| Class 1 Device Recall smiths medical Medfusion | |
Date Initiated by Firm | December 19, 2023 |
Date Posted | February 01, 2024 |
Recall Status1 |
Open3, Classified |
Recall Number | Z-0864-2024 |
Recall Event ID |
93757 |
510(K)Number | K111386 |
Product Classification |
Pump, infusion - Product Code FRN
|
Product | smiths medical Medfusion Model 4000 syringe pump |
Code Information |
All Medfusion Model 4000 with software versions: v1.0.0, v1.1.0, v1.1.1, v1.1.2, v1.5.0, v1.5.1, 1.6.0, v1.6.1, v1.6.4, v2.3, v2.4, v2.5, and all versions prior to v1.6.5. |
Recalling Firm/ Manufacturer |
Smiths Medical ASD Inc. 6000 Nathan Ln N Minneapolis MN 55442-1690
|
For Additional Information Contact | icumed.custhelp.com/app/market-action 763-383-3072 |
Manufacturer Reason for Recall | Medfusion Model 4000 syringe infusion pumps, with the following software versions, may have the following issues: 1. Delivery During Motor Not Running High Priority Alarm (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 2. Re-administered Loading Dose (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 3. Incorrect Critical Data Failure Alarm, 4. Interruption of Bolus or Loading Dose Delivery (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 5. Incorrect Total Bolus/Loading Dose Displayed (v1.1.0, v1.1.1, v1.1.2), 6. Volume Limit Before Bolus/Loading Dose Complete (v1.1.2), 7. Drug Library Lower Limit Displayed Incorrectly (v1.6.0, v1.6.1), 8. Depleted Battery Alarm (All versions previous to v1.6.5), 9. Loss of Wireless Connectivity (v1.5.0, v1.5.1, v1.6.0, v1.6.1, v1.6.4), 10. PharmGuard Server Password (v2.3, v2.4, v2.5).
Smiths Medical corrected all issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases.
Please ensure you have the most recent Medfusion software installed on your pumps.
|
FDA Determined Cause 2 | Software Design Change |
Action | Smiths Medical issued an URGENT MEDICAL DEVICE CORRRECTION notice to its consignees on 12/19/2023 by letter delivered via traceable means. This notification details the issues and the affected software versions. Please note that the pump displays the installed software version on the startup screen after the pump is powered on. The notice explained the issues with the pumps, the potential risks involved with their use, and requested the following actions be taken:
1. Locate all affected pumps in your possession and ensure all users or potential users
of these devices are immediately made aware of this notification and proposed
mitigations. Please ensure you have the most recent Medfusion software installed on
your pumps.
2. Complete and return the attached Response Form to smithsmedical8045@sedgwick.com within ten days of receipt to acknowledge your understanding of this notification.
3. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to smithsmedical8045@sedgwick.com.
To report adverse events or product complaints: Global Complaint Management, globalcomplaints@icumed.com, 1-(866)-216-8806
Additional information or technical assistance: Technical Support, icumed.custhelp.com/app/market-action
Questions about this Field Correction Notice: Field Corrections, icumed.custhelp.com/app/market-action |
Quantity in Commerce | 60146 pumps |
Distribution | US nationwide, Canada, Great Britain, Bermuda, Australia, Germany, Singapore, Saudi Arabia, Lebanon, and Malaysia
|
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|