• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall medtronic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall medtronicsee related information
Date Initiated by FirmMay 10, 2023
Date PostedJune 16, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1754-2023
Recall Event ID 92328
PMA NumberP010031S550 
Product Classification Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) - Product Code NIK
ProductCRTD DTMA1D4 CLARIA MRI US DF4, Model Number DTMA1D4; Implantable Cardioverter Defibrillators
Code Information GTIN 00643169720657, Lot Serial Numbers: RPC203306H, RPC205335H, RPC207546H, RPC210227H, RPC205865H, RPC205916H, RPC205937H, RPC206296H, RPC206416H, RPC206665H, RPC207309H, RPC207864H, RPC208393H, RPC209253H, RPC209691H, RPC209707H, RPC209745H, RPC209762H, RPC209847H, RPC209918H, RPC209843H, RPC203399H, RPC204295H, RPC202725H, RPC207338H, RPC204122H, RPC210083H, RPC207179H, RPC207493H, RPC207715H, RPC208839H, RPC210208H, RPC210209H, RPC210279H, RPC202502H, RPC206685H, RPC207835H, RPC202337H, RPC203913H, RPC206604H, RPC206605H, RPC206608H, RPC206609H, RPC207319H, RPC207608H, RPC209946H, RPC207476H, RPC208004H, RPC205236H, RPC207257H, RPC202491H, RPC202610H, RPC203628H, RPC203989H, RPC204238H, RPC204412H, RPC204523H, RPC205585H, RPC210162H, RPC202353H, RPC202434H, RPC202549H, RPC202947H, RPC203116H, RPC207575H, RPC202513H, RPC205862H, RPC207849H, RPC209388H, RPC210124H, RPC203780H, RPC204065H, RPC204066H, RPC210111H, RPC204504H, RPC203114H, RPC205516H, RPC205847H, RPC207025H, RPC207175H, RPC204556H, RPC207683H, RPC208161H, RPC208598H, RPC208896H, RPC209484H, RPC203280H, RPC203999H, RPC204866H, RPC204869H, RPC204871H, RPC204874H, RPC205657H, RPC205859H, RPC205860H, RPC206559H, RPC206713H, RPC207077H, RPC207577H, RPC208079H, RPC208366H, RPC209001H, RPC209018H, RPC209067H, RPC209386H, RPC202652H, RPC203629H, RPC203149H, RPC204932H, RPC206900H, RPC208153H, RPC203547H, RPC203646H, RPC203649H, RPC203650H, RPC203657H, RPC204651H, RPC204964H, RPC205201H, RPC205208H, RPC205243H, RPC205257H, RPC205324H, RPC205418H, RPC205450H, RPC205886H, RPC206253H, RPC206303H, RPC206363H, RPC206405H, RPC206570H, RPC207011H, RPC207013H, RPC207164H, RPC207189H, RPC207223H, RPC207299H, RPC207317H, RPC207734H, RPC207944H, RPC208173H, RPC208687H, RPC208727H, RPC209119H, RPC209560H, RPC209751H, RPC210185H, RPC203519H, RPC204384H, RPC204719H, RPC208146H, RPC205717H, RPC206158H, RPC207008H, RPC207445H, RPC207612H, RPC207680H, RPC207685H, RPC209617H, RPC209618H, RPC204693H, RPC205822H, RPC205823H, RPC210256H, RPC207238H, RPC202405H, RPC205892H, RPC204308H, RPC206420H, RPC206768H, RPC209390H, RPC210007H, RPC202487H, RPC202490H, RPC202768H, RPC203159H, RPC203590H, RPC203876H, RPC204890H, RPC204891H, RPC206020H, RPC206954H, RPC207621H, RPC207623H, RPC207663H, RPC208020H, RPC208030H, RPC208081H, RPC208089H, RPC208132H, RPC208690H, RPC208692H, RPC209610H, RPC209914H, RPC209986H, RPC202800H, RPC209058H, RPC205273H, RPC202436H, RPC203284H, RPC203422H, RPC204104H, RPC206143H, RPC208006H, RPC210219H, RPC208606H, RPC209145H, RPC203499H, RPC203578H, RPC207574H, RPC202903H, RPC203763H, RPC204545H, RPC204971H, RPC205198H, RPC206098H, RPC207921H, RPC203330H, RPC205224H, RPC206462H, RPC207857H, RPC208082H, RPC208210H, RPC208256H, RPC202424H, RPC202650H, RPC202779H, RPC204250H, RPC204713H, RPC209333H, RPC203501H, RPC205234H, RPC203051H, RPC207629H, RPC208340H, RPC208652H, RPC209003H, RPC203480H, RPC205968H, RPC208556H, RPC208730H, RPC208793H, RPC208804H, RPC209252H, RPC203474H, RPC205364H, RPC206256H, RPC203908H, RPC204751H, RPC208722H, RPC201367H, RPC202661H, RPC203001H, RPC203014H, RPC203095H, RPC203419H, RPC203706H, RPC203727H, RPC203960H, RPC204006H, RPC204149H, RPC204439H, RPC204476H, RPC204709H, RPC205138H, RPC205353H, RPC205445H, RPC205448H, RPC205498H, RPC205535H, RPC205620H, RPC206028H, RPC206063H, RPC206071H, RPC206216H, RPC206218H, RPC206283H, RPC206318H, RPC206334H, RPC206367H, RPC206394H, RPC206422H, RPC207301H, RPC207388H, RPC207566H, RPC207820H, RPC208497H, RPC209361H, RPC209380H, RPC209902H, RPC210112H, RPC203745H, RPC206526H, RPC209877H, RPC203122H, RPC203277H, RPC205025H, RPC205910H, RPC206478H, RPC206893H, RPC206898H, RPC206903H, RPC208272H, RPC209424H, RPC206558H, RPC209332H, RPC202705H, RPC203112H, RPC203840H, RPC203842H, RPC204000H, RPC204321H, RPC206194H, RPC207505H, RPC209195H, RPC209215H, RPC203030H, RPC202546H, RPC203652H, RPC202680H, RPC203569H, RPC204416H, RPC204435H, RPC205287H, RPC205292H, RPC205529H, RPC205727H, RPC206550H, RPC207564H, RPC207866H, RPC207928H, RPC208277H, RPC209343H, RPC210122H, RPC203291H, RPC203568H, RPC202715H, RPC202395H, RPC203401H, RPC203540H, RPC205466H, RPC207199H, RPC204151H, RPC206221H, RPC209578H, RPC209607H, RPC202469H, RPC202753H, RPC203327H, RPC203625H, RPC203857H, RPC203916H, RPC203922H, RPC203924H, RPC204831H, RPC204833H, RPC204834H, RPC206439H, RPC206947H, RPC206964H, RPC207779H, RPC207780H, RPC207781H, RPC207782H, RPC203598H, RPC205099H, RPC204551H, RPC205084H, RPC205165H, RPC203529H, RPC203693H, RPC203934H, RPC203936H, RPC206633H, RPC207022H, RPC207023H, RPC207084H, RPC209062H, RPC202607H, RPC202691H, RPC203028H, RPC203263H, RPC204605H, RPC205621H, RPC206436H, RPC208191H, RPC209431H, RPC210156H, RPC206668H, RPC207602H, RPC202386H, RPC202917H, RPC203125H, RPC203156H, RPC203321H, RPC203322H, RPC204146H, RPC204147H, RPC204975H, RPC204978H, RPC204979H, RPC204982H, RPC204989H, RPC204990H, RPC204991H, RPC204992H, RPC204995H, RPC206356H, RPC207688H, RPC207692H, RPC207693H, RPC207966H, RPC209976H, RPC204061H, RPC204064H, RPC204079H, RPC204904H, RPC205019H, RPC205039H, RPC205220H, RPC205223H, RPC205403H, RPC205474H, RPC205475H, RPC205979H, RPC205981H, RPC205984H, RPC205986H, RPC205990H, RPC208700H, RPC208782H, RPC208786H, RPC208788H, RPC208791H, RPC208794H, RPC208795H, RPC208801H, RPC210082H, RPC205454H, RPC205455H, RPC205457H, RPC208869H, RPC203226H, RPC204264H, RPC205347H, RPC206417H, RPC207828H, RPC202411H, RPC204327H, RPC205094H, RPC208155H, RPC209198H, RPC209490H, RPC205349H, RPC203101H, RPC203825H, RPC205333H, RPC205472H, RPC206628H, RPC207290H, RPC207560H, RPC207627H, RPC208179H, RPC208220H, RPC208536H, RPC210102H, RPC203182H, RPC208717H, RPC203970H, RPC206281H, RPC206909H, RPC208119H, RPC209703H, RPC210266H, RPC209977H, RPC202670H, RPC205076H, RPC207410H, RPC209611H, RPC202566H, RPC203047H, RPC203308H, RPC203539H, RPC203948H, RPC203951H, RPC204071H, RPC205090H, RPC205998H, RPC205999H, RPC206512H, RPC206520H, RPC206897H, RPC207712H, RPC207957H, RPC208409H, RPC209183H, RPC209347H, RPC209723H, RPC202964H, RPC202965H, RPC203883H, RPC203885H, RPC204926H, RPC205412H, RPC206039H, RPC206049H, RPC206058H, RPC206059H, RPC206061H, RPC206065H, RPC206066H, RPC208419H, RPC209434H, RPC204036H, RPC204592H, RPC209148H, RPC202333H, RPC202378H, RPC202527H, RPC202831H, RPC204117H, RPC204275H, RPC204813H, RPC205006H, RPC205299H, RPC206304H, RPC206335H, RPC206830H, RPC207870H, RPC208029H, RPC209039H, RPC209581H, RPC210221H, RPC210239H, RPC204738H, RPC204746H, RPC206217H, RPC203006H, RPC205599H, RPC206031H, RPC206284H, RPC206368H, RPC207125H, RPC207976H, RPC209054H, RPC209487H, RPC209997H, RPC207846H, RPC210166H, RPC210168H, RPC203588H, RPC203896H, RPC204814H, RPC204816H, RPC205254H, RPC207522H, RPC203462H, RPC208240H, RPC208826H, RPC209954H, RPC206916H, RPC203485H, RPC209444H, RPC207314H, RPC210206H, RPC203836H, RPC204382H, RPC204684H, RPC206459H, RPC206749H, RPC207241H, RPC207495H, RPC208806H, RPC209020H, RPC210282H, RPC202505H, RPC202510H, RPC202853H, RPC203354H, RPC204717H, RPC206142H, RPC207217H, RPC209008H, RPC202441H, RPC206393H, RPC208547H, RPC203341H, RPC203654H, RPC207153H, RPC207557H, RPC207894H, RPC208515H, RPC205228H, RPC203461H, RPC203944H, RPC204302H, RPC209072H, RPC204129H, RPC205438H, RPC206571H, RPC206575H, RPC207303H, RPC207595H, RPC208832H, RPC208549H, RPC206931H, RPC207161H, RPC208346H, RPC208141H, RPC202440H, RPC203215H, RPC203538H, RPC207063H, RPC209676H, RPC209783H, RPC203528H, RPC205514H, RPC206339H, RPC208203H, RPC208281H, RPC209339H, RPC209582H, RPC202366H, RPC202664H, RPC205205H, RPC205277H, RPC207880H, RPC209376H, RPC207720H, RPC203100H, RPC206556H, RPC204585H, RPC209015H, RPC207590H, RPC207965H, RPC203506H, RPC204136H, RPC204607H, RPC204634H, RPC205017H, RPC205699H, RPC207016H, RPC207837H, RPC208937H, RPC209170H, RPC209246H, RPC209508H, RPC209878H, RPC203516H, RPC207284H, RPC208397H, RPC202567H, RPC208045H, RPC205741H, RPC209412H, RPC207923H, RPC207992H, RPC204613H, RPC205963H, RPC208835H, RPC209614H, RPC209139H, RPC202474H, RPC202757H, RPC202927H, RPC203000H, RPC203076H, RPC203345H, RPC203370H, RPC203653H, RPC204428H, RPC205042H, RPC205375H, RPC205588H, RPC206269H, RPC206396H, RPC206573H, RPC206630H, RPC206758H, RPC206939H, RPC206990H, RPC206993H, RPC207035H, RPC207163H, RPC207230H, RPC207271H, RPC207384H, RPC207475H, RPC207484H, RPC207551H, RPC208037H, RPC208147H, RPC208217H, RPC208247H, RPC208609H, RPC208661H, RPC208921H, RPC209134H, RPC209451H, RPC209608H, RPC210049H, RPC210214H, RPC202354H, RPC202412H, RPC202508H, RPC202816H, RPC202834H, RPC202879H, RPC202958H, RPC202971H, RPC203208H, RPC203217H, RPC203475H, RPC203509H, RPC203644H, RPC203759H, RPC203892H, RPC203938H, RPC204067H, RPC204157H, RPC204297H, RPC204356H, RPC204586H, RPC205168H, RPC205400H, RPC205401H, RPC205492H, RPC205586H, RPC205658H, RPC205750H, RPC206344H, RPC206686H, RPC207449H, RPC207841H, RPC207893H, RPC207972H, RPC208080H, RPC208098H, RPC208314H, RPC208564H, RPC209094H, RPC209769H, RPC209798H, RPC210170H, RPC205826H, RPC207390H, RPC207142H, RPC207478H, RPC208311H, RPC203647H, RPC206359H, RPC207078H, RPC207396H, RPC208381H, RPC208947H, RPC209378H, RPC202428H, RPC202683H, RPC202726H, RPC203027H, RPC203220H, RPC203867H, RPC204096H, RPC204178H, RPC204208H, RPC204396H, RPC204397H, RPC205261H, RPC205262H, RPC205270H, RPC205509H, RPC205814H, RPC207024H, RPC207456H, RPC207616H, RPC207863H, RPC207884H, RPC208170H, RPC208662H, RPC208862H, RPC209755H, RPC210022H, RPC210237H, RPC202891H, RPC203576H, RPC204120H, RPC205635H, RPC202828H, RPC203564H, RPC205507H, RPC208333H, RPC209121H, RPC209435H, RPC208305H, RPC203404H, RPC203453H, RPC203856H, RPC203902H, RPC204525H, RPC205127H, RPC205154H, RPC205369H, RPC205897H, RPC205980H, RPC206408H, RPC210137H, RPC203017H, RPC203018H, RPC203019H, RPC203020H, RPC203021H, RPC203022H, RPC205488H, RPC205835H, RPC206960H, RPC206961H, RPC206968H, RPC206971H, RPC206972H, RPC206977H, RPC209834H, RPC210125H, RPC210127H, RPC210135H, RPC203858H, RPC203991H, RPC204893H, RPC208422H, RPC207517H, RPC209504H, RPC210265H, RPC202497H, RPC203352H, RPC203385H, RPC204392H, RPC209059H, RPC202370H, RPC202390H, RPC204475H, RPC204480H, RPC204631H, RPC205912H, RPC207783H, RPC207790H, RPC207853H, RPC203671H, RPC204240H, RPC204414H, RPC206474H, RPC209270H, RPC202619H, RPC203225H, RPC204443H, RPC204688H, RPC208571H, RPC204310H, RPC204351H, RPC205048H, RPC206938H, RPC202912H, RPC203339H, RPC203359H, RPC203582H, RPC206617H, RPC204190H, RPC204437H, RPC209381H, RPC206926H, RPC204828H, RPC204835H, RPC205976H, RPC207597H, RPC208504H, RPC202674H, RPC203070H, RPC203331H, RPC203742H, RPC205091H, RPC205162H, RPC205290H, RPC205444H, RPC205589H, RPC206450H, RPC206601H, RPC206824H, RPC207052H, RPC207100H, RPC207428H, RPC207662H, RPC208860H, RPC209178H, RPC209384H, RPC209897H, RPC203868H, RPC204106H, RPC205404H, RPC205842H, RPC207180H, RPC207558H, RPC208453H, RPC208489H, RPC203289H, RPC204502H, RPC204969H, RPC205572H, RPC209078H, RPC209573H, RPC203571H, RPC208946H, RPC206151H, RPC202408H, RPC203703H, RPC204220H, RPC206301H, RPC207465H, RPC205289H, RPC207176H, RPC207391H, RPC203760H, RPC207892H, RPC203664H, RPC204100H, RPC204119H, RPC204255H, RPC204565H, RPC206026H, RPC206899H, RPC207220H, RPC207645H, RPC203585H, RPC204590H, RPC204641H, RPC209470H, RPC210106H, RPC204224H, RPC204662H, RPC205081H, RPC205662H, RPC206527H, RPC207173H, RPC209415H, RPC209937H, RPC205943H, RPC209416H, RPC209967H, RPC208126H, RPC208271H, RPC203187H, RPC204030H, RPC205642H, RPC205795H, RPC206116H, RPC206595H, RPC207138H, RPC207582H, RPC207989H, RPC208936H, RPC209326H, RPC209407H, RPC209426H, RPC207379H, RPC207521H, RPC207830H, RPC208452H, RPC209526H, RPC209671H, RPC209960H, RPC210088H, RPC208443H, RPC209286H, RPC204508H, RPC205372H, RPC206975H, RPC208222H, RPC209095H, RPC210017H, RPC202689H, RPC206858H, RPC204424H, RPC202488H, RPC204207H, RPC205294H, RPC205380H, RPC206878H, RPC207098H, RPC207254H, RPC208585H, RPC209681H, RPC210183H, RPC204273H, RPC205248H, RPC207074H, RPC203817H, RPC204285H, RPC204967H, RPC206149H, RPC207946H, RPC208330H, RPC209082H, RPC202852H, RPC206088H, RPC209087H, RPC207541H, RPC209613H, RPC202771H, RPC203204H, RPC203606H, RPC204063H, RPC204206H, RPC204234H, RPC205382H, RPC205654H, RPC205799H, RPC205855H, RPC206463H, RPC206772H, RPC206773H, RPC207021H, RPC207255H, RPC207496H, RPC208136H, RPC208388H, RPC208587H, RPC208831H, RPC208911H, RPC209323H, RPC209379H, RPC209460H, RPC209552H, RPC209661H, RPC209732H, RPC209856H, RPC210020H, RPC204043H, RPC207784H, RPC203154H, RPC203250H, RPC203510H, RPC204496H, RPC205105H, RPC205122H, RPC205458H, RPC205462H, RPC205722H, RPC206843H, RPC207169H, RPC207203H, RPC207964H, RPC207997H, RPC208971H, RPC209117H, RPC209167H, RPC209708H, RPC210089H, RPC206598H, RPC209895H, RPC203271H, RPC203429H, RPC203533H, RPC204579H, RPC206200H, RPC206688H, RPC206717H, RPC206784H, RPC208379H, RPC209237H, RPC209862H, RPC202599H, RPC203492H, RPC203844H, RPC204847H, RPC205018H, RPC206551H, RPC207006H, RPC208242H, RPC208707H, RPC209366H, RPC209965H, RPC204425H, RPC210250H, RPC203886H, RPC205429H, RPC206849H, RPC208810H, RPC209820H, RPC208399H, RPC210086H, RPC203638H, RPC204543H, RPC205008H, RPC206164H, RPC207829H, RPC209606H, RPC202645H, RPC209832H, RPC202539H, RPC206322H, RPC207208H, RPC209228H, RPC202881H, RPC205283H, RPC208102H, RPC204222H, RPC205255H, RPC208268H, RPC208713H, RPC209530H, RPC209534H, RPC203691H, RPC207351H, RPC207881H, RPC210006H, RPC202387H, RPC203378H, RPC203561H, RPC204338H, RPC206419H, RPC207283H, RPC207927H, RPC208105H, RPC208386H, RPC209930H, RPC209970H, RPC210016H, RPC202399H, RPC206904H, RPC207128H, RPC209177H, RPC209990H, RPC203313H, RPC205078H, RPC209501H, RPC207039H, RPC207819H, RPC202880H, RPC209966H, RPC203394H, RPC204787H, RPC207183H, RPC208447H, RPC206319H, RPC208317H, RPC209656H, RPC203172H, RPC203520H, RPC204417H, RPC203398H, RPC204542H, RPC206574H, RPC207289H, RPC207520H, RPC208610H, RPC208763H, RPC210149H, RPC205898H, RPC207472H, RPC207567H, RPC208542H, RPC210152H, RPC210154H, RPC205013H, RPC205688H, RPC203143H, RPC203163H, RPC203870H, RPC204636H, RPC204755H, RPC204812H, RPC205041H, RPC206673H, RPC207010H, RPC207228H, RPC208073H, RPC208331H, RPC208350H, RPC208805H, RPC208950H, RPC209663H, RPC209664H, RPC210141H, RPC205543H, RPC205875H, RPC205771H, RPC202885H, RPC202844H, RPC203645H, RPC203900H, RPC204026H, RPC205857H, RPC205858H, RPC208777H, RPC209480H, RPC209509H, RPC209669H, RPC209739H, RPC209741H, RPC203435H, RPC204781H, RPC205064H, RPC207062H, RPC207311H, RPC207507H, RPC207670H, RPC207673H, RPC208844H, RPC207005H, RPC202444H, RPC205071H, RPC205746H, RPC207215H, RPC209482H, RPC205759H, RPC207258H, RPC208931H, RPC209457H, RPC203242H, RPC209045H, RPC206125H, RPC206307H, RPC206709H, RPC207019H, RPC208007H, RPC203732H, RPC207273H, RPC208349H, RPC202475H, RPC204103H, RPC206627H, RPC207745H, RPC207948H, RPC208600H, RPC210261H, RPC202672H, RPC203651H, RPC204889H, RPC208863H, RPC209091H, RPC209989H, RPC210095H, RPC202568H, RPC208201H, RPC209395H, RPC209824H, RPC204503H, RPC208112H, RPC205844H, RPC207122H, RPC207374H, RPC207831H, RPC208520H, RPC209165H, RPC210129H, RPC202618H, RPC203395H, RPC203819H, RPC203860H, RPC204562H, RPC207945H, RPC208495H, RPC209425H, RPC209545H, RPC207339H, RPC210145H, RPC204660H, RPC204888H, RPC203425H, RPC204446H, RPC207020H, RPC202477H, RPC204614H, RPC207951H, RPC203243H, RPC203296H, RPC203517H, RPC203537H, RPC203583H, RPC204914H, RPC205379H, RPC205470H, RPC206309H, RPC207028H, RPC208206H, RPC209234H, RPC210160H, RPC202723H, RPC203058H, RPC203072H, RPC203266H, RPC203535H, RPC203548H, RPC203574H, RPC203595H, RPC203596H, RPC203667H, RPC203694H, RPC203724H, RPC203861H, RPC203914H, RPC204228H, RPC204260H, RPC204278H, RPC204448H, RPC204499H, RPC204710H, RPC204793H, RPC205928H, RPC206168H, RPC206317H, RPC206433H, RPC206460H, RPC206711H, RPC206860H, RPC206917H, RPC207009H, RPC207037H, RPC207046H, RPC207123H, RPC207150H, RPC207166H, RPC207297H, RPC207308H, RPC207318H, RPC207376H, RPC207461H, RPC207659H, RPC207695H, RPC207855H, RPC207983H, RPC207985H, RPC208063H, RPC208097H, RPC208099H, RPC208109H, RPC208172H, RPC208193H, RPC208259H, RPC208280H, RPC208306H, RPC208355H, RPC208625H, RPC208629H, RPC208710H, RPC208882H, RPC209016H, RPC209036H, RPC209079H, RPC209222H, RPC209265H, RPC209293H, RPC209316H, RPC209331H, RPC209464H, RPC209675H, RPC209764H, RPC209863H, RPC210046H, RPC210186H, RPC210228H, RPC210251H, RPC204913H, RPC204759H, RPC206820H, RPC204597H, RPC205456H, RPC205818H, RPC206083H, RPC206382H, RPC208463H, RPC208468H, RPC208599H, RPC208790H, RPC208842H, RPC209311H, RPC210084H, RPC205625H, RPC206641H, RPC207664H, RPC204387H, RPC206432H, RPC209891H, RPC204137H, RPC206349H, RPC208634H, RPC209106H, RPC202742H, RPC208266H, RPC209489H, RPC202458H, RPC202461H, RPC202830H, RPC204469H, RPC204700H, RPC206308H, RPC206323H, RPC206992H, RPC207236H, RPC208886H, RPC209284H, RPC209440H, RPC209694H, RPC209949H, RPC210119H, RPC210144H, RPC203450H, RPC207825H, RPC207840H, RPC209397H, RPC207014H, RPC203396H, RPC204059H, RPC204060H, RPC206791H, RPC206794H, RPC207253H, RPC207821H, RPC208766H, RPC209550H, RPC209556H, RPC203434H, RPC204131H, RPC202642H, RPC202877H, RPC202936H, RPC203234H, RPC204357H, RPC204367H, RPC204368H, RPC204394H, RPC204800H, RPC205300H, RPC205496H, RPC205504H, RPC205756H, RPC206064H, RPC206279H, RPC206379H, RPC206607H, RPC206624H, RPC207113H, RPC207365H, RPC207381H, RPC207386H, RPC207971H, RPC208764H, RPC208955H, RPC209521H, RPC205360H, RPC208994H, RPC202627H, RPC203954H, RPC204506H, RPC206544H, RPC206689H, RPC208773H, RPC203278H, RPC203403H, RPC205336H, RPC208433H, RPC202359H, RPC203534H, RPC207509H, RPC209174H, RPC204279H, RPC207101H, RPC207424H, RPC207528H, RPC209027H, RPC209619H, RPC202980H, RPC203211H, RPC204980H, RPC206249H, RPC209602H, RPC204438H, RPC202357H, RPC203415H, RPC203428H, RPC203680H, RPC205558H, RPC209255H, RPC209377H, RPC205130H, RPC206112H, RPC207583H, RPC202476H, RPC203407H, RPC204153H, RPC204704H, RPC206842H, RPC207500H, RPC207519H, RPC208974H, RPC210150H, RPC202933H, RPC203975H, RPC204685H, RPC205302H, RPC205441H, RPC205974H, RPC206086H