| | Class 2 Device Recall Exactech Equinoxe |  |
| Date Initiated by Firm | March 06, 2024 |
| Date Posted | March 29, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1404-2024 |
| Recall Event ID |
94092 |
| 510(K)Number | K042021 |
| Product Classification |
Prosthesis, shoulder, semi-constrained, metal/polymer cemented - Product Code KWS
|
| Product | Exactech Equinoxe GLENOID,KEELED, CEMENTED, mates with 47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-02-02, b) Medium, Item Number 314-02-03, c) Large, Item Number 314-02-04, Shoulder Arthroplasty |
| Code Information |
a) Item Number 314-02-02, UDI/DI 10885862084477, Serial Numbers: 0747871, 0747855, 0747853, 0747858, 0747827, 0747805, 0747814, 0747845, 0747781, 0747779, 0747823, 0747835, 0747804, 0747813, 0747839, 0747775, 0747838, 0747798, 0747846, 0747825, 0747841, 0747832, 0747837, 0747776, 0768035, 0747816, 0747833, 0747872, 0747844, 0747847, 0747803, 0747809, 0747818, 0747860, 0747808, 0747824, 0747859, 0747801, 0747840, 0747829, 0747831, 0747799, 0747794, 0747782, 0768047, 0768042, 0768033, 0768068, 0768058, 0747784, 0747857, 0747778, 0747797, 0747802, 0747810, 0747777, 0747865, 0768012, 0768064, 0768041, 0768086, 0747787, 0747830, 0768031, 0768063, 0768097, 0747828, 0768092, 0768006, 0768004, 0747863, 0767999, 0768076, 0768080, 0747826, 0747850, 0768087, 0747783, 0768010, 0768034, 0768038, 0747870, 0768060, 0768043, 0768089, 0768077, 0747812, 0768093, 0768073, 0768025, 0747852, 0747780, 0747843, 0747791, 0747854, 0747795, 0768065, 0747864, 0768020, 0768101, 0768011, 0747820, 0768051, 0768098, 0768015, 0768008, 0768062, 0768071, 0768028, 0747817, 0747867, 0747869, 0747806, 0768100, 0768102, 0768040, 0768036, 0768085, 1088787, 1088875, 1088794, 1088864, 1088783, 1088847, 0768096, 0768052, 0768079, 0768027, 0768082, 0747821, 0768061, 0747822, 0768078, 0747868, 0768026, 0768054, 0768057, 0747834, 0768084, 0747788, 0747807, 0768067, 0768017, 0768095, 0768029, 0768049, 0768032, 0768069, 0768018, 0768107, 0768037, 0768013, 0768021, 0768099, 0768002, 0768083, 0747790, 0768056, 0768016, 0768001, 0768044, 0768050, 0768055, 1088834, 1088821, 1088876, 1088807, 1088818, 1088799, 1088868, 1088879, 1088858, 1088877, 1088811, 1088777, 1088785, 1088857, 1088854, 1088830, 1088833, 1088862, 1088869, 1088816, 1088824, 1088866, 1088805, 1088846, 1088872, 1088837, 1088801, 1088781, 1088802, 1088836, 1088796, 1088870, 1088827, 1088845, 1088878, 1088848, 1088850, 1088778, 1088812, 1088803, 1088792, 1088826, 1088806, 1088793, 1088822, 1088786, 1088873, 1088817, 1088843, 1088800, 1088819, 1088829, 1088808, 1088828, 1088838, 1088809, 1088852, 1088825, 1088831, 1088865, 1088776, 1259732, 1343942, 1383864, 1259646, 1259633, 1259718, 1361257, 1259670, 1259729, 1383861, 1259716, 1361293, 1259733, 1259667, 1259695, 1259647, 1259668, 1259710, 1259725, 1259704, 1259690, 1343956, 1361267, 1361288, 1361294, 1361295, 1259681, 1259683, 1259721, 1259713, 1361274, 1259630, 1259638, 1259641, 1259649, 1259656, 1259657, 1259658, 1259661, 1259666, 1259669, 1259671, 1259673, 1259680, 1259692, 1259698, 1259699, 1259709, 1259728, 0768106, 0747851, 0768007, 0768003, 0768088, 0768023, 0768046, 0768009, 0768091, 0747789, 0768005, 0747815, 0768075, 0747792, 0768039, 0768094, 0747862, 1088789, 1088784, 1088860, 1088867, 1088855, 1088815, 1088841, 1088810, 1088782, 1088851, 1088779, 1088791, 1259730, 1343952, 1343967, 1497063, 1259664, 1259639, 1394913, 1361272, 1259634, 1361271, 1553619, 1553635, 1545990, 1394926, 1497047, 1361279, 1394930, 1424757, 1259689, 1433230, 1424762, 1553637, 1259675, 1361276, 1259662, 1361270, 1343968, 1343963, 1497059, 1545989, 1343960, 1343966, 1361287, 1343946, 1259707, 1424740, 1259731, 1361269, 1361282, 1424742, 1259717, 1361292, 1361259, 1343945, 1361261, 1394938, 1433232, 1343951, 1424759, 1343941, 1424756, 1497075, 1424744, 1259663, 1361260, 1343965, 1259702, 1394936, 1259650, 1259644, 1497068, 1259645, 1545991, 1259714, 1259676, 1259632, 1343957, 1361286, 1394927, 1259674, 1343964, 1361277, 1394904, 1488038, 1424739, 1259653, 1488042, 1497070, 1394924, 1394907, 1424743, 1488049, 1488035, 1424735, 1394920, 1394915, 1259696, 1394908, 1361263, 1383863, 1488040, 1343961, 1394909, 1361273, 1424748, 1394921, 1259726, 1343959, 1433231, 1361266, 1259701, 1497053, 1497051, 1497071, 1553640, 1361284, 1343953, 1394934, 1488046, 1497061, 1424749, 1394937, 1343955, 1553653, 1488041, 1343948, 1394929, 1424734, 1259686, 1424754, 1343958, 1259715, 1259708, 1259648, 1259719, 1424751, 1424750, 1424745, 1424741, 1553651, 0747819, 0747811, 0768074, 0768024, 0768090, 0768104, 0768105, 1088844, 1088840, 1088814, 1088813, 1088823, 1088820, 1088798, 1088859, 1088790, 1259693, 1497050, 1553649, 1553622, 1553638, 1497044, 1394918, 1424737, 1497067, 1488048, 1488039, 1259703, 1553625, 1553646, 1553623, 1553650, 1394935, 1497066, 1497069, 1361296, 1394922, 1497057, 1455358, 1259665, 1455355, 1455359, 1394917, 1394928, 1259679, 1343947, 1259705, 1424747, 1394916, 1424761, 1553657, 1553641, 1394911, 1488037, 1361278, 1553647, 1488036, 1259685, 1383860, 1383857, 1259687, 1394910, 1361290, 1553627, 1394925, 1259700, 1497074, 1259631, 1545993, 1259711, 1383859, 1497055, 1553644, 1553624, 1343962, 1455356, 1497054, 1497046, 1497060, 1343944, 1488047, 1553631, 1545994, 1383855, 1553655, 1361258, 1497062, 1497045, 1553648, 1394914, 1424753, 1553620, 1259677, 1424758, 1433229, 1259655, 1553642, 1394899, 1488034, 1259684, 1394932, 1553652, 1736406, 1628629, 1736418, 1630279, 1627701, 1694080, 1630281, 1736425, 1630274, 1736441, 1774821, 1630271, 1736395, 1694065, 1736404, 1628617, 1774830, 1736416, 1802952, 1802944, 1736431, 1736426, 1802927, 1827093, 1630275, 1827097, 1628622, 1888476, 1774840, 1694061, 1630288, 1736398, 1774839, 1630270, 1627707, 1736436, 1630285, 1627703, 1630280, 1827103, 1736414, 1827081, 1694079, 1630287, 1694071, 1827083, 1888478, 1802931, 1630262, 1627709, 1694054, 1736397, 1630264, 1827098, 1802961, 1774817, 1736440, 1694067, 1827106, 1628627, 1802937, 1802962, 1802939, 1774827, 1694056, 1774832, 1630276, 1802929, 1827105, 1802953, 1627704, 1802946, 1827104, 1694059, 1737447, 1630284, 1627708, 1888484, 1736429, 1628619, 1774833, 1694076, 1694077, 1694063, 1694069, 1694070, 1694082, 1630282, 1694074, 1736409, 1736412, 1774810, 1774825, 1774845, 1774837, 1736432, 1802934, 1802947, 1802924, 1802925, 1802954, 1827079, 1827080, 1827085, 1802969, 1802970, 1827082, 1827107, 1827109, 1802932, 1694075, 1736442, 1628624, 1628625, 1630283, 1627705, 1628615, 1736433, 1774836, 1627700, 1694066, 1630266, 1694060, 1628626, 1694073, 1736427, 1628623, 1827084, 1088797, 1088832, 1088804, 1088795, 1088874, 1088853, 1088839, 1361262, 1361275, 1361289, 1394912, 1259640, 1361268, 1424763, 1553621, 1361264, 1394900, 1259635, 1343943, 1424755, 1497073, 1394933, 1553632, 1424736, 1455357, 1488044, 1553633, 1259637, 1497065, 1553639, 1394919, 1343969, 1259636, 1488045, 1394905, 1361280, 1259722, 1553645, 1497048, 1488043, 1553634, 1343970, 1553630, 1433233, 1259659, 1259678, 1343950, 1553656, 1424760, 1497064, 1259672, 1982274, 1827094, 1774812, 1956100, 2042756, 2042742, 1774834, 1774814, 1802950, 2042728, 1938009, 1802933, 1974999, 1888491, 1736402, 1736428, 1956096, 1937997, 1888480, 1982281, 1974989, 2042737, 1827087, 1888490, 2042740, 2042757, 1974991, 2042727, 2042750, 1956105, 1888492, 1827086, 1736422, 1774823, 1938015, 1802940, 1774809, 1736419, 1630289, 1888485, 1802938, 1827088, 1827115, 1736399, 1774831, 1736443, 1802964, 1802951, 1774828, 1694072, 2071728, 1630290, 1956089, 2042766, 1956093, 1956088, 1774815, 1827099, 1774818, 1888495, 1736438, 1630265, 1802936, 2001596, 1827114, 1737442, 1938018, 2042730, 1938000, 1827116, 1982276, 1827100, 2001588, 1802959, 1774813, 1774829, 1982280, 1694083, 1802942, 2042733, 1774808, 2071731, 1974997, 1974981, 