• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall medtronic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 1 Device Recall medtronic see related information
Date Initiated by Firm May 10, 2023
Date Posted June 16, 2023
Recall Status1 Open3, Classified
Recall Number Z-1710-2023
Recall Event ID 92328
PMA Number P980016S382 
Product Classification Implantable cardioverter defibrillator (non-CRT) - Product Code LWS
Product ICD-DR DDBC3D4 EVERA S IS1/DF4 GLOB, Model Number DDBC3D4; Implantable Cardioverter Defibrillators

Code Information GTIN 00643169720381, Lot Serial Numbers: BWF206541H, BWF206448H, BWF206403H, BWF206382H, BWF206552H, BWF206320H, BWF206323H, BWF206420H, BWF206376H, BWF206458H, BWF206459H, BWF206430H, BWF206453H, BWF206359H, BWF206478H, BWF206333H, BWF206365H, BWF206370H, BWF206557H, BWF206599H, BWF206607H, BWF206551H, BWF206367H, BWF206396H, BWF206421H, BWF206554H, BWF206355H, BWF206461H, BWF206604H, BWF206371H, BWF206373H, BWF206374H, BWF206378H, BWF206380H, BWF206381H, BWF206383H, BWF206387H, BWF206388H, BWF206390H, BWF206449H, BWF206473H, BWF206474H, BWF206475H, BWF206500H, BWF206526H, BWF206527H, BWF206528H, BWF206529H, BWF206535H, BWF206563H, BWF206565H, BWF206567H, BWF206568H, BWF206573H, BWF206574H, BWF206575H, BWF206597H, BWF206598H, BWF206175H, BWF206419H, BWF206327H, BWF206491H, BWF206496H, BWF206401H, BWF206321H, BWF206334H, BWF206338H, BWF206344H, BWF206432H, BWF206470H, BWF206489H, BWF206507H, BWF206558H, BWF206577H, BWF206583H, BWF206570H, BWF206581H, BWF206342H, BWF206410H, BWF206492H, BWF206417H, BWF206424H, BWF206476H, BWF206539H, BWF206580H, BWF206485H, BWF206608H, BWF206601H, BWF206317H, BWF206384H, BWF206395H, BWF206490H, BWF206585H, BWF206590H, BWF206322H, BWF206431H, BWF206443H, BWF206463H, BWF206484H, BWF206487H, BWF206497H, BWF206502H, BWF206505H, BWF206534H, BWF206553H, BWF206569H, BWF206571H, BWF206415H, BWF206451H, BWF206452H, BWF206544H, BWF206546H, BWF206460H, BWF206360H, BWF206440H, BWF206462H, BWF206340H, BWF206435H, BWF206510H, BWF206495H, BWF206375H, BWF206405H, BWF206418H, BWF206454H, BWF206465H, BWF206488H, BWF206358H, BWF206353H, BWF206397H, BWF206582H, BWF206176H, BWF206456H, BWF206589H, BWF206412H, BWF206369H, BWF206394H, BWF206426H, BWF206398H, BWF206578H, BWF206425H, BWF206318H, BWF206402H, BWF206584H, BWF206357H, BWF206400H, BWF206409H, BWF206341H, BWF206428H, BWF206392H, BWF206393H, BWF206416H, BWF206427H, BWF206437H, BWF206441H, BWF206518H, BWF206530H, BWF206533H, BWF206345H, BWF206455H, BWF206457H, BWF206594H, BWF206602H, BWF206548H, BWF206536H, BWF206347H, BWF206349H, BWF206408H, BWF206429H, BWF206434H, BWF206506H, BWF206572H, BWF206576H, BWF206587H, BWF206588H, BWF206361H, BWF206362H; GTIN 00643169720398, Lot Serial Numbers: BWF606744S, BWF606741S, BWF606219S, BWF606448S, BWF606453S, BWF606410S, BWF605918S, BWF605872S, BWF606271S, BWF605623S, BWF606336S, BWF606345S, BWF605908S, BWF605763S, BWF605654S, BWF606234S, BWF605707S, BWF605912S, BWF606049S, BWF606062S, BWF606087S, BWF606199S, BWF606352S, BWF606469S, BWF606509S, BWF605447S, BWF605448S, BWF605678S, BWF605680S, BWF606689S, BWF606695S, BWF607268S, BWF605619S, BWF606242S, BWF606472S, BWF606562S, BWF606761S, BWF605624S, BWF605625S, BWF605662S, BWF605665S, BWF606343S, BWF606613S, BWF606749S, BWF605703S, BWF605762S, BWF606269S, BWF606353S, BWF605636S, BWF606233S, BWF606703S, BWF606763S, BWF606220S, BWF605870S, BWF605664S, BWF605767S, BWF606481S, BWF605630S, BWF605917S, BWF606115S, BWF606223S, BWF606211S, BWF606239S, BWF606532S, BWF605867S, BWF605668S, BWF605628S, BWF605629S, BWF606104S, BWF606106S, BWF606327S, BWF606330S, BWF606441S, BWF606442S, BWF606737S, BWF606766S, BWF605700S, BWF606624S, BWF606750S, BWF605673S, BWF605681S, BWF606622S, BWF605754S, BWF606270S, BWF606385S, BWF606685S, BWF606625S, BWF605635S, BWF605663S, BWF605909S, BWF606493S, BWF606351S, BWF606094S, BWF606447S, BWF606454S, BWF606767S, BWF606768S, BWF605670S, BWF605877S, BWF605904S, BWF606341S, BWF606222S, BWF606152S, BWF606183S, BWF606185S, BWF606186S, BWF606285S, BWF606294S, BWF606318S, BWF606321S, BWF606322S, BWF606324S, BWF606512S, BWF606530S, BWF606542S, BWF606544S, BWF606597S, BWF606610S, BWF606455S, BWF606457S, BWF606163S, BWF606177S, BWF606179S, BWF606510S, BWF606516S, BWF606531S, BWF606534S, BWF606539S, BWF606541S, BWF605671S, BWF605756S, BWF605688S, BWF605689S, BWF606244S, BWF606187S, BWF606762S, BWF607611S, BWF607573S, BWF605959S, BWF606249S, BWF606333S, BWF606411S, BWF605702S, BWF606759S, BWF606764S, BWF606753S, BWF606758S, BWF605659S, BWF605660S, BWF605661S, BWF605704S, BWF605900S, BWF606235S, BWF606511S, BWF606440S, BWF606700S, BWF606739S, BWF605691S, BWF605631S, BWF605634S, BWF605672S, BWF606342S, BWF606523S, BWF605738S, BWF606438S, BWF605706S, BWF605901S, BWF605757S, BWF605758S, BWF605879S, BWF605880S, BWF605915S, BWF605919S, BWF605921S, BWF605923S, BWF606491S, BWF606492S, BWF605666S, BWF606101S, BWF605910S, BWF605718S, BWF605868S, BWF605674S, BWF606096S, BWF607589S, BWF606295S, BWF606334S, BWF606350S, BWF606757S, BWF605978S, BWF606557S, BWF605677S, BWF605679S, BWF605683S, BWF605687S, BWF605692S, BWF605822S, BWF605838S, BWF606589S, BWF606596S, BWF606600S, BWF606614S, BWF606960S, BWF607227S, BWF605824S, BWF605825S, BWF606421S, BWF606601S, BWF606602S, BWF606605S, BWF606812S, BWF606815S, BWF606889S, BWF606890S, BWF606948S, BWF606959S, BWF606962S, BWF607132S, BWF607134S, BWF607240S, BWF607270S, BWF607319S, BWF605806S, BWF605808S, BWF605809S, BWF605810S, BWF605821S, BWF605827S, BWF605834S, BWF605835S, BWF605840S, BWF606891S, BWF606903S, BWF605627S, BWF606496S, BWF605699S, BWF605713S, BWF605714S, BWF605715S, BWF605759S, BWF605760S, BWF605761S, BWF605764S, BWF605765S, BWF605826S, BWF605828S, BWF605831S, BWF605924S, BWF605953S, BWF605987S, BWF605989S, BWF605990S, BWF605991S, BWF606010S, BWF606011S, BWF606020S, BWF606021S, BWF606025S, BWF606029S, BWF606034S, BWF606044S, BWF606045S, BWF606046S, BWF606047S, BWF606048S, BWF606071S, BWF606078S, BWF606082S, BWF606084S, BWF606085S, BWF606107S, BWF606128S, BWF606133S, BWF606134S, BWF606140S, BWF606141S, BWF606142S, BWF606147S, BWF606148S, BWF606149S, BWF606150S, BWF606157S, BWF606171S, BWF606174S, BWF606175S, BWF606176S, BWF606178S, BWF606201S, BWF606204S, BWF606207S, BWF606691S, BWF606692S, BWF606693