• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall medtronic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 1 Device Recall medtronic see related information
Date Initiated by Firm May 10, 2023
Date Posted June 16, 2023
Recall Status1 Open3, Classified
Recall Number Z-1749-2023
Recall Event ID 92328
PMA Number P010031S442 
Product Classification Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) - Product Code NIK
Product CRT-D DTBC2D1 BRAVA IS1/DF1 INTL, Model Number DTBC2D1; Implantable Cardioverter Defibrillators

Code Information GTIN 00643169720367, Lot Serial Numbers: BLW607464S, BLW607601S, BLW607643S, BLW607705S, BLW607921S, BLW608132S, BLW608133S, BLW608134S, BLW608135S, BLW608235S, BLW608746S, BLW609769S, BLW606989S, BLW607809S, BLW607902S, BLW607917S, BLW607284S, BLW607004S, BLW607851S, BLW607173S, BLW607320S, BLW607015S, BLW607319S, BLW607257S, BLW607269S, BLW607346S, BLW607361S, BLW607031S, BLW607243S, BLW606959S, BLW607058S, BLW607285S, BLW607580S, BLW607394S, BLW607249S, BLW607505S, BLW607573S, BLW606999S, BLW607254S, BLW607835S, BLW607159S, BLW606948S, BLW606961S, BLW607044S, BLW607113S, BLW607292S, BLW607406S, BLW607498S, BLW607661S, BLW607758S, BLW607759S, BLW607865S, BLW607248S, BLW607866S, BLW607021S, BLW607598S, BLW607665S, BLW607823S, BLW607493S, BLW607001S, BLW607817S, BLW607841S, BLW607883S, BLW607747S, BLW608268S, BLW607287S, BLW607356S, BLW607440S, BLW607473S, BLW607636S, BLW607709S, BLW607943S, BLW606941S, BLW607330S, BLW607029S, BLW607205S, BLW607207S, BLW607818S, BLW607896S, BLW607847S, BLW606952S, BLW607639S, BLW607640S, BLW607209S, BLW607230S, BLW607859S, BLW607300S, BLW606991S, BLW607037S, BLW607252S, BLW607468S, BLW607477S, BLW607513S, BLW607863S, BLW607927S, BLW607946S, BLW607188S, BLW607268S, BLW607360S, BLW607412S, BLW607478S, BLW607602S, BLW607679S, BLW608772S, BLW609755S, BLW607064S, BLW607428S, BLW607613S, BLW607055S, BLW607169S, BLW606979S, BLW606932S, BLW606963S, BLW607041S, BLW607099S, BLW607104S, BLW607274S, BLW607275S, BLW607616S, BLW607617S, BLW607723S, BLW607807S, BLW606949S, BLW606967S, BLW607253S, BLW607299S, BLW607318S, BLW607711S, BLW606958S, BLW607701S, BLW607837S, BLW607879S, BLW607920S, BLW607514S, BLW607596S, BLW607547S, BLW607646S, BLW606945S, BLW606946S, BLW606993S, BLW607053S, BLW607304S, BLW607327S, BLW607604S, BLW607717S, BLW607293S, BLW607165S, BLW607298S, BLW607374S, BLW607649S, BLW607754S, BLW607432S, BLW607195S, BLW607641S, BLW607813S, BLW606929S, BLW607106S, BLW607164S, BLW607942S, BLW607683S, BLW607901S, BLW606997S, BLW607481S, BLW607279S, BLW607521S, BLW607704S, BLW607297S, BLW607761S, BLW607471S, BLW606982S, BLW607007S, BLW607211S, BLW607583S, BLW607634S, BLW606983S, BLW607002S, BLW607090S, BLW607225S, BLW607465S, BLW607516S, BLW607560S, BLW607700S, BLW607810S, BLW606930S, BLW606995S, BLW607171S, BLW607331S, BLW607357S, BLW607398S, BLW607429S, BLW607476S, BLW607497S, BLW607504S, BLW607506S, BLW607445S, BLW607864S, BLW606947S, BLW606953S, BLW607036S, BLW607877S, BLW607899S, BLW607718S, BLW607753S, BLW607938S, BLW607940S, BLW606981S, BLW607157S, BLW607393S, BLW607495S, BLW607520S, BLW607567S, BLW607651S, BLW607740S, BLW607751S, BLW607941S, BLW607699S, BLW606955S, BLW606971S, BLW606996S, BLW607193S, BLW606944S, BLW607335S, BLW607517S, BLW607666S, BLW607051S, BLW607372S, BLW607061S, BLW607158S, BLW607166S, BLW607523S, BLW607798S, BLW606939S, BLW607196S, BLW607433S, BLW606966S, BLW607849S, BLW607265S, BLW607451S, BLW607614S, BLW607190S, BLW607027S, BLW607282S, BLW607309S, BLW607881S, BLW607884S, BLW607886S, BLW607020S, BLW607006S, BLW607529S, BLW607101S, BLW607745S, BLW607748S, BLW608201S, BLW608316S, BLW607491S, BLW607577S, BLW608118S, BLW608431S, BLW608077S, BLW608188S, BLW608426S, BLW608429S, BLW608773S, BLW607277S, BLW607328S, BLW607422S, BLW607235S, BLW607789S, BLW607097S, BLW607175S, BLW607524S, BLW607937S, BLW607030S, BLW607162S, BLW607013S, BLW607038S, BLW607888S, BLW607349S, BLW607480S, BLW607858S, BLW607396S, BLW607834S, BLW606992S, BLW607682S, BLW607933S, BLW606974S, BLW607042S, BLW607250S, BLW607716S, BLW607010S, BLW607203S, BLW607251S, BLW607310S, BLW607334S, BLW607345S, BLW607452S, BLW607572S, BLW607635S, BLW606980S, BLW606984S, BLW607059S, BLW607105S, BLW607150S, BLW607483S, BLW607484S, BLW607684S, BLW607685S, BLW607928S, BLW607929S, BLW607930S, BLW607201S, BLW607561S, BLW607680S, BLW607035S, BLW607091S, BLW607172S, BLW607296S, BLW607399S, BLW607472S, BLW607502S, BLW607003S, BLW607170S, BLW607202S, BLW607337S, BLW607400S, BLW607402S, BLW607404S, BLW607431S, BLW607475S, BLW607507S, BLW607611S, BLW607050S, BLW607386S, BLW607388S, BLW607389S, BLW607390S, BLW607391S, BLW607267S, BLW607644S, BLW607714S, BLW607935S, BLW607426S, BLW607515S, BLW607612S, BLW607742S, BLW607785S, BLW607631S, BLW607853S, BLW607290S, BLW607364S, BLW607063S, BLW607336S, BLW607595S, BLW607844S, BLW607922S, BLW607161S, BLW607365S, BLW607600S, BLW607827S, BLW607874S, BLW607427S, BLW607924S, BLW607317S, BLW607518S, BLW607288S, BLW607289S, BLW607308S, BLW607474S, BLW607485S, BLW607610S, BLW607932S, BLW607633S, BLW607341S, BLW607840S, BLW607301S, BLW607829S, BLW607414S, BLW607842S, BLW606935S, BLW606940S, BLW607040S, BLW607107S, BLW607187S, BLW607220S, BLW607223S, BLW607295S, BLW607326S, BLW607415S, BLW607416S, BLW607437S, BLW607489S, BLW607574S, BLW607609S, BLW607624S, BLW607625S, BLW607629S, BLW607630S, BLW607667S, BLW607725S, BLW607726S, BLW607446S, BLW607655S, BLW607872S, BLW607219S, BLW607662S, BLW607291S, BLW607526S, BLW607548S, BLW607825S, BLW606987S, BLW607066S, BLW607070S, BLW606978S, BLW607034S, BLW607778S, BLW607923S, BLW607790S, BLW607903S, BLW607273S, BLW607712S, BLW607046S, BLW607054S, BLW607755S, BLW606950S, BLW607008S, BLW607508S, BLW607562S, BLW607581S, BLW607706S, BLW607710S, BLW607784S, BLW607362S, BLW607397S, BLW607425S, BLW606964S, BLW607821S, BLW607167S, BLW607174S, BLW607638S, BLW607875S, BLW607944S, BLW609765S, BLW606968S, BLW607093S, BLW607102S, BLW607756S, BLW607312S, BLW607232S, BLW607025S, BLW607850S, BLW609208S, BLW607448S, BLW606960S, BLW607466S, BLW607757S, BLW606973S, BLW607531S, BLW607392S, BLW607811S, BLW607244S, BLW607079S, BLW607080S, BLW607860S, BLW607018S, BLW607148S, BLW607168S, BLW609766S, BLW609789S, BLW609790S, BLW606956S, BLW607419S, BLW607421S, BLW606598S, BLW606969S, BLW606972S, BLW606986S, BLW606994S, BLW606998S, BLW607000S, BLW607089S, BLW607311S, BLW607342S, BLW607405S, BLW607418S, BLW607439S, BLW607469S, BLW607470S, BLW607512S, BLW607525S, BLW607594S, BLW607599S, BLW607668S, BLW607819S, BLW607873S, BLW607925S, BLW607043S, BLW607132S, BLW607417S, BLW607519S, BLW607664S, BLW607892S, BLW607098S, BLW607278S, BLW607045S, BLW607047S, BLW607049S, BLW607619S, BLW607620S, BLW607621S, BLW607622S, BLW607623S, BLW607669S, BLW607670S, BLW607671S, BLW607678S, BLW607067S, BLW607069S, BLW607071S, BLW607072S, BLW607073S, BLW607075S, BLW607076S, BLW607077S, BLW607078S, BLW607088S, BLW607237S, BLW607238S, BLW607762S, BLW607764S, BLW607766S, BLW607768S, BLW607769S, BLW607771S, BLW607773S, BLW607774S, BLW607775S, BLW607776S, BLW607815S, BLW607065S, BLW607467S, BLW607593S, BLW607824S, BLW607026S, BLW607094S, BLW607286S, BLW607347S, BLW608237S, BLW609756S, BLW609757S, BLW609759S, BLW609764S, BLW606597S, BLW607739S, BLW607017S, BLW606599S, BLW606954S, BLW607033S, BLW607153S, BLW607160S, BLW607236S, BLW607294S, BLW607344S, BLW607447S, BLW607522S, BLW607565S, BLW607647S, BLW606942S, BLW607022S, BLW607118S, BLW607189S, BLW607438S, BLW607490S, BLW607570S, BLW607703S, BLW607283S, BLW607242S, BLW607500S, BLW607741S, BLW607887S, BLW607005S, BLW607845S, BLW606976S, BLW607487S, BLW607501S, BLW607648S, BLW607713S, BLW607803S, BLW607804S, BLW607828S, BLW607836S, BLW606957S, BLW607509S, BLW607028S, BLW607082S, BLW607096S, BLW607057S, BLW607401S, BLW607442S, BLW607450S, BLW607494S, BLW607645S, BLW607720S, BLW607434S, BLW607530S, BLW606938S, BLW606977S, BLW607435S, BLW607436S, BLW607443S, BLW607449S, BLW607839S, BLW606943S, BLW607885S, BLW606975S, BLW607831S, BLW607482S, BLW607569S, BLW607882S, BLW607212S, BLW607338S, BLW607413S, BLW606970S, BLW607204S, BLW607423S, BLW607453S, BLW607563S, BLW607603S, BLW607805S, BLW607891S, BLW607900S, BLW607496S, BLW607628S, BLW607637S, BLW607650S, BLW607846S, BLW607322S, BLW607147S, BLW607131S, BLW607133S, BLW607134S, BLW607144S, BLW607149S, BLW607151S, BLW607152S, BLW607154S, BLW607155S, BLW607156S, BLW607656S, BLW607657S, BLW607658S, BLW607659S, BLW607660S, BLW607904S, BLW607905S, BLW607906S, BLW607907S, BLW607908S, BLW607339S, BLW606990S, BLW607627S, BLW608066S, BLW608074S, BLW607936S, BLW607229S, BLW607245S, BLW607302S, BLW607348S, BLW607642S, BLW607681S, BLW607686S, BLW607693S, BLW607746S, BLW607760S, BLW607765S, BLW607772S, BLW607812S, BLW607833S, BLW607897S, BLW607914S, BLW608272S, BLW608275S, BLW608295S, BLW608366S, BLW608730S, BLW608735S, BLW608737S, BLW608739S, BLW608742S, BLW608887S, BLW608889S, BLW608891S, BLW608892S, BLW608903S, BLW608904S, BLW609210S, BLW607228S, BLW607246S, BLW607307S, BLW607350S, BLW607363S, BLW607663S, BLW607707S, BLW607744S, BLW607763S, BLW607770S, BLW607880S, BLW607893S, BLW607895S, BLW607931S, BLW608271S, BLW608273S, BLW608427S, BLW608428S, BLW608430S, BLW608432S, BLW608444S, BLW608552S, BLW608732S, BLW608741S, BLW608896S, BLW608897S, BLW608905S, BLW608906S, BLW607210S, BLW607215S, BLW607217S, BLW607224S, BLW607214S, BLW607303S, BLW607315S, BLW607675S, BLW607838S, BLW607909S, BLW607915S, BLW607919S, BLW608266S, BLW608281S, BLW608433S, BLW608442S, BLW608731S, BLW608744S, BLW608899S, BLW608901S, BLW608902S, BLW607332S, BLW607511S, BLW607571S, BLW607424S, BLW607441S, BLW607324S, BLW607340S, BLW607375S, BLW607385S, BLW607539S, BLW607552S, BLW607553S, BLW607692S, BLW607696S, BLW607722S, BLW607728S, BLW608291S, BLW608547S, BLW608550S, BLW608551S, BLW608553S, BLW607532S, BLW607537S, BLW607550S, BLW607691S, BLW607738S, BLW607313S, BLW607409S, BLW607410S, BLW607727S, BLW607916S, BLW608079S, BLW608555S, BLW607371S, BLW607549S, BLW607555S, BLW607695S, BLW608541S, BLW607729S, BLW607370S, BLW607889S, BLW608197S, BLW608267S, BLW607721S, BLW607316S, BLW607407S, BLW607541S, BLW607590S, BLW607687S, BLW607688S, BLW607730S, BLW607731S, BLW608081S, BLW608189S, BLW608190S, BLW608241S, BLW608242S, BLW608243S, BLW608435S, BLW608544S, BLW607367S, BLW607218S, BLW607540S, BLW607698S, BLW608087S, BLW608202S, BLW607551S, BLW607216S, BLW608198S, BLW608240S, BLW607369S, BLW607535S, BLW607534S, BLW607586S, BLW607587S, BLW607697S, BLW608436S, BLW608093S, BLW607213S, BLW607533S, BLW607538S, BLW607690S, BLW607736S, BLW607890S, BLW608192S, BLW608193S, BLW608234S, BLW608239S, BLW608368S, BLW608434S, BLW608548S, BLW607592S, BLW607737S, BLW608290S, BLW608437S, BLW607048S, BLW607052S, BLW607366S, BLW607373S, BLW607376S, BLW609761S, BLW609762S, BLW609792S, BLW609793S, BLW609794S, BLW607544S, BLW608373S, BLW607430S, BLW607444S, BLW607456S, BLW607458S, BLW607543S, BLW607781S, BLW608086S, BLW608090S, BLW608091S, BLW608092S, BLW608094S, BLW608122S, BLW608124S, BLW608362S, BLW608364S, BLW608374S, BLW608425S, BLW608542S, BLW608767S, BLW608770S, BLW608774S, BLW607359S, BLW607545S, BLW608445S, BLW607780S, BLW608085S, BLW608089S, BLW608771S, BLW608084S, BLW608119S, BLW607103S, BLW607241S, BLW608131S, BLW607176S, BLW608323S, BLW608324S, BLW607014S, BLW607110S, BLW607119S, BLW607231S, BLW609119S, BLW609121S, BLW607226S, BLW608363S, BLW607122S, BLW607016S, BLW607108S, BLW607180S, BLW608370S, BLW609116S, BLW607112S, BLW607114S, BLW607109S, BLW607178S, BLW607009S, BLW607111S, BLW607126S, BLW608738S, BLW607011S, BLW607115S, BLW607023S, BLW607024S, BLW607177S, BLW608294S, BLW607128S, BLW607129S, BLW607185S, BLW607270S, BLW607271S, BLW608321S, BLW608361S, BLW608372S, BLW608438S, BLW608439S, BLW608440S, BLW608443S, BLW607019S, BLW608441S, BLW607183S, BLW607117S, BLW607182S, BLW607179S, BLW608369S, BLW607861S, BLW607074S, BLW607377S, BLW607379S, BLW607380S, BLW607382S, BLW607383S, BLW607384S, BLW608740S, BLW607479S, BLW607145S, BLW607558S, BLW607783S, BLW608095S, BLW607615S, BLW607116S, BLW607092S, BLW607272S, BLW607355S, BLW607420S, BLW