• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Centurion

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 1 Device Recall Centurion see related information
Date Initiated by Firm November 15, 2023
Date Posted February 25, 2024
Recall Status1 Open3, Classified
Recall Number Z-1132-2024
Recall Event ID 93800
Product Classification Tray, surgical - Product Code LRP
Product Centurion Kits, trays, and packs labeled as follows:
a) BASIC PROCEDURE TRAY, REF MNS13600;
b) BWH DAILY DRIVELINE MGMT KIT, REF DM640A;
c) CIRCUMCISION TRAY, REF CIT6180;
d) INCISION & DRAINAGE TRAY, REF ID1690;
e) LACERATION TRAY, REF SUT19510;
f) LACERATION TRAY, REF SUT19510;
g) LACERATION TRAY, REF SUT8565;
h) LACERATION TRAY, REF SUT8565;
i) LINE REPAIR KIT, REF MNS12045A;
j) MMC DAILY DRIVELINE MGMT KIT, REF DM850A;
k) NEWBORN THORACOTOMY TRAY, REF MNS3470;
l) NICU STERILE CENTRAL LINE CHANGE KIT, REF DT18520A;
m) OMMAYA KITS - LATEX FREE, REF HT2375;
n) PEDIATRIC ULTRASOUND KIT, REF MNS12995A;
o) PORT ACCESS INFUSION KIT - 1IN NEEDLE, REF DYNDC2817C;
p) WOUND CLOSURE AND IRRIGATION TRAY, REF SUT14785;
Code Information a) MNS13600, 00653160349887 (case), 10653160349884 (unit), Lot Numbers: 2022022350, 2022041880; b) DM640A, 00653160357271 (case), 10653160357278 (unit), Lot Numbers: 2022122090, 2022122890, 2023073190; c) CIT6180, 00653160290691 (case), 10653160290698 (unit), Lot Numbers: 2023080290; d) ID1690, 10653160235903 (case), 00653160235906 (unit), Lot Numbers: 2022071390, 2022120790, 2023062690; e) SUT19510, 00653160352436 (case), 10653160352433 (unit), Lot Numbers: 2022041880, 2022071180, 2022071280; f) SUT19510, 00653160356199 (case), 10653160356196 (unit), Lot Numbers: 2022120290, 2023033090, 2023052290, 2023080190; g) SUT8565, 50653160016845 (case), 00653160016840 (unit), Lot Numbers: 2022042580; h) SUT8565, 00653160356182 (case), 10653160356189 (unit), Lot Numbers: 2023020690, 2023030390, 2023050390, 2023091890; i) MNS12045A, 00653160350340 (case), 10653160350347 (unit), Lot Numbers: 2022032850, 2022041180, 2022041280; j) DM850A, 00653160354096 (case), 10653160354093 (unit), Lot Numbers: 2022072890, 2022090690, 2022112290, 2023030390, 2023070390; k) MNS3470, 10653160106029 (case), 00653160106022 (unit), Lot Numbers: 2022031790, 2022091390, 2022110890, 2023011690, 2023052290, 2023080790; l) DT18520A, 00653160351491 (case), 10653160351498 (unit), Lot Numbers: 2022082280, 2022082980; m) HT2375, 50653160050221 (case), 00653160050226 (unit), Lot Numbers: 2023061290, 023092190; n) MNS12995A, 00653160349818 (case), 10653160349815 (unit), Lot Numbers: 2022031750, 2022031850, 2022082980; o) DYNDC2817C, 00653160348361 (case), 10653160348368 (unit), Lot Numbers: 2022012480, 2022050980; p) SUT14785, 10653160193081 (case), 00653160193084 (unit), Lot Numbers: 2022060290, 2022110490, 2023040390, 2023052490, 2023061390, 2023091490
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information Contact Karin Johnson
886-359-1704
Manufacturer Reason
for Recall
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
FDA Determined
Cause 2
Nonconforming Material/Component
Action MEDLINE issued an IMMEDIATE ACTION REQUIRED NURSE ASSIST PRODUCT RECALL notice to its consignees on 11/16/2023 via USPS first class mail and email. The notice explained the problem, requested log on to their portal to identify the products and check their inventory so the product could be quarantined. Once Medline receives the completed response form, over-labels will be shipped to the consignee. Medline Industries, LP instructed customers to over-label the convenience kits with a warning label indicating to remove the sodium chloride irrigation or Sterile water and replace from a different sterile supply. Distributors or consignees who further distributed the affected products were directed to notify their customers. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity in Commerce 14234 units
Distribution Distribution US nationwide including Puerto Rico, Canada, UAE, and Panama
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-