S, BWF606745S, BWF606746S, BWF606805S, BWF606806S, BWF606807S, BWF606810S, BWF607008S, BWF607009S, BWF607010S, BWF607076S, BWF607077S, BWF607078S, BWF607097S, BWF607098S, BWF607099S, BWF607100S, BWF607101S, BWF607127S, BWF607128S, BWF607130S, BWF607131S, BWF607133S, BWF607144S, BWF607185S, BWF607186S, BWF607192S, BWF607193S, BWF607194S, BWF607225S, BWF607226S, BWF607229S, BWF607230S, BWF607231S, BWF607242S, BWF607243S, BWF607244S, BWF607247S, BWF607252S, BWF607253S, BWF607254S, BWF607255S, BWF607256S, BWF607285S, BWF607286S, BWF607287S, BWF607318S, BWF607320S, BWF607321S, BWF607322S, BWF607323S, BWF607334S, BWF607447S, BWF607490S, BWF607491S, BWF607495S, BWF607496S, BWF607497S, BWF607498S, BWF607499S, BWF607518S, BWF607519S, BWF607520S, BWF607521S, BWF607522S, BWF607545S, BWF607546S, BWF607547S, BWF607550S, BWF607551S, BWF607552S, BWF607553S, BWF607554S, BWF607557S, BWF607558S, BWF607559S, BWF607560S, BWF607564S, BWF607567S, BWF607568S, BWF607569S, BWF607570S, BWF607571S, BWF607577S, BWF607578S, BWF607579S, BWF607580S, BWF607581S, BWF607583S, BWF607584S, BWF607585S, BWF607586S, BWF607587S, BWF607588S, BWF607590S, BWF607591S, BWF607592S, BWF607593S, BWF607594S, BWF607595S, BWF607597S, BWF607598S, BWF607599S, BWF607600S, BWF607601S, BWF605643S, BWF605637S, BWF605684S, BWF605708S, BWF605710S, BWF605733S, BWF605734S, BWF605735S, BWF605736S, BWF605737S, BWF605739S, BWF605740S, BWF605741S, BWF605742S, BWF605743S, BWF605744S, BWF605745S, BWF605746S, BWF605747S, BWF605748S, BWF605751S, BWF605766S, BWF605784S, BWF605802S, BWF605803S, BWF605903S, BWF605906S, BWF605911S, BWF605922S, BWF605949S, BWF605960S, BWF605961S, BWF605962S, BWF605965S, BWF605966S, BWF605967S, BWF605968S, BWF605969S, BWF606012S, BWF606013S, BWF606036S, BWF606037S, BWF606090S, BWF606098S, BWF606105S, BWF606113S, BWF606114S, BWF606116S, BWF606117S, BWF606118S, BWF606119S, BWF606120S, BWF606122S, BWF606123S, BWF606124S, BWF606125S, BWF606126S, BWF606143S, BWF606144S, BWF606146S, BWF606156S, BWF606168S, BWF606169S, BWF606180S, BWF606182S, BWF606184S, BWF606188S, BWF606189S, BWF606190S, BWF606686S, BWF606687S, BWF606688S, BWF606690S, BWF606697S, BWF606698S, BWF606704S, BWF606735S, BWF606743S, BWF606823S, BWF606824S, BWF606885S, BWF606886S, BWF606888S, BWF607012S, BWF607013S, BWF607014S, BWF607015S, BWF607016S, BWF607017S, BWF607018S, BWF607079S, BWF607080S, BWF607089S, BWF607090S, BWF607091S, BWF607092S, BWF607135S, BWF607141S, BWF607142S, BWF607260S, BWF607261S, BWF607265S, BWF607266S, BWF607267S, BWF607269S, BWF607271S, BWF607272S, BWF607273S, BWF607274S, BWF607278S, BWF607280S, BWF607281S, BWF607282S, BWF607283S, BWF607284S, BWF607291S, BWF607292S, BWF607293S, BWF607294S, BWF607327S, BWF607328S, BWF607331S, BWF607332S, BWF607449S, BWF607451S, BWF607452S, BWF607454S, BWF607455S, BWF607456S, BWF607458S, BWF607460S, BWF607461S, BWF607462S, BWF606306S, BWF605652S, BWF605626S, BWF605633S, BWF605645S, BWF606515S, BWF606519S, BWF606520S, BWF606521S, BWF606528S, BWF606536S, BWF606546S, BWF606547S, BWF606549S, BWF606556S, BWF606558S, BWF605640S, BWF605642S, BWF605650S, BWF606524S, BWF606517S, BWF606518S, BWF606099S, BWF606145S, BWF606153S, BWF605770S, BWF605805S, BWF605807S, BWF605820S, BWF605823S, BWF605830S, BWF605836S, BWF605837S, BWF605845S, BWF606240S, BWF606537S, BWF606559S, BWF606181S, BWF606543S, BWF605685S, BWF605907S, BWF605769S, BWF605789S, BWF605793S, BWF605871S, BWF605925S, BWF605938S, BWF605939S, BWF605941S, BWF605942S, BWF605944S, BWF605946S, BWF605957S, BWF606050S, BWF606051S, BWF606052S, BWF606053S, BWF606054S, BWF606055S, BWF606056S, BWF606057S, BWF606058S, BWF606059S, BWF606191S, BWF606205S, BWF606255S, BWF606256S, BWF606289S, BWF606291S, BWF606293S, BWF606298S, BWF606299S, BWF606300S, BWF606303S, BWF606305S, BWF606307S, BWF606347S, BWF606431S, BWF606433S, BWF606434S, BWF606436S, BWF605669S, BWF605778S, BWF605783S, BWF605785S, BWF605787S, BWF605878S, BWF605931S, BWF605932S, BWF605933S, BWF605935S, BWF605936S, BWF605937S, BWF605947S, BWF605954S, BWF605955S, BWF605956S, BWF606348S, BWF606354S, BWF606355S, BWF606356S, BWF606357S, BWF606358S, BWF606359S, BWF606360S, BWF606361S, BWF606362S, BWF606363S, BWF606364S, BWF605620S, BWF605720S, BWF606482S, BWF606162S, BWF607315S, BWF605913S, BWF606751S, BWF605771S, BWF605772S, BWF605774S, BWF605777S, BWF605779S, BWF605797S, BWF605798S, BWF605981S, BWF605982S, BWF606014S, BWF606015S, BWF606016S, BWF606017S, BWF606026S, BWF606035S, BWF606074S, BWF606076S, BWF606111S, BWF606112S, BWF606132S, BWF606138S, BWF606139S, BWF606216S, BWF606218S, BWF606694S, BWF606696S, BWF606736S, BWF606752S, BWF606754S, BWF606898S, BWF606899S, BWF606900S, BWF606964S, BWF607005S, BWF607022S, BWF607065S, BWF607066S, BWF607067S, BWF607085S, BWF607086S, BWF607087S, BWF607088S, BWF607093S, BWF607094S, BWF607095S, BWF607277S, BWF607288S, BWF607289S, BWF607324S, BWF607325S, BWF607326S, BWF607329S, BWF607444S, BWF607459S, BWF607463S, BWF607465S, BWF607472S, BWF607474S, BWF607475S, BWF607486S, BWF607492S, BWF607493S, BWF607494S, BWF607506S, BWF607507S, BWF607509S, BWF607517S, BWF607565S, BWF607566S, BWF605786S, BWF605788S, BWF605792S, BWF605795S, BWF605972S, BWF605973S, BWF605976S, BWF605977S, BWF605979S, BWF606007S, BWF606009S, BWF606038S, BWF606041S, BWF606042S, BWF606043S, BWF606086S, BWF606088S, BWF606089S, BWF606108S, BWF606109S, BWF606110S, BWF606164S, BWF606165S, BWF606135S, BWF606893S, BWF606945S, BWF605446S, BWF605632S, BWF605622S, BWF605676S, BWF605693S, BWF605694S, BWF605696S, BWF605811S, BWF605814S, BWF605817S, BWF605829S, BWF605832S, BWF606383S, BWF606406S, BWF606427S, BWF606591S, BWF606607S, BWF606774S, BWF606808S, BWF606809S, BWF606818S, BWF606822S, BWF606887S, BWF606894S, BWF606895S, BWF606897S, BWF606946S, BWF606951S, BWF606963S, BWF605841S, BWF605799S, BWF605800S, BWF605801S, BWF605812S, BWF606022S, BWF606030S, BWF606033S, BWF606039S, BWF606040S, BWF606081S, BWF606103S, BWF606151S, BWF606154S, BWF606155S, BWF606170S, BWF606699S, BWF606901S, BWF606902S, BWF607125S, BWF607126S, BWF607136S, BWF607137S, BWF607241S, BWF607245S, BWF607246S, BWF607251S, BWF607448S, BWF607450S, BWF607453S, BWF607470S, BWF607471S, BWF607612S, BWF605639S, BWF605641S, BWF605644S, BWF605646S, BWF605647S, BWF605648S, BWF605649S, BWF605651S, BWF605653S, BWF605655S, BWF605656S, BWF605