607554S, BLW607779S, BLW607492S, BLW608776S, BLW607408S, BLW607652S, BLW608286S, BLW608287S, BLW607715S, BLW609767S, BLW607358S, BLW607486S, BLW607488S, BLW607605S, BLW607618S, BLW607862S, BLW609122S, BLW609218S, BLW608769S, BLW607782S, BLW607559S, BLW608088S, BLW608322S, BLW608360S, BLW607146S, BLW607280S, BLW607395S, BLW607546S, BLW608320S, BLW606937S, BLW607060S, BLW608326S, BLW608367S, BLW608325S, BLW608191S, BLW607791S, BLW607848S, BLW608545S, BLW607786S, BLW607787S, BLW607876S, BLW608116S, BLW608117S, BLW608120S, BLW608327S, BLW608328S, BLW608359S, BLW608129S, BLW606988S, BLW608080S, BLW608073S, BLW608125S, BLW608187S, BLW608199S, BLW608200S, BLW608203S, BLW608204S, BLW608076S, BLW608123S, BLW608126S, BLW608127S, BLW607221S, BLW607234S, BLW607247S, BLW607306S, BLW607314S, BLW607323S, BLW607343S, BLW607351S, BLW607352S, BLW607354S, BLW607378S, BLW607767S, BLW607814S, BLW607822S, BLW607826S, BLW607910S, BLW607911S, BLW607912S, BLW607939S, BLW607945S, BLW608540S, BLW608728S, BLW608743S, BLW608888S, BLW608890S, BLW608894S, BLW608895S, BLW608898S, BLW609215S, BLW609219S, BLW607918S, BLW609117S, BLW609127S, BLW609128S, BLW609134S, BLW609212S, BLW609213S, BLW607081S, BLW607083S, BLW607084S, BLW607085S, BLW607086S, BLW607087S, BLW607127S, BLW607130S, BLW607135S, BLW607136S, BLW607137S, BLW607138S, BLW607139S, BLW607140S, BLW607141S, BLW607142S, BLW607181S, BLW607191S, BLW607197S, BLW607198S, BLW607199S, BLW607233S, BLW607240S, BLW607255S, BLW607256S, BLW607258S, BLW607259S, BLW607260S, BLW607261S, BLW607262S, BLW607263S, BLW607264S, BLW607266S, BLW607672S, BLW607673S, BLW607674S, BLW607676S, BLW607677S, BLW607852S, BLW607854S, BLW607855S, BLW607856S, BLW607857S, BLW607867S, BLW607868S, BLW607869S, BLW607870S, BLW607871S, BLW608071S, BLW608072S, BLW609118S, BLW609123S, BLW609132S, BLW609133S, BLW607281S, BLW607564S, BLW608075S, BLW607143S, BLW607454S, BLW607455S, BLW607457S, BLW607459S, BLW607460S, BLW607606S, BLW607608S, BLW607733S, BLW607734S, BLW607735S, BLW607749S, BLW608069S, BLW608070S, BLW608205S, BLW608270S, BLW608274S, BLW608276S, BLW608277S, BLW608278S, BLW608279S, BLW608282S, BLW608283S, BLW608284S, BLW608288S, BLW608289S, BLW608543S, BLW607032S, BLW607557S, BLW608121S, BLW608365S, BLW607222S, BLW607227S, BLW607239S, BLW607816S, BLW607820S, BLW607830S, BLW607832S, BLW607894S, BLW607898S, BLW607913S, BLW607934S, BLW608549S, BLW608727S, BLW608733S, BLW608734S, BLW608736S, BLW608893S, BLW608900S, BLW607536S, BLW608128S, BLW607012S, BLW607607S, BLW608082S, BLW608269S, BLW608186S, BLW608194S, BLW609124S, BLW609125S, BLW609126S, BLW609129S, BLW609130S, BLW609131S, BLW609135S, BLW609206S, BLW609207S, BLW607056S, BLW607095S, BLW607702S, BLW608317S, BLW608375S, BLW608536S, BLW608554S, BLW608768S, BLW607653S, BLW608067S, BLW608068S, BLW608285S, BLW608083S, BLW609120S, BLW609209S, BLW609211S, BLW609216S, BLW609217S, BLW609220S, BLW609221S, BLW609222S, BLW609223S, BLW609224S, BLW609225S, BLW609791S; GTIN 00643169980129, Lot Serial Numbers: BLW610806S, BLW613813S, BLW615082S, BLW615321S, BLW608105S, BLW608252S, BLW608748S, BLW608829S, BLW608867S, BLW609063S, BLW609247S, BLW609264S, BLW609355S, BLW609665S, BLW610234S, BLW610622S, BLW610850S, BLW610904S, BLW612442S, BLW612659S, BLW613241S, BLW613702S, BLW614591S, BLW608994S, BLW610886S, BLW608168S, BLW608170S, BLW608399S, BLW608557S, BLW608560S, BLW608949S, BLW608051S, BLW608298S, BLW608530S, BLW608983S, BLW608563S, BLW608616S, BLW608938S, BLW608527S, BLW609080S, BLW610214S, BLW608846S, BLW607957S, BLW607960S, BLW608174S, BLW608388S, BLW608836S, BLW609718S, BLW608213S, BLW609904S, BLW608265S, BLW608474S, BLW608647S, BLW608153S, BLW608643S, BLW609856S, BLW608486S, BLW608140S, BLW608262S, BLW608479S, BLW609074S, BLW609109S, BLW609933S, BLW610773S, BLW610816S, BLW610912S, BLW612154S, BLW608875S, BLW608876S, BLW609378S, BLW610410S, BLW611296S, BLW611297S, BLW610015S, BLW610627S, BLW610274S, BLW610780S, BLW611024S, BLW611046S, BLW608221S, BLW608314S, BLW608988S, BLW609163S, BLW609177S, BLW609183S, BLW608019S, BLW608059S, BLW608151S, BLW608300S, BLW608446S, BLW608453S, BLW608469S, BLW608498S, BLW608501S, BLW608577S, BLW608579S, BLW608618S, BLW608620S, BLW608658S, BLW608690S, BLW608694S, BLW609040S, BLW609061S, BLW609391S, BLW609497S, BLW609511S, BLW609708S, BLW610041S, BLW610120S, BLW610424S, BLW610615S, BLW610626S, BLW610670S, BLW610834S, BLW610945S, BLW611105S, BLW611291S, BLW611568S, BLW612010S, BLW608691S, BLW610280S, BLW608003S, BLW609281S, BLW609700S, BLW609893S, BLW609914S, BLW610179S, BLW609153S, BLW609241S, BLW609263S, BLW610309S, BLW611298S, BLW611491S, BLW611875S, BLW611958S, BLW610153S, BLW610154S, BLW610156S, BLW610159S, BLW607969S, BLW608058S, BLW608400S, BLW608837S, BLW609093S, BLW610929S, BLW611204S, BLW609838S, BLW610206S, BLW610212S, BLW610900S, BLW607962S, BLW609753S, BLW609833S, BLW610011S, BLW610036S, BLW610071S, BLW610164S, BLW610220S, BLW610624S, BLW610647S, BLW610667S, BLW610986S, BLW611197S, BLW611494S, BLW608004S, BLW608264S, BLW608907S, BLW610228S, BLW608655S, BLW609103S, BLW609341S, BLW610482S, BLW610686S, BLW611315S, BLW610210S, BLW608162S, BLW608457S, BLW608576S, BLW608629S, BLW609559S, BLW610150S, BLW608017S, BLW608570S, BLW609450S, BLW609504S, BLW608663S, BLW609451S, BLW609452S, BLW610216S, BLW608062S, BLW608332S, BLW608333S, BLW608531S, BLW609193S, BLW609238S, BLW609280S, BLW609465S, BLW609687S, BLW609938S, BLW609963S, BLW610135S, BLW610774S, BLW610785S, BLW610841S, BLW609661S, BLW608172S, BLW608688S, BLW608953S, BLW608960S, BLW609068S, BLW609393S, BLW610213S, BLW610285S, BLW610533S, BLW610534S, BLW610568S, BLW610569S, BLW610662S, BLW610663S, BLW610666S, BLW610688S, BLW610689S, BLW610690S, BLW610869S, BLW610902S, BLW611227S, BLW611803S, BLW612143S, BLW609262S, BLW608916S, BLW609630S, BLW611097S, BLW608156S, BLW608211S, BLW608297S, BLW608502S, BLW608521S, BLW608830S, BLW608990S, BLW609490S, BLW609908S, BLW610128S, BLW610299S, BLW613175S, BLW613182S, BLW610177S, BLW610827S, BLW607956S, BLW607958S, BLW607972S, BLW607974S, BLW609801S, BLW611382S, BLW609853S, BLW608166S, BLW609227S, BLW610486S, BLW611591S, BLW608175S, BLW608247S, BLW608338S, BLW608352S, BLW608394S, BLW608448S, BLW608495S, BLW608832S, BLW608852S, BLW608941S, BLW609141S, BLW609195S, BLW609432S, BLW609514S, BLW609852S, BLW610057S, BLW608872S, BLW608978S, BLW609460S, BLW609876S, BLW609936S, BLW610052S, BLW611851S, BLW610010S, BLW611749S, BLW611806S, BLW612047S, BLW609147S, BLW609148S, BLW609149S, BLW609150S, BLW609151S, BLW609363S, BLW609367S, BLW609368S, BLW609369S, BLW609370S, BLW608231S, BLW610919S, BLW608631S, BLW610631S, BLW608301S, BLW608330S, BLW609491S, BLW608857S, BLW608028S, BLW608115S, BLW608309S, BLW608313S, BLW608494S, BLW608827S, BLW608854S, BLW609507S, BLW609740S, BLW611022S, BLW608500S, BLW608507S, BLW611162S, BLW608588S, BLW611069S, BLW608178S, BLW609057S, BLW610625S, BLW608155S, BLW608849S, BLW607955S, BLW607981S, BLW608142S, BLW608306S, BLW608402S, BLW608456S, BLW608489S, BLW608508S, BLW608509S, BLW608564S, BLW608565S, BLW608924S, BLW608998S, BLW609076S, BLW609358S, BLW609509S, BLW609524S, BLW609561S, BLW609563S, BLW609981S, BLW610168S, BLW610236S, BLW610237S, BLW610866S, BLW610913S, BLW610993S, BLW612555S, BLW609010S, BLW609019S, BLW609180S, BLW609197S, BLW609233S, BLW609252S, BLW609339S, BLW609340S, BLW609389S, BLW609474S, BLW609476S, BLW609598S, BLW609870S, BLW609920S, BLW610027S, BLW610127S, BLW610200S, BLW610238S, BLW610239S, BLW610303S, BLW610784S, BLW610860S, BLW610862S, BLW610864S, BLW611124S, BLW611148S, BLW611153S, BLW611168S, BLW611199S, BLW611229S, BLW611349S, BLW611496S, BLW611756S, BLW611772S, BLW611964S, BLW612037S, BLW609152S, BLW609513S, BLW610284S, BLW611320S, BLW611597S, BLW611770S, BLW611776S, BLW608145S, BLW608356S, BLW608518S, BLW608855S, BLW610345S, BLW607964S, BLW609834S, BLW610984S, BLW608259S, BLW608349S, BLW608396S, BLW608409S, BLW608447S, BLW609086S, BLW609088S, BLW609185S, BLW610926S, BLW610980S, BLW611293S, BLW609142S, BLW609448S, BLW609464S, BLW610055S, BLW607993S, BLW609077S, BLW608951S, BLW609362S, BLW610116S, BLW610138S, BLW610865S, BLW609335S, BLW609419S, BLW609191S, BLW611360S, BLW607996S, BLW608138S, BLW608490S, BLW608504S, BLW608982S, BLW609331S, BLW609840S, BLW610132S, BLW610169S, BLW610270S, BLW609742S, BLW608386S, BLW609374S, BLW610623S, BLW611312S, BLW608000S, BLW610139S, BLW608587S, BLW608628S, BLW609079S, BLW609493S, BLW609871S, BLW610167S, BLW610197S, BLW610494S, BLW610049S, BLW610346S, BLW609384S, BLW610434S, BLW611089S, BLW610059S, BLW610131S, BLW608648S, BLW609979S, BLW608561S, BLW608874S, BLW609062S, BLW609110S, BLW609111S, BLW609312S, BLW609449S, BLW610819S, BLW611322S, BLW608014S, BLW608214S, BLW608315S, BLW608664S, BLW608781S, BLW608931S, BLW609361S, BLW609929S, BLW608169S, BLW608493S, BLW609146S, BLW609690S, BLW610368S, BLW607965S, BLW607966S, BLW607967S, BLW608600S, BLW608871S, BLW609438S, BLW611238S, BLW611239S, BLW611361S, BLW611006S, BLW609699S, BLW610060S, BLW610995S, BLW611318S, BLW608940S, BLW611276S, BLW611560S, BLW608989S, BLW608455S, BLW608389S, BLW609155S, BLW609269S, BLW610133S, BLW608002S, BLW610761S, BLW609682S, BLW610211S, BLW610269S, BLW611206S, BLW611419S, BLW608851S, BLW609454S, BLW610054S, BLW610535S, BLW610654S, BLW612540S, BLW608569S, BLW608413S, BLW608482S, BLW608483S, BLW608499S, BLW609877S, BLW609891S, BLW610673S, BLW608249S, BLW608699S, BLW609916S, BLW610048S, BLW610411S, BLW610817S, BLW610918S, BLW607952S, BLW610480S, BLW611171S, BLW611769S, BLW612562S, BLW608165S, BLW608246S, BLW608329S, BLW608391S, BLW609002S, BLW609084S, BLW609087S, BLW609097S, BLW609381S, BLW609382S, BLW609860S, BLW610843S, BLW611562S, BLW611563S, BLW611564S, BLW611565S, BLW611566S, BLW610058S, BLW609337S, BLW608345S, BLW610766S, BLW608562S, BLW608612S, BLW608880S, BLW609333S, BLW609364S, BLW608645S, BLW609372S, BLW609671S, BLW609980S, BLW611552S, BLW611583S, BLW611775S, BLW611998S, BLW608261S, BLW608257S, BLW608215S, BLW608401S, BLW607984S, BLW609098S, BLW609394S, BLW610190S, BLW609431S, BLW611000S, BLW611490S, BLW610042S, BLW610296S, BLW608865S, BLW608701S, BLW609653S, BLW608143S, BLW608842S, BLW609145S, BLW609192S, BLW609505S, BLW609562S, BLW609875S, BLW610685S, BLW610821S, BLW611005S, BLW611170S, BLW611177S, BLW609311S, BLW607961S, BLW607989S, BLW607998S, BLW608001S, BLW608136S, BLW608626S, BLW608650S, BLW609242S, BLW610137S, BLW610263S, BLW611395S, BLW611590S, BLW612067S, BLW612070S, BLW612144S, BLW612386S, BLW613061S, BLW610650S, BLW610905S, BLW611047S, BLW611207S, BLW611346S, BLW611808S, BLW611990S, BLW612062S, BLW613388S, BLW613404S, BLW613820S, BLW614108S, BLW614126S, BLW610191S, BLW610513S, BLW612297S, BLW612569S, BLW609902S, BLW610273S, BLW608260S, BLW609254S, BLW609918S, BLW609924S, BLW611088S, BLW612295S, BLW613407S, BLW607953S, BLW608006S, BLW608668S, BLW608669S, BLW608936S, BLW608958S, BLW609095S, BLW609274S, BLW609276S, BLW609467S, BLW609930S, BLW610053S, BLW610147S, BLW610148S, BLW610207S, BLW610209S, BLW610221S, BLW610244S, BLW610245S, BLW610298S, BLW610510S, BLW610636S, BLW610893S, BLW610895S, BLW611424S, BLW611426S, BLW608673S, BLW608674S, BLW608799S, BLW608804S, BLW608826S, BLW608913S, BLW608992S, BLW609931S, BLW609932S, BLW610879S, BLW611586S, BLW612080S, BLW612051S, BLW613796S, BLW609357S, BLW611587S, BLW610247S, BLW607463S, BLW608336S, BLW609009S, BLW610471S, BLW610764S, BLW609429S, BLW608007S, BLW608258S, BLW608697S, BLW609265S, BLW609745S, BLW609894S, BLW609896S, BLW610607S, BLW611101S, BLW611165S, BLW609187S, BLW610873S, BLW611649S, BLW611650S, BLW611815S, BLW608476S, BLW610531S, BLW611542S, BLW609427S, BLW609848S, BLW608417S, BLW608615S, BLW609858S, BLW610367S, BLW610466S, BLW610859S, BLW610924S, BLW611135S, BLW608011S, BLW608357S, BLW608613S, BLW608839S, BLW608850S, BLW609652S, BLW607992S, BLW609423S, BLW609971S, BLW610355S, BLW609691S, BLW610136S, BLW610931S, BLW611184S, BLW608212S, BLW608331S, BLW608693S, BLW608917S, BLW609181S, BLW610616S, BLW610937S, BLW610970S, BLW611319S, BLW608687S, BLW609468S, BLW609854S, BLW611094S, BLW608148S, BLW608594S, BLW609083S, BLW608164S, BLW608256S, BLW609296