1888487, 1736417, 1956087, 2001595, 1938004, 1974998, 2071723, 1628618, 2042761, 1774843, 1956098, 1736424, 1736420, 1630272, 1694064, 1827119, 1802943, 1956090, 2071729, 1802930, 1736437, 1736403, 1774822, 1827117, 1628621, 1888488, 1937996, 2042754, 1694057, 2042767, 1736413, 1627706, 1630277, 1982284, 1736400, 1630269, 1974986, 1802958, 1736396, 1888489, 1802949, 1937994, 1982289, 1774819, 1888493, 1802966, 1802963, 1630263, 2042735, 1627702, 1974990, 2042745, 2042758, 1736407, 1736415, 1956086, 1956094, 1974988, 1982287, 1982286, 2071722, 2071724, 2071732, 1802941, 1694062, 1802955, 2071726, 2042731, 1827101, 1736401, 1736410, 1827096, 1802935, 1774824, 1937991, 1938017, 1802968, 2042749, 1938011, 1736421, 1982296, 2001594, 1736394, 1888494, 1827108, 2042738, 2042759, 1982293, 1956099, 1938002, 1888477, 2001593, 2137573, 2137567, 2108211, 2159768, 2108195, 2145602, 2145578, 2159758, 2137569, 2108214, 2089644, 2145590, 2145600, 2137578, 2137584, 2145580, 2089633, 2089634, 2183125, 2183113, 2089629, 2089631, 2159767, 2089640, 2145601, 2089632, 2089627, 2108201, 2137586, 2137582, 2089643, 2089645, 2108193, 2108198, 2108199, 2108205, 2108206, 2108210, 2108215, 2183109, 2089639, 2108204, 1088861, 1088849, 1088788, 1088856, 1361291, 1259712, 1361283, 1394906, 1553629, 1553643, 1394902, 1394931, 1259688, 1259642, 1259660, 1394901, 1424738, 1553658, 1259694, 1982285, 1956102, 1938010, 1827110, 1630286, 1827118, 1736434, 2001590, 2042762, 1974992, 2042760, 1982294, 1628628, 1694068, 1938013, 1630261, 1937986, 1774820, 2071725, 1827102, 1974982, 1982277, 1938003, 1888483, 1956101, 1737448, 1938001, 1937990, 1974984, 1974993, 2042753, 2042768, 2071735, 2001589, 1737441, 1827112, 1974996, 1630267, 1974994, 1802923, 2042734, 1974983, 1827090, 1938012, 1774811, 1774826, 1736430, 1982283, 2071736, 1937998, 1774844, 2042763, 1774816, 1827113, 1888479, 1938008, 1937987, 2042739, 2071727, 2071738, 1956104, 2071734, 1956091, 1802965, 2071739, 2042736, 1938014, 1802972, 2042752, 2042747, 2042729, 1982282, 1802971, 2001587, 2159748, 2230872, 2108213, 2251846, 2277690, 2393012, 2318952, 2183108, 2288233, 2230887, 2145606, 2145582, 2251869, 2288216, 2222091, 2159747, 2183120, 2137585, 2277678, 2358646, 2393016, 2421128, 2159749, 2339793, 2506125, 2137576, 2183104, 2318982, 2137587, 2145594, 2230876, 2288235, 2393027, 2506150, 2137583, 2277679, 2159756, 2277692, 2393005, 2288215, 2339791, 2159751, 2251855, 2421110, 2421114, 2392995, 2393007, 2451317, 2421137, 2277684, 2230883, 2251854, 2318971, 2089637, 2318950, 2451339, 2089630, 2159764, 2277676, 2339800, 2421098, 2393006, 2393029, 2159760, 2277675, 2318943, 2230863, 2108216, 2108203, 2421135, 2277688, 2421106, 2230896, 2230864, 2251847, 2358668, 2288226, 2145586, 2288230, 2183110, 2222094, 2145599, 2222096, 2230862, 2137574, 2421133, 2251849, 2108208, 2222089, 2183122, 2251862, 2159763, 2230903, 2159753, 2137570, 2145583, 2230877, 2318953, 2392990, 2392994, 2183114, 2230873, 2288214, 2451336, 2421094, 2421129, 2392985, 2145593, 2145609, 2159762, 2145610, 2108202, 2393028, 2230890, 2230875, 2251864, 2145577, 2288223, 2393011, 2288231, 2230880, 2251865, 2288236, 2339802, 2108197, 2277689, 2318945, 2230869, 2159761, 2183106, 2230897, 2318949, 2339803, 2318975, 2251845, 2339799, 2137568, 2145598, 2230878, 2222093, 2145595, 2251866, 2230870, 2145584, 2159746, 2318965, 2339804, 2358661, 2230885, 2145603, 2318967, 2288220, 2393013, 2421100, 2230874, 2159766, 2145608, 2318948, 2230861, 2159744, 2251857, 2108196, 2145591, 2339797, 2358669, 2339796, 2358656, 2230879, 2230902, 2251848, 2277683, 2277685, 2318954, 2318958, 2288225, 2318947, 2318970, 2318984, 2277682, 2421104, 2421108, 2421112, 2421097, 2421134, 2358652, 2421095, 2421096, 2421116, 2222092, 2318969, 2392996, 2393024, 2421115, 2451328, 2451334, 2451337, 2451368, 2451369, 2451381, 2451316, 2451330, 2451323, 2451326, 2522768, 2522771, 2522772, 2522775, 2522785, 2522786, 2288227, 2089638, 2108200, 2089642, 2183111, 2089641, 2358650, 2421105, 2159750, 2230894, 2318973, 2506123, 2230898, 2230906, 2318980, 2358658, 2318960, 2451376, 2506149, 2137579, 2222095, 2159757, 2318983, 2451324, 2451371, 2451327, 2421132, 2451360, 2522749, 2145579, 2183117, 2137572, 2183107, 2183116, 2318966, 2230899, 2230908, 2108207, 2145581, 2145585, 2137566, 2230893, 2230886, 2288218, 2318962, 2318957, 2318963, 2451329, 2288224, 2421099, 1424752, 1361281, 1259643, 1497072, 1553654, 1424746, 1497049, 2042755, 2042732, 2042748, 1827095, 1774842, 1827089, 1982288, 1937989, 1802945, 1937995, 1982290, 1802928, 1630268, 1774835, 1802956, 1938007, 1827091, 1694058, 2042744, 2042751, 2001592, 1982279, 1982273, 1628620, 2042741, 2318968, 2251861, 2522755, 2318951, 2392998, 2421123, 2358670, 2451351, 2230895, 2542796, 2506127.2358651, 2522757, 2451354, 2421136, 2421127, 2451365, 2451340, 2522745, 2542786, 2451341, 2318981, 2451366, 2251853, 2251852, 2421130, 2451379, 2358664, 2421113, 2506135, 2230900, 2542809, 2506121, 2542805, 2393026, 2230865, 2506131, 2251844, 2542792, 2506136, 2230882, 2318976, 2392992, 2230871, 2318977, 2522740, 2089646, 2451344, 2542798, 2339789, 2522793, 2542793, 2393004, 2393014, 2145597, 2522787, 2392999, 2145588, 2393019, 2251860, 2392989, 2230867, 2522769, 2393018, 2542799, 2451312, 2522750, 2393025, 2506145, 2251863, 2451325, 2358648, 2451373, 2339798, 2183119, 2421117, 2451370, 2506115, 2522794, 2392988, 2393020, 2137571, 2392993, 2451342, 2451338, 2137575, 2222098, 2522739, 2451364, 2542804, 2089628, 2137580, 2542800, 2288232, 2358655, 2506129, 2542802, 2451346, 2145604, 2318979, 2251850, 2421118, 2506139, 2542808, 2522738, 2451358, 2542787, 2230881, 2089635, 2159754, 2522758, 2542788, 2522746, 2522778, 2288229, 2451313, 2522744, 2108209, 2318956, 2506122, 2506134, 2230868, 2542807, 2542803, 2522752, 2288221, 2358666, 2506148, 2522763, 2522776, 2230905, 2522783, 2522773, 2393000, 2522780, 2451353, 2393003, 2358659, 2522784, 2522762, 2145607, 2506147, 2522765, 2451372, 2522747, 2648842, 2736252, 2629772, 2629726, 2629752, 2767191, 2552753, 2648816, 2552749, 2648861, 2825657, 2552765, 2638941, 2629731, 2736260, 2629725, 2552747, 2767177, 2790552, 2629753, 2648849, 2629765, 2648818, 2648858, 2629762, 2790562, 2790548, 2552745, 2736251, 2648810, 2790557, 2648839, 2648841, 2825648, 2844206, 2648868, 2790556, 2552763, 2552770, 2854324, 2629740, 2629754, 2629757, 2629760, 2629761, 2629763, 2629766, 2552751, 2552756, 2629728, 2648813, 2648814, 2648827, 2648834, 2648851, 2648863, 2648864, 2648867, 2638925, 2638937, 2629737, 2629745, 2629751, 2629764, 2629770, 2638933, 2638935, 2638939, 2648857, 2736245, 2736250, 2736253, 2736262, 2736268, 2736269, 2736279, 2736273, 2767174, 2767176, 2767183, 2767184, 2767198, 2767179, 2767175, 2767178, 2767180, 2767200, 2790539, 2767189, 2767192, 2767199, 2790535, 2790538, 2790543, 2825632, 2767193, 2854313, 2854321, 2825635, 2825637, 2825649, 2825651, 2825655, 2825656, 2854322, 2854323, 2854325, 2854328, 2854331, 2854333, 2854318, 2854320, 2854334, 2854330, 2854339, 2854312, 2854314, 2854315, 2854317, 2854319, 2854326, 2854327, 2854329, 2854332, 2854337, 2854340, 2854341, 2736261, 2825636, 2886741, 2629738, 2638942, 2648825, 2767188, 2767190, 2767185, 2767187, 2629729, 2629736, 2790545, 2790563, 2629759, 2736258, 2648824, 2736259, 2552762, 2629767, 2844202, 2844183, 2767196, 2767197, 2629747, 2736267, 2844179, 2736276, 2790561, 2790551, 2629733, 1774841, 1938016, 1802967, 1736423, 2042769, 2042743, 1982278, 1737443, 1982275, 2042746, 1802926, 1736411, 2277680, 2277673, 2358649, 2506133, 2542789, 2522790, 2451332, 2522743, 2358662, 2451350, 2145605, 2145589, 2288213, 2522741, 2451359, 2542806, 2542811, 2108194, 2288222, 2451322, 2506140, 2542785, 2451314, 2451319, 2522748, 2318978, 2421125, 2506116, 2506119, 2542797, 2506142, 2451352, 2451321, 2288228, 2451380, 2451320, 2421121, 2421120, 