695S, BWF605749S, BWF605750S, BWF605752S, BWF605753S, BWF605846S, BWF605847S, BWF605848S, BWF605849S, BWF605850S, BWF605851S, BWF605852S, BWF605853S, BWF605854S, BWF605855S, BWF605856S, BWF605857S, BWF605859S, BWF605865S, BWF605866S, BWF605869S, BWF605873S, BWF605881S, BWF605882S, BWF605883S, BWF605884S, BWF605885S, BWF605886S, BWF605887S, BWF605888S, BWF605889S, BWF605890S, BWF605891S, BWF605892S, BWF605893S, BWF605894S, BWF605895S, BWF606225S, BWF606227S, BWF606228S, BWF606229S, BWF606232S, BWF606259S, BWF606260S, BWF606261S, BWF606262S, BWF606263S, BWF606264S, BWF606265S, BWF606266S, BWF606267S, BWF606268S, BWF606386S, BWF606387S, BWF606429S, BWF606449S, BWF606450S, BWF606452S, BWF606456S, BWF606459S, BWF606460S, BWF606461S, BWF606462S, BWF606463S, BWF606464S, BWF606465S, BWF606466S, BWF606467S, BWF606468S, BWF606470S, BWF606471S, BWF606473S, BWF606474S, BWF606475S, BWF606476S, BWF606477S, BWF606478S, BWF606479S, BWF606480S, BWF606483S, BWF606485S, BWF606486S, BWF606487S, BWF606488S, BWF606489S, BWF606490S, BWF606495S, BWF606522S, BWF606527S, BWF606529S, BWF606553S, BWF606555S, BWF606563S, BWF606564S, BWF606565S, BWF606566S, BWF606567S, BWF606568S, BWF606569S, BWF606570S, BWF606571S, BWF606572S, BWF606573S, BWF606574S, BWF606586S, BWF606251S, BWF606254S, BWF606257S, BWF606258S, BWF606275S, BWF606280S, BWF606282S, BWF605709S, BWF605729S, BWF605730S, BWF605732S, BWF605813S, BWF605815S, BWF605816S, BWF605818S, BWF605839S, BWF605842S, BWF605899S, BWF605902S, BWF605929S, BWF605934S, BWF605984S, BWF605985S, BWF605986S, BWF605993S, BWF605999S, BWF606000S, BWF606001S, BWF606002S, BWF606003S, BWF606005S, BWF606091S, BWF606092S, BWF606093S, BWF606095S, BWF606097S, BWF606100S, BWF606102S, BWF606136S, BWF606137S, BWF606159S, BWF606160S, BWF606166S, BWF606167S, BWF606172S, BWF606173S, BWF606192S, BWF606193S, BWF606202S, BWF606203S, BWF606208S, BWF606209S, BWF606221S, BWF606226S, BWF606230S, BWF606231S, BWF606237S, BWF606238S, BWF606738S, BWF606740S, BWF606742S, BWF606748S, BWF606811S, BWF606813S, BWF606814S, BWF606816S, BWF606817S, BWF606819S, BWF606820S, BWF606821S, BWF606892S, BWF606947S, BWF606949S, BWF606950S, BWF606952S, BWF606953S, BWF606954S, BWF606955S, BWF606956S, BWF606957S, BWF606958S, BWF606961S, BWF607006S, BWF607007S, BWF607011S, BWF607023S, BWF607024S, BWF607068S, BWF607069S, BWF607070S, BWF607071S, BWF607072S, BWF607083S, BWF607084S, BWF607096S, BWF607102S, BWF607103S, BWF607139S, BWF607140S, BWF607187S, BWF607188S, BWF607189S, BWF607190S, BWF607191S, BWF607228S, BWF607232S, BWF607233S, BWF607234S, BWF607235S, BWF607236S, BWF607237S, BWF607238S, BWF607257S, BWF607258S, BWF607259S, BWF607262S, BWF607263S, BWF607264S, BWF607275S, BWF607276S, BWF607316S, BWF607317S, BWF607330S, BWF607333S, BWF607443S, BWF607445S, BWF607446S, BWF607457S, BWF607464S, BWF607466S, BWF607467S, BWF607468S, BWF607469S, BWF607476S, BWF607477S, BWF607478S, BWF607479S, BWF607480S, BWF607481S, BWF607482S, BWF607483S, BWF607484S, BWF607485S, BWF607487S, BWF607488S, BWF607489S, BWF607502S, BWF607503S, BWF607504S, BWF607505S, BWF607523S, BWF607531S, BWF607532S, BWF607533S, BWF607544S, BWF607555S, BWF607556S, BWF607561S, BWF607562S, BWF607563S, BWF607572S, BWF607574S, BWF607582S, BWF607602S, BWF607603S, BWF607604S, BWF607605S, BWF607606S, BWF606904S, BWF605833S, BWF605952S, BWF606006S, BWF605755S, BWF605667S, BWF605697S, BWF605698S, BWF605725S, BWF605728S, BWF605731S, BWF606121S, BWF606236S, BWF605768S, BWF605773S, BWF605775S, BWF605780S, BWF605858S, BWF605860S, BWF605861S, BWF605862S, BWF605863S, BWF605958S, BWF605864S, BWF605657S, BWF605711S, BWF605712S, BWF605727S, BWF606027S, BWF606028S, BWF606127S, BWF606161S, BWF605675S, BWF606422S, BWF606592S, BWF606593S, BWF606594S, BWF606599S, BWF606604S, BWF606609S, BWF606612S, BWF606896S, BWF607250S, BWF605781S, BWF605930S, BWF605940S, BWF605943S, BWF605951S, BWF605963S, BWF605964S, BWF605970S, BWF605975S, BWF605980S, BWF605983S, BWF605992S, BWF606008S, BWF606031S, BWF606032S, BWF606060S, BWF606061S, BWF606063S, BWF606064S, BWF606065S, BWF606067S, BWF606068S, BWF606069S, BWF606070S, BWF606072S, BWF606073S, BWF606075S, BWF606077S, BWF606278S, BWF606279S, BWF606281S, BWF606283S, BWF606320S, BWF606323S, BWF606326S, BWF606328S, BWF606329S, BWF606337S, BWF606338S, BWF606339S, BWF606340S, BWF606349S, BWF606368S, BWF606369S, BWF606370S, BWF606371S, BWF606372S, BWF606373S, BWF606374S, BWF606375S, BWF606376S, BWF606377S, BWF606378S, BWF606379S, BWF606380S, BWF606381S, BWF606382S, BWF606384S, BWF606388S, BWF606389S, BWF606390S, BWF606391S, BWF606392S, BWF606393S, BWF606394S, BWF606395S, BWF606396S, BWF606400S, BWF606401S, BWF606402S, BWF606403S, BWF606404S, BWF606405S, BWF606412S, BWF606413S, BWF606414S, BWF606415S, BWF605658S, BWF605717S, BWF605719S, BWF605722S, BWF605723S, BWF605726S, BWF605945S, BWF605948S, BWF605950S, BWF605994S, BWF605995S, BWF605996S, BWF605997S, BWF605998S, BWF606018S, BWF606019S, BWF606024S, BWF606080S, BWF606129S, BWF606130S, BWF606131S, BWF606158S, BWF606194S, BWF606196S, BWF606197S, BWF606198S, BWF606210S, BWF606212S, BWF606215S, BWF607019S, BWF607020S, BWF607021S, BWF607073S, BWF607074S, BWF607075S, BWF607081S, BWF607082S, BWF607104S, BWF607248S, BWF607249S, BWF607290S, BWF607138S, BWF607239S, BWF605844S, BWF607473S, BWF607500S, BWF607501S, BWF607508S, BWF607524S, BWF607525S, BWF607540S, BWF607541S, BWF607542S, BWF607510S, BWF607511S, BWF607512S, BWF607513S, BWF607514S, BWF607515S, BWF607516S, BWF607526S, BWF607528S, BWF607529S, BWF607530S, BWF607534S, BWF607535S, BWF607537S, BWF607538S, BWF607539S, BWF607607S, BWF607608S, BWF607609S, BWF607610S, BWF607536S, BWF607548S, BWF607549S; GTIN 00643169980235, Lot Serial numbers: