• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall smiths medical Medfusion

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall smiths medical Medfusionsee related information
Date Initiated by FirmDecember 19, 2023
Date PostedFebruary 01, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0864-2024
Recall Event ID 93757
510(K)NumberK111386 
Product Classification Pump, infusion - Product Code FRN
Productsmiths medical Medfusion Model 4000 syringe pump
Code Information All Medfusion Model 4000 with software versions: v1.0.0, v1.1.0, v1.1.1, v1.1.2, v1.5.0, v1.5.1, 1.6.0, v1.6.1, v1.6.4, v2.3, v2.4, v2.5, and all versions prior to v1.6.5.
FEI Number 3012307300
Recalling Firm/
Manufacturer
Smiths Medical ASD Inc.
6000 Nathan Ln N
Minneapolis MN 55442-1690
For Additional Information Contacticumed.custhelp.com/app/market-action
763-383-3072
Manufacturer Reason
for Recall
Medfusion Model 4000 syringe infusion pumps, with the following software versions, may have the following issues: 1. Delivery During Motor Not Running High Priority Alarm (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 2. Re-administered Loading Dose (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 3. Incorrect Critical Data Failure Alarm, 4. Interruption of Bolus or Loading Dose Delivery (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 5. Incorrect Total Bolus/Loading Dose Displayed (v1.1.0, v1.1.1, v1.1.2), 6. Volume Limit Before Bolus/Loading Dose Complete (v1.1.2), 7. Drug Library Lower Limit Displayed Incorrectly (v1.6.0, v1.6.1), 8. Depleted Battery Alarm (All versions previous to v1.6.5), 9. Loss of Wireless Connectivity (v1.5.0, v1.5.1, v1.6.0, v1.6.1, v1.6.4), 10. PharmGuard Server Password (v2.3, v2.4, v2.5). Smiths Medical corrected all issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases. Please ensure you have the most recent Medfusion software installed on your pumps.
FDA Determined
Cause 2
Software Design Change
ActionSmiths Medical issued an URGENT MEDICAL DEVICE CORRRECTION notice to its consignees on 12/19/2023 by letter delivered via traceable means. This notification details the issues and the affected software versions. Please note that the pump displays the installed software version on the startup screen after the pump is powered on. The notice explained the issues with the pumps, the potential risks involved with their use, and requested the following actions be taken: 1. Locate all affected pumps in your possession and ensure all users or potential users of these devices are immediately made aware of this notification and proposed mitigations. Please ensure you have the most recent Medfusion software installed on your pumps. 2. Complete and return the attached Response Form to smithsmedical8045@sedgwick.com within ten days of receipt to acknowledge your understanding of this notification. 3. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to smithsmedical8045@sedgwick.com. To report adverse events or product complaints: Global Complaint Management, globalcomplaints@icumed.com, 1-(866)-216-8806 Additional information or technical assistance: Technical Support, icumed.custhelp.com/app/market-action Questions about this Field Correction Notice: Field Corrections, icumed.custhelp.com/app/market-action
Quantity in Commerce60146 pumps
DistributionUS nationwide, Canada, Great Britain, Bermuda, Australia, Germany, Singapore, Saudi Arabia, Lebanon, and Malaysia
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = FRN
-
-