, RPC206310H, RPC206321H, RPC206599H, RPC207462H, RPC207463H, RPC207488H, RPC202814H, RPC204601H, RPC205124H, RPC205303H, RPC207485H, RPC208290H, RPC208369H, RPC206166H, RPC210274H, RPC205075H, RPC206470H, RPC209890H, RPC206554H, RPC209447H, RPC202738H, RPC203440H, RPC205308H, RPC205315H, RPC206229H, RPC206517H, RPC202493H, RPC205812H, RPC202944H, RPC204247H, RPC206725H, RPC207443H, RPC207839H, RPC202909H, RPC203452H, RPC209076H, RPC203224H, RPC202376H, RPC202668H, RPC208408H, RPC208617H, RPC208688H, RPC209491H, RPC209481H, RPC209873H, RPC210060H, RPC209144H, RPC208859H, RPC209197H, RPC209524H, RPC208262H, RPC208966H, RPC203282H, RPC203587H, RPC204215H, RPC204269H, RPC210195H, RPC204699H, RPC202942H, RPC203635H, RPC203443H, RPC204332H, RPC206295H, RPC207115H, RPC207474H, RPC208494H, RPC208528H, RPC208824H, RPC208996H, RPC209209H, RPC209310H, RPC207581H, RPC207588H, RPC208297H, RPC209218H, RPC210087H, RPC202471H, RPC203024H, RPC203103H, RPC203157H, RPC203672H, RPC204197H, RPC204413H, RPC204689H, RPC204754H, RPC204776H, RPC204777H, RPC205377H, RPC205624H, RPC206053H, RPC206093H, RPC206442H, RPC206457H, RPC206625H, RPC206714H, RPC206807H, RPC206859H, RPC206884H, RPC207015H, RPC207129H, RPC207160H, RPC207211H, RPC207234H, RPC207264H, RPC207554H, RPC207984H, RPC208121H, RPC208199H, RPC208284H, RPC208309H, RPC208332H, RPC208352H, RPC208630H, RPC208668H, RPC208726H, RPC209060H, RPC209104H, RPC209169H, RPC209180H, RPC210240H, RPC205020H, RPC203307H, RPC203930H, RPC204680H, RPC207770H, RPC207882H, RPC206817H, RPC205112H, RPC206406H, RPC208194H, RPC210199H, RPC205485H, RPC205486H, RPC204040H, RPC204623H, RPC203375H, RPC204373H, RPC206548H, RPC210009H, RPC206224H, RPC210164H, RPC210231H, RPC204897H, RPC205414H, RPC206631H, RPC206635H, RPC205134H, RPC206274H, RPC208461H, RPC207165H, RPC203993H, RPC204218H, RPC204291H, RPC204977H, RPC204988H, RPC205129H, RPC205663H, RPC207655H, RPC208301H, RPC208619H, RPC209476H, RPC202609H, RPC208808H, RPC202684H, RPC204853H, RPC205216H, RPC206163H, RPC210213H, RPC203521H, RPC203527H, RPC204788H, RPC205405H, RPC207836H, RPC208926H, RPC209143H, RPC209393H, RPC210172H, RPC204923H, RPC207615H, RPC209758H, RPC206618H, RPC203046H, RPC203626H, RPC204239H, RPC204548H, RPC205093H, RPC206157H, RPC207349H, RPC209074H, RPC209088H, RPC209315H, RPC209367H, RPC202688H, RPC209936H, RPC209837H, RPC202517H, RPC202590H, RPC203392H, RPC203436H, RPC205421H, RPC206910H, RPC207334H, RPC207513H, RPC202384H, RPC203342H, RPC208865H, RPC202946H, RPC203879H, RPC205374H, RPC205390H, RPC205988H, RPC206004H, RPC206347H, RPC207053H, RPC207093H, RPC207914H, RPC208055H, RPC208531H, RPC208780H, RPC209463H, RPC209506H, RPC203056H, RPC204132H, RPC202511H, RPC202681H, RPC202923H, RPC203048H, RPC203446H, RPC203730H, RPC203765H, RPC203862H, RPC204005H, RPC204345H, RPC204864H, RPC204999H, RPC205003H, RPC205012H, RPC205815H, RPC206254H, RPC206298H, RPC206614H, RPC206615H, RPC207105H, RPC207371H, RPC207401H, RPC207725H, RPC207727H, RPC207728H, RPC208834H, RPC209753H, RPC210193H, RPC207990H, RPC202646H, RPC203129H, RPC205242H, RPC206160H, RPC206561H, RPC206698H, RPC207272H, RPC207733H, RPC208005H, RPC203139H, RPC203140H, RPC208072H, RPC209111H, RPC204881H, RPC204882H, RPC208481H, RPC204226H, RPC204333H, RPC206370H, RPC206623H, RPC207354H, RPC207915H, RPC202330H, RPC204549H, RPC210217H, RPC203275H, RPC203818H, RPC204602H, RPC205411H, RPC205714H, RPC206266H, RPC206276H, RPC208899H, RPC203466H, RPC203673H, RPC204632H, RPC205176H, RPC206374H, RPC206844H, RPC208138H, RPC208215H, RPC208263H, RPC209191H, RPC209821H, RPC205113H, RPC205152H, RPC205420H, RPC208124H, RPC208870H, RPC209041H, RPC202720H, RPC202901H, RPC203002H, RPC203113H, RPC203244H, RPC203320H, RPC203524H, RPC203526H, RPC203959H, RPC204326H, RPC204403H, RPC204792H, RPC205101H, RPC206543H, RPC210038H, RPC202961H, RPC203402H, RPC204360H, RPC204679H, RPC205473H, RPC206092H, RPC209916H, RPC202655H, RPC202829H, RPC204742H, RPC204918H, RPC205177H, RPC206203H, RPC208370H, RPC208622H, RPC209034H, RPC209116H, RPC210151H, RPC206958H, RPC205820H, RPC205821H, RPC206657H, RPC207625H, RPC207630H, RPC207631H, RPC207632H, RPC207634H, RPC205318H, RPC209338H, RPC204198H, RPC203800H, RPC203801H, RPC207420H, RPC208058H, RPC209808H, RPC210077H, RPC203349H, RPC208807H, RPC209131H, RPC203045H, RPC204406H, RPC204785H, RPC208197H, RPC203483H, RPC205490H, RPC205582H, RPC206541H, RPC202908H, RPC203295H, RPC203426H, RPC206896H, RPC207066H, RPC207275H, RPC207929H, RPC209040H, RPC202730H, RPC202918H, RPC203319H, RPC206139H, RPC207363H, RPC208118H, RPC209314H, RPC209372H, RPC209373H, RPC209978H, RPC202693H, RPC202776H, RPC203640H, RPC203735H, RPC205329H, RPC205685H, RPC205850H, RPC202454H, RPC202520H, RPC202732H, RPC202794H, RPC202795H, RPC203005H, RPC203171H, RPC203421H, RPC204018H, RPC204022H, RPC204827H, RPC204986H, RPC205092H, RPC205098H, RPC205111H, RPC205370H, RPC205574H, RPC205659H, RPC205715H, RPC206197H, RPC206205H, RPC206285H, RPC206476H, RPC206581H, RPC206774H, RPC206862H, RPC206865H, RPC207212H, RPC207245H, RPC208042H, RPC208056H, RPC208295H, RPC208326H, RPC208425H, RPC208559H, RPC208654H, RPC208959H, RPC209080H, RPC209572H, RPC209909H, RPC209941H, RPC202676H, RPC203110H, RPC203180H, RPC203767H, RPC204068H, RPC204140H, RPC204143H, RPC204187H, RPC204449H, RPC204513H, RPC204954H, RPC205413H, RPC205432H, RPC205653H, RPC205788H, RPC208046H, RPC208456H, RPC208488H, RPC209073H, RPC209510H, RPC209559H, RPC209948H, RPC202692H, RPC203484H, RPC204696H, RPC204799H, RPC206189H, RPC207911H, RPC208614H, RPC208873H, RPC210019H, RPC210068H, RPC204970H, RPC206678H, RPC206681H, RPC206841H, RPC207088H, RPC208145H, RPC208395H, RPC208448H, RPC208470H, RPC209108H, RPC209244H, RPC209568H, RPC210052H, RPC210223H, RPC202840H, RPC203365H, RPC205366H, RPC206246H, RPC209225H, RPC210287H, RPC204254H, RPC206891H, RPC202686H, RPC206653H, RPC206745H, RPC207073H, RPC208200H, RPC208586H, RPC205153H, RPC206969H, RPC208021H, RPC207506H, RPC203562H, RPC206211H, RPC208131H, RPC209477H, RPC209360H, RPC206514H, RPC203887H, RPC204340H, RPC204362H, RPC207978H, RPC208496H, RPC209069H, RPC209337H, RPC209423H, RPC202417H, RPC205355H, RPC206077H, RPC207369H, RPC207438H, RPC207464H, RPC207705H, RPC207707H, RPC207711H, RPC207879H, RPC208672H, RPC208877H, RPC208924H, RPC209030H, RPC210236H, RPC205617H, RPC207291H, RPC203823H, RPC203851H, RPC204930H, RPC205339H, RPC205550H, RPC206327H, RPC206870H, RPC207479H, RPC207795H, RPC207940H, RPC208234H, RPC208427H, RPC208720H, RPC209408H, RPC209896H, RPC203207H, RPC206449H, RPC206456H, RPC209450H, RPC209828H, RPC202414H, RPC203523H, RPC206427H, RPC204118H, RPC204370H, RPC204658H, RPC208428H, RPC208477H, RPC208573H, RPC208977H, RPC203400H, RPC205626H, RPC205911H, RPC206207H, RPC208357H, RPC203198H, RPC203873H, RPC204352H, RPC204943H, RPC205351H, RPC206144H, RPC209442H, RPC209962H, RPC209943H, RPC204311H, RPC204894H, RPC204939H, RPC205672H, RPC206137H, RPC207174H, RPC207409H, RPC207660H, RPC208466H, RPC208491H, RPC208995H, RPC209210H, RPC209212H, RPC209394H, RPC209391H, RPC209680H, RPC203698H, RPC204243H, RPC204244H, RPC206913H, RPC208976H, RPC209304H, RPC202540H, RPC208025H, RPC208880H, RPC203756H, RPC206819H, RPC207677H, RPC205614H, RPC205879H, RPC206195H, RPC206261H, RPC208018H, RPC209171H, RPC209207H, RPC209264H, RPC209289H, RPC209305H, RPC209403H, RPC202382H, RPC203055H, RPC203061H, RPC203063H, RPC204578H, RPC207534H, RPC207753H, RPC210163H, RPC202823H, RPC203866H, RPC203942H, RPC204731H, RPC204885H, RPC205265H, RPC206191H, RPC206385H, RPC208291H, RPC208811H, RPC208814H, RPC208815H, RPC210138H, RPC203725H, RPC208094H, RPC202437H, RPC202717H, RPC203015H, RPC203273H, RPC203559H, RPC203998H, RPC204098H, RPC204115H, RPC205622H, RPC206336H, RPC206732H, RPC207991H, RPC208925H, RPC209028H, RPC209070H, RPC209525H, RPC209744H, RPC209867H, RPC202914H, RPC209259H, RPC202847H, RPC203168H, RPC205446H, RPC205495H, RPC206399H, RPC207064H, RPC207785H, RPC208083H, RPC209549H, RPC210040H, RPC206485H, RPC207961H, RPC203102H, RPC203828H, RPC203963H, RPC205087H, RPC205760H, RPC205770H, RPC206090H, RPC206300H, RPC206924H, RPC207382H, RPC207708H, RPC208523H, RPC209827H, RPC209841H, RPC209885H, RPC206187H, RPC208405H, RPC205309H, RPC206136H, RPC206583H, RPC209647H, RPC204485H, RPC208903H, RPC202592H, RPC203148H, RPC203431H, RPC203432H, RPC203550H, RPC203758H, RPC204101H, RPC204102H, RPC204114H, RPC204389H, RPC204391H, RPC205282H, RPC205314H, RPC205419H, RPC206082H, RPC206839H, RPC206846H, RPC206848H, RPC206850H, RPC208525H, RPC208633H, RPC208944H, RPC209021H, RPC209115H, RPC210155H, RPC202930H, RPC204553H, RPC207213H, RPC209968H, RPC206723H, RPC208473H, RPC209375H, RPC206080H, RPC207703H, RPC207713H, RPC210035H, RPC202928H, RPC203777H, RPC206741H, RPC207471H, RPC207699H, RPC208139H, RPC208554H, RPC208735H, RPC209236H, RPC209132H, RPC202643H, RPC202810H, RPC205484H, RPC207366H, RPC208167H, RPC209517H, RPC209576H, RPC210281H, RPC209456H, RPC207397H, RPC210043H, RPC203053H, RPC203054H, RPC203057H, RPC203059H, RPC203060H, RPC203064H, RPC207182H, RPC207185H, RPC208853H, RPC209951H, RPC204121H, RPC208095H, RPC209420H, RPC209848H, RPC207340H, RPC202331H, RPC202862H, RPC204701H, RPC205560H, RPC205832H, RPC206952H, RPC207639H, RPC209136H, RPC209998H, RPC210004H, RPC210100H, RPC202666H, RPC202792H, RPC207288H, RPC207436H, RPC208505H, RPC209097H, RPC202430H, RPC202769H, RPC202770H, RPC203236H, RPC203361H, RPC204007H, RPC204192H, RPC204386H, RPC204786H, RPC205295H, RPC205348H, RPC205740H, RPC205816H, RPC206441H, RPC206467H, RPC206500H, RPC206580H, RPC207418H, RPC208214H, RPC208647H, RPC209068H, RPC209458H, RPC209654H, RPC209731H, RPC209817H, RPC209919H, RPC210003H, RPC210033H, RPC202605H, RPC204790H, RPC205184H, RPC206042H, RPC206180H, RPC206426H, RPC206528H, RPC206645H, RPC206650H, RPC206692H, RPC206693H, RPC207029H, RPC207336H, RPC207691H, RPC207888H, RPC208475H, RPC208951H, RPC209013H, RPC209064H, RPC209262H, RPC209274H, RPC209507H, RPC209774H, RPC210196H, RPC203123H, RPC203965H, RPC204588H, RPC205358H, RPC205359H, RPC205361H, RPC205362H, RPC205387H, RPC207738H, RPC208759H, RPC209206H, RPC209574H, RPC209577H, RPC209579H, RPC209589H, RPC209822H, RPC209851H, RPC210098H, RPC203205H, RPC203441H, RPC203457H, RPC204225H, RPC206482H, RPC206740H, RPC206991H, RPC207789H, RPC208236H, RPC208872H, RPC208986H, RPC209285H, RPC209492H, RPC209872H, RPC204180H, RPC204196H, RPC205204H, RPC207417H, RPC207480H, RPC208033H, RPC208374H, RPC209093H, RPC209099H, RPC209864H, RPC210258H, RPC210285H, RPC202416H, RPC207231H, RPC208451H, RPC207562H, RPC206222H, RPC207890H, RPC209309H, RPC209844H, RPC209931H, RPC203852H, RPC208327H, RPC208830H, RPC209086H, RPC202334H, RPC204175H, RPC204861H, RPC205518H, RPC209514H, RPC209756H, RPC202518H, RPC202906H, RPC204606H, RPC205079H, RPC205651H, RPC206126H, RPC206181H, RPC206337H, RPC206341H, RPC206377H, RPC206634H, RPC206857H, RPC208066H, RPC208411H, RPC209235H, RPC209277H, RPC202385H, RPC208897H, RPC208898H, RPC208905H, RPC209567H, RPC204052H, RPC204268H, RPC204427H, RPC205015H, RPC205325H, RPC205721H, RPC206134H, RPC207168H, RPC207514H, RPC207872H, RPC207987H, RPC209419H, RPC209929H, RPC205511H, RPC206621H, RPC206955H, RPC206996H, RPC207118H, RPC210283H, RPC203744H, RPC208653H, RPC207267H, RPC203909H, RPC205266H, RPC206378H, RPC206511H, RPC207368H, RPC207591H, RPC208400H, RPC209049H, RPC207360H, RPC202883H, RPC206759H, RPC206760H, RPC206762H, RPC206763H, RPC206766H, RPC206767H, RPC206771H, RPC206780H, RPC206783H, RPC206785H, RPC207102H, RPC207756H, RPC208064H, RPC208415H, RPC208416H, RPC208430H, RPC208432H, RPC208437H, RPC208575H, RPC208658H, RPC208665H, RPC208666H, RPC208669H, RPC208674H, RPC208682H, RPC208683H, RPC208684H, RPC208685H, RPC208789H, RPC209512H, RPC209541H, RPC209543H, RPC209547H, RPC209548H, RPC210013H, RPC202401H, RPC203174H, RPC203449H, RPC203543H, RPC203752H, RPC203815H, RPC204139H, RPC204420H, RPC204569H, RPC204609H, RPC204720H, RPC206096H, RPC206373H, RPC207302H, RPC207328H, RPC202598H, RPC202698H, RPC204133H, RPC204298H, RPC204595H, RPC205007H, RPC205031H, RPC205160H, RPC206369H, RPC206612H, RPC203270H, RPC203290H, RPC204804H, RPC205037H, RPC205597H, RPC205719H, RPC205938H, RPC206000H, RPC206007H, RPC206045H, RPC206046H, RPC206488H, RPC209023H, RPC209725H, RPC204084H, RPC204580H, RPC206654H, RPC206885H, RPC207141H, RPC208923H, RPC208973H, RPC208991H, RPC209261H, RPC210024H, RPC204931H, RPC204933H, RPC207269H, RPC208876H, RPC203188H, RPC203895H, RPC207412H, RPC210133H, RPC202969H, RPC203309H, RPC203911H, RPC204012H, RPC204307H, RPC206593H, RPC207372H, RPC208613H, RPC209924H, RPC202521H, RPC202656H, RPC202858H, RPC202863H, RPC203195H, RPC203690H, RPC203736H, RPC203808H, RPC204203H, RPC204741H, RPC204892H, RPC205131H, RPC205271H, RPC205393H, RPC205961H, RPC206132H, RPC206452H, RPC207087H, RPC207455H, RPC208364H, RPC208506H, RPC209107H, RPC209637H, RPC202783H, RPC204795H, RPC206431H, RPC202636H, RPC210153H, RPC209926H, RPC210048H, RPC203709H, RPC203912H, RPC204165H, RPC205186H, RPC210042H, RPC202328H, RPC202432H, RPC202905H, RPC203448H, RPC203491H, RPC204522H, RPC204748H, RPC204873H, RPC205114H, RPC205810H, RPC206696H, RPC206737H, RPC208027H, RPC208130H, RPC208424H, RPC210205H, RPC202744H, RPC203085H, RPC203097H, RPC203098H, RPC204764H, RPC204767H, RPC204768H, RPC205088H, RPC206815H, RPC206829H, RPC207076H, RPC207686H, RPC208741H, RPC208744H, RPC203107H, RPC203108H, RPC203367H, RPC203754H, RPC203850H, RPC204183H, RPC204706H, RPC204997H, RPC205755H, RPC206047H, RPC206050H, RPC206130H, RPC206855H, RPC206962H, RPC207067H, RPC207696H, RPC207751H, RPC207752H, RPC208521H, RPC210177H, RPC203253H, RPC207027H, RPC208796H, RPC208884H, RPC209221H, RPC202938H, RPC203405H, RPC204054H, RPC204055H, RPC204056H, RPC204058H, RPC206881H, RPC207431H, RPC207802H, RPC207803H, RPC207804H, RPC207805H, RPC208127H, RPC209004H, RPC209720H, RPC209724H, RPC209726H, RPC207643H, RPC208486H, RPC209318H, RPC209523H, RPC207672H, RPC208382H, RPC202611H, RPC203158H, RPC203502H, RPC206364H, RPC206889H, RPC202876H, RPC202999H, RPC204664H, RPC205422H, RPC205505H, RPC205991H, RPC206666H, RPC209538H, RPC209907H, RPC202976H, RPC203042H, RPC203043H, RPC203356H, RPC204276H, RPC206147H, RPC208049H, RPC208601H, RPC209051H, RPC210271H, RPC203929H, RPC205605H, RPC206821H, RPC206838H, RPC206209H, RPC204802H, RPC207871H, RPC203144H, RPC203573H, RPC203008H, RPC203832H, RPC205307H, RPC206235H, RPC207901H, RPC209638H, RPC203423H, RPC202394H, RPC204410H, RPC205096H, RPC205431H, RPC206586H, RPC206987H, RPC206314H, RPC206620H, RPC206908H, RPC204654H, RPC208140H, RPC208375H, RPC209410H, RPC202772H, RPC203310H, RPC203433H, RPC204471H, RPC205174H, RPC206477H, RPC207676H, RPC203658H, RPC204536H, RPC205789H, RPC207143H, RPC208091H, RPC208454H, RPC210014H, RPC203257H, RPC206578H, RPC203267H, RPC204878H, RPC205824H, RPC206270H, RPC206518H, RPC209714H, RPC209846H, RPC209850H, RPC210031H, RPC206532H, RPC207151H, RPC202626H, RPC202818H, RPC203487H, RPC203669H, RPC204105H, RPC204479H, RPC204794H, RPC205021H, RPC205562H, RPC206343H, RPC207749H, RPC207877H, RPC209738H, RPC203007H, RPC204477H, RPC202415H, RPC204494H, RPC204594H, RPC205022H, RPC205191H, RPC206140H, RPC206292H, RPC202438H, RPC202937H, RPC202940H, RPC203845H, RPC203846H, RPC204775H, RPC208245H, RPC208351H, RPC208913H, RPC203560H, RPC206230H, RPC207003H, RPC209127H, RPC209925H, RPC203153H, RPC206423H, RPC207529H, RPC207373H, RPC209288H, RPC202364H, RPC202507H, RPC202582H, RPC203439H, RPC204045H, RPC204046H, RPC204048H, RPC205368H, RPC205813H, RPC207675H, RPC207682H, RPC207690H, RPC207694H, RPC207700H, RPC207717H, RPC208972H, RPC204727H, RPC207225H, RPC206487H, RPC208465H, RPC203106H, RPC204763H, RPC205551H, RPC203127H, RPC204694H, RPC208592H, RPC209679H, RPC209801H, RPC209974H, RPC203377H, RPC203381H, RPC203382H, RPC203383H, RPC203384H, RPC203388H, RPC203389H, RPC203391H, RPC204570H, RPC207499H, RPC207786H, RPC207787H, RPC207788H, RPC207791H, RPC207792H, RPC207794H, RPC207796H, RPC207797H, RPC207798H, RPC207799H, RPC207800H, RPC206019H, RPC209000H, RPC209203H, RPC209736H, RPC209740H, RPC203067H, RPC202335H, RPC207681H, RPC209623H, RPC208218H, RPC203161H, RPC205166H, RPC205397H, RPC206727H, RPC208115H, RPC208404H, RPC206179H, RPC208739H, RPC209421H, RPC209818H, RPC202363H, RPC202855H, RPC203206H, RPC207584H, RPC203248H, RPC204577H, RPC206708H, RPC207170H, RPC207285H, RPC207407H, RPC207563H, RPC207568H, RPC209401H, RPC209427H, RPC210002H, RPC202413H, RPC202984H, RPC203292H, RPC203414H, RPC204145H, RPC204381H, RPC204721H, RPC205313H, RPC206484H, RPC207968H, RPC208529H, RPC208917H, RPC206935H, RPC206989H, RPC207530H, RPC208156H, RPC202379H, RPC203033H, RPC203036H, RPC203038H, RPC203040H, RPC203761H, RPC204123H, RPC204134H, RPC204167H, RPC204317H, RPC204349H, RPC204376H, RPC204452H, RPC204667H, RPC204783H, RPC205836H, RPC205838H, RPC206656H, RPC206659H, RPC208000H, RPC208689H, RPC208724H, RPC209307H, RPC209312H, RPC209493H, RPC209775H, RPC209845H, RPC203967H, RPC205063H, RPC206204H, RPC204257H, RPC204490H, RPC204603H, RPC206245H, RPC206690H, RPC207216H, RPC207826H, RPC208182H, RPC208282H, RPC208648H, RPC209700H, RPC201366H, RPC202624H, RPC203302H, RPC204088H, RPC205506H, RPC207524H, RPC202819H, RPC203222H, RPC204457H, RPC204735H, RPC206440H, RPC207055H, RPC207331H, RPC207370H, RPC207468H, RPC208353H, RPC209804H, RPC202817H, RPC203500H, RPC203919H, RPC203973H, RPC204728H, RPC205245H, RPC206579H, RPC207408H, RPC207832H, RPC207833H, RPC209233H, RPC202660H, RPC203062H, RPC203393H, RPC203663H, RPC203741H, RPC203766H, RPC204456H, RPC204829H, RPC206701H, RPC207144H, RPC208137H, RPC203584H, RPC204857H, RPC208129H, RPC209564H, RPC204323H, RPC204458H, RPC205060H, RPC205430H, RPC206190H, RPC207450H, RPC207771H, RPC210012H, RPC210029H, RPC202404H, RPC203643H, RPC203928H, RPC204042H, RPC204921H, RPC205553H, RPC206445H, RPC206888H, RPC207079H, RPC207652H, RPC208657H, RPC209303H, RPC206212H, RPC205785H, RPC202763H, RPC202658H, RPC204626H, RPC208664H, RPC207626H, RPC209224H, RPC203737H, RPC206444H, RPC202558H, RPC203213H, RPC204756H, RPC206153H, RPC206742H, RPC209231H, RPC210182H, RPC210290H, RPC202974H, RPC209778H, RPC202342H, RPC202960H, RPC203044H, RPC203091H, RPC203946H, RPC204141H, RPC204261H, RPC204659H, RPC205043H, RPC205296H, RPC205350H, RPC206471H, RPC206540H, RPC206963H, RPC206984H, RPC207218H, RPC207249H, RPC207709H, RPC207850H, RPC207913H, RPC208069H, RPC209319H, RPC209465H, RPC210081H, RPC204087H, RPC206324H, RPC202538H, RPC203086H, RPC203261H, RPC203968H, RPC207460H, RPC207653H, RPC208190H, RPC208635H, RPC209583H, RPC210245H, RPC202764H, RPC203749H, RPC203904H, RPC207842H, RPC209643H, RPC202537H, RPC203726H, RPC206