| Class 2 Device Recall Exactech Equinoxe |  |
Date Initiated by Firm | March 06, 2024 |
Date Posted | March 29, 2024 |
Recall Status1 |
Open3, Classified |
Recall Number | Z-1413-2024 |
Recall Event ID |
94092 |
510(K)Number | K113309 |
Product Classification |
Prosthesis, shoulder, semi-constrained, metal/polymer cemented - Product Code KWS
|
Product | Exactech Equinoxe CAGE GLENOID, CEMENTED, beta curvature, Mates with:
a) 47, 50, 53 head, Medium, Item Number 314-13-13,
b) 47, 50, 53 head, Large, Item Number 314-13-14,
c) 47, 50, 53 head, Extra Large, Item Number 314-13-15; Shoulder Arthroplasty |
Code Information |
a) Item Number 314-13-13, UDI/DI 10885862172693, Serial Numbers: 2204820, 2204812, 2204813, 2204815, 2452178, 2500820, 2499638, 2331072, 2427399, 2331064, 2204830, 2336229, 2427393, 2204831, 2204818, 2336244, 2336253, 2336241, 2427397, 2427387, 2499624, 2499632, 2500803, 2336258, 2336245, 2336257, 2406667, 2406673, 2427404, 2427392, 2460064, 2406653, 2499640, 2331078, 2336249, 2336235, 2204821, 2500818, 2204827, 2331062, 2204824, 2204822, 2331068, 2331074, 2406665, 2406664, 2406658, 2452182, 2406650, 2500804, 2452181, 2460057, 2406663, 2406649, 2499625, 2336254, 2427402, 2336230, 2204829, 2204825, 2204814, 2406656, 2336251, 2204816, 2406661, 2204819, 2331077, 2331080, 2331073, 2339033, 2499636, 2331076, 2336252, 2460058, 2406659, 2336240, 2336232, 2204817, 2460062, 2500813, 2665906, 2499647, 2452179, 2562206, 2499637, 2331075, 2562191, 2582335, 2562212, 2562209, 2665770, 2460060, 2500802, 2336256, 2622345, 2204828, 2622349, 2336236, 2406651, 2562189, 2665897, 2499633, 2499652, 2499629, 2582323, 2665785, 2499649, 2499626, 2639486, 2639498, 2427388, 2406646, 2562205, 2622361, 2665766, 2562203, 2609433, 2665778, 2452180, 2622342, 2665794, 2665903, 2665911, 2331061, 2427398, 2499630, 2622336, 2665925, 2665774, 2609430, 2609428, 2609432, 2460061, 2331063, 2500806, 2562186, 2622359, 2665791, 2500816, 2499648, 2460063, 2562202, 2582341, 2582310, 2665921, 2500811, 2427400, 2562207, 2406648, 2427403, 2582339, 2609426, 2609429, 2500819, 2665899, 2665908, 2665783, 2622348, 2331070, 2582332, 2665780, 2562214, 2562194, 2500821, 2460059, 2331079, 2582325, 2622339, 2639508, 2582318, 2665779, 2499627, 2562196, 2336250, 2582313, 2582316, 2499646, 2622356, 2582309, 2665930, 2562187, 2622363, 2639488, 2639482, 2639506, 2639497, 2639479, 2639500, 2406647, 2427395, 2427394, 2582333, 2500814, 2622360, 2665935, 2609427, 2639507, 2500798, 2331069, 2562199, 2562222, 2609425, 2665933, 2639509, 2622353, 2500815, 2582315, 2622338, 2499634, 2639481, 2406670, 2460055, 2499645, 2665768, 2499641, 2639503, 2622364, 2665907, 2562198, 2582343, 2622325, 2665917, 2582336, 2339036, 2562185, 2562195, 2665769, 2339032, 2622333, 2499628, 2582314, 2582311, 2622324, 2665776, 2562211, 2562218, 2582344, 2622362, 2582321, 2622343, 2622354, 2406660, 2622366, 2582342, 2639480, 2609424, 2562193, 2582324, 2406669, 2500799, 2665792, 2582330, 2665920, 2452183, 2639492, 2622337, 2622335, 2582331, 2336237, 2622328, 2336246, 2665781, 2406652, 2336233, 2336243, 2336247, 2622331, 2582326, 2665924, 2406655, 2331071, 2622341, 2609423, 2331065, 2582334, 2665902, 2406668, 2406672, 2500817, 2622367, 2582328, 2582317, 2622332, 2622369, 2665928, 2562204, 2562190, 2339034, 2665927, 2622344, 2460056, 2665913, 2639504, 2622334, 2562221, 2406674, 2665931, 2582319, 2665915, 2406662, 2562182, 2499635, 2665893, 2622357, 2499639, 2499623, 2499651, 2562217, 2582345, 2336238, 2639501, 2582322, 2665786, 2339035, 2562184, 2562183, 2406666, 2336231, 2499631, 2500812, 2500807, 2665793, 2499650, 2499642, 2812129, 2777436, 2698269, 2770473, 2700850, 2770480, 2689212, 2812169, 2812204, 2770472, 2777608, 2698262, 2698264, 2689216, 2777617, 2770494, 2770492, 2777611, 2689205, 2770497, 2812153, 2700845, 2770467, 2812155, 2812207, 2812149, 2812199, 2770477, 2770483, 2777457, 2812136, 2770482, 2770481, 2770493, 2812130, 2812133, 2812135, 2812160, 2812167, 2812134, 2812170, 2812118, 2812148, 2770460, 2812205, 2698273, 2812168, 2812146, 2812197, 2689210, 2777614, 2770490, 2770484, 2700848, 2812123, 2698278, 2777616, 2812200, 2812164, 2689199, 2812162, 2812141, 2812131, 2812152, 2700849, 2698276, 2812119, 2698274, 2689215, 2698266, 2812159, 2700846, 2777430, 2812124, 2812156, 2770485, 2812202, 2689214, 2698272, 2698265, 2700844, 2700851, 2770462, 2777441, 2689202, 2770496, 2770461, 2777435, 2777429, 2689207, 2700843, 2770479, 2777618, 2812196, 2812201, 2812213, 2812215, 2689200, 2689211, 2812194, 2562219, 2622350, 2665892, 2562188, 2665929, 2665901, 2562216, 2336234, 2639487, 2562215, 2665905, 2500808, 2622340, 2406671, 2622330, 2622351, 2665916, 2639485, 2639505, 2639495, 2562220, 2622370, 2639491, 2622327, 2500805, 2622352, 2500801, 2639484, 2499643, 2500809, 2622372, 2665894, 2622326, 2665891, 2582329, 2639493, 2427389, 2562201, 2665923, 2665926, 2665787, 2499644, 2609431, 2665895, 2427390, 2665919, 2639494, 2427396, 2562213, 2665914, 2665904, 2639502, 2500797, 2562200, 2500810, 2622368, 2639490, 2665889, 2582312, 2639483, 2899491, 2698279, 2777451, 2689209, 3693370, 3693368, 3693388, 2812117, 2899474, 2952006, 2968667, 2952009, 3636135, 2698263, 2979059, 2952027, 3636214, 2899520, 2777609, 3587596, 3610042, 2975773, 2979090, 2952082, 2967002, 2966997, 3593900, 3636174, 3636111, 2812198, 2899478, 2777431, 2968647, 2770468, 2812210, 3577196, 3636216, 2899482, 3593927, 2979061, 2952007, 2967003, 3593922, 2968661, 2968658, 3593925, 2952085, 2952023, 2952081, 2979055, 2975746, 2979057, 3577186, 3672145, 3672162, 3636222, 2777619, 2812147, 2899500, 2899498, 2777449, 2952004, 2952021, 2777439, 2952072, 2975743, 2966977, 2966985, 2979085, 3672157, 2812126, 3636150, 3593915, 3636164, 3587602, 3636124, 2968654, 2689201, 2952018, 2952003, 2979066, 2812165, 2975766, 3587603, 3693375, 2899509, 2899522, 2952087, 3636219, 2975748, 2975755, 3593920, 2899495, 2952025, 3636113, 2966980, 3593905, 2770476, 2966987, 3636154, 3636155, 3636127, 2952064, 3693363, 3587587, 3672159, 3636146, 2770489, 2698270, 2812192, 2968660, 3593917, 3636137, 3750089, 3796861, 3796843, 2952002, 2975735, 2899490, 3636130, 2966983, 3750056, 2967005, 2966999, 2966991, 2975762, 2979100, 2967006, 3672142, 3636163, 3693359, 3693371, 3750069, 2777434, 2952024, 2812166, 3577192, 2968635, 3587585, 3636145, 3672169, 3693357, 2812144, 2968653, 2899521, 2952014, 2770486, 2975734, 3636208, 2812138, 2899514, 2968641, 3587582, 2899505, 2698267, 2700847, 3577189, 3593930, 2770463, 2952065, 3636173, 2698277, 2952012, 2777446, 2777437, 2975738, 3577197, 2952073, 2979054, 2979062, 2975761, 3672139, 2777442, 2899501, 2968652, 2975757, 2952070, 2777426, 2968656, 3750062, 2966978, 2952084, 3672171, 2899484, 2952013, 3636138, 2777445, 2899504, 2975764, 2777448, 3593908, 2899489, 2899516, 3672179, 3693353, 3636206, 3636144, 3750052, 3750063, 2979092, 3636136, 3636203, 2770469, 2975732, 2979096, 2777444, 2979078, 3636209, 2812158, 2899503, 2899508, 2812116, 2966993, 2975765, 3577194, 2966994, 3610038, 3672143, 2812120, 3593929, 3672161, 3693345, 3636225, 3693346, 3750078, 3672141, 3672135, 3636162, 3636172, 3693373, 3796888, 3587592, 2770498, 2979094, 3636212, 2979103, 3750060, 2899487, 2975740, 2812157, 3610044, 2777454, 3636115, 2979088, 3610048, 2777607, 2899499, 2777440, 2899481, 2899510, 2899480, 2952005, 2952067, 2975745, 2979087, 2770470, 3636151, 3610032, 3636117, 3672151, 3693358, 2899519, 3636159, 3693389, 3750057, 3750088, 3587583, 2966996, 2777438, 2975744, 2777453, 2979072, 2979071, 3587599, 2975774, 2968663, 3636118, 2979079, 2966988, 3672158, 3636153, 3672166, 3750071, 3636120, 2966986, 2975733, 2968636, 2812114, 2979084, 3593909, 2968637, 2777447, 2899496, 2979098, 2968651, 2952076, 2952091, 2899515, 3636161, 2770466, 3610055, 3610040, 3593931, 2899512, 2966979, 2812151, 2968650, 3796853, 2952031, 2952020, 2952078, 3593901, 3636125, 3593910, 3587590, 3593912, 3636217, 3636169, 2777459, 2899513, 3593916, 2812209, 2975770, 2698271, 2979081, 2979075, 3587593, 2968648, 2967000, 2899497, 2899492, 3750084, 3593904, 2952026, 2979069, 3672148, 2899506, 2770495, 2968655, 3636106, 2812122, 2975772, 2952086, 2966992, 2899485, 2966981, 3636139, 3610049, 2952075, 2770491, 2979063, 3577200, 3636211, 2952066, 2975737, 2777433, 3636221, 2979097, 2812111, 2975780, 3593898, 2777452, 2770464, 3693344, 2899507, 2975767, 2968666, 2968662, 2968657, 3672180, 3636220, 3636218, 2812143, 3610041, 3636226, 3636141, 2777432, 2689206, 2698268, 3610046, 2689208, 2975747, 2968644, 2812145, 2899473, 3672175, 3693343, 2777427, 2899479, 2975754, 3636160, 2812113, 3577190, 2770475, 2812206, 2975778, 2979093, 2952068, 2979068, 2812211, 2812139, 2777450, 3610034, 3636112, 2975749, 2979102, 3593906, 2975758, 3672156, 2979091, 2770487, 2812137, 2952010, 2968659, 2689213, 2689217, 2812115, 2899476, 2952011, 2952030, 2952017, 2975739, 2968642, 2975763, 2975742, 2979083, 3577199, 2812208, 2812125, 3610031, 2812195, 2899477, 2952008, 3693342, 3587594, 2899486, 2968645, 2812140, 2812142, 2899511, 2952028, 2899517, 2777612, 3636122, 3672170, 3636158, 3750058, 2952029, 2899502, 2899518, 2975768, 3750048, 2770471, 2952071, 2812112, 2952079, 3610033, 2689204, 2979089, 3672160, 2979076, 2777455, 3577198, 3636175, 3593911, 3577195, 3693391, 3750067, 2777615, 2975750, 2979095, 3636140, 3672174, 3636148, 3796887, 3636223, 3672173, 3577193, 3636224, 3672152, 3693380, 2975753, 2979099, 3672164, 3693347, 3693379, 2975760, 3693367, 2665767, 2622358, 2665890, 2639496, 2665777, 2639510, 2562223, 2622365, 2562210, 2562192, 2562197, 2204823, 2582327, 2500800, 