A51946-060, A51946-061, A51946-062, A51946-063, A51946-064, A51946-065, A51946-066, A51946-067, A51946-068, A51946-069, A51946-070, A51946-071, A51946-072, A51946-073, A51946-074, A51946-075, A51946-076, A51946-077, A51946-078, A51946-079, A51946-080, A51946-081, A51946-082, A51946-083, A51946-084, A51946-085, A51946-086, A51946-087, A51946-088, A51946-089, A51946-090, A51946-091, A51946-092, A51946-093, A51946-094, A51946-095, A51946-096, A51946-097, A51946-098, A51946-099, A51946-100, A51946-101, A51946-102, A51946-103, A51970-011, A51970-012, A51970-013, A51970-014, A51970-015, A51970-016, A51970-017, A51970-018, A51970-019, A51970-020, A51970-021, A51970-022, A51970-023, A51970-024, A51970-025, A51970-026, A51970-027, A51970-028, A51970-029, A51970-030, A51970-031, A51970-032, A51970-033, A51970-034, A51970-035, A51970-036, A51970-037, A51970-038, A51970-039, A51970-040, A51970-041, A51970-042, A51970-043, A51970-044, A51970-045, A51970-046, A51970-047, A51970-048, A51970-049, A51970-050, A51970-051, A51970-052, A51970-053, A51970-054, A51970-055, A51970-056, A51970-057, A51970-058, A51970-059, A51970-060, A51970-061, A51970-062, A51970-063, A51970-064, A51970-065, A51970-066, A51970-077, A51970-078, A51970-079, A51970-080, A51970-081, A51970-082, A51970-083, A51970-084, A51970-085, A52212-049, A52212-050, A52212-051, A52212-052, A52212-053, A52212-054, A52212-055, A52212-056, A52212-057, A52212-058, A52212-059, A52212-060, A52212-061, A52212-062, A52212-063, A52212-064, A52212-065, A52212-066, A52212-067, A52212-068, A52212-069, A52212-070, A52212-071, A52212-072, A52212-073, A52212-074, A52212-075, A52212-076, A52212-077, A52212-078, A52212-079, A52212-080, A52212-081, A52212-082, A52212-083, A52212-084, A52212-085, A52212-086, A52212-087, A52212-088, A52212-089, A52212-090, A52212-091, A52212-092, A52212-093, A52212-094, A52212-095, A52212-096, A52212-097, A52212-098, A52212-099, A52212-100, A52212-101, A52212-102, A52212-103, A52212-104, A52212-105, A52212-106, A52212-107, A52212-108, A52212-109, A52212-110, A52212-111, A52212-112, A52212-113, A52212-114, A52212-115, A52212-116, A52212-117, A52212-118, A52212-119, A52212-120, A52212-121, A52212-122, A52212-123, A52212-124, A52212-125, A52212-126, A52212-127, A52212-128, A52212-129, A52212-130, A52212-131, A52212-132, A52212-133, A52212-134, A52212-135, A52212-136, A52212-137, A52212-138, A52212-139, A52212-140, A52212-141, A52212-142, A52212-143, A52212-144, A52212-145, A52212-146, A52212-147, A52212-148, A52275-039, A52275-040, A52275-041, A52275-042, A52275-043, A52275-044, A52275-045, A52275-046, A52275-047, A52275-048, A52275-049, A52275-050, A52275-051, A52275-052, A52275-053, A52275-054, A52275-055, A52275-056, A52275-057, A52275-058, A52275-059, A52275-060, A52275-061, A52275-062, A52275-063, A52275-077, A52275-078, A52275-079, A52275-080, A52275-081, A52275-082, A52275-083, A52275-084, A52275-085, A52275-086, A52275-087, A52275-088, A52860-021, A52860-022, A52860-023, A52860-024, A52860-025, A52860-026, A52860-027, A52860-028, A52860-029, A52860-030, A52860-031, A52860-032, A52860-033, A52860-034, A52860-035, A52860-036, A52860-037, A52860-038, A52860-039, A52860-040, A52860-041, A52860-042, A52860-043, A52860-044, A52860-045, A52860-046, A52860-047, A52860-048, A52860-049, A52860-050, A52860-051, A52860-052, A52860-053, A52860-054, A52860-055, A52860-056, A52860-057, A52860-058, A52860-059, A52860-060, A52860-061, A52860-062, A52860-063, A52860-064, A52860-065, A52860-066, A52860-067, A52860-068, A52860-069, A52860-070, A52860-071, A52860-072, A52860-073, A52860-074, A52860-075, A52860-076, A52860-077, A52860-078, A52860-079, A52860-080, A52860-081, A52860-082, A52860-083, A52860-084, A52860-085, A52860-086, A52860-087, A52860-088, A52860-089, A52860-090, A52860-091, A52860-092, A52860-093, A52860-094, A52860-095, A52860-096, A52860-097, A52860-098, A52860-099, A52860-100, A52860-101, A52860-102, A52860-103, A52860-104, A52860-105, A52860-106, A52860-107, A52860-108, A52860-109, A52860-110, A52860-111, A52860-112, A52860-113, A52860-114, A52860-115, A52860-116, A52860-117, A52860-118, A52860-119, A52860-120, A52860-121, A52860-122, A52860-123, A52860-124, A52860-125, A52860-126, A52860-127, A53118-082, A53118-083, A53118-084, A53118-085, A53118-086, A53118-087, A53118-088, A53118-089, A53118-090, A53118-091, A53118-092, A53118-093, A53118-094, A53118-095, A53118-096, A53118-097, A53118-098, A53118-099, A53118-100, A53118-101, A53118-102, A53118-103, A53118-104, A53118-105, A53118-106, A53118-107, A53118-108, A53118-109, A53118-110, A53118-111, A53118-112, A53118-113, A53118-114, A53118-115, A53118-116, A53118-117, A53118-118, A53118-119, A53118-120, A53118-121, A53118-122, A53118-123, A53118-124, A53118-125, A53118-126, A53118-127, A53118-128, A53118-129, A53118-130, A53118-131, A53118-132, A53118-133, A53118-134, A53118-135, A53118-136, A53118-137, A53118-138, A53118-139, A53118-140, A53118-141, A53118-142, A53118-143, A53118-144, A53118-145, A53118-146, A53118-147, A53118-148, A53118-149, A53118-150, A53118-151, A53118-152, A53118-153, A53118-154, A53118-155, A53118-156, A53118-157, A53118-158, A53118-159, A53118-160, A53118-161, A53118-162, A53118-163, A53118-164, A53118-165, A53118-166, A53118-167, A53177-027, A53177-028, A53177-029, A53177-030, A53177-031, A53177-032, A53177-033, A53177-034, A53177-035, A53177-036, A53177-037, A53177-038, A53177-039, A53177-040, A53177-041, A53177-042, A53177-043, A53177-044, A53177-045, A53177-046, A53177-047, A53177-048, A53177-049, A53177-050, A53177-051, A53177-052, A53177-053, A53177-054, A53177-055, A53177-062, A53177-063, A53177-064, A53177-065, A53177-066, A53177-067, A53177-068, A53177-069, A53177-070, A53177-071, A53177-072, A53177-073, A53177-074, A53177-075, A53177-076, A53177-077, A53177-078, A53177-079, A53177-080, A53177-081