297H, RPC206496H, RPC207467H, RPC208185H, RPC208294H, RPC208550H, RPC208762H, RPC209411H, RPC209433H, RPC209698H, RPC202848H, RPC202849H, RPC205386H, RPC205649H, RPC206861H, RPC206869H, RPC208538H, RPC208552H, RPC208563H, RPC208566H, RPC204619H, RPC206024H, RPC208594H, RPC210113H, RPC203601H, RPC204008H, RPC208012H, RPC208075H, RPC208283H, RPC209010H, RPC209017H, RPC209133H, RPC209498H, RPC209505H, RPC209666H, RPC209729H, RPC209730H, RPC209733H, RPC209922H, RPC209947H, RPC204328H, RPC205249H, RPC205402H, RPC205690H, RPC208561H, RPC208941H, RPC203923H, RPC203665H, RPC209984H, RPC202846H, RPC203169H, RPC203442H, RPC204082H, RPC207274H, RPC207298H, RPC208211H, RPC203833H, RPC207704H, RPC209563H, RPC209866H, RPC202630H, RPC202867H, RPC207908H, RPC208524H, RPC209154H, RPC203611H, RPC204726H, RPC205631H, RPC207219H, RPC207768H, RPC208417H, RPC208533H, RPC209765H, RPC208500H, RPC209313H, RPC210194H, RPC202932H, RPC207605H, RPC203957H, RPC204366H, RPC209185H, RPC206455H, RPC209418H, RPC206753H, RPC207268H, RPC204821H, RPC206585H, RPC206796H, RPC209616H, RPC205331H, RPC206675H, RPC208593H, RPC209972H, RPC206085H, RPC209077H, RPC209519H, RPC207503H, RPC209599H, RPC207059H, RPC203494H, RPC206567H, RPC208040H, RPC208540H, RPC210248H, RPC202445H, RPC209566H, RPC210174H, RPC202406H, RPC202897H, RPC203891H, RPC204152H, RPC204599H, RPC204663H, RPC205611H, RPC206154H, RPC206312H, RPC206981H, RPC207061H, RPC207130H, RPC208047H, RPC208219H, RPC209187H, RPC205580H, RPC209422H, RPC205225H, RPC205286H, RPC206506H, RPC205519H, RPC207600H, RPC207750H, RPC209276H, RPC209406H, RPC209478H, RPC204984H, RPC205590H, RPC210234H, RPC204003H, RPC206069H, RPC209689H, RPC204364H, RPC203987H, RPC204212H, RPC204453H, RPC204823H, RPC206925H, RPC207668H, RPC207669H, RPC208778H, RPC209699H, RPC209701H, RPC209704H, RPC203796H, RPC210220H, RPC202551H, RPC205542H, RPC207082H, RPC207083H, RPC209906H, RPC208930H, RPC203075H, RPC203733H, RPC204737H, RPC206428H, RPC205982H, RPC207970H, RPC209292H, RPC202332H, RPC206700H, RPC208484H, RPC209927H, RPC202346H, RPC203605H, RPC203610H, RPC205312H, RPC206589H, RPC206592H, RPC207466H, RPC207896H, RPC208017H, RPC208088H, RPC208144H, RPC205817H, RPC207442H, RPC206155H, RPC207057H, RPC208026H, RPC208396H, RPC203117H, RPC203338H, RPC203410H, RPC203682H, RPC204354H, RPC207415H, RPC208758H, RPC208760H, RPC208919H, RPC202629H, RPC207844H, RPC209364H, RPC209761H, RPC204909H, RPC207139H, RPC202460H, RPC204459H, RPC202637H, RPC202822H, RPC203286H, RPC204772H, RPC206932H, RPC206237H, RPC204981H, RPC207628H, RPC208464H, RPC206983H, RPC210216H, RPC203073H, RPC204665H, RPC206746H, RPC203681H, RPC204301H, RPC207361H, RPC209241H, RPC202755H, RPC202758H, RPC202878H, RPC203197H, RPC203820H, RPC204284H, RPC206868H, RPC207569H, RPC207886H, RPC209181H, RPC209247H, RPC210044H, RPC203427H, RPC203493H, RPC203586H, RPC204725H, RPC205513H, RPC205883H, RPC207389H, RPC207859H, RPC208103H, RPC209240H, RPC209646H, RPC202934H, RPC203351H, RPC205689H, RPC205899H, RPC208522H, RPC204872H, RPC204976H, RPC206736H, RPC208895H, RPC209712H, RPC203364H, RPC206664H, RPC203566H, RPC207592H, RPC209475H, RPC206652H, RPC208478H, RPC203133H, RPC203142H, RPC203468H, RPC203734H, RPC204049H, RPC204057H, RPC204337H, RPC204388H, RPC204903H, RPC204916H, RPC204917H, RPC205010H, RPC205221H, RPC205502H, RPC205989H, RPC206008H, RPC206943H, RPC207772H, RPC207774H, RPC207776H, RPC207777H, RPC207919H, RPC208344H, RPC210262H, RPC207585H, RPC202922H, RPC203196H, RPC204316H, RPC206388H, RPC207922H, RPC208983H, RPC204924H, RPC204032H, RPC205023H, RPC205026H, RPC205932H, RPC205933H, RPC205936H, RPC202519H, RPC205637H, RPC208989H, RPC208304H, RPC209688H, RPC203714H, RPC203976H, RPC207048H, RPC207126H, RPC208335H, RPC202907H, RPC204752H, RPC206545H, RPC207050H, RPC207343H, RPC207545H, RPC207607H, RPC207873H, RPC209245H, RPC209441H, RPC209938H, RPC206584H, RPC202780H, RPC202356H, RPC202449H, RPC203465H, RPC203469H, RPC203839H, RPC204236H, RPC204624H, RPC204744H, RPC205253H, RPC205463H, RPC205619H, RPC205895H, RPC205954H, RPC206146H, RPC206486H, RPC206591H, RPC207002H, RPC207033H, RPC207159H, RPC207399H, RPC207993H, RPC208168H, RPC208507H, RPC209179H, RPC205169H, RPC209904H, RPC206632H, RPC207058H, RPC204223H, RPC204342H, RPC204402H, RPC205644H, RPC205678H, RPC202455H, RPC204491H, RPC205726H, RPC206287H, RPC209981H, RPC204256H, RPC202348H, RPC202352H, RPC202422H, RPC202699H, RPC203593H, RPC204516H, RPC204697H, RPC205133H, RPC205383H, RPC206325H, RPC207026H, RPC207860H, RPC208044H, RPC208134H, RPC208171H, RPC209803H, RPC205212H, RPC206918H, RPC207851H, RPC202402H, RPC202563H, RPC202983H, RPC202987H, RPC203052H, RPC203572H, RPC203687H, RPC203803H, RPC204038H, RPC204039H, RPC204335H, RPC204455H, RPC205054H, RPC205055H, RPC205056H, RPC205057H, RPC205058H, RPC206027H, RPC206035H, RPC207865H, RPC208380H, RPC208784H, RPC208785H, RPC209702H, RPC209706H, RPC209964H, RPC208418H, RPC208879H, RPC204233H, RPC206569H, RPC206832H, RPC209193H, RPC209565H, RPC202798H, RPC202919H, RPC203482H, RPC202815H, RPC202890H, RPC204107H, RPC202560H, RPC202562H, RPC202573H, RPC202576H, RPC207999H, RPC208823H, RPC208845H, RPC208847H, RPC208850H, RPC208964H, RPC209346H, RPC209499H, RPC207706H, RPC204217H, RPC205278H, RPC206905H, RPC206906H, RPC209320H, RPC205885H, RPC206012H, RPC206272H, RPC208813H, RPC209812H, RPC204393H, RPC208498H, RPC208875H, RPC208902H, RPC202555H, RPC202782H, RPC203080H, RPC203335H, RPC203486H, RPC203660H, RPC203972H, RPC204093H, RPC204094H, RPC204112H, RPC204418H, RPC204718H, RPC204745H, RPC205044H, RPC205965H, RPC206100H, RPC206171H, RPC206214H, RPC206305H, RPC206418H, RPC206464H, RPC206521H, RPC207117H, RPC207136H, RPC207145H, RPC207167H, RPC207190H, RPC207726H, RPC207848H, RPC207998H, RPC208128H, RPC208178H, RPC208279H, RPC208371H, RPC208378H, RPC208391H, RPC209071H, RPC209184H, RPC209773H, RPC209797H, RPC203366H, RPC209554H, RPC202865H, RPC203917H, RPC206454H, RPC202682H, RPC203424H, RPC203460H, RPC204404H, RPC204645H, RPC205244H, RPC205384H, RPC205539H, RPC205691H, RPC206764H, RPC207548H, RPC207926H, RPC208008H, RPC208792H, RPC210142H, RPC210229H, RPC203488H, RPC204963H, RPC207158H, RPC207854H, RPC210026H, RPC210215H, RPC204887H, RPC203953H, RPC205258H, RPC207075H, RPC204334H, RPC204825H, RPC206084H, RPC207068H, RPC207827H, RPC205385H, RPC209781H, RPC204528H, RPC204531H, RPC208502H, RPC202374H, RPC205268H, RPC203170H, RPC202911H, RPC202857H, RPC202347H, RPC202722H, RPC203980H, RPC204004H, RPC206755H, RPC208565H, RPC209794H, RPC207552H, RPC203688H, RPC206936H, RPC210072H, RPC203299H, RPC204395H, RPC206233H, RPC207448H, RPC208589H, RPC208518H, RPC209138H, RPC203221H, RPC204161H, RPC206765H, RPC206769H, RPC207941H, RPC208302H, RPC208334H, RPC208572H, RPC209280H, RPC210179H, RPC202446H, RPC202731H, RPC203202H, RPC203827H, RPC203165H, RPC205340H, RPC206818H, RPC208205H, RPC204899H, RPC207429H, RPC207702H, RPC208160H, RPC208551H, RPC209019H, RPC209130H, RPC210140H, RPC210184H, RPC202435H, RPC202472H, RPC203082H, RPC204369H, RPC204629H, RPC206014H, RPC206974H, RPC207137H, RPC207934H, RPC208711H, RPC209084H, RPC209092H, RPC209200H, RPC209232H, RPC209597H, RPC209870H, RPC209887H, RPC209953H, RPC210005H, RPC210045H, RPC210071H, RPC210094H, RPC210130H, RPC210169H, RPC208322H, RPC204166H, RPC204849H, RPC205085H, RPC208987H, RPC208999H, RPC202483H, RPC204116H, RPC205269H, RPC206950H, RPC208360H, RPC202604H, RPC204089H, RPC204126H, RPC204495H, RPC206539H, RPC207206H, RPC207358H, RPC208945H, RPC208949H, RPC209175H, RPC205947H, RPC205949H, RPC205951H, RPC205952H, RPC206371H, RPC208982H, RPC209446H, RPC209591H, RPC202842H, RPC204478H, RPC204957H, RPC206822H, RPC207350H, RPC207550H, RPC207757H, RPC208519H, RPC209038H, RPC209135H, RPC209156H, RPC209263H, RPC209437H, RPC209562H, RPC202721H, RPC203294H, RPC206129H, RPC206170H, RPC206649H, RPC207950H, RPC208821H, RPC209915H, RPC210123H, RPC202421H, RPC203239H, RPC203305H, RPC204330H, RPC204736H, RPC207281H, RPC207868H, RPC208265H, RPC210181H, RPC202523H, RPC203556H, RPC206997H, RPC207393H, RPC208940H, RPC210091H, RPC204358H, RPC205280H, RPC208588H, RPC203276H, RPC204162H, RPC204174H, RPC204576H, RPC205136H, RPC205494H, RPC207766H, RPC207769H, RPC208195H, RPC210120H, RPC209975H, RPC204432H, RPC204675H, RPC205388H, RPC205708H, RPC207330H, RPC207364H, RPC207876H, RPC207900H, RPC207938H, RPC208440H, RPC208602H, RPC209357H, RPC209595H, RPC210278H, RPC202340H, RPC204086H, RPC204078H, RPC204081H, RPC204336H, RPC205190H, RPC208843H, RPC209374H, RPC203705H, RPC206489H, RPC204554H, RPC205246H, RPC209085H, RPC202484H, RPC202623H, RPC204282H, RPC205159H, RPC205674H, RPC206366H, RPC203193H, RPC204201H, RPC205487H, RPC204440H, RPC202841H, RPC206415H, RPC202640H, RPC204497H, RPC204505H, RPC204463H, RPC209632H, RPC208910H, RPC205443H, RPC204426H, RPC207565H, RPC209043H, RPC209110H, RPC203283H, RPC204716H, RPC203132H, RPC203532H, RPC204015H, RPC204020H, RPC204021H, RPC204028H, RPC204817H, RPC204839H, RPC204843H, RPC204846H, RPC205863H, RPC205864H, RPC203656H, RPC206988H, RPC206186H, RPC206397H, RPC207242H, RPC202633H, RPC203192H, RPC206472H, RPC207054H, RPC206809H, RPC208781H, RPC210201H, RPC203746H, RPC204587H, RPC205575H, RPC206995H, RPC207049H, RPC208410H, RPC208916H, RPC209472H, RPC202528H, RPC202886H, RPC204033H, RPC204041H, RPC204050H, RPC204905H, RPC204908H, RPC204915H, RPC206036H, RPC206037H, RPC206043H, RPC206044H, RPC206674H, RPC206833H, RPC206836H, RPC206942H, RPC207714H, RPC207716H, RPC208009H, RPC208376H, RPC208698H, RPC209275H, RPC209511H, RPC209513H, RPC209515H, RPC207222H, RPC209063H, RPC203409H, RPC209594H, RPC202542H, RPC204400H, RPC205680H, RPC205682H, RPC205737H, RPC207040H, RPC207097H, RPC207201H, RPC208479H, RPC208978H, RPC209081H, RPC209768H, RPC202597H, RPC209540H, RPC209992H, RPC210050H, RPC203252H, RPC203565H, RPC203577H, RPC204286H, RPC205121H, RPC206182H, RPC206662H, RPC206695H, RPC207606H, RPC208402H, RPC208656H, RPC209075H, RPC209090H, RPC209358H, RPC209830H, RPC208907H, RPC204378H, RPC204884H, RPC207017H, RPC207613H, RPC202596H, RPC202602H, RPC202648H, RPC203119H, RPC203120H, RPC203121H, RPC203969H, RPC204184H, RPC204293H, RPC204314H, RPC204582H, RPC204934H, RPC204937H, RPC204938H, RPC204942H, RPC204947H, RPC204950H, RPC205915H, RPC205919H, RPC205920H, RPC205921H, RPC205925H, RPC205927H, RPC205931H, RPC205934H, RPC205935H, RPC207665H, RPC207666H, RPC209260H, RPC209853H, RPC209858H, RPC210064H, RPC210069H, RPC210078H, RPC203272H, RPC206448H, RPC208631H, RPC202341H, RPC203346H, RPC203515H, RPC204266H, RPC204281H, RPC204598H, RPC204687H, RPC204910H, RPC205326H, RPC205647H, RPC205873H, RPC205876H, RPC205877H, RPC205978H, RPC206013H, RPC206148H, RPC206413H, RPC206414H, RPC206524H, RPC206531H, RPC206533H, RPC206534H, RPC206661H, RPC206669H, RPC206677H, RPC206682H, RPC206691H, RPC206699H, RPC206710H, RPC206716H, RPC206719H, RPC206720H, RPC206722H, RPC206823H, RPC207434H, RPC208621H, RPC208776H, RPC209217H, RPC209625H, RPC209627H, RPC209628H, RPC209630H, RPC209631H, RPC209633H, RPC209634H, RPC209635H, RPC209737H, RPC202601H, RPC207967H, RPC203031H, RPC204195H, RPC206886H, RPC207060H, RPC202762H, RPC203984H, RPC210247H, RPC204996H, RPC207697H, RPC208389H, RPC208704H, RPC209522H, RPC203003H, RPC204510H, RPC207856H, RPC208368H, RPC208671H, RPC209065H, RPC202400H, RPC202427H, RPC202535H, RPC202581H, RPC202893H, RPC202959H, RPC203023H, RPC203183H, RPC203223H, RPC203241H, RPC203740H, RPC203764H, RPC203829H, RPC203880H, RPC203894H, RPC204219H, RPC204270H, RPC204467H, RPC204838H, RPC204842H, RPC205104H, RPC205819H, RPC205973H, RPC206231H, RPC206278H, RPC206577H, RPC206594H, RPC207032H, RPC207755H, RPC207909H, RPC207925H, RPC208237H, RPC208241H, RPC208243H, RPC208407H, RPC209157H, RPC209202H, RPC209387H, RPC209430H, RPC209528H, RPC209529H, RPC210053H, RPC210212H, RPC202381H, RPC203848H, RPC203814H, RPC203447H, RPC208825H, RPC208438H, RPC208485H, RPC210085H, RPC204674H, RPC208446H, RPC202838H, RPC202861H, RPC203071H, RPC203081H, RPC203511H, RPC203712H, RPC204524H, RPC204558H, RPC204870H, RPC204876H, RPC204879H, RPC204880H, RPC205196H, RPC205426H, RPC205561H, RPC206565H, RPC206610H, RPC206790H, RPC206853H, RPC207109H, RPC207134H, RPC207895H, RPC208070H, RPC208204H, RPC208714H, RPC208732H, RPC209296H, RPC209297H, RPC209306H, RPC209603H, RPC209685H, RPC210158H, RPC205549H, RPC206094H, RPC202641H, RPC205809H, RPC209151H, RPC203728H, RPC209958H, RPC205232H, RPC206099H, RPC206185H, RPC205571H, RPC203753H, RPC206380H, RPC207611H, RPC209810H, RPC203666H, RPC204862H, RPC205867H, RPC206299H, RPC208616H, RPC209267H, RPC202962H, RPC207070H, RPC203675H, RPC207671H, RPC208176H, RPC204253H, RPC204434H, RPC207490H, RPC208889H, RPC206600H, RPC207656H, RPC210159H, RPC202529H, RPC202973H, RPC203932H, RPC205035H, RPC207378H, RPC207556H, RPC207635H, RPC207640H, RPC207641H, RPC207646H, RPC208471H, RPC208754H, RPC208755H, RPC209555H, RPC209558H, RPC202797H, RPC203164H, RPC203915H, RPC206255H, RPC206429H, RPC208202H, RPC208329H, RPC209147H, RPC209658H, RPC209659H, RPC202479H, RPC202594H, RPC202677H, RPC202953H, RPC203397H, RPC203420H, RPC203570H, RPC204852H, RPC204993H, RPC205499H, RPC206232H, RPC206286H, RPC206320H, RPC206404H, RPC206438H, RPC207233H, RPC207457H, RPC207834H, RPC207995H, RPC208001H, RPC208852H, RPC209258H, RPC202574H, RPC204431H, RPC206922H, RPC207112H, RPC207282H, RPC207525H, RPC207960H, RPC208390H, RPC208510H, RPC208603H, RPC209163H, RPC209886H, RPC210161H, RPC202365H, RPC203597H, RPC203865H, RPC205310H, RPC205392H, RPC205641H, RPC205839H, RPC207571H, RPC208164H, RPC208435H, RPC209355H, RPC204773H, RPC204774H, RPC203498H, RPC206590H, RPC207540H, RPC207905H, RPC207949H, RPC208086H, RPC208702H, RPC209813H, RPC204766H, RPC203943H, RPC203950H, RPC204205H, RPC209913H, RPC202373H, RPC205197H, RPC206684H, RPC207419H, RPC208482H, RPC208929H, RPC208998H, RPC201365H, RPC203685H, RPC204037H, RPC204520H, RPC206282H, RPC207657H, RPC209113H, RPC209839H, RPC206479H, RPC207637H, RPC208827H, RPC206705H, RPC208772H, RPC203882H, RPC208347H, RPC203137H, RPC203470H, RPC204796H, RPC204798H, RPC204803H, RPC206811H, RPC206812H, RPC206813H, RPC206386H, RPC204486H, RPC206330H, RPC206754H, RPC208060H, RPC209859H, RPC210148H, RPC205497H, RPC207405H, RPC203477H, RPC210171H, RPC204488H, RPC205940H, RPC205941H, RPC207250H, RPC207251H, RPC208221H, RPC208229H, RPC208231H, RPC208235H, RPC208474H, RPC206536H, RPC208935H, RPC203230H, RPC203717H, RPC205569H, RPC207300H, RPC207539H, RPC207244H, RPC202850H, RPC203884H, RPC204611H, RPC204782H, RPC206376H, RPC210109H, RPC210233H, RPC206113H, RPC207947H, RPC209910H, RPC202391H, RPC203281H, RPC203617H, RPC204514H, RPC205796H, RPC207549H, RPC208567H, RPC208618H, RPC208818H, RPC208885H, RPC208238H, RPC209042H, RPC204051H, RPC203332H, RPC204546H, RPC207148H, RPC207932H, RPC208641H, RPC209471H, RPC209932H, RPC202388H, RPC203994H, RPC203995H, RPC203997H, RPC204002H, RPC205028H, RPC205029H, RPC205222H, RPC205914H, RPC205918H, RPC205924H, RPC205926H, RPC208057H, RPC209226H, RPC202600H, RPC203175H, RPC203186H, RPC203940H, RPC204566H, RPC204671H, RPC205209H, RPC205481H, RPC205962H, RPC207512H, RPC209155H, RPC209842H, RPC203633H, RPC203692H, RPC203697H, RPC203966H, RPC204401H, RPC204409H, RPC207885H, RPC207910H, RPC204616H, RPC207447H, RPC207421H, RPC205103H, RPC207593H, RPC209214H, RPC209385H, RPC203111H, RPC203721H, RPC203722H, RPC207644H, RPC209281H, RPC209389H, RPC209945H, RPC206240H, RPC203088H, RPC206208H, RPC207933H, RPC209923H, RPC207178H, RPC208900H, RPC204610H, RPC207286H, RPC203079H, RPC204013H, RPC204265H, RPC205945H, RPC206048H, RPC207555H, RPC207570H, RPC207912H, RPC208181H, RPC208186H, RPC209044H, RPC209605H, RPC202982H, RPC202991H, RPC202992H, RPC202993H, RPC202998H, RPC204621H, RPC205276H, RPC205427H, RPC205437H, RPC205440H, RPC206076H, RPC206776H, RPC206779H, RPC206782H, RPC206786H, RPC206787H, RPC206788H, RPC206789H, RPC207069H, RPC207131H, RPC207132H, RPC207135H, RPC207256H, RPC207305H, RPC207718H, RPC208574H, RPC208576H, RPC208578H, RPC208579H, RPC208580H, RPC208582H, RPC208583H, RPC209160H, RPC209162H, RPC209334H, RPC209648H, RPC209649H, RPC209650H, RPC209651H, RPC209652H, RPC209814H, RPC206800H, RPC209359H, RPC203792H, RPC204125H, RPC205538H, RPC205960H, RPC205966H, RPC205971H, RPC209098H, RPC209208H, RPC209243H, RPC209584H, RPC209585H, RPC209590H, RPC206847H, RPC202835H, RPC203285H, RPC203297H, RPC205014H, RPC205595H, RPC205743H, RPC205758H, RPC206176H, RPC206196H, RPC207578H, RPC207599H, RPC208065H, RPC208458H, RPC208459H, RPC208492H, RPC208733H, RPC208734H, RPC209325H, RPC209330H, RPC209413H, RPC209660H, RPC209662H, RPC209668H, RPC209673H, RPC209718H, RPC205433H, RPC207306H, RPC209893H, RPC209956H, RPC202543H, RPC202544H, RPC202714H, RPC204596H, RPC204630H, RPC205297H, RPC205407H, RPC205452H, RPC206280H, RPC207157H, RPC209055H, RPC209266H, RPC209533H, RPC209713H, RPC209784H, RPC209894H, RPC209908H, RPC209934H, RPC210041H, RPC210272H, RPC210273H, RPC203620H, RPC205563H, RPC203797H, RPC205143H, RPC206054H, RPC206251H, RPC208318H, RPC208568H, RPC208604H, RPC210132H, RPC203609H, RPC204009H, RPC204010H, RPC204034H, RPC204095H, RPC204489H, RPC204691H, RPC204824H, RPC206508H, RPC206792H, RPC207398H, RPC208002H, RPC208011H, RPC208207H, RPC208774H, RPC208803H, RPC208864H, RPC209857H, RPC209900H, RPC209987H, RPC209988H, RPC204025H, RPC209153H, RPC209445H, RPC204815H, RPC207307H, RPC208258H, RPC208457H, RPC202473H, RPC203655H, RPC204209H, RPC204715H, RPC206169H, RPC206372H, RPC206557H, RPC207103H, RPC207969H, RPC208893H, RPC209196H, RPC209569H, RPC209759H, RPC210093H, RPC210157H, RPC210286H, RPC209103H, RPC205046H, RPC205530H, RPC205531H, RPC205532H, RPC205533H, RPC206032H, RPC206040H, RPC207486H, RPC208285H, RPC209024H, RPC209102H, RPC209290H, RPC209690H, RPC209692H, RPC209696H, RPC209697H, RPC209705H, RPC209735H, RPC204643H, RPC208962H, RPC205089H, RPC204185H, RPC204433H, RPC208292H, RPC207047H, RPC209037H, RPC210246H, RPC202407H, RPC208074H, RPC202895H, RPC205752H, RPC205754H, RPC205977H, RPC206726H, RPC206750H, RPC206752H, RPC209273H, RPC203661H, RPC202329H, RPC203229H, RPC203247H, RPC203631H, RPC204171H, RPC204211H, RPC204272H, RPC204411H, RPC204627H, RPC204668H, RPC205917H, RPC206851H, RPC206930H, RPC207034H, RPC207858H, RPC208023H, RPC208157H, RPC208943H, RPC209344H, RPC210023H, RPC206907H, RPC202786H, RPC204069H, RPC204070H, RPC204415H, RPC204807H, RPC205167H, RPC206403H, RPC209620H, RPC202559H, RPC202712H, RPC203990H, RPC204142H, RPC204252H, RPC204347H, RPC204958H, RPC205970H, RPC206184H, RPC206564H, RPC206738H, RPC206949H, RPC207878H, RPC208269H, RPC208620H, RPC208736H, RPC208838H, RPC208894H, RPC208908H, RPC209096H, RPC209414H, RPC209428H, RPC209855H, RPC209875H, RPC210180H, RPC202536H, RPC202871H, RPC202872H, RPC203495H, RPC203553H, RPC204283H, RPC204519H, RPC205415H, RPC206114H, RPC206202H, RPC207395H, RPC207425H, RPC207473H, RPC208003H, RPC208092H, RPC208165H, RPC208196H, RPC208954H, RPC208956H, RPC209770H, RPC205345H, RPC206131H, RPC203430H, RPC205703H, RPC204249H, RPC202453H, RPC202492H, RPC202728H, RPC202970H, RPC203632H, RPC203871H, RPC203872H, RPC205995H, RPC206680H, RPC208849H, RPC209268H, RPC209609H, RPC209940H, RPC207089H, RPC207094H, RPC208532H, RPC203473H, RPC204789H, RPC205381H, RPC205803H, RPC206402H, RPC206915H, RPC207325H, RPC202685H, RPC203255H, RPC205508H, RPC206629H, RPC209137H, RPC204441H, RPC209829H, RPC202925H, RPC203757H, RPC205591H, RPC205856H, RPC209186H, RPC210222H, RPC203011H, RPC203246H, RPC203949H, RPC204135H, RPC204150H, RPC204474H, RPC204912H, RPC204966H, RPC205150H, RPC205416H, RPC205435H, RPC205483H, RPC206074H, RPC206976H, RPC206979H, RPC207147H, RPC207171H, RPC207237H, RPC207822H, RPC208013H, RPC208059H, RPC208361H, RPC208365H, RPC208367H, RPC208423H, RPC208846H, RPC209048H, RPC209254H, RPC209257H, RPC209352H, RPC209353H, RPC209502H, RPC209792H, RPC209800H, RPC209802H, RPC209811H, RPC209823H, RPC209882H, RPC209884H, RPC209892H, RPC209899H, RPC209939H, RPC209969H, RPC209994H, RPC210275H, RPC210284H, RPC203549H, RPC208493H, RPC208615H, RPC209982H, RPC204907H, RPC205240H, RPC208816H, RPC206866H, RPC202653H, RPC203489H, RPC204241H, RPC204584H, RPC204642H, RPC204848H, RPC205581H, RPC206473H, RPC207092H, RPC207547H, RPC207889H, RPC207994H, RPC208010H, RPC208152H, RPC208398H, RPC209047H, RPC209799H, RPC202478H, RPC205317H, RPC202383H, RPC209449H, RPC207498H, RPC202443H, RPC203961H, RPC208612H, RPC203092H, RPC204810H, RPC207735H, RPC210235H, RPC202498H, RPC209114H, RPC209370H, RPC204422H, RPC208526H, RPC210167H, RPC206437H, RPC208135H, RPC201369H, RPC205476H, RPC208948H, RPC209350H, RPC202887H, RPC206120H, RPC207536H, RPC203355H, RPC205417H, RPC205669H, RPC207527H, RPC207931H, RPC208985H, RPC209066H, RPC209742H, RPC209748H, RPC202442H, RPC204287H, RPC208257H, RPC206980H, RPC202820H, RPC204633H, RPC207426H, RPC202926H, RPC205579H, RPC206985H, RPC209324H, RPC210075H, RPC208650H, RPC204193H, RPC208927H, RPC203026H, RPC206965H, RPC210267H, RPC204637H, RPC209865H, RPC202336H, RPC202565H, RPC202616H, RPC202702H, RPC202952H, RPC203614H, RPC203630H, RPC203854H, RPC204091H, RPC204176H, RPC204181H, RPC204470H, RPC205151H, RPC205521H, RPC205609H, RPC206128H, RPC206291H, RPC206311H, RPC206315H, RPC206672H, RPC206775H, RPC206837H, RPC207232H, RPC208336H, RPC208337H, RPC208414H, RPC208596H, RPC208709H, RPC208892H, RPC209033H, RPC209150H, RPC209189H, RPC209362H, RPC209461H, RPC209760H, RPC209793H, RPC209835H, RPC209880H, RPC209881H, RPC210103H, RPC210121H, RPC203545H, RPC203821H, RPC204288H, RPC204322H, RPC204517H, RPC204820H, RPC205650H, RPC206095H, RPC206872H, RPC208841H, RPC210039H, RPC210257H, RPC209604H, RPC206381H, RPC208444H, RPC209993H, RPC204468H, RPC207348H, RPC210260H, RPC209686H, RPC209757H, RPC210074H, RPC204669H, RPC205320H, RPC209190H, RPC205263H, RPC208836H, RPC204822H, RPC205423H, RPC207459H, RPC207504H, RPC208014H, RPC208312H, RPC209329H, RPC210061H, RPC203374H, RPC204974H, RPC202888H, RPC203743H, RPC204380H, RPC205180H, RPC206072H, RPC206553H, RPC204230H, RPC209876H, RPC202793H, RPC202801H, RPC209598H, RPC203642H, RPC203841H, RPC208076H, RPC202496H, RPC203546H, RPC207559H, RPC203379H, RPC205652H, RPC207080H, RPC206948H, RPC205592H, RPC202694H, RPC202695H, RPC204952H, RPC205408H, RPC205632H, RPC206871H, RPC209035H, RPC202747H, RPC202839H, RPC202913H, RPC203093H, RPC203503H, RPC203619H, RPC204109H, RPC204329H, RPC206106H, RPC206542H, RPC207508H, RPC208019H, RPC208233H, RPC208276H, RPC209050H, RPC209248H, RPC209961H, RPC202743H, RPC202943H, RPC204445H, RPC204447H, RPC206430H, RPC206921H, RPC208968H, RPC209152H, RPC206994H, RPC209399H, RPC210101H, RPC206596H, RPC208319H, RPC208595H, RPC209112H, RPC209911H, RPC203166H, RPC205009H, RPC202809H, RPC202533H, RPC206127H, RPC208133H, RPC208142H, RPC208548H, RPC202884H, RPC207114H, RPC203579H, RPC208906H, RPC204325H, RPC206236H, RPC207906H, RPC207907H, RPC210176H, RPC206289H, RPC208177H, RPC208980H, RPC208125H, RPC202990H, RPC204512H, RPC205338H, RPC205702H, RPC206178H, RPC206572H, RPC206854H, RPC207262H, RPC207935H, RPC208300H, RPC208413H, RPC208445H, RPC208775H, RPC203009H, RPC203312H, RPC204493H, RPC202824H, RPC203099H, RPC204758H, RPC205052H, RPC208742H, RPC209271H, RPC209596H, RPC208150H, RPC202570H, RPC202967H, RPC205482H, RPC207188H, RPC203360H, RPC203623H, RPC203670H, RPC204090H, RPC204472H, RPC208699H, RPC208939H, RPC209354H, RPC202339H, RPC203679H, RPC204080H, RPC204173H, RPC204200H, RPC204703H, RPC205618H, RPC205890H, RPC206398H, RPC206933H, RPC208151H, RPC210090H, RPC202615H, RPC202651H, RPC202799H, RPC203445H, RPC204309H, RPC204672H, RPC207454H, RPC207867H, RPC207955H, RPC202860H, RPC203130H, RPC204740H, RPC206198H, RPC209371H, RPC202638H, RPC203602H, RPC203696H, RPC203704H, RPC203710H, RPC205956H, RPC206563H, RPC206873H, RPC206875H, RPC206876H, RPC206877H, RPC206879H, RPC206880H, RPC208767H, RPC208768H, RPC209432H, RPC210255H, RPC204168H, RPC205038H, RPC205230H, RPC205233H, RPC205391H, RPC206729H, RPC206733H, RPC206734H, RPC206739H, RPC207435H, RPC208812H, RPC209535H, RPC209539H, RPC208232H, RPC209787H, RPC206133H, RPC208159H, RPC206313H, RPC208740H, RPC206951H, RPC209933H, RPC207501H, RPC203472H, RPC208802H, RPC208981H, RPC202550H, RPC204348H, RPC207689H, RPC203497H, RPC208990H, RPC209905H, RPC207207H, RPC209950H, RPC204487H, RPC205120H, RPC207377H, RPC208681H, RPC206911H, RPC202586H, RPC203126H, RPC203637H, RPC203855H, RPC204169H, RPC205616H, RPC206033H, RPC206223H, RPC206228H, RPC207469H, RPC208278H, RPC209921H, RPC203907H, RPC210062H, RPC202854H, RPC204267H, RPC208348H, RPC209046H, RPC206177H, RPC207624H, RPC202665H, RPC203809H, RPC206104H, RPC203065H, RPC204113H, RPC205666H, RPC210037H, RPC210047H, RPC203194H, RPC203260H, RPC204111H, RPC205179H, RPC205638H, RPC206257H, RPC208224H, RPC208225H, RPC208227H, RPC208228H, RPC208254H, RPC208933H, RPC206162H, RPC208737H, RPC208738H, RPC209227H, RPC209448H, RPC209693H, RPC209754H, RPC203264H, RPC203333H, RPC203775H, RPC203788H, RPC205378H, RPC205525H, RPC207392H, RPC208439H, RPC208442H, RPC208958H, RPC209282H, RPC203799H, RPC204296H, RPC204673H, RPC205200H, RPC210008H, RPC204155H, RPC208623H, RPC208626H, RPC208627H, RPC208636H, RPC209454H, RPC209455H, RPC209462H, RPC209466H, RPC209467H, RPC209473H, RPC209485H, RPC209486H, RPC203962H, RPC204259H, RPC206504H, RPC202553H, RPC205807H, RPC207518H, RPC208969H, RPC210232H, RPC207149H, RPC207413H, RPC207437H, RPC207439H, RPC208748H, RPC208912H, RPC210139H, RPC210191H, RPC210280H, RPC202564H, RPC203921H, RPC209903H, RPC204555H, RPC205279H, RPC204779H, RPC208110H, RPC208799H, RPC209056H, RPC209805H, RPC202963H, RPC208123H, RPC208605H, RPC208819H, RPC209544H, RPC209665H, RPC209973H, RPC203069H, RPC204819H, RPC207579H, RPC203025H, RPC203029H, RPC203490H, RPC204407H, RPC204498H, RPC204739H, RPC205070H, RPC205139H, RPC205142H, RPC205149H, RPC205156H, RPC205157H, RPC205195H, RPC205425H, RPC205993H, RPC206333H, RPC206502H, RPC207292H, RPC207293H, RPC207294H, RPC207647H, RPC207954H, RPC208035H, RPC208288H, RPC208362H, RPC208756H, RPC208761H, RPC208901H, RPC209002H, RPC209005H, RPC209006H, RPC209007H, RPC209009H, RPC209011H, RPC209012H, RPC210025H, RPC210027H, RPC210203H, RPC206894H, RPC209256H, RPC210225H, RPC205615H, RPC204612H, RPC205217H, RPC209991H, RPC208753H, RPC202750H, RPC202752H, RPC202756H, RPC202765H, RPC202766H, RPC202767H, RPC202777H, RPC202781H, RPC202784H, RPC202789H, RPC202796H, RPC202803H, RPC203899H, RPC204646H, RPC204647H, RPC204648H, RPC204650H, RPC204655H, RPC204657H, RPC204661H, RPC205866H, RPC205868H, RPC205869H, RPC205870H, RPC205872H, RPC205884H, RPC205893H, RPC205894H, RPC205896H, RPC205900H, RPC205901H, RPC205903H, RPC205905H, RPC205906H, RPC205907H, RPC208747H, RPC208749H, RPC208750H, RPC208751H, RPC208779H, RPC209639H, RPC209641H, RPC209642H, RPC209644H, RPC202345H, RPC203068H, RPC203464H, RPC203802H, RPC203838H, RPC204182H, RPC204186H, RPC204363H, RPC204841H, RPC205073H, RPC206117H, RPC208483H, RPC208539H, RPC208645H, RPC209396H, RPC203258H, RPC204557H, RPC205327H, RPC205765H, RPC206434H, RPC207000H, RPC209438H, RPC207120H, RPC204372H, RPC202585H, RPC202949H, RPC204011H, RPC204073H, RPC204074H, RPC204075H, RPC204604H, RPC204640H, RPC204666H, RPC204761H, RPC204769H, RPC204797H, RPC204805H, RPC204806H, RPC206248H, RPC206447H, RPC206515H, RPC207038H, RPC207270H, RPC207806H, RPC207809H, RPC207810H, RPC207811H, RPC207952H, RPC208223H, RPC209719H, RPC206089H, RPC208745H, RPC209746H, RPC209747H, RPC210188H, RPC204575H, RPC202464H, RPC202470H, RPC202514H, RPC202515H, RPC202522H, RPC202577H, RPC202587H, RPC202622H, RPC202735H, RPC203200H, RPC204304H, RPC204622H, RPC205439H, RPC206345H, RPC206346H, RPC206348H, RPC206350H, RPC206353H, RPC206354H, RPC206497H, RPC206761H, RPC206803H, RPC206804H, RPC206840H, RPC206882H, RPC206928H, RPC206929H, RPC207007H, RPC207091H, RPC208513H, RPC209503H, RPC209840H, RPC209852H, RPC210099H, RPC210105H, RPC210107H, RPC204216H, RPC205110H, RPC205118H, RPC205259H, RPC202704H, RPC202706H, RPC202708H, RPC202711H, RPC202920H, RPC202921H, RPC202924H, RPC203986H, RPC204341H, RPC204421H, RPC204521H, RPC204530H, RPC204538H, RPC204539H, RPC205311H, RPC205763H, RPC205764H, RPC205801H, RPC205994H, RPC206001H, RPC206006H, RPC206010H, RPC206022H, RPC206529H, RPC206530H, RPC206535H, RPC206537H, RPC209025H, RPC209173H, RPC209497H, RPC209575H, RPC209963H, RPC210277H, RPC202552H, RPC202736H, RPC203078H, RPC203219H, RPC203898H, RPC204263H, RPC204451H, RPC204560H, RPC204561H, RPC204564H, RPC204574H, RPC204940H, RPC204941H, RPC204944H, RPC204946H, RPC204949H, RPC204953H, RPC204955H, RPC204956H, RPC204960H, RPC205002H, RPC205304H, RPC205305H, RPC205564H, RPC205565H, RPC205568H, RPC205570H, RPC206259H, RPC206260H, RPC206262H, RPC206264H, RPC206267H, RPC206268H, RPC206410H, RPC206492H, RPC206494H, RPC206495H, RPC206498H, RPC206499H, RPC206501H, RPC206503H, RPC206637H, RPC206638H, RPC206639H, RPC206640H, RPC206642H, RPC206643H, RPC206644H, RPC206647H, RPC206651H, RPC207617H, RPC207619H, RPC207620H, RPC207897H, RPC208114H, RPC208120H, RPC208249H, RPC208394H, RPC208441H, RPC208499H, RPC208501H, RPC208508H, RPC208509H, RPC208511H, RPC208607H, RPC208608H, RPC208637H, RPC208695H, RPC208701H, RPC208703H, RPC208904H, RPC209101H, RPC209298H, RPC209299H, RPC209300H, RPC209302H, RPC210211H, RPC204361H, RPC205045H, RPC206156H, RPC206206H, RPC202631H, RPC203368H, RPC204062H, RPC204906H, RPC205334H, RPC207260H, RPC207891H, RPC210116H, RPC206060H, RPC208122H, RPC205950H, RPC207862H, RPC207758H, RPC207812H, RPC207813H, RPC207814H, RPC207815H, RPC207816H, RPC207817H, RPC208649H, RPC209710H, RPC203373H, RPC207427H, RPC203729H, RPC207587H, RPC206555H, RPC208324H, RPC207276H, RPC207279H, RPC207326H, RPC207492H, RPC207743H, RPC207744H, RPC207746H, RPC207747H, RPC207748H, RPC208856H, RPC208857H, RPC208858H, RPC208861H, RPC208868H, RPC208871H, RPC209995H, RPC209996H, RPC210000H, RPC210001H, RPC210146H, RPC210147H, RPC210187H, RPC210192H, RPC210268H, RPC210269H, RPC204248H, RPC204500H, RPC206097H, RPC206956H, RPC207845H, RPC208670H, RPC208993H, RPC208328H, RPC207986H, RPC208212H, RPC208746H, RPC209695H, RPC206801H, RPC209640H, RPC209849H, RPC206941H, RPC207071H, RPC207204H, RPC207205H, RPC207210H, RPC207214H, RPC207235H, RPC207246H, RPC207263H, RPC207433H, RPC207440H, RPC207444H, RPC207446H, RPC207451H, RPC207481H, RPC207482H, RPC207531H, RPC207572H, RPC207594H, RPC207638H, RPC207678H, RPC207684H, RPC207731H, RPC207732H, RPC207764H, RPC207767H, RPC208310H, RPC208345H, RPC208387H, RPC208516H, RPC208517H, RPC208527H, RPC208541H, RPC208545H, RPC208558H, RPC208591H, RPC208624H, RPC208638H, RPC208642H, RPC208660H, RPC208676H, RPC208677H, RPC208678H, RPC208679H, RPC208680H, RPC208691H, RPC208721H, RPC208723H, RPC208822H, RPC208829H, RPC208837H, RPC208878H, RPC208909H, RPC208934H, RPC208965H, RPC208988H, RPC208997H, RPC209125H, RPC209126H, RPC209321H, RPC209340H, RPC209342H, RPC209363H, RPC209368H, RPC209369H, RPC209453H, RPC209570H, RPC209777H, RPC209779H, RPC209789H, RPC209819H, RPC209825H, RPC209952H, RPC209955H, RPC209957H, RPC209959H, RPC209999H, RPC210015H, RPC210021H, RPC210063H, RPC210065H, RPC210080H, RPC210118H, RPC210131H, RPC210134H, RPC210238H, RPC210242H, RPC210243H, RPC210244H, RPC210252H, RPC210254H, RPC210259H, RPC210263H, RPC210289H; GTIN 00763000059583, Lot Serial Numbers: RPC602804S, RPC606143S, RPC606233S, RPC600478S, RPC600541S, RPC607389S, RPC604078S, RPC603092S, RPC604623S, RPC605127S, RPC606506S, RPC601525S, RPC607802S, RPC608105S, RPC608239S, RPC605355S, RPC605465S, RPC605647S, RPC605649S, RPC605665S, RPC605777S, RPC605840S, RPC606310S, RPC606312S, RPC607275S, RPC602140S, RPC603096S, RPC604001S, RPC605657S, RPC603990S, RPC605048S, RPC605049S, RPC605321S, RPC605322S, RPC605323S, RPC605324S, RPC605336S, RPC605354S, RPC605369S, RPC605370S, RPC605425S, RPC605491S, RPC605492S, RPC605571S, RPC605574S, RPC605781S, RPC605782S, RPC605783S, RPC606420S, RPC606447S, RPC606544S, RPC606679S, RPC607053S, RPC607054S, RPC607055S, RPC607056S, RPC607105S, RPC607686S, RPC607691S, RPC600468S, RPC601165S, RPC601487S, RPC601624S, RPC601791S, RPC602122S, RPC602148S, RPC602325S, RPC602386S, RPC602387S, RPC602388S, RPC602463S, RPC602474S, RPC602643S, RPC602650S, RPC602699S, RPC602747S, RPC602760S, RPC603311S, RPC603316S, RPC603363S, RPC603365S, RPC603367S, RPC603368S, RPC603812S, RPC603813S, RPC604182S, RPC604212S, RPC604251S, RPC604340S, RPC604341S, RPC604519S, RPC604601S, RPC604602S, RPC604604S, RPC607683S, RPC607684S, RPC607687S, RPC607688S, RPC607106S, RPC605469S, RPC606350S, RPC600818S, RPC601506S, RPC601857S, RPC602762S, RPC605790S, RPC605791S, RPC607533S, RPC600017S, RPC600019S, RPC600023S, RPC600031S, RPC601940S, RPC603412S, RPC607456S, RPC600840S, RPC600957S, RPC601048S, RPC601812S, RPC601924S, RPC602096S, RPC602363S, RPC602436S, RPC602845S, RPC607002S, RPC605371S, RPC601167S, RPC601701S, RPC603649S, RPC603732S, RPC602214S, RPC604037S, RPC605578S, RPC607618S, RPC601762S, RPC605462S, RPC605687S, RPC606154S, RPC600128S, RPC600200S, RPC602308S, RPC602880S, RPC602889S, RPC603680S, RPC606358S, RPC605542S, RPC607303S, RPC602561S, RPC602581S, RPC603625S, RPC605662S, RPC608029S, RPC600922S, RPC603360S, RPC604621S, RPC605330S, RPC600585S, RPC600893S, RPC601292S, RPC601526S, RPC601726S, RPC602829S, RPC601842S, RPC600777S, RPC601923S, RPC602715S, RPC602990S, RPC605479S, RPC605953S, RPC605954S, RPC600884S, RPC601053S, RPC601155S, RPC604088S, RPC604409S, RPC605244S, RPC606636S, RPC605000S, RPC605175S, RPC605176S, RPC605262S, RPC605379S, RPC605697S, RPC605703S, RPC605704S, RPC605765S, RPC606235S, RPC607038S, RPC607045S, RPC601401S, RPC601523S, RPC601564S, RPC602306S, RPC602413S, RPC602866S, RPC603024S, RPC603510S, RPC604383S, RPC604616S, RPC607203S, RPC607204S, RPC607311S, RPC608300S, RPC606951S, RPC602684S, RPC605054S, RPC605055S, RPC601077S, RPC601304S, RPC605280S, RPC605282S, RPC600881S, RPC600882S, RPC601629S, RPC602667S, RPC603208S, RPC603212S, RPC603213S, RPC603614S, RPC603671S, RPC603673S, RPC601416S, RPC606272S, RPC604017S, RPC603912S, RPC604396S, RPC604463S, RPC604812S, RPC601748S, RPC603030S, RPC600760S, RPC603983S, RPC607884S, RPC607885S, RPC607888S, RPC605625S, RPC601785S, RPC603167S, RPC603170S, RPC603206S, RPC603297S, RPC603356S, RPC603659S, RPC603675S, RPC603690S, RPC603838S, RPC603862S, RPC603958S, RPC601118S, RPC605533S, RPC605713S, RPC602035S, RPC602657S, RPC604704S, RPC602427S, RPC600048S, RPC600049S, RPC600224S, RPC600587S, RPC608184S, RPC606817S, RPC605611S, RPC605824S, RPC605847S, RPC604044S, RPC604345S, RPC606075S, RPC606949S, RPC602332S, RPC603396S, RPC605216S, RPC607408S, RPC602006S, RPC605327S, RPC601630S, RPC604211S, RPC604799S, RPC604888S, RPC605499S, RPC605837S, RPC607647S, RPC602826S, RPC602850S, RPC602162S, RPC607514S, RPC600259S, RPC602319S, RPC601950S, RPC602645S, RPC602662S, RPC605835S, RPC600604S, RPC602394S, RPC604053S, RPC604213S, RPC605337S, RPC603011S, RPC606074S, RPC600726S, RPC600768S, RPC600979S, RPC601031S, RPC601228S, RPC602601S, RPC603448S, RPC604089S, RPC605294S, RPC607230S, RPC600714S, RPC601227S, RPC602376S, RPC602742S, RPC604908S, RPC600877S, RPC601121S, RPC601638S, RPC603049S, RPC603155S, RPC604342S, RPC605375S, RPC606534S, RPC600163S, RPC600302S, RPC602242S, RPC602888S, RPC603436S, RPC604807S, RPC608062S, RPC608063S, RPC607609S, RPC603925S, RPC605822S, RPC601370S, RPC604877S, RPC605052S, RPC605070S, RPC607422S, RPC600254S, RPC600562S, RPC601094S, RPC601454S, RPC601990S, RPC602208S, RPC603405S, RPC603652S, RPC603714S, RPC604370S, RPC605894S, RPC606123S, RPC606290S, RPC606411S, RPC600455S, RPC601795S, RPC601796S, RPC602327S, RPC604773S, RPC602788S, RPC602532S, RPC603055S, RPC604762S, RPC600701S, RPC601050S, RPC601218S, RPC601605S, RPC601662S, RPC602256S, RPC602400S, RPC603305S, RPC603678S, RPC604148S, RPC607865S, RPC600621S, RPC602076S, RPC603868S, RPC600231S, RPC606144S, RPC600539S, RPC600962S, RPC606209S, RPC606118S, RPC601781S, RPC602090S, RPC602422S, RPC602879S, RPC603861S, RPC602071S, RPC603142S, RPC603237S, RPC603278S, RPC607487S, RPC607819S, RPC600050S, RPC600051S, RPC600052S, RPC600364S, RPC603272S, RPC604073S, RPC607818S, RPC600139S, RPC604000S, RPC604060S, RPC604066S, RPC604067S, RPC604244S, RPC604249S, RPC604686S, RPC604689S, RPC604702S, RPC608133S, RPC603952S, RPC604084S, RPC607050S, RPC601854S, RPC602265S, RPC603323S, RPC607577S, RPC600325S, RPC601535S, RPC604903S, RPC605518S, RPC606568S, RPC606624S, RPC600378S, RPC600526S, RPC601136S, RPC602441S, RPC604970S, RPC606009S, RPC606136S, RPC607383S, RPC601951S, RPC602705S, RPC604321S, RPC605256S, RPC606604S, RPC600227S, RPC600823S, RPC601295S, RPC602049S, RPC602087S, RPC602241S, RPC603095S, RPC603125S, RPC603496S, RPC603848S, RPC604186S, RPC604187S, RPC604346S, RPC605493S, RPC603964S, RPC604827S, RPC604851S, RPC604859S, RPC604861S, RPC605009S, RPC606804S, RPC606806S, RPC606807S, RPC606808S, RPC606809S, RPC607833S, RPC607834S, RPC601543S, RPC601544S, RPC601545S, RPC601546S, RPC601855S, RPC601863S, RPC602330S, RPC603168S, RPC603169S, RPC603721S, RPC603722S, RPC603723S, RPC603724S, RPC603725S, RPC604237S, RPC604410S, RPC604424S, RPC604440S, RPC604487S, RPC607830S, RPC607831S, RPC607832S, RPC604847S, RPC604963S, RPC604986S, RPC606064S, RPC606065S, RPC606066S, RPC606067S, RPC600655S, RPC600685S, RPC600854S, RPC600865S, RPC600867S, RPC600880S, RPC601139S, RPC601426S, RPC601911S, RPC602528S, RPC602