BWF606855S, BWF606721S, BWF606825S, BWF607821S, BWF606879S, BWF609295S, BWF606865S, BWF606936S, BWF607044S, BWF606657S, BWF606776S, BWF606778S, BWF606779S, BWF606783S, BWF607025S, BWF607308S, BWF607372S, BWF607624S, BWF607827S, BWF606853S, BWF606874S, BWF606854S, BWF609815S, BWF607736S, BWF607989S, BWF608784S, BWF607899S, BWF608521S, BWF609015S, BWF609341S, BWF607037S, BWF607206S, BWF606710S, BWF607814S, BWF607819S, BWF606671S, BWF606797S, BWF606877S, BWF606909S, BWF606716S, BWF606859S, BWF606860S, BWF607375S, BWF609071S, BWF609072S, BWF606719S, BWF606828S, BWF606918S, BWF607052S, BWF606715S, BWF606784S, BWF606826S, BWF606827S, BWF606985S, BWF607029S, BWF607030S, BWF606578S, BWF606656S, BWF607036S, BWF608785S, BWF606646S, BWF606664S, BWF607377S, BWF606670S, BWF606880S, BWF606986S, BWF606991S, BWF606992S, BWF607109S, BWF607110S, BWF607207S, BWF606910S, BWF606647S, BWF606780S, BWF606849S, BWF606884S, BWF606995S, BWF606994S, BWF607047S, BWF607054S, BWF607735S, BWF606873S, BWF606725S, BWF606788S, BWF606506S, BWF606579S, BWF606585S, BWF606654S, BWF606876S, BWF606878S, BWF606965S, BWF606937S, BWF606628S, BWF607033S, BWF606502S, BWF606504S, BWF606872S, BWF606881S, BWF606648S, BWF608806S, BWF608809S, BWF609021S, BWF609022S, BWF609026S, BWF609036S, BWF609145S, BWF608134S, BWF608148S, BWF608274S, BWF608275S, BWF608318S, BWF608321S, BWF608323S, BWF608326S, BWF608340S, BWF608432S, BWF608471S, BWF608610S, BWF608733S, BWF608734S, BWF608782S, BWF608786S, BWF608840S, BWF609018S, BWF609020S, BWF609033S, BWF609073S, BWF609190S, BWF609476S, BWF609478S, BWF609482S, BWF609483S, BWF609485S, BWF609689S, BWF609724S, BWF609730S, BWF608260S, BWF608261S, BWF608485S, BWF608685S, BWF609687S, BWF606650S, BWF606651S, BWF606658S, BWF606660S, BWF606709S, BWF606718S, BWF606724S, BWF606786S, BWF606798S, BWF606837S, BWF606851S, BWF606919S, BWF606922S, BWF606966S, BWF606967S, BWF607049S, BWF609355S, BWF609775S, BWF607035S, BWF607829S, BWF608085S, BWF608212S, BWF608928S, BWF608988S, BWF609002S, BWF609348S, BWF609395S, BWF609713S, BWF608474S, BWF608611S, BWF608707S, BWF609042S, BWF609144S, BWF609403S, BWF609405S, BWF609438S, BWF609350S, BWF609692S, BWF609726S, BWF608607S, BWF608705S, BWF606649S, BWF606652S, BWF606653S, BWF606655S, BWF606659S, BWF606661S, BWF606667S, BWF606717S, BWF606841S, BWF606971S, BWF607050S, BWF607106S, BWF607107S, BWF607108S, BWF607111S.BWF607112S, BWF607209S, BWF607295S, BWF607296S, BWF607297S, BWF607298S, BWF607340S, BWF607341S, BWF607342S, BWF607343S, BWF607618S, BWF607623S, BWF607743S, BWF607750S, BWF607755S, BWF607815S, BWF607853S, BWF607864S, BWF607900S, BWF607902S, BWF607982S, BWF607987S, BWF607988S, BWF608080S, BWF608081S, BWF608083S, BWF608117S, BWF608118S, BWF608135S, BWF608136S, BWF608149S, BWF608150S, BWF608158S, BWF608162S, BWF608167S, BWF608169S, BWF608468S, BWF608469S, BWF608470S, BWF608475S, BWF608491S, BWF608594S, BWF608595S, BWF608600S, BWF608601S, BWF608602S, BWF608603S, BWF608604S, BWF608605S, BWF608612S, BWF608619S, BWF608702S, BWF608704S, BWF608732S, BWF608787S, BWF608843S, BWF608849S, BWF608861S, BWF608872S, BWF608918S, BWF608975S, BWF608977S, BWF608993S, BWF609016S, BWF609043S, BWF609045S, BWF609046S, BWF609047S, BWF609048S, BWF609049S, BWF609058S, BWF609080S, BWF609204S, BWF609362S, BWF609369S, BWF609370S, BWF609375S, BWF609378S, BWF609382S, BWF609393S, BWF609399S, BWF609415S, BWF609418S, BWF609420S, BWF609421S, BWF609424S, BWF609427S, BWF609433S, BWF609439S, BWF609441S, BWF609470S, BWF609474S, BWF609475S, BWF609693S, BWF609694S, BWF609698S, BWF609704S, BWF609705S, BWF609707S, BWF609711S, BWF609715S, BWF609716S, BWF609717S, BWF609718S, BWF609719S, BWF609720S, BWF609721S, BWF609722S, BWF606787S, BWF606799S, BWF606800S, BWF606830S, BWF606846S, BWF606970S, BWF607051S, BWF607053S, BWF607893S, BWF607983S, BWF607985S, BWF608133S, BWF608168S, BWF608269S, BWF608270S, BWF608615S, BWF609041S, BWF609074S, BWF609242S, BWF609377S, BWF609380S, BWF609690S, BWF609691S, BWF609723S, BWF609725S, BWF609727S, BWF609729S, BWF609188S, BWF609479S, BWF608759S, BWF609447S, BWF606875S, BWF607368S, BWF609029S, BWF606867S, BWF606883S, BWF606866S, BWF606850S, BWF607043S, BWF606905S, BWF606990S, BWF607991S, BWF606576S, BWF606580S, BWF606584S, BWF607048S, BWF606627S, BWF606796S, BWF609039S, BWF606583S, BWF608142S, BWF608116S, BWF608200S, BWF608171S, BWF606499S, BWF606500S, BWF606501S, BWF606636S, BWF606637S, BWF606640S, BWF606641S, BWF606642S, BWF606643S, BWF606869S, BWF607026S, BWF607027S, BWF607038S, BWF607300S, BWF607302S, BWF607804S, BWF607807S, BWF607808S, BWF607810S, BWF607811S, BWF607813S, BWF607990S, BWF608019S, BWF608199S, BWF608204S, BWF608228S, BWF608466S, BWF608543S, BWF608801S, BWF608933S, BWF608938S, BWF608941S, BWF608963S, BWF608967S, BWF608969S, BWF609429S, BWF609448S, BWF609526S, BWF609665S, BWF606864S, BWF606497S, BWF606498S, BWF606629S, BWF606630S, BWF608208S, BWF607751S, BWF609203S, BWF608467S, BWF608031S, BWF607374S, BWF607816S, BWF607041S, BWF606968S, BWF609392S, BWF609397S, BWF607210S, BWF608137S, BWF608139S, BWF608472S, BWF608539S, BWF608606S, BWF608608S, BWF608790S, BWF608805S, BWF608842S, BWF609014S, BWF609017S, BWF609044S, BWF609181S, BWF609335S, BWF609345S, BWF609389S, BWF609477S, BWF609481S, BWF609688S, BWF609731S, BWF609732S, BWF608322S, BWF607699S, BWF607700S, BWF607879S, BWF607968S, BWF607974S, BWF608064S, BWF608140S, BWF608252S, BWF608392S, BWF608395S, BWF608413S, BWF608427S, BWF608503S, BWF608716S, BWF608232S, BWF607369S, BWF607389S, BWF607408S, BWF608290S, BWF608291S, BWF608418S, BWF608877S, BWF607868S, BWF607708S, BWF607721S, BWF607726S, BWF607730S, BWF607367S, BWF607376S, BWF607403S, BWF607407S, BWF607421S, BWF607435S, BWF607440S, BWF607441S, BWF607617S, BWF607619S, BWF607620S, BWF607621S, BWF607637S, BWF607641S, BWF607662S, BWF607694S, BWF607701S, BWF607707S, BWF607733S, BWF607761S, BWF607874S, BWF607897S, BWF607996S, BWF608043S, BWF608074S, BWF608075S, BWF608123S, BWF608161S, BWF608189S, BWF608194S, BWF608220S, BWF608224S, BWF608241S, BWF608243S, BWF608263S, BWF608292S, BWF608294S, BWF608303S, BWF608399S, BWF607632S, BWF607655S, BWF607767S, BWF607862S, BWF608036S, BWF608072S, BWF608266S, BWF606505S, BWF606868S, BWF608838S, BWF608839S, BWF608847S, BWF607702S, BWF607703S, BWF607704S, BWF607706S, BWF607709S, BWF607741S, BWF607772S, BWF607812S, BWF607045S, BWF606913S, BWF608138S, BWF608141S, BWF608259S, BWF608265S, BWF608267S, BWF608703S, BWF608760S, BWF608763S, BWF608789S, BWF608791S, BWF609193S, BWF609195S, BWF609197S, BWF609198S, BWF609225S, BWF609227S, BWF609241S, BWF609290S, BWF609294S, BWF609296S, BWF609298S, BWF609342S, BWF609344S, BWF607400S, BWF607402S, BWF607404S, BWF607405S, BWF607410S, BWF607411S, BWF607412S, BWF607413S, BWF607414S, BWF607415S, BWF607416S, BWF607417S, BWF607418S, BWF607419S, BWF607420S, BWF607422S, BWF608174S, BWF608175S, BWF608176S, BWF608177S, BWF608178S, BWF608179S, BWF608180S, BWF