2183121, 2522760, 2522788, 2251858, 2421119, 2522764, 2506144, 2542813, 2393002, 2288234, 2522791, 2339792, 2137581, 2522737, 2522782, 2451382, 2108212, 2451333, 2451375, 2393022, 2183123, 2451357, 2137565, 2522767, 2648832, 2825631, 2934346, 2934343, 2910091, 2736266, 2934347, 2629741, 2960080, 2844201, 2736263, 2736257, 2960072, 2886726, 2736277, 2844199, 2844192, 2790549, 2736254, 2648848, 2825644, 2648854, 2886735, 2767194, 3673069, 2844197, 2790555, 3552979, 2960071, 2648847, 2648845, 2629735, 2844180, 2648860, 2648859, 2648835, 2648840, 2638936, 2844205, 2825633, 3591803, 3552937, 3552987, 3552959, 3640580, 2552752, 2638929, 3673067, 2736270, 3673072, 2934328, 3591791, 2886742, 2910108, 2910092, 2910094, 2910095, 2910107, 2934330, 2934332, 2648837, 2934319, 2934321, 2934329, 2934333, 2934334, 2934335, 2934337, 2934340, 2934344, 2854316, 2854335, 2854336, 2854338, 2886736, 2886740, 2910089, 2910093, 2910100, 2910103, 2910106, 2910109, 2960082, 2960084, 2960073, 2960081, 2648828, 2886723, 2886730, 2960067, 2960076, 2960079, 2960069, 2960088, 2960068, 2960074, 3552932, 3552933, 3552939, 3552950, 3552952, 3552963, 3552976, 2960070, 3575515, 3575516, 3552940, 3552961, 3575522, 3575524, 3673040, 3673043, 3575542, 3673037, 2629771, 2736275, 2825653, 2825658, 2825652, 2910101, 2934320, 3575514, 3706883, 2552766, 2910099, 2910102, 3673032, 3673047, 3673034, 3673045, 3706839, 3706885, 3706871, 3706878, 2790536, 2790553, 2648819, 2844204, 2934327, 3591785, 3575538, 3591799, 2934339, 3552966, 2648831, 2767181, 2648812, 3575533, 2886738, 2844203, 2960085, 2886725, 3591801, 3706837, 3706860, 2910088, 3673029, 2790547, 2886719, 2629723, 3673042, 2886716, 3575523, 2886743, 2886728, 3591823, 3706849, 2825639, 2844184, 3575527, 2960086, 3706863, 2886729, 2910098, 2910090, 3575528, 2629742, 3706850, 2910105, 2629769, 2886734, 3591786, 3575525, 3575553, 2960089, 3575535, 3591827, 3673039, 2767173, 2886739, 2886722, 3673063, 2886717, 2910097, 2648866, 1974985, 2071730, 1982272, 1956092, 1937999, 2071721, 2071737, 2318972, 2542784, 2522742, 2230907, 2451355, 2451310, 2451343, 2506117, 2251870, 2145592, 2358657, 2451347, 2522777, 2421109, 2251859, 2159755, 2451315, 2358647, 2451318, 2451349, 2506120, 2451335, 2277691, 2358667, 2318964, 2392997, 2222097, 2629727, 2886737, 2844200, 2934336, 2629732, 3575539, 3552954, 3591790, 2844207, 3552936, 3591798, 3673057, 2844196, 3591813, 2790541, 2638932, 2648865, 3673062, 2736255, 2552746, 3591814, 3706840, 2844194, 2910111, 2552771, 3575529, 2736246, 3673055, 3591828, 2886718, 3591804, 3591817, 2934341, 3591782, 3575546, 2960075, 2638928, 3575552, 2844188, 3552930, 3575554, 3575545, 3591806, 3552962, 3591816, 2790550, 3552944, 3640585, 2790544, 3552985, 3591818, 3673024, 3673031, 3673033, 3673035, 3673044, 3552938, 3552942, 3552953, 3552968, 3673060, 3673070, 3552975, 3575551, 2934338, 2736274, 3706886, 3673038, 3706873, 3706865, 3706884, 3798934, 3798947, 3798918, 3798931, 3798945, 3798929, 3798915, 3673059, 3798906, 3798907, 3798908, 3798909, 3798913, 3798917, 3798922, 3798924, 3798926, 3798927, 3798928, 3798930, 3798935, 3798936, 3798937, 3798941, 3798948, 3798919, 3706879, 3706874, 3706866, 3706842, 3706852, 3706876, 3706881, 3706887, 3706843, 3798911, 3798944, 3798946, 3798952, 3798920, 3706872, 3706875, 3673065, 3575520, 3706847, 3552967, 2886727, 3575534, 2910096, 3798943, 3798916, 3706854, 3706857, 3706848, 3706841, 3706859, 3798923, 3798932, 3591809, 3706867, 2960066, 3798950, 3798912, 3706870, 2934322, 2790546, 3591784, 3706862, 3706889, 3706851, 3915805, 4053648, 3915816, 4053670, 3915795, 3915835, 3915773, 3915743, 3915799, 3915806, 3915813, 3915793, 3915761, 3915744, 3915810, 3915755, 3915800, 3915811, 3915749, 3915759, 3915752, 3915764, 3915762, 3915796, 3915780, 3915798, 3915792, 2071740, 2451348, 2522779, 2451363, 2542790, 2542794, 2230891, 2318959, 2288217, 2277686, 2393010, 2230909, 2451378, 2358653, 3706892, 3575555, 3591789, 2736272, 2960087, 3575530, 3552957, 2629750, 2638924, 2552757, 3673058, 3575519, 3673068, 2648856, 2736265, 3673050, 3591820, 3591812, 2629756, 2648843, 3591821, 3591811, 2629758, 3552971, 3575518, 2552769, 2825654, 3552949, 3673049, 3552969, 2934324, 3552947, 3552981, 2552758, 3552931, 3552948, 3552984, 3552978, 3552943, 3552973, 3640584, 3552941, 3640579, 3640582, 3640578, 3673023, 3673046, 3552980, 3640577, 3552955, 3591830, 3640583, 3673054, 2629743, 3575544, 3591781, 3706882, 3798942, 3706888, 3706877, 3706891, 3706895, 2825660, 3706864, 3706861, 3798951, 3673056, 3673030, 3706856, 3706853, 3706869, 3591783, 3798953, 3798949, 3706855, 3706880, 2767195, 4053647, 3915769, 3915833, 3915777, 4053694, 3915803, 4053675, 4265854, 4265848, 4091800, 3915768, 4053656, 4265852, 4053646, 3915784, 4053666, 4091804, 3915823, 3915814, 4439180, 4265878, 3915820, 4053681, 4053671, 4346139, 4265842, 4439185, 4439193, 4439179, 4439196, 4439191, 4439195, 4439182, 4439183, 4439184, 4439188, 3915742, 3915819, 3915794, 3915786, 4091793, 4053672, 4053659, 4053674, 4265843, 4091795, 4091806, 4346143, 4053689, 4053673, 4053695, 4346147, 4346145, 4091792, 4265837, 4346148, 4346149, 4346133, 4439190, 3915809, 3915751, 4265874, 4439198, 4265885, 4346156, 4346135, 4346167, 4091803, 3915791, 4265864, 3915750, 4346119, 4265867, 4346150, 4346153, 4053684, 4265884, 4346161, 3915765, 4346166, 4053657, 4053692, 3915747, 3915753, 2522761, 2522792, 3575513, 2825638, 2552767, 3591793, 2886731, 2638926, 3552951, 2648862, 2790560, 2648822, 2648829, 2648844, 3591819, 3552988, 2552755, 3673041, 3552934, 3591808, 3640576, 3640581, 2910112, 3798938, 3798954, 2790558, 3591826, 3706838, 3673051, 3706894, 4053662, 4346124, 4053679, 4091796, 4439197, 4265861, 4265866, 4053650, 4265873, 3915774, 3915788, 3915771, 4053685, 3915778, 3915756, 4053687, 3915757, 4346137, 3915767, 4091797, 4091794, 4346146, 4439199, 4053682, 4439187, 4439186, 4091788, 4346160, 4439202, 4346157, 4091807, 4346159, 4091805, 3915802, 4053661, 4843363, 4971264, 4903117, 4705774, 4971252, 4802559, 4705770, 4802561, 4802568, 4705779, 4705780, 4705783, 4802564, 4802566, 4802567, 4824797, 4843353, 4843367, 4843362, 4843366, 4892361, 4892362, 4892356, 4892357, 4892360, 4892372, 4971254, 4971262, 4971266, 4843358, 4843359, 4843361, 4843364, 5054074, 5075663, 5075657, 4931731, 4931737, 4931739, 4971265, 4853695, 4853701, 4892363, 4802551, 4802563, 4824813, 4824811, 4705762, 4705765, 4853699, 4925028, 4843357, 4925042, 5075654, 5054084, 5075662, 5075672, 4705768, 4892359, 4903114, 5054073, 4971249, 4971251, 4705782, 4824807, 4892370, 4705781, 4824800, 4705778, 4705771, 4824812, 4903128, 4843351, 4705758, 4971248, 4824810, 4802558, 2825659, 3575547, 2648836, 2736264, 2934323, 2790559, 3673052, 2886720, 3575548, 2790540, 3552960, 3798914, 2934331, 3575550, 2844181, 4439200, 4053676, 4053664, 4091790, 4439192, 4265869, 4053677, 4346131, 3915745, 4265872, 4265879, 4439201, 4346118, 4346134, 4265844, 4265858, 4346142, 4053690, 4346155, 4265838, 4346136, 3915787, 4346130, 4903123, 4843349, 4903132, 4931732, 4853693, 4853686, 4971257, 4824806, 4892369, 4892366, 4802552, 4903113, 4903119, 5075673, 4853700, 5237807, 4853691, 5245837, 5237810, 5237788, 4802565, 4925034, 4903129, 5245847, 5237800, 5075664, 5075671, 4853694, 5054075, 5054080, 5237792, 4931734, 5245852, 4824808, 4705785, 5245851, 4892371, 4843352, 5245849, 4971263, 4824802, 5075658, 4824805, 4971258, 4931736, 5237793, 4802562, 4802555, 4705769, 4925025, 4802550, 4802557, 4903131, 4931749, 4903124, 5237797, 2522753, 3915797, 3915812, 4053665, 3915782, 4053658, 4053669, 4346129, 4346151, 4346158, 4265882, 4265857, 3915776, 4265846, 4903125, 4705777, 5245841, 5237783, 4925041, 4903120, 4931743, 4892367, 5245836, 5245842, 4931747, 5237789, 4705784, 4931745, 5054078, 5237805, 5245843, 4925033, 5245835, 5934857, 5936423, 5934871, 5934841, 6204388, 