• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall medtronic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall medtronicsee related information
Date Initiated by FirmMay 10, 2023
Date PostedJune 16, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1758-2023
Recall Event ID 92328
PMA NumberP010031S550 
Product Classification Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) - Product Code NIK
ProductCRTD DTMA2D4 CLARIA MRI OUS DF4, Model Number DTMA2D4; Implantable Cardioverter Defibrillators
Code Information GTIN 00643169720664, Lot Serial Numbers: RPD605666S, RPD602978S, RPD602979S, RPD603034S, RPD604592S, RPD601674S, RPD603271S, RPD604163S, RPD605225S, RPD604970S, RPD601851S, RPD602346S, RPD602637S, RPD603019S, RPD605081S, RPD605207S, RPD604558S, RPD604927S, RPD604980S, RPD601671S, RPD601672S, RPD601677S, RPD601678S, RPD602218S, RPD602277S, RPD602303S, RPD602369S, RPD602439S, RPD602443S, RPD602581S, RPD602724S, RPD602728S, RPD602781S, RPD602793S, RPD602910S, RPD602940S, RPD603033S, RPD603173S, RPD603180S, RPD603254S, RPD603677S, RPD603733S, RPD603756S, RPD603966S, RPD603967S, RPD604091S, RPD604124S, RPD604201S, RPD604897S, RPD605386S, RPD605444S, RPD605446S, RPD605524S, RPD605667S, RPD605679S, RPD605428S, RPD604405S, RPD604954S, RPD604941S, RPD605596S, RPD601962S, RPD602692S, RPD602918S, RPD602973S, RPD602982S, RPD605595S, RPD605601S, RPD603067S, RPD602895S, RPD604445S, RPD602694S, RPD604443S, RPD604552S, RPD605084S, RPD603878S, RPD602612S, RPD602015S, RPD602203S, RPD602244S, RPD602348S, RPD602512S, RPD602649S, RPD602813S, RPD602835S, RPD602924S, RPD603046S, RPD603061S, RPD603062S, RPD603483S, RPD603567S, RPD603785S, RPD603849S, RPD603850S, RPD603860S, RPD603870S, RPD603888S, RPD604408S, RPD604562S, RPD604598S, RPD604616S, RPD604823S, RPD604923S, RPD605011S, RPD605053S, RPD605395S, RPD605403S, RPD605624S, RPD605673S, RPD601592S, RPD603971S, RPD604019S, RPD604497S, RPD604806S, RPD604924S, RPD605172S, RPD605252S, RPD605645S, RPD603584S, RPD605643S, RPD602423S, RPD604221S, RPD601924S, RPD602113S, RPD602179S, RPD602635S, RPD602963S, RPD603087S, RPD603108S, RPD603310S, RPD604038S, RPD604062S, RPD604349S, RPD604412S, RPD604563S, RPD604795S, RPD604990S, RPD605006S, RPD605122S, RPD605124S, RPD605140S, RPD605267S, RPD605337S, RPD605588S, RPD605674S, RPD601625S, RPD602073S, RPD604042S, RPD604347S, RPD604727S, RPD602999S, RPD604462S, RPD605042S, RPD602526S, RPD604432S, RPD602723S, RPD603149S, RPD603672S, RPD605231S, RPD601933S, RPD602094S, RPD602872S, RPD603885S, RPD604355S, RPD605208S, RPD605286S, RPD603649S, RPD603719S, RPD603729S, RPD603758S, RPD604197S, RPD602590S, RPD603135S, RPD603268S, RPD603297S, RPD604365S, RPD604528S, RPD604556S, RPD604565S, RPD605108S, RPD605158S, RPD603980S, RPD604184S, RPD604270S, RPD604577S, RPD605480S, RPD605039S, RPD601471S, RPD603435S, RPD603900S, RPD603976S, RPD604361S, RPD604467S, RPD605247S, RPD605636S, RPD603985S, RPD605298S, RPD605495S, RPD605614S, RPD603740S, RPD603450S, RPD603598S, RPD603613S, RPD604148S, RPD604194S, RPD604302S, RPD604485S, RPD604517S, RPD604606S, RPD604626S, RPD604401S, RPD604464S, RPD604532S, RPD605178S, RPD605204S, RPD605471S, RPD605531S, RPD605533S, RPD603775S, RPD604305S, RPD604609S, RPD605034S, RPD602289S, RPD602968S, RPD603504S, RPD603550S, RPD603862S, RPD604407S, RPD605061S, RPD605062S, RPD605063S, RPD605184S, RPD605249S, RPD605522S, RPD602469S, RPD603038S, RPD603523S, RPD604359S, RPD604391S, RPD604984S, RPD603615S, RPD603722S, RPD604617S, RPD602566S, RPD602571S, RPD603654S, RPD603757S, RPD604193S, RPD602115S, RPD602169S, RPD602516S, RPD603327S, RPD604266S, RPD604634S, RPD604826S, RPD605515S, RPD602118S, RPD601903S, RPD603680S, RPD603869S, RPD604409S, RPD604862S, RPD605050S, RPD605101S, RPD605259S, RPD601638S, RPD603717S, RPD602047S, RPD602192S, RPD602240S, RPD602598S, RPD602646S, RPD602687S, RPD602747S, RPD602812S, RPD602946S, RPD603074S, RPD603910S, RPD603969S, RPD603992S, RPD602145S, RPD602705S, RPD603749S, RPD604151S, RPD604217S, RPD604350S, RPD605163S, RPD605195S, RPD605209S, RPD605246S, RPD605589S, RPD602018S, RPD602364S, RPD602446S, RPD602626S, RPD603024S, RPD603650S, RPD603829S, RPD603972S, RPD604252S, RPD604298S, RPD604484S, RPD604650S, RPD605176S, RPD605292S, RPD605490S, RPD605501S, RPD601868S, RPD601930S, RPD602120S, RPD602320S, RPD602358S, RPD602464S, RPD603103S, RPD603156S, RPD603166S, RPD603187S, RPD603605S, RPD603981S, RPD603990S, RPD604035S, RPD604045S, RPD604611S, RPD605030S, RPD605248S, RPD605482S, RPD605492S, RPD601778S, RPD602250S, RPD604268S, RPD604289S, RPD604476S, RPD603927S, RPD603072S, RPD603812S, RPD603952S, RPD604324S, RPD604685S, RPD605427S, RPD605468S, RPD605532S, RPD602397S, RPD603264S, RPD603510S, RPD603660S, RPD603747S, RPD602792S, RPD603815S, RPD602396S, RPD602436S, RPD602776S, RPD603028S, RPD603699S, RPD603769S, RPD604150S, RPD605334S, RPD602187S, RPD601474S, RPD602065S, RPD601603S, RPD601610S, RPD602124S, RPD602184S, RPD602915S, RPD603754S, RPD602037S, RPD602053S, RPD604050S, RPD602359S, RPD601608S, RPD601705S, RPD601754S, RPD602183S, RPD602313S, RPD602315S, RPD602409S, RPD602557S, RPD602775S, RPD602778S, RPD603582S, RPD603827S, RPD604089S, RPD604212S, RPD604274S, RPD604312S, RPD604470S, RPD604509S, RPD605200S, RPD605212S, RPD601612S, RPD601639S, RPD601686S, RPD601950S, RPD602076S, RPD602123S, RPD602797S, RPD603208S, RPD603610S, RPD603723S, RPD603755S, RPD604510S, RPD604654S, RPD604683S, RPD601618S, RPD604456S, RPD604995S, RPD605377S, RPD601478S, RPD601503S, RPD601598S, RPD602798S, RPD604637S, RPD604855S, RPD605049S, RPD605236S, RPD605455S, RPD601606S, RPD602375S, RPD602938S, RPD603143S, RPD604939S, RPD605283S, RPD605637S, RPD601600S, RPD604186S, RPD603489S, RPD603592S, RPD603617S, RPD603806S, RPD601553S, RPD601684S, RPD602783S, RPD602998S, RPD603499S, RPD604105S, RPD604204S, RPD604458S, RPD602371S, RPD601917S, RPD603714S, RPD601642S, RPD601643S, RPD601616S, RPD601798S, RPD605475S, RPD603300S, RPD604154S, RPD604725S, RPD605676S, RPD603296S, RPD602204S, RPD602207S, RPD601796S, RPD602374S, RPD603279S, RPD603452S, RPD603500S, RPD603562S, RPD603702S, RPD604211S, RPD604265S, RPD604303S, RPD604313S, RPD604785S, RPD604979S, RPD605307S, RPD605362S, RPD605627S, RPD602606S, RPD603877S, RPD602485S, RPD604292S, RPD605133S, RPD605299S, RPD605390S, RPD602210S, RPD601477S, RPD601755S, RPD601952S, RPD605275S, RPD602200S, RPD602202S, RPD602208S, RPD601499S, RPD601685S, RPD602064S, RPD602133S, RPD602519S, RPD603530S, RPD604189S, RPD604196S, RPD604207S, RPD604219S, RPD604687S, RPD604819S, RPD605187S, RPD605234S, RPD605251S, RPD605341S, RPD605372S, RPD605474S, RPD605638S, RPD605659S, RPD602231S, RPD603874S, RPD603975S, RPD604435S, RPD605573S, RPD601607S, RPD602532S, RPD604001S, RPD604531S, RPD605215S, RPD602558S, RPD603940S, RPD603941S, RPD603943S, RPD601954S, RPD602111S, RPD602952S, RPD603202S, RPD603893S, RPD601759S, RPD601972S, RPD602082S, RPD604267S, RPD602140S, RPD602225S, RPD601835S, RPD601843S, RPD601899S, RPD601986S, RPD601987S, RPD602071S, RPD602430S, RPD603236S, RPD603653S, RPD604253S, RPD604