S, BLW611493S, BLW607976S, BLW608985S, BLW609332S, BLW609424S, BLW611122S, BLW609064S, BLW609273S, BLW609290S, BLW607954S, BLW608050S, BLW608157S, BLW608226S, BLW608244S, BLW608310S, BLW608341S, BLW608488S, BLW608505S, BLW608649S, BLW608660S, BLW608885S, BLW609037S, BLW609139S, BLW609143S, BLW609179S, BLW609188S, BLW609202S, BLW609420S, BLW609453S, BLW609461S, BLW609560S, BLW609572S, BLW610122S, BLW610427S, BLW610695S, BLW610870S, BLW610871S, BLW610909S, BLW611080S, BLW611140S, BLW611198S, BLW609137S, BLW611220S, BLW608935S, BLW608955S, BLW608956S, BLW608977S, BLW609234S, BLW609246S, BLW609248S, BLW609957S, BLW610019S, BLW610051S, BLW608932S, BLW609244S, BLW609480S, BLW610358S, BLW609972S, BLW611317S, BLW609901S, BLW609416S, BLW611467S, BLW609008S, BLW608492S, BLW609249S, BLW609418S, BLW609496S, BLW608046S, BLW608139S, BLW608220S, BLW608665S, BLW607973S, BLW608397S, BLW608223S, BLW609359S, BLW609421S, BLW609495S, BLW609748S, BLW610684S, BLW610702S, BLW610958S, BLW610988S, BLW611019S, BLW611231S, BLW611240S, BLW611303S, BLW611332S, BLW611374S, BLW608700S, BLW611295S, BLW609313S, BLW610854S, BLW611145S, BLW608465S, BLW608695S, BLW609144S, BLW609937S, BLW610034S, BLW610065S, BLW610125S, BLW610145S, BLW610793S, BLW610830S, BLW608337S, BLW608556S, BLW609058S, BLW609162S, BLW609463S, BLW609895S, BLW610170S, BLW610532S, BLW610867S, BLW610938S, BLW610940S, BLW611232S, BLW610655S, BLW611200S, BLW608311S, BLW609014S, BLW608580S, BLW608918S, BLW607950S, BLW608154S, BLW608997S, BLW608512S, BLW609754S, BLW610275S, BLW608308S, BLW609628S, BLW609673S, BLW609674S, BLW609676S, BLW609799S, BLW609826S, BLW610026S, BLW610189S, BLW610271S, BLW610311S, BLW610339S, BLW610450S, BLW610484S, BLW610485S, BLW610501S, BLW610632S, BLW610677S, BLW610697S, BLW610874S, BLW611007S, BLW611161S, BLW611477S, BLW611538S, BLW611638S, BLW611639S, BLW611886S, BLW611956S, BLW612153S, BLW611215S, BLW608339S, BLW608825S, BLW609256S, BLW610472S, BLW610815S, BLW610903S, BLW610964S, BLW608454S, BLW609250S, BLW609375S, BLW609425S, BLW609689S, BLW610001S, BLW610063S, BLW610066S, BLW608222S, BLW608039S, BLW608406S, BLW609503S, BLW610162S, BLW610878S, BLW611383S, BLW611793S, BLW608606S, BLW610608S, BLW610983S, BLW611250S, BLW611548S, BLW611981S, BLW611996S, BLW608395S, BLW608047S, BLW608491S, BLW609847S, BLW610916S, BLW610481S, BLW608939S, BLW609336S, BLW610467S, BLW610961S, BLW611500S, BLW608173S, BLW608466S, BLW610687S, BLW611422S, BLW611963S, BLW607970S, BLW607983S, BLW607985S, BLW607991S, BLW608407S, BLW608503S, BLW608873S, BLW608909S, BLW608954S, BLW609070S, BLW609138S, BLW609178S, BLW609295S, BLW609666S, BLW609855S, BLW609915S, BLW610447S, BLW610448S, BLW610649S, BLW610820S, BLW611225S, BLW608970S, BLW610840S, BLW611497S, BLW612565S, BLW611026S, BLW608404S, BLW609512S, BLW611093S, BLW611085S, BLW607980S, BLW610050S, BLW607949S, BLW610165S, BLW609182S, BLW607986S, BLW608013S, BLW608403S, BLW608415S, BLW608416S, BLW608656S, BLW608831S, BLW609030S, BLW609071S, BLW609260S, BLW609268S, BLW609270S, BLW609439S, BLW609651S, BLW609969S, BLW610062S, BLW610143S, BLW610279S, BLW610334S, BLW610408S, BLW610409S, BLW610477S, BLW610676S, BLW610778S, BLW611662S, BLW611746S, BLW612000S, BLW612556S, BLW612595S, BLW612958S, BLW613027S, BLW613029S, BLW613158S, BLW608161S, BLW608698S, BLW610037S, BLW610201S, BLW609140S, BLW608863S, BLW608995S, BLW609004S, BLW609157S, BLW609165S, BLW609869S, BLW610240S, BLW611201S, BLW608478S, BLW610056S, BLW610174S, BLW610665S, BLW608335S, BLW608340S, BLW609863S, BLW608263S, BLW610312S, BLW611292S, BLW611316S, BLW611805S, BLW608137S, BLW608843S, BLW608946S, BLW609055S, BLW609392S, BLW609459S, BLW609660S, BLW609667S, BLW609839S, BLW609909S, BLW609964S, BLW610160S, BLW610268S, BLW610503S, BLW610570S, BLW610671S, BLW611216S, BLW611301S, BLW611335S, BLW611339S, BLW611483S, BLW611543S, BLW611544S, BLW608410S, BLW608681S, BLW610304S, BLW610946S, BLW611584S, BLW608347S, BLW610657S, BLW608034S, BLW608411S, BLW608449S, BLW608835S, BLW608847S, BLW609073S, BLW609272S, BLW609360S, BLW610064S, BLW610199S, BLW610219S, BLW610412S, BLW610974S, BLW611187S, BLW611294S, BLW611849S, BLW608141S, BLW609851S, BLW610941S, BLW609258S, BLW611371S, BLW608506S, BLW608472S, BLW608510S, BLW608999S, BLW609007S, BLW609259S, BLW609960S, BLW610121S, BLW610965S, BLW611486S, BLW611502S, BLW611600S, BLW609036S, BLW609925S, BLW610858S, BLW611040S, BLW611269S, BLW611001S, BLW610161S, BLW608355S, BLW608522S, BLW609090S, BLW609913S, BLW610266S, BLW610336S, BLW610499S, BLW610833S, BLW611585S, BLW608307S, BLW610768S, BLW610857S, BLW610672S, BLW609506S, BLW609054S, BLW608475S, BLW610344S, BLW610196S, BLW611210S, BLW611362S, BLW611370S, BLW611620S, BLW612152S, BLW612381S, BLW613194S, BLW608883S, BLW609113S, BLW609231S, BLW609356S, BLW611222S, BLW611384S, BLW608005S, BLW608010S, BLW608147S, BLW608414S, BLW608458S, BLW608644S, BLW608861S, BLW608933S, BLW608973S, BLW608974S, BLW609006S, BLW609085S, BLW609092S, BLW609743S, BLW609864S, BLW610114S, BLW610195S, BLW610509S, BLW610910S, BLW611159S, BLW611241S, BLW608459S, BLW608566S, BLW608582S, BLW608884S, BLW608886S, BLW609069S, BLW609201S, BLW609494S, BLW609502S, BLW609557S, BLW609892S, BLW610243S, BLW610248S, BLW610249S, BLW610648S, BLW610802S, BLW610915S, BLW610971S, BLW611152S, BLW611271S, BLW611340S, BLW608572S, BLW609385S, BLW608040S, BLW608227S, BLW608353S, BLW608805S, BLW608957S, BLW608986S, BLW608991S, BLW609314S, BLW609387S, BLW609433S, BLW609477S, BLW609697S, BLW609836S, BLW610451S, BLW610479S, BLW610567S, BLW610967S, BLW611123S, BLW611208S, BLW610044S, BLW610887S, BLW607982S, BLW608571S, BLW608915S, BLW608972S, BLW610242S, BLW610618S, BLW608245S, BLW610612S, BLW608218S, BLW609091S, BLW609373S, BLW609500S, BLW610067S, BLW608573S, BLW610944S, BLW608228S, BLW608229S, BLW608230S, BLW608578S, BLW608678S, BLW608840S, BLW608844S, BLW608966S, BLW609105S, BLW609156S, BLW609279S, BLW609417S, BLW609492S, BLW609696S, BLW610012S, BLW610038S, BLW610115S, BLW610172S, BLW610621S, BLW608217S, BLW608343S, BLW608412S, BLW608464S, BLW608485S, BLW608670S, BLW608680S, BLW608760S, BLW608761S, BLW608763S, BLW608764S, BLW608765S, BLW608766S, BLW608798S, BLW608800S, BLW608801S, BLW608802S, BLW608803S, BLW608806S, BLW608807S, BLW608808S, BLW608809S, BLW608813S, BLW608816S, BLW608817S, BLW608818S, BLW608819S, BLW608820S, BLW608822S, BLW608824S, BLW612133S, BLW609065S, BLW609844S, BLW610614S, BLW611130S, BLW610129S, BLW610925S, BLW612098S, BLW612940S, BLW612941S, BLW612942S, BLW612943S, BLW612944S, BLW612948S, BLW612949S, BLW608167S, BLW608392S, BLW609897S, BLW610235S, BLW608219S, BLW608462S, BLW608853S, BLW609501S, BLW610024S, BLW610045S, BLW611226S, BLW612543S, BLW612563S, BLW608210S, BLW609186S, BLW609903S, BLW610406S, BLW607975S, BLW610433S, BLW609747S, BLW608393S, BLW609525S, BLW608348S, BLW608979S, BLW609056S, BLW609094S, BLW609261S, BLW609294S, BLW609704S, BLW610994S, BLW608012S, BLW610502S, BLW608568S, BLW608591S, BLW608652S, BLW609243S, BLW610126S, BLW610934S, BLW608514S, BLW608567S, BLW609108S, BLW609236S, BLW609510S, BLW609521S, BLW609849S, BLW609850S, BLW609880S, BLW609919S, BLW609927S, BLW609928S, BLW609939S, BLW610013S, BLW610014S, BLW610039S, BLW610069S, BLW610070S, BLW610123S, BLW610175S, BLW610176S, BLW610187S, BLW610188S, BLW610205S, BLW610208S, BLW610278S, BLW610340S, BLW610341S, BLW610362S, BLW610443S, BLW610476S, BLW610675S, BLW610679S, BLW610776S, BLW610805S, BLW610809S, BLW610837S, BLW610846S, BLW610932S, BLW610933S, BLW610968S, BLW610982S, BLW611027S, BLW611071S, BLW611079S, BLW611141S, BLW611188S, BLW611191S, BLW611219S, BLW611311S, BLW611341S, BLW611376S, BLW611378S, BLW611385S, BLW611420S, BLW611492S, BLW611540S, BLW611727S, BLW611771S, BLW611807S, BLW611991S, BLW612041S, BLW612044S, BLW612296S, BLW612384S, BLW612409S, BLW612545S, BLW612573S, BLW612608S, BLW612763S, BLW612829S, BLW612831S, BLW612834S, BLW612955S, BLW613060S, BLW613145S, BLW613239S, BLW613822S, BLW613935S, BLW613936S, BLW614019S, BLW614441S, BLW614684S, BLW614686S, BLW614687S, BLW609481S, BLW611480S, BLW607948S, BLW608418S, BLW610166S, BLW610241S, BLW610250S, BLW610693S, BLW611304S, BLW611414S, BLW608841S, BLW609230S, BLW611364S, BLW611365S, BLW611366S, BLW611368S, BLW611369S, BLW610699S, BLW610779S, BLW611546S, BLW611659S, BLW609528S, BLW610198S, BLW610682S, BLW610880S, BLW610185S, BLW611637S, BLW609059S, BLW609436S, BLW607977S, BLW607978S, BLW608008S, BLW608015S, BLW608144S, BLW608208S, BLW608225S, BLW608299S, BLW608354S, BLW608379S, BLW608385S, BLW608480S, BLW608496S, BLW608559S, BLW608607S, BLW608610S, BLW608627S, BLW608651S, BLW608667S, BLW608677S, BLW608692S, BLW608868S, BLW608928S, BLW608943S, BLW608944S, BLW608947S, BLW608948S, BLW608959S, BLW609164S, BLW609189S, BLW609255S, BLW609267S, BLW609297S, BLW609298S, BLW609328S, BLW609488S, BLW609520S, BLW609558S, BLW609564S, BLW609698S, BLW609703S, BLW609719S, BLW609841S, BLW609846S, BLW609923S, BLW610073S, BLW610134S, BLW610171S, BLW610217S, BLW610423S, BLW610425S, BLW610628S, BLW610668S, BLW610674S, BLW610694S, BLW610795S, BLW610882S, BLW610972S, BLW610973S, BLW611133S, BLW611189S, BLW611213S, BLW611555S, BLW611642S, BLW611695S, BLW611751S, BLW611765S, BLW611773S, BLW611993S, BLW612066S, BLW608342S, BLW608532S, BLW608787S, BLW609016S, BLW609100S, BLW609237S, BLW609414S, BLW609695S, BLW610763S, BLW610808S, BLW609075S, BLW609746S, BLW611211S, BLW608250S, BLW608251S, BLW608253S, BLW608254S, BLW608255S, BLW608702S, BLW608703S, BLW608704S, BLW608705S, BLW608706S, BLW608756S, BLW608757S, BLW608758S, BLW608759S, BLW608762S, BLW609041S, BLW609042S, BLW609043S, BLW609045S, BLW609046S, BLW609047S, BLW609048S, BLW609049S, BLW609050S, BLW609051S, BLW609052S, BLW609308S, BLW609309S, BLW609326S, BLW609327S, BLW609565S, BLW610638S, BLW611144S, BLW611244S, BLW611245S, BLW611246S, BLW611379S, BLW609067S, BLW609081S, BLW609406S, BLW609407S, BLW609408S, BLW609409S, BLW609410S, BLW609411S, BLW609412S, BLW609413S, BLW610456S, BLW610457S, BLW610458S, BLW610459S, BLW610461S, BLW610462S, BLW610463S, BLW610464S, BLW610465S, BLW610888S, BLW610889S, BLW611142S, BLW611158S, BLW611173S, BLW611182S, BLW611185S, BLW607461S, BLW611179S, BLW610949S, BLW608160S, BLW608515S, BLW610952S, BLW609857S, BLW610619S, BLW610875S, BLW608390S, BLW608619S, BLW608662S, BLW609099S, BLW609136S, BLW609865S, BLW609198S, BLW611999S, BLW608045S, BLW608609S, BLW610812S, BLW609996S, BLW607959S, BLW607988S, BLW611488S, BLW608063S, BLW608463S, BLW608661S, BLW608833S, BLW609204S, BLW609862S, BLW609907S, BLW610701S, BLW612599S, BLW609245S, BLW609983S, BLW610620S, BLW610899S, BLW608021S, BLW609900S, BLW611557S, BLW611622S, BLW608696S, BLW609683S, BLW608926S, BLW609961S, BLW608296S, BLW608450S, BLW608623S, BLW608624S, BLW608942S, BLW609038S, BLW609158S, BLW609194S, BLW609570S, BLW609802S, BLW609813S, BLW610223S, BLW610252S, BLW610360S, BLW611083S, BLW611131S, BLW611205S, BLW611334S, BLW610917S, BLW607947S, BLW607968S, BLW609102S, BLW609447S, BLW611212S, BLW608659S, BLW610942S, BLW611067S, BLW611499S, BLW611697S, BLW608149S, BLW610496S, BLW611214S, BLW611333S, BLW608617S, BLW608030S, BLW608621S, BLW608838S, BLW609741S, BLW611228S, BLW608877S, BLW609705S, BLW610276S, BLW610852S, BLW611086S, BLW608112S, BLW608812S, BLW608821S, BLW609878S, BLW610661S, BLW610705S, BLW611224S, BLW608233S, BLW608646S, BLW608968S, BLW609455S, BLW609926S, BLW609956S, BLW610301S, BLW611209S, BLW611249S, BLW611464S, BLW611468S, BLW611503S, BLW609861S, BLW609899S, BLW610803S, BLW610826S, BLW607971S, BLW608689S, BLW610267S, BLW610617S, BLW611504S, BLW608344S, BLW608657S, BLW610194S, BLW610307S, BLW608642S, BLW608111S, BLW608452S, BLW608470S, BLW608590S, BLW608848S, BLW609017S, BLW609527S, BLW609677S, BLW609707S, BLW609842S, BLW609845S, BLW610046S, BLW610047S, BLW610202S, BLW610277S, BLW610629S, BLW610664S, BLW610928S, BLW611021S, BLW611375S, BLW608146S, BLW608451S, BLW608460S, BLW608484S, BLW609266S, BLW609721S, BLW609967S, BLW609751S, BLW610256S, BLW610960S, BLW611462S, BLW607995S, BLW608031S, BLW608224S, BLW608312S, BLW608882S, BLW609013S, BLW609440S, BLW609516S, BLW610140S, BLW610186S, BLW610652S, BLW610849