5925011, 5925013, 5925019, 5925009, 5925015, 5925016, 5925021, 5936463, 5936414, 5936418, 5936419, 5936424, 5936436, 5934805, 5934809, 5934840, 5934849, 5934851, 5934853, 5934855, 5934852, 5934844, 6204372, 6204375, 6204378, 6204394, 6204312, 6204314, 5925017, 5936461, 5936422, 5936450, 6204320, 6204321, 6204323, 5936425, 5934811, 5934876, 5936452, 5936457, 5925122, 5936421, 5934864, 5936433, 5934808, 6204315, 5934810, 5925028, 5936448, 5936434, 5936445, 5925111, 5925018, 1259682, 4265847, 4346121, 4265865, 4705786, 4824804, 4903126, 4931738, 5237791, 4925026, 4853697, 4931746, 5934804, 5936467, 5936449, 5936415, 6309293, 6204397, 6204322, 6505434, 6204310, 6309282, 5936420, 5936432, 5936437, 5936440, 6204383, 6204409, 6505450, 6309287, 6204413, 6309289, 6309290, 6505442, 6309279, 6505437, 6204407, 6204392, 6204402, 6204425, 6204385, 6204313, 5925012, 6204400, 6204395, 6204398, 5925010, 6462906, 6418168, 6418182, 4853698, 5075656, 5054072, 4903118, 5237795, 5934842, 6505443, 6309297, 6505435, 5925119, 5936451, 6520989, 6520968, 6505451, 6505441, 6520971, 6520980, 6520984, 6505452, 6505453, 5934867, 5936460, 6505447, 6204408, 6204422, 5934843, 5936456, 6505439, 6204379, 6204319, 6505446, 6833446, 6833324, 6833330, 6833331, 6833348, 6833350, 6833354, 6462921, 6833357, 6833341, 6833342, 6418163, 6833604, 6853851, 6418176, 6462915, 6833360, 4931744, 5054088, 5054076, 5054086, 5237809, 5934850, 6309286, 6520987, 5934869, 6204316, 6204424, 5925027, 5936446, 6520969, 5925115, 6505448, 5936413, 6309288, 7041509, 6833462, 7041535, 7042219, 7042221, 7042222, 7042225, 7042226, 7042243, 7042245, 7042246, 7042247, 6847150, 6847151, 6847153, 6833599, 6847133, 6833339, 6833458, 6847139, 6847142, 6847147, 6847156, 6853824, 6853828, 7041520, 7041538, 7042229, 7042249, 7042250, 7042253, 7042255, 7042256, 7042258, 7042267, 7042262, 7042282, 7042285, 7042269, 6833453, 6833589, 7042244, 7041508, 6833595, 6833356, 6833355, 6418164, 6847149, 6853825, 6853830, 7042220, 5934845, 6520973, 6520970, 6520992, 6204393, 6505436, 5934848, 6204387, 6520965, 5925125, 6833597, 6847138, 7042223, 7042272, 7041540, 7042261, 7041534, 6833594, 6853827, 7041514, 7042227, 7041516, 7041507, 7042276, 6853831, 7041505, 6833605, 6847125, 6847128, 6833336, 6462907, 6833461, 6462908, 6520974, 6204406, 6204325, 6853840, 6462920;
b) Item Number 314-02-03, UDI/DI 10885862084484, Serial Numbers: 0777367, 0745391, 0777366, 0777351, 0774821, 0745373, 0777363, 0745384, 0774826, 0747902, 0774820, 0777359, 0747919, 0745390, 0747912, 0774852, 0774803, 0774818, 0774842, 0774817, 0774840, 0747911, 0774811, 0774806, 0745389, 0774807, 0774832, 0774847, 0774812, 0774838, 0774839, 0774827, 0747914, 0745383, 0774844, 0745386, 0774846, 0924471, 0777355, 0777350, 0774850, 0774814, 0924497, 0774831, 0774801, 0747905, 0774800, 0745380, 0745385, 0745388, 0774809, 0747908, 0777362, 0774841, 0747918, 0747900, 0924480, 0774851, 0774804, 0774815, 0774823, 0774825, 0774843, 0777353, 0747915, 0924473, 0924467, 0745377, 0924484, 0774802, 0924457, 0777361, 0924466, 0924488, 0924481, 0924494, 0924483, 0747916, 0924496, 0747901, 0924472, 0924478, 0774824, 0747913, 0768485, 0768486, 0924498, 0924448, 0774848, 0768487, 0777358, 0924464, 0777352, 0924487, 0924485, 0774829, 0924447, 0924452, 0924451, 0924469, 0924482, 0774835, 0924446, 0924455, 0924491, 0924461, 0924479, 0924499, 0774805, 0774837, 0774816, 0745381, 0924475, 0924490, 0774836, 0924474, 0924492, 0774845, 0924465, 0777360, 0747910, 0745379, 0774830, 1002427, 1002430, 1002380, 1002358, 1002403, 1002431, 1002414, 1002420, 1002425, 1002422, 1002372, 1002367, 1002381, 1002398, 1002394, 1002388, 1002411, 1002361, 1004722, 1002378, 1002424, 1002357, 1002426, 0924495, 0745374, 0924489, 0924468, 0774834, 0745376, 0774808, 0924477, 0924450, 0747906, 0747917, 0777368, 0924476, 0924456, 0774819, 0924470, 0747907, 0924459, 0924463, 0745375, 0747904, 0747909, 0924493, 0777354, 0745387, 0774828, 0924454, 1004719, 1004718, 1002364, 1002370, 1002405, 1002379, 1002419, 1002404, 1002417, 1002385, 1002401, 1002366, 1002382, 1002389, 1002391, 1002399, 1002410, 1002362, 1002434, 1002368, 1002415, 1002371, 1002386, 1002423, 1004714, 1002408, 1002421, 1004713, 1004716, 1002428, 1004726, 1002413, 1002433, 1002384, 1002373, 1004721, 1002397, 1002402, 1004717, 1004725, 1002375, 1187203, 1158674, 1187186, 1158693, 1158679, 1158676, 1187204, 1158678, 1158675, 1187205, 1187197, 1158683, 1158672, 1158689, 1158685, 1158677, 1187202, 1187199, 1187196, 1158684, 1158681, 1187183, 1158686, 1158673, 1187184, 1187187, 1187191, 1187193, 1187194, 1158690, 1158682, 1187188, 1234722, 1321294, 1355699, 1234720, 1282375, 1355673, 1234719, 1234730, 1234727, 1234707, 1282363, 1321309, 1355674, 1282364, 1321313, 1234757, 1234740, 1282362, 1234742, 1321310, 1234752, 1321300, 1234723, 1321295, 1321308, 1234697, 1234693, 1341720, 1321305, 1321312, 1290555, 1234713, 1234701, 1341734, 1282381, 1282384, 1234696, 1234714, 1234704, 1234725, 1290557, 1234709, 1234756, 1341731, 1341723, 1321314, 1234716, 1321297, 1355681, 1321291, 1282366, 1234738, 1282379, 1234706, 1234728, 1234726, 1234721, 1355666, 1355667, 1355670, 1355700, 1355715, 1282380, 1290551, 1234751, 1355676, 1234739, 1234744, 1234746, 1282365, 1282369, 1282371, 1282372, 1282373, 1282376, 1282383, 1282385, 1290553, 1290554, 1290559, 1234743, 0774822, 0777357, 0777364, 0924500, 0924453, 0747903, 0777365, 0777349, 0924486, 0774833, 0774810, 1004727, 1002376, 1002392, 1004712, 1004724, 1002395, 1002377, 1002406, 1002429, 1002409, 1002416, 1002393, 1002383, 1002418, 1004723, 1002356, 1158671, 1158691, 1187200, 1187201, 1187189, 1187190, 1282374, 1355696, 1408154, 1355695, 1408147, 1341725, 1425199, 1341732, 1426207, 1487896, 1521303, 1536890, 1521304, 1355677, 1408161, 1321306, 1408150, 1477892, 1234695, 1553933, 1487914, 1355688, 1408155, 1408156, 1487904, 1536882, 1408128, 1234712, 1511431, 1487889, 1536871, 1341724, 1384107, 1355671, 1425196, 1457667, 1355720, 1355705, 1355679, 1426210, 1536877, 1234741, 1408162, 1426199, 1457659, 1384104, 1408125, 1521311, 1234758, 1487895, 1321299, 1408170, 1355694, 1408167, 1234691, 1408153, 1234748, 1477893, 1477897, 1355714, 1234754, 1234737, 1408140, 1341726, 1425201, 1553947, 1425178, 1355672, 1553950, 1536883, 1426204, 1321311, 1384108, 1450598, 1426213, 1234731, 1408135, 1321304, 1536876, 1426202, 1234694, 1355683, 1355668, 1408132, 1341715, 1321290, 1234700, 1425190, 1553949, 1355703, 1425193, 1425182, 1536869, 1341721, 1425183, 1408120, 1234715, 1234698, 1408124, 1355690, 1408157, 1408151, 1521312, 1355684, 1321298, 1355717, 1408121, 1355697, 1511437, 1487908, 1536880, 1487907, 1355685, 1384099, 1511430, 1425194, 1553946, 1384100, 1426196, 1521309, 1408152, 1425185, 1355709, 1341728, 1426208, 1425188, 1457660, 1355701, 1341736, 1408166, 1384103, 1234732, 1450595, 1290552, 1355711, 1384105, 1426205, 1457670, 1487899, 1521316, 1408126, 1408172, 1536889, 1553948, 1553934, 1487912, 1408133, 1355686, 1355669, 1321315, 1408143, 1408160, 1341719, 1321302, 1408136, 1355702, 1282368, 1355692, 1321303, 1408168, 1282377, 1341717, 1425177, 1408130, 1355689, 1384112, 1477896, 1355707, 1511429, 1511435, 1521308, 1487917, 1477891, 1487891, 1553974, 1408123, 1408127, 1457669, 1511425, 1511427, 1521310, 1536891, 1536888, 1553964, 1553973, 1457661, 1457668, 1477890, 1487915, 1234708, 1426215, 1384109, 1553971, 1355675, 1425179, 1425180, 1355704, 1355665, 1384111, 1408159, 1426193, 1426212, 1425187, 1521307, 1521314, 1536870, 1536878, 1536884, 1234736, 1426203, 1426209, 1426206, 1321307, 1234734, 1234755, 1553944, 0777356, 1002396, 1002435, 1002359, 1002400, 1002412, 1004710, 1004711, 1002363, 1002390, 1004720, 1158680, 1187195, 1187192, 1158692, 1426201, 1536872, 1426197, 1290556, 1553940, 1355682, 1457671, 1355664, 1425181, 1477899, 1355713, 1384102, 1408134, 1426200, 