337S, RPD605013S, RPD605510S, RPD605609S, RPD605422S, RPD605548S, RPD605591S, RPD602339S, RPD604000S, RPD605302S, RPD605477S, RPD602642S, RPD603614S, RPD604224S, RPD604520S, RPD603589S, RPD605146S, RPD603022S, RPD604832S, RPD604877S, RPD605094S, RPD605670S, RPD605649S, RPD603081S, RPD603289S, RPD605131S, RPD605205S, RPD605605S, RPD603021S, RPD603112S, RPD604455S, RPD604742S, RPD604817S, RPD605544S, RPD601824S, RPD602475S, RPD603673S, RPD603655S, RPD604320S, RPD601998S, RPD602007S, RPD602543S, RPD602710S, RPD602749S, RPD602753S, RPD602995S, RPD603070S, RPD603228S, RPD603318S, RPD603549S, RPD603661S, RPD603902S, RPD604123S, RPD604220S, RPD604474S, RPD604597S, RPD604781S, RPD603897S, RPD603983S, RPD604160S, RPD605496S, RPD605576S, RPD602639S, RPD604066S, RPD604542S, RPD602641S, RPD603153S, RPD605107S, RPD601655S, RPD601825S, RPD601827S, RPD601855S, RPD602146S, RPD602174S, RPD602333S, RPD602843S, RPD602845S, RPD604065S, RPD605279S, RPD605285S, RPD605546S, RPD602366S, RPD602426S, RPD602427S, RPD602734S, RPD602790S, RPD602819S, RPD602931S, RPD603088S, RPD603834S, RPD603842S, RPD602628S, RPD602664S, RPD603521S, RPD603690S, RPD603851S, RPD604861S, RPD601734S, RPD601828S, RPD601859S, RPD601860S, RPD603181S, RPD603278S, RPD605206S, RPD601536S, RPD602961S, RPD602994S, RPD602818S, RPD604064S, RPD604280S, RPD604442S, RPD604461S, RPD604780S, RPD604815S, RPD605075S, RPD605303S, RPD605304S, RPD605542S, RPD605497S, RPD602002S, RPD602253S, RPD602951S, RPD602981S, RPD603090S, RPD603950S, RPD605652S, RPD605083S, RPD605593S, RPD601791S, RPD601981S, RPD603899S, RPD604417S, RPD605005S, RPD605309S, RPD601715S, RPD602148S, RPD602338S, RPD604643S, RPD605628S, RPD605633S, RPD602196S, RPD602334S, RPD602561S, RPD602811S, RPD604063S, RPD604255S, RPD601640S, RPD605106S, RPD605141S, RPD602980S, RPD604315S, RPD603788S, RPD602052S, RPD602083S, RPD602084S, RPD602841S, RPD602922S, RPD603146S, RPD603443S, RPD603444S, RPD604638S, RPD604739S, RPD604740S, RPD601254S, RPD601511S, RPD601912S, RPD602323S, RPD603267S, RPD603627S, RPD603642S, RPD603796S, RPD605610S, RPD601469S, RPD601864S, RPD601988S, RPD602652S, RPD602972S, RPD603831S, RPD603852S, RPD604041S, RPD604127S, RPD605026S, RPD605103S, RPD605626S, RPD605640S, RPD602848S, RPD603837S, RPD604282S, RPD603177S, RPD604390S, RPD601708S, RPD602957S, RPD603069S, RPD603397S, RPD603441S, RPD603602S, RPD603603S, RPD603608S, RPD604336S, RPD604400S, RPD604788S, RPD604813S, RPD602449S, RPD603114S, RPD601532S, RPD601571S, RPD601661S, RPD602009S, RPD602170S, RPD603344S, RPD603345S, RPD604814S, RPD601832S, RPD602324S, RPD602954S, RPD602955S, RPD601955S, RPD602638S, RPD605668S, RPD601829S, RPD602172S, RPD601852S, RPD602028S, RPD602242S, RPD602291S, RPD602379S, RPD602444S, RPD603082S, RPD603234S, RPD603502S, RPD603743S, RPD604181S, RPD604291S, RPD604341S, RPD604397S, RPD604686S, RPD604765S, RPD605143S, RPD605467S, RPD605528S, RPD601865S, RPD602155S, RPD602156S, RPD602163S, RPD602331S, RPD602854S, RPD602867S, RPD604300S, RPD604667S, RPD604717S, RPD605662S, RPD602305S, RPD603216S, RPD603448S, RPD603449S, RPD603451S, RPD603453S, RPD603462S, RPD604049S, RPD604459S, RPD604475S, RPD604503S, RPD604561S, RPD604594S, RPD604645S, RPD604682S, RPD604732S, RPD604851S, RPD605192S, RPD605244S, RPD605502S, RPD605584S, RPD605631S, RPD602697S, RPD602822S, RPD604656S, RPD604782S, RPD605223S, RPD605347S, RPD605481S, RPD605574S, RPD605360S, RPD601786S, RPD601949S, RPD602034S, RPD602161S, RPD602402S, RPD602500S, RPD603376S, RPD603474S, RPD603830S, RPD603889S, RPD603924S, RPD604052S, RPD605007S, RPD602209S, RPD605213S, RPD601491S, RPD601569S, RPD601736S, RPD601990S, RPD602171S, RPD602173S, RPD602844S, RPD601487S, RPD602592S, RPD603492S, RPD604843S, RPD605393S, RPD605648S, RPD601465S, RPD601502S, RPD603121S, RPD603125S, RPD603663S, RPD603664S, RPD605296S, RPD604879S, RPD602177S, RPD602831S, RPD603063S, RPD604967S, RPD605201S, RPD605366S, RPD604074S, RPD603999S, RPD601943S, RPD602185S, RPD602478S, RPD602699S, RPD602751S, RPD602864S, RPD603974S, RPD604128S, RPD605306S, RPD605367S, RPD601878S, RPD602384S, RPD602576S, RPD602611S, RPD603335S, RPD604460S, RPD604850S, RPD605211S, RPD605289S, RPD605644S, RPD602540S, RPD602644S, RPD602743S, RPD605029S, RPD602414S, RPD603505S, RPD604719S, RPD604722S, RPD604891S, RPD605041S, RPD605287S, RPD603179S, RPD603390S, RPD603421S, RPD603595S, RPD603669S, RPD604079S, RPD604082S, RPD604226S, RPD604242S, RPD604244S, RPD604583S, RPD604816S, RPD604909S, RPD604910S, RPD604921S, RPD605370S, RPD605408S, RPD601897S, RPD602932S, RPD602936S, RPD605268S, RPD605269S, RPD605326S, RPD605517S, RPD605654S, RPD605291S, RPD604540S, RPD605647S, RPD603391S, RPD603456S, RPD605077S, RPD603576S, RPD603577S, RPD603578S, RPD604017S, RPD601876S, RPD602021S, RPD602181S, RPD602666S, RPD602740S, RPD602828S, RPD605635S, RPD605093S, RPD603544S, RPD601523S, RPD601682S, RPD601738S, RPD602035S, RPD602085S, RPD602144S, RPD602245S, RPD602340S, RPD602341S, RPD602380S, RPD602506S, RPD602509S, RPD602912S, RPD602987S, RPD602989S, RPD603048S, RPD603367S, RPD603369S, RPD603853S, RPD603854S, RPD604342S, RPD604343S, RPD604631S, RPD604774S, RPD604779S, RPD601995S, RPD605549S, RPD601766S, RPD601773S, RPD602919S, RPD603838S, RPD601548S, RPD603394S, RPD601940S, RPD602107S, RPD602584S, RPD602585S, RPD603346S, RPD603347S, RPD603383S, RPD603384S, RPD603426S, RPD604628S, RPD601964S, RPD602038S, RPD602226S, RPD602352S, RPD603190S, RPD603248S, RPD601963S, RPD604824S, RPD604543S, RPD601466S, RPD601578S, RPD602091S, RPD602599S, RPD602643S, RPD602674S, RPD602684S, RPD602815S, RPD602816S, RPD602817S, RPD602884S, RPD602885S, RPD603035S, RPD603093S, RPD602328S, RPD603052S, RPD603486S, RPD603832S, RPD604175S, RPD604182S, RPD601831S, RPD601968S, RPD602112S, RPD602251S, RPD602412S, RPD603047S, RPD603105S, RPD603365S, RPD603524S, RPD604022S, RPD604338S, RPD604709S, RPD605162S, RPD605264S, RPD605270S, RPD605357S, RPD605489S, RPD601793S, RPD602489S, RPD603272S, RPD603428S, RPD604764S, RPD604898S, RPD604999S, RPD605059S, RPD604493S, RPD602587S, RPD603580S, RPD602003S, RPD602127S, RPD602804S, RPD603765S, RPD601549S, RPD601550S, RPD603566S, RPD603587S, RPD604743S, RPD604746S, RPD604882S, RPD602550S, RPD601802S, RPD602030S, RPD604069S, RPD604744S, RPD605175S, RPD605197S, RPD605585S, RPD605663S, RPD604998S, RPD602745S, RPD602898S, RPD602902S, RPD603559S, RPD603630S, RPD603906S, RPD603997S, RPD603998S, RPD604053S, RPD604729S, RPD604730S, RPD604731S, RPD604973S, RPD604974S, RPD604975S, RPD604977S, RPD604735S, RPD604784S, RPD604917S, RPD602199S, RPD602308S, RPD602337S, RPD602857S, RPD603060S, RPD603995S, RPD604055S, RPD604153S, RPD604323S, RPD605224S, RPD605563S, RPD601574S, RPD601861S, RPD601862S, RPD603127S, RPD602667S, RPD602862S, RPD603101S, RPD603241S, RPD603395S, RPD604327S, RPD604383S, RPD605074S, RPD605177S, RPD605186S, RPD601513S, RPD601833S, RPD601934S, RPD602259S, RPD603656S, RPD603876S, RPD604893S, RPD603920S, RPD604745S, RPD605487S, RPD601889S, RPD601891S, RPD602307S, RPD602330S, RPD602425S, RPD602560S, RPD602658S, RPD602683S, RPD603059S, RPD603097S, RPD603427S, RPD603564S, RPD603586S, RPD603696S, RPD603791S, RPD603968S, RPD604395S, RPD604410S, RPD604411S, RPD604894S, RPD605045S, RPD605099S, RPD605123S, RPD605155S, RPD605227S, RPD603861S, RPD604083S, RPD603162S, RPD603163S, RPD603688S, RPD603689S, RPD604272S, RPD604870S, RPD601573S, RPD602217S, RPD602746S, RPD605620S, RPD605436S, RPD601974S, RPD602162S, RPD602675S, RPD602913S, RPD602294S, RPD603569S, RPD603585S, RPD603836S, RPD604007S, RPD604553S, RPD604868S, RPD604994S, RPD605044S, RPD605339S, RPD604134S, RPD602777S, RPD604618S, RPD602100S, RPD602447S, RPD602607S, RPD602945S, RPD603171S, RPD603332S, RPD603890S, RPD604161S, RPD604254S, RPD604449S, RPD604787S, RPD604297S, RPD604608S, RPD602771S, RPD601782S, RPD601593S, RPD604831S, RPD602278S, RPD603265S, RPD604198S, RPD604345S, RPD604771S, RPD604935S, RPD604958S, RPD605226S, RPD605233S, RPD605368S, RPD605401S, RPD605447S, RPD605545S, RPD601836S, RPD605432S, RPD601688S, RPD601960S, RPD602297S, RPD602807S, RPD603056S, RPD603085S, RPD603309S, RPD603380S, RPD604724S, RPD605020S, RPD605575S, RPD605675S, RPD601874S, RPD602275S, RPD602276S, RPD602316S, RPD602322S, RPD602513S, RPD602537S, RPD602618S, RPD602750S, RPD602933S, RPD603123S, RPD603138S, RPD603189S, RPD603431S, RPD603458S, RPD603465S, RPD603479S, RPD603512S, RPD603575S, RPD603855S, RPD603894S, RPD603895S, RPD604146S, RPD604162S, RPD604356S, RPD604450S, RPD604453S, RPD604483S, RPD604491S, RPD604666S, RPD604865S, RPD605019S, RPD605024S, RPD605090S, RPD605202S, RPD605313S, RPD605315S, RPD605332S, RPD605348S, RPD605349S, RPD605359S, RPD605382S, RPD605385S, RPD605406S, RPD605442S, RPD605537S, RPD605587S, RPD605642S, RPD605646S, RPD605664S, RPD605671S, RPD605672S, RPD602165S, RPD603546S, RPD604386S, RPD604871S, RPD604940S, RPD601978S, RPD602921S, RPD603865S, RPD601481S, RPD601488S, RPD603277S, RPD602407S, RPD602541S, RPD602542S, RPD602736S, RPD603029S, RPD603030S, RPD603031S, RPD603148S, RPD603947S, RPD604152S, RPD604366S, RPD604835S, RPD604838S, RPD604839S, RPD604841S, RPD604929S, RPD604930S, RPD605506S, RPD605507S, RPD605508S, RPD601658S, RPD601743S, RPD601776S, RPD602080S, RPD602081S, RPD602106S, RPD602212S, RPD602220S, RPD603430S, RPD602851S, RPD602870S, RPD602871S, RPD603206S, RPD602754S, RPD603197S, RPD601604S, RPD602455S, RPD605381S, RPD601501S, RPD601585S, RPD601619S, RPD601629S, RPD601631S, RPD602186S, RPD603503S, RPD603731S, RPD604325S, RPD605371S, RPD605379S, RPD605429S, RPD604690S, RPD602800S, RPD602903S, RPD603291S, RPD603616S, RPD603737S, RPD603875S, RPD604179S, RPD604680S, RPD604738S, RPD604792S, RPD605319S, RPD605384S, RPD605409S, RPD605425S, RPD605479S, RPD604906S, RPD602024S, RPD602299S, RPD605284S, RPD601622S, RPD601665S, RPD602036S, RPD602434S, RPD602900S, RPD603588S, RPD603611S, RPD603760S, RPD605174S, RPD605375S, RPD601636S, RPD602548S, RPD605036S, RPD605043S, RPD605435S, RPD605456S, RPD605463S, RPD605519S, RPD604737S, RPD605033S, RPD601614S, RPD601637S, RPD601784S, RPD601894S, RPD602549S, RPD602791S, RPD602799S, RPD602937S, RPD603124S, RPD603259S, RPD603591S, RPD604515S, RPD604799S, RPD605232S, RPD605594S, RPD605603S, RPD602466S, RPD604187S, RPD601780S, RPD602001S, RPD602075S, RPD602451S, RPD602967S, RPD602296S, RPD604821S, RPD604059S, RPD602262S, RPD602850S, RPD602966S, RPD603042S, RPD604614S, RPD601615S, RPD602142S, RPD602318S, RPD601546S, RPD601590S, RPD601758S, RPD602476S, RPD602959S, RPD602970S, RPD603175S, RPD603263S, RPD603399S, RPD603741S, RPD603752S, RPD603911S, RPD603953S, RPD603987S, RPD604142S, RPD604223S, RPD604514S, RPD604546S, RPD605173S, RPD605242S, RPD605318S, RPD605518S, RPD605555S, RPD605590S, RPD605678S, RPD604487S, RPD602460S, RPD605650S, RPD603203S, RPD604247S, RPD601666S, RPD601787S, RPD604423S, RPD601495S, RPD602906S, RPD603581S, RPD603710S, RPD604096S, RPD605540S, RPD601725S, RPD602101S, RPD602600S, RPD604257S, RPD604438S, RPD604463S, RPD605198S, RPD602383S, RPD605465S, RPD602221S, RPD602554S, RPD603045S, RPD602428S, RPD602837S, RPD602863S, RPD603051S, RPD603301S, RPD603411S, RPD603574S, RPD603859S, RPD603922S, RPD604501S, RPD604593S, RPD604692S, RPD604755S, RPD601696S, RPD603375S, RPD604251S, RPD604286S, RPD604354S, RPD604385S, RPD604676S, RPD601667S, RPD604036S, RPD604279S, RPD604496S, RPD603261S, RPD603331S, RPD603432S, RPD603464S, RPD603879S, RPD604056S, RPD604133S, RPD604416S, RPD604715S, RPD604955S, RPD604978S, RPD605134S, RPD605156S, RPD605164S, RPD605239S, RPD605311S, RPD605504S, RPD602102S, RPD603846S, RPD605353S, RPD602019S, RPD602040S, RPD602043S, RPD602523S, RPD602824S, RPD602896S, RPD603835S, RPD605220S, RPD605301S, RPD601475S, RPD601520S, RPD601521S, RPD601529S, RPD601540S, RPD601547S, RPD601559S, RPD601716S, RPD601718S, RPD602254S, RPD602255S, RPD602256S, RPD602336S, RPD602362S, RPD602591S, RPD602676S, RPD602677S, RPD602689S, RPD602820S, RPD602865S, RPD603058S, RPD603223S, RPD603229S, RPD603406S, RPD603501S, RPD601858S, RPD602847S, RPD604734S, RPD601813S, RPD601911S, RPD602830S, RPD602964S, RPD605430S, RPD605499S, RPD602563S, RPD602868S, RPD605443S, RPD602726S, RPD602343S, RPD601805S, RPD602491S, RPD602625S, RPD602782S, RPD602823S, RPD602859S, RPD603548S, RPD603563S, RPD603872S, RPD601819S, RPD602230S, RPD602238S, RPD602361S, RPD602650S, RPD602657S, RPD602808S, RPD603054S, RPD604393S, RPD604647S, RPD604892S, RPD605095S, RPD605378S, RPD605564S, RPD602462S, RPD603200S, RPD603728S, RPD604021S, RPD604214S, RPD604340S, RPD604655S, RPD604736S, RPD605240S, RPD605139S, RPD601741S, RPD601901S, RPD602309S, RPD602700S, RPD603572S, RPD603706S, RPD604299S, RPD601969S, RPD602302S, RPD603094S, RPD603568S, RPD603698S, RPD603703S, RPD603704S, RPD603705S, RPD604023S, RPD604024S, RPD604029S, RPD604030S, RPD604173S, RPD604227S, RPD604228S, RPD604230S, RPD604231S, RPD604232S, RPD604233S, RPD604234S, RPD604236S, RPD604237S, RPD604239S, RPD604392S, RPD604523S, RPD604982S, RPD604985S, RPD604987S, RPD604988S, RPD605129S, RPD605179S, RPD605181S, RPD605182S, RPD605271S, RPD605333S, RPD605389S, RPD605656S, RPD605669S, RPD601935S, RPD602104S, RPD602105S, RPD602274S, RPD602349S, RPD602490S, RPD602577S, RPD602996S, RPD603023S, RPD603126S, RPD603131S, RPD603313S, RPD603360S, RPD603533S, RPD603534S, RPD601919S, RPD602545S, RPD602552S, RPD604936S, RPD605397S, RPD605416S, RPD603488S, RPD603867S, RPD601794S, RPD602679S, RPD602738S, RPD603129S, RPD603192S, RPD603414S, RPD604418S, RPD604981S, RPD602269S, RPD601900S, RPD602033S, RPD602283S, RPD602321S, RPD602515S, RPD603036S, RPD603596S, RPD603742S, RPD603962S, RPD604398S, RPD604513S, RPD604569S, RPD603847S, RPD604248S, RPD604858S, RPD602213S, RPD604362S, RPD602154S, RPD602821S, RPD603958S, RPD604922S, RPD605012S, RPD601907S, RPD602095S, RPD602098S, RPD605196S, RPD605484S, RPD604560S, RPD601253S, RPD601531S, RPD602284S, RPD602401S, RPD602435S, RPD603225S, RPD604658S, RPD604845S, RPD604962S, RPD605100S, RPD605150S, RPD605276S, RPD605651S, RPD601767S, RPD601768S, RPD602616S, RPD602621S, RPD603037S, RPD603362S, RPD603424S, RPD603526S, RPD603561S, RPD603644S, RPD603726S, RPD603864S, RPD604012S, RPD604468S, RPD604482S, RPD