2665898, 3593923, 3593926, 3796841, 3892410, 3610045, 3892428, 3892429, 4059221, 2952015, 3823780, 3823814, 3880194, 3892409, 3750043, 3693383, 2812163, 3832684, 3823787, 3750085, 3750075, 3832663, 2777428, 3587584, 3880174, 4018029, 3985503, 3913009, 4049247, 4099787, 3610043, 3796860, 3587588, 3913013, 4005395, 4054509, 3750051, 3985516, 3823786, 2770465, 2698280, 2968638, 2975759, 3823800, 4193460, 3587605, 3832656, 2968668, 3913017, 3913039, 3880186, 2812127, 4005397, 4005382, 4091557, 4054505, 4141968, 4153715, 4155651, 4153714, 2979073, 4059220, 3636132, 3636142, 4150414, 3880129, 3832678, 3636114, 3636119, 3985512, 3587586, 3880199, 3750046, 3880180, 3892421, 3796889, 3610039, 3880162, 3880132, 4068813, 4091555, 3832652, 4020688, 3587600, 3880110, 3832664, 3880125, 3880176, 3892388, 3880156, 3880190, 4128060, 4150413, 3880113, 3636213, 2975779, 3672177, 3880160, 3880165, 3892415, 3892390, 2975752, 3796863, 4075933, 2966984, 3823789, 3750074, 3880188, 2975741, 3750061, 3636157, 3913026, 3593919, 3823783, 3832648, 3636147, 3880122, 3832685, 3880123, 3892408, 3892403, 3832649, 3880128, 3892430, 3880106, 3913010, 3880155, 3985511, 3892419, 3913035, 3880171, 4020689, 4020685, 4020677, 4049243, 4153712, 4091565, 4182764, 3880119, 4005385, 4155664, 2975781, 3892397, 3823846, 2968646, 3796855, 3823802, 3832653, 3823831, 2979077, 3636166, 4091560, 4103636, 4020693, 3843217, 2777425, 3823776, 3796859, 3832674, 3880109, 3880121, 4121991, 4092807, 3796867, 3796865, 3823804, 3823849, 3823837, 3880120, 3796874, 3832671, 4005391, 3823828, 3985496, 3796882, 3913001, 2979056, 2812212, 3672140, 3892392, 3892435, 3832680, 3750081, 3832645, 4150428, 2899488, 2777456, 3832691, 3587604, 4035622, 4121999, 3913011, 4091551, 2968664, 4128051, 4097241, 4150418, 2899493, 3880172, 3672144, 4128064, 3985514, 3587589, 3672137, 3672147, 3593913, 2966990, 3913015, 3672181, 3832668, 3796856, 3892420, 2968640, 3636165, 3823840, 3832665, 2899483, 3892391, 3913020, 3892407, 4150409, 3672155, 4091552, 3672133, 3985502, 2812154, 2975736, 3880150, 3796864, 3913005, 3912992, 4020684, 4155662, 4182757, 3610050, 4097244, 3823779, 3880201, 3636156, 3693385, 3593903, 3636215, 3796858, 3796844, 3823791, 2975777, 2952077, 3823784, 3610052, 3880179, 3912994, 3880111, 3750055, 3880173, 3985495, 4005381, 4005387, 4035624, 4020682, 4063701, 3892418, 4121998, 4121986, 4091561, 4150416, 4122000, 4193444, 3796842, 2812171, 3880168, 2777613, 3693382, 3985504, 4068809, 4005393, 4182768, 4059223, 3892433, 3796881, 3636123, 3636107, 3823782, 3892404, 3593914, 3610054, 3892432, 3693348, 3796883, 3593921, 3823827, 3985493, 3636205, 3796885, 3823799, 4005383, 4049240, 4005380, 3796872, 3796886, 3832672, 3823810, 3832661, 3880192, 3892414, 3892394, 3823824, 4049252, 4091556, 4020678, 4150432, 4155660, 3796884, 2952083, 3892387, 3636108, 3636105, 3672167, 3823848, 3823823, 3880167, 3985501, 4020674, 4018028, 4128047, 4141974, 2975769, 3912995, 3693349, 3823820, 3693364, 3823839, 2812161, 4005384, 4020672, 3880143, 4099790, 3636116, 4018027, 4075946, 4122001.2952022, 3880182, 2979082, 3880198, 4068812, 4121987, 4155650, 3796866, 4075931, 3880159, 2812150, 3672132, 3823826, 3750054, 3636167, 3823772, 3892396, 3843215, 3892398, 4049251, 3823832, 3750066, 4005392, 3672163, 3823794, 3693355, 3693356, 3796846, 4150424, 4091559, 4020671, 4054506, 2966995, 4049239, 3693372, 3750082, 2952088, 3880178, 3823830, 3823775, 3880195, 3750086, 4059217, 4182774, 3587601, 4054510, 2952074, 2975771, 2966998, 4091558, 4141961, 4165418, 3832683, 2979074, 3750073, 3672176, 2952063, 3913030, 4075944, 4165415, 4035623, 4155671, 3832646, 4066127, 2777458, 3672178, 3843214, 2698275, 3796876, 3892411, 3880127, 3913023, 3892427, 3750053, 3880189, 3892395, 3750065, 2979058, 2952089, 3672172, 3823798, 3796851, 2952080, 3593918, 3823822, 3823821, 3593902, 4091563, 3750044, 3693374, 3880138, 3892436, 3913028, 3913019, 3823836, 3796868, 4091818, 3610036, 3577191, 3796854, 3823844, 3587606, 3587595, 3750072, 3880114, 4075939, 3880112, 2979065, 2966982, 3823806, 3832655, 3672146, 3750080, 3925470, 3925472, 3832659, 3750045, 3832679, 3880196, 3880204, 3693362, 3864536, 3925471, 3864534, 3925474, 3880117, 4103637, 4091554, 3880115, 3864535, 3880193, 4153711, 3610051, 3796880, 3880108, 3636143, 3880145, 3672165, 2975756, 3880126, 3636170, 3985497, 4075943, 3672150, 3913025, 3636104, 3593897, 2975775, 3823818, 3832651, 4075930, 4075940, 3672168, 3823847, 3880177, 4005379, 3750087, 4099786, 3823809, 4155663, 3823834, 2979086, 2975776, 3832686, 3593928, 3693350, 3750083, 3796870, 3843213, 3913032, 3880181, 3892412, 3892401, 3892406, 3912991, 4075936, 3913012, 4054508, 4150425, 4182770, 3593907, 3832677, 4049245, 3672149, 4075937, 3832676, 3823790, 3880157, 3823813, 3913000, 3985500, 3750059, 3672134, 3823816, 3823843, 3913021, 3823817, 2770478, 2770488, 3796878, 3892434, 2812128, 3823801, 4005398, 4150410, 3587591, 3636109, 3796873, 3880130, 3796871, 3880163, 2952090, 3913003, 3912996, 3832689, 4020670, 4020676, 4049238, 3913036, 3880200, 3985505, 3985507, 4150427, 3750050, 3832669, 3832660, 3880124, 3880140, 2968669, 3672136, 3636171, 3832650, 3880146, 2967001, 3880131, 3823812, 3832693, 3832690, 3832673, 3913033, 3832682, 3823808, 2812132, 2952069, 2975751, 4155658, 3892425, 3693351, 3693387, 3636133, 3672154, 3693366, 3750042, 3693352, 3913004, 3823842, 3832647, 3832687, 3892423, 3823778, 3636177, 3636128, 3823785, 4063696, 4063698, 3796845, 3823773, 3823774, 3823777, 3823797, 3823805, 3823811, 3892417, 3880154, 3636121, 3913031, 3832670, 3693369, 3832667, 3913008, 3636131, 3636129, 3672153, 3577187, 3832654, 4160578, 3693354, 4078841, 3693390, 4160577, 2582337, 2406654, 2622371, 2582320, 2665772, 2452177, 2665922, 2562208, 2665909, 2622347, 4193461, 3636168, 4172913, 4332576, 3823792, 4228108, 4332452, 4172915, 3750090, 4141976, 4128049, 4224153, 4121990, 4224156, 4128056, 4509219, 4018030, 4091564, 4266644, 4339594, 4339562, 4509276, 4428856, 4103634, 4020687, 4250686, 4193456, 4005394, 4122008, 4339570, 4403162, 4428898, 4059219, 4266655, 4054507, 4182751, 3985513, 4298075, 4332595, 4339563, 4216274, 4339617, 4266662, 4059218, 4266627, 4332591, 4439565, 4141978, 4155659, 4428904, 3985509, 4509208, 4128059, 4238963, 4256945, 4049254, 4281088, 4155669, 4403178, 4250693, 4332441, 4403182, 4216277, 4165416, 4141962, 4298083, 3796852, 4298087, 4128048, 3832692, 4216267, 4428851, 4428893, 4020692, 4155674, 4216269, 4332470, 4141960, 4439563, 3823819, 3796849, 4005388, 4121997, 4193453, 4208655, 4193467, 4250689, 4266621, 4281080, 4266635, 4208656, 4332471, 4332578, 4216275, 4339628, 4339630, 4403173, 4428900, 4472703, 4339583, 4509302, 4122005, 4509296, 4242466, 4193466, 3823829, 4049246, 4141971, 4266624, 4281084, 4339625, 4298084, 4403171, 4256930, 4472687, 4208649, 2698261, 3796857, 3587598, 4059224, 4155657, 4281070, 4339603, 4428846, 4509275, 4403167, 4141975, 4190762, 4266632, 4216270, 4128055, 4332596, 4266641, 4238962, 4439573, 4509212, 4509213, 4242471, 4228115, 4155654, 4428847, 4005390, 3913027, 3693360, 4266654, 4439581, 3587597, 4256932, 4141965, 4193442, 4224322, 4150408, 3985506, 4332618, 4281079, 4035620, 4256941, 4182771, 4122002, 4509267, 4228103, 4509261, 4224323, 4121994, 4509307, 4332606, 4193449, 4250687, 4266651, 4298072, 4509262, 4128054, 4242463, 4439574, 4141977, 4332435, 4097243, 4250685, 4332457, 4339612, 4403174, 4121996, 3880166, 4256925, 4332466, 4121995, 4332468, 3832694, 4339623, 3636176, 4332442, 4224155, 4099789, 4472699, 4428849, 3636110, 3796869, 3913002, 3880187, 4049248, 4049249, 4228101, 4224327, 4128063, 3880205, 2968639, 4266639, 3796848, 4238961, 4281064, 4339568, 4472702, 4472698, 4281061, 4266636, 4224325, 4256933, 4150430, 4428908, 3985498, 4141973, 4228110, 4266664, 4005386, 4339611, 4256940, 4155665, 4128065, 4266647, 4332580, 4439567, 4141969, 4266625, 4216278, 4428842, 4208657, 4509216, 4121989, 4266634, 4332598, 4332445, 4509303, 3636210, 4238957, 4238956, 3880148, 4128052, 4339615, 3880147, 4155667, 4078840, 3750049, 4238952, 3892413, 4509227, 4108504, 4193451, 2968665, 4216263, 4216273, 4224320, 4332448, 4332439, 4339609, 4266660, 4339618, 4428896, 4428902, 4472688, 4068810, 4509215, 4332588, 3912998, 4185971, 3912993, 4256934, 4428906, 4216272, 4238950, 3913007, 4228113, 4332585, 4256935, 3823788, 4242469, 3610035, 4005396, 4428857, 3750047, 4091550, 4256931, 4141967, 3593932, 4005378, 4332589, 4332610, 4298076, 4128057, 4165419, 4020691, 4172917, 4020683, 4228099, 4266629, 3913034, 4332447, 4150433, 4250692, 4266645, 4228107, 4509312, 4185968, 3636204, 4190760, 3610053, 4108499, 4141979, 4182766, 2777443, 4256942, 4216264, 4141972, 4256922, 4298067, 4403181, 4472694, 4428911, 4339577, 4339558, 4155668, 4339584, 4298068, 4190761, 4150417, 4228112, 4256943, 4266657, 4256936, 4298071, 4128062, 4298070, 4339629, 3880175, 4128061, 4339559, 4281066, 4332459, 4153716, 4020673, 4020681, 3892422, 4332581, 4332621, 3880152, 4428912, 4224154, 4150421, 4224329, 4208652, 4097242, 4228104, 4281074, 4339557, 4266640, 4182765, 4182758, 4266626, 4332586, 4403176, 4266666, 4332458, 4091819, 3985508, 4128058, 2777610, 4403164, 2952062, 4075942, 4256918, 4182760, 4182753, 4250688, 3823815, 4238953, 4509277, 4439580, 3880107, 4121992, 4122004, 4266628, 4185969, 4182763, 4059222, 3913029, 3913022, 4193448, 4339626, 4193465, 4332463, 4339560, 4242472, 4428903, 4150419, 4428841, 3796847, 4182754, 4281076, 4208650, 3750076, 4339601, 4332573, 4509280, 4075941, 4193450, 4266619, 4108498, 4339622, 4339598, 4403165, 4428860, 4472689, 4099788, 