, A53177-082, A53177-083, A53177-084, A53177-085, A53177-086, A53177-087, A53177-088, A53177-089, A53177-090, A53177-091, A53177-092, A53177-093, A53177-094, A53177-095, A53177-096, A53177-097, A53177-098, A53177-099, A53177-100, A53177-101, A53177-102, A53177-103, A53177-104, A53177-105, A53177-106, A53177-107, A53177-108, A53177-109, A53177-110, A53177-111, A53177-112, A53177-113, A53177-114, A53177-115, A53177-116, A53177-117, A53177-118, A53177-119, A53177-120, A53177-121, A53177-122, A53177-124, A53177-125, A53177-126, A53177-127, A53385-011, A53385-012, A53385-013, A53385-014, A53385-015, A53385-016, A53385-017, A53385-018, A53385-019, A53385-020, A53385-021, A53385-022, A53385-023, A53385-024, A53385-025, A53385-026, A53385-027, A53385-028, A53385-029, A53385-030, A53385-031, A53385-032, A53385-033, A53385-034, A53385-035, A53385-036, A53385-037, A53385-038, A53385-039, A53385-040, A53385-041, A53385-042, A53385-043, A53385-044, A53385-045, A53385-046, A53385-047, A53385-048, A53385-049, A53385-050, A53385-051, A53385-052, A53385-053, A53385-054, A53385-055, A53385-056, A53385-057, A53385-058, A53385-059, A53385-060, A53385-061, A53385-062, A53385-063, A53385-064, A53385-065, A53385-066, A53385-067, A53385-068, A53385-069, A53385-070, A53385-071, A53385-072, A53385-073, A53385-074, A53385-075, A53385-076, A53385-077, A53385-078, A53385-079, A53385-080, A53385-081, A53385-082, A53385-083, A53385-084, A53385-085, A53387-011, A53387-012, A53387-013, A53387-014, A53387-015, A53387-016, A53387-017, A53387-018, A53387-019, A53387-020, A53387-021, A53387-022, A53387-023, A53387-024, A53387-025, A53387-026, A53387-027, A53387-028, A53387-029, A53387-030, A53387-031, A53387-032, A53387-033, A53387-034, A53387-035, A53387-036, A53387-037, A53387-038, A53387-039, A53387-040, A53387-041, A53387-042, A53387-043, A53387-044, A53387-045, A53387-046, A53387-047, A53387-048, A53387-049, A53387-050, A53387-051, A53491-082, A53491-083, A53491-084, A53491-085, A53491-086, A53491-087, A53491-088, A53491-089, A53491-090, A53491-091, A53491-092, A53491-093, A53491-094, A53491-095, A53491-096, A53491-097, A53491-098, A53491-099, A53491-100, A53491-101, A53491-102, A53491-103, A53491-104, A53491-105, A53491-106, A53491-107, A53491-108, A53491-109, A53491-110, A53491-111, A53491-112, A53491-113, A53491-114, A53491-115, A53491-116, A53491-117, A53491-118, A53491-119, A53491-120, A53491-121, A53491-122, A53491-123, A53491-124, A53491-125, A53491-126, A53491-127, A53491-128, A53491-129, A53491-130, A53491-131, A53491-132, A53491-133, A53491-134, A53491-135, A53491-136, A53491-137, A53491-138, A53491-139, A53491-140, A53491-141, A53491-142, A53491-143, A53491-144, A53491-145, A53491-146, A53491-147, A53491-148, A53970-051, A53970-052, A53970-053, A53970-054, A53970-055, A53970-056, A53970-057, A53970-058, A53970-059, A53970-060, A53970-061, A53970-062, A53970-063, A53970-064, A53970-065, A53970-066, A53970-067, A53970-068, A53970-069, A53970-070, A53970-071, A53970-072, A53970-073, A53970-074, A53970-075, A53970-076, A53970-077, A53970-078, A53970-079, A53970-080, A53970-081, A53970-082, A53970-083, A53970-084, A53970-085, A53970-086, A53970-087, A53970-088, A53970-089, A53970-090, A53970-091, A53970-092, A53970-093, A53970-094, A53970-095, A53970-096, A53970-097, A53970-098, A53970-099, A53970-100, A53970-101, A53970-102, A53970-103, A53970-104, A53970-105, A53970-106, A53970-107, A53970-108, A53970-109, A53970-110, A53970-111, A53970-112, A53970-113, A53970-114, A53970-115, A53970-116, A53970-117, A53970-118, A53970-119, A53970-120, A53970-121, A53970-122, A53970-123, A53970-124, A53970-125, A53970-126, A53970-127, A53970-128, A53970-129, A53970-130, A54671-048, A54671-049, A54671-050, A54671-051, A54671-052, A54671-053, A54671-054, A54671-055, A54671-056, A54671-057, A54671-058, A54671-059, A54671-060, A54671-061, A54671-062, A54671-063, A54671-064, A54671-065, A54671-066, A54671-067, A54671-068, A54671-069, A54671-070, A54671-071, A54671-072, A54671-073, A54671-074, A54671-075, A54671-076, A55131-026, A55131-027, A55131-028, A55131-029, A55131-030, A55131-031, A55131-032, A55131-033, A55131-034, A55131-035, A55131-036, A55131-037, A55131-038, A55131-039, A55131-040, A55131-041, A55131-042, A55131-043, A55131-044, A55131-045, A55131-046, A55131-047, A55131-048, A55131-049, A55131-050, A55131-051, A55131-052, A55131-053, A55131-054, A55131-055, A55131-056, A55131-057, A55131-058, A55131-059, A55131-060, A55131-061, A55131-062, A55131-063, A55131-064, A55131-065, A55131-066, A55131-067, A55131-068, A55131-069, A55131-070, A55131-071, A55131-072, A55131-073, A55131-074, A55131-075, A55131-076, A55131-077, A55131-078, A55131-079, A55131-080, A55131-081, A55131-082, A55131-083, A55131-084, A55131-085, A55131-086, A55131-087, A55131-088, A55131-089, A55131-090, A55131-091, A55131-092, A55131-093, A55131-094, A55131-095, A55131-096, A55131-097, A55131-098, A55131-099, A55131-100, A55131-101, A55131-102, A55131-103, A55131-104;