1536879, 1553939, 1425192, 1553956, 1553970, 1553932, 1487898, 1408119, 1234702, 1408149, 1290558, 1282370, 1553961, 1355663, 1487894, 1426191, 1408164, 1408148, 1553957, 1408142, 1553938, 1487905, 1553953, 1553965, 1234724, 1553968, 1341727, 1341713, 1536886, 1355687, 1536867, 1536875, 1553935, 1487906, 1521301, 1487910, 1290550, 1426214, 1553952, 1487903, 1234753, 1487900, 1282367, 1408158, 1536885, 1408131, 1536868, 1487911, 1355712, 1553941, 1553955, 1553966, 1487897, 1282386, 1234711, 1408137, 1321292, 1408163, 1425186, 1341716, 1457665, 1521299, 1408169, 1511434, 1355691, 1408138, 1425184, 1408146, 1450597, 1355708, 1536895, 1553963, 1521315, 1553972, 1341730, 1384110, 1553958, 1425191, 1234733, 1521306, 1536874, 1384101, 1408171, 1426195, 1408129, 1341733, 1536881, 1511433, 1536892, 1553954, 1426192, 1425197, 1282387, 1341714, 1457662, 1426194, 1553969, 1536894, 1553937, 1625978, 1734634, 1679476, 1718026, 1718038, 1734635, 1909507, 1909485, 1758788, 1734646, 1734641, 1627067, 1679461, 1734637, 1802713, 1909358, 1627064, 1758787, 1814005, 1758786, 1909375, 1625989, 1625960, 1625977, 1734652, 1758804, 1802719, 1814030, 1625972, 1625968, 1625971, 1625992, 1758808, 1794961, 1734656, 1794981, 1794953, 1802699, 1722249, 1625985, 1909381, 1625973, 1814012, 1794965, 1814011, 1679464, 1625958, 1814031, 1627059, 1814017, 1625974, 1625975, 1758802, 1734648, 1679468, 1802721, 1625993, 1734642, 1909495, 1679479, 1758778, 1814008, 1625986, 1625981, 1625962, 1794975, 1794967, 1625995, 1718034, 1734627, 1679480, 1758798, 1802722, 1802712, 1734636, 1794973, 1679484, 1758784, 1627066, 1627075, 1679471, 1814029, 1814033, 1679463, 1625967, 1758779, 1909498, 1679466, 1679470, 1794980, 1734647, 1794976, 1814027, 1758780, 1734649, 1814014, 1814010, 1794956, 1625979, 1625961, 1718037, 1936627, 1627072, 1758782, 1679488, 1679486, 1718027, 1794957, 1625984, 1802709, 1734630, 1627060, 1814022, 1625965, 1802720, 1625963, 1625970, 1814021, 1627065, 1814026, 1814016, 1909488, 1679474, 1794963, 1802697, 1627061, 1679483, 1625964, 1718040, 1718041, 1625966, 1734653, 1758791, 1758795, 1758799, 1758811, 1758790, 1758803, 1794948, 1794964, 1802704, 1802708, 1802710, 1802695, 1802706, 1802714, 1758800, 1802715, 1734650, 1758812, 1679481, 1679469, 1718043, 1718024, 1625982, 1718033, 1718031, 1679459, 1625983, 1679458, 1679478, 1794952, 0924458, 1004715, 1002407, 1002369, 1002387, 1187185, 1158687, 1282382, 1477895, 1234718, 1536873, 1234749, 1487892, 1341729, 1536887, 1487893, 1487890, 1408145, 1384106, 1553943, 1487909, 1234717, 1408122, 1355719, 1426198, 1450599, 1408141, 1408139, 1426211, 1425189, 1487901, 1511426, 1553959, 1234735, 1234692, 1487913, 1355693, 1425195, 1487902, 1477898, 1282378, 1234703, 1457663, 1408144, 1487916, 1234710, 1234745, 1457666, 1321301, 1341718, 1341712, 1450596, 1734655, 1718028, 1982431, 1814003, 2005097, 1814020, 1758792, 1909380, 1718044, 1734640, 1758789, 1734651, 1909482, 1982441, 1734624, 1909361, 1625987, 2056256, 1936641, 1625959, 1802717, 1982423, 1982435, 1734644, 1909469, 2056275, 2056266, 1936617, 1909483, 1627070, 1814019, 1758810, 1802711, 1909484, 1936607, 1814023, 1936614, 1936619, 1909372, 2056276, 2048775, 1936624, 1936606, 1802716, 1627068, 2056260, 1909371, 1625980, 1936633, 2005093, 1814013, 1734645, 1909468, 1909479, 1909376, 1909363, 1909383, 1909506, 2026629, 2048765, 1909353, 1909359, 1627063, 2056259, 2048770, 1814032, 1909518, 2005087, 1936640, 1802700, 1627071, 2005089, 1909378, 1909400, 1909486, 1909500, 1936634, 1909513, 2026631, 2048776, 2056257, 2026633, 2026632, 1679472, 1718042, 1909505, 2048772, 1936613, 1909504, 1982427, 1758781, 1982430, 1982422, 1909515, 1909348, 1718036, 2056264, 2026621, 1734633, 1734629, 1982437, 1982424, 1802707, 1909474, 1802724, 1909390, 2005096, 2048746, 1794974, 1814015, 1909398, 2048737, 2048749, 1814002, 1794955, 2048763, 1909369, 2048742, 2026619, 1909387, 1982449, 1794979, 2026622, 1909516, 1802698, 1627074, 1909517, 2026634, 1909501, 1982450, 1982443, 1794978, 1679477, 1909477, 1909491, 2026624, 2056269, 1679456, 1936621, 2056255, 2026620, 1909364, 1982434, 2048771, 1758794, 1936618, 1679487, 1802725, 1758783, 1909471, 2056253, 1909362, 1794958, 1909368, 2048738, 2048766, 1679485, 1909481, 1982439, 1758797, 1734631, 1794969, 1909351, 1936626, 1909494, 1936631, 1802703, 1909466, 1909476, 1909386, 1909493, 1679457, 1802694, 1909492, 2056282, 2056279, 2048755, 1909511, 1909352, 1909366, 1909357, 2026635, 1909510, 1794972, 1679473, 1734626, 1936612, 1758809, 1909519, 1909497, 1982448, 1936620, 2056271, 2056258, 1909356, 2005086, 2056281, 1909499, 1909365, 1936635, 1936638, 2048767, 1794951, 1982440, 1758805, 2056267, 1909360, 1909503, 1909374, 1814009, 1814006, 2005084, 1909384, 2005099, 2048753, 1909349, 1909480, 1802702, 1936615, 1758785, 1909391, 1982420, 2026618, 1909346, 1909394, 1909388, 1909470, 1982438, 1982428, 2048740, 2048743, 2056274, 1802718, 2026627, 1734658, 1718030, 2005098, 1909382, 1794966, 1982445, 1802705, 1758807, 1982451, 2005100, 2005091, 1802701, 2026625, 2026623, 2026636, 2005082, 1734628, 1758806, 1909370, 2146074, 2084957, 2117175, 2117176, 2146072, 2172827, 2172850, 2172840, 2140283, 2117203, 2146065, 2117206, 2140267, 2084937, 2084948, 2172842, 2172851, 2140277, 2084945, 2146063, 2117201, 2140276, 2140288, 2146080, 2084944, 2117184, 2084942, 2117174, 2172834, 2084956, 2117173, 2084958, 2117197, 2140272, 2140297, 2172811, 2117190, 2146062, 2084940, 2146086, 2146095, 2117177, 2117192, 2140290, 2140278, 2140266, 2084961, 2172824, 2146069, 2140274, 2117189, 2117183, 2140279, 2117210, 2146067, 2140293, 2146081, 2140294, 2117179, 2117188, 2117196, 2117198, 2140268, 2140295, 2084935, 2140269, 2140286, 2146090, 2182949, 2182951, 2182975, 2182959, 2084950, 2084947, 2084943, 2146085, 1002374, 1002365, 1158688, 1408165, 1355718, 1355680, 1321293, 1355698, 1521302, 1521305, 1553962, 1553960, 1321296, 1511424, 1511428, 1553942, 1536893, 1234729, 1521300, 1355678, 1425200, 2048752, 1909389, 2048754, 1718035, 1734632, 2056254, 1909367, 2056268, 1909347, 1794954, 2048762, 2048769, 2005090, 2048747, 2048744, 2056278, 1936625, 1679489, 2048764, 1936630, 2026617, 1794962, 1734639, 1802723, 2056270, 2048761, 1718032, 1794947, 1679482, 2056277, 1982425, 1909475, 1909512, 2005088, 2005081, 1909354, 1936629, 1909392, 1625969, 1758796, 1909350, 1909377, 1909509, 1794959, 1909508, 1936609, 1814007, 2005083, 2056263, 1909473, 1625994, 1814024, 1982436, 1909489, 2048773, 1936622, 1794977, 1625991, 2056280, 1909393, 1625976, 1909478, 1982429, 2005095, 1909379, 1936623, 2048748, 2048741, 2048745, 1722255, 1936637, 1734625, 1909399, 1936628, 1936636, 2026626, 2048751, 1909496, 1982432, 2359188, 2231787, 2380120, 2172821, 2146093, 2288203, 2359170, 2172841, 2469284, 2508068, 2242576, 2323553, 2182981, 2182972, 2172822, 2254293, 2238816, 2242575, 2288130, 2323542, 2323552, 2242547, 2288186, 2146068, 2172818, 2242559, 2288140, 2146066, 2182957, 2117205, 2254292, 2117180, 2172852, 2433170, 2182976, 2469293, 2146082, 2288147, 2331569, 2140291, 2172830, 2172829, 2182955, 2238813, 2172825, 2331571, 2172812, 2288212, 2288139, 2172831, 2140273, 2182973, 2288188, 2433173, 2084946, 2380103, 2380108, 2508065, 2182960, 2242561, 2288187, 2238822, 2288137, 2469288, 2182968, 2254294, 2359205, 2117199, 2288138, 2359172, 2254284, 2323562, 2182980, 2146091, 2288153, 2323534, 2172814, 2231795, 2331566, 2467464, 2433163, 2238808, 2288155, 2359206, 2331575, 2172844, 2323548, 2359186, 2172815, 2238817, 2288204, 2242549, 2331570, 2380128, 2182941, 2146087, 2231799, 2288129, 2359167, 2288179, 2242574, 2254286, 2238818, 2467462, 2323550, 2182942, 2117182, 2231785, 2323557, 2331561, 2323564, 2467466, 2242567, 2254306, 2182987, 2146077, 2172846, 2231784, 2242571, 2182937, 2288158, 2182944, 2288199, 2140270, 2117204, 2231798, 2242551, 2323531, 