3832657, 4193463, 3610037, 4509264, 4439570, 4128053, 4216268, 4332615, 4128046, 3593924, 4509268, 4182769, 4165420, 4281059, 4049253, 3892400, 4281057, 4281083, 4472701, 2979060, 4165417, 4332444, 3750070, 4020690, 3880164, 4298079, 4155655, 4103633, 3832658, 4509317, 4298074, 4155670, 3832681, 4339561, 4049244, 4339569, 2966989, 4266630, 4266623, 4182762, 4509273, 4472685, 4155647, 3593899, 4281071, 4281082, 4266656, 4224328, 4339585, 4041844, 4172906, 4216276, 4472705, 3864532, 4428894, 4403168, 4075945, 4281073, 4193469, 3880135, 4332438, 4332605, 4428844, 4428859, 3864533, 4172907, 3864538, 3636126, 4224326, 4122481, 4041848, 4256923, 4339610, 4472697, 4256938, 4108495, 4182752, 4298081, 4153713, 4332601, 4339590, 4091566, 4108493, 4020680, 4193455, 4182772, 4266661, 4155652, 4332603, 4339596, 4266650, 4403161, 4165413, 4509306, 4150407, 4153719, 3880149, 4332620, 4049241, 4193454, 4150412, 4165412, 4193462, 4250680, 4266663, 4332464, 4332572, 4339588, 4428890, 4128044, 4165414, 3750068, 4339589, 4256939, 3880118, 4238964, 3880185, 4075938, 4403166, 4238948, 4332465, 4403175, 4216271, 4228105, 3672138, 3636152, 4155653, 4339597, 4332593, 3892393, 4224151, 4339575, 3913016, 4228114, 4193457, 3913024, 4193464, 4238951, 3880203, 4403179, 4339602, 4216266, 4266642, 4128045, 4103635, 4339586, 4035619, 4224321, 4509311, 4509297, 4128043, 4428864, 3892416, 3636207, 3880116, 4339571, 4150423, 4332611, 4298078, 4228098, 4108497, 4238949, 4153717, 4250682, 4281065, 4339607, 4339606, 3985515, 3880197, 4472693, 4128050, 4332454, 2979067, 4208654, 4256917, 4256926, 4298066, 4509272, 4266617, 4193441, 4150426, 3823845, 3577188, 4281060, 4193445, 3880151, 4250683, 4266658, 4266659, 3823771, 4281085, 4108494, 3880191, 4122009, 4428840, 4242464, 4242468, 4242470, 4298069, 4298080, 4332619, 4509206, 4428909, 4509224, 4339613, 4339573, 4509298, 4250684, 4250694, 4428854, 4428862, 4428863, 4332574, 4228106, 3912999, 3693361, 4281075, 4228111, 4228102, 4228116, 4155673, 4150441, 3880161, 4172920, 3693378, 3880169, 4224157, 4627272, 4627287, 4627259, 4627258, 4644943, 4627225, 4627217, 4627266, 4627246, 4637050, 4644915, 4627276, 4644931, 4627212, 4627216, 4627257, 4644902, 4666407, 4666459, 4644936, 4644910, 4644934, 4644929, 4644896, 4627265, 4644926, 4627221, 4627208, 4627239, 4627245, 4627294, 4627209, 4644920, 4637052, 4627286, 4644917, 4627270, 4627281, 4666393, 4644895, 2622346, 2665888, 2665912, 4228109, 4224160, 4172911, 4172912, 4332600, 4439579, 4141964, 4063702, 4339564, 4121988, 2812121, 4281062, 4281081, 3880137, 4208648, 4509226, 4509300, 4172919, 4332599, 4332577, 4332575, 4108502, 4339555, 4439575, 4122007, 2968643, 3796862, 4020686, 4242467, 4339608, 4339600, 3750079, 4339567, 2968649, 4472692, 4208651, 4439568, 4509217, 4509210, 4266620, 4403184, 2979080, 4509308, 4242465, 4266618, 4332607, 4266653, 4298085, 4332614, 4193468, 3832666, 2770474, 4256924, 4266646, 4472706, 4472707, 3823835, 3823781, 4332602, 4509260, 4339614, 4063700, 4339566, 4332434, 3880158, 4428907, 4155675, 4266649, 4339572, 3796877, 4509225, 4049242, 4428905, 3880183, 4155656, 3796890, 4238965, 4332579, 4339591, 4332436, 4332613, 4281067, 4141966, 4224158, 4403172, 4298082, 4439564, 4509316, 4509318, 4155649, 4472704, 4266637, 4256946, 2979064, 4332455, 4281086, 4332433, 4339578, 4509279, 3832688, 3880142, 4339624, 4256928, 4238946, 4256912, 4403163, 3823825, 4332584, 4339616, 4472691, 4332592, 4509263, 4509223, 2979101, 4281078, 4509314, 4332467, 4035627, 4193440, 4509274, 4216279, 4332460, 4332590, 4172916, 3880184, 4193443, 2952016, 4339587, 3610047, 4121985, 2952019, 2812214, 4141959, 4238955, 4141963, 4428914, 4339592, 4428861, 4428913, 4298086, 4150415, 4256920, 4509313, 4403185, 4339579, 4266643, 4339574, 4332587, 4439578, 4509257, 4256921, 4256937, 3892424, 4298073, 3750064, 4238958, 3913014, 4403169, 4509301, 4439569, 4428853, 4150422, 4108500, 4281058, 4035630, 4035625, 4041849, 4472690, 4509310, 4428850, 4041845, 4428858, 4472695, 4185972, 4075934, 4428899, 4509207, 4509221, 4121993, 4332451, 4339599, 4266652, 4332472, 4509214, 3985492, 3823841, 4332432, 3892431, 4509281, 4155672, 3913038, 4122003, 4332453, 4472696, 4332608, 4509220, 4049250, 4332582, 4509266, 4428897, 4509209, 4150411, 4509218, 4256919, 4035629, 4332604, 4332437, 4439571, 4298077, 4509204, 4332583, 3880139, 3880144, 4193446, 4256929, 4228097, 4182759, 4172921, 4216281, 4428895, 4428910, 4193459, 3823796, 4228100, 3693381, 4224159, 3693376, 3880141, 4753251, 4627316, 4676827, 4644945, 4744366, 4757009, 4808038, 4666395, 4975662, 4808037, 4627253, 4627315, 4644909, 4897923, 4925818, 4795024, 4627204, 4771653, 4744364, 4925831, 4771645, 4808036, 4888753, 4808040, 5022974, 4771646, 4795030, 4896554, 4676822, 4627236, 4637065, 4637053, 4744358, 4925828, 4666413, 4627238, 4666408, 4753254, 4753248, 4897930, 4627244, 4627256, 4627279, 4756987, 4627291, 4811661, 4936475, 4627255, 4756986, 4757008, 4627321, 4637064, 4666439, 4753258, 4666463, 4771651, 4955395, 4744365, 4777183, 4771649, 5037458, 4888730, 4644925, 4848275, 4848276, 4896550, 4808044, 4897919, 4637067, 4897925, 4997538, 4808039, 5037436, 4627303, 4666441, 4644932, 4627318, 4690953, 4690957, 4753240, 4666461, 4811678, 4777203, 4666452, 4888749, 4848273, 4861763, 4897914, 4925816, 4997536, 4975661, 4861761, 5022972, 4896539, 5037451, 4777207, 4997544, 5022987, 5102766, 4936470, 4757002, 4627297, 4861773, 4897913, 4627248, 4627242, 4888736, 5120341, 5120332, 4666406, 4627229, 4644908, 4753242, 4644933, 4627288, 4753260, 4771642, 4666394, 4666401, 4771650, 4808057, 5161712, 5156924, 4690961, 4955398, 4896556, 4771656, 4795027, 4897927, 4637054, 4676825, 4666454, 4777212, 4896538, 4627305, 4627298, 4832828, 4666399, 4644922, 4753246, 5120337, 4676821, 4753245, 4861772, 5037442, 4832812, 5022978, 4627314, 4644923, 4627260, 4975651, 4997539, 4848264, 4744369, 4666446, 4811663, 4644918, 4753243, 4637063, 4627301, 4644916, 4744362, 4795043, 4808056, 5037460, 4771665, 4811674, 4888743, 4888735, 4744361, 4848269, 4975636, 4808046, 4666460, 4997547, 5022960, 4627218, 4744359, 4848279, 4975642, 4925817, 4897934, 4627326, 4690948, 4896560, 4861760, 4811660, 4756997, 4627284, 4666414, 4771663, 4644906, 4627241, 4744350, 4777210, 4832820, 5102770, 4808045, 4666412, 4690970, 4627302, 4757007, 4897911, 4753239, 4975640, 4666402, 4997532, 4627295, 4861764, 4644952, 4627274, 4811670, 5102753, 4644899, 4644937, 4666448, 4811668, 4811681, 4795023, 4757001, 4627254, 4627290, 4888734, 5037443, 5102772, 4690964, 4744374, 4753249, 4666456, 5037449, 4925802, 4666462, 4777206, 4612906, 4811656, 4832819, 4627247, 4627205, 4666392, 4832824, 4808048, 4861749, 4896563, 4627311, 4975643, 4627299, 4676811, 4690971, 4644935, 4811679, 4627309, 4936454, 4771657, 4811671, 4955391, 4936474, 4795037, 4777198, 4627324, 4795029, 4690952, 4896545, 4861771, 4936465, 4925808, 4753255, 4848262, 4975638, 4888733, 4753252, 4811657, 4644907, 4666453, 4848280, 5022959, 4955393, 4848274, 5037439, 4896546, 4848268, 5037455, 5102771, 4690958, 4676815, 4690950, 4676810, 4777184, 4861762, 4644949, 4848271, 4975655, 4676809, 4832813, 4627283, 4771664, 4666396, 4777190, 4627206, 4888747, 4861750, 4771660, 5102761, 4627307, 4925813, 4627308, 4690954, 4861759, 4861746, 4795020, 4637056, 4811662, 4627214, 4690949, 4861770, 4676820, 4666440, 4644954, 4771668, 4997537, 4848267, 4777200, 4644897, 4897929, 4975644, 4676819, 4644900, 4771667, 4771670, 4811669, 4888742, 4627232, 4896552, 4637062, 4666410, 4896559, 4897917, 5022969, 4975652, 4666400, 4777194, 5022189, 4627312, 4627319, 4753256, 4896555, 4997527, 5022980, 4771658, 4777201, 4666405, 4627222, 4644924, 4690956, 4644942, 4644928, 4627211, 4676829, 4744367, 4861755, 4848261, 4936449, 4627215, 4627243, 4627292, 4897931, 4888738, 4627203, 4627219, 4676808, 4627252, 4756991, 4936453, 4627304, 5022983, 4690951, 4676828, 4811680, 4848270, 4975641, 4612907, 4637060, 4744371, 4975665, 4896558, 5157256, 4627226, 4627207, 5037435, 4771659, 4644950, 4644944, 4771655, 4777199, 4811664, 4795018, 4896565, 4811672, 5037440, 4627220, 4637066, 4666391, 4690959, 4690962, 4771643, 4777195, 4861754, 4861766, 4666457, 4811655, 4666409, 4627249, 4896549, 4644904, 4644941, 4925829, 5022979, 4896561, 4744355, 4897922, 4925812, 4997524, 4777204, 4666444, 4925805, 4896542, 4756988, 4897915, 4627227, 4627228, 4771641, 4832817, 4832818, 4811653, 4925815, 4644898, 4627235, 4795031, 4808050, 4690966, 4795025, 4690947, 4757005, 4676823, 4777211, 4627231, 4627213, 4771661, 4644930, 4811665, 4666404, 4795034, 4753261, 4888740, 4644951, 4795026, 4753241, 4795040, 4897933, 4896544, 4897926, 4936473, 4936461, 4997540, 4637057, 4676814, 4676830, 4753247, 4753238, 4888746, 4666411, 4795036, 4744370, 4888739, 5157241, 5037459, 4666445, 4676812, 4757006, 4888725, 4795044, 5102765, 4861756, 4997545, 4676807, 4832816, 5037454, 5120340, 4808031, 4676817, 4832836, 4848265, 4690955, 4848266, 4666443, 4339619, 3892426, 4063703, 4250690, 4182755, 4208653, 3880134, 4108496, 3823833, 4509295, 4238967, 4063695, 4238947, 3912997, 4439566, 4005377, 4332461, 4256913, 4428848, 4266648, 4509299, 4332616, 3880153, 4266622, 4332450, 4403177, 4509259, 4238966, 4439577, 4332597, 4224324, 4020679, 4332617, 2812193, 4403180, 4428855, 4150431, 4182761, 4509305, 4509271, 4339582, 3796879, 4509304, 3864537, 4041846, 4035621, 4153718, 4332449, 4339581, 4108503, 4193452, 4428843, 3750077, 4005389, 3985494, 4091562, 4472684, 4332594, 4020675, 4509211, 4403170, 4509265, 4439576, 4339621, 3693365, 4472700, 4339627, 4160579, 3796850, 4936469, 