604537S, RPD604673S, RPD604708S, RPD604800S, RPD604801S, RPD604818S, RPD604825S, RPD604895S, RPD604901S, RPD604937S, RPD604993S, RPD605023S, RPD605027S, RPD605092S, RPD604896S, RPD601518S, RPD602086S, RPD602527S, RPD602661S, RPD603956S, RPD604600S, RPD602514S, RPD603107S, RPD603404S, RPD603776S, RPD602286S, RPD602365S, RPD601611S, RPD601623S, RPD601816S, RPD602197S, RPD603139S, RPD603901S, RPD604524S, RPD605117S, RPD605168S, RPD605293S, RPD605294S, RPD605660S, RPD605661S, RPD601690S, RPD602458S, RPD603570S, RPD602897S, RPD604293S, RPD603882S, RPD603909S, RPD604351S, RPD601697S, RPD601742S, RPD602239S, RPD603104S, RPD603311S, RPD605543S, RPD601522S, RPD601525S, RPD601953S, RPD602441S, RPD602583S, RPD603009S, RPD603973S, RPD604448S, RPD604625S, RPD601869S, RPD602026S, RPD602031S, RPD602070S, RPD602168S, RPD602252S, RPD602405S, RPD602432S, RPD602471S, RPD602575S, RPD602688S, RPD602755S, RPD602925S, RPD603247S, RPD603357S, RPD603423S, RPD603469S, RPD603495S, RPD603555S, RPD603609S, RPD603657S, RPD603732S, RPD603977S, RPD604020S, RPD604147S, RPD604157S, RPD604281S, RPD604421S, RPD604703S, RPD604837S, RPD604916S, RPD605144S, RPD603517S, RPD603565S, RPD604913S, RPD602672S, RPD602969S, RPD603077S, RPD605116S, RPD605615S, RPD602167S, RPD602624S, RPD603161S, RPD603612S, RPD603925S, RPD604502S, RPD605535S, RPD602394S, RPD603459S, RPD604477S, RPD604945S, RPD603529S, RPD604853S, RPD602496S, RPD602497S, RPD602499S, RPD602610S, RPD602653S, RPD602752S, RPD602814S, RPD602838S, RPD603040S, RPD603111S, RPD603409S, RPD603675S, RPD603782S, RPD604070S, RPD604145S, RPD604250S, RPD604856S, RPD604926S, RPD601545S, RPD601921S, RPD602046S, RPD603133S, RPD603298S, RPD604599S, RPD604957S, RPD605102S, RPD605335S, RPD605352S, RPD604097S, RPD602263S, RPD602511S, RPD603466S, RPD603658S, RPD604141S, RPD604314S, RPD604387S, RPD604466S, RPD604504S, RPD604566S, RPD605010S, RPD605258S, RPD605305S, RPD605336S, RPD605363S, RPD602125S, RPD602326S, RPD602403S, RPD602732S, RPD602233S, RPD602234S, RPD602421S, RPD603285S, RPD603437S, RPD603442S, RPD604046S, RPD604358S, RPD604419S, RPD604533S, RPD604904S, RPD604911S, RPD604933S, RPD605260S, RPD605344S, RPD605358S, RPD605419S, RPD605426S, RPD605514S, RPD605655S, RPD602241S, RPD604907S, RPD604508S, RPD605470S, RPD605550S, RPD601693S, RPD602228S, RPD603091S, RPD601563S, RPD602332S, RPD604653S, RPD601844S, RPD603164S, RPD604433S, RPD603535S, RPD603841S, RPD602702S, RPD604308S, RPD604310S, RPD601965S, RPD602741S, RPD601956S, RPD601967S, RPD602090S, RPD602468S, RPD602559S, RPD602965S, RPD603281S, RPD603884S, RPD604294S, RPD604185S, RPD601628S, RPD601706S, RPD601790S, RPD602927S, RPD603681S, RPD605364S, RPD601560S, RPD601675S, RPD601691S, RPD601707S, RPD601764S, RPD601905S, RPD602054S, RPD602311S, RPD604478S, RPD601621S, RPD602440S, RPD601627S, RPD601751S, RPD605038S, RPD605486S, RPD605399S, RPD605476S, RPD605488S, RPD605464S, RPD605469S, RPD605613S, RPD601918S, RPD602110S, RPD602784S, RPD603750S, RPD601594S, RPD601599S, RPD601922S, RPD602556S, RPD603304S, RPD604966S, RPD604997S, RPD601992S, RPD602935S, RPD601617S, RPD601753S, RPD601789S, RPD602141S, RPD602219S, RPD602387S, RPD602707S, RPD602887S, RPD603174S, RPD603217S, RPD603619S, RPD603751S, RPD604054S, RPD605068S, RPD605180S, RPD605214S, RPD605376S, RPD604903S, RPD604912S, RPD605405S, RPD605282S, RPD603734S, RPD601883S, RPD602314S, RPD604820S, RPD601508S, RPD601896S, RPD602032S, RPD602258S, RPD602325S, RPD602479S, RPD602493S, RPD602615S, RPD602680S, RPD603076S, RPD603307S, RPD604339S, RPD604389S, RPD604529S, RPD604632S, RPD604705S, RPD604869S, RPD604968S, RPD605025S, RPD605338S, RPD605513S, RPD601700S, RPD602215S, RPD602310S, RPD603543S, RPD603096S, RPD602473S, RPD602780S, RPD602914S, RPD603207S, RPD601634S, RPD601635S, RPD601701S, RPD601722S, RPD601957S, RPD603102S, RPD603287S, RPD601609S, RPD601630S, RPD601633S, RPD602128S, RPD602191S, RPD602194S, RPD603020S, RPD603700S, RPD604317S, RPD602329S, RPD604002S, RPD604488S, RPD605130S, RPD605485S, RPD601976S, RPD601977S, RPD602411S, RPD602562S, RPD602686S, RPD602974S, RPD603032S, RPD603055S, RPD603075S, RPD603089S, RPD603325S, RPD603594S, RPD603949S, RPD603954S, RPD604249S, RPD604646S, RPD604834S, RPD604844S, RPD604949S, RPD605016S, RPD602004S, RPD602622S, RPD602901S, RPD603244S, RPD603485S, RPD603883S, RPD604452S, RPD604956S, RPD605066S, RPD605228S, RPD602300S, RPD602671S, RPD603068S, RPD603366S, RPD603475S, RPD603624S, RPD603707S, RPD604218S, RPD604364S, RPD604948S, RPD605345S, RPD601973S, RPD602099S, RPD602849S, RPD601923S, RPD602386S, RPD602852S, RPD603106S, RPD602665S, RPD603382S, RPD603905S, RPD603948S, RPD604642S, RPD604752S, RPD604857S, RPD604548S, RPD605009S, RPD602265S, RPD602928S, RPD603951S, RPD602223S, RPD602400S, RPD604810S, RPD605277S, RPD605312S, RPD602904S, RPD603218S, RPD603507S, RPD604131S, RPD604723S, RPD605194S, RPD605297S, RPD605356S, RPD604564S, RPD604494S, RPD602006S, RPD602433S, RPD602593S, RPD602651S, RPD602829S, RPD603013S, RPD603116S, RPD603157S, RPD603455S, RPD603482S, RPD603659S, RPD603840S, RPD604549S, RPD604961S, RPD605167S, RPD601783S, RPD602232S, RPD605374S, RPD601509S, RPD601714S, RPD601804S, RPD601830S, RPD601994S, RPD602049S, RPD602121S, RPD602205S, RPD602477S, RPD602744S, RPD602853S, RPD603115S, RPD603447S, RPD605606S, RPD601886S, RPD605230S, RPD601584S, RPD601586S, RPD602603S, RPD602604S, RPD602950S, RPD602020S, RPD602345S, RPD603540S, RPD601662S, RPD601770S, RPD602016S, RPD602017S, RPD602975S, RPD604060S, RPD605136S, RPD601946S, RPD602261S, RPD602855S, RPD603026S, RPD605161S, RPD605216S, RPD605253S, RPD605365S, RPD605418S, RPD602399S, RPD601939S, RPD603219S, RPD604644S, RPD604662S, RPD604860S, RPD604932S, RPD605014S, RPD605031S, RPD605493S, RPD602008S, RPD601850S, RPD602010S, RPD602759S, RPD602836S, RPD602840S, RPD601255S, RPD601468S, RPD601485S, RPD601486S, RPD601498S, RPD601515S, RPD601516S, RPD601517S, RPD602222S, RPD602246S, RPD602247S, RPD602248S, RPD602249S, RPD602388S, RPD602389S, RPD602390S, RPD602391S, RPD602454S, RPD602456S, RPD602457S, RPD602544S, RPD602564S, RPD602698S, RPD602725S, RPD602727S, RPD602943S, RPD602944S, RPD603388S, RPD603389S, RPD603532S, RPD604319S, RPD604321S, RPD604322S, RPD604414S, RPD604457S, RPD604753S, RPD604754S, RPD604805S, RPD605241S, RPD605572S, RPD601747S, RPD601748S, RPD601749S, RPD601750S, RPD602372S, RPD602373S, RPD602376S, RPD602378S, RPD602873S, RPD602875S, RPD602876S, RPD602877S, RPD602878S, RPD602879S, RPD602880S, RPD602881S, RPD602882S, RPD602883S, RPD603319S, RPD603320S, RPD603321S, RPD603322S, RPD603324S, RPD604304S, RPD604306S, RPD604307S, RPD604309S, RPD604311S, RPD604559S, RPD604263S, RPD603249S, RPD605082S, RPD601479S, RPD601490S, RPD601492S, RPD601494S, RPD601500S, RPD601510S, RPD601519S, RPD601526S, RPD601527S, RPD601535S, RPD601551S, RPD601566S, RPD601568S, RPD601572S, RPD601699S, RPD601840S, RPD601841S, RPD601892S, RPD601893S, RPD601916S, RPD601925S, RPD601999S, RPD602000S, RPD602014S, RPD602055S, RPD602159S, RPD602176S, RPD602224S, RPD602287S, RPD602288S, RPD602424S, RPD602445S, RPD602760S, RPD602958