g) DBM T41130 5CC GRAFTON GEL, REF T41130, GTIN 00643169122970, Serial Numbers: A37724-127, A37724-128, A37724-129, A37724-130, A37724-131, A37724-133, A37724-134, A37724-135, A37724-136, A37724-137, A37724-138, A37724-139, A37724-140, A37724-141, A37724-142, A37724-143, A37724-144, A37724-145, A37724-146, A37724-148, A37724-149, A37724-150, A37724-151, A37724-152, A37724-153, A37724-154, A37724-155, A37724-156, A37724-157, A41163-076, A41163-077, A41163-078, A41163-079, A41163-080, A41163-081, A41163-082, A41163-083, A41163-084, A41163-085, A41163-086, A41163-087, A41163-088, A41163-089, A41678-092, A41678-093, A41678-094, A41678-095, A41678-096, A41678-097, A41678-098, A41678-099, A41678-100, A41678-101, A41678-102, A41678-103, A41678-104, A41678-105, A41678-106, A41678-107, A41678-108, A41678-109, A41678-110, A41678-111, A41678-112, A41678-113, A41678-114, A41678-115, A41678-116, A41678-117, A41678-118, A41678-119, A41678-120, A41678-121, A41678-122, A41678-123, A41678-124, A41678-125, A41678-126, A41678-127, A41678-128, A41678-129, A41879-051, A41879-052, A41879-053, A41879-054, A41879-055, A41879-056, A41879-057, A41879-058, A41879-059, A41879-060, A41879-061, A41879-062, A41879-063, A41879-064, A41879-065, A41879-066, A41879-067, A41879-068, A41879-069, A41879-070, A41879-071, A41879-072, A41879-073, A41879-074, A41879-075, A41879-076, A41879-078, A41879-079, A41879-080, A41879-081, A41879-082, A41879-083, A41879-084, A41879-085, A41879-086, A41879-087, A41879-088, A41879-089, A41879-090, A41909-153, A41909-154, A41909-155, A41909-156, A41909-157, A41909-158, A41909-159, A41909-160, A41909-161, A41909-162, A41909-163, A41909-164, A41909-165, A41909-167, A41909-168, A41909-169, A41909-170, A41909-171, A41909-172, A41909-173, A41909-174, A41909-175, A41909-176, A41909-177, A41909-178, A41909-179, A41909-180, A41909-181, A41909-182, A41909-183, A41909-184, A41909-185, A41909-186, A41909-187, A41909-188, A41909-189, A41909-190, A41909-191, A41909-192, A41909-193, A41909-194, A41909-195, A41909-196, A41909-197, A41909-198, A41909-199, A41909-200, A41909-201, A41909-202, A41909-203, A41909-204, A41909-205, A41909-206, A41909-207, A41935-054, A41935-055, A41935-056, A41935-059, A41935-060, A41935-061, A41935-062, A41935-063, A41935-064, A41935-065, A41935-066, A41935-067, A41935-068, A41935-069, A41935-070, A41935-071, A41935-072, A41935-073, A41935-074, A41935-075, A41935-076, A41935-077, A41935-078, A41935-079, A41935-080, A41935-081, A41935-082, A41935-083, A41935-084, A41935-085, A41935-086, A41935-087, A41935-088, A41935-089, A41935-090, A41935-091, A41935-092, A41935-093, A41935-094, A41935-095, A41935-096, A41935-097, A41935-098, A41935-099, A41935-100, A42095-105, A42095-106, A42095-107, A42095-108, A42095-109, A42095-110, A42095-111, A42095-112, A42095-113, A42095-114, A42095-115, A42095-116, A42095-117, A42095-118, A42095-119, A42095-120, A42095-121, A42095-122, A42095-123, A42095-124, A42095-125, A42095-126, A42095-127, A42619-037, A42619-038, A42619-039, A42619-040, A42619-041, A42619-042, A42619-043, A42619-045, A42619-046, A42619-047, A42619-049, A42619-050, A42619-051, A42619-052, A42619-053, A42619-054, A42619-055, A42619-056, A42619-057, A42619-058, A42619-059, A42662-096, A42662-097, A42662-098, A42662-099, A42662-100, A42662-101, A42662-102, A42662-103, A42662-104, A42662-105, A42662-106, A42662-107, A42662-108, A42662-109, A42662-110, A42662-111, A42662-112, A42662-113, A42662-114, A42662-115, A42662-116, A42662-117, A42662-118, A42662-119, A43331-066, A43331-067, A43331-068, A43331-069, A43331-071, A43331-072, A43331-073, A43331-074, A43331-075, A43331-076, A43331-077, A43331-078, A43331-079, A43331-080, A43331-081, A43331-082, A43331-083, A43331-084, A43331-085, A43331-086, A43331-087, A43331-088, A43331-089, A43331-090, A43331-091, A43331-092, A43331-093, A43331-094, A43331-095, A43331-096, A43331-097, A44260-026, A44260-027, A44260-028, A44260-029, A44260-030, A44260-031, A44260-032, A44260-033, A44260-034, A44260-035, A44260-036, A44260-037, A44260-038, A44260-039, A44260-040, A44260-041, A44260-042, A44260-043, A44260-044, A44260-045, A44260-046, A44260-047, A44260-048, A44260-049, A44260-050, A44260-051, A44260-052, A44260-053, A44260-054, A44260-055, A44260-056, A44260-057, A44260-058, A44260-059, A44260-060, A44260-061, A44260-062, A44260-063, A44260-064, A44260-065, A44405-055, A44405-056, A44405-057, A44405-058, A44405-060, A44405-061, A44405-062, A44405-063, A44405-064, A44405-065, A44583-011, A44583-012, A44583-013, A44583-014, A44583-015, A44583-016, A44583-017, A44583-018, A44583-019, A44583-020, A44583-021, A44583-022, A44583-024, A44583-025, A44583-026, A44583-027, A44583-028, A44583-029, A44583-030, A44583-031, A44583-032, A44583-033, A45121-108, A45121-109, A45121-112, A45121-113, A45121-114, A45121-115, A45121-116, A45121-117, A45121-118, A45121-119, A45121-120, A45121-121, A45121-122, A45121-123, A45121-124, A45224-011, A45224-012, A45224-013, A45224-014, A45224-015, A45224-016, A45224-017, A45224-018, A45224-019, A45224-020, A45224-021, A45224-022, A45224-023, A45224-024, A45224-025, A45224-026, A45224-027, A45224-028, A45224-029, A45224-030, A45224-031, A45224-032, A45224-033, A45224-034, A45224-035, A45224-036, A45224-037, A45224-038, A45224-039, A45224-040, A45224-041, A45224-042, A45224-043, A45224-044, A45607-021, A45607-022, A45607-023, A45607-024, A45607-025, A45607-026, A45607-027, A45607-028, A45607-029, A45607-030, A45607-031, A45607-032, A45607-033, A45607-034, A45607-035, A45607-036, A45607-037, A45607-038, A45607-039, A45607-040, A45607-041, A45607-042, A45607-043, A45607-044, A45607-045, A45607-046, A46158-067, A46158-068, A46158-069, A46158-070, A46158-071, A46158-072, A46158-073, A46158-074, A46158-075, A46158-076, A46158-077, A46158-078, A46158-079, A46158-080, A46158-081, A47127-210, A47127-211, A47127-212, A47127-213, A47127-214, A47127-215, A47127-216, A47127-217, A47127-218, A47127-219, A47127-220, A47394-039, A47661-031, A47661-032, A47661-033, A47661-034, A47661-035, A47661-036, A47661-037, A47661-038, A47661-040, A47661-041, A47661-042, A47661-043, A47661-044, A47661-045, A47661-047, A47661-048, A47661-049, A47661-050, A47661-051, A47661-052, A47661-053, A47661-054, A47661-055, A47661-056, A47661-057, A47661-058, A47661-059, A47661-060, A47661-061, A47661-062, A47661-063, A47661-064, A47661-065, A47661-066, A47672-095, A47672-096, A47672-097, A47672-098, A47672-099, A47672-100, A47672-101, A47672-102, A47672-103, A47672-104, A47672-105, A47672-106, A47672-107, A47672-108, A47672-109, A47672-110, A47672-111, A47672-112, A47672-113, A47672-114, A47672-115, A47672-116, A47672-117, A47672-118, A47672-119, A47725-050, A47725