5102751, 5362653, 5102749, 4832834, 5204831, 4757003, 4936447, 5303558, 5301488, 5453510, 5321739, 5303535, 4627223, 5364460, 5342637, 5161724, 4955399, 4744351, 5161705, 5157250, 5022981, 5157229, 5364439, 5431124, 5022986, 5389703, 5364453, 5389696, 4888728, 4936464, 4744373, 5156920, 5364452, 5342659, 4832814, 4808042, 5321736, 5321714, 5157233, 4888732, 5453489, 4777189, 5303556, 5342658, 4756985, 4777185, 4744356, 5453503, 4997529, 5432825, 5156926, 4888737, 5178574, 4897937, 4888751, 5120317, 5303555, 5161717, 5451826, 5431125, 5464611, 4627210, 4832829, 5161727, 5037452, 5364434, 5022967, 4832830, 5157249, 4997548, 4888741, 5226292, 5321716, 5321738, 5342677, 5389694, 5204830, 5211313, 5342654, 5226291, 5226280, 5453483, 5364441, 4861745, 4861769, 4861774, 5178570, 4848281, 5161714, 4861758, 5178566, 5321701, 5022202, 5321908, 5037444, 5037433, 5161721, 5120338, 4997549, 5102759, 4896548, 5157254, 5453491, 5432835, 5161703, 4777205, 5204812, 5120326, 5321915, 5178552, 5321926, 5321730, 5335916, 5453506, 5342689, 5342609, 5389700, 5204806, 5204833, 5022211, 5389698, 4832823, 5204807, 5211301, 5102776, 5156919, 5321709, 4744372, 5022973, 4925820, 4925822, 5342681, 5342692, 5211308, 5161716, 5178557, 5178569, 5321719, 5211320, 5321914, 4777191, 5157245, 4627261, 4888731, 5432836, 5453485, 5301510, 5362658, 5301492, 4795046, 4997543, 5157235, 5301507, 5301500, 5464596, 5156903, 4975649, 5301501, 5157234, 5389712, 4848258, 4744353, 5362670, 4644905, 5301486, 5453484, 5453499, 5531805, 5157247, 5226277, 4676813, 5321899, 4627251, 5389714, 5342634, 5342639, 5022214, 4795035, 4666449, 5120330, 4975639, 5226295, 4627237, 5364443, 5321920, 4808047, 4666458, 4808030, 5022193, 5226298, 5037434, 5161707, 4627234, 4896562, 5531806, 5211300, 5362660, 5464602, 5451832, 5362671, 5157244, 5431126, 5342627, 5451824, 5342670, 5102764, 4832837, 4627262, 4997533, 4936460, 4808058, 5464593, 4756998, 4955392, 4666398, 5157236, 4897939, 4811659, 5022965, 5321933, 5321905, 4936463, 5226300, 5022217, 5022976, 5451831, 5321901, 5178562, 5364449, 4753262, 5421526, 5022971, 5120323, 5321735, 5342638, 5120336, 4925807, 4777197, 5156916, 5389710, 5022975, 5342617, 5022961, 5022201, 5022205, 5204815, 5157246, 5303557, 5335920, 5342662, 5342614, 5157238, 4690969, 4888744, 5178555, 5226285, 4997522, 5303549, 5022213, 5335910, 5335903, 5421519, 5464613, 5022206, 5335897, 5335902, 5211306, 5342653, 4997526, 4861747, 5037438, 5321703, 5321731, 5156906, 5102758, 5161725, 4848272, 5421542, 4888750, 5211328, 5362673, 4753250, 4832832, 5102763, 5157242, 4811682, 5421533, 4955401, 5178577, 5453488, 5022195, 4690968, 5156917, 5178564, 4808033, 5303550, 5321712, 5335922, 5362663, 5362655, 5421524, 5432830, 5451827, 5431112, 4637055, 4897935, 4627313, 5321733, 5178559, 5421543, 5342625, 4756999, 5037441, 5022191, 5157239, 5178578, 5204820, 5211314, 4848257, 5303534, 5303547, 4897921, 4936458, 5204810, 5431113, 4936471, 5120331, 4848260, 5321912, 5120342, 5453512, 4936448, 5156907, 5431120, 5161710, 4627278, 4936457, 5451840, 4896551, 5211317, 5342690, 5102752, 5531797, 4888727, 5531817, 5531814, 5342655, 4955396, 5178563, 5321932, 5022963, 5022210, 5301514, 4896537, 5335905, 5421529, 5120333, 5362667, 5301489, 5342591, 5335923, 5362665, 5022985, 4896543, 5226296, 4795045, 5301512, 5303541, 5531815, 5161713, 4777208, 5204827, 5226284, 5342622, 5303544, 4936452, 5321711, 5301504, 5453487, 5362657, 5102773, 4744360, 5022207, 5156910, 4955394, 4861752, 4832809, 5321929, 5364438, 5321931, 5421523, 5531807, 4644948, 5211305, 5037450, 5421545, 4666450, 5161709, 4997541, 4744357, 4637061, 4666403, 4627280, 4795047, 4795042, 5362656, 4795041, 5362661, 5451825, 5120328, 5204825, 5022209, 5335918, 5178579, 5226283, 5022984, 5362669, 4997542, 5432818, 5389705, 4861757, 4936455, 5037437, 5431114, 5464608, 5421536, 5211322, 5211327, 5022966, 4888726, 5389691, 4753244, 4861751, 5321917, 4811677, 5120327, 4777188, 4997531, 5178556, 4925825, 5362674, 5432829, 4808051, 5451841, 4666397, 4848284, 5102760, 5321723, 4832835, 5451839, 5321710, 5421522, 4753253, 5431119, 5321732, 4997523, 5226287, 5157243, 5156901, 5156912, 5157232, 5156902, 5321717, 4936451, 4925804, 4997550, 4777186, 4666442, 5335926, 5321737, 5342665, 5178550, 5342675, 5161699, 4771652, 5211324, 5161698, 4808052, 5364457, 5531804, 5226289, 5226288, 5321898, 4861768, 5226279, 5321708, 5157248, 5120319, 5157258, 5022199, 5321705, 4861748, 5342679, 4644919, 4811654, 5161715, 5451835, 5364454, 5157257, 4936462, 5120334, 4808049, 5453502, 5161708, 5161701, 4777202, 5120329, 5451836, 4925803, 4975657, 4897912, 4897920, 5037453, 5156923, 5178571, 5178568, 5464603, 5022216, 4925821, 5389709, 5303538, 4744368, 5178553, 5335917, 5335901, 5303540, 5531792, 5362650, 4808032, 4896547, 5342673, 5156914, 5157240, 5022968, 5022192, 5156904, 4795033, 4644953, 5389708, 5342600, 5120335, 5022215, 5178575, 5204817, 5204808, 5120321, 5178573, 4936466, 4627293, 5022964, 5022190, 4832808, 4975660, 4795032, 4627250, 5335912, 4627267, 5321900, 5022208, 4644947, 5102774, 5321907, 5364437, 5301496, 5037445, 4771644, 5342668, 5421528, 5421537, 4888752, 5464587, 5022196, 4795022, 5362654, 5156922, 4975654, 4975659, 5211326, 5156927, 5211310, 5389699, 4808054, 5389716, 5421544, 4756996, 4757000, 5432822, 5362666, 5464591, 5301493, 5303542, 4771669, 4756989, 5156909, 5156913, 5156928, 5178561, 5335906, 5453482, 5389704, 5451821, 5342596, 5342597, 5321930, 5342664, 5362672, 5335898, 5342685, 5321720, 5102777, 5178576, 5102775, 5102757, 5156915, 4777209, 5321724, 5342667, 5161706, 5342663, 4861753, 5102755, 5453500, 5321925, 5321910, 4925809, 5321923, 5335907, 4936446, 5321902, 5321903, 4644901, 4832826, 5335899, 4936468, 5586520, 5566099, 5572575, 5572579, 5586523, 5566092, 5572591, 5586528, 5572589, 5566098, 5572569, 5586543, 5572583, 5572574, 5615315, 5566083, 5572566, 5572576, 5586525, 5586517, 5586540, 5571945, 5572584, 5571958, 5571953, 5571947, 5667368, 5572585, 5667370, 5617497, 5586531, 5566077, 5572572, 5586530, 5586516, 5566095, 5566085, 5586518, 5566104, 5572587, 5566079, 5571951, 5617505, 5572571, 5615327, 5616849, 5586544, 5586532, 5615317, 5616858, 4332469, 3892405, 4182756, 3636134, 4108492, 4068811, 4238959, 4216265, 4150429, 3636227, 4035628, 4172909, 4509258, 4509319, 4182767, 3880133, 4122480, 4428891, 4509222, 4182773, 4339593, 4266665, 4256916, 4332612, 4428845, 4266638, 4172908, 4428852, 4256927, 3832662, 4332609, 4509315, 4216280, 3985499, 3823793, 4193458, 5321921, 5204804, 4936467, 5421525, 5321911, 5178567, 5321918, 4896541, 5120339, 5303536, 5342592, 5451814, 5364451, 5226297, 5389706, 4897916, 5342608, 5531798, 5342647, 4777193, 5303546, 5342672, 5342644, 4925810, 4975645, 5389695, 5364436, 5453496, 5342598, 4848282, 4627275, 5421540, 4795039, 5301498, 4756995, 4637059, 5364440, 5431131, 5204822, 5156905, 5342618, 5022982, 5431136, 5342676, 5156908, 5342623, 5362652, 5204813, 5120316, 5389701, 5211302, 5453497, 5389690, 4896540, 4861767, 4896536, 5342588, 5022970, 5342606, 5157251, 5301494, 5301495, 5364444, 5421538, 5389713, 5453508, 5321906, 5531816, 5431111, 5389707, 5321706, 5301499, 5301508, 5321728, 5421541, 4627230, 5531796, 5531801, 5303559, 5321734, 5156911, 4627269, 4975656, 5342635, 5342674, 4936450, 5421534, 5335925, 5161722, 5453501, 5453509, 5204805, 5037457, 5531802, 5451833, 5451828, 5161720, 5531803, 5120324, 5531794, 4925826, 4848255, 5022197, 5157255, 5301511, 5342633, 4897918, 4997535, 5022194, 5301509, 5157237, 5464594, 5335921, 4690963, 4975647, 5531812, 4690967, 5342641, 5453513, 5342593, 5342602, 5342661, 5421531, 5364445, 5432833, 4627277, 4676816, 5389718, 5531818, 5335908, 4955400, 5389715, 5464610, 4925819, 5321718, 5464592, 4666447, 5362664, 5431134, 4808055, 5342642, 5303545, 5120320, 5389697, 5531793, 5364456, 5102762, 5432820, 5342687, 4644946, 4808043, 5157253, 5431109, 5431123, 4676818, 5531813, 5342678, 5342691, 5301487, 5321727, 5102768, 5301491, 5342632, 5432834, 5342628, 5226290, 4644940, 5432819, 5453492, 5431115, 5342601, 4637051, 5431130, 5389719, 4832810, 4627240, 4627289, 5464595, 5464588, 5226278, 5335924, 5364455, 5431138, 5120322, 5342611, 4756992, 5321922, 4897938, 5531819, 5204818, 5464612, 4936459, 5464607, 5451837, 5022962, 4808041, 5432831, 5303553, 5464597, 5451817, 5421539, 5389711, 5389702, 5451829, 5531809, 5389692, 5211309, 5421530, 5342652, 5156925, 5531808, 4975664, 5431135, 5451834, 5531811, 5431121, 5531795, 5531800, 5421520, 5211325, 5431129, 5211311, 5453493, 5431133, 5421517, 5453505, 5453498, 5364459, 4897928, 4676824, 5321924, 5362649, 4896553, 5431132, 5531799, 5178565, 5342612, 5303552, 5303539, 5211318, 5421518, 5421535, 5451830, 5464601, 5161718, 5211321, 4955397, 5342682, 5342666, 5432826, 5342649, 4897936, 5204824, 4777196, 5464598, 5321934, 4771647, 4644903, 4832811, 4811666, 5389693, 5342630, 5431110, 5022200, 5362659, 5204819, 5303554, 5204832, 5451818, 4936472, 5335911, 5421546, 5342643, 5451838, 4896564, 5572582, 5616846, 5616841, 5667375, 5615324, 5850840, 5615301, 5848000, 5912999, 5897791, 5937606, 6051724, 5897783, 6130288, 5956373, 6088087, 6088105, 5912992, 6087928, 6087920, 6130302, 6130290, 5951207, 6130537, 6088081, 6130282, 6088113, 6087911, 5571967, 6045263, 5850850, 5667365, 5951195, 6130291, 5937794, 6179494, 5572570, 5937808, 5566091, 5586535, 5621264, 5912977, 6087912, 5937628, 6088128, 5951196, 6088120, 5571942, 6087923, 5572564, 5621279, 5667391, 5667372, 5937796, 5617526, 5937792, 5951199, 5950811, 5937637, 6051683, 6051680, 5956379, 6087910, 6087929, 6087909, 6087892, 6130289, 5950813, 