S, BLW611125S, BLW608305S, BLW609866S, BLW609868S, BLW610703S, BLW610855S, BLW610999S, BLW611136S, BLW608585S, BLW609672S, BLW609921S, BLW610969S, BLW608024S, BLW608810S, BLW607951S, BLW609670S, BLW610343S, BLW609843S, BLW610976S, BLW611034S, BLW608654S, BLW608209S, BLW608589S, BLW608653S, BLW609160S, BLW609377S, BLW609498S, BLW609596S, BLW609658S, BLW609679S, BLW609692S, BLW610613S, BLW611774S, BLW608302S, BLW609161S, BLW610310S, BLW610366S, BLW608608S, BLW608945S, BLW610204S, BLW610842S, BLW611128S, BLW612561S, BLW608858S, BLW608061S, BLW609750S, BLW609966S, BLW610473S, BLW611116S, BLW608158S, BLW608171S, BLW608096S, BLW608468S, BLW608473S, BLW608666S, BLW609379S, BLW610478S, BLW607963S, BLW610669S, BLW610040S, BLW610281S, BLW610975S, BLW609235S, BLW608097S, BLW608101S, BLW608103S, BLW608104S, BLW608106S, BLW608107S, BLW608420S, BLW608421S, BLW608422S, BLW608423S, BLW608424S, BLW608707S, BLW608708S, BLW608709S, BLW608710S, BLW608711S, BLW608717S, BLW608718S, BLW608719S, BLW608720S, BLW608721S, BLW609166S, BLW609167S, BLW609168S, BLW609169S, BLW609170S, BLW609171S, BLW609172S, BLW609173S, BLW609174S, BLW609175S, BLW609532S, BLW609534S, BLW609536S, BLW609537S, BLW609538S, BLW609539S, BLW609540S, BLW609541S, BLW609542S, BLW609543S, BLW609984S, BLW609986S, BLW609987S, BLW610030S, BLW610031S, BLW610033S, BLW610288S, BLW610289S, BLW610290S, BLW610291S, BLW610292S, BLW610293S, BLW610294S, BLW610302S, BLW610305S, BLW610306S, BLW608054S, BLW608055S, BLW608056S, BLW612136S, BLW612137S, BLW612138S, BLW612140S, BLW612141S, BLW612142S, BLW608304S, BLW608350S, BLW608351S, BLW608517S, BLW608790S, BLW608793S, BLW609292S, BLW609305S, BLW609323S, BLW609386S, BLW609867S, BLW609973S, BLW610731S, BLW610863S, BLW610979S, BLW611070S, BLW611195S, BLW611425S, BLW611458S, BLW611783S, BLW609322S, BLW609324S, BLW609349S, BLW609352S, BLW609395S, BLW610683S, BLW611479S, BLW611501S, BLW611588S, BLW611616S, BLW609301S, BLW609310S, BLW609329S, BLW609365S, BLW609947S, BLW609977S, BLW610698S, BLW610725S, BLW610733S, BLW610839S, BLW610845S, BLW611032S, BLW608574S, BLW608387S, BLW613013S, BLW613699S, BLW614628S, BLW615278S, BLW615280S, BLW611615S, BLW607997S, BLW608622S, BLW609066S, BLW609962S, BLW610035S, BLW610985S, BLW610130S, BLW610577S, BLW608525S, BLW608526S, BLW608529S, BLW608419S, BLW608523S, BLW608524S, BLW608528S, BLW610192S, BLW608152S, BLW608177S, BLW608676S, BLW608779S, BLW608921S, BLW609712S, BLW611959S, BLW610254S, BLW610255S, BLW610573S, BLW609443S, BLW609487S, BLW609535S, BLW609567S, BLW609955S, BLW609959S, BLW610522S, BLW610575S, BLW610576S, BLW611541S, BLW611553S, BLW611554S, BLW611570S, BLW611876S, BLW611877S, BLW611881S, BLW611882S, BLW611900S, BLW611901S, BLW611902S, BLW611908S, BLW611909S, BLW611968S, BLW611971S, BLW611973S, BLW611980S, BLW612050S, BLW612059S, BLW612091S, BLW612108S, BLW612110S, BLW612125S, BLW612566S, BLW612577S, BLW612578S, BLW612597S, BLW612598S, BLW612600S, BLW612607S, BLW612619S, BLW612621S, BLW612623S, BLW612624S, BLW612625S, BLW612626S, BLW612634S, BLW612639S, BLW612653S, BLW612655S, BLW612656S, BLW612675S, BLW612682S, BLW612683S, BLW612684S, BLW612702S, BLW612737S, BLW612738S, BLW612739S, BLW612751S, BLW612772S, BLW612773S, BLW612786S, BLW612798S, BLW612826S, BLW612833S, BLW612895S, BLW612912S, BLW612921S, BLW612929S, BLW612931S, BLW612932S, BLW612937S, BLW612984S, BLW612989S, BLW612999S, BLW613045S, BLW613149S, BLW613155S, BLW613171S, BLW613172S, BLW613176S, BLW613186S, BLW613187S, BLW613193S, BLW613197S, BLW613201S, BLW613252S, BLW613253S, BLW613255S, BLW613256S, BLW613257S, BLW613258S, BLW613289S, BLW613445S, BLW613460S, BLW613464S, BLW613566S, BLW613571S, BLW613596S, BLW613744S, BLW613745S, BLW613746S, BLW613747S, BLW613757S, BLW613782S, BLW613804S, BLW613808S, BLW613809S, BLW613862S, BLW613863S, BLW613927S, BLW614012S, BLW614042S, BLW614065S, BLW614438S, BLW614478S, BLW614479S, BLW614480S, BLW614509S, BLW614510S, BLW614592S, BLW614613S, BLW614614S, BLW614615S, BLW614720S, BLW614721S, BLW614752S, BLW614753S, BLW615302S, BLW615308S, BLW615311S, BLW615312S, BLW615314S, BLW615316S, BLW615317S, BLW615319S, BLW608604S, BLW609199S, BLW609444S, BLW608856S, BLW608967S, BLW608980S, BLW609000S, BLW609662S, BLW609663S, BLW609668S, BLW609669S, BLW609686S, BLW609701S, BLW609702S, BLW609803S, BLW609989S, BLW609997S, BLW610265S, BLW610460S, BLW609484S, BLW609577S, BLW609578S, BLW609579S, BLW609735S, BLW609744S, BLW610163S, BLW610178S, BLW610258S, BLW610646S, BLW612036S, BLW613291S, BLW614461S, BLW615371S, BLW612035S, BLW610523S, BLW610524S, BLW611804S, BLW608975S, BLW608981S, BLW611892S, BLW611893S, BLW611894S, BLW612810S, BLW612938S, BLW612982S, BLW613565S, BLW613647S, BLW613667S, BLW613672S, BLW613706S, BLW613708S, BLW613737S, BLW613759S, BLW613769S, BLW613770S, BLW613773S, BLW613866S, BLW613868S, BLW613869S, BLW614275S, BLW614276S, BLW614281S, BLW614394S, BLW614409S, BLW614410S, BLW614412S, BLW614433S, BLW614436S, BLW614437S, BLW614475S, BLW614481S, BLW614507S, BLW614508S, BLW614549S, BLW614599S, BLW614702S, BLW614723S, BLW614724S, BLW614751S, BLW614960S, BLW615306S, BLW615307S, BLW615309S, BLW615310S, BLW615315S, BLW608952S, BLW609053S, BLW609435S, BLW609482S, BLW609483S, BLW609526S, BLW609614S, BLW609680S, BLW609715S, BLW609716S, BLW610180S, BLW610257S, BLW610449S, BLW610452S, BLW610455S, BLW610488S, BLW610489S, BLW610574S, BLW610584S, BLW611489S, BLW611811S, BLW611814S, BLW611816S, BLW611839S, BLW611845S, BLW611880S, BLW611899S, BLW611907S, BLW611984S, BLW612014S, BLW612100S, BLW612290S, BLW612385S, BLW612438S, BLW612571S, BLW612606S, BLW612651S, BLW612652S, BLW612657S, BLW612658S, BLW612718S, BLW612765S, BLW612788S, BLW612966S, BLW612967S, BLW612969S, BLW612976S, BLW613030S, BLW613031S, BLW613085S, BLW613139S, BLW613142S, BLW613243S, BLW613292S, BLW613298S, BLW613714S, BLW613765S, BLW614307S, BLW614309S, BLW610729S, BLW610734S, BLW612286S, BLW612587S, BLW613092S, BLW613636S, BLW613697S, BLW613723S, BLW613743S, BLW613749S, BLW613750S, BLW613783S, BLW614467S, BLW609684S, BLW612131S, BLW613000S, BLW613165S, BLW609568S, BLW613088S, BLW612591S, BLW612794S, BLW613724S, BLW613802S, BLW614408S, BLW614434S, BLW614439S, BLW614468S, BLW613297S, BLW613789S, BLW610264S, BLW612579S, BLW612584S, BLW613162S, BLW608747S, BLW608823S, BLW608930S, BLW610184S, BLW610261S, BLW609911S, BLW609107S, BLW610193S, BLW609722S, BLW609797S, BLW609798S, BLW609800S, BLW609806S, BLW609807S, BLW609808S, BLW609809S, BLW609810S, BLW609811S, BLW609812S, BLW609814S, BLW609815S, BLW609817S, BLW609819S, BLW609821S, BLW609823S, BLW610079S, BLW610080S, BLW610081S, BLW610082S, BLW610083S, BLW610084S, BLW610085S, BLW610088S, BLW610089S, BLW610090S, BLW610091S, BLW610092S, BLW610093S, BLW610096S, BLW610108S, BLW610113S, BLW610352S, BLW610353S, BLW610354S, BLW610440S, BLW610441S, BLW610445S, BLW610790S, BLW610792S, BLW610798S, BLW610799S, BLW610811S, BLW610813S, BLW610814S, BLW610828S, BLW610885S, BLW609710S, BLW609257S, BLW609383S, BLW609529S, BLW608879S, BLW614754S, BLW608593S, BLW608625S, BLW609106S, BLW609253S, BLW609456S, BLW609519S, BLW609530S, BLW609556S, BLW609681S, BLW609709S, BLW609470S, BLW609906S, BLW609471S, BLW609472S, BLW609278S, BLW610470S, BLW610474S, BLW610519S, BLW610520S, BLW611780S, BLW612269S, BLW612276S, BLW612285S, BLW612821S, BLW613033S, BLW613043S, BLW613050S, BLW613251S, BLW613319S, BLW613321S, BLW613416S, BLW613441S, BLW613867S, BLW611762S, BLW612103S, BLW612115S, BLW612124S, BLW612130S, BLW612610S, BLW612647S, BLW612933S, BLW609576S, BLW612925S, BLW609828S, BLW610487S, BLW610740S, BLW610743S, BLW611826S, BLW612061S, BLW612104S, BLW612564S, BLW612710S, BLW612776S, BLW612792S, BLW612964S, BLW613028S, BLW613658S, BLW613663S, BLW612382S, BLW612387S, BLW612743S, BLW613318S, BLW611347S, BLW608870S, BLW608961S, BLW608962S, BLW608963S, BLW608964S, BLW608965S, BLW609229S, BLW609232S, BLW609390S, BLW609397S, BLW609398S, BLW609400S, BLW609401S, BLW609402S, BLW609403S, BLW609415S, BLW609422S, BLW609426S, BLW609428S, BLW609430S, BLW609437S, BLW609441S, BLW609522S, BLW609523S, BLW609664S, BLW609713S, BLW609714S, BLW609872S, BLW609873S, BLW609874S, BLW612750S, BLW613190S, BLW613191S, BLW613196S, BLW613200S, BLW613202S, BLW613203S, BLW613299S, BLW613302S, BLW613304S, BLW613308S, BLW613312S, BLW613314S, BLW613326S, BLW613334S, BLW613338S, BLW613381S, BLW613391S, BLW613392S, BLW613397S, BLW613422S, BLW613428S, BLW613434S, BLW613504S, BLW613508S, BLW613524S, BLW613525S, BLW613528S, BLW613538S, BLW613539S, BLW613236S, BLW609485S, BLW609566S, BLW609879S, BLW610251S, BLW610335S, BLW610692S, BLW610997S, BLW610181S, BLW610497S, BLW610579S, BLW613052S, BLW611791S, BLW612567S, BLW612664S, BLW610765S, BLW610517S, BLW611975S, BLW612117S, BLW613758S, BLW612516S, BLW613207S, BLW609442S, BLW611273S, BLW611274S, BLW611275S, BLW611278S, BLW611279S, BLW611280S, BLW611281S, BLW611282S, BLW611829S, BLW611832S, BLW611833S, BLW611843S, BLW611967S, BLW611969S, BLW611974S, BLW611977S, BLW611979S, BLW612005S, BLW612006S, BLW612008S, BLW612009S, BLW612660S, BLW612805S, BLW608878S, BLW609434S, BLW609466S, BLW610072S, BLW610634S, BLW610892S, BLW610896S, BLW610996S, BLW611985S, BLW611648S, BLW614431S, BLW607979S, BLW609196S, BLW612583S, BLW609096S, BLW612469S, BLW612472S, BLW613623S, BLW613624S, BLW613690S, BLW613691S, BLW615833S, BLW615837S, BLW608937S, BLW609200S, BLW611461S, BLW611466S, BLW611572S, BLW611573S, BLW611574S, BLW612306S, BLW612309S, BLW612314S, BLW612338S, BLW612339S, BLW612342S, BLW612344S, BLW612389S, BLW612392S, BLW612393S, BLW612394S, BLW612395S, BLW612517S, BLW612576S, BLW614330S, BLW614331S, BLW614482S, BLW614483S, BLW614484S, BLW614485S, BLW608037S, BLW608303S, BLW608346S, BLW608477S, BLW608481S, BLW608516S, BLW608584S, BLW608611S, BLW608630S, BLW608811S, BLW608996S, BLW609388S, BLW609837S, BLW609898S, BLW610144S, BLW610232S, BLW610297S, BLW613333S, BLW614924S, BLW609827S, BLW610508S, BLW610529S, BLW610507S, BLW608914S, BLW612574S, BLW613814S, BLW608984S, BLW609012S, BLW609072S, BLW609518S, BLW611735S, BLW611736S, BLW611740S, BLW611741S, BLW611743S, BLW614555S, BLW614558S, BLW614767S, BLW614773S, BLW614774S, BLW614778S, BLW614779S, BLW612914S, BLW613328S, BLW610959S, BLW610966S, BLW611108S, BLW611175S, BLW611194S, BLW611344S, BLW611441S, BLW611465S, BLW611482S, BLW611484S, BLW611831S, BLW611838S, BLW611857S, BLW611863S, BLW612767S, BLW613153S, BLW613169S, BLW613344S, BLW613498S, BLW613545S, BLW613642S, BLW613659S, BLW613680S, BLW613681S, BLW613682S, BLW613828S, BLW614244S, BLW614353S, BLW614605S, BLW614606S, BLW614745S, BLW614810S, BLW615619S, BLW615621S, BLW615854S, BLW611729S, BLW611730S, BLW611745S, BLW609675S, BLW610506S, BLW611739S, BLW611800S, BLW611903S, BLW611904S, BLW612132S, BLW612596S, BLW612611S, BLW612891S, BLW612972S, BLW613038S, BLW613206S, BLW613792S, BLW614101S, BLW610468S, BLW610521S, BLW610578S, BLW610797S, BLW610810S, BLW611430S, BLW611487S, BLW611767S, BLW611777S, BLW611888S, BLW611890S, BLW611911S, BLW611914S, BLW611915S, BLW611965S, BLW611966S, BLW612116S, BLW612126S, BLW612129S, BLW612134S, BLW612415S, BLW612426S, BLW612602S, BLW612603S, BLW612681S, BLW612686S, BLW612688S, BLW612690S, BLW612695S, BLW612696S, BLW612704S, BLW612706S, BLW612745S, BLW612748S, BLW613208S, BLW613218S, BLW613224S, BLW613225S, BLW613226S, BLW613228S, BLW613231S, BLW613242S, BLW613245S, BLW613246S, BLW613249S, BLW613250S, BLW613254S, BLW613261S, BLW613264S, BLW613265S, BLW613283S, BLW613475S, BLW613497S, BLW613536S, BLW613721S, BLW613726S, BLW613727S, BLW613729S, BLW613732S, BLW613734S, BLW613740S, BLW613748S, BLW613756S, BLW613761S, BLW613766S, BLW613768S, BLW613784S, BLW613807S, BLW613810S, BLW613864S, BLW613865S, BLW614401S, BLW614402S, BLW614407S, BLW614550S, BLW614551S, BLW614552S, BLW614688S, BLW614689S, BLW614690S, BLW614693S, BLW614707S, BLW614708S, BLW614709S, BLW614711S, BLW614713S, BLW614714S, BLW614726S, BLW614727S, BLW614732S, BLW614742S, BLW609685S, BLW610581S, BLW610582S, BLW610596S, BLW610739S, BLW610794S, BLW611822S, BLW612289S, BLW612604S, BLW612605S, BLW613742S, BLW613762S, BLW613764S, BLW614456S, BLW614458S, BLW614459S, BLW614462