-051, A47725-052, A47725-053, A47781-030, A47781-031, A47781-032, A47781-033, A47781-034, A47781-035, A47781-036, A47781-037, A47781-039, A47781-040, A47781-041, A47781-042, A47781-043, A47781-044, A47781-045, A47781-046, A47781-047, A47781-048, A47781-049, A47781-050, A47781-051, A47781-052, A47781-053, A47781-054, A47781-055, A47781-056, A47781-057, A47781-058, A48114-102, A48114-103, A48114-104, A48114-105, A48114-106, A48114-107, A48114-108, A48114-109, A48114-110, A48114-111, A48114-112, A48114-113, A48114-114, A48114-116, A48114-117, A48114-118, A48114-119, A48114-120, A48114-121, A48114-122, A48114-123, A48114-125, A48114-126, A48114-127, A48114-128, A48114-129, A48114-130, A48114-131, A48114-132, A48114-133, A48114-134, A48114-135, A48114-136, A48114-137, A48114-138, A48114-139, A48438-074, A48438-075, A48438-076, A48438-077, A48438-078, A48438-079, A48438-080, A48438-081, A48438-082, A48438-083, A48438-084, A48438-085, A48438-086, A48438-087, A48438-088, A48438-089, A48438-091, A48438-092, A48438-093, A48438-094, A48438-095, A48438-096, A49710-128, A49710-129, A49710-130, A49710-131, A49710-132, A49710-133, A49710-134, A49710-135, A49710-136, A49710-137, A49710-138, A49710-139, A49710-140, A49710-141, A49710-142, A49710-143, A49710-144, A49710-145, A49710-146, A49751-052, A49751-053, A49751-054, A49751-055, A49751-056, A49751-057, A49751-058, A49751-059, A49751-060, A49751-061, A49751-062, A49751-063, A49751-064, A49751-065, A49751-066, A49751-067, A49751-068, A49751-069, A49751-070, A49751-071, A49751-072, A49751-073, A49751-074, A49751-075, A49751-076, A49751-077, A49751-078, A50408-031, A50408-032, A50408-033, A50408-034, A50408-035, A50408-036, A50408-037, A50408-038, A50408-039, A50408-040, A50408-041, A50408-042, A50408-043, A50408-044, A50408-045, A50408-046, A50408-047, A50408-048, A50408-049, A50408-050, A50408-051, A50408-052, A50408-053, A50408-054, A50408-055, A50408-056, A50408-057, A50408-058, A50408-059, A50910-109, A50910-110, A50910-111, A50910-112, A50910-113, A50910-114, A50910-115, A50910-116, A50910-118, A50910-119, A50910-120, A50910-121, A50910-122, A50910-123, A50910-124, A50910-125, A50910-126, A50910-127, A50910-128, A50910-129, A50910-130, A50910-131, A50910-132, A50910-133, A50910-134, A50910-136, A52369-093, A52369-094, A52369-095, A52369-096, A52369-097, A52369-098, A52369-099, A52369-100, A52369-101, A52369-102, A52369-103, A52369-104, A52369-105, A52369-106, A52369-107, A52369-108, A52369-109, A52369-110, A52369-111, A52369-112, A52369-113, A52369-114, A52369-115, A52369-116, A52369-117, A52369-118, A52425-056, A52425-057, A52425-058, A52425-059, A52425-060, A52425-061, A52425-064, A52425-065, A52425-066, A52425-067, A52425-068, A52848-022, A52848-023, A52848-024, A52848-025, A52848-026, A52848-027, A52848-028, A52848-029, A52848-030, A52848-031, A52848-032, A52848-033, A52848-034, A52848-035, A52848-036, A52848-037, A52848-038, A52848-039, A52848-040, A52848-041, A52848-042, A52848-043, A52848-044, A52848-045, A52848-046, A52848-047, A52848-048, A52848-049, A52848-050, A52848-051, A52848-052, A52848-053, A52848-054, A52848-055, A52848-056, A52848-057, A52848-058, A52848-059, A52848-060, A52848-061, A52848-062, A52848-063, A52848-064, A52848-065, A52848-066, A52848-067, A52848-068, A52848-069, A52928-082, A52928-083, A52928-084, A52928-085, A52928-086, A52928-087, A52928-088, A52928-089, A52928-090, A52928-091, A52928-092, A52928-093, A52928-094, A52928-095, A52928-096, A52928-097, A52928-098, A52928-099, A52928-100, A52928-101, A52928-102, A53149-041, A53149-042, A53149-043, A53149-044, A53149-045, A53149-046, A53149-047, A53149-048, A53149-049, A53149-050, A53149-051, A53149-052, A53149-053, A53149-054, A53149-055, A53386-098, A53386-099, A53386-100, A53386-101, A53386-102, A53386-103, A53386-104, A53386-105, A53386-106, A53386-107, A53386-108, A53386-109, A53386-110, A53386-111, A53471-067, A53471-068, A53471-069, A53471-070, A53471-071, A53471-072, A53471-073, A53471-074, A53471-075, A53471-076, A53471-077, A53471-078, A53471-079, A53471-080, A53471-081, A53471-082, A53471-083, A53471-084, A53471-085, A53597-051, A53597-052, A53597-053, A53597-054, A53597-055, A53597-056, A53597-057, A53597-058, A53597-059, A53597-060, A53597-061, A53597-062, A53597-063, A53597-064, A53597-065, A53597-066, A53597-067, A54123-045, A54123-046, A54123-047, A54123-048, A54123-049, A54123-050, A54123-051, A54123-052, A54123-053, A54123-054, A54123-055, A54123-056, A54123-057, A54123-058, A54123-061, A54123-062, A54123-069, A54123-070, A54238-061, A54244-074, A54244-075, A54244-076, A54244-077, A54244-078, A54244-079, A54244-080, A54244-081, A54244-082, A54244-083, A54244-084, A54244-085, A54244-086, A54244-088, A54244-089, A54244-090, A54244-091, A54244-092, A54244-093, A54373-093, A54373-094, A54373-095, A54373-096, A54373-097, A54373-098, A54373-099, A54373-100, A54373-101, A54373-102, A54373-103, A54373-104, A54373-105, A54373-106, A54373-107, A54373-108, A54373-109, A54373-110, A54373-111, A54650-048;