608181S, BWF608182S, BWF608183S, BWF608184S, BWF608185S, BWF608187S, BWF608188S, BWF608190S, BWF608198S, BWF608229S, BWF608230S, BWF608231S, BWF608233S, BWF608234S, BWF608235S, BWF608236S, BWF608238S, BWF608239S, BWF608240S, BWF608242S, BWF608244S, BWF608245S, BWF608247S, BWF606803S, BWF606804S, BWF606829S, BWF606834S, BWF606852S, BWF606856S, BWF606857S, BWF606920S, BWF606969S, BWF607423S, BWF607424S, BWF607860S, BWF607984S, BWF607986S, BWF609191S, BWF609472S, BWF609406S, BWF607314S, BWF606907S, BWF607861S, BWF607366S, BWF607613S, BWF607616S, BWF607858S, BWF608165S, BWF608264S, BWF607032S, BWF608147S, BWF607757S, BWF607758S, BWF607760S, BWF608030S, BWF608054S, BWF608084S, BWF608086S, BWF608087S, BWF608089S, BWF608090S, BWF608091S, BWF608092S, BWF608096S, BWF608102S, BWF608105S, BWF608110S, BWF608112S, BWF608114S, BWF608115S, BWF608125S, BWF608130S, BWF608131S, BWF608160S, BWF608186S, BWF608205S, BWF608206S, BWF608207S, BWF608213S, BWF608214S, BWF608216S, BWF608217S, BWF608218S, BWF608219S, BWF608221S, BWF608222S, BWF608223S, BWF608225S, BWF608268S, BWF608276S, BWF608277S, BWF608278S, BWF608279S, BWF608280S, BWF608281S, BWF608282S, BWF608283S, BWF608284S, BWF608286S, BWF608293S, BWF608295S, BWF608296S, BWF608297S, BWF608298S, BWF608299S, BWF608300S, BWF608301S, BWF608302S, BWF608304S, BWF608305S, BWF608306S, BWF608307S, BWF608312S, BWF608431S, BWF608434S, BWF608436S, BWF608437S, BWF608440S, BWF608442S, BWF608445S, BWF608446S, BWF608447S, BWF608448S, BWF608450S, BWF608452S, BWF608453S, BWF608454S, BWF608455S, BWF608456S, BWF608460S, BWF608461S, BWF608462S, BWF608463S, BWF608464S, BWF608509S, BWF608510S, BWF608511S, BWF608512S, BWF608513S, BWF608515S, BWF608516S, BWF608518S, BWF608519S, BWF608520S, BWF608522S, BWF608523S, BWF608524S, BWF608525S, BWF608527S, BWF608528S, BWF608529S, BWF608530S, BWF608531S, BWF608536S, BWF608538S, BWF608540S, BWF608544S, BWF608546S, BWF608548S, BWF608577S, BWF608579S, BWF608590S, BWF608591S, BWF608596S, BWF608597S, BWF608609S, BWF608618S, BWF608633S, BWF608642S, BWF608643S, BWF608646S, BWF608648S, BWF608650S, BWF608654S, BWF608655S, BWF608656S, BWF608661S, BWF608662S, BWF608664S, BWF608665S, BWF608735S, BWF608736S, BWF608737S, BWF608738S, BWF608739S, BWF608741S, BWF608742S, BWF608743S, BWF608744S, BWF608745S, BWF608746S, BWF608747S, BWF608748S, BWF608749S, BWF608750S, BWF608751S, BWF608753S, BWF608754S, BWF608755S, BWF608756S, BWF608757S, BWF608758S, BWF608761S, BWF608762S, BWF608765S, BWF608766S, BWF608767S, BWF608769S, BWF608770S, BWF608771S, BWF608772S, BWF608773S, BWF608774S, BWF608775S, BWF608777S, BWF608778S, BWF608779S, BWF608780S, BWF608781S, BWF608794S, BWF608795S, BWF608796S, BWF608798S, BWF608800S, BWF608802S, BWF608803S, BWF608804S, BWF608807S, BWF608808S, BWF608811S, BWF608812S, BWF608813S, BWF608818S, BWF608820S, BWF608821S, BWF608822S, BWF608823S, BWF608825S, BWF608827S, BWF608830S, BWF608831S, BWF608832S, BWF608835S, BWF608836S, BWF608994S, BWF609003S, BWF609004S, BWF609005S, BWF609006S, BWF609007S, BWF609008S, BWF609009S, BWF609010S, BWF609011S, BWF609012S, BWF609013S, BWF609023S, BWF609024S, BWF609051S, BWF609052S, BWF609054S, BWF609056S, BWF609057S, BWF609060S, BWF609061S, BWF609062S, BWF609063S, BWF609064S, BWF609078S, BWF609079S, BWF609194S, BWF609209S, BWF609210S, BWF609214S, BWF609215S, BWF609216S, BWF609217S, BWF609218S, BWF609219S, BWF609221S, BWF609334S, BWF605897S, BWF605898S, BWF606633S, BWF608164S, BWF608342S, BWF608792S, BWF608810S, BWF609030S, BWF609075S, BWF609083S, BWF609186S, BWF609196S, BWF609383S, BWF609387S, BWF609059S, BWF609553S, BWF606577S, BWF606775S, BWF606863S, BWF606917S, BWF607762S, BWF607763S, BWF607823S, BWF608048S, BWF608119S, BWF608124S, BWF608172S, BWF608173S, BWF608210S, BWF608211S, BWF608324S, BWF608325S, BWF608533S, BWF608535S, BWF608631S, BWF608634S, BWF609360S, BWF609401S, BWF609359S, BWF607822S, BWF607835S, BWF607970S, BWF608155S, BWF608351S, BWF608384S, BWF608385S, BWF608414S, BWF608424S, BWF608487S, BWF608507S, BWF608559S, BWF608613S, BWF608691S, BWF608720S, BWF607749S, BWF607994S, BWF609223S, BWF609408S, BWF609465S, BWF609142S, BWF606632S, BWF606634S, BWF606635S, BWF606638S, BWF606639S, BWF606988S, BWF607784S, BWF607785S, BWF607786S, BWF607789S, BWF607790S, BWF607793S, BWF607794S, BWF607796S, BWF607349S, BWF607393S, BWF607399S, BWF607401S, BWF607406S, BWF607409S, BWF607425S, BWF607615S, BWF607622S, BWF607630S, BWF607633S, BWF607640S, BWF607693S, BWF607705S, BWF607870S, BWF607884S, BWF608073S, BWF608215S, BWF608246S, BWF608271S, BWF608272S, BWF608273S, BWF608285S, BWF607720S, BWF607722S, BWF607728S, BWF607742S, BWF607744S, BWF607746S, BWF607748S, BWF607809S, BWF607981S, BWF606672S, BWF606673S, BWF606674S, BWF606675S, BWF606677S, BWF606678S, BWF606680S, BWF606681S, BWF606682S, BWF606683S, BWF606705S, BWF606706S, BWF606707S, BWF606708S, BWF606713S, BWF606714S, BWF606726S, BWF606727S, BWF606728S, BWF606729S, BWF606730S, BWF606731S, BWF606732S, BWF606733S, BWF606734S, BWF606791S, BWF606792S, BWF606793S, BWF606794S, BWF606795S, BWF606831S, BWF606833S, BWF606835S, BWF606836S, BWF606839S, BWF606840S, BWF606842S, BWF606844S, BWF606845S, BWF606847S, BWF606921S, BWF606923S, BWF606924S, BWF606925S, BWF606926S, BWF606927S, BWF606929S, BWF606930S, BWF606931S, BWF606932S, BWF606933S, BWF606934S, BWF606935S, BWF606938S, BWF606939S, BWF606941S, BWF606942S, BWF606943S, BWF606944S, BWF606973S, BWF606974S, BWF606975S, BWF606976S, BWF606977S, BWF606978S, BWF606979S, BWF606980S, BWF606981S, BWF606982S, BWF606983S, BWF606984S, BWF606996S, BWF606997S, BWF606998S, BWF606999S, BWF607000S, BWF607001S, BWF607002S, BWF607003S, BWF607004S, BWF607055S, BWF607056S, BWF607057S, BWF607058S, BWF607059S, BWF607060S, BWF607061S, BWF607062S, BWF607063S, BWF607064S, BWF607113S, BWF607114S, BWF607115S, BWF607116S, BWF607117S, BWF607118S, BWF607119S, BWF607120S, BWF607121S, BWF607122S, BWF607123S, BWF607124S, BWF607145S, BWF607146S, BWF607147S, BWF607148S, BWF607149S, BWF607150S, BWF607151S, BWF607152S, BWF607153S, BWF607154S, BWF607155S, BWF607156S, BWF607157S, BWF607158S, BWF607159S, BWF607160S, BWF607161S, BWF607162S, BWF607163S, BWF607164S, BWF607165S, BWF607166S, BWF607167S, BWF607168S, BWF607169S, BWF607170S, BWF607171S, BWF607172S, BWF607173S, BWF607174S, BWF607175S, BWF607176S, BWF607177S, BWF607178S, BWF607179S, BWF607180S, BWF607181S, BWF607182S, BWF607183S, BWF607184S, BWF607195S, BWF607196S, BWF607197S, BWF607198S, BWF607199S, BWF607200S, BWF607201S, BWF607202S, BWF607211S, BWF607212S, BWF607213S, BWF607214S, BWF607215S, BWF607216S, BWF607217S, BWF607218S, BWF607219S, BWF607220S, BWF607221S, BWF607222S, BWF607223S, BWF607299S, BWF607301S, BWF607303S, BWF607304S, BWF607305S, BWF607306S, BWF607307S, BWF607309S, BWF607310S, BWF607311S, BWF607312S, BWF607313S, BWF607344S, BWF607345S, BWF607346S, BWF607347S, BWF607348S, BWF607379S, BWF607380S, BWF607381S, BWF607382S, BWF607383S, BWF607384S, BWF607385S, BWF607386S, BWF607387S, BWF607388S, BWF607390S, BWF607391S, BWF607392S, BWF607394S, BWF607395S, BWF607396S, BWF607397S, BWF607398S, BWF607426S, BWF607427S, BWF607428S, BWF607429S, BWF607430S, BWF607431S, BWF607432S, BWF607434S, BWF607436S, BWF607437S, BWF607438S, BWF607439S, BWF607442S, BWF607625S, BWF607626S, BWF607627S, BWF607628S, BWF607629S, BWF607631S, BWF607634S, BWF607635S, BWF607636S, BWF607638S, BWF607639S, BWF607642S, BWF607643S, BWF607644S, BWF607645S, BWF607646S, BWF607648S, BWF607649S, BWF607651S, BWF607652S, BWF607653S, BWF607654S, BWF607656S, BWF607657S, BWF607660S, BWF607663S, BWF607664S, BWF607665S.BWF607666S, BWF607669S, BWF607670S, BWF607673S, BWF607674S, BWF607675S, BWF607676S, BWF607677S, BWF607678S, BWF607679S, BWF607680S, BWF607681S, BWF607682S, BWF607683S, BWF607684S, BWF607685S, BWF607686S, BWF607689S, BWF607690S, BWF607692S, BWF607764S, BWF607765S, BWF607766S, BWF607768S, BWF607769S, BWF607770S, BWF607831S, BWF607833S, BWF607834S, BWF607836S, BWF607838S, BWF607841S, BWF607842S, BWF607847S, BWF607849S, BWF607850S, BWF607852S, BWF607854S, BWF607855S, BWF607865S, BWF607866S, BWF607869S, BWF607871S, BWF607872S, BWF607873S, BWF607875S, BWF607876S, BWF607877S, BWF607878S, BWF607880S, BWF607883S, BWF607885S, BWF607886S, BWF607887S, BWF607888S, BWF607890S, BWF607891S, BWF607892S, BWF607908S, BWF607909S, BWF607912S, BWF607913S, BWF607916S, BWF607917S, BWF607918S, BWF607919S, BWF607923S, BWF607924S, BWF607925S, BWF607926S, BWF607927S, BWF607928S, BWF607929S, BWF607931S, BWF607934S, BWF607936S, BWF607938S, BWF607939S, BWF607969S, BWF607995S, BWF607997S, BWF607998S, BWF607999S, BWF608000S, BWF608001S, BWF608002S, BWF608003S, BWF608004S, BWF608005S, BWF608006S, BWF608007S, BWF608008S, BWF608009S, BWF608010S, BWF608011S, BWF608012S, BWF608013S, BWF608014S, BWF608015S, BWF608016S, BWF608017S, BWF608018S, BWF608020S, BWF608021S, BWF608022S, BWF608023S, BWF608024S, BWF608025S, BWF608026S, BWF608027S, BWF608028S, BWF608029S, BWF608032S, BWF608034S, BWF608035S, BWF608037S, BWF608038S, BWF608039S, BWF608040S, BWF608041S, BWF608042S, BWF608044S, BWF608045S, BWF608046S, BWF608047S, BWF608050S, BWF608051S, BWF608052S, BWF608053S, BWF608055S, BWF608056S, BWF608057S, BWF608058S, BWF608059S, BWF608060S, BWF608061S, BWF608062S, BWF608063S, BWF608066S, BWF608067S, BWF608068S, BWF608069S, BWF608071S, BWF608077S, BWF608078S, BWF608079S, BWF608093S, BWF608094S, BWF608095S, BWF608097S, BWF608098S, BWF608099S, BWF608100S, BWF608101S, BWF608103S, BWF608104S, BWF608106S, BWF608107S, BWF608108S, BWF608109S, BWF608226S, BWF608227S, BWF608248S, BWF608249S, BWF608251S, BWF608253S, BWF608254S, BWF608255S, BWF608256S, BWF608287S, BWF608288S, BWF608308S, BWF608309S, BWF608310S, BWF608311S, BWF608327S, BWF608330S, BWF608331S, BWF608332S, BWF608333S, BWF608334S, BWF608335S, BWF608336S, BWF608337S, BWF608338S, BWF608339S, BWF608343S, BWF608344S, BWF608345S, BWF608346S, BWF608347S, BWF608532S, BWF608541S, BWF608545S, BWF608555S, BWF608556S, BWF608557S, BWF608558S, BWF608560S, BWF608561S, BWF608562S, BWF608563S, BWF608564S, BWF608568S, BWF608569S, BWF608620S, BWF608621S, BWF608623S, BWF608625S, BWF608626S, BWF608629S, BWF608635S, BWF608636S, BWF608637S, BWF608640S, BWF608641S, BWF608659S, BWF608666S, BWF608979S, BWF608984S, BWF608987S, BWF609025S, BWF609027S, BWF609028S, BWF609031S, BWF609032S, BWF609034S, BWF609035S, BWF609037S, BWF609038S, BWF609040S, BWF609065S, BWF609067S, BWF609131S, BWF609147S, BWF609148S, BWF609149S, BWF609150S, BWF609152S, BWF609153S, BWF609155S, BWF609158S, BWF609159S, BWF609161S, BWF609162S, BWF609163S, BWF609164S, BWF609166S, BWF609168S, BWF609170S, BWF609172S, BWF609174S, BWF609177S, BWF609185S, BWF609187S, BWF609199S, BWF609201S, BWF609207S, BWF609208S, BWF609213S, BWF609243S, BWF609244S, BWF609245S, BWF609248S, BWF609249S, BWF609252S, BWF609255S, BWF609256S, BWF609258S, BWF609259S, BWF609260S, BWF609261S, BWF609262S, BWF609264S, BWF609266S, BWF609267S, BWF609268S, BWF609269S, BWF609271S, BWF609272S, BWF609275S, BWF609277S, BWF609278S, BWF609283S, BWF609284S, BWF609291S, BWF609292S, BWF609293S, BWF609306S, BWF609311S, BWF609315S, BWF609351S, BWF609352S, BWF609353S, BWF609354S, BWF609356S, BWF609357S, BWF609366S, BWF609371S, BWF609372S, BWF609376S, BWF609379S, BWF609381S, BWF609386S, BWF609413S, BWF609445S, BWF609462S, BWF609467S, BWF609468S, BWF609469S, BWF609471S, BWF609480S, BWF609484S, BWF609489S, BWF609490S, BWF609492S, BWF609495S, BWF607659S, BWF607667S, BWF607668S, BWF607691S, BWF607695S, BWF607696S, BWF607698S, BWF607729S, BWF607731S, BWF607732S, BWF607773S, BWF607775S, BWF607776S, BWF607783S, BWF607787S, BWF607788S, BWF607791S, BWF607792S, BWF607795S, BWF607797S, BWF607798S, BWF607799S, BWF607800S, BWF607802S, BWF607803S, BWF607805S, BWF607806S, BWF607824S, BWF607851S, BWF607910S, BWF607911S, BWF607914S, BWF607937S, BWF607941S, BWF607944S, BWF607945S, BWF607948S, BWF607949S, BWF607956S, BWF607961S, BWF607966S, BWF607971S, BWF607972S, BWF607976S, BWF607980S, BWF608126S, BWF608128S, BWF608144S, BWF608153S, BWF608154S, BWF608156S, BWF608348S, BWF608349S, BWF608350S, BWF608359S, BWF608362S, BWF608369S, BWF608372S, BWF608382S, BWF608383S, BWF608390S, BWF608407S, BWF608410S, BWF608421S, BWF608422S, BWF608429S, BWF608441S, BWF608443S, BWF608444S, BWF608449S, BWF608473S, BWF608476S, BWF608478S, BWF608479S, BWF608480S, BWF608492S, BWF608526S, BWF608534S, BWF608542S, BWF608554S, BWF608565S, BWF608566S, BWF608574S, BWF608575S, BWF608576S, BWF608578S, BWF608580S, BWF608582S, BWF608583S, BWF608586S, BWF608587S, BWF608588S, BWF608589S, BWF608592S, BWF608653S, BWF608657S, BWF608663S, BWF608668S, BWF608678S, BWF608679S, BWF608680