S, BLW614463S, BLW614548S, BLW614621S, BLW614622S, BLW614763S, BLW614765S, BLW614766S, BLW611218S, BLW608114S, BLW608163S, BLW608232S, BLW608461S, BLW608471S, BLW608519S, BLW608520S, BLW608788S, BLW608789S, BLW608791S, BLW608794S, BLW608795S, BLW608016S, BLW608018S, BLW608020S, BLW608022S, BLW608023S, BLW608025S, BLW608026S, BLW608027S, BLW608029S, BLW608032S, BLW608033S, BLW608035S, BLW608036S, BLW608038S, BLW608041S, BLW608042S, BLW608043S, BLW608044S, BLW608048S, BLW608049S, BLW608057S, BLW608098S, BLW608099S, BLW608109S, BLW608113S, BLW609499S, BLW609508S, BLW609717S, BLW609720S, BLW609752S, BLW610222S, BLW610475S, BLW610556S, BLW608511S, BLW609457S, BLW608777S, BLW608780S, BLW608782S, BLW608784S, BLW608785S, BLW608786S, BLW611221S, BLW611233S, BLW611575S, BLW611578S, BLW611820S, BLW612580S, BLW612582S, BLW612588S, BLW612590S, BLW612592S, BLW612612S, BLW612614S, BLW612615S, BLW612618S, BLW612642S, BLW612643S, BLW612644S, BLW612649S, BLW612650S, BLW612654S, BLW614486S, BLW614487S, BLW614489S, BLW614490S, BLW614491S, BLW614492S, BLW614493S, BLW614494S, BLW614495S, BLW614496S, BLW614497S, BLW614498S, BLW614577S, BLW612917S, BLW609203S, BLW609321S, BLW609338S, BLW609376S, BLW609404S, BLW609970S, BLW609975S, BLW609978S, BLW609982S, BLW610735S, BLW610757S, BLW610848S, BLW610877S, BLW610891S, BLW610922S, BLW611726S, BLW611790S, BLW611987S, BLW609820S, BLW609829S, BLW610074S, BLW610106S, BLW610107S, BLW610112S, BLW610418S, BLW610742S, BLW610744S, BLW610771S, BLW610796S, BLW610800S, BLW610829S, BLW610884S, BLW608060S, BLW608064S, BLW608065S, BLW608179S, BLW608180S, BLW608181S, BLW608182S, BLW608183S, BLW608184S, BLW608185S, BLW608206S, BLW608207S, BLW608216S, BLW608358S, BLW608371S, BLW608376S, BLW608377S, BLW608378S, BLW608380S, BLW608381S, BLW608382S, BLW608383S, BLW608384S, BLW608595S, BLW608596S, BLW608597S, BLW608598S, BLW608599S, BLW608712S, BLW608713S, BLW608714S, BLW608715S, BLW608716S, BLW608722S, BLW608723S, BLW608724S, BLW608725S, BLW608726S, BLW608749S, BLW608750S, BLW608751S, BLW608752S, BLW608753S, BLW608754S, BLW608755S, BLW609020S, BLW609021S, BLW609022S, BLW609023S, BLW609024S, BLW609025S, BLW609026S, BLW609027S, BLW609028S, BLW609029S, BLW609031S, BLW609032S, BLW609033S, BLW609034S, BLW609035S, BLW609114S, BLW609115S, BLW609282S, BLW609283S, BLW609284S, BLW609285S, BLW609286S, BLW609287S, BLW609288S, BLW609289S, BLW609291S, BLW609293S, BLW609315S, BLW609316S, BLW609318S, BLW609320S, BLW609342S, BLW609343S, BLW609344S, BLW609345S, BLW609346S, BLW609347S, BLW609348S, BLW609350S, BLW609351S, BLW609353S, BLW609354S, BLW609531S, BLW609544S, BLW609545S, BLW609546S, BLW609547S, BLW609548S, BLW609550S, BLW609551S, BLW609552S, BLW609553S, BLW609554S, BLW609555S, BLW609573S, BLW609574S, BLW609575S, BLW609881S, BLW609882S, BLW609883S, BLW609884S, BLW609885S, BLW609886S, BLW609887S, BLW609888S, BLW609889S, BLW609890S, BLW609940S, BLW609941S, BLW609943S, BLW609944S, BLW609945S, BLW609946S, BLW609948S, BLW609950S, BLW609952S, BLW609953S, BLW610286S, BLW610287S, BLW610314S, BLW610315S, BLW610316S, BLW610317S, BLW610318S, BLW610319S, BLW610320S, BLW610321S, BLW610322S, BLW610323S, BLW610324S, BLW610325S, BLW610326S, BLW610328S, BLW610329S, BLW610330S, BLW610331S, BLW610332S, BLW610333S, BLW610640S, BLW610642S, BLW610643S, BLW610644S, BLW610645S, BLW611342S, BLW611343S, BLW611345S, BLW611389S, BLW611390S, BLW611391S, BLW611392S, BLW611393S, BLW611396S, BLW611397S, BLW611428S, BLW611437S, BLW611856S, BLW611858S, BLW611859S, BLW611860S, BLW611861S, BLW611862S, BLW611878S, BLW611986S, BLW612056S, BLW612057S, BLW612058S, BLW612063S, BLW612064S, BLW612223S, BLW612258S, BLW612261S, BLW611757S, BLW612214S, BLW612218S, BLW612278S, BLW612280S, BLW609366S, BLW609486S, BLW608614S, BLW609935S, BLW609804S, BLW609816S, BLW609942S, BLW609949S, BLW610094S, BLW610103S, BLW610104S, BLW610105S, BLW610831S, BLW610832S, BLW610981S, BLW610989S, BLW610990S, BLW610991S, BLW610998S, BLW611002S, BLW611003S, BLW611031S, BLW611077S, BLW611084S, BLW611087S, BLW611090S, BLW611092S, BLW611127S, BLW611129S, BLW611146S, BLW611147S, BLW611190S, BLW611193S, BLW611259S, BLW611260S, BLW611263S, BLW611270S, BLW611272S, BLW611277S, BLW611286S, BLW611299S, BLW611302S, BLW611305S, BLW611398S, BLW611399S, BLW611402S, BLW611408S, BLW611409S, BLW611416S, BLW611418S, BLW611438S, BLW611550S, BLW611556S, BLW611559S, BLW611580S, BLW611581S, BLW611627S, BLW611628S, BLW611629S, BLW611634S, BLW611635S, BLW611693S, BLW611696S, BLW611698S, BLW611700S, BLW611701S, BLW611702S, BLW611784S, BLW611835S, BLW611836S, BLW611837S, BLW611840S, BLW611841S, BLW611895S, BLW611896S, BLW611898S, BLW611910S, BLW611912S, BLW611913S, BLW611955S, BLW611960S, BLW611978S, BLW612060S, BLW612083S, BLW612146S, BLW612149S, BLW612151S, BLW612155S, BLW612156S, BLW612157S, BLW612167S, BLW612170S, BLW612191S, BLW612192S, BLW612193S, BLW612246S, BLW612254S, BLW612256S, BLW612263S, BLW612266S, BLW612277S, BLW612279S, BLW612302S, BLW612445S, BLW612446S, BLW612460S, BLW612485S, BLW612486S, BLW612496S, BLW612500S, BLW612539S, BLW612541S, BLW612547S, BLW612548S, BLW612551S, BLW612552S, BLW612553S, BLW612570S, BLW612667S, BLW612668S, BLW612669S, BLW612670S, BLW612671S, BLW612672S, BLW612717S, BLW612721S, BLW612722S, BLW612723S, BLW612724S, BLW612725S, BLW612770S, BLW612771S, BLW612774S, BLW612777S, BLW612780S, BLW612782S, BLW612804S, BLW612806S, BLW612808S, BLW612811S, BLW612812S, BLW612815S, BLW612816S, BLW612818S, BLW612877S, BLW612953S, BLW612988S, BLW612991S, BLW612992S, BLW612994S, BLW613035S, BLW613042S, BLW613049S, BLW613067S, BLW613068S, BLW613071S, BLW613078S, BLW613079S, BLW613081S, BLW613082S, BLW613103S, BLW613104S, BLW613105S, BLW613115S, BLW613116S, BLW613117S, BLW613137S, BLW613138S, BLW613178S, BLW613179S, BLW613184S, BLW613185S, BLW613210S, BLW613213S, BLW613370S, BLW613371S, BLW613372S, BLW613373S, BLW613374S, BLW613377S, BLW613425S, BLW613427S, BLW613429S, BLW613431S, BLW613433S, BLW613436S, BLW613437S, BLW613438S, BLW613440S, BLW613442S, BLW613449S, BLW613450S, BLW613452S, BLW613456S, BLW613457S, BLW613459S, BLW613462S, BLW613469S, BLW613474S, BLW613483S, BLW613484S, BLW613486S, BLW613487S, BLW613496S, BLW613531S, BLW613533S, BLW613537S, BLW613540S, BLW613546S, BLW613547S, BLW613548S, BLW613549S, BLW613550S, BLW613579S, BLW613619S, BLW613620S, BLW613621S, BLW613685S, BLW613686S, BLW613687S, BLW613955S, BLW613956S, BLW613957S, BLW613958S, BLW613959S, BLW613960S, BLW613961S, BLW614044S, BLW614045S, BLW614046S, BLW614047S, BLW614050S, BLW614053S, BLW614055S, BLW614058S, BLW614059S, BLW614061S, BLW614062S, BLW614064S, BLW614066S, BLW614091S, BLW614096S, BLW614100S, BLW614111S, BLW614127S, BLW614132S, BLW614146S, BLW614147S, BLW614158S, BLW614163S, BLW614164S, BLW614168S, BLW614170S, BLW614171S, BLW614172S, BLW614176S, BLW614177S, BLW614182S, BLW614183S, BLW614191S, BLW614212S, BLW614219S, BLW614229S, BLW614231S, BLW614232S, BLW614233S, BLW614234S, BLW614237S, BLW614239S, BLW614243S, BLW614251S, BLW614252S, BLW614263S, BLW614264S, BLW614265S, BLW614266S, BLW614267S, BLW614268S, BLW614274S, BLW614277S, BLW614278S, BLW614279S, BLW614280S, BLW614282S, BLW614286S, BLW614287S, BLW614288S, BLW614289S, BLW614290S, BLW614291S, BLW614292S, BLW614293S, BLW614294S, BLW614296S, BLW614318S, BLW614355S, BLW614358S, BLW614588S, BLW614671S, BLW614771S, BLW614772S, BLW614787S, BLW614788S, BLW614789S, BLW614790S, BLW614791S, BLW614792S, BLW614793S, BLW614794S, BLW614845S, BLW614846S, BLW614847S, BLW614848S, BLW614871S, BLW614872S, BLW614876S, BLW614878S, BLW614879S, BLW614884S, BLW614887S, BLW614901S, BLW614952S, BLW614953S, BLW614954S, BLW615001S, BLW615009S, BLW615011S, BLW615012S, BLW615013S, BLW615014S, BLW615015S, BLW615016S, BLW615017S, BLW615018S, BLW615019S, BLW615020S, BLW615029S, BLW615030S, BLW615177S, BLW615178S, BLW615201S, BLW615202S, BLW615203S, BLW615204S, BLW615384S, BLW615385S, BLW615386S, BLW615387S, BLW615388S, BLW615389S, BLW615391S, BLW615399S, BLW615400S, BLW615401S, BLW615402S, BLW615403S, BLW615404S, BLW615405S, BLW615406S, BLW615407S, BLW615618S, BLW615625S, BLW615629S, BLW615630S, BLW615631S, BLW615668S, BLW615669S, BLW615670S, BLW615671S, BLW615673S, BLW615839S, BLW615843S, BLW615847S, BLW615848S, BLW610504S, BLW608603S, BLW608602S, BLW610760S, BLW610557S, BLW611759S, BLW611957S, BLW612205S, BLW612646S, BLW612663S, BLW612778S, BLW612779S, BLW612822S, BLW612823S, BLW612850S, BLW612852S, BLW612853S, BLW612892S, BLW612893S, BLW612897S, BLW613340S, BLW613375S, BLW613376S, BLW613378S, BLW613410S, BLW613418S, BLW613489S, BLW613512S, BLW613574S, BLW613577S, BLW613578S, BLW613683S, BLW613684S, BLW613763S, BLW613817S, BLW613818S, BLW613972S, BLW613973S, BLW613974S, BLW613975S, BLW614074S, BLW614076S, BLW614078S, BLW614080S, BLW614086S, BLW614087S, BLW614088S, BLW614089S, BLW614090S, BLW614092S, BLW614283S, BLW614284S, BLW614315S, BLW614317S, BLW614322S, BLW614323S, BLW614604S, BLW614642S, BLW614682S, BLW614685S, BLW614691S, BLW614692S, BLW614807S, BLW614808S, BLW614864S, BLW614966S, BLW614998S, BLW614999S, BLW615003S, BLW615031S, BLW615156S, BLW615157S, BLW615195S, BLW615409S, BLW615472S, BLW615474S, BLW615571S, BLW615611S, BLW615614S, BLW615615S, BLW615620S, BLW615835S, BLW615842S, BLW615844S, BLW615846S, BLW610920S, BLW611733S, BLW611785S, BLW611788S, BLW612069S, BLW612388S, BLW612521S, BLW612531S, BLW612532S, BLW613324S, BLW613330S, BLW613689S, BLW609905S, BLW609405S, BLW609590S, BLW609626S, BLW609724S, BLW609732S, BLW610540S, BLW610548S, BLW610551S, BLW610552S, BLW610553S, BLW610555S, BLW610597S, BLW610598S, BLW610599S, BLW610600S, BLW610601S, BLW610602S, BLW610603S, BLW610604S, BLW610653S, BLW610656S, BLW610678S, BLW610696S, BLW610700S, BLW610754S, BLW610755S, BLW610758S, BLW610759S, BLW611372S, BLW611380S, BLW611732S, BLW611798S, BLW612053S, BLW612077S, BLW612150S, BLW612158S, BLW612203S, BLW612281S, BLW612762S, BLW612997S, BLW613003S, BLW613010S, BLW613720S, BLW614457S, BLW614474S, BLW614865S, BLW611906S, BLW610158S, BLW610738S, BLW612631S, BLW612736S, BLW612919S, BLW613715S, BLW614385S, BLW608248S, BLW610728S, BLW612609S, BLW612147S, BLW613827S, BLW614015S, BLW614440S, BLW610505S, BLW613037S, BLW611761S, BLW612101S, BLW612105S, BLW612568S, BLW612628S, BLW612693S, BLW613083S, BLW613095S, BLW613098S, BLW613129S, BLW614432S, BLW614435S, BLW614465S, BLW614516S, BLW614517S, BLW614518S, BLW614519S, BLW614694S, BLW614695S, BLW614698S, BLW614701S, BLW609678S, BLW609830S, BLW610182S, BLW610525S, BLW611760S, BLW612915S, BLW612916S, BLW612934S, BLW612936S, BLW614476S, BLW610514S, BLW614249S, BLW613806S, BLW610260S, BLW610838S, BLW609302S, BLW609306S, BLW609317S, BLW609319S, BLW609325S, BLW609976S, BLW610583S, BLW610853S, BLW610907S, BLW610911S, BLW610921S, BLW611017S, BLW611058S, BLW611151S, BLW611167S, BLW611186S, BLW609251S, BLW609571S, BLW613173S, BLW609910S, BLW609912S, BLW612956S, BLW613211S, BLW613791S, BLW613199S, BLW613751S, BLW610518S, BLW610338S, BLW611797S, BLW613815S, BLW611864S, BLW609445S, BLW608814S, BLW608815S, BLW609082S, BLW609205S, BLW609299S, BLW609300S, BLW609958S, BLW608405S, BLW608052S, BLW608053S, BLW608100S, BLW608102S, BLW608108S, BLW608535S, BLW608558S, BLW608575S, BLW608601S, BLW608672S, BLW608679S, BLW608796S, BLW608797S, BLW608834S, BLW608845S, BLW608881S, BLW609159S, BLW609458S, BLW609462S, BLW609469S, BLW609599S, BLW610043S, BLW610183S, BLW610283S, BLW610300S, BLW610308S, BLW610527S, BLW610528S, BLW610530S, BLW610704S, BLW609795S, BLW609805S, BLW609825S, BLW610347S, BLW610348S, BLW610442S, BLW610444S, BLW610861S, BLW610868S, BLW610876S, BLW610890S, BLW610906S, BLW611018S, BLW611037S, BLW611038S, BLW611104S, BLW611109S, BLW611174S, BLW611367S, BLW611579S, BLW611582S, BLW611611S, BLW611612S, BLW611623S, BLW611647S, BLW611674S, BLW611742S, BLW611750S, BLW611865S, BLW611970S, BLW612020S, BLW612026S, BLW612029S, BLW612195S, BLW612199S, BLW612200S, BLW612232S, BLW612251S, BLW612292S, BLW612311S, BLW612390S, BLW612391S, BLW612396S, BLW612400S, BLW612401S, BLW612459S, BLW612466S, BLW612487S, BLW612488S, BLW612489S, BLW612503S, BLW612504S, BLW612507S, BLW612508S, BLW612509