• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Nova Stat Profile Prime Plus Analyzer

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Nova Stat Profile Prime Plus Analyzersee related information
Date Initiated by FirmSeptember 18, 2023
Date PostedNovember 01, 2023
Recall Status1 Open3, Classified
Recall NumberZ-0208-2024
Recall Event ID 93095
510(K)NumberK200403 
Product Classification Electrode, ion specific, sodium - Product Code JGS
ProductNova Stat Profile Prime Plus POC Analyzer (CE)-Intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. (1) PN 63368 Stat Profile Prime Plus POC Analyzer (2) PN 63369 Stat Profile Prime Plus POC Analyzer, Remanufactured P/N: 57400, 63368
Code Information UDI-DI: N/A All software versions: International Except China: W1.17.1123.0 and W3.2.1123.0 V1.50.1123.0 China: V1.50.1123.0
Recalling Firm/
Manufacturer
Nova Biomedical Corporation
200 Prospect St
Waltham MA 02453-3407
For Additional Information ContactTheresa Austin
781-647-3700
Manufacturer Reason
for Recall
Potential risk of reporting erroneous creatinine patient results associated with the initial start-up of a new BUN/Creatinine Sensor Cartridge. The potential risk is within the first 2-4 hours after cartridge installation that may not be identified by traditional Quality Control practices.
FDA Determined
Cause 2
Process design
ActionNova Biomedical Customer Success Managers starting 9/18/23 contacted consignees directly via telephone, to review the bulletin with the consignee DRP, verify an email to send the bulletin to them, and then immediately forward the bulletin to them so they can share within their own healthcare facility. Letter states the reason for recall, health risk and action to take: Nova Biomedical recommends that your healthcare facility take one of the following risk mitigation steps after a BUN/Creatinine Sensor Cartridge replacement to reduce the possibility of this issue occurring in your facility until your analyzers have received a software update. " If your analyzer tests patient samples frequently, initiate a manual analyzer calibration every 30 minutes within the first 2 hours, and then every hour for the next 2 hours after cartridge replacement. " If your analyzer tests patient samples infrequently, within the first 4 hours after cartridge replacement, initiate a manual analyzer calibration prior to performing a patient test. Until your Prime Plus analyzer(s) software is updated, if your healthcare facility experiences an erroneous creatinine result or adverse event after a new sensor cartridge installation, Nova encourages you to contact Nova Biomedical or your authorized distributor to properly document the incident. For adverse event(s), Nova encourages you to report the details using the MedWatch Online Voluntary Reporting Form, MedWatch Online Voluntary Reporting Form (fda.gov). INTERNAL HEALTHCARE FACILITY COMMUNICATION: This Field Correction Notice should be distributed to all those within your organization who need to be aware of this issue. Please notify other parties affiliated with your organization, which are affected by this notice. Please maintain awareness of this notice until the issue has been corrected with a scheduled software update. If you have any questions on the contents of this notice, would like to discuss other risk mitigation options
Quantity in Commerce161 units
DistributionNationwide Foreign: Country Algeria Australia Austria Bangladesh Benin Bolivia Brazil Canada Chile China Colombia Croatia CYPRUS Czech Republic Ecuador Egypt El Salvador Finland France Germany Greece Guatemala India Indonesia Ireland Israel Italy Japan JORDAN Kazakhstan Kenya Kurdistan, Iraq Kuwait Malaysia Malta Mexico Moldova Morocco Nepal Netherlands New Zealand Oman Pakistan Palestine Paraguay Paraquay Peru Philippines Poland Puerto Rico Qatar Republic of Panama Romania Russia Rwanda Saudi Arabia Serbia Slovenia South Africa South Korea Spain Sri Lanka Sudan Sweden Switzerland Syria Thailand Trinidad Trinidad,W.I. Tunisia UK Ukraine United Arab Emerites Uruguay Vietnam
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = JGS
-
-