S, RPD602960S, RPD603083S, RPD603084S, RPD603144S, RPD603145S, RPD603415S, RPD603417S, RPD603508S, RPD603509S, RPD603552S, RPD603554S, RPD603891S, RPD603892S, RPD603978S, RPD603979S, RPD603982S, RPD603984S, RPD604188S, RPD604190S, RPD604273S, RPD604348S, RPD604406S, RPD604441S, RPD604454S, RPD604596S, RPD604712S, RPD604867S, RPD604094S, RPD604762S, RPD603276S, RPD605526S, RPD604886S, RPD604905S, RPD603239S, RPD603240S, RPD603242S, RPD603545S, RPD604013S, RPD604014S, RPD604177S, RPD604590S, RPD604623S, RPD604624S, RPD604943S, RPD605080S, RPD605109S, RPD605110S, RPD605112S, RPD605199S, RPD601681S, RPD601975S, RPD602013S, RPD602048S, RPD602298S, RPD602304S, RPD602404S, RPD602530S, RPD602531S, RPD602535S, RPD602908S, RPD602909S, RPD602911S, RPD603113S, RPD604525S, RPD604527S, RPD601777S, RPD601779S, RPD603693S, RPD603929S, RPD601533S, RPD601660S, RPD603073S, RPD604619S, RPD605308S, RPD601867S, RPD601505S, RPD601507S, RPD601512S, RPD602237S, RPD605051S, RPD605529S, RPD605530S, RPD601489S, RPD601823S, RPD601826S, RPD601853S, RPD601857S, RPD602347S, RPD602640S, RPD604880S, RPD605126S, RPD605127S, RPD605261S, RPD602467S, RPD604808S, RPD602272S, RPD604402S, RPD605021S, RPD605583S, RPD605617S, RPD601483S, RPD601484S, RPD601542S, RPD601543S, RPD601752S, RPD601980S, RPD602108S, RPD602149S, RPD602150S, RPD602227S, RPD602461S, RPD602594S, RPD602673S, RPD602991S, RPD603099S, RPD603158S, RPD603284S, RPD603694S, RPD604222S, RPD604269S, RPD604283S, RPD604427S, RPD602976S, RPD602977S, RPD601544S, RPD603392S, RPD603393S, RPD603418S, RPD603419S, RPD603420S, RPD605592S, RPD604225S, RPD604713S, RPD601557S, RPD601872S, RPD605407S, RPD605541S, RPD605411S, RPD601470S, RPD601556S, RPD601756S, RPD601765S, RPD602147S, RPD602335S, RPD602342S, RPD602344S, RPD602382S, RPD602385S, RPD602398S, RPD602488S, RPD602534S, RPD602578S, RPD602627S, RPD602704S, RPD602714S, RPD603014S, RPD603092S, RPD603128S, RPD603210S, RPD603374S, RPD603387S, RPD603405S, RPD603843S, RPD604099S, RPD604352S, RPD604396S, RPD604444S, RPD604584S, RPD604633S, RPD604866S, RPD604885S, RPD604890S, RPD605002S, RPD605048S, RPD605142S, RPD605221S, RPD605612S, RPD605619S, RPD605653S, RPD603560S, RPD603873S, RPD604138S, RPD604331S, RPD604451S, RPD605113S, RPD604659S, RPD605114S, RPD605448S, RPD601712S, RPD601713S, RPD603477S, RPD603515S, RPD603516S, RPD603518S, RPD603637S, RPD603679S, RPD604028S, RPD604034S, RPD604039S, RPD604136S, RPD603856S, RPD604073S, RPD605193S, RPD605254S, RPD605255S, RPD605256S, RPD605569S, RPD605604S, RPD605623S, RPD605288S, RPD604040S, RPD601472S, RPD601579S, RPD601587S, RPD601680S, RPD601881S, RPD602216S, RPD602290S, RPD602670S, RPD602772S, RPD602834S, RPD603095S, RPD603130S, RPD603178S, RPD603429S, RPD603511S, RPD603618S, RPD603727S, RPD603898S, RPD604620S, RPD604756S, RPD601467S, RPD601480S, RPD601744S, RPD601879S, RPD601913S, RPD602027S, RPD602079S, RPD602166S, RPD602327S, RPD602623S, RPD602953S, RPD603041S, RPD603141S, RPD603151S, RPD603212S, RPD603214S, RPD603457S, RPD603839S, RPD603886S, RPD604043S, RPD605415S, RPD602271S, RPD602453S, RPD602662S, RPD603857S, RPD604718S, RPD605160S, RPD602888S, RPD604859S, RPD601493S, RPD605412S, RPD601888S, RPD602574S, RPD602827S, RPD603844S, RPD603871S, RPD604469S, RPD604505S, RPD604530S, RPD604707S, RPD604996S, RPD605121S, RPD601473S, RPD602370S, RPD603662S, RPD603930S, RPD604512S, RPD604661S, RPD604938S, RPD605355S, RPD605402S, RPD602005S, RPD603514S, RPD603833S, RPD604278S, RPD605539S, RPD601788S, RPD601898S, RPD601910S, RPD601914S, RPD601915S, RPD602078S, RPD602206S, RPD602282S, RPD602350S, RPD602368S, RPD602381S, RPD602502S, RPD602619S, RPD602659S, RPD602833S, RPD605120S, RPD605132S, RPD605394S, RPD605421S, RPD605618S, RPD601941S, RPD601942S, RPD602151S, RPD602614S, RPD602656S, RPD604011S, RPD605188S, RPD601928S, RPD602429S, RPD602448S, RPD602806S, RPD604301S, RPD604440S, RPD605040S, RPD605057S, RPD605449S, RPD605580S, RPD605607S, RPD601719S, RPD601792S, RPD602117S, RPD602351S, RPD602660S, RPD602739S, RPD602926S, RPD604058S, RPD604388S, RPD604884S, RPD605503S, RPD601948S, RPD602178S, RPD602279S, RPD604048S, RPD601906S, RPD602114S, RPD603079S, RPD604180S, RPD604260S, RPD601721S, RPD601947S, RPD604500S, RPD604695S, RPD603579S, RPD604711S, RPD601561S, RPD602089S, RPD602985S, RPD603471S, RPD604277S, RPD601497S, RPD601504S, RPD601589S, RPD601601S, RPD601602S, RPD601679S, RPD602874S, RPD603481S, RPD603583S, RPD603763S, RPD604698S, RPD604798S, RPD602520S, RPD601668S, RPD602214S, RPD604490S, RPD602029S, RPD602413S, RPD604245S, RPD605570S, RPD601757S, RPD605317S, RPD603778S, RPD604629S, RPD604842S, RPD605329S, RPD605434S, RPD605383S, RPD601524S, RPD602060S, RPD602136S, RPD602317S, RPD602356S, RPD602431S, RPD602507S, RPD602565S, RPD603701S, RPD603709S, RPD603753S, RPD604368S, RPD604426S, RPD604554S, RPD604602S, RPD604694S, RPD605281S, RPD601694S, RPD602295S, RPD603342S, RPD601936S, RPD604648S, RPD602134S, RPD602733S, RPD602025S, RPD602536S, RPD602712S, RPD601702S, RPD601931S, RPD601932S, RPD601993S, RPD602899S, RPD603401S, RPD601703S, RPD603541S, RPD603558S, RPD604796S, RPD605157S, RPD605373S, RPD605462S, RPD602742S, RPD601769S, RPD602109S, RPD601745S, RPD601845S, RPD602629S, RPD603182S, RPD604346S, RPD605321S, RPD602292S, RPD602588S, RPD604100S, RPD604328S, RPD605460S, RPD605527S, RPD601605S, RPD601698S, RPD602063S, RPD603604S, RPD603643S, RPD604018S, RPD604199S, RPD604430S, RPD604522S, RPD603996S, RPD604126S, RPD604899S, RPD605088S, RPD605151S, RPD605170S, RPD605262S, RPD604657S, RPD604511S, RPD603730S, RPD602051S, RPD602949S, RPD603172S, RPD604143S, RPD604622S, RPD601908S, RPD604183S, RPD604243S, RPD601771S, RPD602367S, RPD602442S, RPD603472S, RPD604033S, RPD602093S, RPD602188S, RPD602773S, RPD603043S, RPD603071S, RPD603858S, RPD601641S, RPD601649S, RPD604763S, RPD605064S, RPD601595S, RPD602706S, RPD603305S, RPD604369S, RPD602729S, RPD602730S, RPD603402S, RPD602193S, RPD603282S, RPD604636S, RPD605153S, RPD605210S, RPD605238S, RPD601541S, RPD601626S, RPD601920S, RPD602474S, RPD604989S, RPD605274S, RPD605512S, RPD602126S, RPD602130S, RPD602768S, RPD604677S, RPD605420S, RPD605521S, RPD605149S, RPD602096S, RPD602180S, RPD603970S, RPD603122S, RPD602050S, RPD602392S, RPD602395S, RPD602917S, RPD602920S, RPD602923S, RPD603349S, RPD603720S, RPD603721S, RPD601476S, RPD601596S, RPD601597S, RPD604550S, RPD605147S, RPD601720S, RPD601774S, RPD605634S, RPD605424S, RPD605478S, RPD602074S, RPD602119S, RPD602189S, RPD602481S, RPD602703S, RPD603461S, RPD603557S, RPD603735S, RPD604213S, RPD604465S, RPD604479S, RPD604908S, RPD605280S, RPD605354S, RPD605611S, RPD605616S, RPD605571S, RPD604827S, RPD604950S, RPD605125S, RPD605325S, RPD605538S, RPD601797S, RPD602045S, RPD602059S, RPD602701S, RPD604195S, RPD604809S, RPD605003S, RPD605273S, RPD605629S, RPD604472S, RPD601613S, RPD601909S, RPD602190S, RPD605534S, RPD605622S, RPD602715S, RPD604519S, RPD605091S, RPD603525S, RPD603528S, RPD604178S, RPD601669S, RPD604934S, RPD605437S, RPD605445S, RPD605498S, RPD604447S, RPD604652S, RPD605119S, RPD605396S, RPD602508S, RPD602528S, RPD602737S, RPD604431S, RPD602285S, RPD602525S, RPD603328S, RPD602630S, RPD604335S, RPD604424S, RPD604778S, RPD604852S, RPD602357S, RPD602568S, RPD602668S, RPD602685S, RPD602971S, RPD603001S, RPD603931S, RPD604701S, RPD604767S, RPD602810S, RPD602605S, RPD603354S, RPD603358S, RPD603818S, RPD603716S, RPD603963S, RPD602132S, RPD603725S, RPD603918S, RPD601815S, RPD603017S, RPD603170S, RPD604702S, RPD604420S, RPD601729S, RPD601951S, RPD603018S, RPD603794S, RPD603805S, RPD604229S, RPD604607S, RPD604965S, RPD601735S, RPD601740S, RPD602198S, RPD603777S, RPD603783S, RPD603808S, RPD604171S, RPD602632S, RPD603599S, RPD603821S, RPD604603S, RPD603913S, RPD601659S, RPD601730S, RPD601817S, RPD603011S, RPD603012S, RPD603015S, RPD603919S, RPD603945S, RPD604027S, RPD604208S, RPD604259S, RPD604881S, RPD604883S, RPD601762S, RPD601812S, RPD601814S, RPD601944S, RPD602522S, RPD602801S, RPD603006S, RPD603057S, RPD603165S, RPD603626S, RPD604262S, RPD604330S, RPD601657S, RPD601800S, RPD601997S, RPD602039S, RPD602354S, RPD602533S, RPD603822S, RPD604570S, RPD601801S, RPD602761S, RPD603795S, RPD603807S, RPD604010S, RPD604536S, RPD604641S, RPD604660S, RPD601645S, RPD604088S, RPD604095S, RPD601647S, RPD602860S, RPD603802S, RPD604087S, RPD604555S, RPD604240S, RPD601644S, RPD601670S, RPD601683S, RPD602438S, RPD602716S, RPD602718S, RPD602719S, RPD603531S, RPD603597S, RPD603715S, RPD603736S, RPD603988S, RPD604068S, RPD604446S, RPD604518S, RPD604535S, RPD604693S, RPD604696S, RPD604699S, RPD604802S, RPD605453S, RPD605500S, RPD603547S, RPD603787S, RPD604061S, RPD604329S, RPD605391S, RPD605665S, RPD605578S, RPD605219S, RPD604836S, RPD604579S, RPD603674S, RPD605222S, RPD604635S, RPD605085S, RPD601996S, RPD602567S, RPD603632S, RPD604367S, RPD601656S, RPD602856S, RPD603293S, RPD603771S, RPD603932S, RPD604098S, RPD604101S, RPD604084S, RPD601808S, RPD603917S, RPD604235S, RPD604379S, RPD604918S, RPD604960S, RPD605148S, RPD602131S, RPD602483S, RPD602495S, RPD602758S, RPD602984S, RPD603147S, RPD603819S, RPD603960S, RPD601650S, RPD601739S, RPD601849S, RPD601961S, RPD602484S, RPD602597S, RPD602787S, RPD602803S, RPD602942S, RPD603008S, RPD603134S, RPD603724S, RPD603825S, RPD603933S, RPD603986S, RPD603991S, RPD604006S, RPD604080S, RPD604205S, RPD604376S, RPD604378S, RPD604601S, RPD604674S, RPD604710S, RPD604777S, RPD604748S, RPD602503S, RPD603786S, RPD604326S, RPD604726S, RPD601646S, RPD602501S, RPD603955S, RPD604759S, RPD605632S, RPD602939S, RPD604720S, RPD603463S, RPD604684S, RPD605630S, RPD605079S, RPD604090S, RPD604840S, RPD605189S, RPD604864S, RPD602521S, RPD602858S, RPD603772S, RPD603804S, RPD602319S, RPD604952S, RPD605556S, RPD605056S, RPD605229S, RPD601654S, RPD601854S, RPD601979S, RPD602068S, RPD602763S, RPD602983S, RPD603004S, RPD603186S, RPD603685S, RPD603774S, RPD604568S, RPD604613S, RPD603593S, RPD602613S, RPD603684S, RPD603944S, RPD602717S, RPD605218S, RPD604108S, RPD601809S, RPD603683S, RPD603942S, RPD604486S, RPD601652S, RPD601761S, RPD602655S, RPD602947S, RPD603002S, RPD603167S, RPD603676S, RPD603915S, RPD603921S, RPD604786S, RPD601763S, RPD601971S, RPD602057S, RPD602312S, RPD602524S, RPD602579S, RPD602602S, RPD602654S, RPD602767S, RPD603183S, RPD603185S, RPD603193S, RPD603803S, RPD603820S, RPD603989S, RPD603993S, RPD604203S, RPD604422S, RPD604534S, RPD604758S, RPD604791S, RPD602072S, RPD602487S, RPD604382S, RPD604539S, RPD604575S, RPD604580S, RPD604582S, RPD604663S, RPD604876S, RPD602510S, RPD602620S, RPD602809S, RPD603687S, RPD603965S, RPD601929S, RPD603665S, RPD603738S, RPD603744S, RPD603809S, RPD604106S, RPD604567S, RPD604697S, RPD604769S, RPD603010S, RPD604704S, RPD603790S, RPD602494S, RPD602498S, RPD603691S, RPD602504S, RPD603798S, RPD603934S, RPD604571S, RPD603926S, RPD602042S, RPD602720S, RPD602786S, RPD603007S, RPD603168S, RPD603768S, RPD603779S, RPD603912S, RPD604071S, RPD604572S, RPD602774S, RPD604773S, RPD601728S, RPD602636S, RPD604615S, RPD604741S, RPD604863S, RPD605165S, RPD601799S, RPD604156S, RPD604621S, RPD604946S, RPD601651S, RPD601821S, RPD602648S, RPD602764S, RPD601818S, RPD601959S, RPD603152S, RPD603761S, RPD603959S, RPD604538S, RPD604766S, RPD603299S, RPD601806S, RPD602486S, RPD602762S, RPD604547S, RPD602789S, RPD603824S, RPD603682S, RPD601820S, RPD602586S, RPD603159S, RPD603671S, RPD603764S, RPD603814S, RPD603964S, RPD604009S, RPD604110S, RPD601966S, RPD602116S, RPD602569S, RPD602572S, RPD602785S, RPD603003S, RPD603745S, RPD603811S, RPD603817S, RPD604078S, RPD604111S, RPD604155S, RPD604192S, RPD604200S, RPD604588S, RPD604605S, RPD604750S, RPD604770S, RPD604775S, RPD603766S, RPD602663S, RPD603638S, RPD604116S, RPD604604S, RPD604751S, RPD604789S, RPD603813S, RPD601704S, RPD602802S, RPD601982S, RPD602757S, RPD602788S, RPD604413S, RPD604760S, RPD602997S, RPD604768S, RPD601723S, RPD603607S, RPD601731S, RPD601856S, RPD602794S, RPD603150S, RPD603169S, RPD604716S, RPD602721S, RPD603154S, RPD603209S, RPD603746S, RPD603801S, RPD603916S, RPD604115S, RPD601726S, RPD603666S, RPD602796S, RPD604714S, RPD601737S, RPD602589S, RPD603828S, RPD605015S, RPD602153S, RPD604158S, RPD605387S, RPD601732S, RPD601895S, RPD601653S, RPD601727S, RPD601760S, RPD604003S, RPD603204S, RPD604238S, RPD602756S, RPD602916S, RPD603780S, RPD603816S, RPD604016S, RPD604215S, RPD605417S, RPD601591S, RPD604206S, RPD604384S, RPD601663S, RPD601664S, RPD601689S, RPD604067S, RPD604846S, RPD604847S, RPD604849S, RPD604963S, RPD605032S, RPD605097S, RPD605115S, RPD605128S, RPD605263S, RPD605410S, RPD601496S, RPD601528S, RPD601534S, RPD601554S, RPD601575S, RPD601576S, RPD601875S, RPD601882S, RPD601884S, RPD601885S, RPD602069S, RPD602087S, RPD602182S, RPD602415S, RPD602416S, RPD602418S, RPD602450S, RPD602452S, RPD602546S, RPD602553S, RPD602711S, RPD602825S, RPD602826S, RPD603527S, RPD603551S, RPD604086S, RPD604287S, RPD604332S, RPD604649S, RPD604797S, RPD604854S, RPD605553S, RPD602551S, RPD603274S, RPD604586S, RPD601873S, RPD601902S, RPD602092S, RPD602570S, RPD602573S, RPD603348S, RPD603848S, RPD603868S, RPD604129S, RPD604353S, RPD604363S, RPD604495S, RPD601837S, RPD601870S, RPD601871S, RPD602138S, RPD602266S, RPD602306S, RPD602547S, RPD603205S, RPD603496S, RPD603697S, RPD604489S, RPD605398S, RPD605423S, RPD603520S, RPD604285S, RPD604574S, RPD604578S, RPD604630S, RPD604947S, RPD605098S, RPD605185S, RPD605237S, RPD605340S, RPD605008S, RPD605183S, RPD605413S, RPD604671S, RPD605069S, RPD604318S, RPD605451S, RPD604670S, RPD604371S, RPD601846S, RPD603625S, RPD603908S, RPD604403S, RPD605452S, RPD605461S, RPD602264S, RPD603253S, RPD603255S, RPD604258S, RPD604668S, RPD602842S, RPD603016S, RPD604541S, RPD604316S, RPD601984S, RPD601985S, RPD602062S, RPD602268S, RPD602529S, RPD603053S, RPD603652S, RPD604246S, RPD604334S, RPD604875S, RPD605047S, RPD605058S, RPD605509S, RPD605658S, RPD602795S, RPD601710S, RPD601711S, RPD601775S, RPD602211S, RPD605217S, RPD603629S, RPD604669S, RPD601945S, RPD605022S, RPD601648S, RPD601847S, RPD602805S, RPD604665S, RPD603110S, RPD602175S, RPD602866S, RPD603142S, RPD603386S, RPD604375S, RPD604627S, RPD604942S, RPD604959S, RPD605438S, RPD604544S, RPD605323S, RPD605324S, RPD602691S, RPD602894S, RPD604008S, RPD602869S, RPD602539S, RPD602678S, RPD603880S, RPD604573S, RPD605525S, RPD602948S, RPD603160S, RPD603251S, RPD604296S, RPD604394S, RPD604439S, RPD604749S, RPD604812S, RPD604830S, RPD604972S, RPD605000S, RPD605071S, RPD603907S, RPD604374S, RPD605520S, RPD605380S, RPD604772S, RPD604889S, RPD604551S, RPD604887S, RPD604888S, RPD605137S, RPD605138S, RPD605243S, RPD605431S, RPD605454S, RPD605441S, RPD605599S, RPD604122S, RPD604437S, RPD602595S; GTIN 