h) DBM T41130INT 5CC GRAFTON GEL, REF T41130INT, GTIN 00643169964853, Serial Numbers: A38910-064, A38910-065, A38910-066, A38910-067, A38910-068, A38910-069, A38910-070, A38910-071, A38910-072, A38910-073, A38910-074, A38910-075, A38910-076, A38910-077, A38910-078, A38910-079, A38910-080, A38910-081, A38910-082, A38910-083, A38910-084, A38910-085, A38910-086, A38910-087, A38910-088, A38910-089, A38910-090, A38910-091, A38910-092, A38910-093, A38910-094, A41224-077, A41224-078, A41224-079, A41224-080, A41224-081, A41224-082, A41224-083, A41224-084, A41224-085, A41224-086, A41224-087, A41224-088, A41224-089, A41224-090, A41224-091, A41224-092, A41224-093, A41224-094, A41224-095, A41224-096, A41224-097, A41224-098, A41224-099, A41224-100, A41626-060, A41626-061, A41626-062, A41626-063, A41626-064, A41626-065, A41626-066, A41626-067, A41626-068, A41626-069, A41626-070, A41626-071, A41626-072, A41626-073, A41626-074, A41626-075, A41627-040, A41627-041, A41627-042, A41627-043, A41627-044, A41627-045, A41627-046, A41627-047, A41627-048, A41627-049, A41627-050, A41627-051, A41627-052, A41627-053, A41627-054, A41627-055, A41627-056, A41627-057, A41627-058, A41627-059, A41627-060, A41627-061, A41627-062, A41627-063, A41627-064, A41627-065, A41627-066, A41627-067, A41627-068, A42484-078, A42484-079, A42484-080, A42484-081, A42484-082, A42484-083, A42484-084, A42484-085, A42484-086, A42484-087, A42484-088, A42484-089, A42484-090, A42484-091, A42484-092, A42571-110, A42571-111, A42571-112, A42571-113, A42571-114, A42571-115, A42571-116, A42571-117, A42571-118, A42571-119, A42571-120, A42571-121, A42571-122, A42571-123, A42571-124, A42571-125, A42571-126, A42571-127, A42571-128, A42571-129, A42571-130, A42571-131, A42662-120, A42662-121, A42662-122, A42662-123, A42662-124, A42662-125, A42662-126, A42662-127, A42662-128, A42780-102, A42780-103, A42780-104, A42780-105, A42780-106, A42780-107, A42780-108, A42780-109, A42780-110, A42780-111, A42780-112, A42780-113, A42780-114, A42780-115, A42780-116, A42780-117, A42780-118, A42780-119, A42780-120, A42780-121, A42780-122, A42780-123, A43107-021, A43107-022, A43107-023, A43107-024, A43107-025, A43107-026, A43107-027, A43107-028, A43107-029, A43107-030, A43107-031, A43107-032, A43107-033, A43107-034, A43107-035, A43107-036, A43107-037, A43107-038, A43107-039, A43107-040, A43107-041, A43107-042, A43107-043, A43107-044, A43107-045, A43107-046, A43107-047, A43107-048, A43939-044, A43939-045, A43939-046, A43939-047, A43939-048, A43939-049, A43939-050, A43939-051, A43939-052, A43939-053, A43939-054, A43939-056, A46333-083, A46333-084, A46333-085, A46333-086, A46333-087, A46333-088, A46333-089, A46333-090, A46333-091, A46333-092, A46333-093, A46333-094, A46333-095, A46333-096, A46333-097, A46333-098, A46333-099, A46333-100, A46333-101, A46333-102, A46333-103, A46333-104, A46333-105, A46333-106, A46333-107, A46333-108, A46333-109, A46333-110, A46333-111, A46333-112, A46333-113, A46333-114, A47115-041, A47115-042, A47115-043, A47115-044, A47115-045, A47115-046, A47115-047, A47115-048, A47115-049, A47115-050, A47115-051, A47115-052, A47115-053, A47115-054, A47115-055, A47115-056, A47115-057, A47115-058, A47115-059, A47115-060, A47115-061, A47115-062, A47115-063, A47115-064, A47115-065, A47115-066, A47115-067, A47115-068, A47115-069, A47115-070, A47115-071, A47115-072, A47115-073, A47115-074, A47127-236, A47127-237, A47127-238, A47127-239, A47127-240, A47127-241, A47127-242, A47127-243, A47127-244, A47127-245, A47127-247, A47127-248, A47127-249, A47127-250, A47338-101, A47338-102, A47338-103, A47338-104, A47338-105, A47338-106, A47338-107, A47338-108, A47338-109, A47338-110, A47338-111, A47338-112, A47338-113, A47338-114, A47338-115, A47338-116, A47338-117, A47338-118, A47338-119, A47338-120, A47338-121, A47338-122, A47338-123, A47338-124, A47338-125, A47338-126, A47338-127, A47338-128, A47338-129, A47338-130, A47338-131, A47338-133, A47338-134, A47391-101, A47391-102, A47391-103, A47391-104, A47391-105, A47391-106, A47391-107, A47391-108, A47391-109, A47391-110, A47391-111, A47391-112, A47391-113, A47391-114, A47391-115, A47391-116, A47391-117, A47391-118, A47391-119, A47391-120, A47391-121, A47391-122, A47391-123, A47391-124, A47391-125, A47391-126, A47391-127, A47391-128, A47391-130, A47391-131, A47391-132, A47391-133, A47391-134, A47391-135, A47391-136, A47431-083, A47431-084, A47431-085, A47431-086, A47431-087, A47431-088, A47431-089, A47431-090, A47431-091, A47431-092, A47431-093, A47431-094, A47431-095, A47431-096, A47431-097, A47431-098, A47431-099, A47431-100, A47431-101, A47431-102, A47431-103, A47431-104, A47431-105, A47431-106, A47431-107, A47431-108, A47431-109, A47431-110, A47431-111, A47431-112, A47431-113, A47431-114, A47431-115, A47573-096, A47573-097, A47573-098, A47573-099, A47573-100, A47573-101, A47573-102, A47573-103, A47573-104, A47573-105, A47573-106, A47573-107, A47573-108, A47573-109, A47573-110, A47573-111, A47573-112, A47573-113, A47573-114, A47573-115, A47573-116, A47573-117, A47573-118, A47573-119, A47573-120, A47573-121, A47725-054, A47725-055, A47725-056, A47725-057, A47725-058, A47725-059, A47725-060, A47725-061, A47725-062, A47725-063, A47725-064, A47725-065, A47725-066, A47725-067, A47725-068, A47725-069, A47725-070, A47725-071, A47725-072, A47725-073, A47725-074, A47939-031, A47939-032, A47939-033, A47939-034, A47939-035, A47939-036, A47939-037, A47939-038, A47939-039, A47939-040, A47939-041, A47939-042, A47939-043, A47939-044, A47939-045, A47939-046, A47939-047, A47939-048, A47939-049, A47939-050, A47939-051, A47939-052, A47939-053, A47939-054, A47939-055, A47939-056, A47939-057, A47939-058, A47939-059, A47939-060, A47939-061, A47939-062, A47939-063, A47939-064, A47939-065, A47939-066, A47939-067, A47939-068, A47939-069, A47939-070, A47939-071, A47939-072, A47939-073, A47939-074, A47939-075, A47939-076, A48378-049, A48378-050, A48378-051, A48378-052, A48378-053, A48378-054, A48378-055, A48378-056, A48378-057, A48378-058, A48378-059, A48378-060, A48378-061, A48378-062, A48378-063, A48378-064, A48378-065, A48378-066, A48378-067, A48378-068, A48378-069, A48378-070, A48378-071, A48378-072, A48378-073, A48378-074, A48378-075, A48378-076, A48378-077, A48378-078, A48835-027, A48835-028, A48835-029, A48835-030, A48835-031, A48835-032, A48835-033, A48835-038, A48835-039, A48835-040, A48835-041, A48835-042, A48835-043, A48835-044, A48835-045, A48835-046, A48835-047, A48835-048, A48835-049, A48835-050, A48835-051, A48835-052, A48835-053, A48835-054, A48835-055, A48835-056, A48835-057, A48835-058, A48835-059, A48835-060, A48835-061, A48835-062, A48835-063, A48835-064, A48835-065, A49912-094, A49912-095, A49912