S, BWF608681S, BWF608686S, BWF608687S, BWF608698S, BWF608699S, BWF608731S, BWF608793S, BWF608819S, BWF608824S, BWF608826S, BWF608828S, BWF608829S, BWF608841S, BWF608858S, BWF608863S, BWF608869S, BWF608870S, BWF608871S, BWF608873S, BWF608894S, BWF608900S, BWF608901S, BWF608902S, BWF608903S, BWF608904S, BWF608906S, BWF608912S, BWF608913S, BWF608914S, BWF608915S, BWF608923S, BWF608926S, BWF608927S, BWF608929S, BWF608930S, BWF608970S, BWF608971S, BWF608976S, BWF608978S, BWF608990S, BWF608991S, BWF609001S, BWF609069S, BWF609171S, BWF609257S, BWF609316S, BWF609319S, BWF609324S, BWF609325S, BWF609327S, BWF609337S, BWF609338S, BWF609339S, BWF609358S, BWF609404S, BWF609414S, BWF609416S, BWF609443S, BWF609444S, BWF609446S, BWF609449S, BWF609450S, BWF609510S, BWF609512S, BWF609513S, BWF609514S, BWF609515S, BWF609531S, BWF609542S, BWF609543S, BWF609544S, BWF609545S, BWF609570S, BWF609571S, BWF609572S, BWF609573S, BWF609574S, BWF609597S, BWF609600S, BWF609603S, BWF609604S, BWF609605S, BWF609606S, BWF609608S, BWF609609S, BWF609624S, BWF609625S, BWF609626S, BWF609635S, BWF609636S, BWF609638S, BWF609641S, BWF609642S, BWF609658S, BWF609659S, BWF609661S, BWF609662S, BWF609663S, BWF609664S, BWF609666S, BWF609667S, BWF609668S, BWF609669S, BWF609670S, BWF609671S, BWF609673S, BWF609674S, BWF609675S, BWF609676S, BWF609682S, BWF609683S, BWF609684S, BWF609697S, BWF609708S, BWF609709S, BWF609710S, BWF609776S, BWF609790S, BWF609791S, BWF609794S, BWF609795S, BWF609796S, BWF609805S, BWF609806S, BWF609855S, BWF609856S, BWF608387S, BWF608412S, BWF609055S, BWF609169S, BWF609364S, BWF609508S, BWF609598S, BWF609601S, BWF609637S, BWF609685S, BWF609700S, BWF609706S, BWF609737S, BWF609738S, BWF609799S, BWF609851S, BWF609852S, BWF606663S, BWF606781S, BWF606782S, BWF606871S, BWF609287S, BWF609288S, BWF609299S, BWF609308S, BWF609309S, BWF609310S, BWF609313S, BWF609314S, BWF609317S, BWF609318S, BWF607335S, BWF607905S, BWF607907S, BWF607915S, BWF607920S, BWF608191S, BWF608192S, BWF608196S, BWF608197S, BWF608257S, BWF608258S, BWF608313S, BWF608712S, BWF608713S, BWF608718S, BWF607781S, BWF607373S, BWF607614S, BWF607647S, BWF607697S, BWF607856S, BWF607881S, BWF607882S, BWF608033S, BWF608193S, BWF608201S, BWF608262S, BWF608403S, BWF608406S, BWF608409S, BWF608624S, BWF608697S, BWF607336S, BWF607337S, BWF607338S, BWF607339S, BWF607350S, BWF607351S, BWF607352S, BWF607353S, BWF607354S, BWF607921S, BWF607922S, BWF607930S, BWF607935S, BWF607940S, BWF607942S, BWF607943S, BWF607947S, BWF607950S, BWF607959S, BWF607962S, BWF607965S, BWF607973S, BWF607975S, BWF607979S, BWF608672S, BWF608673S, BWF608674S, BWF608675S, BWF608676S, BWF608677S, BWF608683S, BWF608684S, BWF608688S, BWF608689S, BWF608690S, BWF608693S, BWF608694S, BWF608844S, BWF608845S, BWF608846S, BWF608394S, BWF608415S, BWF608416S, BWF608417S, BWF608420S, BWF608425S, BWF608428S, BWF608430S, BWF608433S, BWF608435S, BWF608438S, BWF608439S, BWF608459S, BWF608465S, BWF608481S, BWF608495S, BWF608496S, BWF608497S, BWF608498S, BWF608499S, BWF608500S, BWF608501S, BWF608502S, BWF608614S, BWF608616S, BWF608617S, BWF608622S, BWF608627S, BWF608628S, BWF608630S, BWF608632S, BWF608638S, BWF608644S, BWF608645S, BWF608647S, BWF608649S, BWF608651S, BWF608814S, BWF608816S, BWF608817S, BWF608862S, BWF608867S, BWF608868S, BWF608886S, BWF608887S, BWF608890S, BWF608891S, BWF608895S, BWF608896S, BWF608897S, BWF608898S, BWF608899S, BWF608910S, BWF608924S, BWF608925S, BWF608935S, BWF608936S, BWF608939S, BWF608944S, BWF608950S, BWF608951S, BWF608952S, BWF608961S, BWF608962S, BWF608964S, BWF608981S, BWF608982S, BWF608983S, BWF608985S, BWF608996S, BWF608997S, BWF608998S, BWF608999S, BWF609000S, BWF609066S, BWF609068S, BWF609100S, BWF609104S, BWF609107S, BWF609156S, BWF609157S, BWF609160S, BWF609165S, BWF609167S, BWF609189S, BWF609205S, BWF609206S, BWF609211S, BWF609246S, BWF609247S, BWF609250S, BWF609388S, BWF609390S, BWF609391S, BWF609394S, BWF609396S, BWF609422S, BWF609430S, BWF609431S, BWF609432S, BWF609440S, BWF609442S, BWF609457S, BWF609458S, BWF609459S, BWF609461S, BWF609463S, BWF609464S, BWF609487S, BWF609500S, BWF609501S, BWF609502S, BWF609503S, BWF609504S, BWF609506S, BWF609507S, BWF609532S, BWF609533S, BWF609534S, BWF609535S, BWF609536S, BWF609537S, BWF609538S, BWF609539S, BWF609540S, BWF609541S, BWF609546S, BWF609547S, BWF609548S, BWF609549S, BWF609550S, BWF609559S, BWF609560S, BWF609561S, BWF609562S, BWF609563S, BWF609564S, BWF609575S, BWF609576S, BWF609577S, BWF609578S, BWF609579S, BWF609580S, BWF609581S, BWF609582S, BWF609583S, BWF609584S, BWF609585S, BWF609586S, BWF609587S, BWF609588S, BWF609589S, BWF609590S, BWF609592S, BWF609593S, BWF609594S, BWF609614S, BWF609615S, BWF609616S, BWF609617S, BWF609618S, BWF609630S, BWF609632S, BWF609633S, BWF609644S, BWF609645S, BWF609646S, BWF609647S, BWF609648S, BWF609649S, BWF609650S, BWF609651S, BWF609652S, BWF609653S, BWF609654S, BWF609655S, BWF609672S, BWF609677S, BWF609678S, BWF609679S, BWF609680S, BWF609714S, BWF609728S, BWF609742S, BWF609744S, BWF609746S, BWF609749S, BWF609752S, BWF609753S, BWF609754S, BWF609760S, BWF609762S, BWF609765S, BWF609812S, BWF609813S, BWF609814S, BWF609817S, BWF609819S, BWF609820S, BWF609824S, BWF609825S, BWF609826S, BWF609845S, BWF609846S, BWF609847S, BWF609854S, BWF609861S, BWF609862S, BWF609863S, BWF609864S, BWF609865S, BWF609866S, BWF609873S, BWF609874S, BWF609875S, BWF609879S, BWF609881S, BWF609883S, BWF607203S, BWF607204S, BWF607224S, BWF607365S, BWF608888S, BWF609081S, BWF609097S, BWF609251S, BWF609423S, BWF609516S, BWF609596S, BWF607718S, BWF607719S, BWF607740S, BWF607754S, BWF607756S, BWF607774S, BWF607801S, BWF607830S, BWF607832S, BWF607955S, BWF608120S, BWF608129S, BWF608143S, BWF608353S, BWF608354S, BWF608360S, BWF608361S, BWF608374S, BWF608380S, BWF608386S, BWF608396S, BWF608397S, BWF608477S, BWF608482S, BWF608508S, BWF608514S, BWF608517S, BWF608570S, BWF608571S, BWF608852S, BWF608854S, BWF608864S, BWF608865S, BWF608940S, BWF608980S, BWF609050S, BWF609053S, BWF609154S, BWF609200S, BWF609300S, BWF609304S, BWF609328S, BWF609340S, BWF609385S, BWF609425S, BWF609436S, BWF609528S, BWF609529S, BWF609530S, BWF609599S, BWF609602S, BWF609607S, BWF609702S, BWF609703S, BWF609712S, BWF609750S, BWF609774S, BWF609784S, BWF609785S, BWF609869S, BWF609870S, BWF609876S, BWF607671S, BWF