00763000206277, Lot Serial Numbers: RPD606206S, RPD606101S, RPD605877S, RPD605780S, RPD606953S, RPD606951S, RPD606675S, RPD606663S, RPD606086S, RPD606660S, RPD607039S, RPD606423S, RPD607136S, RPD606292S, RPD607184S, RPD607205S, RPD607104S, RPD606707S, RPD607162S, RPD607070S, RPD606430S, RPD607085S, RPD607152S, RPD606242S, RPD606258S, RPD607175S, RPD606480S, RPD607202S, RPD606779S, RPD606578S, RPD606327S, RPD606504S, RPD606622S, RPD607089S, RPD607170S, RPD605732S, RPD605738S, RPD605813S, RPD605834S, RPD605998S, RPD606015S, RPD606173S, RPD606243S, RPD606312S, RPD606328S, RPD606515S, RPD606640S, RPD606641S, RPD607124S, RPD605767S, RPD606014S, RPD606120S, RPD606132S, RPD606136S, RPD606237S, RPD606349S, RPD606357S, RPD606572S, RPD605756S, RPD605944S, RPD606728S, RPD605805S, RPD605806S, RPD605909S, RPD606109S, RPD606356S, RPD606384S, RPD606442S, RPD606453S, RPD606588S, RPD606649S, RPD606712S, RPD606734S, RPD606757S, RPD606802S, RPD606176S, RPD605695S, RPD606061S, RPD606610S, RPD606188S, RPD605773S, RPD605748S, RPD606161S, RPD606330S, RPD606459S, RPD606211S, RPD606185S, RPD606104S, RPD606183S, RPD606539S, RPD606540S, RPD606807S, RPD605747S, RPD605984S, RPD605987S, RPD606042S, RPD606074S, RPD606381S, RPD605685S, RPD605770S, RPD605887S, RPD605981S, RPD606005S, RPD606268S, RPD606372S, RPD606389S, RPD606390S, RPD606603S, RPD606965S, RPD605860S, RPD606043S, RPD606393S, RPD606447S, RPD606484S, RPD606488S, RPD606518S, RPD605929S, RPD605958S, RPD605979S, RPD606361S, RPD605830S, RPD605914S, RPD605994S, RPD606122S, RPD606358S, RPD606360S, RPD606623S, RPD606956S, RPD607119S, RPD606410S, RPD605757S, RPD606160S, RPD605925S, RPD605959S, RPD605985S, RPD606168S, RPD606253S, RPD606409S, RPD606552S, RPD606711S, RPD605787S, RPD605826S, RPD605882S, RPD606190S, RPD606353S, RPD606398S, RPD606565S, RPD606753S, RPD606806S, RPD606964S, RPD605760S, RPD606230S, RPD606252S, RPD606045S, RPD606090S, RPD606351S, RPD605836S, RPD606506S, RPD606650S, RPD606544S, RPD606594S, RPD606569S, RPD605718S, RPD605950S, RPD606076S, RPD606228S, RPD606234S, RPD606256S, RPD606708S, RPD606788S, RPD605792S, RPD605846S, RPD606073S, RPD606494S, RPD606732S, RPD606770S, RPD607038S, RPD606195S, RPD606369S, RPD605997S, RPD606232S, RPD606407S, RPD606408S, RPD606321S, RPD606200S, RPD606426S, RPD606781S, RPD606265S, RPD606595S, RPD605939S, RPD606092S, RPD606803S, RPD606009S, RPD606214S, RPD606235S, RPD605861S, RPD605844S, RPD606940S, RPD606085S, RPD606087S, RPD606100S, RPD606150S, RPD606264S, RPD605938S, RPD606202S, RPD606450S, RPD606279S, RPD605843S, RPD606001S, RPD605774S, RPD606201S, RPD606933S, RPD605777S, RPD605458S, RPD606449S, RPD606320S, RPD606046S, RPD606547S, RPD607054S, RPD606524S, RPD606726S, RPD606226S, RPD606297S, RPD606392S, RPD606937S, RPD605775S, RPD605818S, RPD606551S, RPD606970S, RPD605866S, RPD605891S, RPD606018S, RPD606530S, RPD606029S, RPD606182S, RPD605771S, RPD605902S, RPD605786S, RPD605986S, RPD605859S, RPD605863S, RPD605868S, RPD605879S, RPD605884S, RPD606115S, RPD606117S, RPD606146S, RPD606147S, RPD606584S, RPD606585S, RPD606586S, RPD606627S, RPD606796S, RPD606797S, RPD606800S, RPD606934S, RPD606935S, RPD606968S, RPD606969S, RPD607044S, RPD607045S, RPD607046S, RPD605926S, RPD605809S, RPD605831S, RPD605832S, RPD605923S, RPD605924S, RPD606397S, RPD607139S, RPD605942S, RPD605943S, RPD605731S, RPD606520S, RPD606591S, RPD606597S, RPD606784S, RPD605758S, RPD605815S, RPD606112S, RPD605721S, RPD606193S, RPD605682S, RPD606078S, RPD606303S, RPD606304S, RPD606305S, RPD605829S, RPD606221S, RPD606503S, RPD606138S, RPD605821S, RPD605960S, RPD605871S, RPD605917S, RPD606296S, RPD606205S, RPD606225S, RPD606064S, RPD606066S, RPD605802S, RPD606059S, RPD606125S, RPD606425S, RPD606941S, RPD607033S, RPD605895S, RPD605912S, RPD606002S, RPD606224S, RPD606567S, RPD605736S, RPD605755S, RPD605906S, RPD606008S, RPD606035S, RPD606041S, RPD606103S, RPD606141S, RPD606145S, RPD606153S, RPD606215S, RPD606267S, RPD606290S, RPD606299S, RPD606316S, RPD606366S, RPD606432S, RPD606439S, RPD606771S, RPD606975S, RPD607057S, RPD607058S, RPD607079S, RPD607083S, RPD607107S, RPD606926S, RPD606774S, RPD605692S, RPD605729S, RPD605681S, RPD605705S, RPD605857S, RPD606485S, RPD606513S, RPD605457S, RPD605697S, RPD605934S, RPD606808S, RPD605980S, RPD605724S, RPD605842S, RPD606097S, RPD606218S, RPD605766S, RPD606399S, RPD605708S, RPD605712S, RPD606044S, RPD606094S, RPD606805S, RPD606332S, RPD605878S, RPD606343S, RPD605875S, RPD606051S, RPD606140S, RPD606382S, RPD605900S, RPD606134S, RPD606217S, RPD607077S, RPD605963S, RPD605964S, RPD606110S, RPD606313S, RPD606433S, RPD606444S, RPD606063S, RPD606164S, RPD606284S, RPD605948S, RPD606024S, RPD606531S, RPD606039S, RPD606095S, RPD605811S, RPD605794S, RPD605797S, RPD606413S, RPD606414S, RPD606714S, RPD606790S, RPD605949S, RPD606311S, RPD605881S, RPD606058S, RPD606071S, RPD606159S, RPD606255S, RPD606599S, RPD606609S, RPD605768S, RPD606020S, RPD606093S, RPD605983S, RPD606546S, RPD606838S, RPD607122S, RPD607176S, RPD607187S, RPD605898S, RPD606124S, RPD607043S, RPD605965S, RPD605966S, RPD605933S, RPD606111S, RPD606370S, RPD606479S, RPD606760S, RPD605982S, RPD605810S, RPD605858S, RPD605955S, RPD606262S, RPD606394S, RPD605968S, RPD605817S, RPD605928S, RPD606114S, RPD606162S, RPD606363S, RPD606616S, RPD605931S, RPD605936S, RPD605865S, RPD605978S, RPD606391S, RPD607037S, RPD607086S, RPD607093S, RPD607094S, RPD606048S, RPD605855S, RPD606013S, RPD605688S, RPD605839S, RPD605873S, RPD606248S, RPD606011S, RPD606386S, RPD606269S, RPD606342S, RPD605687S, RPD606314S, RPD606010S, RPD606130S, RPD606346S, RPD606613S, RPD606804S, RPD606181S, RPD606946S, RPD605689S, RPD605759S, RPD605913S, RPD605991S, RPD606007S, RPD606034S, RPD606152S, RPD606729S, RPD605812S, RPD606568S, RPD607071S, RPD605907S, RPD605961S, RPD606102S, RPD606119S, RPD606179S, RPD606239S, RPD606285S, RPD606359S, RPD606493S, RPD606938S, RPD607051S, RPD605916S, RPD606289S, RPD606505S, RPD606529S, RPD606961S, RPD606223S, RPD606574S, RPD606643S, RPD605864S, RPD605799S, RPD606785S, RPD606801S, RPD606033S, RPD605781S, RPD606135S, RPD606958S, RPD606016S, RPD606295S, RPD606500S, RPD606636S, RPD606089S, RPD606395S, RPD606455S, RPD606203S, RPD606080S, RPD606038S, RPD605735S, RPD606501S, RPD606763S, RPD607040S, RPD605823S, RPD606065S, RPD606131S, RPD606244S, RPD606245S, RPD606318S, RPD606562S, RPD607146S, RPD606713S, RPD606792S, RPD605977S, RPD605856S, RPD605749S, RPD605751S, RPD605753S, RPD605754S, RPD605819S, RPD605820S, RPD605862S, RPD605897S, RPD606027S, RPD606028S, RPD606030S, RPD606031S, RPD606