-096, A49912-097, A49912-098, A49912-099, A49912-100, A49912-101, A49912-102, A49912-103, A49912-104, A49912-105, A49912-106, A49912-107, A49912-108, A49912-109, A49912-110, A49912-111, A49912-112, A49912-113, A49912-114, A49912-115, A49912-116, A49912-117, A49912-118, A49912-119, A49912-120, A49912-121, A49912-122, A49912-123, A49912-124, A49912-125, A49912-126, A49912-127, A49912-128, A49912-129, A49912-130, A49912-131, A49912-132, A49912-133, A49912-134, A49912-135, A49912-136, A49912-137, A49975-080, A49975-081, A49975-082, A49975-083, A49975-084, A49975-085, A49975-086, A49975-087, A49975-088, A49975-089, A49975-090, A49975-091, A49975-092, A49975-093, A49975-094, A49975-095, A49975-096, A49975-097, A49975-098, A49975-099, A49975-100, A49975-101, A49975-102, A49975-103, A49975-104, A49975-105, A49975-106, A49975-107, A49975-108, A49975-109, A49975-110, A49975-111, A49975-112, A49975-113, A49975-114, A49975-115, A50311-071, A50311-072, A50311-073, A50311-074, A50311-075, A50311-076, A50311-077, A50311-078, A50311-079, A50311-080, A50311-081, A50311-082, A50311-083, A50311-084, A50311-085, A50311-086, A50311-087, A50311-088, A50317-011, A50317-012, A50317-013, A50317-014, A50317-015, A50317-016, A50317-017, A50317-018, A50317-019, A50317-020, A50317-021, A50317-022, A50317-023, A50317-024, A50317-025, A50317-026, A50317-027, A50317-028, A50317-029, A50317-030, A50317-031, A50317-032, A50317-033, A50317-034, A50317-035, A50317-036, A50317-037, A50317-038, A50317-039, A50317-040, A50317-041, A50317-042, A50317-043, A50317-044, A50317-045, A50317-046, A50317-047, A50317-048, A50317-049, A50317-050, A50317-051, A50467-026, A50467-027, A50467-028, A50467-029, A50467-030, A50467-031, A50467-032, A50467-033, A50467-034, A50467-035, A50467-036, A50467-037, A50467-038, A50467-039, A50467-040, A50467-041, A50467-042, A50467-043, A50467-044, A50467-045, A50467-046, A50467-047, A50467-048, A50467-049, A50467-050, A50467-051, A50467-052, A50467-053, A50467-054, A50467-055, A50467-056, A50467-057, A50467-058, A50467-059, A50467-060, A50467-061, A50467-062, A50467-063, A50467-064, A51706-050, A51706-051, A51706-052, A51706-053, A51706-054, A51706-055, A51706-056, A51706-057, A51706-058, A51706-059, A51706-060, A51706-061, A51706-062, A51706-063, A51706-064, A51706-065, A51706-066, A51706-067, A51706-068, A51706-069, A51706-070, A51706-071, A51706-072, A51706-073, A51706-074, A51706-075, A51706-076, A51706-077, A51706-078, A51706-079, A51706-080, A51706-081, A51706-082, A51706-083, A52101-011, A52101-012, A52101-013, A52101-014, A52101-015, A52101-016, A52101-017, A52101-018, A52101-019, A52101-020, A52101-021, A52101-022, A52101-023, A52101-024, A52101-025, A52101-026, A52101-027, A52101-028, A52101-029, A52101-030, A52101-031, A52101-032, A52101-033, A52101-034, A52101-035, A52101-036, A52101-037, A52216-011, A52216-012, A52216-013, A52216-014, A52216-015, A52216-016, A52216-017, A52216-018, A52216-019, A52216-020, A52216-021, A52216-022, A52216-023, A52216-024, A52216-025, A52216-026, A52216-027, A52216-028, A52216-029, A52216-030, A52216-031, A52216-032, A52216-033, A52216-034, A52216-035, A52216-036, A52216-037, A52627-029, A52627-030, A52627-031, A52627-032, A52627-033, A52627-034, A52627-035, A52627-036, A52627-037, A52627-038, A52627-039, A52627-040, A52627-041, A52627-042, A52627-043, A52627-044, A52627-045, A52627-046, A52627-047, A52627-048, A52627-049, A52627-050, A52627-051, A52627-052, A52627-053, A52627-054, A53172-037, A53172-038, A53172-039, A53172-040, A53172-041, A53172-042, A53172-043, A53172-044, A53172-045, A53172-046, A53172-047, A53172-048, A53172-049, A53172-050, A53172-051, A53172-052, A53172-053, A53172-054, A53172-055, A53172-056, A53172-057, A53185-017, A53185-018, A53185-019, A53185-020, A53185-021, A53185-022, A53185-023, A53185-024, A53185-025, A53185-026, A53185-027, A53185-028, A53185-029, A53185-030, A53185-031, A53185-032, A53185-033, A53185-034, A53185-035, A53185-036, A53185-037, A53185-038, A53185-039, A53185-040, A53185-041, A53389-091, A53389-092, A53389-093, A53389-094, A53389-095, A53389-096, A53389-097, A53389-098, A53389-099, A53389-100, A53389-101, A53389-102, A53389-103, A53389-104, A53389-105, A53389-106, A53389-107, A53389-108, A53438-040, A53438-041, A53438-042, A53438-043, A53438-044, A53438-045, A53438-046, A53438-047, A53438-048, A53438-049, A53438-050, A53438-051, A53438-052, A53438-053, A53438-054, A53438-055, A53438-056, A53438-057, A53438-058, A53438-059, A53438-060, A53647-066, A53647-067, A53647-068, A53647-069, A53647-070, A53647-071, A53647-072, A53647-073, A53647-074, A53647-075, A53647-076, A53647-077, A53647-078, A53647-079, A53647-080, A54236-031, A54236-032, A54236-033, A54236-034, A54236-035, A54236-036, A54236-037, A54236-038, A54236-039, A54236-040, A54236-041, A54236-042, A54236-043, A54236-044, A54236-045, A54236-046, A54236-047, A54236-048, A54236-049, A54236-050, A54236-051, A54236-052, A54236-053, A54236-054, A54236-055, A54236-056, A54236-057, A54236-058, A54236-059, A54236-060, A54236-061, A54236-062, A54671-021, A54671-022, A54671-023, A54671-024, A54671-025, A54671-026, A54671-027, A54671-028, A54671-029, A54671-030, A54671-031, A54671-032, A54671-033, A54671-034, A54671-035, A54671-036, A54671-037, A54671-038, A54671-039, A54671-040, A54671-041, A54671-042, A54671-043, A54671-044, A54671-045, A54671-046, A54671-047;