853S, RPC602861S, RPC603077S, RPC604018S, RPC605950S, RPC606063S, RPC604627S, RPC604931S, RPC605263S, RPC604516S, RPC602706S, RPC602818S, RPC606278S, RPC606397S, RPC602133S, RPC602637S, RPC605884S, RPC606923S, RPC600222S, RPC600990S, RPC602923S, RPC605187S, RPC606127S, RPC606160S, RPC600711S, RPC601311S, RPC602048S, RPC604785S, RPC601007S, RPC600062S, RPC600063S, RPC605850S, RPC606626S, RPC603752S, RPC602414S, RPC603706S, RPC600324S, RPC602253S, RPC604919S, RPC604928S, RPC605013S, RPC605113S, RPC605141S, RPC605301S, RPC605302S, RPC605311S, RPC605312S, RPC606224S, RPC606322S, RPC606384S, RPC607425S, RPC600659S, RPC601264S, RPC602557S, RPC603191S, RPC608146S, RPC600074S, RPC605160S, RPC605419S, RPC605420S, RPC605789S, RPC606298S, RPC606299S, RPC606704S, RPC607193S, RPC607486S, RPC607623S, RPC607639S, RPC607705S, RPC607905S, RPC600510S, RPC600571S, RPC600605S, RPC600661S, RPC600807S, RPC600927S, RPC600954S, RPC600964S, RPC601055S, RPC601379S, RPC601458S, RPC601498S, RPC601598S, RPC601618S, RPC601888S, RPC601889S, RPC602067S, RPC602120S, RPC602199S, RPC602348S, RPC602356S, RPC602487S, RPC602548S, RPC602567S, RPC602640S, RPC602873S, RPC603041S, RPC603093S, RPC603196S, RPC603198S, RPC603217S, RPC603432S, RPC603517S, RPC604076S, RPC604791S, RPC607925S, RPC608138S, RPC608243S, RPC600344S, RPC600345S, RPC600105S, RPC601276S, RPC601329S, RPC601333S, RPC601489S, RPC601712S, RPC603935S, RPC605011S, RPC605234S, RPC605487S, RPC605977S, RPC606215S, RPC606346S, RPC606524S, RPC606752S, RPC606850S, RPC606851S, RPC602178S, RPC602324S, RPC602646S, RPC603400S, RPC603414S, RPC603711S, RPC604365S, RPC604540S, RPC605388S, RPC605650S, RPC606407S, RPC606937S, RPC601993S, RPC602673S, RPC604803S, RPC600811S, RPC602644S, RPC602733S, RPC602884S, RPC605095S, RPC607127S, RPC607579S, RPC602062S, RPC604151S, RPC604653S, RPC607128S, RPC600119S, RPC601140S, RPC601970S, RPC603426S, RPC602576S, RPC604983S, RPC605938S, RPC605964S, RPC606457S, RPC607342S, RPC607782S, RPC600682S, RPC601145S, RPC601213S, RPC601286S, RPC602020S, RPC602622S, RPC603611S, RPC603795S, RPC604085S, RPC604086S, RPC604415S, RPC604693S, RPC605878S, RPC605879S, RPC606295S, RPC607267S, RPC607510S, RPC600288S, RPC601331S, RPC602163S, RPC603834S, RPC603835S, RPC604234S, RPC606828S, RPC600096S, RPC600193S, RPC600424S, RPC600888S, RPC601242S, RPC602198S, RPC602273S, RPC602659S, RPC602691S, RPC603298S, RPC603485S, RPC603736S, RPC603989S, RPC604593S, RPC604600S, RPC604657S, RPC605015S, RPC603210S, RPC604056S, RPC604165S, RPC604728S, RPC603344S, RPC605016S, RPC606708S, RPC602603S, RPC603457S, RPC604544S, RPC606273S, RPC601459S, RPC604669S, RPC601551S, RPC602124S, RPC603458S, RPC606120S, RPC606965S, RPC600202S, RPC605107S, RPC605208S, RPC605471S, RPC605483S, RPC605605S, RPC606518S, RPC600323S, RPC600947S, RPC601881S, RPC602527S, RPC603081S, RPC604214S, RPC604430S, RPC605084S, RPC605616S, RPC606110S, RPC606367S, RPC606694S, RPC607893S, RPC602410S, RPC602746S, RPC603382S, RPC604181S, RPC604287S, RPC604776S, RPC605623S, RPC605774S, RPC602778S, RPC603310S, RPC604317S, RPC604323S, RPC601963S, RPC607118S, RPC600462S, RPC602311S, RPC605814S, RPC604495S, RPC604496S, RPC604577S, RPC604711S, RPC607731S, RPC606675S, RPC605121S, RPC606676S, RPC603166S, RPC607730S, RPC605714S, RPC604996S, RPC605223S, RPC605342S, RPC605603S, RPC606007S, RPC606183S, RPC606184S, RPC607355S, RPC600759S, RPC601170S, RPC601241S, RPC601279S, RPC601570S, RPC602524S, RPC604597S, RPC600234S, RPC600850S, RPC605976S, RPC600435S, RPC601079S, RPC601319S, RPC603086S, RPC603306S, RPC604735S, RPC604813S, RPC604354S, RPC600331S, RPC600504S, RPC600706S, RPC600805S, RPC602776S, RPC603710S, RPC605871S, RPC604909S, RPC603509S, RPC604638S, RPC606846S, RPC605776S, RPC600625S, RPC601574S, RPC601648S, RPC601938S, RPC601590S, RPC601183S, RPC601713S, RPC602948S, RPC606645S, RPC606902S, RPC604588S, RPC605224S, RPC601917S, RPC603329S, RPC605352S, RPC604811S, RPC606157S, RPC607035S, RPC601807S, RPC601886S, RPC605245S, RPC607132S, RPC601205S, RPC604324S, RPC600508S, RPC601253S, RPC601478S, RPC602792S, RPC604584S, RPC604957S, RPC607134S, RPC601718S, RPC602872S, RPC604091S, RPC607720S, RPC604451S, RPC603453S, RPC605553S, RPC605688S, RPC606100S, RPC606103S, RPC606135S, RPC607817S, RPC600328S, RPC600853S, RPC601024S, RPC603302S, RPC603317S, RPC603867S, RPC605033S, RPC605737S, RPC605754S, RPC605755S, RPC605952S, RPC606147S, RPC606566S, RPC606588S, RPC607131S, RPC607169S, RPC607172S, RPC607394S, RPC607472S, RPC607516S, RPC600084S, RPC600085S, RPC600333S, RPC600602S, RPC600676S, RPC600829S, RPC600924S, RPC602988S, RPC603283S, RPC603376S, RPC603406S, RPC603463S, RPC603464S, RPC604395S, RPC604730S, RPC603888S, RPC604005S, RPC606254S, RPC606468S, RPC607380S, RPC600204S, RPC604123S, RPC606475S, RPC606819S, RPC606820S, RPC606821S, RPC606822S, RPC604058S, RPC604834S, RPC605190S, RPC605192S, RPC605194S, RPC605409S, RPC605410S, RPC606265S, RPC606266S, RPC606352S, RPC606617S, RPC606618S, RPC600493S, RPC600509S, RPC600934S, RPC603506S, RPC604967S, RPC602729S, RPC606132S, RPC604875S, RPC604878S, RPC604879S, RPC605050S, RPC605547S, RPC606091S, RPC606455S, RPC607548S, RPC607708S, RPC601514S, RPC601808S, RPC603941S, RPC604072S, RPC604447S, RPC604580S, RPC604796S, RPC605530S, RPC601172S, RPC601734S, RPC603101S, RPC603338S, RPC603870S, RPC607866S, RPC600395S, RPC600396S, RPC605019S, RPC605020S, RPC605021S, RPC605023S, RPC605024S, RPC605028S, RPC607208S, RPC607257S, RPC607261S, RPC601492S, RPC602639S, RPC603576S, RPC603577S, RPC603578S, RPC604022S, RPC604024S, RPC604026S, RPC607079S, RPC603932S, RPC604833S, RPC604925S, RPC602117S, RPC601422S, RPC603314S, RPC603670S, RPC601075S, RPC602773S, RPC603515S, RPC604057S, RPC607633S, RPC607185S, RPC600363S, RPC604014S, RPC604766S, RPC607033S, RPC601890S, RPC603756S, RPC605689S, RPC606449S, RPC606546S, RPC601461S, RPC600368S, RPC605296S, RPC607321S, RPC600603S, RPC603059S, RPC608069S, RPC600223S, RPC605606S, RPC601691S, RPC605056S, RPC605069S, RPC605103S, RPC606326S, RPC606532S, RPC600664S, RPC601880S, RPC603071S, RPC603992S, RPC605094S, RPC600187S, RPC605283S, RPC602580S, RPC603215S, RPC603564S, RPC603572S, RPC603700S, RPC603735S, RPC607417S, RPC602569S, RPC607792S, RPC606439S, RPC605716S, RPC606058S, RPC606086S, RPC606087S, RPC606088S, RPC606325S, RPC606764S, RPC607074S, RPC608025S, RPC600440S, RPC600521S, RPC600915S, RPC601013S, RPC601980S, RPC601981S, RPC602066S, RPC602164S, RPC602553S, RPC602611S, RPC602779S, RPC602935S, RPC603287S, RPC603595S, RPC603610S, RPC603979S, RPC604825S, RPC606613S, RPC605179S, RPC606115S, RPC600616S, RPC600859S, RPC601949S, RPC603351S, RPC607960S, RPC606072S, RPC606899S, RPC606925S, RPC603053S, RPC604367S, RPC606090S, RPC600061S, RPC603926S, RPC603393S, RPC603865S, RPC604175S, RPC606201S, RPC605080S, RPC604935S, RPC605495S, RPC605920S, RPC600665S, RPC601289S, RPC601301S, RPC601944S, RPC602182S, RPC602206S, RPC602506S, RPC602757S, RPC602932S, RPC603052S, RPC603322S, RPC603504S, RPC603940S, RPC604229S, RPC604295S, RPC604299S, RPC605577S, RPC605760S, RPC607390S, RPC607891S, RPC600280S, RPC601501S, RPC603008S, RPC603501S, RPC603552S, RPC607421S, RPC605497S, RPC600843S, RPC601288S, RPC601608S, RPC602377S, RPC603904S, RPC604353S, RPC607431S, RPC607432S, RPC601556S, RPC602707S, RPC604236S, RPC604301S, RPC602842S, RPC604173S, RPC605415S, RPC605418S, RPC605780S, RPC600568S, RPC603468S, RPC605636S, RPC605674S, RPC605979S, RPC606915S, RPC606934S, RPC608244S, RPC600186S, RPC603898S, RPC604137S, RPC604705S, RPC604809S, RPC604727S, RPC602027S, RPC605728S, RPC606542S, RPC606733S, RPC607339S, RPC607340S, RPC608005S, RPC600783S, RPC601061S, RPC602375S, RPC603042S, RPC605325S, RPC606670S, RPC606697S, RPC601006S, RPC603171S, RPC605335S, RPC606182S, RPC606210S, RPC606408S, RPC606952S, RPC607047S, RPC604376S, RPC606418S, RPC607544S, RPC607545S, RPC607032S, RPC600812S, RPC602653S, RPC607640S, RPC605252S, RPC600599S, RPC600920S, RPC600942S, RPC601587S, RPC601945S, RPC602119S, RPC603845S, RPC605563S, RPC606167S, RPC606815S, RPC607445S, RPC602930S, RPC603878S, RPC605320S, RPC608209S, RPC604849S, RPC605143S, RPC606102S, RPC607059S, RPC601496S, RPC602310S, RPC604082S, RPC604208S, RPC605901S, RPC601164S, RPC604839S, RPC605255S, RPC600930S, RPC602767S, RPC603591S, RPC604197S, RPC607989S, RPC606572S, RPC606982S, RPC607490S, RPC607736S, RPC604599S, RPC607824S, RPC608192S, RPC602619S, RPC603776S, RPC604830S, RPC604892S, RPC605251S, RPC605253S, RPC605756S, RPC605770S, RPC605905S, RPC606170S, RPC606515S, RPC606720S, RPC607192S, RPC607443S, RPC607717S, RPC600348S, RPC600523S, RPC600762S, RPC600939S, RPC601026S, RPC601217S, RPC601296S, RPC601383S, RPC601919S, RPC602704S, RPC603650S, RPC603800S, RPC603804S, RPC603945S, RPC603999S, RPC604092S, RPC604369S, RPC604431S, RPC604466S, RPC604521S, RPC604611S, RPC604806S, RPC607908S, RPC605631S, RPC605826S, RPC603424S, RPC603425S, RPC604814S, RPC605383S, RPC601285S, RPC602114S, RPC605926S, RPC606558S, RPC600520S, RPC600688S, RPC604298S, RPC607393S, RPC605569S, RPC600295S, RPC600967S, RPC602693S, RPC602914S, RPC604074S, RPC604118S, RPC604411S, RPC602813S, RPC604844S, RPC604854S, RPC606641S, RPC601757S, RPC601817S, RPC602714S, RPC603951S, RPC600382S, RPC600722S, RPC603163S, RPC602759S, RPC603111S, RPC603260S, RPC603263S, RPC603507S, RPC603796S, RPC605400S, RPC605823S, RPC601866S, RPC604458S, RPC605918S, RPC606895S, RPC608003S, RPC608004S, RPC600458S, RPC601093S, RPC601122S, RPC601424S, RPC604717S, RPC607692S, RPC608293S, RPC606547S, RPC602898S, RPC604359S, RPC606109S, RPC603560S, RPC603561S, RPC604185S, RPC600164S, RPC603980S, RPC607538S, RPC607651S, RPC605154S, RPC607039S, RPC605193S, RPC606316S, RPC603307S, RPC603567S, RPC606315S, RPC605405S, RPC605498S, RPC605607S, RPC605919S, RPC606671S, RPC606883S, RPC601868S, RPC601907S, RPC603069S, RPC603264S, RPC603460S, RPC604149S, RPC604745S, RPC606768S, RPC602150S, RPC602815S, RPC606962S, RPC605092S, RPC605909S, RPC605955S, RPC606655S, RPC600892S, RPC601042S, RPC601393S, RPC602104S, RPC602247S, RPC602520S, RPC602974S, RPC604042S, RPC604356S, RPC604456S, RPC604457S, RPC600091S, RPC607732S, RPC601480S, RPC603090S, RPC601675S, RPC605043S, RPC605345S, RPC606526S, RPC606578S, RPC606748S, RPC604102S, RPC604271S, RPC604337S, RPC603777S, RPC602755S, RPC600488S, RPC601932S, RPC604231S, RPC605422S, RPC605523S, RPC605549S, RPC600175S, RPC600974S, RPC602074S, RPC602551S, RPC606759S, RPC607635S, RPC604009S, RPC607801S, RPC606903S, RPC602669S, RPC607291S, RPC603162S, RPC603954S, RPC604947S, RPC605034S, RPC605045S, RPC605404S, RPC605412S, RPC606083S, RPC606630S, RPC607968S, RPC600584S, RPC603438S, RPC603600S, RPC604168S, RPC607168S, RPC600077S, RPC604270S, RPC604786S, RPC605508S, RPC600024S, RPC606666S, RPC600503S, RPC603518S, RPC604779S, RPC600406S, RPC601580S, RPC601204S, RPC601761S, RPC603141S, RPC603582S, RPC602354S, RPC607418S, RPC607099S, RPC603957S, RPC605240S, RPC600710S, RPC604198S, RPC606237S, RPC607254S, RPC601493S, RPC603031S, RPC603906S, RPC605753S, RPC606203S, RPC600878S, RPC603773S, RPC604777S, RPC607707S, RPC600069S, RPC601142S, RPC602921S, RPC604015S, RPC604822S, RPC607160S, RPC605125S, RPC605126S, RPC605184S, RPC605517S, RPC605820S, RPC605859S, RPC606112S, RPC606268S, RPC606303S, RPC607065S, RPC607189S, RPC601377S, RPC604442S, RPC604722S, RPC604781S, RPC604783S, RPC604784S, RPC605860S, RPC606623S, RPC606996S, RPC605586S, RPC605702S, RPC605757S, RPC606456S, RPC600192S, RPC600607S, RPC601208S, RPC601565S, RPC601885S, RPC602564S, RPC603324S, RPC603442S, RPC605124S, RPC605170S, RPC601467S, RPC603045S, RPC603899S, RPC605478S, RPC605490S, RPC605733S, RPC605827S, RPC606202S, RPC606231S, RPC606659S, RPC606966S, RPC607067S, RPC607912S, RPC600312S, RPC601119S, RPC601290S, RPC601298S, RPC601351S, RPC601388S, RPC601533S, RPC602335S, RPC602452S, RPC602453S, RPC602498S, RPC602642S, RPC602685S, RPC603200S, RPC603201S, RPC603202S, RPC603203S, RPC603692S, RPC603826S, RPC608130S, RPC608139S, RPC605720S, RPC603762S, RPC600366S, RPC601308S, RPC601715S, RPC604678S, RPC600816S, RPC602913S, RPC604188S, RPC604660S, RPC605514S, RPC605570S, RPC606594S, RPC607162S, RPC600060S, RPC600461S, RPC601860S, RPC602863S, RPC603893S, RPC604103S, RPC604158S, RPC605351S, RPC605851S, RPC605887S, RPC605892S, RPC606261S, RPC606395S, RPC606598S, RPC600592S, RPC600980S, RPC601716S, RPC601953S, RPC602358S, RPC603015S, RPC603262S, RPC603459S, RPC603698S, RPC603771S, RPC603887S, RPC604614S, RPC604718S, RPC604719S, RPC604721S, RPC606392S, RPC606473S, RPC600239S, RPC604048S, RPC607037S, RPC606537S, RPC606559S, RPC607760S, RPC608041S, RPC601946S, RPC603068S, RPC604283S, RPC605416S, RPC606179S, RPC601212S, RPC601576S, RPC606737S, RPC605139S, RPC600253S, RPC605276S, RPC605363S, RPC605365S, RPC605373S, RPC605715S, RPC605744S, RPC607370S, RPC607401S, RPC607582S, RPC604399S, RPC605857S, RPC606323S, RPC606698S, RPC606712S, RPC607329S, RPC607699S, RPC607807S, RPC607880S, RPC607971S, RPC608065S, RPC608107S, RPC603301S, RPC603521S, RPC604531S, RPC603536S, RPC603537S, RPC603539S, RPC603544S, RPC603545S, RPC604932S, RPC605060S, RPC605096S, RPC605326S, RPC605377S, RPC605397S, RPC605681S, RPC605682S, RPC605686S, RPC605763S, RPC605883S, RPC606242S, RPC606246S, RPC606313S, RPC606388S, RPC606434S, RPC606435S, RPC606446S, RPC606491S, RPC606662S, RPC606761S, RPC606927S, RPC606929S, RPC606930S, RPC606931S, RPC607034S, RPC607403S, RPC607575S, RPC607752S, RPC607894S, RPC607926S, RPC607973S, RPC607993S, RPC608024S, RPC608108S, RPC608194S, RPC600516S, RPC600532S, RPC600555S, RPC600557S, RPC600597S, RPC600627S, RPC600660S, RPC600731S, RPC600871S, RPC600886S, RPC600921S, RPC600937S, RPC601087S, RPC601107S, RPC601315S, RPC601360S, RPC601511S, RPC601538S, RPC601616S, RPC601642S, RPC601660S, RPC601892S, RPC602047S, RPC602108S, RPC602127S, RPC602137S, RPC602168S, RPC602173S, RPC602192S, RPC602268S, RPC602337S, RPC602339S, RPC602359S, RPC602373S, RPC602460S, RPC602512S, RPC603117S, RPC603216S, RPC603409S, RPC603530S, RPC603535S, RPC603541S, RPC603779S, RPC603907S, RPC604013S, RPC604343S, RPC604420S, RPC604640S, RPC604649S, RPC604724S, RPC604756S, RPC604760S, RPC604761S, RPC604802S, RPC606928S, RPC606760S, RPC608312S, RPC603810S, RPC604289S, RPC607845S, RPC608076S, RPC608238S, RPC604736S, RPC607569S, RPC601171S, RPC602690S, RPC603783S, RPC604132S, RPC600441S, RPC600588S, RPC600966S, RPC601064S, RPC601080S, RPC601216S, RPC601300S, RPC601686S, RPC601896S, RPC602094S, RPC602243S, RPC602331S, RPC602502S, RPC602689S, RPC603842S, RPC604373S, RPC606084S, RPC605344S, RPC605654S, RPC605529S, RPC604753S, RPC606124S, RPC600517S, RPC603823S, RPC604145S, RPC604150S, RPC600140S, RPC600174S, RPC606300S, RPC603696S, RPC600528S, RPC605275S, RPC605278S, RPC605279S, RPC600449S, RPC600544S, RPC600577S, RPC600773S, RPC601182S, RPC601312S, RPC601408S, RPC601823S, RPC601937S, RPC601974S, RPC602070S, RPC602110S, RPC602433S, RPC602439S, RPC602440S, RPC602497S, RPC602724S, RPC602860S, RPC602867S, RPC603428S, RPC607648S, RPC608140S, RPC607253S, RPC605771S, RPC600692S, RPC600969S, RPC604381S, RPC604486S, RPC605481S, RPC600680S, RPC605135S, RPC602316S, RPC604262S, RPC607874S, RPC607876S, RPC600372S, RPC604220S, RPC604361S, RPC602207S, RPC600420S, RPC606860S, RPC606861S, RPC606678S, RPC605066S, RPC605241S, RPC606194S, RPC606195S, RPC606416S, RPC606555S, RPC607223S, RPC600090S, RPC600794S, RPC602228S, RPC603146S, RPC604310S, RPC604312S, RPC607879S, RPC600889S, RPC606492S, RPC606926S, RPC600398S, RPC603968S, RPC604703S, RPC604708S, RPC607520S, RPC604995S, RPC607190S, RPC601891S, RPC602820S, RPC605463S, RPC606513S, RPC606650S, RPC607498S, RPC607499S, RPC607745S, RPC607746S, RPC602367S, RPC602805S, RPC603130S, RPC603621S, RPC603676S, RPC604406S, RPC604634S, RPC604808S, RPC603020S, RPC604525S, RPC600151S, RPC600515S, RPC600618S, RPC600620S, RPC604951S, RPC605038S, RPC605865S, RPC600015S, RPC600667S, RPC600686S, RPC601330S, RPC601752S, RPC602244S, RPC603138S, RPC603320S, RPC606971S, RPC601772S, RPC603339S, RPC603882S, RPC602465S, RPC606763S, RPC604651S, RPC604871S, RPC604897S, RPC604959S, RPC606563S, RPC606589S, RPC600611S, RPC600622S, RPC604117S, RPC604582S, RPC606581S, RPC604031S, RPC604032S, RPC604105S, RPC604386S, RPC604387S, RPC606582S, RPC607536S, RPC601259S, RPC604050S, RPC604144S, RPC604862S, RPC605406S, RPC606642S, RPC606653S, RPC606750S, RPC607559S, RPC602598S, RPC603568S, RPC603573S, RPC603820S, RPC604083S, RPC604665S, RPC605360S, RPC602905S, RPC603553S, RPC603555S, RPC606914S, RPC605222S, RPC605265S, RPC605407S, RPC601900S, RPC601954S, RPC602008S, RPC602014S, RPC602167S, RPC602688S, RPC604673S, RPC605017S, RPC605691S, RPC600684S, RPC603785S, RPC604125S, RPC604338S, RPC604758S, RPC603256S, RPC603417S, RPC604065S, RPC604363S, RPC605910S, RPC606870S, RPC607301S, RPC607605S, RPC600724S, RPC602680S, RPC602777S, RPC602830S, RPC602843S, RPC603110S, RPC603788S, RPC601281S, RPC601248S, RPC602632S, RPC602652S, RPC603372S, RPC603373S, RPC606255S, RPC602317S, RPC602890S, RPC602899S, RPC603584S, RPC602037S, RPC607107S, RPC607528S, RPC601806S, RPC601995S, RPC602083S, RPC602438S, RPC602719S, RPC603797S, RPC604349S, RPC601521S, RPC606735S, RPC606738S, RPC606739S, RPC606740S, RPC606741S, RPC604904S, RPC605393S, RPC605395S, RPC605396S, RPC605411S, RPC605761S, RPC605839S, RPC606011S, RPC606721S, RPC600662S, RPC601282S, RPC602658S, RPC603066S, RPC603554S, RPC604033S, RPC604404S, RPC604405S, RPC604810S, RPC608131S, RPC606062S, RPC602851S, RPC604650S, RPC605474S, RPC605841S, RPC606057S, RPC606980S, RPC606981S, RPC607222S, RPC607260S, RPC607287S, RPC601250S, RPC601550S, RPC601589S, RPC601977S, RPC602196S, RPC603579S, RPC603631S, RPC604030S, RPC604641S, RPC600284S, RPC606207S, RPC607591S, RPC604216S, RPC600456S, RPC601161S, RPC602019S, RPC600252S, RPC605332S, RPC605349S, RPC605540S, RPC605680S, RPC607784S, RPC600453S, RPC600949S, RPC601005S, RPC601210S, RPC601381S, RPC601399S, RPC601529S, RPC601972S, RPC601973S, RPC602938S, RPC602999S, RPC603192S, RPC603763S, RPC604426S, RPC601438S, RPC603250S, RPC604209S, RPC604681S, RPC605205S, RPC605552S, RPC601592S, RPC602372S, RPC602648S, RPC603503S, RPC600044S, RPC600156S, RPC601572S, RPC601841S, RPC602379S, RPC602912S, RPC602986S, RPC603350S, RPC605317S, RPC605536S, RPC602941S, RPC604435S, RPC604441S, RPC602837S, RPC604130S, RPC601353S, RPC600311S, RPC600481S, RPC607335S, RPC607495S, RPC607743S, RPC607930S, RPC607987S, RPC600289S, RPC600294S, RPC601294S, RPC602798S, RPC605118S, RPC605693S, RPC605845S, RPC606538S, RPC602181S, RPC603001S, RPC604380S, RPC605180S, RPC600589S, RPC602891S, RPC607508S, RPC607718S, RPC600092S, RPC603398S, RPC603697S, RPC607658S, RPC604894S, RPC602054S, RPC604591S, RPC605546S, RPC605558S, RPC601736S, RPC605198S, RPC605387S, RPC606336S, RPC606703S, RPC600689S, RPC601020S, RPC601220S, RPC601738S, RPC602868S, RPC603750S, RPC603846S, RPC603875S, RPC605072S, RPC605356S, RPC605384S, RPC605880S, RPC605881S, RPC606652S, RPC607391S, RPC601280S, RPC601486S, RPC601567S, RPC601689S, RPC601706S, RPC601729S, RPC601802S, RPC602156S, RPC602420S, RPC602485S, RPC602721S, RPC602902S, RPC603019S, RPC603226S, RPC604433S, RPC604464S, RPC604465S, RPC604598S, RPC606244S, RPC601784S, RPC603288S, RPC603995S, RPC606941S, RPC602766S, RPC604701S, RPC603387S, RPC605962S, RPC606637S, RPC601698S, RPC602612S, RPC602927S, RPC603431S, RPC604675S, RPC605014S, RPC605408S, RPC605560S, RPC605844S, RPC600256S, RPC601041S, RPC601078S, RPC601083S, RPC601695S, RPC601827S, RPC602084S, RPC603437S, RPC603604S, RPC604344S, RPC604483S, RPC606672S, RPC607005S, RPC606138S, RPC600742S, RPC605207S, RPC605266S, RPC600575S, RPC601070S, RPC601071S, RPC601072S, RPC601166S, RPC601502S, RPC601763S, RPC601767S, RPC602559S, RPC605972S, RPC607174S, RPC607364S, RPC601234S, RPC601348S, RPC601588S, RPC603656S, RPC607795S, RPC605331S, RPC605904S, RPC606525S, RPC606529S, RPC606682S, RPC606695S, RPC606856S, RPC601994S, RPC602302S, RPC602307S, RPC603469S, RPC603531S, RPC603814S, RPC603837S, RPC603892S, RPC604196S, RPC604246S, RPC603799S, RPC607381S, RPC600973S, RPC601341S, RPC601934S, RPC604111S, RPC604829S, RPC602176S, RPC606230S, RPC602544S, RPC608310S, RPC605812S, RPC602007S, RPC603408S, RPC604322S, RPC604920S, RPC602803S, RPC603148S, RPC603397S, RPC606337S, RPC607082S, RPC607119S, RPC607247S, RPC600339S, RPC604388S, RPC604706S, RPC600400S, RPC600402S, RPC601768S, RPC602550S, RPC602604S, RPC603225S, RPC603240S, RPC603420S, RPC600391S, RPC600392S, RPC600393S, RPC600394S, RPC602661S, RPC602666S, RPC603434S, RPC603512S, RPC603613S, RPC604964S, RPC604977S, RPC605318S, RPC605319S, RPC600297S, RPC603194S, RPC605646S, RPC604253S, RPC607346S, RPC600276S, RPC600336S, RPC604401S, RPC605867S, RPC606317S, RPC600847S, RPC602750S, RPC603498S, RPC608037S, RPC601178S, RPC604019S, RPC605617S, RPC600454S, RPC600808S, RPC600857S, RPC601255S, RPC601571S, RPC602409S, RPC600844S, RPC601801S, RPC602450S, RPC604305S, RPC600796S, RPC603470S, RPC603471S, RPC603473S, RPC604658S, RPC606583S, RPC607696S, RPC601667S, RPC601724S, RPC602179S, RPC603346S, RPC603361S, RPC603607S, RPC601197S, RPC601548S, RPC601313S, RPC602200S, RPC602398S, RPC602807S, RPC603199S, RPC607258S, RPC603879S, RPC600912S, RPC600913S, RPC606545S, RPC602801S, RPC605610S, RPC605708S, RPC605832S, RPC607537S, RPC601168S, RPC602267S, RPC602399S, RPC603145S, RPC604629S, RPC604747S, RPC604104S, RPC604578S, RPC600408S, RPC600409S, RPC600410S, RPC606454S, RPC603472S, RPC607603S, RPC607756S, RPC605182S, RPC605364S, RPC606305S, RPC606306S, RPC606960S, RPC607375S, RPC607581S, RPC607755S, RPC603451S, RPC603809S, RPC607737S, RPC607917S, RPC606523S, RPC600290S, RPC600463S, RPC600861S, RPC605359S, RPC605504S, RPC605849S, RPC606003S, RPC606227S, RPC606283S, RPC603182S, RPC603359S, RPC600327S, RPC601778S, RPC604939S, RPC604940S, RPC604941S, RPC605353S, RPC605428S, RPC605711S, RPC605786S, RPC605846S, RPC606093S, RPC606145S, RPC606146S, RPC606859S, RPC607093S, RPC607167S, RPC607173S, RPC607214S, RPC607446S, RPC607541S, RPC607542S, RPC600086S, RPC600158S, RPC600286S, RPC601223S, RPC601335S, RPC602814S, RPC602816S, RPC603074S, RPC603547S, RPC603791S, RPC603967S, RPC604071S, RPC604143S, RPC604238S, RPC604250S, RPC604494S, RPC607540S, RPC607130S, RPC601673S, RPC601864S, RPC608301S, RPC608302S, RPC608303S, RPC608304S, RPC606674S, RPC606681S, RPC607226S, RPC601599S, RPC601760S, RPC601875S, RPC602034S, RPC602624S, RPC602765S, RPC602823S, RPC602936S, RPC603292S, RPC603293S, RPC607225S, RPC607227S, RPC607716S, RPC600340S, RPC600842S, RPC603749S, RPC607518S, RPC607709S, RPC600612S, RPC601112S, RPC600177S, RPC602626S, RPC603768S, RPC604436S, RPC601849S, RPC603286S, RPC604069S, RPC606483S, RPC600247S, RPC602717S, RPC604423S, RPC606137S, RPC601688S, RPC602953S, RPC604594S, RPC606285S, RPC601463S, RPC603563S, RPC604697S, RPC606111S, RPC601363S, RPC602334S, RPC603808S, RPC600819S, RPC602607S, RPC602722S, RPC603039S, RPC603934S, RPC603516S, RPC604927S, RPC605029S, RPC605466S, RPC605741S, RPC606223S, RPC607028S, RPC607256S, RPC600656S, RPC600785S, RPC600800S, RPC600820S, RPC600982S, RPC600986S, RPC601052S, RPC601243S, RPC601560S, RPC601965S, RPC601967S, RPC602189S, RPC602654S, RPC602903S, RPC603300S, RPC603633S, RPC603677S, RPC603858S, RPC604122S, RPC604479S, RPC604608S, RPC608248S, RPC602723S, RPC606225S, RPC606873S, RPC600774S, RPC601594S, RPC601717S, RPC603088S, RPC603366S, RPC606872S, RPC606602S, RPC605598S, RPC601045S, RPC601913S, RPC602064S, RPC602345S, RPC602623S, RPC603332S, RPC606603S, RPC603402S, RPC605922S, RPC605923S, RPC606597S, RPC606877S, RPC601898S, RPC602315S, RPC603618S, RPC606805S, RPC607574S, RPC601503S, RPC602015S, RPC602159S, RPC602920S, RPC602922S, RPC603318S, RPC603991S, RPC607166S, RPC605210S, RPC600337S, RPC602397S, RPC603642S, RPC604008S, RPC604190S, RPC605236S, RPC607350S, RPC600405S, RPC600641S, RPC605308S, RPC605854S, RPC605882S, RPC606374S, RPC606479S, RPC606540S, RPC604315S, RPC604400S, RPC605804S, RPC606355S, RPC606707S, RPC607938S, RPC602486S, RPC604687S, RPC605108S, RPC605212S, RPC605695S, RPC605718S, RPC606964S, RPC607986S, RPC600221S, RPC600322S, RPC601021S, RPC601336S, RPC601541S, RPC601643S, RPC601674S, RPC602055S, RPC602769S, RPC602907S, RPC606150S, RPC606404S, RPC607104S, RPC607643S, RPC608010S, RPC600896S, RPC600965S, RPC601601S, RPC601735S, RPC601743S, RPC601759S, RPC602677S, RPC604003S, RPC604309S, RPC604520S, RPC607411S, RPC605712S, RPC606386S, RPC606496S, RPC606612S, RPC607373S, RPC602819S, RPC604205S, RPC604331S, RPC607561S, RPC600451S, RPC600566S, RPC602445S, RPC600617S, RPC600767S, RPC603884S, RPC605067S, RPC605051S, RPC606421S, RPC606687S, RPC607180S, RPC607181S, RPC607270S, RPC607550S, RPC607675S, RPC607828S, RPC608014S, RPC600640S, RPC600970S, RPC601153S, RPC602197S, RPC602291S, RPC602338S, RPC602562S, RPC602566S, RPC602748S, RPC602770S, RPC602775S, RPC602822S, RPC603048S, RPC603085S, RPC603308S, RPC603389S, RPC603446S, RPC603447S, RPC603548S, RPC603688S, RPC603695S, RPC603703S, RPC603751S, RPC603822S, RPC603853S, RPC603872S, RPC604679S, RPC605476S, RPC605898S, RPC606181S, RPC606417S, RPC606562S, RPC607276S, RPC607428S, RPC602390S, RPC604152S, RPC602521S, RPC602627S, RPC602631S, RPC602732S, RPC604166S, RPC604176S, RPC604522S, RPC601287S, RPC604853S, RPC605669S, RPC606275S, RPC600650S, RPC600743S, RPC601163S, RPC602023S, RPC602366S, RPC602464S, RPC602763S, RPC604552S, RPC603817S, RPC605692S, RPC605848S, RPC607115S, RPC600266S, RPC600810S, RPC600830S, RPC600943S, RPC601116S, RPC601123S, RPC601413S, RPC601765S, RPC601862S, RPC602012S, RPC602854S, RPC603064S, RPC603258S, RPC604038S, RPC604138S, RPC604139S, RPC604154S, RPC604603S, RPC604633S, RPC607729S, RPC605178S, RPC600147S, RPC600511S, RPC602098S, RPC603949S, RPC604838S, RPC604845S, RPC605634S, RPC607665S, RPC600638S, RPC603313S, RPC604824S, RPC600501S, RPC600542S, RPC601921S, RPC602078S, RPC602812S, RPC603043S, RPC605740S, RPC601510S, RPC603615S, RPC600036S, RPC600595S, RPC600761S, RPC601720S, RPC601903S, RPC602079S, RPC602143S, RPC602202S, RPC602352S, RPC600586S, RPC601658S, RPC601328S, RPC601344S, RPC601790S, RPC602676S, RPC602783S, RPC602848S, RPC602950S, RPC603330S, RPC603994S, RPC604120S, RPC605218S, RPC602641S, RPC602741S, RPC602749S, RPC600218S, RPC603918S, RPC606208S, RPC604320S, RPC602768S, RPC603232S, RPC603375S, RPC604028S, RPC603880S, RPC603013S, RPC606615S, RPC603630S, RPC606450S, RPC605414S, RPC606601S, RPC607248S, RPC607790S, RPC607851S, RPC608205S, RPC601203S, RPC601374S, RPC601479S, RPC602424S, RPC602523S, RPC604055S, RPC600537S, RPC607397S, RPC604842S, RPC605159S, RPC605209S, RPC605963S, RPC606343S, RPC606344S, RPC606345S, RPC606424S, RPC600590S, RPC600836S, RPC601325S, RPC601585S, RPC601783S, RPC602051S, RPC602056S, RPC602245S, RPC602248S, RPC602269S, RPC602577S, RPC602875S, RPC603078S, RPC603135S, RPC603238S, RPC603452S, RPC603581S, RPC604407S, RPC604778S, RPC607029S, RPC604011S, RPC601375S, RPC605559S, RPC600080S, RPC600593S, RPC600831S, RPC600864S, RPC600872S, RPC600923S, RPC602177S, RPC603921S, RPC604221S, RPC604222S, RPC604277S, RPC604635S, RPC603798S, RPC602725S, RPC602261S, RPC605747S, RPC607466S, RPC600203S, RPC600732S, RPC601676S, RPC602255S, RPC602546S, RPC604040S, RPC604805S, RPC608241S, RPC605372S, RPC605382S, RPC607358S, RPC607903S, RPC603950S, RPC604408S, RPC605086S, RPC605097S, RPC605098S, RPC605099S, RPC605793S, RPC606939S, RPC607502S, RPC600073S, RPC600260S, RPC600804S, RPC600837S, RPC600981S, RPC600991S, RPC603765S, RPC603767S, RPC607503S, RPC607504S, RPC604961S, RPC604966S, RPC604975S, RPC605106S, RPC605423S, RPC605596S, RPC605597S, RPC605706S, RPC606444S, RPC606477S, RPC606493S, RPC606494S, RPC607440S, RPC608019S, RPC601231S, RPC601405S, RPC602274S, RPC602584S, RPC602825S, RPC603112S, RPC603175S, RPC603585S, RPC603603S, RPC603754S, RPC603854S, RPC604360S, RPC604642S, RPC601566S, RPC605635S, RPC605787S, RPC606243S, RPC606660S, RPC606705S, RPC606888S, RPC607004S, RPC607159S, RPC600565S, RPC600863S, RPC601266S, RPC601617S, RPC601930S, RPC602322S, RPC602594S, RPC602969S, RPC603401S, RPC603998S, RPC604286S, RPC604772S, RPC605027S, RPC605305S, RPC605464S, RPC605545S, RPC605752S, RPC605947S, RPC607322S, RPC607727S, RPC601307S, RPC601749S, RPC602633S, RPC602697S, RPC602904S, RPC603004S, RPC603134S, RPC603482S, RPC604079S, RPC604080S, RPC608204S, RPC602885S, RPC600929S, RPC600963S, RPC601233S, RPC601257S, RPC601361S, RPC601561S, RPC602105S, RPC602292S, RPC602312S, RPC602371S, RPC602574S, RPC602846S, RPC602951S, RPC603511S, RPC604302S, RPC604304S, RPC607478S, RPC604655S, RPC607025S, RPC604887S, RPC605012S, RPC605058S, RPC605082S, RPC605102S, RPC605162S, RPC605164S, RPC605831S, RPC605890S, RPC606453S, RPC606458S, RPC606459S, RPC606754S, RPC606840S, RPC606853S, RPC607044S, RPC601138S, RPC602681S, RPC603014S, RPC603021S, RPC603022S, RPC603023S, RPC603571S, RPC603605S, RPC603811S, RPC603972S, RPC604127S, RPC604131S, RPC604136S, RPC604357S, RPC607018S, RPC607026S, RPC607867S, RPC607012S, RPC604164S, RPC606464S, RPC606706S, RPC607036S, RPC602500S, RPC604515S, RPC604684S, RPC606320S, RPC607249S, RPC601524S, RPC601815S, RPC601859S, RPC602026S, RPC602068S, RPC603486S, RPC605473S, RPC606220S, RPC600472S, RPC600705S, RPC603505S, RPC603520S, RPC606353S, RPC606647S, RPC600766S, RPC603018S, RPC603379S, RPC600033S, RPC600037S, RPC600387S, RPC600388S, RPC600471S, RPC600931S, RPC601577S, RPC601661S, RPC601719S, RPC601192S, RPC606403S, RPC606437S, RPC606924S, RPC607476S, RPC600386S, RPC604699S, RPC603953S, RPC600971S, RPC604161S, RPC604162S, RPC606535S, RPC606293S, RPC601957S, RPC601821S, RPC602565S, RPC604873S, RPC605374S, RPC601082S, RPC601152S, RPC607423S, RPC607424S, RPC607062S, RPC607395S, RPC600858S, RPC605348S, RPC605683S, RPC605731S, RPC604484S, RPC604835S, RPC608148S, RPC600124S, RPC605232S, RPC605233S, RPC605933S, RPC606148S, RPC606152S, RPC606204S, RPC606297S, RPC606329S, RPC606398S, RPC606425S, RPC606657S, RPC606658S, RPC606685S, RPC607271S, RPC607290S, RPC607330S, RPC607331S, RPC607712S, RPC600482S, RPC600483S, RPC600484S, RPC600485S, RPC600486S, RPC600489S, RPC601291S, RPC601306S, RPC601460S, RPC601465S, RPC601681S, RPC603508S, RPC603825S, RPC604160S, RPC604872S, RPC605619S, RPC605833S, RPC606333S, RPC606519S, RPC606570S, RPC607525S, RPC601878S, RPC602416S, RPC603315S, RPC603622S, RPC604285S, RPC608116S, RPC608117S, RPC606113S, RPC606715S, RPC607294S, RPC603364S, RPC607188S, RPC605119S, RPC605730S, RPC605855S, RPC606978S, RPC607184S, RPC607187S, RPC600216S, RPC600546S, RPC601596S, RPC601625S, RPC602380S, RPC603475S, RPC603480S, RPC607748S, RPC608135S, RPC601349S, RPC602549S, RPC603058S, RPC606338S, RPC606339S, RPC606340S, RPC606609S, RPC606726S, RPC606848S, RPC604156S, RPC608031S, RPC604740S, RPC600370S, RPC601895S, RPC600438S, RPC600636S, RPC602597S, RPC603038S, RPC603821S, RPC606379S, RPC605338S, RPC606081S, RPC607657S, RPC605390S, RPC606140S, RPC606247S, RPC606498S, RPC607629S, RPC600160S, RPC600928S, RPC600976S, RPC601610S, RPC601611S, RPC602529S, RPC602743S, RPC602874S, RPC603114S, RPC603574S, RPC603686S, RPC605026S, RPC605656S, RPC601025S, RPC603149S, RPC603422S, RPC605554S, RPC605738S, RPC606015S, RPC606724S, RPC607630S, RPC608093S, RPC600157S, RPC600243S, RPC600866S, RPC601552S, RPC601649S, RPC601882S, RPC602138S, RPC602396S, RPC602887S, RPC603490S, RPC603778S, RPC603801S, RPC603911S, RPC604450S, RPC604782S, RPC607673S, RPC607266S, RPC603381S, RPC603679S, RPC604620S, RPC601941S, RPC604489S, RPC604239S, RPC605426S, RPC602877S, RPC601038S, RPC603719S, RPC603529S, RPC606984S, RPC607578S, RPC601017S, RPC601169S, RPC601340S, RPC604330S, RPC604537S, RPC606098S, RPC606516S, RPC600747S, RPC601159S, RPC601507S, RPC601687S, RPC602191S, RPC602789S, RPC604049S, RPC604770S, RPC604800S, RPC605734S, RPC605751S, RPC600367S, RPC601109S, RPC601530S, RPC603569S, RPC603780S, RPC606507S, RPC607441S, RPC603908S, RPC603728S, RPC604195S, RPC604503S, RPC606241S, RPC606256S, RPC606239S, RPC602886S, RPC604199S, RPC604826S, RPC604860S, RPC604956S, RPC605183S, RPC606078S, RPC607580S, RPC607694S, RPC603065S, RPC603643S, RPC603847S, RPC604041S, RPC604106S, RPC604445S, RPC604643S, RPC601373S, RPC601456S, RPC601908S, RPC602100S, RPC604566S, RPC605215S, RPC601189S, RPC601481S, RPC603609S, RPC604469S, RPC607513S, RPC601347S, RPC604169S, RPC607264S, RPC605120S, RPC604467S, RPC606462S, RPC601084S, RPC604255S, RPC607010S, RPC604923S, RPC605158S, RPC605858S, RPC606864S, RPC606989S, RPC608002S, RPC602405S, RPC602508S, RPC602946S, RPC603035S, RPC603606S, RPC604264S, RPC604472S, RPC608095S, RPC605651S, RPC608125S, RPC603378S, RPC605366S, RPC604473S, RPC603289S, RPC604002S, RPC604677S, RPC608123S, RPC605398S, RPC600736S, RPC602401S, RPC604193S, RPC604233S, RPC605226S, RPC600779S, RPC601111S, RPC602418S, RPC605053S, RPC605343S, RPC606175S, RPC606341S, RPC600129S, RPC600856S, RPC602432S, RPC602444S, RPC602467S, RPC603336S, RPC600781S, RPC604061S, RPC601224S, RPC601411S, RPC601789S, RPC602490S, RPC603017S, RPC603098S, RPC603116S, RPC605584S, RPC604129S, RPC600596S, RPC605869S, RPC605870S, RPC601154S, RPC606436S, RPC601254S, RPC604266S, RPC600240S, RPC603466S, RPC604659S, RPC608284S, RPC603863S, RPC603864S, RPC603886S, RPC603938S, RPC603939S, RPC604093S, RPC604094S, RPC604201S, RPC602383S, RPC604412S, RPC606349S, RPC607455S, RPC601201S, RPC601417S, RPC601534S, RPC601751S, RPC602811S, RPC603737S, RPC604385S, RPC608028S, RPC601905S, RPC600860S, RPC603807S, RPC602488S, RPC606149S, RPC600875S, RPC601260S, RPC601303S, RPC602635S, RPC601404S, RPC605161S, RPC604968S, RPC600474S, RPC600610S, RPC604134S, RPC605148S, RPC606487S, RPC607218S, RPC607232S, RPC606592S, RPC607066S, RPC606442S, RPC601902S, RPC603731S, RPC606625S, RPC607999S, RPC604898S, RPC604929S, RPC604972S, RPC605128S, RPC605130S, RPC607089S, RPC607090S, RPC607726S, RPC602153S, RPC602570S, RPC602708S, RPC602716S, RPC603060S, RPC603234S, RPC603702S, RPC603942S, RPC604043S, RPC607724S, RPC602314S, RPC604843S, RPC607407S, RPC600977S, RPC601091S, RPC601342S, RPC602215S, RPC602992S, RPC603540S, RPC608085S, RPC606105S, RPC606089S, RPC606134S, RPC606999S, RPC603242S, RPC603249S, RPC604723S, RPC602236S, RPC604953S, RPC602931S, RPC600938S, RPC605629S, RPC606605S, RPC608075S, RPC600846S, RPC606213S, RPC603774S, RPC604276S, RPC604529S, RPC604901S, RPC605189S, RPC605303S, RPC605599S, RPC600196S, RPC600465S, RPC601297S, RPC602022S, RPC603418S, RPC603455S, RPC603620S, RPC603713S, RPC603745S, RPC603969S, RPC604077S, RPC604242S, RPC604398S, RPC606363S, RPC603178S, RPC603556S, RPC603915S, RPC606868S, RPC602501S, RPC605534S, RPC606619S, RPC600918S, RPC605358S, RPC601682S, RPC602149S, RPC600529S, RPC600754S, RPC601318S, RPC601737S, RPC603388S, RPC604046S, RPC605723S, RPC605724S, RPC602947S, RPC605698S, RPC605842S, RPC606394S, RPC600265S, RPC604268S, RPC604547S, RPC602519S, RPC604247S, RPC604368S, RPC605889S, RPC606701S, RPC602134S, RPC604075S, RPC607924S, RPC604994S, RPC604135S, RPC604886S, RPC602443S, RPC600270S, RPC601822S, RPC602031S, RPC600524S, RPC600567S, RPC600646S, RPC600894S, RPC601636S, RPC601732S, RPC600305S, RPC600034S, RPC604922S, RPC604962S, RPC604982S, RPC605022S, RPC605105S, RPC605538S, RPC606068S, RPC600874S, RPC601766S, RPC602183S, RPC603349S, RPC603467S, RPC603744S, RPC603978S, RPC604146S, RPC604548S, RPC604819S, RPC603124S, RPC605759S, RPC600826S, RPC600078S, RPC603391S, RPC605214S, RPC602588S, RPC602858S, RPC604252S, RPC608215S, RPC608216S, RPC608217S, RPC606539S, RPC607231S, RPC600263S, RPC601392S, RPC603948S, RPC606629S, RPC606463S, RPC601283S, RPC601787S, RPC603815S, RPC605961S, RPC606258S, RPC606262S, RPC606263S, RPC607434S, RPC608120S, RPC602790S, RPC604462S, RPC608064S, RPC605430S, RPC608023S, RPC600013S, RPC600152S, RPC600809S, RPC600849S, RPC608269S, RPC606972S, RPC605862S, RPC604885S, RPC606467S, RPC607589S, RPC607706S, RPC604227S, RPC603371S, RPC603629S, RPC605618S, RPC606451S, RPC606734S, RPC601482S, RPC602687S, RPC600081S, RPC600082S, RPC600083S, RPC605063S, RPC605122S, RPC605513S, RPC606314S, RPC607197S, RPC607980S, RPC600514S, RPC601096S, RPC601472S, RPC601515S, RPC602009S, RPC602011S, RPC602495S, RPC603239S, RPC604695S, RPC604795S, RPC605594S, RPC603805S, RPC605576S, RPC602726S, RPC603668S, RPC604524S, RPC607562S, RPC605087S, RPC600155S, RPC601606S, RPC603304S, RPC603766S, RPC604804S, RPC607942S, RPC600178S, RPC602395S, RPC605796S, RPC606517S, RPC606690S, RPC600687S, RPC600835S, RPC600944S, RPC600948S, RPC601134S, RPC601323S, RPC601350S, RPC601407S, RPC601415S, RPC601725S, RPC602044S, RPC602046S, RPC602151S, RPC602186S, RPC602201S, RPC602223S, RPC602355S, RPC602423S, RPC603157S, RPC603435S, RPC603589S, RPC604333S, RPC604334S, RPC604335S, RPC604336S, RPC607274S, RPC607177S, RPC607336S, RPC607597S, RPC607598S, RPC603683S, RPC605151S, RPC605171S, RPC605828S, RPC606415S, RPC604523S, RPC605836S, RPC606289S, RPC607027S, RPC600756S, RPC600852S, RPC604095S, RPC607840S, RPC607841S, RPC607852S, RPC607853S, RPC607854S, RPC605779S, RPC606574S, RPC602341S, RPC604470S, RPC604605S, RPC606661S, RPC607092S, RPC607492S, RPC600150S, RPC604934S, RPC605727S, RPC606983S, RPC607449S, RPC602600S, RPC603312S, RPC604700S, RPC604714S, RPC603832S, RPC605111S, RPC604096S, RPC605590S, RPC605277S, RPC606094S, RPC606099S, RPC606321S, RPC606548S, RPC608186S, RPC605101S, RPC605185S, RPC605196S, RPC605243S, RPC606226S, RPC606327S, RPC601539S, RPC601983S, RPC603781S, RPC604636S, RPC604691S, RPC604752S, RPC606471S, RPC608157S, RPC604883S, RPC600306S, RPC601635