2238814, 2182978, 2323555, 2231794, 2242560, 2288209, 2288160, 2172820, 2254302, 2231797, 2146096, 2182958, 2182939, 2323539, 2182943, 2359169, 2084939, 2140282, 2498458, 2238805, 2467472, 2146079, 2498469, 2288157, 2242556, 2231796, 2288127, 2380140, 2433138, 2146075, 2140284, 2238820, 2242555, 2359194, 2140296, 2140285, 2288159, 2498439, 2182979, 2182989, 2182990, 2182950, 2117212, 2146089, 2242568, 2331560, 2172805, 2530232, 2231783, 2469282, 2117208, 2146078, 2182953, 2323535, 2182965, 2288183, 2331564, 2433146, 2331584, 2433148, 2323547, 2238811, 2469265, 2433143, 2140281, 2530242, 2182991, 2140280, 2288202, 2140287, 2242580, 2288146, 2238810, 2288134, 2231802, 2359173, 2433160, 2117185, 2182954, 2288141, 2140275, 2254283, 2182977, 2288206, 2359183, 2359196, 2433171, 2172806, 2172838, 2238812, 2359181, 2323560, 2288205, 2359185, 2146070, 2380114, 2380122, 2146084, 2238807, 2433133, 2182945, 2469287, 2254299, 2323559, 2117187, 2242566, 2182938, 2172809, 2172810, 2172843, 2238823, 2084959, 2242553, 2242562, 2380123, 2380130, 2380105, 2172817, 2172847, 2182935, 2182946, 2182956, 2182961, 2182963, 2182983, 2146064, 2172845, 2172848, 2182940, 2182971, 2182985, 2172813, 2254305, 2254274, 2254285, 2242573, 2288128, 2254273, 2254276, 2254281, 2254301, 2288196, 2288208, 2254295, 2323554, 2331558, 2331559, 2331567, 2331568, 2288145, 2288149, 2288151, 2288154, 2331563, 2331576, 2331581, 2331582, 2359171, 2359203, 2172804, 2182948, 2380138, 2359198, 2380136, 2288197, 2331556, 2380110, 2380119, 2182967, 2380142, 2433176, 2467455, 2467461, 2467467, 2469271, 2467468, 2469275, 2469261, 2469272, 2433139, 2433149, 2508043, 2508045, 2469262, 2469263, 2469286, 2469306, 2469307, 2508064, 2140289, 2117194, 2117178, 2117181, 2182969, 2231793, 2288142, 2331585, 2172837, 2380139, 2433136, 2359175, 2359189, 2359190, 2433151, 2433152, 2469298, 2469299, 2469270, 2469283, 2467449, 2146076, 2242565, 2238821, 2242550, 2359176, 2380106, 2380115, 2433161, 2433177, 2084962, 2231792, 2242581, 2242552, 2172823, 2433135, 2433162, 2380117, 2469295, 2172826, 2117193, 2172832, 2182984, 2380112, 2380104, 2084938, 2359191, 2469278, 2172816, 2254279, 2238809, 2254288, 2433168, 1553936, 1341722, 1355706, 1814025, 2005092, 1625988, 1625990, 1936605, 2048759, 1982444, 2048774, 2048739, 1679467, 2048750, 1982442, 2048757, 1936611, 1722252, 2005085, 1936642, 2026628, 1794982, 1982447, 1909487, 1718039, 1982446, 2583730, 2469285, 2583542, 2359184, 2583726, 2530250, 2288211, 2288178, 2288150, 2608948, 2433153, 2498448, 2380135, 2380131, 2530231, 2583590, 2433141, 2583580, 2433164, 2583594, 2323538, 2498443, 2498441, 2467450, 2288192, 2146094, 2323537, 2583587, 2608935, 2323532, 2331562, 2380121, 2498472, 2323551, 2530237, 2467457, 2469267, 2433178, 2467451, 2608931, 2608885, 2331565, 2530245, 2467473, 2172808, 2608905, 2608889, 2508055, 2380113, 2433165, 2583724, 2508062, 2608907, 2331572, 2359168, 2380109, 2323528, 2433130, 2231791, 2530248, 2359178, 2583545, 2323561, 2288193, 2359179, 2254289, 2583608, 2583599, 2583610, 2583736, 2508040, 2583582, 2508053, 2583541, 2530252, 2498460, 2498455, 2380111, 2583554, 2238806, 2583721, 2583548, 2583543, 2288194, 2172807, 2331573, 2323543, 2498466, 2242577, 2583595, 2583612, 2433131, 2469300, 2467456, 2583598, 2498462, 2254297, 2498473, 2254278, 2254296, 2467474, 2583615, 2608944, 2530238, 2583556, 2608932, 2583553, 2323565, 2608909, 2608898, 2608920, 2323529, 2608887, 2583583, 2530240, 2433172, 2146073, 2146083, 2583737, 2608930, 2469297, 2508050, 2231789, 2530246, 2508069, 2508060, 2530233, 2242558, 2498456, 2583729, 2583567, 2469294, 2530259, 2380118, 2433145, 2433144, 2288207, 2117191, 2433174, 2508054, 2583714, 2469301, 2530247, 2498461, 2583708, 2508058, 2467465, 2469296, 2608927, 2608896, 2254300, 2498459, 2242569, 2359187, 2498453, 2254303, 2508041, 2331578, 2498450, 2583588, 2583552, 2608895, 2583715, 2583719, 2583727, 2583731, 2583561, 2583570, 2583559, 2583573, 2583544, 2583557, 2583572, 2583566, 2583575, 2182936, 2608924, 2608943, 2608953, 2323544, 2380125, 2498451, 2323536, 2117207, 2117211, 2608928, 2583734, 2608936, 2583577, 2583601, 2583604, 2242570, 2433158, 2254282, 2146092, 2498442, 2583735, 2583614, 2469280, 2498449, 2467469, 2498465, 2508057, 2583738, 2608940, 2254287, 2469291, 2498474, 2467454, 2583576, 2433156, 2467458, 2182934, 2583551, 2608946, 2773578, 2825689, 2690702, 2690726, 2807377, 2627505, 2844258, 2690728, 2627495, 2825674, 2627487, 2627506, 2825681, 2744796, 2627489, 2758972, 2773587, 2773599, 2807376, 2744794, 2844252, 2744800, 2825683, 2773579, 2690727, 2627497, 2627508, 2690700, 2690703, 2690725, 2690731, 2690732, 2690707, 2690718, 2690722, 2690737, 2690699, 2690709, 2690714, 2690740, 2690743, 2773597, 2690701, 2773581, 2773589, 2773590, 2773595, 2773596, 2773600, 2690721, 2807367, 2807368, 2807371, 2807372, 2807380, 2807381, 2825670, 2825679, 2825677, 2825672, 2825675, 2825662, 2825666, 2825671, 2844253, 2844254, 2844262, 2844268, 2854342, 2854346, 2854348, 2854355, 2854365, 2825667, 2854343, 2854345, 2854347, 2854349, 2854350, 2854352, 2854353, 2854359, 2854360, 2854361, 2854364, 2690711, 2627491, 2627517, 2627516, 2690698, 2690719, 2690729, 2690738, 2690742, 2627496, 2627512, 2773577, 2773594, 2773602, 2690717, 2627499, 2627518, 2825685, 2690710, 2690741, 2773585, 2627494, 1553951, 1909490, 2056273, 2056265, 1722251, 1627073, 2056272, 1936639, 1909514, 1758793, 1802696, 1814004, 1679475.2048756, 2508066, 2608947, 2172839, 2608915, 2498445, 2498444, 2508047, 2288184, 2530258, 2583586, 2608942, 2469289, 2288181, 2182970, 2583728, 2608893, 2469274, 2146071, 2530256, 2433159, 2583609, 2583564, 2331577, 2467459, 2433142, 2608918, 2583591, 2288133, 2508044, 2583596, 2583713, 2583718, 2117195, 2380137, 2182947, 2117209, 2331583, 2608883, 2583589, 2323556, 2238824, 2288200, 2608923, 2583565, 2433147, 2498457, 2508061, 2583555, 2231790, 2508051, 2288201, 2583717, 2608886, 2508049, 2583562, 2530257, 2530234, 2608934, 2288198, 2469273, 2583720, 2583603, 2608884, 2583584, 2323563, 2583606, 2608937, 2608951, 2608925, 2583711, 2583560, 2498446, 2583574, 2508063, 2288144, 2288136, 2469302, 2498452, 2172853, 2608926, 2583611, 2242557, 2242572, 2084954, 2608929, 2323540, 2608955, 2469281, 2608960, 2583723, 2583571, 2825673, 3601431, 2627502, 2825691, 2942251, 2844247, 2942258, 3601388, 2807379, 2942250, 2844248, 3563471, 3563462, 2744795, 2744802, 2942272, 2907049, 2967105, 3569877, 3694060, 2690697, 2907058, 3601424, 2967098, 3569873, 2807378, 2942270, 2825686, 2807374, 2844265, 2825690, 3569872, 2773588, 2807373, 2773576, 2744791, 3601432, 2844259, 3601429, 2825688, 2758977, 2758981, 2967091, 2825661, 2844260, 2942263, 2907056, 2690744, 2690706, 2844249, 2690708, 3694089, 3694078, 2690696, 2825676, 2825668, 3563468, 2744797, 3569893, 2744801, 2907045, 2907039, 2942252, 2854370, 2907047, 2773582, 2854367, 2854369, 2854366, 2854358, 2854363, 2854371, 2907060, 2907067, 2967110, 2967119, 3563453, 3563457, 3569878, 3569879, 3569892, 3601393, 2967093, 2854356, 2967109, 2967115, 3601395, 3601427, 2967113, 3649084, 3649098, 3649107, 2967097, 2967107, 3601398, 3601410, 3601412, 3694055, 3694058, 3694080, 3694086, 3694091, 3694094, 3694095, 3694100, 3694088, 2690720, 3569882, 3601385, 3601416, 3601417, 3601413, 2967108, 2967120, 3649099, 3649103, 2907036, 3649094, 3649108, 3649088, 3649087, 3649106, 3694083, 3649101, 2690733, 2942274, 2967103, 3601389, 2942254, 2907048, 3569881, 3569894, 3649091, 3601433, 2942266, 2967114, 2907033, 2773601, 2907042, 2907061, 3694087, 2627490, 2825687, 3649093, 2773598, 3601423, 2854351, 2690716, 3569883, 2773584, 2627510, 2907069, 2942256, 2967101, 3563464, 3601390, 2854354, 3569897, 2627515, 2773575, 3601419, 3569867, 3601406, 2967104, 2967112, 2773591, 2967111, 2773580, 3569884, 1909467, 1909397, 1909502, 1909472, 2005094, 2469290, 2583546, 