6237423, 6068704, 6130560, 6179477, 6051710, 6237429, 6254822, 6154675, 6237421, 6254819, 6254818, 5616853, 5615303, 6254812, 5617519, 5617500, 5617511, 5850831, 5850827, 5850845, 5949268, 6068707, 5850839, 6130300, 6154662, 6179490, 5951206, 6130547, 5571949, 5951200, 5912986, 5912987, 5937807, 5949289, 5617503, 5667379, 5950815, 5667387, 5571948, 5848008, 5897785, 6130295, 5937809, 5615308, 6051714, 6087915, 6179502, 5615306, 5615316, 5951213, 5937605, 6088118, 5586536, 5937624, 5621274, 5615310, 5848011, 6130284, 6068714, 5616845, 5667384, 5617504, 5912988, 5847997, 6154668, 6045262, 5572592, 6051695, 5586538, 5937632, 6088078, 5616854, 5847999, 5897804, 5937602, 5951197, 6068713, 5949280, 5621251, 5956375, 6051677, 5571957, 6045244, 6088089, 5617509, 5617514, 5615325, 5616859, 5897793, 5951187, 6051679, 5937802, 6130296, 6088127, 6254814, 5897803, 5667380, 6251122, 5615321, 6179506, 5615330, 5949282, 5617499, 6088077, 6130549, 6130273, 5615322, 5615320, 6051696, 5621260, 5897782, 5621261, 5667385, 5847996, 5848010, 5571965, 5950802, 6051687, 6051699, 5937788, 6179873, 6237418, 6088116, 6045249, 5621278, 5621273, 5848001, 5912985, 5621254, 6130556, 5586533, 5615318, 6179863, 5566081, 5617522, 5847994, 5912976, 5847988, 5950794, 5956365, 5912997, 6078974, 5897778, 5950797, 6051721, 5897781, 5615323, 5937619, 5937620, 5949274, 5951186, 5949281, 5615326, 6087924, 6130299, 5566094, 5566101, 5897807, 5586519, 5571964, 5667363, 5572565, 5616840, 5617520, 5616856, 5615304, 5883125, 5883129, 5572586, 5937622, 5937608, 5617518, 5937611, 5937613, 5950800, 5951188, 6051682, 6045243, 6045241, 6051709, 6045251, 6068700, 5937817, 6078994, 6088114, 6088132, 6078992, 6087895, 6088076, 6088091, 6237439, 5667374, 6068717, 5950808, 5667386, 5566097, 5883130, 5586521, 5956370, 5615312, 5616855, 5615314, 6087931, 5571961, 5667364, 6088096, 6045254, 5667389, 5572588, 5847990, 5951214, 5883131, 5937814, 5667383, 5566105, 5897805, 6051712, 5615328, 6130301, 6088129, 5937818, 5897796, 5566084, 6078993, 5617517, 5847995, 5615313, 5621257, 6051720, 5897788, 5912978, 6088090, 5937786, 5571954, 5586541, 6045256, 5617510, 5937783, 6051726, 5621269, 5949266, 6087898, 5937791, 6130293, 6130278, 5912984, 5615302, 5617501, 5572573, 5566089, 5950792, 5937820, 6045258, 6237417, 6130275, 5883127, 5847989, 5897795, 5897794, 5586522, 5883126, 5617513, 5617523, 5572567, 5616847, 5586542, 5937635, 5937626, 5616838, 6051702, 6254805, 5667376, 6130538, 5621252, 5566082, 5586537, 5667371, 5571962, 6045248, 5571955, 6088131, 5617498, 5937789, 5937631, 5937638, 6087916, 6068712, 6237415, 6154667, 6130550, 5937618, 5913000, 5571950, 5937639, 6088083, 5950810, 5571963, 5912972, 5621267, 5621262, 5616860, 6051705, 6051717, 5951201, 5621258, 5956377, 6051704, 6051698, 6051691, 5586539, 5617508, 5937616, 5617515, 5912971, 5949285, 5615305, 5937815, 5850846, 5616865, 5850844, 5586526, 6051708, 5621259, 6068716, 6078988, 5850841, 6130534, 5937801, 5912989, 5883124, 5912991, 5912995, 5937603, 5937609, 5912975, 6068725, 6078996, 5621276, 5850833, 6045261, 5937625, 6130286, 6079000, 6179861, 6254806, 6130292, 6078989, 5956358, 6087918, 6087925, 6087917, 5912982, 5617506, 6068724, 6237431, 5616839, 6051703, 5949291, 5950793, 5616842, 5950790, 5951189, 5951193, 5951205, 6087904, 6130535, 5951204, 6078975, 6078983, 5937800, 6237435, 6237440, 6130533, 6237420, 6237430, 5897800, 5897780, 5951192, 5897801, 5950798, 5950816, 6088098, 5566080, 5848009, 5616857, 5848003, 5616863, 5847992, 5848004, 5848005, 5667388, 5616843, 4281089, 4281069, 4339595, 4509205, 4281077, 4509309, 4141970, 4238960, 4256915, 4403183, 4339580, 4193447, 4150434, 4509228, 4155666, 4332462, 4756990, 5342613, 5451813, 5335900, 4925824, 5303560, 5342645, 5464609, 4975646, 5037446, 4997528, 4744354, 5451823, 4627263, 5321726, 5157230, 5342651, 5301490, 5342616, 5178572, 5157252, 5531790, 4897932, 5451812, 5464599, 5037432, 4627296, 5102750, 5178558, 5211304, 5431117, 4808053, 4627271, 4627224, 5342603, 5451815, 5342624, 5464600, 5342683, 5226282, 4644939, 5163509, 4777192, 5303548, 4808034, 4811658, 5431127, 5161719, 5342657, 4771662, 4808059, 5157231, 4832825, 5303543, 4848278, 5321725, 5342686, 5451820, 5342660, 5362668, 5335909, 5120315, 5303533, 5342610, 5364435, 5364450, 5321707, 5531810, 5342688, 5451819, 5226293, 4795021, 5321927, 5161711, 4757004, 5342595, 4756994, 5342589, 4811667, 4627282, 4644938, 5342669, 5335914, 4975658, 5453507, 5321916, 4832833, 5102769, 4627310, 5204816, 6179492, 6087900, 6078995, 5951212, 6130559, 6269452, 6277287, 6269459, 6179478, 6237424, 6328498, 6448110, 6448042, 6448757, 6328484, 6448093, 6448049, 6480656, 6269454, 6339352, 6592425, 6277299, 6448011, 6592465, 6339371, 6480662, 5621271, 5571943, 6448068, 6448097, 6277296, 5951203, 5937805, 6078972, 6154671, 6448070, 6130298, 6448025, 6592440, 5951208, 6251129, 6251104, 5617507, 6087889, 6179489, 6448124, 5617524, 6592431, 6598143, 6251116, 6351381, 6277297, 6385257, 6448054, 6088086, 5616844, 6179864, 6385247, 6339367, 6277284, 6339350, 6448118, 6237426, 6237428, 6480663, 6251102, 5616866, 6254816, 6179869, 6254825, 6237437, 6328475, 6334335, 6334330, 6179850, 6269438, 6334320, 6351388, 6295793, 6295766, 6328494, 6451374, 6351375, 6439920, 6448095, 6295781, 6448746, 6295787, 6428638, 6439932, 5937623, 6595594, 5937803, 6295788, 6237425, 6179868, 5667378, 6277303, 6254807, 6334321, 6448031, 6351383, 6334336, 6448071, 6251120, 6448014, 6448016, 6269456, 6045246, 6351408, 6295792, 6277282, 6277288, 5912996, 5937790, 6334329, 5937810, 6448043, 5617525, 6448067, 5949294, 5621265, 6295777, 6334338, 6179847, 6154673, 6448052, 6592460, 6045247, 5937811, 6277302, 5949288, 6154651, 6269448, 6269453, 6334331, 6078971, 5950799, 6595553, 5912981, 6154652, 6448058, 6448117, 6480644, 6592451, 5949275, 6179848, 6088130, 6428626, 6328478, 6592424, 5586524, 6277280, 6448094, 6595542, 6595538, 6328488, 5937781, 6079001, 6251109, 6295778, 6448104, 5667381, 6351391, 6448010, 6480650, 5950788, 6592435, 6251106, 6087913, 6269450, 6251124, 6130548, 6448066, 5950814, 6295773, 6448080, 5937793, 6451372, 6277294, 6269431, 6068720, 6448062, 5667377, 6154657, 6251111, 6480649, 5937633, 6339373, 6592464, 5949272, 5621250, 5937627, 6351409, 6051700, 6448028, 6448048, 6154677, 6254815, 6051689, 6328491, 6334323, 6448754, 6592467, 6254820, 6269435, 5667366, 5937787, 6595588, 6598262, 6251108, 6251126, 6334334, 6277276, 6068715, 5956360, 6179499, 6385256, 6385248, 6334327, 6179858, 6592433, 6251123, 6334337, 6439937, 6088100, 5950817, 6269434, 5615311, 6295785, 6428632, 6385241, 6130561, 5621266, 6254810, 5616861, 6428621, 6179855, 6339378, 6595577, 6179853, 6328483, 6130551, 6051716, 6269446, 6269433, 6295782, 6269451, 6328482, 6328481, 6334328, 6328485, 6334332, 6351387, 6428622, 6428636, 6428627, 6428629, 6428631, 6428625, 6428623, 5571946, 6439930, 6448021, 6439926, 6439935, 5571956, 6448111, 6439942, 6448051, 6448083, 6592478, 6448119, 6592449, 6448749, 6592482, 6592438, 6595559, 6339351, 6592461, 5949293, 6130274, 6179866, 6045253, 5897779, 6254813, 6351384, 6328474, 6351394, 6179481, 6277292, 6328471, 6351380, 6351402, 6448077, 6592439, 5621272, 5949286, 6334324, 5883128, 6045235, 6269457, 6088093, 6179488, 6237432, 6269460, 6051693, 6269444, 5621256, 6051706, 6428637, 5572590, 6051713, 6448107, 6448125, 5615307, 5912980, 6385252, 5937601, 6179503, 6130558, 6179484, 6309360, 6269458, 6428624, 6439917, 6428620, 6078986, 6154650, 6269439, 6339372, 6334333, 6592452, 6254808, 6051694, 6592476, 6448074, 6592481, 5571944, 6130543, 6130294, 5912994, 6351374, 5912993, 5951209, 5951198, 6130562, 6339356, 5937604, 5937785, 6068728, 5950805, 5937607, 6179491, 6269443, 6277286, 5667369, 6428630, 6448026, 6480660, 6078998, 6480657, 6087930, 5912998, 6130281, 6385251, 6269432, 6078984, 6254826, 6295769, 6448759, 6130280, 6448743, 6448084, 6448015, 6592442, 6448760, 6088104, 6045240, 6277277, 6595565, 6448123, 6237414, 6439943, 5937634, 6254817, 6480652, 6448082, 6179480, 6251114, 6045257, 6088080, 6130555, 6087894, 6339369, 6078990, 6351377, 6339374, 6351396, 6385254, 6385240, 5937621, 5937782, 6448758, 6269437, 5667373, 6179876, 5586545, 6439928, 6254834, 5667362, 6351393, 6448741, 6130546, 6068701, 6448755, 6448745, 6277275, 6068705, 6237422, 6328489, 6351395, 6078980, 6448744, 5615319, 6237433, 5850830, 6269442, 6179487, 6251119, 6339361, 6480653, 6385249, 6339360, 6448092, 6448126, 6448050, 6328476, 6448102, 6480661, 6087903, 6088133, 6480659, 6088095, 6334322, 6295764, 6439938, 5566088, 5566102, 6130285, 6328480, 6448024, 5897786, 5950795, 6130287, 5937797, 5949273, 6051686, 5949267, 6269440, 5956372, 6339354, 6154663, 5848002, 6277295, 6277298, 5848006, 5850843, 6154659, 5897790, 6269441, 6448747, 6269436, 4509269, 4925823, 5451822, 5432832, 5342619, 5211303, 4861765, 5342636, 5335919, 4848283, 5464590, 5335904, 4832822, 4627323, 5432823, 5301505, 5226281, 5022212, 5022203, 4627233, 4690965, 4975648, 5342656, 5022198, 4975653, 4925806, 5211316, 4888754, 4897910, 5161723, 5161702, 5342599, 5432824, 5321919, 5301503, 4848259, 5342646, 5342626, 4832827, 5211319, 5211329, 4888729, 5342587, 4627322, 5102748, 5432821, 5431137, 5342631, 5204823, 5342671, 5342605, 5464614, 5364442, 4771648, 4690960, 5211312, 5335913, 5364458, 5226286, 4811676, 5037447, 5161700, 5362662, 5178554, 5432828, 5951191, 5897792, 6179496, 6592468, 6793361, 6745757, 6598134, 6592454, 5937812, 6745753, 6787954, 6786060, 6787948, 6787917, 6830461, 6787961, 6832583, 6179852, 6078976, 6339358, 6385242, 6830468, 6087897, 6328479, 6068709, 6598277, 