S, BLW612510S, BLW612511S, BLW612514S, BLW612522S, BLW612523S, BLW612542S, BLW612546S, BLW612554S, BLW612557S, BLW612558S, BLW612589S, BLW612617S, BLW612620S, BLW612622S, BLW612635S, BLW612637S, BLW612678S, BLW612679S, BLW612680S, BLW612685S, BLW612687S, BLW612689S, BLW612691S, BLW612692S, BLW612694S, BLW612697S, BLW612698S, BLW612707S, BLW612713S, BLW612714S, BLW612775S, BLW612793S, BLW612824S, BLW612828S, BLW612832S, BLW612835S, BLW612836S, BLW612837S, BLW612838S, BLW612839S, BLW612840S, BLW612841S, BLW612842S, BLW612843S, BLW612844S, BLW612854S, BLW612881S, BLW612883S, BLW612884S, BLW612885S, BLW612894S, BLW612899S, BLW612908S, BLW612909S, BLW612910S, BLW612920S, BLW612923S, BLW612924S, BLW612926S, BLW612935S, BLW612995S, BLW613012S, BLW613014S, BLW613015S, BLW613016S, BLW613018S, BLW613019S, BLW613020S, BLW613025S, BLW613026S, BLW613032S, BLW613086S, BLW613090S, BLW613093S, BLW613096S, BLW613099S, BLW613110S, BLW613111S, BLW613112S, BLW613113S, BLW613114S, BLW613147S, BLW613148S, BLW613150S, BLW613151S, BLW613152S, BLW613154S, BLW613156S, BLW613157S, BLW613159S, BLW613160S, BLW613161S, BLW613163S, BLW613164S, BLW613166S, BLW613167S, BLW613168S, BLW613170S, BLW613174S, BLW613189S, BLW613192S, BLW613195S, BLW613209S, BLW613263S, BLW613273S, BLW613277S, BLW613281S, BLW613294S, BLW613310S, BLW613311S, BLW613313S, BLW613315S, BLW613316S, BLW613317S, BLW613322S, BLW613323S, BLW613325S, BLW613329S, BLW613331S, BLW613332S, BLW613335S, BLW613336S, BLW613337S, BLW613355S, BLW613398S, BLW613403S, BLW613408S, BLW613409S, BLW613411S, BLW613412S, BLW613415S, BLW613420S, BLW613421S, BLW613424S, BLW613430S, BLW613435S, BLW613443S, BLW613444S, BLW613448S, BLW613454S, BLW613458S, BLW613461S, BLW613463S, BLW613468S, BLW613471S, BLW613473S, BLW613478S, BLW613480S, BLW613481S, BLW613485S, BLW613488S, BLW613493S, BLW613495S, BLW613500S, BLW613502S, BLW613503S, BLW613506S, BLW613509S, BLW613510S, BLW613562S, BLW613563S, BLW613570S, BLW613573S, BLW613575S, BLW613576S, BLW613580S, BLW613581S, BLW613582S, BLW613589S, BLW613590S, BLW613591S, BLW613592S, BLW613593S, BLW613594S, BLW613595S, BLW613597S, BLW613598S, BLW613599S, BLW613601S, BLW613602S, BLW613603S, BLW613669S, BLW613670S, BLW613671S, BLW613673S, BLW613674S, BLW613675S, BLW613676S, BLW613677S, BLW613678S, BLW613679S, BLW613692S, BLW613693S, BLW613694S, BLW613695S, BLW613696S, BLW613698S, BLW613700S, BLW613701S, BLW613713S, BLW613717S, BLW613718S, BLW613719S, BLW613722S, BLW613725S, BLW613728S, BLW613730S, BLW613731S, BLW613733S, BLW613735S, BLW613741S, BLW613755S, BLW613760S, BLW613771S, BLW613779S, BLW613787S, BLW613788S, BLW613790S, BLW613798S, BLW613801S, BLW613805S, BLW613812S, BLW613819S, BLW613835S, BLW613837S, BLW613838S, BLW613839S, BLW613842S, BLW613843S, BLW613928S, BLW613929S, BLW613947S, BLW613948S, BLW613950S, BLW613951S, BLW613952S, BLW613953S, BLW613954S, BLW613967S, BLW613971S, BLW613976S, BLW613977S, BLW613978S, BLW613979S, BLW613980S, BLW613991S, BLW613992S, BLW613993S, BLW613994S, BLW613995S, BLW613996S, BLW613997S, BLW613998S, BLW613999S, BLW614000S, BLW614001S, BLW614002S, BLW614003S, BLW614004S, BLW614005S, BLW614006S, BLW614007S, BLW614008S, BLW614009S, BLW614010S, BLW614011S, BLW614013S, BLW614014S, BLW614016S, BLW614017S, BLW614018S, BLW614020S, BLW614021S, BLW614037S, BLW614040S, BLW614048S, BLW614049S, BLW614051S, BLW614052S, BLW614054S, BLW614057S, BLW614060S, BLW614063S, BLW614071S, BLW614072S, BLW614073S, BLW614077S, BLW614085S, BLW614095S, BLW614112S, BLW614114S, BLW614116S, BLW614118S, BLW614121S, BLW614122S, BLW614130S, BLW614145S, BLW614149S, BLW614150S, BLW614151S, BLW614152S, BLW614153S, BLW614154S, BLW614155S, BLW614156S, BLW614157S, BLW614162S, BLW614175S, BLW614178S, BLW614179S, BLW614181S, BLW614184S, BLW614186S, BLW614187S, BLW614188S, BLW614190S, BLW614192S, BLW614193S, BLW614194S, BLW614195S, BLW614196S, BLW614197S, BLW614198S, BLW614199S, BLW614200S, BLW614201S, BLW614202S, BLW614203S, BLW614204S, BLW614205S, BLW614206S, BLW614207S, BLW614208S, BLW614209S, BLW614210S, BLW614213S, BLW614216S, BLW614217S, BLW614218S, BLW614220S, BLW614221S, BLW614222S, BLW614224S, BLW614225S, BLW614226S, BLW614227S, BLW614228S, BLW614230S, BLW614255S, BLW614295S, BLW614298S, BLW614299S, BLW614301S, BLW614302S, BLW614303S, BLW614308S, BLW614325S, BLW614326S, BLW614328S, BLW614332S, BLW614333S, BLW614334S, BLW614335S, BLW614336S, BLW614337S, BLW614342S, BLW614343S, BLW614346S, BLW614347S, BLW614348S, BLW614349S, BLW614363S, BLW614422S, BLW614488S, BLW614608S, BLW614609S, BLW614610S, BLW614617S, BLW614623S, BLW614624S, BLW614625S, BLW614626S, BLW614627S, BLW614630S, BLW614632S, BLW614635S, BLW614637S, BLW614639S, BLW614640S, BLW614647S, BLW614670S, BLW614696S, BLW614697S, BLW614699S, BLW614700S, BLW614710S, BLW614712S, BLW614716S, BLW614717S, BLW614718S, BLW614722S, BLW614744S, BLW614755S, BLW614761S, BLW614762S, BLW614764S, BLW614768S, BLW614769S, BLW614770S, BLW614795S, BLW614796S, BLW614797S, BLW614798S, BLW614799S, BLW614800S, BLW614801S, BLW614811S, BLW614812S, BLW614813S, BLW614814S, BLW614815S, BLW614816S, BLW614817S, BLW614818S, BLW614819S, BLW614820S, BLW614841S, BLW614842S, BLW614843S, BLW614844S, BLW614856S, BLW614857S, BLW614858S, BLW614859S, BLW614860S, BLW614861S, BLW614862S, BLW614866S, BLW614867S, BLW614868S, BLW614869S, BLW614870S, BLW614939S, BLW614940S, BLW614941S, BLW614942S, BLW614959S, BLW614961S, BLW614972S, BLW614974S, BLW614975S, BLW614976S, BLW614977S, BLW614978S, BLW614979S, BLW614980S, BLW614981S, BLW614983S, BLW614984S, BLW614986S, BLW614987S, BLW614988S, BLW614989S, BLW614990S, BLW614991S, BLW614992S, BLW615006S, BLW615021S, BLW615022S, BLW615023S, BLW615024S, BLW615055S, BLW615136S, BLW615137S, BLW615139S, BLW615141S, BLW615142S, BLW615143S, BLW615144S, BLW615145S, BLW615146S, BLW615147S, BLW615148S, BLW615149S, BLW615150S, BLW615152S, BLW615153S, BLW615154S, BLW615155S, BLW615169S, BLW615205S, BLW615206S, BLW615207S, BLW615208S, BLW615209S, BLW615211S, BLW615212S, BLW615213S, BLW615214S, BLW615217S, BLW615218S, BLW615219S, BLW615229S, BLW615230S, BLW615231S, BLW615233S, BLW615234S, BLW615236S, BLW615237S, BLW615242S, BLW615244S, BLW615245S, BLW615246S, BLW615247S, BLW615248S, BLW615249S, BLW615250S, BLW615266S, BLW615274S, BLW615275S, BLW615279S, BLW615313S, BLW615318S, BLW615339S, BLW615351S, BLW615451S, BLW615452S, BLW615453S, BLW615454S, BLW615607S, BLW615840S, BLW608581S, BLW609190S, BLW609917S, BLW613135S, BLW610498S, BLW610515S, BLW610155S, BLW611916S, BLW611256S, BLW611609S, BLW611613S, BLW611747S, BLW611748S, BLW611852S, BLW612321S, BLW612322S, BLW612323S, BLW612327S, BLW612742S, BLW612755S, BLW612801S, BLW614297S, BLW614312S, BLW614316S, BLW614327S, BLW614329S, BLW609275S, BLW609515S, BLW609271S, BLW609749S, BLW611976S, BLW608110S, BLW610369S, BLW610370S, BLW610371S, BLW610372S, BLW610373S, BLW610374S, BLW610375S, BLW610376S, BLW610377S, BLW610378S, BLW610379S, BLW610380S, BLW610381S, BLW610382S, BLW610383S, BLW610384S, BLW610385S, BLW610386S, BLW610387S, BLW610388S, BLW610389S, BLW610390S, BLW610391S, BLW610392S, BLW610393S, BLW610394S, BLW610395S, BLW610396S, BLW610397S, BLW610399S, BLW610400S, BLW610401S, BLW610402S, BLW610403S, BLW610404S, BLW610416S, BLW610417S, BLW610421S, BLW610469S, BLW610537S, BLW610539S, BLW610541S, BLW610544S, BLW610545S, BLW610549S, BLW610550S, BLW610554S, BLW610559S, BLW610562S, BLW610563S, BLW610564S, BLW610565S, BLW610566S, BLW610580S, BLW610585S, BLW610586S, BLW610587S, BLW610588S, BLW610589S, BLW610590S, BLW610591S, BLW610592S, BLW610593S, BLW610594S, BLW610609S, BLW610610S, BLW610611S, BLW610633S, BLW610635S, BLW610992S, BLW611008S, BLW611009S, BLW611010S, BLW611013S, BLW611015S, BLW611020S, BLW611025S, BLW611029S, BLW611035S, BLW611036S, BLW611042S, BLW611044S, BLW611045S, BLW611049S, BLW611050S, BLW611053S, BLW611057S, BLW611060S, BLW611063S, BLW611064S, BLW611065S, BLW611073S, BLW611074S, BLW611075S, BLW611076S, BLW611078S, BLW611091S, BLW611095S, BLW611096S, BLW611098S, BLW611099S, BLW611107S, BLW611117S, BLW611181S, BLW611192S, BLW611234S, BLW611235S, BLW611236S, BLW611237S, BLW611242S, BLW611243S, BLW611251S, BLW611253S, BLW611254S, BLW611255S, BLW611257S, BLW611258S, BLW611261S, BLW611262S, BLW611264S, BLW611266S, BLW611267S, BLW611283S, BLW611284S, BLW611285S, BLW611287S, BLW611288S, BLW611289S, BLW611290S, BLW611306S, BLW611308S, BLW611309S, BLW611310S, BLW611323S, BLW611324S, BLW611325S, BLW611326S, BLW611327S, BLW611328S, BLW611331S, BLW611336S, BLW611337S, BLW611338S, BLW611386S, BLW611387S, BLW611388S, BLW611401S, BLW611403S, BLW611405S, BLW611407S, BLW611410S, BLW611412S, BLW611413S, BLW611415S, BLW611417S, BLW611429S, BLW611432S, BLW611433S, BLW611435S, BLW611440S, BLW611442S, BLW611443S, BLW611444S, BLW611446S, BLW611448S, BLW611450S, BLW611451S, BLW611452S, BLW611453S, BLW611455S, BLW611456S, BLW611459S, BLW611460S, BLW611606S, BLW611625S, BLW611626S, BLW611630S, BLW611631S, BLW611632S, BLW611633S, BLW611636S, BLW611663S, BLW611667S, BLW611675S, BLW611677S, BLW611680S, BLW611681S, BLW611684S, BLW611685S, BLW611688S, BLW611689S, BLW611691S, BLW611694S, BLW611707S, BLW611708S, BLW611716S, BLW611717S, BLW611718S, BLW611719S, BLW611722S, BLW611796S, BLW611992S, BLW611994S, BLW612001S, BLW612002S, BLW612003S, BLW612004S, BLW612011S, BLW612012S, BLW612015S, BLW612016S, BLW612017S, BLW612019S, BLW612021S, BLW612022S, BLW612023S, BLW612024S, BLW612025S, BLW612027S, BLW612028S, BLW612030S, BLW612033S, BLW612034S, BLW612038S, BLW612039S, BLW612040S, BLW612042S, BLW612043S, BLW612049S, BLW612052S, BLW612054S, BLW612071S, BLW612081S, BLW612084S, BLW612085S, BLW612086S, BLW612087S, BLW612088S, BLW612090S, BLW612095S, BLW612097S, BLW612099S, BLW612102S, BLW612118S, BLW612119S, BLW612120S, BLW612121S, BLW612127S, BLW612128S, BLW612139S, BLW612160S, BLW612163S, BLW612166S, BLW612168S, BLW612172S, BLW612173S, BLW612174S, BLW612176S, BLW612177S, BLW612178S, BLW612179S, BLW612183S, BLW612184S, BLW612186S, BLW612189S, BLW612194S, BLW612196S, BLW612197S, BLW612198S, BLW612202S, BLW612206S, BLW612208S, BLW612211S, BLW612213S, BLW612222S, BLW612224S, BLW612227S, BLW612230S, BLW612235S, BLW612239S, BLW612241S, BLW612242S, BLW612243S, BLW612245S, BLW612250S, BLW612252S, BLW612255S, BLW612264S, BLW612300S, BLW612303S, BLW612305S, BLW612315S, BLW612317S, BLW612318S, BLW612320S, BLW612329S, BLW612330S, BLW612331S, BLW612332S, BLW612333S, BLW612336S, BLW612337S, BLW612340S, BLW612397S, BLW612398S, BLW612399S, BLW612402S, BLW612403S, BLW612404S, BLW612405S, BLW612406S, BLW612407S, BLW612408S, BLW612411S, BLW612413S, BLW612416S, BLW612417S, BLW612418S, BLW612419S, BLW612420S, BLW612421S, BLW612422S, BLW612424S, BLW612425S, BLW612427S, BLW612428S, BLW612431S, BLW612432S, BLW612433S, BLW612434S, BLW612435S, BLW612439S, BLW612447S, BLW612627S, BLW612633S, BLW612636S, BLW612638S, BLW612640S, BLW612641S, BLW612819S, BLW612825S, BLW612827S, BLW612830S, BLW612868S, BLW612870S, BLW612876S, BLW612879S, BLW612887S, BLW612922S, BLW612930S, BLW612959S, BLW612963S, BLW612965S, BLW612970S, BLW612971S, BLW612974S, BLW612978S, BLW612983S, BLW612990S, BLW612993S, BLW613021S, BLW613023S, BLW613046S, BLW613058S, BLW613070S, BLW613217S, BLW613230S, BLW613259S, BLW613260S, BLW613262S, BLW613266S, BLW613267S, BLW613268S, BLW613269S, BLW613270S, BLW613271S, BLW613274S, BLW613275S, BLW613276S, BLW613278S, BLW613279S, BLW613280S, BLW613282S, BLW613284S, BLW613287S, BLW613288S, BLW613290S, BLW613293S, BLW613295S, BLW613296S, BLW613668S, BLW613823S, BLW613824S, BLW613825S, BLW613826S, BLW613830S, BLW613831S, BLW613832S, BLW613833S, BLW613834S, BLW613836S, BLW613840S, BLW613841S, BLW613844S, BLW613845S, BLW613846S, BLW613847S, BLW613848S, BLW613849S, BLW613850S, BLW613851S, BLW613852S, BLW613853S, BLW613854S, BLW613855S, BLW613856S, BLW613857S, BLW613858S, BLW613859S, BLW613860S, BLW613861S, BLW613871S, BLW613872S, BLW613873S, BLW613874S, BLW613875S, BLW613876S, BLW613877S, BLW613878S, BLW613879S, BLW613905S, BLW613906S, BLW613908S, BLW613909S, BLW613910S, BLW613911S, BLW613912S, BLW613913S, BLW613914S, BLW613915S, BLW613916