2182964, 2469266, 2530249, 2608950, 2323530, 2583613, 2583602, 2583579, 2254277, 2323566, 2084941, 2583592, 2498471, 2608904, 2467470, 2583607, 2288189, 2508059, 2583558, 2583705, 2359193, 2231788, 2359192, 2288195, 2433134, 2583605, 2433140, 2608891, 2380129, 2583585, 2288132, 2323533, 2508056, 2359180, 2182966, 2608899, 2583725, 2433169, 2467463, 2583550, 2469277, 2608921, 2288191, 2380132, 2380141, 2530235, 2583704, 2242563, 2288210, 2498470, 2117202, 2380116, 2288190, 3694063, 2907041, 2744788, 2690704, 2627493, 3569900, 3563467, 2690730, 3601400, 3563455, 2907066, 2825680, 2627509, 2967102, 2907062, 2907059, 3601397, 3563476, 3563473, 2844267, 2627500, 2744803, 2942267, 3569885, 2627507, 2907071, 2967118, 3601387, 3601401, 2907070, 3601426, 2744799, 2690724, 2844251, 2627504, 3694097, 3694066, 2907030, 2844255, 2844264, 2907063, 3601403, 2907051, 3569876, 3569886, 3601391, 3601405, 3569899, 3569890, 2825682, 3747193, 3747197, 3747203, 3747218, 3649092, 3694068, 3649097, 3649100, 3649095, 3601386, 3747215, 3747202, 3694093, 3694069, 3747199, 3747211, 3747212, 3747219, 3747221, 3747201, 3747208, 3747223, 3694064, 2942253, 3694059, 3694092, 3601428, 2854357, 2773586, 3694076, 3747195, 3694099, 3694073, 3747207, 3694054, 2907043, 3747209, 3694096, 3694079, 3649104, 3601408, 3694067, 3569898, 3747205, 3601399, 3694062, 2967100, 3747198, 3694075, 2690713, 3649086, 3601411, 3747191, 3649089, 2967099, 3694070, 3898859, 3915945, 3915919, 4033250, 4033205, 3898856, 3898849, 4033228, 4033243, 3889632, 3898915, 3915921, 4053623, 3898861, 3898919, 3915925, 3889626, 3915923, 3898855, 3898852, 3898921, 3915916, 3889639, 3889638, 3889644, 3898870, 3898910, 3915944, 4053625, 3898922, 3915928, 4033216, 4033217, 4033224, 3898909, 4053635, 3915934, 4033214, 4033221, 3889641, 3889647, 3898862, 3889623, 3889657, 4053596, 3898871, 3898901, 4033225, 4053616, 4053639, 3889624, 3889631, 3889650, 3889658, 3889659, 3889640, 3898912, 3898854, 3898850, 3898911, 3889652, 3915914, 3915936, 3889649, 3898899, 3898898, 3898916, 3889630, 3889613, 3915942, 3898904, 2583733, 2508046, 2254307, 2323546, 2608911, 2380133, 2508048, 2608933, 2608958, 2467448, 2469279, 2231801, 2469305, 2583706, 2583563, 2242579, 2231786, 2433150, 2608916, 2583722, 2288182, 2433132, 2242548, 2433154, 2084952, 2583707, 2967106, 2907044, 3747213, 3694081, 2758974, 3601394, 2690712, 2967095, 2758975, 3569870, 2744793, 2690734, 3563470, 3694082, 2844266, 2825669, 3694098, 2907053, 3563463, 2627503, 3694071, 3601430, 2627513, 3569888, 3601409, 3694102, 3694074, 3747224, 2690705, 3649102, 3569901, 3694084, 3747210, 3747194, 3747222, 4329518, 4053606, 3889617, 4053628, 4053603, 4053589, 4053588, 4053611, 4095157, 4095154, 4095344, 4346170, 4033227, 4053600, 3898848, 4346174, 4095350, 4033244, 4095161, 4033203, 3915931, 4346211, 3915946, 3915915, 4033207, 4095169, 3889654, 4095168, 4346205, 4346203, 3915924, 4033233, 3889642, 4053601, 4329517, 4346198, 3889612, 4053630, 4033237, 4033234, 4095166, 4053593, 4346176, 4346168, 3898863, 4346187, 4346173, 4033202, 3915939, 4329519, 4346192, 4033230, 4053604, 4053590, 4053595, 4053597, 4053598, 4053612, 4346182, 4412742, 4412746, 4412747, 4412752, 4412763, 4412773, 4412779, 4412780, 4412782, 4412783, 4473976, 4473987, 4473995, 4412735, 4412739, 4412778, 4473973, 4473967, 4473949, 4473950, 4473951, 4473956, 4473965, 4473970, 4473971, 3898900, 4033215, 3889646, 4095145, 4095345, 4095351, 4095151, 4473952, 4053637, 4095359, 4053633, 4095352, 4053607, 4053634, 4033238, 4053605, 4412741, 4473972, 3915930, 3889619, 4053608, 4346199, 3889648, 4095160, 4033226, 4053644, 4053592, 4033213, 4346215, 4346210, 4053638, 4346183, 4053622, 4033229, 3889635, 3915935, 4053610, 4095150, 4095364, 4053620, 3889625, 4033220, 4095363, 4412753, 4346206, 4053621, 4095367, 4095148, 4095355, 4053645, 4053643, 4053614, 4053640, 4095346, 4412751, 4095349, 4473992, 2498468, 2583568, 2608903, 2498454, 2608900, 3601422, 3601407, 3563472, 2907052, 2627511, 2942262, 2942268, 3694077, 3747216, 4346213, 4473968, 4033241, 4473966, 4095361, 4473960, 4473991, 4033235, 4033248, 4346184, 3889629, 3898908, 3898920, 4033223, 3898906, 3898917, 3898867, 3898851, 4473948, 4473980, 4412767, 3915943, 3898858, 4412758, 4346209, 4473958, 4412749, 4412760, 4412744, 4473978, 3889616, 4095353, 4412781, 3889645, 4095167, 4053631, 4473974, 4473977, 4473979, 4473990, 4053632, 3898866, 4412766, 4053615, 4346200, 4346178, 4412740, 4346181, 4095357, 4346189, 3898918, 4346188, 4412772, 4473994, 4683065, 4683063, 4847344, 4956967, 4683023, 4683038, 4683043, 4683045, 4683069, 4683044, 4847333, 4847346, 4683020, 4683072, 4825412, 4825418, 4825420, 4847327, 4847345, 4683030, 4683054, 4683062, 4683056, 4683061, 4683067, 4903164, 4903167, 4903169, 4903173, 4683068, 4921453, 4921457, 4921461, 4903166, 4903171, 4921452, 4903165, 4903168, 4921450, 4921454, 4861338, 4861340, 4861348, 4903180, 5075627, 5075630, 4903172, 4683019, 4683046, 4932978, 4932985, 4683073, 4683042, 4847338, 4683021, 5064788, 5075617, 4683049, 4921449, 4956958, 4683058, 4683033, 4861356, 4861346, 4903162, 4903176, 4683055, 4683036, 4683074, 4861351, 4921448, 4921451, 4683071, 4683041, 2744787, 2942275, 3569868, 3569875, 2907031, 3601402, 3694072, 2942269, 3694085, 3694061, 4412757, 4473975, 4329530, 3889620, 3889661, 4346214, 4053609, 4412768, 3898860, 3915932, 4473982, 4053629, 4329536, 4412775, 4329533, 4033222, 4412771, 3889651, 4473947, 3915922, 3898905, 4346216, 4473981, 4095368, 4329537, 4346191, 4033204, 4095146, 4412759, 5237018, 4847335, 4903170, 5064796, 4903163, 4825422, 4932970, 5241176, 4847334, 4683050, 4956965, 5075624, 5075626, 4956962, 5064800, 5075621, 5237007, 4903179, 4683066, 5241149, 4932973, 5237019, 4932984, 4932979, 5241177, 4956973, 5064790, 4847342, 5064803, 5241150, 5241154, 5241172, 4903177, 5241167, 5241168, 4683059, 4921456, 4921445, 4683052, 5075629, 4903161, 4825414, 4921444, 4903174, 5064792, 4956959, 5075625, 5064787, 4956955, 4861337, 4847343, 5064794, 2690723, 3601434, 4412756, 4053594, 4033212, 3915917, 3915918, 4329529, 4346194, 4329535, 4412761, 3915913, 4053617, 4346180, 4095360, 4329524, 4053619, 4473954, 4033209, 3889621, 4412738, 4932986, 5241153, 4956957, 5064799, 5237006, 4825416, 5236999, 5237000, 5237012, 5237013, 5237020, 4683070, 5237003, 4921459, 4861341, 5075628, 4683028, 5241173, 5241165, 5854573, 5854626, 5854673, 5854633, 5854634, 5854636, 5854638, 5854574, 5854690, 5854695, 5854679, 5854557, 5854676, 5854643, 5854645, 5854646, 5854650, 5854651, 5854656, 5854658, 5854668, 5854664, 6204080, 6204088, 6204100, 6204047, 6204053, 6204062, 6204138, 5854631, 5854592, 5854597, 5854602, 5854616, 5854614, 5854601, 5854568, 5854559, 5854615, 5854625, 5854671, 5854569, 5854576, 5854660, 5854667, 5854639, 5854644, 5854553, 5854662, 6204086, 6204105, 5854630, 5854593, 5854580, 5854617, 5854659, 5854605, 5854691, 5854696, 5854586, 5854588, 5854641, 4329521, 4346204, 4473955, 4412748, 4412765, 4861349, 5075614, 5064785, 5854653, 5854579, 5854665, 5854619, 6204078, 6204091, 5854561, 5854558, 5854648, 5854652, 5854663, 5854666, 5854681, 6204099, 6204052, 6204102, 6204065, 6204050, 6204051, 6204073, 6204120, 6204122, 6204109, 6204092, 6204095, 6204054, 6204061, 5854661, 6204126, 5854595, 6204108, 5854706, 5854627, 5854670, 6204066, 5854618, 5854608, 6204084, 5854554, 6461531, 6572922, 6572939, 6572943, 6572910, 6461520, 6572925, 6577948, 6461528, 6461533, 6461534, 5075622, 5241158, 5237016, 4932977, 5075633, 4683022, 4932971, 4921443, 4932969, 4956972, 5064791, 4921447, 4956961, 4956974, 6204104, 5854655, 6204076, 5854654, 5854694, 6204134, 6204121, 6204072, 5854600, 6204103, 6204101, 6883182, 6577947, 6572918, 6883184, 6572942, 6461521, 6577960, 6577934, 6577935, 6577937, 6577939, 6577944, 6700514, 6577940, 6883113, 6883116, 6883122, 6883125, 6883128, 6883130, 6883131, 6577929, 6634007, 6853131, 6461525, 6634002, 6572938, 6572900, 6634005, 6634060, 6577949, 6633998, 6634019, 6572907, 6572930, 6572923, 6633997, 6634061, 6634050, 5237022, 6204119, 5854677, 5854611, 5854563, 5854567, 5854621, 5854622, 5854623, 5854672, 5854632, 6204132, 6204123, 6204113, 5854594, 5854703, 6204133, 5854628, 5854587, 5854698, 6853139, 6634079, 6634080, 6634066, 6634058, 6572926, 6634011, 6577959, 6882521, 6882522, 6882525, 6882533, 6634042, 6700505, 6882534, 6634048, 6634049, 6634051, 6883189, 6883178, 6883188, 6883114, 6883132, 6883282, 6853113, 6853132, 6853103, 6853122, 6883153, 6883154, 6883136, 6883174, 6883145, 6883163, 6883273, 6634008, 6572906, 6634074, 6634057, 6633999, 6700499, 6634069, 6634006, 6577936, 6634053, 6634001, 6883281, 6634015, 6634013, 6883287, 6883159, 6577956, 5854612, 5854577, 6204112, 5854585, 5854688, 5854708, 6572920, 6883150, 6634016, 6572927, 6883170, 6883121, 6883173, 6853140, 7021852, 6883157, 6883166, 7021851, 6700501, 6577932, 6853109, 6883183, 6634078, 6572940, 6634024, 6883168, 6572905, 6853107, 6853129, 6700494, 6634071, 6204106;