i) DBM T41150 10CC GRAFTON GEL, REF T41150, GTIN 00643169122987, Serial Numbers: A41213-056, A41213-057, A41213-058, A41213-059, A41213-060, A41213-061, A41213-062, A41213-063, A41213-064, A41213-065, A41213-066, A41213-067, A41215-036, A41215-037, A41215-038, A41215-039, A41215-040, A41215-041, A41215-042, A41215-043, A41911-115, A41911-116, A41911-117, A41911-118, A41911-119, A41911-120, A41911-121, A41911-122, A41911-123, A41911-124, A41911-125, A41911-126, A41911-127, A41911-128, A41911-129, A41911-130, A41911-131, A42285-026, A42285-027, A42285-028, A42285-029, A42285-030, A42285-031, A42285-032, A42285-033, A42285-034, A42285-035, A42285-036, A42285-037, A42285-039, A42285-040, A42285-041, A42285-042, A42307-011, A42307-012, A42307-013, A42307-014, A42307-015, A42307-016, A42307-017, A42307-018, A42307-019, A42307-020, A42307-021, A42307-022, A42307-023, A42307-024, A42307-025, A42307-026, A42307-027, A42307-028, A42307-029, A42307-030, A42307-031, A42307-032, A42307-033, A42444-053, A42444-054, A42444-055, A42444-056, A42444-057, A42444-058, A42444-059, A42444-060, A42444-061, A42444-062, A42663-070, A42663-071, A42663-072, A42663-073, A42663-074, A42663-075, A42663-076, A42663-077, A42663-078, A42663-079, A42663-080, A42663-081, A42663-082, A42663-083, A42895-094, A42895-095, A42895-096, A42895-097, A43013-063, A43013-064, A43013-065, A43013-066, A43013-067, A43013-068, A43013-069, A43013-070, A43013-071, A43013-072, A43013-073, A43013-074, A43013-075, A43013-076, A43013-077, A43013-078, A43013-079, A43286-062, A43286-063, A43286-064, A43286-065, A43286-066, A43328-040, A43328-041, A43328-042, A43328-043, A43328-044, A43328-045, A43328-046, A43328-047, A43328-048, A43328-049, A43328-050, A43328-051, A43328-052, A43328-053, A43338-121, A43338-122, A43338-123, A43338-124, A43338-125, A43338-126, A43338-127, A43338-128, A43338-129, A43338-130, A43338-131, A43338-132, A43338-133, A43338-134, A43338-135, A43338-136, A43338-137, A45084-076, A45084-077, A45084-078, A45084-079, A45084-080, A45084-081, A45084-082, A45084-083, A45084-084, A45084-085, A45413-047, A45413-048, A45413-049, A45413-050, A45413-051, A45413-052, A45413-053, A45413-054, A45413-055, A45413-056, A45413-057, A45413-058, A45413-059, A45413-060, A45413-061, A45413-062, A45690-011, A45690-012, A45690-013, A45690-014, A45690-015, A45690-016, A45690-017, A45690-018, A45690-019, A45690-020, A45690-021, A45690-022, A45690-023, A45690-024, A45690-025, A45690-026, A45690-027, A45690-028, A45717-011, A45717-012, A45717-013, A45717-014, A45717-015, A45717-016, A45717-017, A45717-018, A45717-019, A45717-020, A45717-021, A45717-022, A45717-023, A45717-024, A45717-025, A45717-026, A45757-016, A45757-017, A45757-018, A45757-019, A45757-020, A45757-021, A45757-022, A45757-023, A45757-024, A45757-025, A45757-026, A45757-027, A45757-028, A45757-029, A45757-030, A45757-031, A45757-032, A45757-033, A45757-034, A45892-011, A45892-012, A45892-013, A45892-014, A45892-015, A45892-016, A45892-017, A45892-018, A45892-019, A45892-020, A45892-021, A45892-022, A45892-023, A46846-067, A46846-068, A46846-069, A46846-070, A46846-071, A46846-072, A46846-073, A46846-074, A46846-075, A47225-011, A47225-012, A47225-013, A47225-014, A47225-015, A47433-145, A47433-146, A47433-147, A47433-148, A47433-149, A47433-150, A47433-151, A47433-152, A47934-016, A47934-017, A47934-018, A47934-019, A47934-020, A47934-021, A47934-022, A47934-023, A47934-024, A47934-025, A47934-026, A47934-027, A47934-028, A47934-029, A47934-030, A47934-031, A47934-032, A47934-033, A47934-034, A47934-035, A47934-036, A47934-038, A47934-039, A47934-040, A47934-041, A47934-042, A47934-043, A48441-092, A48441-093, A48441-094, A48441-095, A48441-096, A48441-097, A48441-098, A48441-099, A48441-100, A48441-101, A48441-102, A48441-103, A48441-104, A48441-105, A48441-106, A48441-107, A48441-108, A48441-109, A48441-110, A48441-111, A48781-176, A48781-177, A48781-178, A48781-179, A48781-180, A48781-181, A48781-182, A48781-183, A48781-184, A48781-185, A48781-186, A48781-187, A48781-188, A48985-016, A48985-017, A48985-018, A48985-019, A48985-020, A48985-021, A48985-022, A48985-023, A48985-024, A48985-025, A48985-026, A48985-027, A48985-028, A48985-029, A48985-030, A48985-031, A48985-032, A48985-033, A48985-034, A48985-035, A48985-036, A48985-037, A48985-038, A48985-039, A48985-040, A49002-074, A49002-075, A49002-076, A49002-077, A49002-078, A49002-079, A49002-080, A49002-081, A49002-082, A49002-083, A49002-084, A50232-060, A50232-061, A50232-062, A50232-063, A50232-064, A50232-065, A50232-066, A50232-067, A50232-068, A50232-069, A50232-070, A50232-071, A50232-074, A50232-075, A50232-076, A50232-077, A50232-078, A50317-052, A50317-053, A50317-054, A50317-056, A50317-057, A50317-058, A50489-038, A50489-039, A50489-040, A50489-041, A50489-042, A50489-043, A50489-044, A50489-045, A50489-046, A50489-047, A50489-048, A50968-056, A50968-057, A50968-058, A50968-059, A50968-060, A51247-026, A51247-027, A51247-028, A51247-029, A51247-030, A51247-031, A51247-032, A51247-033, A51247-034, A51247-035, A51247-036, A51247-037, A51247-038, A51247-039, A51247-040, A51247-041, A51251-032, A51251-033, A51251-034, A51251-035, A51251-036, A51251-037, A51251-038, A51251-039, A51251-040, A51251-041, A51251-042, A51251-043, A51251-044, A51251-045, A51251-046, A51289-011, A51289-012, A51289-013, A51289-014, A51289-015, A51289-016, A51289-017, A51289-018, A51289-019, A51289-020, A51289-021, A51289-022, A51289-023, A51289-024, A51289-025, A51289-026, A51289-027, A51289-028, A51547-049, A51547-050, A51547-051, A51547-052, A51547-053, A51547-054, A51547-055, A51547-056, A51547-057, A51547-058, A51547-059, A51547-060, A51547-061, A51547-062, A51547-063, A51547-064, A51547-065, A52028-060, A52028-061, A52028-062, A52028-063, A52028-064, A52028-065, A52028-066, A52028-067, A52028-068, A52028-069, A52028-070, A52028-071, A52028-072, A52028-073, A52237-011, A52237-012, A52237-013, A52237-014, A52237-015, A52237-016, A52237-017, A52237-018, A52237-019, A52237-020, A52237-021, A52237-022, A52237-023, A52237-024, A52237-025, A52237-026, A52237-027, A52237-028, A52914-052, A52914-053, A52914-054, A52914-055, A53117-011, A53117-012, A53117-013, A53117-014, A53117-015, A53117-016, A53117-017, A53117-018, A53117-019, A53117-020, A53117-021, A53117-022, A