S, BLW613917S, BLW613918S, BLW614138S, BLW614144S, BLW614165S, BLW614173S, BLW614185S, BLW614189S, BLW614214S, BLW614215S, BLW614344S, BLW614345S, BLW614354S, BLW614366S, BLW614367S, BLW614370S, BLW614378S, BLW614383S, BLW614414S, BLW614415S, BLW614423S, BLW614428S, BLW614450S, BLW614451S, BLW614452S, BLW614453S, BLW614454S, BLW614455S, BLW614499S, BLW614500S, BLW614501S, BLW614502S, BLW614503S, BLW614504S, BLW614505S, BLW614506S, BLW614520S, BLW614530S, BLW614531S, BLW614532S, BLW614534S, BLW614535S, BLW614536S, BLW614537S, BLW614538S, BLW614539S, BLW614540S, BLW614541S, BLW614542S, BLW614557S, BLW614559S, BLW614560S, BLW614561S, BLW614562S, BLW614563S, BLW614564S, BLW614565S, BLW614566S, BLW614567S, BLW614568S, BLW614569S, BLW614570S, BLW614571S, BLW614572S, BLW614573S, BLW614574S, BLW614575S, BLW614576S, BLW614578S, BLW614579S, BLW614580S, BLW614863S, BLW613024S, BLW612817S, BLW612913S, BLW613040S, BLW613220S, BLW614616S, BLW610788S, BLW611012S, BLW612586S, BLW613777S, BLW608605S, BLW609330S, BLW609380S, BLW609581S, BLW609582S, BLW609584S, BLW609585S, BLW609586S, BLW609587S, BLW609588S, BLW609589S, BLW609591S, BLW609592S, BLW609593S, BLW609594S, BLW609595S, BLW609597S, BLW609600S, BLW609601S, BLW609602S, BLW609603S, BLW609604S, BLW609605S, BLW609606S, BLW609607S, BLW609608S, BLW609609S, BLW609610S, BLW609611S, BLW609612S, BLW609613S, BLW609616S, BLW609617S, BLW609618S, BLW609619S, BLW609620S, BLW609621S, BLW609622S, BLW609623S, BLW609624S, BLW609627S, BLW609629S, BLW609631S, BLW609632S, BLW609633S, BLW609634S, BLW609635S, BLW609636S, BLW609637S, BLW609638S, BLW609639S, BLW609640S, BLW609641S, BLW609642S, BLW609643S, BLW609644S, BLW609645S, BLW609646S, BLW609647S, BLW609648S, BLW609649S, BLW609650S, BLW609654S, BLW609655S, BLW609656S, BLW609657S, BLW609659S, BLW609706S, BLW609998S, BLW609999S, BLW610005S, BLW610023S, BLW610025S, BLW610032S, BLW610117S, BLW610118S, BLW610119S, BLW610141S, BLW610142S, BLW610146S, BLW610224S, BLW610225S, BLW610226S, BLW610227S, BLW610229S, BLW610230S, BLW610231S, BLW610233S, BLW610407S, BLW610415S, BLW610419S, BLW610422S, BLW610426S, BLW610430S, BLW610431S, BLW610432S, BLW610435S, BLW610446S, BLW610490S, BLW610516S, BLW610526S, BLW610538S, BLW610542S, BLW610543S, BLW610546S, BLW610547S, BLW610560S, BLW610561S, BLW610571S, BLW610572S, BLW610605S, BLW610606S, BLW610637S, BLW610641S, BLW610745S, BLW610746S, BLW610747S, BLW610748S, BLW610749S, BLW610750S, BLW610751S, BLW610752S, BLW610753S, BLW608408S, BLW608513S, BLW608534S, BLW608792S, BLW609479S, BLW610558S, BLW612601S, BLW613907S, BLW614413S, BLW610495S, BLW612031S, BLW612107S, BLW613738S, BLW613767S, BLW613772S, BLW613774S, BLW613775S, BLW614285S, BLW614471S, BLW614513S, BLW614514S, BLW614748S, BLW614757S, BLW615151S, BLW611593S, BLW611872S, BLW612171S, BLW612645S, BLW612703S, BLW612865S, BLW613870S, BLW609694S, BLW609723S, BLW609725S, BLW609726S, BLW609727S, BLW609728S, BLW609729S, BLW609730S, BLW609731S, BLW609733S, BLW609734S, BLW609736S, BLW609737S, BLW609738S, BLW609739S, BLW609831S, BLW609832S, BLW610075S, BLW610076S, BLW610077S, BLW610086S, BLW610097S, BLW610098S, BLW610099S, BLW610100S, BLW610101S, BLW610109S, BLW610110S, BLW610111S, BLW610350S, BLW610356S, BLW610357S, BLW610359S, BLW610361S, BLW610363S, BLW610398S, BLW610413S, BLW610414S, BLW610420S, BLW610428S, BLW610429S, BLW610436S, BLW610437S, BLW610730S, BLW610732S, BLW610756S, BLW610772S, BLW610781S, BLW610818S, BLW610822S, BLW610823S, BLW610824S, BLW610835S, BLW610881S, BLW610894S, BLW610898S, BLW610901S, BLW610936S, BLW610951S, BLW610953S, BLW611028S, BLW611030S, BLW611033S, BLW611048S, BLW611059S, BLW611072S, BLW611106S, BLW611120S, BLW611132S, BLW611163S, BLW611172S, BLW611196S, BLW611202S, BLW611252S, BLW611265S, BLW611423S, BLW611439S, BLW611445S, BLW611447S, BLW611449S, BLW611454S, BLW611457S, BLW611463S, BLW611576S, BLW611577S, BLW611602S, BLW611617S, BLW611655S, BLW611656S, BLW611661S, BLW611664S, BLW611665S, BLW611666S, BLW611703S, BLW611704S, BLW611705S, BLW611706S, BLW611713S, BLW611714S, BLW611738S, BLW611789S, BLW611792S, BLW611802S, BLW611809S, BLW611824S, BLW611827S, BLW611828S, BLW611830S, BLW611842S, BLW611844S, BLW611846S, BLW611847S, BLW611982S, BLW612068S, BLW612073S, BLW612074S, BLW612075S, BLW612082S, BLW612114S, BLW612135S, BLW612175S, BLW612180S, BLW612234S, BLW612236S, BLW612262S, BLW612275S, BLW612294S, BLW612298S, BLW612326S, BLW612328S, BLW612455S, BLW612456S, BLW612457S, BLW612464S, BLW612468S, BLW612476S, BLW612491S, BLW612492S, BLW612493S, BLW612499S, BLW612512S, BLW612513S, BLW612560S, BLW612708S, BLW612709S, BLW612712S, BLW612720S, BLW612726S, BLW612728S, BLW612731S, BLW612732S, BLW612733S, BLW612734S, BLW612735S, BLW612741S, BLW612744S, BLW612746S, BLW612747S, BLW612749S, BLW612752S, BLW612753S, BLW612754S, BLW612756S, BLW612757S, BLW612758S, BLW612760S, BLW612761S, BLW612764S, BLW612783S, BLW612784S, BLW612785S, BLW612803S, BLW612845S, BLW612846S, BLW612847S, BLW612848S, BLW612849S, BLW612855S, BLW612856S, BLW612857S, BLW612862S, BLW612863S, BLW612864S, BLW612866S, BLW612869S, BLW612873S, BLW612874S, BLW613001S, BLW613004S, BLW613051S, BLW613057S, BLW613059S, BLW613062S, BLW613063S, BLW613064S, BLW613065S, BLW613066S, BLW613107S, BLW613300S, BLW613301S, BLW613303S, BLW613305S, BLW613306S, BLW613307S, BLW613309S, BLW613357S, BLW613380S, BLW613382S, BLW613383S, BLW613384S, BLW613385S, BLW613386S, BLW613387S, BLW613389S, BLW613390S, BLW613393S, BLW613394S, BLW613395S, BLW613396S, BLW613399S, BLW613400S, BLW613401S, BLW613402S, BLW613543S, BLW613553S, BLW613557S, BLW613558S, BLW613559S, BLW613560S, BLW613561S, BLW613564S, BLW613567S, BLW613568S, BLW613569S, BLW613572S, BLW613588S, BLW613644S, BLW613645S, BLW613646S, BLW613648S, BLW613649S, BLW613664S, BLW613665S, BLW613666S, BLW613703S, BLW613704S, BLW613705S, BLW613709S, BLW613710S, BLW613711S, BLW613712S, BLW613821S, BLW613981S, BLW613982S, BLW613983S, BLW613984S, BLW613985S, BLW613986S, BLW613987S, BLW613988S, BLW613989S, BLW613990S, BLW614022S, BLW614023S, BLW614024S, BLW614025S, BLW614082S, BLW614084S, BLW614250S, BLW614254S, BLW614256S, BLW614258S, BLW614261S, BLW614319S, BLW614320S, BLW614321S, BLW614351S, BLW614352S, BLW614373S, BLW614374S, BLW614643S, BLW614644S, BLW614645S, BLW614646S, BLW614664S, BLW614665S, BLW614666S, BLW614667S, BLW614668S, BLW614669S, BLW614725S, BLW614728S, BLW614729S, BLW614747S, BLW614750S, BLW614780S, BLW614781S, BLW614782S, BLW614802S, BLW614803S, BLW614809S, BLW614830S, BLW614831S, BLW614832S, BLW614833S, BLW614835S, BLW614836S, BLW614837S, BLW614838S, BLW614839S, BLW614840S, BLW614853S, BLW614854S, BLW614855S, BLW614936S, BLW614937S, BLW614938S, BLW614947S, BLW614948S, BLW614950S, BLW614951S, BLW615004S, BLW615007S, BLW615008S, BLW615084S, BLW615163S, BLW615164S, BLW615165S, BLW615375S, BLW615376S, BLW615377S, BLW615378S, BLW615379S, BLW615599S, BLW615612S, BLW615613S, BLW615622S, BLW615623S, BLW615624S, BLW615627S, BLW615628S, BLW615672S, BLW615693S, BLW615701S, BLW615702S, BLW615703S, BLW615705S, BLW615711S, BLW615712S, BLW615850S, BLW615851S, BLW615852S, BLW615853S, BLW609796S, BLW609951S, BLW610095S, BLW610102S, BLW610342S, BLW610364S, BLW610438S, BLW610762S, BLW610801S, BLW610807S, BLW611039S, BLW611068S, BLW611157S, BLW611248S, BLW611400S, BLW611478S, BLW611545S, BLW611558S, BLW611601S, BLW611608S, BLW611640S, BLW611641S, BLW611643S, BLW611657S, BLW611658S, BLW611660S, BLW611671S, BLW611672S, BLW611673S, BLW611834S, BLW612032S, BLW612046S, BLW612148S, BLW612190S, BLW612231S, BLW612233S, BLW612238S, BLW612240S, BLW612244S, BLW612248S, BLW612257S, BLW612265S, BLW612308S, BLW612310S, BLW612334S, BLW612383S, BLW612461S, BLW612467S, BLW612470S, BLW612477S, BLW612478S, BLW612479S, BLW612505S, BLW612506S, BLW612661S, BLW612662S, BLW612700S, BLW612701S, BLW612715S, BLW612716S, BLW612740S, BLW612795S, BLW612851S, BLW612858S, BLW612860S, BLW612861S, BLW612867S, BLW612878S, BLW612880S, BLW612882S, BLW612886S, BLW612888S, BLW612889S, BLW612939S, BLW612945S, BLW612946S, BLW612947S, BLW612975S, BLW612980S, BLW612981S, BLW612985S, BLW612986S, BLW613054S, BLW613077S, BLW613108S, BLW613109S, BLW613130S, BLW613131S, BLW613144S, BLW613146S, BLW613188S, BLW613223S, BLW613227S, BLW613229S, BLW613232S, BLW613240S, BLW613346S, BLW613347S, BLW613350S, BLW613353S, BLW613356S, BLW613455S, BLW613513S, BLW613517S, BLW613518S, BLW613519S, BLW613522S, BLW613534S, BLW613535S, BLW613541S, BLW613542S, BLW613544S, BLW613551S, BLW613638S, BLW613639S, BLW613640S, BLW613641S, BLW613643S, BLW613650S, BLW613653S, BLW613654S, BLW613656S, BLW613657S, BLW613964S, BLW613966S, BLW613968S, BLW613969S, BLW613970S, BLW614124S, BLW614125S, BLW614135S, BLW614137S, BLW614140S, BLW614141S, BLW614142S, BLW614159S, BLW614167S, BLW614169S, BLW614174S, BLW614180S, BLW614211S, BLW614269S, BLW614270S, BLW614272S, BLW614273S, BLW614359S, BLW614360S, BLW614361S, BLW614672S, BLW614673S, BLW614674S, BLW614730S, BLW614825S, BLW614826S, BLW614827S, BLW614828S, BLW614829S, BLW614834S, BLW614873S, BLW614874S, BLW614875S, BLW614885S, BLW614886S, BLW614943S, BLW614944S, BLW614945S, BLW614946S, BLW614962S, BLW614963S, BLW614964S, BLW614970S, BLW614971S, BLW615005S, BLW615025S, BLW615026S, BLW615032S, BLW615033S, BLW615034S, BLW615035S, BLW615036S, BLW615047S, BLW615048S, BLW615049S, BLW615050S, BLW615051S, BLW615053S, BLW615054S, BLW615071S, BLW615083S, BLW615088S, BLW615089S, BLW615090S, BLW615096S, BLW615097S, BLW615159S, BLW615160S, BLW615161S, BLW615162S, BLW615167S, BLW615196S, BLW615197S, BLW615198S, BLW615199S, BLW615200S, BLW615555S, BLW615608S, BLW615831S, BLW615834S, BLW615836S, BLW615855S, BLW609580S, BLW614719S, BLW609822S, BLW609824S, BLW610078S, BLW610087S, BLW610349S, BLW610365S, BLW610439S, BLW610737S, BLW610804S, BLW610825S, BLW610897S, BLW610914S, BLW610923S, BLW610927S, BLW610935S, BLW610939S, BLW610943S, BLW610950S, BLW610954S, BLW610955S, BLW610956S, BLW610957S, BLW610977S, BLW611011S, BLW611016S, BLW611023S, BLW611043S, BLW611051S, BLW611052S, BLW611054S, BLW611055S, BLW611056S, BLW611061S, BLW611062S, BLW611066S, BLW611081S, BLW611082S, BLW611110S, BLW611112S, BLW611113S, BLW611114S, BLW611115S, BLW611118S, BLW611119S, BLW611121S, BLW611126S, BLW611134S, BLW611138S, BLW611139S, BLW611143S, BLW611149S, BLW611154S, BLW611156S, BLW611164S, BLW611166S, BLW611169S, BLW611176S, BLW611180S, BLW611183S, BLW611313S, BLW611314S, BLW611330S, BLW611350S, BLW611351S, BLW611354S, BLW611358S, BLW611381S, BLW611404S, BLW611406S, BLW611411S, BLW611421S, BLW611485S, BLW611495S, BLW611539S, BLW611547S, BLW611549S, BLW611551S, BLW611567S, BLW611589S, BLW611592S, BLW611594S, BLW611595S, BLW611598S, BLW611603S, BLW611604S, BLW611605S, BLW611610S, BLW611614S, BLW611618S, BLW611619S, BLW611621S, BLW611624S, BLW611644S, BLW611645S, BLW611646S, BLW611651S, BLW611652S, BLW611654S, BLW611668S, BLW611669S, BLW611676S, BLW611678S, BLW611679S, BLW611682S, BLW611683S, BLW611686S, BLW611687S, BLW611690S, BLW611692S, BLW611709S, BLW611710S, BLW611712S, BLW611715S, BLW611720S, BLW611721S, BLW611723S, BLW611724S, BLW611725S, BLW611728S, BLW611731S, BLW611734S, BLW611737S, BLW611752S, BLW611753S, BLW611755S, BLW611758S, BLW611763S, BLW611764S, BLW611781S, BLW611786S, BLW611787S, BLW611794S, BLW611795S, BLW611799S, BLW611801S, BLW611810S, BLW611812S, BLW611813S, BLW611817S, BLW611818S, BLW611821S, BLW611823S, BLW611848S, BLW611850S, BLW611853S, BLW611854S, BLW611855S, BLW611869S, BLW611870S, BLW611879S, BLW611883S, BLW611889S, BLW611891S, BLW611983S, BLW611988S, BLW611989S, BLW612007S, BLW612089S, BLW612092S, BLW612096S, BLW612106S, BLW612111S, BLW612122S, BLW612123S, BLW612159S, BLW612162S, BLW612164S, BLW612165S, BLW612181S, BLW612187S, BLW612188S, BLW612207S, BLW612209S, BLW612212S, BLW612215S, BLW612216S, BLW612217S, BLW612219S, BLW612220S, BLW612221S, BLW612225S, BLW612226S, BLW612228S, BLW612229S, BLW612237S, BLW612247S, BLW612249S, BLW612253S, BLW612259S, BLW612260S, BLW612267S, BLW612268S, BLW612270S, BLW612271S, BLW612272S, BLW612273S, BLW612274S, BLW612282S, BLW612284S, BLW612287S, BLW612288S, BLW612293S, BLW612299