c) Item Number 314-02-04, UDI/DI 10885862084491, Serial Numbers: 0744227, 0744208, 0744215, 0744239, 0797758, 0744248, 0744243, 0744202, 0797743, 0744197, 0797754, 0744199, 0744225, 0744216, 0744230, 0744200, 0744231, 0744203, 0744206, 0797756, 0797759, 0744223, 0797751, 0797766, 0797741, 0797763, 0797765, 1356686, 1257617, 1257663, 1257641, 1257654, 1257621, 1257627, 1257632, 1257650, 1257619, 1257656, 1356675, 1356682, 1356685, 1356687, 1394151, 1257616, 1257625, 1257626, 1257636, 1257640, 1257655, 1257662, 1257664, 0744233, 0744236, 0797739, 0744238, 0797740, 1423345, 1257651, 1433276, 1257628, 1423348, 1535182, 1257614, 1257645, 1404841, 1433275, 1450616, 1590188, 1590190, 1394143, 1394126, 1423352, 1523026, 1523027, 1394135, 1404852, 0797767, 0744212, 1450615, 1394131, 1404844, 1257638, 1535180, 1590173, 1535179, 1590189, 1423343, 1590196, 1590172, 1394137, 1404848, 1590176, 1590180, 1590194, 1590202, 1783107, 1697910, 1783105, 1765962, 1793700, 1757552, 1697893, 1783104, 1793709, 1697905, 1765960, 1765946, 1793682, 1697896, 1697906, 1697914, 1757557, 1757549, 1757554, 1757560, 1757562, 1765963, 1783098, 1697903, 1765943, 1765958, 1765934, 1783096, 1783106, 1765948, 1793687, 1793707, 1783100, 1793708, 1697899, 1697918, 1535172, 1394159, 1535177, 1257661, 1394152, 1257660, 1394155, 1590186, 1257649, 1590175, 1590199, 1394160, 1450613, 1257618, 1257637, 1394129, 1590187, 1703771, 1757548, 1765954, 1697909, 1933417, 1697907, 2000281, 1765942, 1933433, 1697908, 1765965, 1757544, 1793699, 1783097, 1765945, 1793680, 1935051, 1933432, 1793690, 1933434, 1933419, 1984480, 2000289, 1783101, 1935052, 1765956, 1933430, 1783092, 2117349, 2117347, 2117351, 2117366, 2117368, 1394141, 1590204, 1404849, 1257630, 1257659, 1404847, 1535186, 1590171, 1590206, 1433279, 1356684, 1535183, 1590192, 1423355, 1590183, 1433278, 1765953, 1984475, 1765964, 2000270, 1697919, 1935060, 2000274, 1933448, 1935058, 2000272, 1793694, 1783094, 1933410, 1933429, 2159870, 2258723, 2258739, 2258748, 2258754, 2258728, 2258742, 2258721, 2301149, 2301160, 2301161, 2301176, 2301151, 2301171, 2301180, 2469317, 2469324, 2301157, 2301169, 2536917, 2117357, 2258725, 2301172, 1394149, 1394121, 1404836, 1535178, 1423350, 1590207, 1404845, 1356674, 1590184, 1404838, 1697917, 1765929, 1765959, 2000287, 1697897, 1703774, 2469313, 2536897, 2506173, 2536894, 2117346, 2301152, 2301148, 2469308, 2506163, 2536913, 2536912, 2536918, 2506162, 2469315, 2506177, 2117365, 2557781, 2669720, 2669723, 2669738, 2669739, 2669741, 2669743, 2669746, 2751650, 2557771, 2557773, 2557774, 2557776, 2669733, 2669735, 2751640, 2751644, 2751645, 2751648, 2751635, 2751649, 2669718, 2669719, 2669729, 2669736, 2669737, 2779100, 2669716, 2669725, 2779092, 2779094, 2779095, 2779096, 2779099, 2779102, 2795325, 2795326, 2795327, 2795329, 2795330, 2795334, 2825703, 2795328, 2825695, 2795333, 2669726, 2669728, 2669734, 2669744, 2751636, 2779093, 2779097, 2779098, 2779101, 2779103, 2779104, 2795332, 2825717, 2669721, 2669740, 2871324, 2871326, 2871327, 2871328, 2871329, 2871330, 2871331, 2871332, 2871333, 2871334, 2669732, 2669727, 2669745, 2669731, 1697904, 1765940, 2301173, 2258731, 2536888, 2117370, 2825702, 2825699, 2825709, 2825696, 2795324, 2898377, 2898369, 2898370, 2898375, 2898395, 2958923, 2961292, 2958917, 2961287, 2980545, 2980556, 2958919, 3587091, 2961291, 2980546, 2958915, 2958920, 2958924, 2961286, 2825718, 1935054, 2159871, 2536908, 2536916, 3629140, 2557784, 3587097, 2898393, 2958916, 2958918, 2898372, 2871325, 2898378, 2889226, 3660464, 2958921, 3660471, 3660455, 3660460, 2825727, 4040084, 4040101, 4040103, 4091968, 4091979, 4091969, 4091986, 4092033, 4092036, 4092044, 4091973, 4092018, 3868746, 3875711, 3886333, 3886334, 3886336, 3886322, 3886323, 3950929, 3875710, 2301144, 2506178, 2258745, 3587083, 2889228, 4092022, 4368706, 4414851, 4414860, 4414868, 4414862, 4368707, 4368702, 4368720, 3974300, 3886330, 4475405, 2825722, 2889225, 3660463, 3660470, 2825692, 2825721, 2825704, 2961288, 4368721, 4091980, 4475416, 4475420, 3868744, 4092009, 4092001, 4040086, 3974304, 4475386, 4475410, 4475409, 4860551, 4837258, 4860548, 4932788, 4971276, 4971277, 5075603, 4368725, 3950914, 4040089, 4040092, 3868742, 5158835, 5158856, 4887770, 4932796, 5075613, 5158847, 4837241, 4887771, 4887778, 4887780, 5075957, 5158849, 5117327, 4837249, 4932782, 5137877, 5075955, 4837263, 4860549, 5158857, 5075610, 4887764, 4887776, 4091985, 3868743, 4932781, 4837246, 5117329, 5925031, 5925043, 6149706, 6149713, 6149714, 6149721, 5967383, 6005815, 6005817, 5967385, 6005811, 6005826, 6005834, 5925033, 5967374, 5967375, 6005812, 6149719, 5158852, 6005831, 6151187, 6149711, 6005832, 6149705, 6633463, 6633500, 6633470, 6147713, 6147718, 6147720, 6151184, 6151167, 6633458, 6633472, 6633479, 6645682, 6645712, 6645714, 6005829, 6147769, 6147709, 6633460, 6645680, 7023021, 7023008, 7023029, 6147772, 5967386, 6149707, 7023005, 6861112, 6645704, 7023027, 7023013 |
| FEI Number |
1038671
|
Recalling Firm/ Manufacturer |
Exactech, Inc.
2320 Nw 66th Ct
Gainesville
FL
32653-1630
|
| For Additional Information Contact | Thien Doan 352-377-1140 |
Manufacturer Reason for Recall | The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH). |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Beginning 03/06/2024, Exactech issued an URGENT MEDICAL DEVICE RECALL notice and an URGENT DEAR HEALTHCARE PROFESSIONAL COMMUNICATION to its consignees via email. The recall notice explained the issue, clinical impact, and requested the following actions be taken:
"" CAREFULLY REVIEW THIS RECALL NOTIFICATION to ensure that you fully understand the issue identified, the recall strategy, and all actions required.
" IMMEDIATELY IDENTIFY AND QUARANTINE any of the subject devices in your inventory and/or customer s inventory listed Product Scope Listing.
" EXTEND THE DESCRIPTION OF ISSUE AND CLINICAL IMPACT as described in the recall notification to your accounts that may have this product in their possession.
" COMPLETE AND RETURN the attached Recall Inventory Response Form to Exactech via email at recalls@exac.com.
The communication to healthcare providers referenced the FDA Safety Communication issued on 01/16/2024, explained the issue, potential health risks, and suggested patient monitoring for those with implanted affected devices for potential device wear, failure, or bone loss and should consider performing X-rays to further evaluate the patient and their implanted device if you suspect a failed device.
Contact your local Exactech Representative if you have any questions regarding this communication. |
| Quantity in Commerce | 6678 units |
| Distribution | Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = KWS
|
|
|
|