S, RPC602919S, RPC604099S, RPC606277S, RPC600335S, RPC602821S, RPC606470S, RPC600765S, RPC601156S, RPC601777S, RPC602304S, RPC602782S, RPC603183S, RPC603419S, RPC600666S, RPC604794S, RPC605041S, RPC605930S, RPC601825S, RPC606280S, RPC606643S, RPC606762S, RPC606897S, RPC602694S, RPC603522S, RPC603616S, RPC603715S, RPC603775S, RPC603802S, RPC604263S, RPC604358S, RPC604899S, RPC605916S, RPC606096S, RPC600522S, RPC600775S, RPC604366S, RPC604607S, RPC604609S, RPC607823S, RPC604889S, RPC605032S, RPC602369S, RPC603369S, RPC603462S, RPC603558S, RPC603747S, RPC603909S, RPC604177S, RPC600932S, RPC607016S, RPC607612S, RPC601015S, RPC602254S, RPC605863S, RPC604952S, RPC607601S, RPC602050S, RPC600806S, RPC601040S, RPC601782S, RPC603623S, RPC603626S, RPC603627S, RPC603628S, RPC604007S, RPC601270S, RPC604421S, RPC604788S, RPC605511S, RPC606683S, RPC600329S, RPC600608S, RPC600885S, RPC600905S, RPC601384S, RPC603687S, RPC606071S, RPC600261S, RPC601824S, RPC603497S, RPC600712S, RPC603325S, RPC604226S, RPC604613S, RPC604993S, RPC601936S, RPC603960S, RPC604179S, RPC605872S, RPC607003S, RPC602075S, RPC602305S, RPC602491S, RPC606422S, RPC603758S, RPC606944S, RPC605797S, RPC605798S, RPC606234S, RPC606751S, RPC606757S, RPC606758S, RPC600087S, RPC603099S, RPC603523S, RPC603525S, RPC603526S, RPC607413S, RPC607899S, RPC600475S, RPC600476S, RPC601569S, RPC602021S, RPC604863S, RPC606979S, RPC607608S, RPC603943S, RPC604583S, RPC600793S, RPC600985S, RPC601214S, RPC603977S, RPC604850S, RPC605707S, RPC606191S, RPC607549S, RPC600258S, RPC603916S, RPC602382S, RPC605123S, RPC604362S, RPC604769S, RPC605939S, RPC601206S, RPC601238S, RPC605168S, RPC605085S, RPC605690S, RPC604397S, RPC604954S, RPC605290S, RPC605503S, RPC605524S, RPC606380S, RPC607406S, RPC601484S, RPC601563S, RPC602123S, RPC602272S, RPC602634S, RPC607409S, RPC607753S, RPC603946S, RPC602692S, RPC605519S, RPC601115S, RPC602419S, RPC603152S, RPC604126S, RPC605658S, RPC606881S, RPC607892S, RPC607763S, RPC600148S, RPC600900S, RPC604793S, RPC606368S, RPC606370S, RPC606869S, RPC606963S, RPC607216S, RPC607279S, RPC607947S, RPC601909S, RPC602112S, RPC603587S, RPC600199S, RPC600535S, RPC601033S, RPC601252S, RPC601829S, RPC602832S, RPC602952S, RPC604319S, RPC607091S, RPC602473S, RPC602916S, RPC603040S, RPC605145S, RPC605700S, RPC606732S, RPC603321S, RPC603828S, RPC604460S, RPC600098S, RPC600815S, RPC602839S, RPC604917S, RPC605044S, RPC600961S, RPC601723S, RPC602321S, RPC602412S, RPC603164S, RPC604632S, RPC607362S, RPC605622S, RPC605664S, RPC606573S, RPC606593S, RPC600940S, RPC604100S, RPC604801S, RPC605573S, RPC605705S, RPC605888S, RPC606476S, RPC605076S, RPC605144S, RPC605199S, RPC605200S, RPC605592S, RPC605593S, RPC606085S, RPC606139S, RPC606236S, RPC606364S, RPC606371S, RPC600518S, RPC600519S, RPC600527S, RPC601059S, RPC601141S, RPC601209S, RPC601912S, RPC602758S, RPC603233S, RPC603326S, RPC603790S, RPC603924S, RPC604225S, RPC604532S, RPC605736S, RPC603046S, RPC606970S, RPC607183S, RPC600166S, RPC600681S, RPC605600S, RPC606271S, RPC603997S, RPC606308S, RPC606985S, RPC604622S, RPC601798S, RPC604707S, RPC607292S, RPC605090S, RPC600502S, RPC600799S, RPC601095S, RPC601956S, RPC602128S, RPC605701S, RPC603844S, RPC606178S, RPC606536S, RPC606947S, RPC600490S, RPC600814S, RPC601229S, RPC601663S, RPC601670S, RPC602656S, RPC602696S, RPC602824S, RPC603136S, RPC603195S, RPC603370S, RPC603897S, RPC604207S, RPC604328S, RPC607359S, RPC605675S, RPC605676S, RPC605766S, RPC600834S, RPC602028S, RPC602434S, RPC603012S, RPC603047S, RPC604035S, RPC604210S, RPC604261S, RPC604325S, RPC604443S, RPC607121S, RPC601578S, RPC603461S, RPC604570S, RPC607457S, RPC600127S, RPC600786S, RPC602764S, RPC603883S, RPC600717S, RPC603856S, RPC606260S, RPC600550S, RPC600933S, RPC601002S, RPC604347S, RPC606060S, RPC604865S, RPC604974S, RPC602751S, RPC603488S, RPC605061S, RPC605116S, RPC606189S, RPC601322S, RPC601491S, RPC601804S, RPC601943S, RPC602862S, RPC603716S, RPC606004S, RPC604713S, RPC603900S, RPC605268S, RPC600057S, RPC600165S, RPC600279S, RPC604543S, RPC603394S, RPC603430S, RPC606302S, RPC605539S, RPC600749S, RPC605310S, RPC605677S, RPC605678S, RPC600561S, RPC601439S, RPC605149S, RPC600338S, RPC604855S, RPC604864S, RPC604946S, RPC605329S, RPC605564S, RPC605585S, RPC605696S, RPC605810S, RPC605940S, RPC606155S, RPC606156S, RPC606222S, RPC606318S, RPC606335S, RPC606369S, RPC606401S, RPC606553S, RPC606921S, RPC600138S, RPC600450S, RPC600512S, RPC600623S, RPC600628S, RPC600639S, RPC600784S, RPC601120S, RPC601249S, RPC601356S, RPC601357S, RPC601622S, RPC601672S, RPC601710S, RPC601846S, RPC601933S, RPC602259S, RPC602361S, RPC602470S, RPC602503S, RPC602672S, RPC602720S, RPC603229S, RPC603230S, RPC603243S, RPC603244S, RPC603334S, RPC603546S, RPC603593S, RPC603608S, RPC603632S, RPC603709S, RPC603843S, RPC604163S, RPC604183S, RPC604206S, RPC604332S, RPC604425S, RPC604676S, RPC604755S, RPC604767S, RPC604771S, RPC606541S, RPC600968S, RPC603602S, RPC604311S, RPC604625S, RPC608077S, RPC606252S, RPC600101S, RPC600161S, RPC604841S, RPC606730S, RPC607212S, RPC600753S, RPC601474S, RPC601495S, RPC601704S, RPC603524S, RPC603528S, RPC607213S, RPC605368S, RPC601274S, RPC602039S, RPC604979S, RPC605775S, RPC605800S, RPC605809S, RPC606590S, RPC607386S, RPC600065S, RPC600191S, RPC601346S, RPC601446S, RPC601490S, RPC601935S, RPC602125S, RPC602250S, RPC602251S, RPC603061S, RPC603235S, RPC603280S, RPC603281S, RPC603701S, RPC603985S, RPC604036S, RPC604274S, RPC604275S, RPC604308S, RPC604774S, RPC608122S, RPC604491S, RPC604930S, RPC605081S, RPC606497S, RPC604528S, RPC608098S, RPC604726S, RPC603440S, RPC603829S, RPC605389S, RPC605941S, RPC605943S, RPC606288S, RPC601062S, RPC603177S, RPC601047S, RPC603296S, RPC603849S, RPC608035S, RPC605908S, RPC605975S, RPC606180S, RPC606228S, RPC606232S, RPC606717S, RPC606718S, RPC606719S, RPC606753S, RPC606953S, RPC606955S, RPC606958S, RPC607278S, RPC607554S, RPC602106S, RPC602139S, RPC602166S, RPC602170S, RPC603107S, RPC603245S, RPC603395S, RPC603930S, RPC604254S, RPC604316S, RPC604475S, RPC604797S, RPC606954S, RPC607280S, RPC607555S, RPC607764S, RPC607765S, RPC607766S, RPC607767S, RPC607768S, RPC607769S, RPC605100S, RPC605978S, RPC606366S, RPC600208S, RPC600209S, RPC600210S, RPC601568S, RPC601810S, RPC603132S, RPC603273S, RPC603936S, RPC604217S, RPC604219S, RPC604477S, RPC604662S, RPC605147S, RPC602833S, RPC602970S, RPC603054S, RPC601929S, RPC602942S, RPC606474S, RPC606505S, RPC604918S, RPC600167S, RPC601387S, RPC603143S, RPC604097S, RPC605729S, RPC606656S, RPC601893S, RPC602895S, RPC603327S, RPC604948S, RPC605238S, RPC605347S, RPC605660S, RPC605746S, RPC605829S, RPC606332S, RPC606399S, RPC606423S, RPC606500S, RPC606621S, RPC606633S, RPC606887S, RPC607096S, RPC607286S, RPC607826S, RPC607849S, RPC600038S, RPC600068S, RPC600447S, RPC602493S, RPC602703S, RPC602831S, RPC603760S, RPC603761S, RPC604259S, RPC604498S, RPC605089S, RPC601382S, RPC602817S, RPC603161S, RPC603185S, RPC603193S, RPC603491S, RPC603494S, RPC604492S, RPC604618S, RPC606606S, RPC606890S, RPC605138S, RPC607360S, RPC607744S, RPC600739S, RPC600995S, RPC602442S, RPC603851S, RPC603855S, RPC603996S, RPC604119S, RPC605362S, RPC606324S, RPC607040S, RPC607194S, RPC607195S, RPC608153S, RPC602756S, RPC603824S, RPC606073S, RPC607560S, RPC600507S, RPC603100S, RPC604059S, RPC602702S, RPC602365S, RPC603299S, RPC603624S, RPC606400S, RPC601699S, RPC601978S, RPC603151S, RPC603645S, RPC603044S, RPC602016S, RPC601395S, RPC605509S, RPC602404S, RPC605769S, RPC604846S, RPC604911S, RPC605873S, RPC600679S, RPC601086S, RPC601497S, RPC604533S, RPC604733S, RPC605506S, RPC607587S, RPC600381S, RPC602033S, RPC603224S, RPC605613S, RPC606164S, RPC606206S, RPC606402S, RPC606410S, RPC606669S, RPC607031S, RPC600668S, RPC602017S, RPC602111S, RPC602362S, RPC605550S, RPC605685S, RPC605719S, RPC600657S, RPC600658S, RPC601110S, RPC601623S, RPC601739S, RPC601999S, RPC603097S, RPC604068S, RPC605788S, RPC606116S, RPC603947S, RPC607754S, RPC607599S, RPC604926S, RPC605211S, RPC606438S, RPC606884S, RPC606885S, RPC607144S, RPC604790S, RPC605030S, RPC607323S, RPC608086S, RPC600654S, RPC602531S, RPC602754S, RPC604039S, RPC604914S, RPC605376S, RPC602945S, RPC603841S, RPC605628S, RPC605732S, RPC602025S, RPC602085S, RPC602408S, RPC602911S, RPC603277S, RPC605956S, RPC605031S, RPC605035S, RPC605036S, RPC605065S, RPC605653S, RPC605914S, RPC606429S, RPC606522S, RPC606648S, RPC606909S, RPC606910S, RPC606911S, RPC600307S, RPC600443S, RPC600614S, RPC600669S, RPC600851S, RPC600899S, RPC600946S, RPC600992S, RPC601191S, RPC601423S, RPC601559S, RPC601703S, RPC601731S, RPC601750S, RPC602246S, RPC602771S, RPC603144S, RPC603253S, RPC603352S, RPC603565S, RPC603594S, RPC603662S, RPC603708S, RPC603986S, RPC604167S, RPC604191S, RPC604513S, RPC601310S, RPC606485S, RPC601758S, RPC602392S, RPC605921S, RPC607365S, RPC602061S, RPC602933S, RPC604468S, RPC605115S, RPC605133S, RPC605315S, RPC605505S, RPC605960S, RPC606205S, RPC606348S, RPC607324S, RPC602496S, RPC602556S, RPC603083S, RPC604016S, RPC604798S, RPC605441S, RPC605467S, RPC606354S, RPC601764S, RPC601814S, RPC602099S, RPC602370S, RPC603137S, RPC603355S, RPC603928S, RPC603944S, RPC604414S, RPC607282S, RPC608285S, RPC605046S, RPC605595S, RPC606936S, RPC600407S, RPC600952S, RPC601143S, RPC601305S, RPC601531S, RPC601697S, RPC602505S, RPC604010S, RPC607847S, RPC600709S, RPC602165S, RPC603197S, RPC601631S, RPC600379S, RPC604116S, RPC600545S, RPC601396S, RPC604882S, RPC601157S, RPC603495S, RPC604852S, RPC605472S, RPC605866S, RPC606993S, RPC607392S, RPC607494S, RPC607937S, RPC601160S, RPC601302S, RPC601742S, RPC601819S, RPC601872S, RPC604284S, RPC604384S, RPC604481S, RPC604538S, RPC603476S, RPC604536S, RPC604902S, RPC604969S, RPC605037S, RPC601558S, RPC604731S, RPC605361S, RPC603705S, RPC605609S, RPC605874S, RPC606279S, RPC606499S, RPC603513S, RPC603806S, RPC602700S, RPC605413S, RPC605924S, RPC606443S, RPC600179S, RPC600251S, RPC603641S, RPC603819S, RPC604171S, RPC604637S, RPC607083S, RPC602069S, RPC602126S, RPC603478S, RPC605117S, RPC604273S, RPC601955S, RPC603265S, RPC606481S, RPC600183S, RPC605830S, RPC604147S, RPC606287S, RPC602141S, RPC603534S, RPC604606S, RPC607522S, RPC604260S, RPC600822S, RPC601488S, RPC602357S, RPC603538S, RPC600268S, RPC603971S, RPC601700S, RPC600299S, RPC607913S, RPC600895S, RPC605535S, RPC604413S, RPC607102S, RPC608100S, RPC604858S, RPC601022S, RPC602381S, RPC607878S, RPC606646S, RPC606756S, RPC606876S, RPC600869S, RPC601018S, RPC602448S, RPC601628S, RPC607356S, RPC606445S, RPC600219S, RPC600416S, RPC601931S, RPC604817S, RPC605246S, RPC605247S, RPC605248S, RPC605249S, RPC605816S, RPC605817S, RPC605818S, RPC605819S, RPC606141S, RPC606185S, RPC606187S, RPC606190S, RPC606826S, RPC603246S, RPC603487S, RPC603684S, RPC604107S, RPC604108S, RPC604109S, RPC604110S, RPC604815S, RPC604816S, RPC604818S, RPC606186S, RPC606188S, RPC606192S, RPC605568S, RPC605811S, RPC601320S, RPC603016S, RPC603920S, RPC604579S, RPC604674S, RPC606214S, RPC607011S, RPC607075S, RPC602158S, RPC602802S, RPC603444S, RPC603787S, RPC604610S, RPC605333S, RPC602737S, RPC600715S, RPC600825S, RPC601265S, RPC601532S, RPC601816S, RPC601979S, RPC602313S, RPC602344S, RPC602522S, RPC602620S, RPC602774S, RPC603247S, RPC603248S, RPC603399S, RPC603492S, RPC603575S, RPC603590S, RPC604192S, RPC604971S, RPC605394S, RPC606012S, RPC607762S, RPC601740S, RPC602909S, RPC603927S, RPC604203S, RPC604204S, RPC604988S, RPC605156S, RPC605633S, RPC606129S, RPC607080S, RPC600215S, RPC600631S, RPC600817S, RPC601240S, RPC601339S, RPC601637S, RPC602275S, RPC602678S, RPC602856S, RPC603290S, RPC603484S, RPC604626S, RPC606169S, RPC607506S, RPC604732S, RPC604912S, RPC604981S, RPC601135S, RPC601858S, RPC603860S, RPC605500S, RPC605501S, RPC605502S, RPC605638S, RPC605639S, RPC605641S, RPC605642S, RPC605643S, RPC600373S, RPC600987S, RPC601010S, RPC601058S, RPC601099S, RPC601100S, RPC601321S, RPC601771S, RPC601964S, RPC602102S, RPC601098S, RPC605018S, RPC605258S, RPC605966S, RPC605967S, RPC605969S, RPC605970S, RPC606787S, RPC602040S, RPC602043S, RPC602169S, RPC602171S, RPC602172S, RPC602174S, RPC602175S, RPC602194S, RPC603158S, RPC603159S, RPC603165S, RPC603173S, RPC603174S, RPC603188S, RPC603190S, RPC603219S, RPC603221S, RPC603231S, RPC604025S, RPC605971S, RPC606773S, RPC606774S, RPC606775S, RPC606776S, RPC606777S, RPC606778S, RPC606779S, RPC606780S, RPC606781S, RPC606782S, RPC606783S, RPC606784S, RPC606785S, RPC606786S, RPC606788S, RPC606789S, RPC606790S, RPC606791S, RPC606792S, RPC604856S, RPC605152S, RPC605334S, RPC605648S, RPC600412S, RPC600421S, RPC600663S, RPC600670S, RPC601410S, RPC601513S, RPC602036S, RPC602368S, RPC602917S, RPC603067S, RPC603207S, RPC603357S, RPC604292S, RPC607209S, RPC607250S, RPC607668S, RPC607669S, RPC607739S, RPC607740S, RPC607855S, RPC607856S, RPC607857S, RPC607858S, RPC607859S, RPC607860S, RPC607861S, RPC607862S, RPC608160S, RPC608161S, RPC608162S, RPC608163S, RPC608164S, RPC608165S, RPC608166S, RPC608167S, RPC608168S, RPC608169S, RPC608171S, RPC608172S, RPC608173S, RPC608174S, RPC608175S, RPC608176S, RPC608177S, RPC608178S, RPC608179S, RPC605572S, RPC606906S, RPC601469S, RPC601263S, RPC601447S, RPC601800S, RPC602555S, RPC604933S, RPC603601S, RPC604062S, RPC605868S, RPC604427S, RPC604958S, RPC601162S, RPC601409S, RPC602472S, RPC604047S, RPC607100S, RPC603465S, RPC604437S, RPC604546S, RPC601239S, RPC601271S, RPC603489S, RPC604235S, RPC605684S, RPC605250S, RPC606347S, RPC606486S, RPC606508S, RPC606509S, RPC606689S, RPC606940S, RPC607872S, RPC607907S, RPC600534S, RPC601027S, RPC602462S, RPC603931S, RPC607952S, RPC607946S, RPC604837S, RPC604915S, RPC605528S, RPC605588S, RPC605627S, RPC605912S, RPC605945S, RPC606296S, RPC606827S, RPC600315S, RPC600346S, RPC600746S, RPC600935S, RPC601219S, RPC601904S, RPC602103S, RPC602364S, RPC602530S, RPC602785S, RPC603036S, RPC603126S, RPC603222S, RPC603267S, RPC604300S, RPC604377S, RPC604378S, RPC607563S, RPC602695S, RPC602606S, RPC601199S, RPC602499S, RPC606159S, RPC600445S, RPC606002S, RPC603328S, RPC604371S, RPC606696S, RPC607672S, RPC605612S, RPC601403S, RPC602994S, RPC603542S, RPC606793S, RPC607450S, RPC604710S, RPC607616S, RPC606882S, RPC606841S, RPC605652S, RPC608128S, RPC602545S, RPC605114S, RPC606177S, RPC607289S, RPC608057S, RPC606561S, RPC606221S, RPC601942S, RPC603340S, RPC604348S, RPC601019S, RPC606259S, RPC607988S, RPC603866S, RPC600182S, RPC605339S, RPC606301S, RPC605583S, RPC604590S, RPC605220S, RPC604474S, RPC604832S, RPC605659S, RPC605799S, RPC606133S, RPC606238S, RPC606560S, RPC606644S, RPC607215S, RPC607304S, RPC607655S, RPC607789S, RPC608001S, RPC608195S, RPC600419S, RPC601011S, RPC601639S, RPC601640S, RPC601991S, RPC601997S, RPC602711S, RPC602780S, RPC603062S, RPC603063S, RPC603354S, RPC603682S, RPC603903S, RPC604215S, RPC604507S, RPC604768S, RPC607887S, RPC608159S, RPC600418S, RPC602655S, RPC602901S, RPC604232S, RPC604989S, RPC602991S, RPC604375S, RPC604402S, RPC604848S, RPC604866S, RPC604906S, RPC606061S, RPC607474S, RPC600043S, RPC600055S, RPC600563S, RPC600698S, RPC601338S, RPC601508S, RPC601873S, RPC601877S, RPC602029S, RPC602575S, RPC603056S, RPC603770S, RPC603913S, RPC604159S, RPC605532S, RPC603257S, RPC603527S, RPC603550S, RPC607593S, RPC604224S, RPC605235S, RPC606121S, RPC603956S, RPC607060S, RPC607357S, RPC604891S, RPC604973S, RPC606387S, RPC607211S, RPC607452S, RPC607453S, RPC603374S, RPC604029S, RPC604596S, RPC605281S, RPC606564S, RPC605621S, RPC600237S, RPC601222S, RPC602590S, RPC603827S, RPC603236S, RPC605548S, RPC606863S, RPC606919S, RPC607465S, RPC607558S, RPC607698S, RPC608056S, RPC600763S, RPC601774S, RPC601928S, RPC602152S, RPC602761S, RPC602963S, RPC604694S, RPC604746S, RPC606381S, RPC605602S, RPC606158S, RPC601378S, RPC601520S, RPC602625S, RPC603343S, RPC601354S, RPC602252S, RPC606077S, RPC603699S, RPC603755S, RPC603793S, RPC603833S, RPC605206S, RPC605482S, RPC605932S, RPC606591S, RPC601106S, RPC601780S, RPC601788S, RPC603499S, RPC604012S, RPC604738S, RPC605259S, RPC605699S, RPC603259S, RPC604449S, RPC604624S, RPC601844S, RPC605386S, RPC605915S, RPC607103S, RPC607566S, RPC601799S, RPC600102S, RPC600103S, RPC600104S, RPC600106S, RPC600107S, RPC600108S, RPC600111S, RPC606816S, RPC606866S, RPC607098S, RPC607927S, RPC607326S, RPC606199S, RPC601376S, RPC605520S, RPC605900S, RPC605475S, RPC607547S, RPC600012S, RPC604070S, RPC604671S, RPC605486S, RPC600195S, RPC602333S, RPC606059S, RPC600790S, RPC600205S, RPC600206S, RPC600211S, RPC607299S, RPC607512S, RPC603403S, RPC604488S, RPC605316S, RPC606452S, RPC604759S, RPC605071S, RPC607507S, RPC600303S, RPC604612S, RPC602943S, RPC603261S, RPC600543S, RPC603876S, RPC603493S, RPC605512S, RPC601540S, RPC603502S, RPC605239S, RPC604459S, RPC604639S, RPC607343S, RPC608121S, RPC604949S, RPC605229S, RPC606504S, RPC600301S, RPC601730S, RPC604314S, RPC601272S, RPC604955S, RPC606107S, RPC606117S, RPC606973S, RPC600735S, RPC601683S, RPC601865S, RPC603027S, RPC604685S, RPC604688S, RPC600029S, RPC600030S, RPC605137S, RPC605391S, RPC605551S, RPC605913S, RPC602115S, RPC602402S, RPC602596S, RPC602939S, RPC603559S, RPC600149S, RPC600573S, RPC601914S, RPC602295S, RPC605778S, RPC601092S, RPC601741S, RPC602428S, RPC603335S, RPC603922S, RPC600426S, RPC605197S, RPC605567S, RPC605965S, RPC606294S, RPC607179S, RPC600750S, RPC602097S, RPC602827S, RPC602906S, RPC603010S, RPC603094S, RPC604821S, RPC600250S, RPC603955S, RPC602799S, RPC604140S, RPC600559S, RPC603896S, RPC604243S, RPC602260S, RPC606097S, RPC606725S, RPC601389S, RPC602258S, RPC602507S, RPC603905S, RPC601054S, RPC601029S, RPC601049S, RPC601056S, RPC601000S, RPC601028S, RPC601068S, RPC601483S, RPC601657S, RPC601794S, RPC601848S, RPC601851S, RPC602045S, RPC602080S, RPC602155S, RPC600505S, RPC601196S, RPC601702S, RPC602587S, RPC602118S, RPC602121S, RPC602185S, RPC602232S, RPC602318S, RPC602320S, RPC602602S, RPC602609S, RPC602610S, RPC602617S, RPC603131S, RPC603654S, RPC603660S, RPC603792S, RPC604327S, RPC605717S, RPC600633S, RPC600649S, RPC600868S, RPC600984S, RPC601066S, RPC601181S, RPC601314S, RPC601634S, RPC601897S, RPC602180S, RPC602187S, RPC602526S, RPC603181S, RPC604820S, RPC601117S, RPC601727S, RPC602908S, RPC606176S, RPC603836S, RPC606274S, RPC601579S, RPC602195S, RPC603993S, RPC601358S, RPC601359S, RPC601542S, RPC601646S, RPC601647S, RPC601744S, RPC601745S, RPC601876S, RPC605367S, RPC605694S, RPC606079S, RPC606249S, RPC606250S, RPC606567S, RPC607014S, RPC607021S, RPC607485S, RPC600185S, RPC600283S, RPC600298S, RPC600446S, RPC600996S, RPC602101S, RPC603474S, RPC603477S, RPC604288S, RPC605203S, RPC602233S, RPC602558S, RPC604663S, RPC604670S, RPC602449S, RPC605079S, RPC600748S, RPC601063S, RPC601581S, RPC601679S, RPC601755S, RPC602203S, RPC602718S, RPC