607672S, BWF607687S, BWF607710S, BWF607711S, BWF607712S, BWF607713S, BWF607714S, BWF607715S, BWF607716S, BWF607717S, BWF607723S, BWF607724S, BWF607725S, BWF607771S, BWF607778S, BWF607779S, BWF607780S, BWF607782S, BWF607817S, BWF607818S, BWF607820S, BWF607828S, BWF607839S, BWF607840S, BWF607844S, BWF607845S, BWF607846S, BWF607848S, BWF607895S, BWF607901S, BWF607946S, BWF607958S, BWF607964S, BWF607978S, BWF608145S, BWF608151S, BWF608320S, BWF608352S, BWF608358S, BWF608367S, BWF608368S, BWF608373S, BWF608379S, BWF608381S, BWF608388S, BWF608389S, BWF608401S, BWF608426S, BWF608451S, BWF608457S, BWF608458S, BWF608483S, BWF608484S, BWF608486S, BWF608488S, BWF608490S, BWF608504S, BWF608505S, BWF608547S, BWF608549S, BWF608581S.BWF608584S, BWF608585S, BWF608723S, BWF608768S, BWF608776S, BWF608797S, BWF608833S, BWF608850S, BWF608853S, BWF608855S, BWF608856S, BWF608857S, BWF608874S, BWF608880S, BWF608881S, BWF608885S, BWF608905S, BWF608907S, BWF608916S, BWF608932S, BWF608946S, BWF608953S, BWF608992S, BWF609091S, BWF609094S, BWF609095S, BWF609096S, BWF609098S, BWF609099S, BWF609115S, BWF609117S, BWF609119S, BWF609130S, BWF609138S, BWF609143S, BWF609146S, BWF609151S, BWF609173S, BWF609175S, BWF609183S, BWF609184S, BWF609320S, BWF609321S, BWF609322S, BWF609323S, BWF609326S, BWF609329S, BWF609361S, BWF609363S, BWF609384S, BWF609417S, BWF609419S, BWF609437S, BWF609497S, BWF609498S, BWF609499S, BWF609524S, BWF609525S, BWF609527S, BWF609558S, BWF609595S, BWF609610S, BWF609611S, BWF609612S, BWF609619S, BWF609620S, BWF609621S, BWF609622S, BWF609627S, BWF609628S, BWF609629S, BWF609643S, BWF609656S, BWF609657S, BWF609660S, BWF609699S, BWF609701S, BWF609733S, BWF609734S, BWF609736S, BWF609757S, BWF609758S, BWF609786S, BWF609787S, BWF609788S, BWF609789S, BWF609792S, BWF609793S, BWF609797S, BWF609798S, BWF609803S, BWF609804S, BWF609843S, BWF609844S, BWF609848S, BWF609849S, BWF609853S, BWF609867S, BWF609868S, BWF609872S, BWF609880S, BWF609884S, BWF607661S, BWF607727S, BWF607734S, BWF607737S, BWF607738S, BWF607825S, BWF607826S, BWF607843S, BWF607903S, BWF607906S, BWF607932S, BWF607933S, BWF607951S, BWF607952S, BWF607953S, BWF607954S, BWF607957S, BWF607960S, BWF608121S, BWF608122S, BWF608146S, BWF608152S, BWF608157S, BWF608159S, BWF608163S, BWF608237S, BWF608250S, BWF608355S, BWF608356S, BWF608357S, BWF608363S, BWF608364S, BWF608365S, BWF608366S, BWF608370S, BWF608371S, BWF608375S, BWF608376S, BWF608377S, BWF608378S, BWF608391S, BWF608398S, BWF608400S, BWF608402S, BWF608405S, BWF608411S, BWF608489S, BWF608493S, BWF608494S, BWF608550S, BWF608551S, BWF608552S, BWF608553S, BWF608567S, BWF608572S, BWF608573S, BWF608593S, BWF608598S, BWF608599S, BWF608639S, BWF608652S, BWF608658S, BWF608660S, BWF608682S, BWF608692S, BWF608696S, BWF608701S, BWF608706S, BWF608710S, BWF608711S, BWF608740S, BWF608752S, BWF608764S, BWF608788S, BWF608815S, BWF608834S, BWF608859S, BWF608875S, BWF608878S, BWF608879S, BWF608882S, BWF608883S, BWF608884S, BWF608892S, BWF608893S, BWF608911S, BWF608917S, BWF608920S, BWF608921S, BWF608922S, BWF608931S, BWF608948S, BWF608949S, BWF608954S, BWF608955S, BWF608957S, BWF608958S, BWF608960S, BWF608965S, BWF608966S, BWF608972S, BWF609070S, BWF609076S, BWF609077S, BWF609082S, BWF609085S, BWF609086S, BWF609087S, BWF609088S, BWF609089S, BWF609111S, BWF609139S, BWF609176S, BWF609253S, BWF609254S, BWF609263S, BWF609265S, BWF609270S, BWF609301S, BWF609302S, BWF609303S, BWF609305S, BWF609307S, BWF609312S, BWF609330S, BWF609336S, BWF609349S, BWF609367S, BWF609368S, BWF609373S, BWF609374S, BWF609398S, BWF609400S, BWF609402S, BWF609407S, BWF609409S, BWF609410S, BWF609411S, BWF609412S, BWF609426S, BWF609428S, BWF609453S, BWF609454S, BWF609455S, BWF609493S, BWF609511S, BWF609517S, BWF609518S, BWF609519S, BWF609520S, BWF609521S, BWF609554S, BWF609555S, BWF609556S, BWF609565S, BWF609566S, BWF609567S, BWF609568S, BWF609569S, BWF609634S, BWF609686S, BWF609695S, BWF609696S, BWF609735S, BWF609751S, BWF609755S, BWF609756S, BWF609759S, BWF609800S, BWF609835S, BWF609836S, BWF609837S, BWF609838S, BWF609839S, BWF609840S, BWF609841S, BWF609842S, BWF609850S, BWF609871S, BWF609877S, BWF609878S, BWF608166S, BWF608314S, BWF608315S, BWF608316S, BWF608317S, BWF608328S, BWF608329S, BWF608537S, BWF608667S, BWF608669S, BWF608670S, BWF608671S, BWF608695S, BWF608700S, BWF608708S, BWF608709S, BWF608714S, BWF608715S, BWF608717S, BWF608719S, BWF608721S, BWF608722S, BWF608724S, BWF608725S, BWF608726S, BWF608727S, BWF608728S, BWF608729S, BWF608730S, BWF609084S, BWF609092S, BWF609093S, BWF609101S, BWF609102S, BWF609103S, BWF609105S, BWF609106S, BWF609108S, BWF609109S, BWF609110S, BWF609112S, BWF609113S, BWF609114S, BWF609116S, BWF609118S, BWF609120S, BWF609121S, BWF609122S, BWF609123S, BWF609124S, BWF609125S, BWF609126S, BWF609127S, BWF609128S, BWF609129S, BWF609132S, BWF609133S, BWF609134S, BWF609135S, BWF609136S, BWF609137S, BWF609140S, BWF609178S, BWF609179S, BWF609180S, BWF609192S, BWF608908S, BWF608909S, BWF608919S, BWF608934S, BWF608945S, BWF608947S, BWF608968S, BWF609182S, BWF609331S, BWF609332S, BWF609333S, BWF609434S, BWF609435S, BWF609456S, BWF609509S, BWF609522S, BWF609523S, BWF609639S, BWF609640S, BWF609681S, BWF609748S, BWF609801S, BWF609802S, BWF609807S, BWF609505S, BWF609740S, BWF609613S, BWF609857S, BWF609858S, BWF609859S, BWF609860S, BWF609957S, BWF609958S, BWF609959S; GTIN 00763000612108, Lot Serial Numbers: BWF609994S, BWF609996S, BWF609998S, BWF609987S, BWF609988S, BWF609989S, BWF609990S, BWF609991S, BWF609997S, BWF610000S, BWF610001S, BWF610012S, BWF610026S, BWF609993S, BWF610020S, BWF610002S, BWF610005S
Recalling Firm/
Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
8200 Coral Sea St Ne
Mounds View MN 55112-4391
For Additional Information Contact Technical Services
800-929-4043
Manufacturer Reason
for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA Determined
Cause 2
Under Investigation by firm
Action Medtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Quantity in Commerce 4056 units
Distribution US Nationwide - Worldwide Distribution
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA Database PMAs with Product Code = LWS and Original Applicant = MEDTRONIC CARDIAC RHYTHM DISEASE MANAGEMENT
PMAs with Product Code = LWS and Original Applicant = Medtronic, Inc.
-
-