6856720, 6598146, 6830447, 6856722, 6856704, 6251100, 6598128, 6439934, 6068702, 6448751, 6787930, 6154665, 6598118, 6595587, 6595567, 6592470, 6595548, 6439918, 6780589, 6780600, 5850828, 5937612, 6856700, 5566090, 6598117, 6598111, 6786043, 6595564, 6592459, 6787922, 5949287, 5950807, 5571959, 6088124, 6598239, 6592450, 6598237, 6251125, 6448128, 6448013, 6087919, 6856726, 5566096, 6448113, 6595582, 6786058, 6745760, 6448112, 6787940, 6787942, 6595571, 6856716, 6787927, 6830438, 6856714, 6448065, 6309364, 6592432, 5912974, 6786062, 6592427, 6448059, 6793359, 6088117, 5617502, 6448108, 6448072, 6787949, 6598155, 6592429, 6154653, 6595572, 6787960, 6598225, 6598230, 6598253, 6351386, 6351407, 6856702, 6251128, 6295765, 6351397, 6087905, 6595556, 6856725, 6068719, 6598227, 5621255, 6277281, 6154676, 6480654.6269449, 6179870, 6179849, 6448105, 6251127, 6787945, 6745754, 6745756, 6787965, 6295767, 6592447, 5616852, 6780602, 6592434, 6598154, 6595579, 6595585, 6830424, 5950806, 6592456, 6598279, 6598266, 6787939, 6787919, 6592443, 6787926, 6448060, 6595586, 6787928, 6130542, 6592453, 6439941, 6480665, 6078982, 6448075, 6780603, 6787929, 6385245, 6595581, 5950801, 6448750, 6787958, 5616850, 6179862, 6179482, 6277293, 6385239, 6339359, 6480643, 6598157, 6428633, 6451378, 6793364, 6428628, 6595584, 6595580, 6598261, 6598251, 6051690, 6793371, 6269445, 6277291, 6595583, 6780605, 6154654, 6787969, 6830432, 5571969, 6787967, 6598132, 6277300, 6598265, 6179874, 6179865, 6179875, 6598242, 6598248, 6448099, 6786055, 6787925, 6451369, 6385260, 6595544, 6295779, 6786066, 6351376, 6448106, 5951202, 6592473, 6154661, 6856705, 6592477, 6595561, 6254833, 6595591, 6787934, 6592428, 6595547, 6351406, 6598276, 6780601, 6448121, 5621277, 6179859, 6328472, 6787947, 6254821, 6351392, 6595573, 6745759, 5847998, 6598156, 6745752, 6787918, 6598133, 6328486, 6786068, 6856707, 6787957, 6385258, 6598139, 6793356, 6339357, 6598119, 6277289, 6786059, 6451371, 6451370, 5566078, 5621263, 6595555, 6595554, 6448036, 6051692, 6448018, 6339375, 6598122, 6154674, 6339377, 6595552, 6595578, 6598223, 6598238, 6598125, 6780597, 6598255, 6598254, 6592462, 6786054, 6787924, 6830426, 6830462, 6793358, 6598247, 6830458, 6328487, 6832577, 5949276, 6598229, 6339379, 6598120, 6592458, 6598137, 6448069, 6428634, 6595551, 6598250, 6598148, 6787935, 6051701, 6787972, 6598144, 6598136, 6339353, 6385238, 6787974, 6237436, 6051715, 6251112, 6448057, 5949271, 6277283, 6088110, 6595557, 5617512, 6598129, 6334319, 6068723, 6087908, 6598127, 6787943, 6179497, 6598141, 6480646, 5937614, 6351400, 6068708, 6787937, 6793367, 6078991, 6780594, 6598241, 6045236, 6787921, 6832585, 6856733, 6595593, 5950804, 6448034, 6328473, 6087887, 5621275, 6130539, 6793362, 6793374, 5571968, 6448091, 6595595, 6130541, 6309363, 6598151, 6154678, 6598112, 6787971, 6598273, 6830453, 6277278, 5951190, 6856727, 6428619, 6598116, 6480658, 6448045, 6598270, 6439929, 6598235, 6592466, 6351403, 6439916, 6448029, 6595541, 6078973, 6088106, 6592480, 5571952, 6130279, 6787956, 6787936, 6598138, 6598159, 6787916, 6448748, 6480669, 6051681, 6088097, 6088082, 6595545, 6087893, 6088075, 6598161, 6780606, 6339362, 6251103, 6439939, 6787975, 6787955, 6787941, 6448012, 6595575, 6309362, 6448022, 6428635, 6309365, 6595589, 6598245, 6598115, 6251117, 6309359, 6439921, 6786050, 6830442, 6830449, 6830469, 6830419, 6448753, 6745755, 6832575, 6832579, 6832584, 6832586, 6448009, 6448047, 6439922, 6786056, 5621268, 6830423, 5956371, 6254809, 6786042, 6045237, 6087906, 6328493, 6385255, 6786044, 5949265, 6339376, 6598224, 5949277, 5615309, 5949279, 5667367, 6787966, 6088115, 6595563, 6439924, 6592455, 6793363, 5956374, 6439923, 6598259, 6385250, 6339363, 6595537, 6830440, 5847993, 6251107, 6339355, 6448076, 6595536, 6448037, 6780598, 6439931, 6598150, 6439936, 6598110, 5897798, 6078979, 6088079, 6967221, 7001551, 6967229, 6974507, 7000987, 7001006, 7032323, 7001000, 7032321, 7033637, 7032340, 7032366, 7032351, 7001012, 7000997, 7001519, 7000983, 6974493, 6967241, 7001537, 6912359, 7033645, 7033648, 7033644, 6912351, 6974480, 6974472, 6974473, 7001525, 7001552, 6967210, 6912347, 7001554, 7001546, 7032367, 6974494, 6967234, 7032328, 7033642, 7032370, 6974497, 6974481, 7000993, 6967228, 7032363, 6967212, 6967217, 7032358, 6974479, 6912352, 7001564, 6912360, 6974495, 7001527, 7001033, 6912356, 7001553, 6967216, 6967235, 7001558, 7001547, 7001522, 7001560, 6967243, 7032337, 6967227, 6967218, 6974474, 6974528, 6967225, 7001545, 7001034, 6974524, 7001556, 7001010, 7001018, 7033635, 7001540, 7001003, 6912346, 7001538, 7033638, 6912355, 7032374, 7001521, 6974526, 7000985, 6967231, 6974508, 6967246, 7001536, 6912358, 6974503, 6974504, 6967244, 7032331, 7001027, 7032353, 7001528, 6967239, 6967219, 7001567, 6974512, 7001004, 6974502, 6967247, 7032360, 7001011, 7001022, 6974483, 6974488, 6974501, 7032322, 7032329, 7032333, 7032362, 7032376, 6967222, 7032361, 7001565, 5321909, 5156921, 5342607, 5321729, 4925811, 5204829, 5342684, 5226299, 4925827, 5389717, 5364448, 5342590, 5362651, 5342680, 5453490, 5342615, 4795038, 5431118, 5421521, 5342640, 4811675, 5421527, 5204828, 5301502, 5161726, 6598272, 6830448, 6830474, 6051711, 5850829, 6598158, 5566103, 5937784, 6786061, 6598152, 6592457, 6351404, 6595590, 5850826, 6787938, 6780604, 6780595, 6595576, 6856719, 6254811, 6439919, 5616851, 6780593, 6448127, 6277290, 6154672, 6179856, 6051697, 6351390, 6309361, 6448088, 6480655, 6832578, 6351389, 6088103, 6385262, 6448756, 6787933, 6787923, 5586534, 6328492, 5912979, 6595540, 6830452, 6451377, 6439940, 6068710, 6448120, 5566100, 6448027, 5571971, 6832582, 6830434, 6480648, 6088123, 6328499, 6595558, 6251105, 6745758, 6078985, 6780590, 6598145, 6351385, 6254829, 6480664, 6045255, 6237413, 6598246, 6254832, 6328490, 6830455, 6595546, 6339368, 6856706, 5848007, 6830437, 6448053, 6786052, 6087921, 6068722, 6451373, 6254828, 6598236, 6598226, 6045234, 6592441, 6130545, 6787932, 6295776, 6448087, 6087922, 6856708, 6088112, 6598121, 6598275, 6598126, 6786047, 6351379, 6448090, 6793372, 5667390, 5937640, 6439927, 6480666, 6154649, 6856709, 6078997, 6237434, 6480647, 6830476, 6254830, 5850847, 5850836, 6295783, 5850842, 5850849, 5850835, 6787963, 6179495, 6448017, 6856728, 6448101, 6787944, 6448019, 6448752, 6328497, 6339364, 6598153, 6088088, 6856734, 7032377, 6974510, 6974506, 7001031, 7032334, 7032343, 7032372, 7032335, 7001019, 7032330, 7001539, 6967220, 7001021, 7001517, 7001563, 7001562, 6974478, 6974496, 6912349, 7001516, 7032369, 7001015, 6912348, 6912350, 6967245, 7032371, 6974482, 7033640, 7001029, 7032346, 7001550, 6967232, 7001543, 7032344, 7032373, 7000989, 7000998, 6974470, 7001529, 6967226, 7000982, 6974489, 7000992, 7032364, 7032355, 7032325, 7032349, 6974471, 7001024, 7032324, 7033639, 6974519, 6974500, 7032378, 7033634, 7001007, 7000991, 7001023, 7032347, 7000994, 6967215, 7032356, 6974487, 7032327, 6051719, 6351398, 6592430, 6598114, 6088125, 5949270, 6448073, 6592463, 6351382, 6088092, 5951215, 6179486, 5949290, 6598149, 6179504, 6598257, 6448086, 6439925, 6451376, 6595574, 6480641, 6448063, 6351405, 6051707, 6856701, 6786041, 6598243, 6051723, 6130540, 6598234, 6592446, 6130297, 6295790, 6351399, 6780599, 5949278, 6480640, 6598256, 6295770, 6598231, 6385244, 6179872, 6598147, 6295780, 6787953, 6793370, 6269455, 5667382, 6786057, 6448079, 6856721, 6595543, 6334326, 6787968, 5850848, 5850837, 5897806, 6087901, 6793365, 5897802, 6793369, 6087899, 5850838, 6595539, 6786064, 5897787, 6385236, 6787959, 6448023, 6179483, 6385243, 6130557, 6351378, 6598260, 6339366, 6295791, 6595562, 6967213, 6974476, 6974491, 7001014, 6967214, 6974520, 7000995, 7001016, 6974492, 7033643, 7001526, 7001020, 7033636, 7001025, 7001531, 7001557, 7000999, 6974485, 6592475, 6830475, 6595566, 6830429, 6830443, 6087914, 6045260, 6856724, 6448064, 6974490;
b) Item Number 314-13-14, UDI/DI 10885862172716, Serial Numbers: 2204764, 2204767, 2452199, 2334147, 2204753, 2204770, 2204766, 2204755, 2334126, 2331025, 2339039, 2331035, 2406555, 2334141, 2331030, 2331032, 2452203, 2334139, 2460120, 2406560, 2497433, 2497452, 2334125, 2406558, 2406557, 2334142, 2452188, 2452209, 2460126, 2499279, 2406556, 2452202, 2452198, 2334134, 2330862, 2452215, 2334151, 2334135, 2331029, 2452185, 2460128, 2406554, 2334138, 2460124, 2499270, 2204765, 2204756, 2334137, 2334136, 2330860, 2331037, 2331023, 2330861, 2406562, 2499259, 2449061, 2452190, 2406552, 2452212, 2339043, 2449062, 2339038, 2499281, 2499271, 2497438, 2460122, 2474455, 2499276, 2406561, 2334145, 2204762, 2204758, 2204763, 2334140, 2406564, 2339042, 2452220, 2474461, 2334128, 2499267, 2497442, 2204768, 2204754, 2204759, 2204761, 2204760, 2204752, 2334129, 2204757, 2334146, 2406551, 2460129, 2452186, 2460125, 2474460, 2499255, 2499280, 2499256, 2497436, 2497439, 2331024, 2474465, 2406548, 2452207, 2499268, 2331031, 2331040, 2331041, 2449058, 2497450, 2497435, 2334144, 2499264, 2334152, 2339041, 2331027, 2339037, 2334130, 2499265, 2474463, 2334149, 2334131, 2562233, 2582353, 2605418, 2499260, 2452191, 2605423, 2452187, 2622589, 2665863, 2605411, 2452195, 2546881, 2605401, 2499262, 2499273, 2497451, 2546875, 2665862, 2546846, 2546857, 2605403, 2600898, 2622590, 2334143, 2562228, 2622594, 2622562, 2331038, 2334133, 2582356, 2331026, 2622567, 2622571, 2582350, 2497460, 2582348, 2499258, 2546849, 2622575, 2622570, 2546853, 2334127, 2605415, 2452189, 2334150, 2546854, 2622595, 2665878, 2665869, 2546887, 2605419, 2622581, 2499274, 2605393, 2497461, 2546859, 2605390, 2582351, 2605405, 2546878, 2406549, 