603127S, RPC603150S, RPC603291S, RPC604422S, RPC606736S, RPC606742S, RPC606161S, RPC600787S, RPC601332S, RPC601966S, RPC602513S, RPC602709S, RPC603450S, RPC603648S, RPC606248S, RPC604870S, RPC606571S, RPC605155S, RPC605614S, RPC605620S, RPC605948S, RPC600530S, RPC602674S, RPC603965S, RPC604064S, RPC600513S, RPC600740S, RPC600745S, RPC600845S, RPC602205S, RPC602838S, RPC602892S, RPC602893S, RPC603251S, RPC607772S, RPC607773S, RPC607774S, RPC607775S, RPC607776S, RPC608042S, RPC608043S, RPC608044S, RPC608045S, RPC608046S, RPC608047S, RPC608048S, RPC608049S, RPC608050S, RPC608051S, RPC605785S, RPC600547S, RPC601554S, RPC602786S, RPC603029S, RPC604170S, RPC604416S, RPC604734S, RPC600738S, RPC601826S, RPC602323S, RPC602698S, RPC603133S, RPC603284S, RPC603707S, RPC604631S, RPC605146S, RPC604916S, RPC604895S, RPC600249S, RPC601067S, RPC601232S, RPC601406S, RPC601705S, RPC602391S, RPC602446S, RPC602560S, RPC602712S, RPC602730S, RPC602738S, RPC602828S, RPC603075S, RPC603214S, RPC604121S, RPC604329S, RPC604350S, RPC605293S, RPC601431S, RPC605292S, RPC600936S, RPC601645S, RPC603227S, RPC603341S, RPC603342S, RPC607666S, RPC600437S, RPC600067S, RPC600326S, RPC605284S, RPC605632S, RPC603333S, RPC603794S, RPC605173S, RPC600848S, RPC607991S, RPC602547S, RPC605710S, RPC605743S, RPC607348S, RPC604200S, RPC604497S, RPC605304S, RPC605591S, RPC605739S, RPC605885S, RPC607019S, RPC600795S, RPC601188S, RPC601976S, RPC602406S, RPC602987S, RPC603269S, RPC603271S, RPC603976S, RPC604595S, RPC606478S, RPC601380S, RPC602107S, RPC600560S, RPC601398S, RPC601455S, RPC603586S, RPC605516S, RPC607543S, RPC601103S, RPC601582S, RPC603720S, RPC606406S, RPC606627S, RPC607006S, RPC602435S, RPC600721S, RPC600824S, RPC601557S, RPC602204S, RPC603205S, RPC605264S, RPC600436S, RPC600606S, RPC600697S, RPC600776S, RPC600821S, RPC601125S, RPC601221S, RPC601948S, RPC602013S, RPC602374S, RPC602426S, RPC603070S, RPC604054S, RPC603637S, RPC607139S, RPC607459S, RPC605307S, RPC606488S, RPC602349S, RPC602350S, RPC602378S, RPC604101S, RPC604619S, RPC607941S, RPC604876S, RPC605093S, RPC605230S, RPC605231S, RPC605541S, RPC607110S, RPC600600S, RPC600755S, RPC600788S, RPC601198S, RPC601708S, RPC601887S, RPC602144S, RPC602264S, RPC602293S, RPC602592S, RPC602649S, RPC602808S, RPC602809S, RPC602810S, RPC602882S, RPC603115S, RPC603319S, RPC603570S, RPC603784S, RPC604230S, RPC604763S, RPC604780S, RPC607111S, RPC607109S, RPC607412S, RPC607918S, RPC608283S, RPC607902S, RPC607077S, RPC601809S, RPC602940S, RPC607398S, RPC607009S, RPC601773S, RPC607957S, RPC606162S, RPC603850S, RPC602092S, RPC602113S, RPC600275S, RPC605403S, RPC605427S, RPC607142S, RPC607143S, RPC603757S, RPC604313S, RPC606319S, RPC603268S, RPC606361S, RPC603032S, RPC602384S, RPC601547S, RPC601671S, RPC601694S, RPC601722S, RPC601852S, RPC605468S, RPC606688S, RPC604290S, RPC605843S, RPC605587S, RPC606975S, RPC600695S, RPC605806S, RPC605929S, RPC605934S, RPC600415S, RPC601869S, RPC604493S, RPC605861S, RPC605864S, RPC608206S, RPC603147S, RPC603254S, RPC604900S, RPC605074S, RPC605075S, RPC605242S, RPC605297S, RPC605298S, RPC605299S, RPC605300S, RPC605663S, RPC605670S, RPC605672S, RPC605673S, RPC605877S, RPC605886S, RPC606502S, RPC606503S, RPC606716S, RPC606727S, RPC606728S, RPC606729S, RPC606731S, RPC607124S, RPC607219S, RPC607399S, RPC607481S, RPC607482S, RPC607483S, RPC607741S, RPC600645S, RPC600700S, RPC600752S, RPC600764S, RPC600780S, RPC600798S, RPC600833S, RPC600841S, RPC600890S, RPC601030S, RPC601065S, RPC601144S, RPC601185S, RPC601187S, RPC601244S, RPC601268S, RPC601427S, RPC601429S, RPC601430S, RPC601432S, RPC601644S, RPC601659S, RPC601692S, RPC601776S, RPC601958S, RPC601959S, RPC601962S, RPC601969S, RPC602147S, RPC602211S, RPC602212S, RPC602213S, RPC602425S, RPC602429S, RPC602593S, RPC602647S, RPC602668S, RPC602671S, RPC602728S, RPC602739S, RPC602796S, RPC602844S, RPC602915S, RPC602995S, RPC603026S, RPC603051S, RPC603091S, RPC603153S, RPC603274S, RPC603276S, RPC603347S, RPC603348S, RPC603377S, RPC603383S, RPC603384S, RPC603385S, RPC603386S, RPC603483S, RPC603592S, RPC603634S, RPC603635S, RPC603636S, RPC603717S, RPC603830S, RPC603831S, RPC603874S, RPC604112S, RPC604113S, RPC604114S, RPC604115S, RPC604142S, RPC604272S, RPC604307S, RPC604379S, RPC604454S, RPC604716S, RPC603975S, RPC607919S, RPC607921S, RPC608196S, RPC604021S, RPC604023S, RPC604339S, RPC600556S, RPC605721S, RPC605838S, RPC606286S, RPC606577S, RPC607017S, RPC600362S, RPC601612S, RPC603782S, RPC604228S, RPC607682S, RPC605768S, RPC602772S, RPC605488S, RPC605544S, RPC606550S, RPC607178S, RPC605745S, RPC601604S, RPC605801S, RPC601207S, RPC602510S, RPC602518S, RPC603223S, RPC608124S, RPC601916S, RPC602336S, RPC606871S, RPC607314S, RPC605624S, RPC606520S, RPC602342S, RPC603392S, RPC603500S, RPC602795S, RPC607783S, RPC604984S, RPC607771S, RPC600978S, RPC603270S, RPC603519S, RPC606070S, RPC605608S, RPC604326S, RPC603840S, RPC603179S, RPC602471S, RPC602469S, RPC603753S, RPC601537S, RPC604153S, RPC604661S, RPC606893S, RPC606976S, RPC607308S, RPC600442S, RPC603852S, RPC600131S, RPC601775S, RPC600171S, RPC605575S, RPC602989S, RPC605601S, RPC605803S, RPC603050S, RPC603362S, RPC603877S, RPC605062S, RPC605257S, RPC600972S, RPC601180S, RPC605555S, RPC606165S, RPC606530S, RPC607013S, RPC607084S, RPC600583S, RPC600778S, RPC600960S, RPC601230S, RPC601337S, RPC601555S, RPC601693S, RPC601850S, RPC601856S, RPC601968S, RPC602347S, RPC602353S, RPC602411S, RPC602591S, RPC602937S, RPC603176S, RPC603303S, RPC603358S, RPC603454S, RPC604256S, RPC604281S, RPC604282S, RPC607085S, RPC608034S, RPC607087S, RPC607088S, RPC605237S, RPC605748S, RPC602003S, RPC606142S, RPC606212S, RPC602351S, RPC603746S, RPC604881S, RPC605260S, RPC605661S, RPC606651S, RPC601317S, RPC601394S, RPC602834S, RPC603674S, RPC603973S, RPC603974S, RPC600634S, RPC601828S, RPC602670S, RPC605341S, RPC605402S, RPC605112S, RPC605565S, RPC605852S, RPC606585S, RPC606922S, RPC606990S, RPC607210S, RPC607557S, RPC607588S, RPC600613S, RPC601245S, RPC601435S, RPC601747S, RPC601797S, RPC602041S, RPC602852S, RPC603113S, RPC603445S, RPC603617S, RPC603704S, RPC603929S, RPC603981S, RPC604764S, RPC605429S, RPC605764S, RPC606069S, RPC607611S, RPC600643S, RPC600696S, RPC601371S, RPC601651S, RPC602142S, RPC601247S, RPC602929S, RPC600637S, RPC605213S, RPC602679S, RPC601433S, RPC604836S, RPC605925S, RPC605928S, RPC603562S, RPC605328S, RPC605931S, RPC603759S, RPC600448S, RPC602918S, RPC600245S, RPC602266S, RPC602835S, RPC601428S, RPC602504S, RPC603694S, RPC603786S, RPC604141S, RPC604189S, RPC600906S, RPC600959S, RPC602437S, RPC602745S, RPC603108S, RPC604355S, RPC604680S, RPC607470S, RPC604020S, RPC607592S, RPC603612S, RPC604052S, RPC602077S, RPC602571S, RPC605722S, RPC601089S, RPC601613S, RPC601915S, RPC601926S, RPC602296S, RPC602878S, RPC604757S, RPC600133S, RPC600572S, RPC601362S, RPC601883S, RPC602088S, RPC602971S, RPC606390S, RPC606393S, RPC603894S, RPC604857S, RPC605309S, RPC604682S, RPC605906S, RPC607176S, RPC604180S, RPC604545S, RPC606950S, RPC607650S, RPC605381S, RPC601677S, RPC604490S, RPC604720S, RPC604765S, RPC603025S, RPC604980S, RPC605057S, RPC606565S, RPC600898S, RPC602095S, RPC602188S, RPC602525S, RPC602840S, RPC605893S, RPC602894S, RPC602568S, RPC603106S, RPC600734S, RPC601593S, RPC602983S, RPC601211S, RPC601843S, RPC602059S, RPC602326S, RPC604615S, RPC605561S, RPC600908S, RPC602663S, RPC603082S, RPC600988S, RPC600341S, RPC601200S, RPC603970S, RPC604683S, RPC605204S, RPC607175S, RPC600904S, RPC601193S, RPC601603S, RPC602086S, RPC603089S, RPC605064S, RPC605136S, RPC607493S, RPC604527S, RPC608292S, RPC600397S, RPC600720S, RPC600723S, RPC601309S, RPC601372S, RPC601680S, RPC603923S, RPC606200S, RPC608230S, RPC604924S, RPC605073S, RPC605485S, RPC606372S, RPC603076S, RPC603407S, RPC604403S, RPC607283S, RPC605424S, RPC605795S, RPC606686S, RPC607081S, RPC607734S, RPC608132S, RPC600702S, RPC600703S, RPC601186S, RPC601397S, RPC601522S, RPC603102S, RPC603104S, RPC603140S, RPC603189S, RPC603209S, RPC603211S, RPC603220S, RPC603449S, RPC603580S, RPC604278S, RPC604480S, RPC604564S, RPC608203S, RPC606829S, RPC606830S, RPC606832S, RPC606834S, RPC606836S, RPC606838S, RPC600011S, RPC600114S, RPC600118S, RPC606833S, RPC606835S, RPC606837S, RPC606839S, RPC606843S, RPC605267S, RPC605772S, RPC605946S, RPC606413S, RPC600548S, RPC602002S, RPC602791S, RPC604828S, RPC606482S, RPC607200S, RPC602928S, RPC604831S, RPC602431S, RPC602595S, RPC601920S, RPC605295S, RPC603653S, RPC607351S, RPC602407S, RPC604184S, RPC605589S, RPC607202S, RPC607436S, RPC603139S, RPC605604S, RPC605758S, RPC607416S, RPC606412S, RPC606330S, RPC601436S, RPC607821S, RPC605562S, RPC606173S, RPC604506S, RPC606533S, RPC607069S, RPC600699S, RPC600891S, RPC602389S, RPC602630S, RPC603255S, RPC604248S, RPC604453S, RPC604617S, RPC604698S, RPC608253S, RPC607313S, RPC601925S, RPC605306S, RPC602063S, RPC605047S, RPC600278S, RPC605773S, RPC600955S, RPC601090S, RPC601237S, RPC601262S, RPC601696S, RPC603404S, RPC603443S, RPC604351S, RPC605392S, RPC606284S, RPC607645S, RPC602492S, RPC605813S, RPC600574S, RPC601614S, RPC601818S, RPC600741S, RPC601652S, RPC602239S, RPC602494S, RPC605856S, RPC600609S, RPC600911S, RPC600999S, RPC601632S, RPC601653S, RPC600189S, RPC603441S, RPC604090S, RPC604945S, RPC600644S, RPC600647S, RPC600648S, RPC600651S, RPC600652S, RPC600653S, RPC601473S, RPC601475S, RPC601476S, RPC605025S, RPC600198S, RPC601626S, RPC601711S, RPC606611S, RPC602468S, RPC602701S, RPC602710S, RPC603072S, RPC605927S, RPC605825S, RPC606640S, RPC607020S, RPC600803S, RPC600910S, RPC601060S, RPC602740S, RPC602744S, RPC603672S, RPC603789S, RPC604372S, RPC604950S, RPC607064S, RPC604294S, RPC604668S, RPC605225S, RPC607349S, RPC604257S, RPC600293S, RPC604869S, RPC604992S, RPC605484S, RPC605726S, RPC605942S, RPC606987S, RPC607932S, RPC601992S, RPC602360S, RPC602599S, RPC602621S, RPC602736S, RPC602993S, RPC603646S, RPC604098S, RPC604775S, RPC606945S, RPC603891S, RPC603557S, RPC606569S, RPC601871S, RPC602944S, RPC603869S, RPC600383S, RPC600384S, RPC602731S, RPC602883S, RPC602972S, RPC603588S, RPC604241S, RPC601769S, RPC602865S, RPC604133S, RPC600797S, RPC605767S, RPC603890S, RPC602234S, RPC600349S, RPC600350S, RPC600351S, RPC600353S, RPC600354S, RPC600355S, RPC600356S, RPC600357S, RPC600693S, RPC600704S, RPC601074S, RPC601256S, RPC602060S, RPC602065S, RPC602218S, RPC602294S, RPC602298S, RPC602289S, RPC602301S, RPC603294S, RPC603295S, RPC604485S, RPC600190S, RPC600452S, RPC604364S, RPC605167S, RPC606092S, RPC607015S, RPC604390S, RPC605417S, RPC602271S, RPC600770S, RPC600771S, RPC601034S, RPC601036S, RPC601101S, RPC601273S, RPC601275S, RPC607714S, RPC603818S, RPC600070S, RPC600071S, RPC600072S, RPC600319S, RPC600564S, RPC601195S, RPC601499S, RPC601500S, RPC602004S, RPC600708S, RPC601051S, RPC601226S, RPC601899S, RPC602554S, RPC600422S, RPC600883S, RPC601793S, RPC602385S, RPC600919S, RPC600926S, RPC604575S, RPC604576S, RPC605340S, RPC607838S, RPC607839S, RPC604867S, RPC604874S, RPC604905S, RPC604910S, RPC604937S, RPC604942S, RPC604943S, RPC604944S, RPC605163S, RPC605165S, RPC605166S, RPC605285S, RPC605286S, RPC605287S, RPC605288S, RPC605289S, RPC605957S, RPC605958S, RPC605959S, RPC605973S, RPC605974S, RPC606013S, RPC606014S, RPC606510S, RPC606512S, RPC606710S, RPC606711S, RPC607846S, RPC607943S, RPC608053S, RPC608055S, RPC600578S, RPC600580S, RPC600581S, RPC600917S, RPC601494S, RPC601830S, RPC601831S, RPC601837S, RPC601884S, RPC601985S, RPC601986S, RPC601987S, RPC601988S, RPC602042S, RPC602136S, RPC602146S, RPC602209S, RPC602249S, RPC602262S, RPC602784S, RPC602787S, RPC602797S, RPC602876S, RPC602926S, RPC603079S, RPC603105S, RPC603481S, RPC603583S, RPC603599S, RPC603655S, RPC603657S, RPC603658S, RPC603663S, RPC603664S, RPC603665S, RPC603666S, RPC603667S, RPC603669S, RPC604417S, RPC604418S, RPC604429S, RPC604434S, RPC604500S, RPC604501S, RPC604648S, RPC604690S, RPC604692S, RPC607944S, RPC608054S, RPC608083S, RPC608084S, RPC608187S, RPC608188S, RPC608256S, RPC608258S, RPC606810S, RPC606812S, RPC606813S, RPC606814S, RPC600909S, RPC601184S, RPC601364S, RPC601365S, RPC601368S, RPC601369S, RPC601468S, RPC603118S, RPC603120S, RPC603121S, RPC603122S, RPC603410S, RPC608316S, RPC608317S, RPC605110S, RPC602489S, RPC600094S, RPC600097S, RPC600495S, RPC600496S, RPC600497S, RPC600498S, RPC600499S, RPC600500S, RPC600879S, RPC600941S, RPC601293S, RPC601448S, RPC601449S, RPC601451S, RPC601505S, RPC601591S, RPC601600S, RPC601607S, RPC601753S, RPC601786S, RPC601811S, RPC601813S, RPC601998S, RPC602000S, RPC602030S, RPC602116S, RPC602225S, RPC602226S, RPC602227S, RPC602235S, RPC602237S, RPC602238S, RPC602466S, RPC602511S, RPC602514S, RPC602515S, RPC602516S, RPC602793S, RPC600314S, RPC603427S, RPC605068S, RPC605784S, RPC600552S, RPC601586S, RPC601792S, RPC602160S, RPC602161S, RPC603685S, RPC604672S, RPC603184S, RPC603279S, RPC602240S, RPC600733S, RPC600873S, RPC601457S, RPC602752S, RPC602900S, RPC603218S, RPC603742S, RPC604004S, RPC606825S, RPC603917S, RPC605313S, RPC605525S, RPC605526S, RPC605709S, RPC606005S, RPC606006S, RPC606151S, RPC606153S, RPC606554S, RPC606634S, RPC606635S, RPC607661S, RPC600672S, RPC600730S, RPC601385S, RPC601386S, RPC601516S, RPC601518S, RPC601519S, RPC602222S, RPC602224S, RPC602297S, RPC602299S, RPC602300S, RPC602660S, RPC602665S, RPC602896S, RPC602897S, RPC603033S, RPC603034S, RPC603739S, RPC603740S, RPC603741S, RPC604585S, RPC604586S, RPC604587S, RPC606556S, RPC606557S, RPC606765S, RPC606766S, RPC607659S, RPC607660S, RPC607825S, RPC608027S, RPC608154S, RPC608155S, RPC608257S, RPC608288S, RPC608289S, RPC605077S, RPC605470S, RPC605557S, RPC605725S, RPC606125S, RPC606163S, RPC606378S, RPC606677S, RPC606845S, RPC608180S, RPC600683S, RPC600691S, RPC600950S, RPC600975S, RPC600994S, RPC601728S, RPC602309S, RPC602864S, RPC602979S, RPC602985S, RPC603005S, RPC603037S, RPC603640S, RPC603961S, RPC604382S, RPC604652S, RPC601269S, RPC601334S, RPC601853S, RPC602072S, RPC602073S, RPC603543S, RPC603772S, RPC602210S, RPC602219S, RPC602221S, RPC600525S, RPC602836S, RPC604896S, RPC605104S, RPC606106S, RPC604223S, RPC604291S, RPC601057S, RPC603734S, RPC607123S, RPC601690S, RPC603889S, RPC606460S, RPC607251S, RPC602869S, RPC603429S, RPC605582S, RPC606311S, RPC606638S, RPC607030S, RPC607319S, RPC607371S, RPC607400S, RPC607471S, RPC604504S, RPC604518S, RPC606913S, RPC602949S, RPC604976S, RPC604985S, RPC604997S, RPC605007S, RPC605083S, RPC605129S, RPC605269S, RPC605273S, RPC605434S, RPC605443S, RPC605447S, RPC605450S, RPC605456S, RPC605457S, RPC605988S, RPC606001S, RPC606026S, RPC606038S, RPC606044S, RPC606045S, RPC606747S, RPC607152S, RPC607239S, RPC601014S, RPC601124S, RPC601130S, RPC601173S, RPC601175S, RPC601177S, RPC601419S, RPC601442S, RPC601443S, RPC601654S, RPC601838S, RPC601839S, RPC601840S, RPC601947S, RPC602052S, RPC602130S, RPC602286S, RPC602287S, RPC602288S, RPC602430S, RPC602458S, RPC602459S, RPC602477S, RPC602478S, RPC602534S, RPC602542S, RPC602578S, RPC602583S, RPC602954S, RPC602956S, RPC602960S, RPC602964S, RPC602966S, RPC602967S, RPC602968S, RPC602977S, RPC603154S, RPC604461S, RPC604505S, RPC604509S, RPC604510S, RPC604563S, RPC604565S, RPC604567S, RPC604568S, RPC604573S, RPC604743S, RPC607151S, RPC607245S, RPC605445S, RPC606042S, RPC606049S, RPC606052S, RPC600624S, RPC600718S, RPC600993S, RPC601418S, RPC602454S, RPC602455S, RPC602535S, RPC602540S, RPC602541S, RPC602586S, RPC602628S, RPC602975S, RPC603160S, RPC604444S, RPC604508S, RPC605435S, RPC605454S, RPC605991S, RPC606029S, RPC607234S, RPC604572S, RPC605271S, RPC605274S, RPC605436S, RPC605440S, RPC605449S, RPC605455S, RPC605458S, RPC605981S, RPC605982S, RPC605984S, RPC605990S, RPC605993S, RPC605994S, RPC606021S, RPC606022S, RPC606031S, RPC606034S, RPC606036S, RPC606039S, RPC606047S, RPC600492S, RPC600862S, RPC601128S, RPC601132S, RPC601133S, RPC601137S, RPC601151S, RPC601194S, RPC601326S, RPC601390S, RPC601445S, RPC601517S, RPC601527S, RPC601684S, RPC601707S, RPC602131S, RPC602132S, RPC602284S, RPC602456S, RPC602457S, RPC602476S, RPC602479S, RPC602480S, RPC602483S, RPC602536S, RPC602614S, RPC602615S, RPC602616S, RPC602978S, RPC602982S, RPC602997S, RPC602998S, RPC603007S, RPC603186S, RPC604551S, RPC604553S, RPC604557S, RPC604751S, RPC605437S, RPC605459S, RPC605980S, RPC606050S, RPC601444S, RPC602579S, RPC606017S, RPC604998S, RPC600491S, RPC601012S, RPC601085S, RPC601656S, RPC602482S, RPC602961S, RPC604438S, RPC604561S, RPC604978S, RPC607146S, RPC605005S, RPC602475S, RPC601131S, RPC605006S, RPC605439S, RPC606041S, RPC606967S, RPC601001S, RPC601008S, RPC601009S, RPC601149S, RPC601174S, RPC601441S, RPC602629S, RPC602962S, RPC602965S, RPC602976S, RPC602984S, RPC603156S, RPC604558S, RPC604739S, RPC606032S, RPC604667S, RPC607996S, RPC603000S, RPC602537S, RPC604439S, RPC601685S, RPC603006S, RPC605010S, RPC605444S, RPC605446S, RPC605451S, RPC605985S, RPC606016S, RPC601126S, RPC602290S, RPC604990S, RPC605004S, RPC605078S, RPC605432S, RPC605433S, RPC605453S, RPC605460S, RPC605986S, RPC606020S, RPC606025S, RPC606033S, RPC606035S, RPC606055S, RPC606428S, RPC606430S, RPC606746S, RPC607150S, RPC607156S, RPC607238S, RPC607462S, RPC601147S, RPC601148S, RPC601150S, RPC601440S, RPC601665S, RPC601668S, RPC601669S, RPC601835S, RPC601836S, RPC602613S, RPC603002S, RPC604549S, RPC604554S, RPC604555S, RPC604559S, RPC604562S, RPC604571S, RPC604666S, RPC604737S, RPC604742S, RPC604744S, RPC604748S, RPC604749S, RPC605431S, RPC605452S, RPC605461S, RPC605998S, RPC606000S, RPC606023S, RPC606056S, RPC607244S, RPC602538S, RPC602582S, RPC604569S, RPC604750S, RPC606037S, RPC605983S, RPC605001S, RPC605002S, RPC605003S, RPC605008S, RPC605270S, RPC605442S, RPC605448S, RPC605989S, RPC605996S, RPC605997S, RPC605999S, RPC606019S, RPC606030S, RPC601127S, RPC601655S, RPC601666S, RPC601833S, RPC602157S, RPC602285S, RPC602481S, RPC602484S, RPC602533S, RPC602585S, RPC602618S, RPC602957S, RPC602958S, RPC602980S, RPC602981S, RPC602996S, RPC603003S, RPC604511S, RPC604512S, RPC604550S, RPC604741S, RPC604987S, RPC606054S, RPC604393S, RPC604394S, RPC601202S, RPC605272S, RPC606027S, RPC606028S, RPC601179S, RPC601420S, RPC601421S, RPC601528S, RPC602543S, RPC602955S, RPC602959S, RPC605228S, RPC603413S, RPC603415S, RPC605995S, RPC607236S, RPC607460S
FEI Number 2182208
Recalling Firm/
Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
8200 Coral Sea St Ne
Mounds View MN 55112-4391
For Additional Information ContactTechnical Services
800-929-4043
Manufacturer Reason
for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Quantity in Commerce12291 units
DistributionUS Nationwide - Worldwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA DatabasePMAs with Product Code = NIK
PMAs with Product Code = NIK
-
-