53208-011, A53208-012, A53208-013, A53208-014, A53208-015, A53208-016, A53208-017, A53208-018, A53208-019, A53208-020, A53289-047, A53289-048, A53289-049, A53289-050, A53289-051, A53289-052, A53289-053, A53289-054, A53289-055, A53289-056, A53289-057, A53505-039, A53505-040, A53505-041, A53505-042, A53505-043, A53505-044, A53505-045, A53505-046, A53505-047, A53505-048, A53505-049, A53505-050, A53656-011, A53656-012, A53656-013, A53656-014, A53656-015, A53656-016, A53656-017, A53656-018, A53656-019, A53656-020, A53656-021, A53656-022, A53656-023, A53724-011, A53724-012, A53724-013, A53724-014, A53724-015, A53724-016, A53724-017, A53724-018, A53724-019, A53724-020, A53724-021, A53724-022, A53724-023, A53724-024, A53724-025, A53724-026, A53724-027, A53724-028, A53724-029, A53894-029, A53894-030, A53894-031, A53894-032, A53894-033, A53894-034, A53894-035, A53894-036, A53894-037, A53894-038, A53894-039, A53894-040, A53894-041, A53894-042, A53894-043, A53894-044, A53894-045, A53894-046, A53894-047, A54026-027, A54026-028, A54026-029, A54026-030, A54026-031, A54026-032, A54026-033, A54026-034, A54026-035, A54026-036, A54026-037, A54026-038, A54026-039, A54026-040, A54026-041, A54026-042, A54040-032, A54040-033, A54040-034, A54040-035, A54040-036, A54040-037, A54040-038, A54040-039, A54040-040, A54350-071, A54350-072, A54350-073, A54350-074, A54350-075, A54350-076, A54350-077, A54350-078, A54350-079, A54350-080, A54350-081, A54350-082, A54350-083, A54350-084, A54350-085, A54350-086, A54350-087, A55129-090, A55129-091, A55129-092, A55129-093, A55129-094, A55129-095, A55129-096, A55129-097, A55129-098, A55129-099, A55129-100, A55129-101, A55129-102, A55129-103, A55129-104, A55129-105, A55129-106, A55129-107;
j) DBM T41150INT 10CC GRAFTON GEL, REF T41150INT, GTIN 00643169964860, Serial Numbers: A41512-065, A41512-066, A41512-067, A41512-068, A41512-069, A41512-070, A41512-071, A41512-072, A41512-073, A41512-074, A41512-075, A41512-076, A41512-077, A41512-078, A41512-079, A41512-080, A41512-081, A41512-082, A41512-083, A41512-084, A41512-085, A41512-086, A41512-087, A41512-088, A41512-089, A41512-090, A41512-091, A42895-098, A42895-099, A42895-100, A42895-101, A42895-102, A42895-103, A42895-104, A42895-105, A42895-106, A42895-107, A42895-108, A43098-065, A43098-066, A43098-067, A43098-068, A43098-069, A43098-070, A43098-071, A43098-072, A43098-073, A43098-074, A43098-075, A43098-076, A43098-077, A43098-078, A44080-036, A44080-037, A44080-038, A44080-039, A44080-040, A44080-041, A44080-042, A44080-043, A44080-044, A44080-045, A44080-046, A45217-169, A45217-170, A45217-171, A45217-172, A45217-173, A46725-042, A46725-043, A46725-044, A46725-045, A46725-046, A46725-047, A46725-048, A47034-063, A47034-064, A47034-065, A47034-066, A47034-067, A47034-068, A47034-069, A47034-070, A47034-071, A47034-072, A47034-073, A47034-074, A47034-075, A47034-076, A47034-077, A47034-078, A47034-079, A47034-080, A47034-081, A47034-082, A47034-083, A47034-084, A48699-033, A48699-034, A48699-035, A48699-036, A48699-037, A48699-038, A48699-039, A48699-040, A48699-041, A48699-042, A48699-043, A48699-044, A48699-045, A48699-046, A48699-047, A48699-048, A48699-049, A48699-050, A48699-051, A48699-052, A48699-053, A48699-054, A48699-055, A50826-031, A50826-032, A50826-033, A50826-035, A50826-036, A50826-037, A50826-038, A50826-039, A50826-044, A51336-016, A51336-017, A51336-018, A51336-019, A51336-020, A51336-021, A51336-022, A51336-023, A51336-024, A51336-025, A51336-026, A51336-027, A51336-028, A51336-029, A51336-030, A51336-031, A51336-032, A51336-033, A51336-034, A51336-035, A51336-036, A51336-037, A51654-011, A51654-012, A51654-013, A51654-014, A51654-015, A51654-016, A51654-017, A51654-018, A51654-019, A51654-020, A51654-021, A51654-022, A51654-023, A51654-024, A51654-025, A52263-032, A52263-033, A52263-034, A52263-035, A52263-036, A52263-037, A52263-038, A52263-039, A52263-040, A52263-041, A52263-042, A52263-043, A52263-044, A52263-045, A52263-046, A52263-047, A52263-048, A52263-049, A52275-089, A52275-090, A52275-091, A52860-128, A52860-129, A52860-130, A52860-131, A52860-132, A52860-133, A52860-134, A52860-135, A52860-136, A52860-137, A52860-138, A52860-139, A52860-140, A52860-141, A52860-142, A52860-143, A53240-041, A53240-042, A54251-158, A54251-159, A54251-160, A54251-161, A54251-162, A54251-163, A54251-164, A54251-165, A54251-166, A54251-167, A54251-168, A54251-173.
|
| FEI Number |
1000477302
|
Recalling Firm/ Manufacturer |
Medtronic Sofamor Danek USA Inc 1800 Pyramid Pl Memphis TN 38132-1703
|
| For Additional Information Contact | Kyra Nead 303-886-2549 |
Manufacturer Reason for Recall | Potential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a
breach in the sterile barrier. |
FDA Determined Cause 2 | Process control |
| Action | Medtronic Sofamor Danek notified customers of the recall on about 02/08/2023, via letter. Customers were instructed to identify and quarantine any affected products, return any affected products, complete and return the Customer Confirmation Form, and provide the letter to all personnel that should be aware or where affected products have been transferred/distributed. |
| Quantity in Commerce | 10090 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Austria, Brunei Darussalam, Canada, Chile, Colombia, Denmark, Dominican Republic, Ecuador, Finland, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Israel, Jamaica, Japan, Jordan, Korea, Republic Of, Kuwait, Lebanon, Liechtenstein, Luxembourg, Macao, Malaysia, Mexico, Mongolia, Namibia, Netherlands, New Zealand, Norway, Oman, Panama, Paraguay, Peru, Philippines, Portugal, Puerto Rico, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Taiwan, Thailand, Trinidad And Tobago, United Arab Emirates, United Kingdom, Viet Nam.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = MBP
|
|
|
|