2605413, 2499275, 2622584, 2460127, 2497445, 2497441, 2605426, 2605392, 2665875, 2605408, 2452193, 2665856, 2449059, 2605400, 2497448, 2622588, 2546847, 2665868, 2562227, 2204769, 2605429, 2665886, 2334154, 2406563, 2452217, 2582346, 2665859, 2499266, 2546877, 2452194, 2605388, 2331039, 2582355, 2499282, 2562225, 2605412, 2605397, 2605410, 2665874, 2665879, 2622565, 2605391, 2452208, 2452210, 2474466, 2499277, 2665884, 2546885, 2499272, 2605422, 2497458, 2452216, 2546864, 2546870, 2497457, 2497444, 2605398, 2622592, 2622580, 2582352, 2665872, 2665871, 2622577, 2600897, 2622574, 2406550, 2622569, 2452218, 2497437, 2546872, 2474462, 2474458, 2546880, 2622587, 2622585, 2452197, 2605406, 2622597, 2452211, 2546852, 2474464, 2622572, 2562234, 2546848, 2582361, 2622564, 2474457, 2497453, 2546856, 2582358, 2605416, 2605417, 2622563, 2600900, 2622582, 2622598, 2605409, 2460121, 2546869, 2605396, 2474456, 2497446, 2546860, 2497443, 2546882, 2334132, 2605407, 2622593, 2546876, 2546883, 2562226, 2546845, 2546866, 2582360, 2497434, 2582359, 2665873, 2622568, 2605424, 2452184, 2497455, 2497440, 2600903, 2406553, 2546888, 2605420, 2562232, 2562231, 2452192, 2452196, 2546884, 2474459, 2665881, 2497432, 2546862, 2622586, 2331036, 2497456, 2546863, 2605402, 2497459, 2562230, 2622583, 2582347, 2582357, 2546850, 2665865, 2499254, 2497449, 2605399, 2605428, 2605427, 2546874, 2562224, 2331034, 2331033, 2499278, 2546855, 2452214, 2605395, 2605425, 2622578, 2665882, 2600899, 2622573, 2665866, 2452201, 2689072, 2700788, 2777460, 2777474, 2770162, 2777517, 2812221, 2689070, 2770142, 2770434, 2812218, 2770140, 2689077, 2700799, 2770437, 2777539, 2770143, 2770150, 2770157, 2873850, 2689069, 2700804, 2689088, 2777466, 2689080, 2770444, 2777520, 2689255, 2777538, 2689071, 2770168, 2770447, 2689092, 2689253, 2689261, 2770160, 2689094, 2700795, 2689089, 2700800, 2770161, 2812222, 2770175, 2698237, 2770152, 2770450, 2777534, 2689083, 2700787, 2700796, 2770439, 2689256, 2777544, 2770456, 2700792, 2689250, 2770446, 2689086, 2777540, 2770174, 2812248, 2777470, 2812234, 2777483, 2777477, 2812231, 2770455, 2689074, 2897365, 2770132, 2777515, 2770130, 2812227, 2770155, 2812233, 2689076, 2770153, 2777528, 2689082, 2812232, 2770149, 2770438, 2770176, 2700794, 2689090, 2689075, 2770443, 2700803, 2770451, 2812236, 2812237, 2770131, 2812220, 2777481, 2777514, 2873853, 2770173, 2777482, 2700802, 2873847, 2689087, 2689252, 2689254, 2770457, 2770128, 2777532, 2777478, 2777531, 2770164, 2700807, 2770172, 2689093, 2689073, 2689068, 2700798, 2777535, 2777480, 2777471, 2777518, 2777475, 2770167, 2700791, 2777541, 2777479, 2777513, 2777464, 2873841, 2689251, 2770165, 2700793, 2777501, 2665864, 2546858, 2546851, 2452219, 2665857, 2452213, 2334153, 2605389, 2546871, 2452205, 2497454, 2474467, 2546886, 2605430, 2665880, 2665885, 2452206, 2600901, 2605404, 2546865, 2497447, 2622596, 2546868, 2449060, 2665860, 2499269, 2499261, 2622591, 2605421, 2582349, 2605394, 2339040, 2474468, 2546879, 2499263, 2951950, 2951942, 3521547, 2951914, 3521557, 3577247, 3670721, 3611307, 2920011, 3670582, 3611335, 3577260, 2920010, 2951922, 2979130, 3577262, 3673202, 2897376, 3670585, 2777468, 2920014, 2777484, 3577248, 2770135, 2770147, 2897383, 2920027, 2979122, 2979128, 2920015, 3564201, 3577252, 3564193, 3577268, 3611315, 3673158, 3610586, 3670714, 3673217, 2689091, 2873838, 2897378, 3577261, 2951927, 3611985, 2979136, 2920005, 3577273, 2777472, 2897372, 2770139, 3611990, 3577256, 3610579, 2951918, 2770435, 2897389, 2951971, 3670690, 3564189, 2689085, 2951929, 3564216, 2951936, 2777510, 2897377, 2897375, 3612000, 3611980, 3611316, 3673163, 2920025, 2951955, 2979106, 3670580, 3670655, 3670656, 2873861, 3670727, 2979129, 3673113, 2920032, 3673222, 3670631, 2777500, 3673215, 2700786, 2873839, 2777465, 3611305, 3670692, 3670644, 3670695, 3724445, 3776230, 3670658, 3670707, 2920022, 3670578, 2897371, 2919985, 2812229, 2951941, 2920018, 3564203, 3673200, 3670565, 3673162, 3724454, 3670612, 3673115, 2777508, 2897367, 3521549, 3521553, 2951953, 2951967, 3577263, 2812230, 3611987, 3673128, 3564204, 3750164, 3611301, 2777463, 3564198, 3577253, 3611319, 3673136, 3673125, 3670653, 3670589, 2873843, 3564205, 2951947, 3577241, 3611331, 3670559, 3670726, 3670639, 3673156, 3670637, 3670723, 3670624, 3756685, 3611328, 2770156, 2770166, 2777462, 2777507, 3521551, 2920006, 3611991, 3611988, 3610583, 3673207, 3724444, 2689259, 3564210, 3670552, 2812239, 2951944, 2897366, 2770448, 3724441, 3556640, 3611989, 3611318, 3673219, 3673132, 3670625, 2770458, 3673130, 3673160, 2897386, 2920024, 2770154, 2689066, 2919988, 2920008, 2920007, 2979124, 3611327, 2951937, 3670712, 3673150, 3724457, 3670729, 3670636, 3670550, 3724452, 3776239, 2897381, 2812243, 2979133, 2979111, 3577267, 3611302, 3756677, 2812235, 3670725, 2919987, 3611332, 2920021, 3611979, 3611984, 3724447, 2951931, 3796813, 2873856, 2897380, 2951915, 2873862, 2951933, 2951935, 3577255, 3670720, 2979114, 3756676, 2951962, 3577254, 3577242, 2812223, 2873837, 2873863, 2951917, 2979125, 3756680, 2777516, 2812247, 2951960, 3673117, 2919990, 3673209, 3673122, 2951934, 2979104, 2873865, 2979109, 2979131, 2812241, 3610580, 3670713, 3564187, 2812228, 2897374, 2919997, 3611992, 3673152, 2770137, 2777529, 2812226, 2812225, 2951952, 2979113, 3577250, 3611993, 3577266, 2777519, 2777537, 2770133, 2873864, 2951957, 2951945, 3577249, 3577259, 2897373, 3611313, 3670687, 2920029, 2920026, 2919983, 3564219, 3724446, 2951916, 2777522, 2812240, 3611306, 3670576, 2777542, 2920031, 2951913, 3564188, 2770138, 2951938, 3556641, 3610574, 3611292, 3670635, 2689065, 3673164, 2770136, 2812216, 2770441, 2979105, 2979121, 3564211, 3611978, 2979127, 3611977, 3611994, 2920020, 2919994, 3670591, 3673154, 2897385, 2873849, 3521556, 3670620, 3673198, 2777524, 3610582, 2812224, 3673210, 3577271, 2920030, 3670598, 3670616, 3673221, 3756674, 3564195, 2979123, 2812244, 2951969, 3556644, 2897382, 2689257, 3611309, 3611329, 3670728, 3670654, 2700801, 3556643, 3673213, 2777461, 2897379, 2812217, 2979132, 3610581, 3556642, 3611304, 3564194, 3611333, 3673204, 3750149, 3521548, 2951924, 2873854, 2979134, 2951920, 3670594, 2919989, 3556646, 3670572, 2979115, 2873858, 2777473, 2979139, 3577251, 3670704, 3673197, 2920000, 3750160, 3611986, 2919995, 3612001, 3673131, 2897387, 2919991, 2897363, 3521552, 3610592, 3611321, 3750163, 3611995, 3670700, 2920004, 3673151, 3577264, 2873840, 2897390, 3564213, 2873845, 2979137, 2689260, 3564215, 3564200, 3611998, 3611323, 2689081, 2777506, 2951928, 3673201, 3670570, 3776226, 2897362, 2812242, 2770141, 2897384, 3564217, 3670574, 3724450, 2979135, 2951954, 3611322, 2770151, 2951946, 3673123, 3521555, 3611334, 3564207, 2770445, 2700805, 2951932, 2919998, 3564197, 3670724, 3610577, 2812245, 3670711, 3670590, 3670715, 2979107, 3577270, 3564212, 3776228, 2777511, 2919984, 2777467, 2689067, 2951956, 2920017, 2979117, 3577243, 3610572, 3724440, 3724455, 3670596, 3670640, 3673135, 3611290, 2873842, 2919999, 2920002, 3670560, 2979120, 3670648, 2951926, 3577257, 2873852, 3670642, 2770158, 3611320, 3556645, 2700789, 2777521, 2979141, 3577269, 2777476, 2873866, 2873844, 2897361, 2873851, 2689084, 2920001, 2951966, 3521550, 2979138, 3564202, 3610585, 3611289, 3611291, 3611298, 3610593, 3670703, 3670709, 3673211, 3673195, 3670659, 3670628, 3670641, 3670649, 3673212, 3796810, 2951964, 2979110, 2777530, 2698247, 3673114, 2770452, 3611326, 2777523, 2873848, 2919986, 3610575, 3564218, 3670650, 3750158, 3611308, 2979126, 3611312, 2770453, 2951930, 2689264, 2920009, 3577244, 2897368, 2812251, 3673205, 2777527, 2777502, 3724438, 3670595, 3670554, 3610589, 2777505, 2920012, 2951965, 3521554, 2777504, 3673225, 3673196, 3611310, 3564190, 3670583, 3673159, 3670697, 3673147, 3670592, 3670634, 2951943, 2951961, 2605414, 2452200, 2331028, 2665877, 2406559, 2546867, 2499257, 2460123, 3611288, 3564208, 3953205, 4054938, 3893506, 3823872, 3823215, 3823240, 4049196, 3796833, 3823752, 4005410, 3823247, 3985523, 3724442, 3673206, 4141996, 3611311, 3611996, 3823229, 3823768, 3823222, 3893509, 3611325, 3953203, 3953217, 4054939, 4054941, 3670717, 3670696, 3724458, 3521546, 3823212, 3823245, 4018026, 3953175, 4041866, 3850702, 3611330, 3832811, 3670694, 3823252, 3832833, 3750162, 3961565, 3985527, 3953193, 3670593, 3985524, 3670573, 4049189, 4075927, 3670688, 4108453, 4054937, 3776227, 3823217, 3832831, 3670586, 4018025, 3670701, 3670579, 3670564, 3876282, 3953207, 3776237, 4032027, 4122642, 3823869, 4005405, 3832830, 3670652, 3961557, 3823759, 3832816, 3823754, 3823873, 3823750, 3776233, 3953189, 3670553, 2951940, 2770163, 3850706, 3832813, 4005406, 4091536, 3724443, 3823760, 3832839, 3876267, 3932410, 3876289, 3876284, 4020699, 3961563, 3953191, 4121953, 4121944, 4121943, 3750154, 3823850, 3750165, 3796827, 3673223, 3611999, 3832822, 3953208, 3832808, 4193493, 3823238, 3823757, 3832806, 3832829, 3832799, 3670568, 3876276, 3876298, 3832807, 3823755, 4005416, 4142074, 4141981, 3670615, 2812249, 3893513, 3893520, 3876275, 3832826, 3932412, 3961558, 3961584, 3961571, 3932408, 3953180, 4020705, 4032024, 3953190, 3985521, 4032031, 4165427, 3953179, 3953187, 3961574, 4175285, 3750147, 3611997, 3673231, 3823211, 3823746, 3832802, 3832804, 3876269, 3832837, 3893524, 3953172, 3823856, 4049195, 4020719, 3750156, 4142054, 4108455, 3832836, 3961581, 3832819, 3756681, 3823766, 3756684, 3953209, 3953202, 4159917, 3823242, 4142050, 3893510, 3893527, 3670630, 4020713, 3953212, 3670622, 4121948, 4020704, 3610594, 3670632, 4149761, 3670563, 3564209, 3985520, 3985525, 3932413, 