S, BLW612301S, BLW612312S, BLW612313S, BLW612319S, BLW612324S, BLW612325S, BLW612335S, BLW612341S, BLW612343S, BLW612410S, BLW612412S, BLW612423S, BLW612429S, BLW612430S, BLW612436S, BLW612437S, BLW612440S, BLW612450S, BLW612458S, BLW612462S, BLW612465S, BLW612471S, BLW612473S, BLW612474S, BLW612475S, BLW612480S, BLW612481S, BLW612482S, BLW612483S, BLW612484S, BLW612490S, BLW612494S, BLW612495S, BLW612498S, BLW612501S, BLW612502S, BLW612515S, BLW612518S, BLW612519S, BLW612520S, BLW612526S, BLW612528S, BLW612529S, BLW612530S, BLW612533S, BLW612534S, BLW612536S, BLW612537S, BLW612538S, BLW612648S, BLW612665S, BLW612666S, BLW612673S, BLW612674S, BLW612676S, BLW612677S, BLW612705S, BLW612711S, BLW612727S, BLW612729S, BLW612730S, BLW612768S, BLW612769S, BLW612787S, BLW612789S, BLW612790S, BLW612791S, BLW612796S, BLW612797S, BLW612799S, BLW612800S, BLW612802S, BLW612871S, BLW612872S, BLW612875S, BLW612898S, BLW612900S, BLW612901S, BLW612902S, BLW612903S, BLW612905S, BLW612906S, BLW612907S, BLW612918S, BLW612950S, BLW612951S, BLW612952S, BLW612954S, BLW612957S, BLW612960S, BLW612961S, BLW612962S, BLW612968S, BLW612973S, BLW612977S, BLW612979S, BLW612987S, BLW612996S, BLW612998S, BLW613002S, BLW613005S, BLW613006S, BLW613007S, BLW613008S, BLW613009S, BLW613017S, BLW613022S, BLW613039S, BLW613044S, BLW613047S, BLW613048S, BLW613055S, BLW613073S, BLW613076S, BLW613080S, BLW613084S, BLW613087S, BLW613089S, BLW613091S, BLW613094S, BLW613097S, BLW613100S, BLW613101S, BLW613119S, BLW613120S, BLW613121S, BLW613124S, BLW613125S, BLW613126S, BLW613127S, BLW613128S, BLW613132S, BLW613133S, BLW613134S, BLW613136S, BLW613140S, BLW613141S, BLW613143S, BLW613180S, BLW613181S, BLW613183S, BLW613198S, BLW613204S, BLW613215S, BLW613216S, BLW613219S, BLW613221S, BLW613233S, BLW613234S, BLW613235S, BLW613237S, BLW613247S, BLW613339S, BLW613341S, BLW613342S, BLW613343S, BLW613345S, BLW613348S, BLW613349S, BLW613351S, BLW613352S, BLW613354S, BLW613358S, BLW613359S, BLW613360S, BLW613361S, BLW613362S, BLW613363S, BLW613364S, BLW613365S, BLW613366S, BLW613368S, BLW613369S, BLW613413S, BLW613414S, BLW613417S, BLW613432S, BLW613499S, BLW613505S, BLW613507S, BLW613511S, BLW613514S, BLW613515S, BLW613516S, BLW613520S, BLW613521S, BLW613523S, BLW613526S, BLW613527S, BLW613552S, BLW613554S, BLW613555S, BLW613556S, BLW613583S, BLW613584S, BLW613585S, BLW613626S, BLW613627S, BLW613628S, BLW613629S, BLW613630S, BLW613631S, BLW613632S, BLW613633S, BLW613634S, BLW613635S, BLW613637S, BLW613651S, BLW613652S, BLW613660S, BLW613661S, BLW613662S, BLW613937S, BLW613938S, BLW613939S, BLW613940S, BLW613941S, BLW613942S, BLW613943S, BLW613944S, BLW613945S, BLW613946S, BLW613962S, BLW613963S, BLW614026S, BLW614027S, BLW614028S, BLW614029S, BLW614030S, BLW614031S, BLW614032S, BLW614033S, BLW614034S, BLW614035S, BLW614036S, BLW614038S, BLW614039S, BLW614041S, BLW614043S, BLW614067S, BLW614068S, BLW614069S, BLW614070S, BLW614075S, BLW614079S, BLW614081S, BLW614093S, BLW614094S, BLW614097S, BLW614098S, BLW614099S, BLW614102S, BLW614103S, BLW614104S, BLW614105S, BLW614109S, BLW614110S, BLW614113S, BLW614115S, BLW614117S, BLW614119S, BLW614120S, BLW614128S, BLW614131S, BLW614133S, BLW614134S, BLW614136S, BLW614139S, BLW614143S, BLW614148S, BLW614160S, BLW614161S, BLW614235S, BLW614236S, BLW614238S, BLW614240S, BLW614241S, BLW614242S, BLW614245S, BLW614246S, BLW614247S, BLW614248S, BLW614253S, BLW614257S, BLW614259S, BLW614260S, BLW614262S, BLW614300S, BLW614305S, BLW614306S, BLW614310S, BLW614311S, BLW614313S, BLW614314S, BLW614338S, BLW614339S, BLW614340S, BLW614341S, BLW614356S, BLW614357S, BLW614364S, BLW614365S, BLW614368S, BLW614369S, BLW614371S, BLW614372S, BLW614375S, BLW614376S, BLW614377S, BLW614379S, BLW614380S, BLW614381S, BLW614382S, BLW614384S, BLW614386S, BLW614387S, BLW614388S, BLW614389S, BLW614390S, BLW614391S, BLW614392S, BLW614393S, BLW614395S, BLW614396S, BLW614397S, BLW614398S, BLW614399S, BLW614400S, BLW614403S, BLW614404S, BLW614406S, BLW614411S, BLW614416S, BLW614417S, BLW614418S, BLW614419S, BLW614420S, BLW614421S, BLW614424S, BLW614425S, BLW614426S, BLW614427S, BLW614429S, BLW614528S, BLW614593S, BLW614595S, BLW614596S, BLW614597S, BLW614598S, BLW614600S, BLW614602S, BLW614603S, BLW614731S, BLW614733S, BLW614734S, BLW614735S, BLW614736S, BLW614737S, BLW614738S, BLW614739S, BLW614740S, BLW614741S, BLW614743S, BLW614746S, BLW614775S, BLW614776S, BLW614777S, BLW614783S, BLW614784S, BLW614785S, BLW614786S, BLW614821S, BLW614823S, BLW614824S, BLW614849S, BLW614850S, BLW614851S, BLW614852S, BLW614877S, BLW614880S, BLW614881S, BLW614882S, BLW614883S, BLW614888S, BLW614889S, BLW614890S, BLW614891S, BLW614892S, BLW614893S, BLW614894S, BLW614895S, BLW614896S, BLW614897S, BLW614898S, BLW614899S, BLW614900S, BLW614903S, BLW614904S, BLW614905S, BLW614906S, BLW614907S, BLW614908S, BLW614909S, BLW614910S, BLW614911S, BLW614912S, BLW614913S, BLW614914S, BLW614915S, BLW614916S, BLW614917S, BLW614918S, BLW614919S, BLW614920S, BLW614921S, BLW614922S, BLW614923S, BLW614925S, BLW614926S, BLW614927S, BLW614928S, BLW614929S, BLW614930S, BLW614931S, BLW614932S, BLW614933S, BLW614934S, BLW614935S, BLW614955S, BLW614956S, BLW614957S, BLW614958S, BLW614965S, BLW614967S, BLW614968S, BLW614969S, BLW614993S, BLW614994S, BLW614995S, BLW614996S, BLW614997S, BLW615000S, BLW615002S, BLW615010S, BLW615027S, BLW615028S, BLW615037S, BLW615038S, BLW615039S, BLW615040S, BLW615056S, BLW615057S, BLW615058S, BLW615059S, BLW615060S, BLW615061S, BLW615062S, BLW615063S, BLW615064S, BLW615065S, BLW615067S, BLW615068S, BLW615069S, BLW615070S, BLW615072S, BLW615073S, BLW615074S, BLW615075S, BLW615076S, BLW615077S, BLW615078S, BLW615079S, BLW615080S, BLW615081S, BLW615085S, BLW615087S, BLW615091S, BLW615092S, BLW615093S, BLW615094S, BLW615095S, BLW615098S, BLW615099S, BLW615100S, BLW615101S, BLW615102S, BLW615103S, BLW615104S, BLW615105S, BLW615106S, BLW615107S, BLW615108S, BLW615110S, BLW615111S, BLW615112S, BLW615113S, BLW615114S, BLW615115S, BLW615116S, BLW615117S, BLW615118S, BLW615119S, BLW615120S, BLW615121S, BLW615122S, BLW615123S, BLW615124S, BLW615125S, BLW615126S, BLW615127S, BLW615128S, BLW615130S, BLW615131S, BLW615132S, BLW615133S, BLW615134S, BLW615135S, BLW615158S, BLW615166S, BLW615168S, BLW615170S, BLW615171S, BLW615172S, BLW615174S, BLW615175S, BLW615176S, BLW615179S, BLW615180S, BLW615181S, BLW615182S, BLW615183S, BLW615184S, BLW615185S, BLW615186S, BLW615187S, BLW615188S, BLW615189S, BLW615190S, BLW615191S, BLW615192S, BLW615193S, BLW615194S, BLW615408S, BLW615410S, BLW615413S, BLW615414S, BLW615417S, BLW615421S, BLW615423S, BLW615424S, BLW615425S, BLW615426S, BLW615427S, BLW615428S, BLW615432S, BLW615433S, BLW615434S, BLW615435S, BLW615436S, BLW615437S, BLW615440S, BLW615441S, BLW615442S, BLW615444S, BLW615448S, BLW615449S, BLW615450S, BLW615455S, BLW615456S, BLW615457S, BLW615458S, BLW615461S, BLW615462S, BLW615463S, BLW615464S, BLW615465S, BLW615466S, BLW615467S, BLW615468S, BLW615469S, BLW615470S, BLW615471S, BLW615483S, BLW615494S, BLW615530S, BLW615616S, BLW615841S, BLW615849S, BLW611874S, BLW612766S, BLW613205S, BLW608675S, BLW613752S, BLW614056S, BLW610770S, BLW610259S, BLW610511S, BLW610722S, BLW610723S, BLW611699S, BLW612911S, BLW613177S, BLW609985S, BLW609988S, BLW610157S, BLW610786S, BLW610789S, BLW613799S, BLW614443S, BLW614444S, BLW614526S, BLW614527S, BLW614529S, BLW610512S, BLW611819S, BLW613776S, BLW610491S, BLW610492S, BLW610493S, BLW610500S, BLW610726S, BLW610727S, BLW610777S, BLW610783S, BLW611436S, BLW611866S, BLW611867S, BLW611868S, BLW611871S, BLW611885S, BLW611887S, BLW612629S, BLW612630S, BLW613320S, BLW613327S, BLW613600S, BLW613622S, BLW613625S, BLW613780S, BLW613781S, BLW614469S, BLW614544S, BLW614601S, BLW614758S, BLW615299S, BLW615305S, BLW610724S, BLW612904S, BLW612441S, BLW612444S, BLW612463S, BLW612497S, BLW612807S, BLW612813S, BLW612814S, BLW614521S, BLW614522S, BLW614523S, BLW614524S, BLW614525S, BLW611155S, BLW611300S, BLW611711S, BLW611768S, BLW612581S, BLW612585S, BLW612616S, BLW612820S, BLW612113S, BLW612079S, BLW610767S, BLW611102S, BLW611103S, BLW611373S, BLW611995S, BLW612065S, BLW612072S, BLW612076S, BLW612078S, BLW612093S, BLW612094S, BLW612161S, BLW612210S, BLW612448S, BLW612449S, BLW612451S, BLW614445S, BLW614446S, BLW614447S, BLW614448S, BLW614449S, BLW613214S, BLW613238S, BLW613379S, BLW613405S, BLW613406S, BLW613419S, BLW613423S, BLW613426S, BLW613439S, BLW613446S, BLW613447S, BLW613451S, BLW613453S, BLW613465S, BLW613470S, BLW613472S, BLW613476S, BLW613479S, BLW613482S, BLW613494S, BLW613501S, BLW613785S, BLW613880S, BLW613881S, BLW613882S, BLW613883S, BLW613884S, BLW613885S, BLW613886S, BLW613887S, BLW613888S, BLW613889S, BLW613890S, BLW613891S, BLW613892S, BLW613893S, BLW613894S, BLW613895S, BLW613896S, BLW613897S, BLW613898S, BLW613899S, BLW613900S, BLW613901S, BLW613902S, BLW613903S, BLW613904S, BLW613919S, BLW613920S, BLW613921S, BLW613922S, BLW613923S, BLW613924S, BLW613926S, BLW613930S, BLW613931S, BLW613932S, BLW613933S, BLW613934S, BLW615210S, BLW615215S, BLW615216S, BLW615273S, BLW615281S, BLW615282S, BLW615283S, BLW615284S, BLW615285S, BLW615286S, BLW615287S, BLW615288S, BLW615289S, BLW615290S, BLW615291S, BLW615292S, BLW615344S, BLW615345S, BLW615346S, BLW615347S, BLW615348S, BLW615349S, BLW615350S, BLW615352S, BLW615353S, BLW615354S, BLW615355S, BLW615356S, BLW615357S, BLW615358S, BLW615359S, BLW615360S, BLW615361S, BLW615362S, BLW615363S, BLW615364S, BLW615365S, BLW615366S, BLW615367S, BLW615368S, BLW611160S, BLW611217S, BLW611321S, BLW611329S, BLW611359S, BLW610736S, BLW611100S, BLW611997S, BLW612013S, BLW612185S, BLW612593S, BLW613716S, BLW614473S, BLW612572S, BLW612575S, BLW612613S, BLW612632S, BLW612809S, BLW613102S, BLW613106S, BLW613222S, BLW613244S, BLW613655S, BLW613736S, BLW613753S, BLW613754S, BLW613786S, BLW613794S, BLW614304S, BLW611873S, BLW611905S, BLW612699S, BLW612781S, BLW612927S, BLW612928S, BLW613034S, BLW613285S, BLW613286S, BLW614106S, BLW614511S, BLW614512S, BLW614618S, BLW614619S, BLW614620S, BLW615138S, BLW612443S, BLW612453S, BLW612454S, BLW612524S, BLW612559S, BLW612859S, BLW613011S, BLW613056S, BLW613118S, BLW613122S, BLW613123S, BLW613248S, BLW613367S, BLW613529S, BLW613530S, BLW613532S, BLW613586S, BLW613587S, BLW613688S, BLW614129S, BLW614324S, BLW614350S, BLW614362S, BLW614804S, BLW614805S, BLW614806S, BLW615086S, BLW615570S, BLW615609S, BLW615626S, BLW615832S, BLW615838S, BLW615845S, BLW614472S, BLW614554S, BLW613797S, BLW613800S, BLW614545S, BLW613707S, BLW613739S, BLW615109S, BLW615221S, BLW615300S, BLW615303S, BLW614582S, BLW614583S, BLW614584S, BLW614680S, BLW615301S, BLW615480S, BLW615482S, BLW615488S, BLW615491S, BLW615492S, BLW615493S, BLW615495S, BLW615522S, BLW615523S, BLW615524S, BLW615525S, BLW615526S, BLW615529S, BLW615533S, BLW615535S, BLW615536S, BLW615537S, BLW615538S, BLW615539S, BLW615540S, BLW615541S, BLW615542S, BLW615544S, BLW615545S, BLW615546S, BLW615547S, BLW615548S, BLW615549S, BLW615550S, BLW615551S, BLW615552S, BLW615553S, BLW615554S, BLW615557S, BLW615558S, BLW615559S, BLW615560S, BLW615561S, BLW615575S, BLW615578S, BLW615579S, BLW615583S, BLW615592S, BLW615600S, BLW615601S, BLW615603S, BLW615605S, BLW615606S, BLW615617S, BLW615674S, BLW615675S, BLW615676S, BLW615677S, BLW615678S, BLW615679S, BLW615680S, BLW615681S, BLW615682S, BLW615683S, BLW615714S, BLW614756S, BLW614949S
Recalling Firm/
Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
8200 Coral Sea St Ne
Mounds View MN 55112-4391
For Additional Information Contact Technical Services
800-929-4043
Manufacturer Reason
for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA Determined
Cause 2
Under Investigation by firm
Action Medtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Quantity in Commerce 7921 units
Distribution US Nationwide - Worldwide Distribution
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA Database PMAs with Product Code = NIK and Original Applicant = MEDTRONIC CARDIAC RHYTHM DISEASE MANAGEMENT
PMAs with Product Code = NIK and Original Applicant = Medtronic, Inc.
-
-