| | Class 2 Device Recall RibLoc U Plus 90 Instrument |  |
| Date Initiated by Firm | November 27, 2023 |
| Date Posted | December 26, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0618-2024 |
| Recall Event ID |
93470 |
| Product Classification |
Orthopedic manual surgical instrument - Product Code LXH
|
| Product | RibLoc U Plus 90 Instrument PRIMARY GUIDE ASSY, LOW PROFILE REF RBL2320
The RBL2320, Low Profile Primary Guide is an instrument used with a power system to
compress the RibLoc plates onto the bone, determine the length of the screw to install, locate the position of the holes to drill, and provide proper orientation for the installation of the screws. |
| Code Information |
Part Number: RBL2320
UDI Code / Batch-Lot Number
10806378118210-538351000000 / 538351
10806378118210-538352000000 / 538352
10806378118210-544036000000 / 544036
10806378118210-554505000000 / 554505
10806378118210-L1810012181031 / L1810012
10806378118210-L1902009190207 / L1902009
10806378118210-L1909017190918 / L1909017
10806378118210-L2001005200130 / L2001005
10806378118210-L2005002200504 / L2005002
10806378118210-L2010046000000 / L2010046
10806378118210-L2010048000000 / L2010048 |
| FEI Number |
1000125930
|
Recalling Firm/ Manufacturer |
Acumed LLC 5885 NW Cornelius Pass Rd Hillsboro OR 97124-9370
|
| For Additional Information Contact | Ellie Wood 214-738-1709 Ext. 293 |
Manufacturer Reason for Recall | Due to potential breakage during use. |
FDA Determined Cause 2 | Unknown/Undetermined by firm |
| Action | On 11/27/2023, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter via FedEx and UPS ground to customers informing them that Acumed is initiating a Recall of all field inventory of Part Number RBL2320, Low Profile Primary Guides due to the potential for breakage during use.
Customers are instructed to:
IMMEDIATELY stop using and/or distributing the product and complete the following actions:
1-The RBL2320, Low Profile Primary Guide is often located with the RBL4030 Instrumentation Tray (as shown in Figure 2 above). Locate the RBL4030 tray and remove the RBL2320, Low Profile Primary guide(s) from the tray. Perform a physical count of their inventory of the RBL2320, Low Profile Primary Guides, and identify any inventory of the batch/lot numbers listed in the table above.
2-Remove the RBL2320, Low Profile Primary Guides removed/segregated in step 1, such that they are taken out of service (i.e., quarantined).
3-Record this data (product, batch/lot number, and quantity) on the Outstanding Field Action Survey found at the end of this recall letter (on page 5). If you do not have product to return, you are still required to complete the survey.
4) Return the completed survey via email to ellie.wood@acumed.net.
5) If you have RBL2320, Low Profile Primary Guide(s) to return, please contact Ellie Wood via email at ellie.wood@acumed.net so that a Return Authorization Number and return instructions can be provided. *Please note, only the RBL2320, Low Profile Primary Guide needs to be returned.
6) If RBL2320, Low Profile Primary Guide product is not returning, please post this Recall Notification Letter on or near the affected products.
|
| Quantity in Commerce | 513 units |
| Distribution | Worldwide - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, NJ, OH, OR, SC, TX, UT, VA, WA, and WI. The country of Australia. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|