068S, RPD606082S, RPD606083S, RPD606108S, RPD606175S, RPD606764S, RPD606765S, RPD606767S, RPD606769S, RPD607061S, RPD607062S, RPD607063S, RPD606056S, RPD607048S, RPD606233S, RPD605778S, RPD605940S, RPD606036S, RPD606154S, RPD606238S, RPD606276S, RPD606277S, RPD606322S, RPD606336S, RPD606428S, RPD606560S, RPD606587S, RPD606986S, RPD606987S, RPD606990S, RPD606991S, RPD606992S, RPD606993S, RPD606994S, RPD606766S, RPD605798S, RPD605993S, RPD606424S, RPD605807S, RPD605707S, RPD605870S, RPD605937S, RPD606128S, RPD606129S, RPD606338S, RPD606555S, RPD606052S, RPD606155S, RPD606050S, RPD606379S, RPD605693S, RPD605730S, RPD605905S, RPD606084S, RPD606624S, RPD605869S, RPD605893S, RPD605927S, RPD606207S, RPD605459S, RPD605765S, RPD606025S, RPD606099S, RPD606236S, RPD606250S, RPD605947S, RPD606067S, RPD606139S, RPD606288S, RPD606754S, RPD606949S, RPD606273S, RPD605988S, RPD605992S, RPD606003S, RPD606004S, RPD606127S, RPD606521S, RPD605680S, RPD605690S, RPD605691S, RPD605714S, RPD605828S, RPD605847S, RPD606174S, RPD606287S, RPD606549S, RPD606601S, RPD605952S, RPD605969S, RPD605970S, RPD605971S, RPD605972S, RPD605973S, RPD606032S, RPD606166S, RPD606308S, RPD605974S, RPD605975S, RPD605976S, RPD606106S, RPD606165S, RPD606374S, RPD606419S, RPD606495S, RPD606554S, RPD606756S, RPD606762S, RPD605716S, RPD605717S, RPD606091S, RPD606260S, RPD606375S, RPD606583S, RPD606309S, RPD606496S, RPD606596S, RPD605922S, RPD606049S, RPD606057S, RPD606310S, RPD606326S, RPD606541S, RPD606625S, RPD606723S, RPD606307S, RPD606335S, RPD606557S, RPD606072S, RPD605711S, RPD605827S, RPD605889S, RPD606337S, RPD605709S, RPD605710S, RPD605918S, RPD606017S, RPD606126S, RPD606133S, RPD606491S, RPD606963S, RPD605727S, RPD605728S, RPD605880S, RPD605885S, RPD605962S, RPD606151S, RPD606752S, RPD606280S, RPD605874S, RPD606445S, RPD605945S, RPD606553S, RPD606422S, RPD606070S, RPD605935S, RPD606502S, RPD606257S, RPD605930S, RPD606365S, RPD606487S, RPD606783S, RPD606054S, RPD606158S, RPD606227S, RPD606261S, RPD606271S, RPD606460S, RPD606037S, RPD606725S, RPD605772S, RPD605845S, RPD606019S, RPD606612S, RPD606942S, RPD606012S, RPD606508S, RPD606510S, RPD606062S, RPD606344S, RPD606298S, RPD606184S, RPD605723S, RPD606882S, RPD605725S, RPD605837S, RPD605990S, RPD605784S, RPD606341S, RPD606352S, RPD606836S, RPD605734S, RPD605752S, RPD605841S, RPD605696S, RPD605849S, RPD606415S, RPD606581S, RPD605901S, RPD606517S, RPD606060S, RPD606247S, RPD606651S, RPD606606S, RPD607116S, RPD606098S, RPD606532S, RPD605698S, RPD605722S, RPD606478S, RPD606676S, RPD606055S, RPD606143S, RPD605761S, RPD605785S, RPD606470S, RPD606177S, RPD606761S, RPD606123S, RPD606105S, RPD606673S, RPD605769S, RPD605800S, RPD606512S, RPD605953S, RPD605956S, RPD605954S, RPD605957S, RPD606121S, RPD606194S, RPD605795S, RPD606144S, RPD606464S, RPD606192S, RPD606404S, RPD605715S, RPD605726S, RPD606254S, RPD605713S, RPD605719S, RPD605808S, RPD605840S, RPD605967S, RPD606208S, RPD606241S, RPD606291S, RPD606325S, RPD606492S, RPD606590S, RPD607026S, RPD607123S, RPD607133S, RPD607134S, RPD607135S, RPD606148S, RPD606229S, RPD606281S, RPD605852S, RPD605853S, RPD606685S, RPD606686S, RPD606737S, RPD606738S, RPD606739S, RPD606817S, RPD606821S, RPD606822S, RPD606824S, RPD606825S, RPD606826S, RPD606827S, RPD606841S, RPD606843S, RPD606844S, RPD606845S, RPD606846S, RPD606847S, RPD606848S, RPD606747S, RPD606748S, RPD606749S, RPD606849S, RPD606850S, RPD606851S, RPD606853S, RPD606855S, RPD606856S, RPD606857S, RPD606998S, RPD607000S, RPD605702S, RPD605700S, RPD605701S, RPD606402S, RPD606180S, RPD606872S, RPD606873S, RPD606874S, RPD606895S, RPD606896S, RPD606897S, RPD605995S, RPD606088S, RPD606317S, RPD606693S, RPD606706S, RPD606854S, RPD606859S, RPD606860S, RPD606861S, RPD606862S, RPD606863S, RPD606864S, RPD606865S, RPD606811S, RPD606812S, RPD606813S, RPD606814S, RPD606815S, RPD606026S, RPD606350S, RPD606441S, RPD606577S, RPD605694S, RPD605699S, RPD607028S, RPD607030S, RPD607035S, RPD606656S, RPD606694S, RPD606695S, RPD606696S, RPD606697S, RPD606698S, RPD606699S, RPD606700S, RPD606702S, RPD606703S, RPD606704S, RPD606705S, RPD606736S, RPD606852S, RPD606858S, RPD606866S, RPD606867S, RPD606868S, RPD606869S, RPD606870S, RPD606871S, RPD607001S, RPD605720S, RPD606558S, RPD606006S, RPD606983S, RPD605999S, RPD605932S, RPD606157S, RPD606209S, RPD606376S, RPD606928S, RPD606403S, RPD606405S, RPD605816S, RPD605850S, RPD605851S, RPD605854S, RPD606628S, RPD606629S, RPD606630S, RPD606631S, RPD606632S, RPD606633S, RPD606634S, RPD606658S, RPD606679S, RPD606680S, RPD606681S, RPD606682S, RPD606683S, RPD606684S, RPD606740S, RPD606741S, RPD606742S, RPD606743S, RPD606744S, RPD606745S, RPD606746S, RPD606875S, RPD606876S, RPD606877S, RPD606878S, RPD606879S, RPD606880S, RPD606884S, RPD606885S, RPD606886S, RPD606887S, RPD606652S, RPD606653S, RPD606654S, RPD606655S, RPD606687S, RPD606688S, RPD606689S, RPD606690S, RPD606691S, RPD606692S, RPD605782S, RPD605804S, RPD605989S, RPD606075S, RPD606113S, RPD606142S, RPD606163S, RPD606169S, RPD606170S, RPD606187S, RPD606213S, RPD606263S, RPD606278S, RPD606282S, RPD606286S, RPD606306S, RPD606315S, RPD606329S, RPD606345S, RPD606347S, RPD606354S, RPD606362S, RPD606367S, RPD606373S, RPD606396S, RPD606418S, RPD606421S, RPD606438S, RPD606451S, RPD606452S, RPD606456S, RPD606461S, RPD606463S, RPD606465S, RPD606469S, RPD606482S, RPD606509S, RPD606511S, RPD606514S, RPD606519S, RPD606523S, RPD606533S, RPD606550S, RPD606559S, RPD606561S, RPD606564S, RPD606570S, RPD606575S, RPD606600S, RPD606619S, RPD606620S, RPD606646S, RPD606647S, RPD606648S, RPD606661S, RPD606670S, RPD606672S, RPD606716S, RPD606717S, RPD606724S, RPD606727S, RPD606730S, RPD606731S, RPD606751S, RPD606758S, RPD606759S, RPD606768S, RPD606775S, RPD606778S, RPD606787S, RPD606789S, RPD606793S, RPD606798S, RPD606839S, RPD606842S, RPD606893S, RPD606936S, RPD606952S, RPD606955S, RPD606962S, RPD606972S, RPD606978S, RPD606996S, RPD607011S, RPD607027S, RPD607036S, RPD607041S, RPD607042S, RPD607052S, RPD607055S, RPD607060S, RPD607074S, RPD607081S, RPD607090S, RPD607091S, RPD607092S, RPD607095S, RPD607096S, RPD607100S, RPD607101S, RPD607102S, RPD607103S, RPD607106S, RPD607125S, RPD607127S, RPD607128S, RPD607129S, RPD607130S, RPD607142S, RPD607153S, RPD607157S, RPD607164S, RPD607165S, RPD607166S, RPD607174S, RPD607177S, RPD607182S, RPD607185S, RPD607192S, RPD607194S, RPD607195S, RPD607197S, RPD607204S, RPD607209S, RPD606525S, RPD607173S, RPD607179S, RPD607196S, RPD607206S, RPD607208S, RPD606251S, RPD606266S, RPD606431S, RPD606466S, RPD606516S, RPD606593S, RPD607120S, RPD607147S, RPD607156S, RPD607178S
FEI Number 2182208
Recalling Firm/
Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
8200 Coral Sea St Ne
Mounds View MN 55112-4391
For Additional Information ContactTechnical Services
800-929-4043
Manufacturer Reason
for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Quantity in Commerce4593 units
DistributionUS Nationwide - Worldwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA DatabasePMAs with Product Code = NIK
PMAs with Product Code = NIK
-
-