3670555, 3796825, 3850697, 3823749, 3850709, 3610591, 3796823, 3750161, 3823239, 3564186, 4005412, 4108462, 4091538, 4165429, 2873859, 3756682, 3673214, 3893512, 3750155, 4121950, 4159921, 2951919, 3823863, 3796821, 3961579, 3670614, 2777543, 3796817, 2979119, 2979140, 3673129, 3670647, 3564191, 3610590, 4193498, 4193476, 3953195, 3673133, 3953182, 3823207, 3876271, 3670705, 2951923, 3724439, 3876272, 2770127, 3673134, 3893505, 3953214, 2689079, 3823857, 3985517, 3823206, 4005414, 4091541, 3673145, 3724434, 2873857, 3832803, 3850713, 2920013, 3670708, 3832798, 3953169, 4091547, 4142065, 4142073, 4142068, 3577275, 2919993, 4159920, 3823861, 4020716, 3670569, 3577265, 4142078, 4142080, 3750159, 3823214, 4159915, 3776234, 3796819, 3796816, 2777525, 3823208, 3673208, 3850711, 3876288, 3823853, 4005418, 3876270, 4020712, 3823864, 3611317, 2770146, 3932411, 3832815, 3564199, 3670551, 3796828, 2700790, 3823249, 3776242, 3611300, 3724448, 3832820, 4032029, 4142063, 4175268, 3577245, 3670698, 4005408, 2897370, 2951958, 3850710, 3823747, 4075929, 2700797, 3756688, 3670645, 3756683, 3823237, 3832809, 3832810, 3823231, 3850694, 3796820, 3953185, 3953196, 4092809, 3876261, 4108459, 3893519, 3670575, 3850700, 4020706, 3611299, 3796829, 3776224, 3832812, 3823745, 3850712, 3673232, 3611314, 3823858, 3823251, 3893526, 3876283, 3876265, 3823748, 3953215, 3876258, 3953216, 3953192, 3876260, 4005417, 3953173, 4020708, 3756679, 4020718, 4041867, 4149766, 4142077, 3953168, 4159918, 3796824, 3673121, 3876294, 2770169, 3823859, 3876254, 3673126, 4141992, 3776236, 3750144, 3893504, 3961556, 3796809, 3823221, 3673118, 3850696, 4005401, 4005402, 3823765, 2770171, 3832801, 4020709, 2812246, 4141989, 3673143, 3823865, 3832838, 3670571, 3673142, 3796822, 3876292, 3823871, 3796826, 3876286, 4005421, 3670657, 3832828, 3953181, 3953188, 3893528, 4066131, 3756678, 3611981, 3670621, 4149760, 4141991, 4175278, 4193486, 3876262, 3893515, 3876274, 4005407, 4193488, 3893518, 3953199, 2770170, 3776240, 2777536, 3796815, 3823241, 3796818, 3876268, 3670566, 2920023, 3832824, 3823751, 3670722, 3673148, 4005409, 3670716, 2919996, 4049184, 3673155, 3823218, 3823224, 3953186, 3670558, 3796811, 3850716, 3823762, 3823851, 4121938, 4059216, 3724451, 3673124, 3876280, 3893508, 3953184, 3670617, 3823870, 4005415, 4049197, 4049194, 4099791, 4020697, 3670619, 3670638, 3823866, 3823259, 3611983, 3673224, 3670623, 3876266, 3724453, 3670627, 3796832, 4005419, 3876281, 3776238, 3865816, 3925843, 3925839, 4032025, 4149754, 4149752, 4149753, 4171143, 3796814, 3564196, 3865814, 3776244, 2951912, 3670597, 3865811, 3865810, 3865812, 3865813, 3865815, 3925840, 3925842, 3925841, 4032022, 4149751, 3611303, 4171142, 3823260, 3876302, 4108450, 3823860, 3961572, 3961583, 3876287, 3610571, 2951948, 3670643, 4020717, 3670706, 3823233, 3611296, 2979112, 3670557, 3832818, 4020696, 4142062, 3670699, 3673230, 3796812, 2770134, 2770145, 2770449, 3823209, 3823769, 3823258, 3823226, 3832805, 3823228, 3832834, 3876299, 3953174, 3832823, 3961580, 3961576, 4091546, 3823243, 4091540, 4121954, 4141983, 4141999, 3832797, 2873855, 2951959, 3670561, 2951925, 3673119, 3876277, 3876279, 3724436, 3610584, 3953204, 2873860, 3610587, 3893522, 3776223, 3611982, 3670588, 3673220, 3670710, 3673120, 3850717, 3750146, 2777533, 3876301, 3985522, 3876264, 4103646, 2777469, 3673116, 4121956, 2951949, 2698258, 3876259, 3670646, 3756675, 2920019, 3673216, 3564214, 3961570, 4049191, 4121939, 3823761, 3823250, 3670587, 3577272, 3673146, 3673153, 3673144, 3673149, 3893507, 3850699, 3992036, 3992037, 3992039, 3992041, 3992038, 3992040, 3992042, 4005422, 4005423, 3673157, 3750142, 3750148, 3750151, 3823254, 3823256, 3832827, 3823220, 3823253, 3823216, 3610570, 2951963, 3577246, 2665876, 2665867, 4141995, 4379759, 4439095, 4334396, 4175282, 4122641, 4121951, 4439093, 4165422, 4165425, 4266688, 4505795, 4175277, 3876303, 4379760, 4505792, 4128069, 4257032, 4142069, 4322384, 4266711, 4266710, 4142052, 4216342, 4121940, 4322374, 4005403, 4439087, 4049198, 3876293, 4334408, 3893511, 4439132, 4322412, 4266715, 4280858, 4322400, 4280860, 4175273, 2700806, 4242434, 4379750, 4334369, 4505793, 4509457, 4091544, 4439080, 4049186, 4165424, 4266713, 4142006, 4257047, 4280854, 4439100, 4439099, 4141982, 4175287, 4122644, 4334407, 4228068, 4257031, 3961582, 4266704, 4266703, 4298102, 4505721, 4334349, 4505776, 4257049, 4322401, 4280869, 4334395, 4266677, 4439120, 4334350, 4439076, 4334364, 4334361, 4091535, 4266696, 4108461, 4149763, 4108452, 4228071, 4257041, 4266716, 4298091, 4257037, 4439130, 4509456, 4439127, 4228062, 4193487, 4238981, 4142075, 4334414, 4257038, 4216322, 3823232, 4228073, 4257036, 4141998, 3673203, 4141980, 3673137, 3776232, 4439129, 4020707, 4439098, 4505781, 4018024, 4216329, 4266683, 4322376, 4298100, 4216338, 4505727, 4242425, 4509455, 4091545, 4216323, 4322382, 4322370, 4379757, 4238977, 4142051, 4439060, 4505780, 4242424, 4439074, 4509448, 4439094, 4439055, 4224165, 4193497, 4091548, 4257042, 4266667, 4280868, 4298101, 4322415, 4238978, 4228060, 4322419, 4159916, 4141994, 4242436, 4322386, 4257048, 4216333, 4020714, 4193474, 4439083, 4439086, 3893521, 4228058, 3611294, 4507304, 4049192, 4242439, 4238968, 4322380, 4224332, 4242430, 4097240, 4439126, 4439096, 3961555, 4121959, 4242447, 4242423, 4322396, 4175291, 4238985, 3953176, 4142072, 4216349, 4379764, 4266676, 4379763, 4075925, 4379749, 4322417, 4439078, 3961567, 4334353, 4439116, 4128071, 4334358, 2951968, 2897369, 4439142, 4509445, 4141985, 4224169, 3670549, 3823862, 3876257, 4193490, 4121949, 3611287, 3823230, 3823868, 3893525, 3832817, 3961562, 4032032, 3850704, 3985519, 4238979, 4165421, 4149762, 4175271, 4505798, 3610588, 3756673, 3876300, 3961577, 4322414, 4193480, 4159922, 4068814, 3953170, 4266687, 4091542, 4242449, 4257057, 4266706, 4266697, 4439134, 4159925, 4322397, 4509450, 4242440, 4439085, 4091821, 4159924, 4242443, 4216339, 4128070, 4128072, 4505732, 4216335, 4032026, 3823225, 4298098, 4175269, 4379745, 4379751, 4216325, 3832814, 4439053, 4280857, 4266668, 4142071, 4322408, 4159919, 4505791, 3823236, 4193479, 4193477, 4242427, 4242426, 4141993, 4242419, 4257059, 4242438, 4439125, 4142083, 4242448, 4153685, 4257050, 4298095, 4142059, 4439122, 4193485, 4280862, 4142082, 4032028, 2689078, 3611297, 4054940, 4242446, 4439124, 4334366, 4505711, 4266699, 3893516, 3893503, 4175279, 3823219, 4005420, 4121955, 4175272, 4280863, 3850693, 4505712, 3876291, 4322402, 4063674, 4193489, 4142064, 4238974, 4193499, 4193495, 4224334, 4266671, 4334415, 4334418, 4439091, 4257058, 4439113, 4439103, 4439107, 4439075, 4505783, 4509449, 4509458, 4224338, 4257035, 3953183, 3724437, 4020715, 4439135, 4505719, 4334355, 4149755, 4216336, 4242442, 4280855, 4257051, 4298097, 4280865, 4322404, 4334397, 4193471, 4379762, 4379743, 4216347, 4439057, 4121946, 4439066, 4439069, 4505720, 4505799, 4509451, 4298096, 4509454, 4216341, 3961564, 4280853, 4322423, 4149758, 4334403, 3850708, 3876295, 4266682, 4439068, 4141987, 4020698, 4216340, 4238971, 3953210, 4266695, 4266678, 4322409, 4193478, 3823763, 3961569, 3796830, 4020711, 4280864, 3876273, 4322390, 4505714, 4032030, 4322395, 4334354, 4216330, 4066128, 4242429, 4091537, 3953211, 4238992, 3823874, 4193475, 4238976, 2770144, 4266669, 4322398, 4334401, 4142000, 4242437, 4142002, 4280867, 4334421, 4149756, 4005413, 3756687, 4216346, 4238987, 3564206, 4257034, 4298099, 4334367, 4334419, 4334356, 4379755, 4507302, 4507299, 4242432, 4505726, 4507298, 4266675, 4509441, 4091539, 4128067, 2951951, 4141984, 4266690, 4266670, 4049193, 3850705, 4439118, 4298106, 4507307, 4175289, 4322388, 3724435, 3611336, 3673127, 3670691, 3670689, 4379761, 4379758, 4505788, 3673226, 4049199, 4439059, 4322410, 4266698, 4175275, 4121937, 4005399, 4228074, 4175292, 4175267, 4099793, 4185973, 4257043, 4242444, 4257039, 4322418, 4334371, 3823855, 4121936, 4322378, 4507318, 4334410, 4228057, 4266672, 4228075, 3670626, 4505729, 3876278, 4322420, 4439117, 4439088, 4228063, 4334412, 4439141, 4334416, 3776231, 4049201, 4175284, 4379756, 4439150, 4216321, 4216337, 4216324, 4228064, 4228066, 4228061, 4228072, 4216345, 4334352, 4266701, 4149759, 4334400, 3823852, 4193473, 4216331, 4439119, 4334392, 4439146, 4505725, 4505715, 4322383, 4334373, 4505785, 4334394, 4507295, 4128066, 4509443, 4175281, 4175286, 4242431, 4266693, 4091549, 4165423, 3673229, 4020694, 4068815, 4121945, 4322407, 4439143, 3893523, 4439114, 4280852, 4228070, 4020695, 3673227, 4238980, 3823854, 4175276, 2777509, 4216334, 4507296, 4041865, 4334417, 4142067, 4142058, 4121941, 3832800, 4238975, 4193483, 4063675, 4280861, 4298105, 4322403, 4379746, 4439071, 4439115, 4257044, 3953171, 4142053, 4141988, 2770454, 4238982, 4257033, 3876285, 3953194, 4257046, 3776241, 4224333, 4232366, 4232364, 4439123, 4266708, 3876255, 4439090, 4439092, 4142003, 4334402, 4238969, 4018023, 3823257, 2777526, 4075924, 4439077, 3953197, 4091543, 3670610, 4041864, 4041862, 4439147, 4439148, 4439128, 4439108, 4505728, 4142055, 3953213, 3610578, 4224330, 3985518, 4216344, 4322422, 4505789, 4175283, 4334406, 4238989, 3670613, 4439058, 4049190, 4322406, 4121935, 2897388, 4266685, 4142057, 4149764, 4322379, 4266700, 4099792, 3776229, 4505787, 3776243, 4322399, 4216332, 4509461, 4322377, 4322373, 4224167, 3876263, 4298090, 4298094, 4322416, 4322372, 4322421, 3750157, 3961578, 3961559, 3961566, 4334363, 4334393, 3673199, 4238986, 4224163, 4224161, 4193494, 4238973, 4224162, 3673161, 3670633, 4175270, 4224164, 4238991, 4661123, 4627082, 4661078, 4661125, 4627136, 4627151, 4628785, 4627132, 4627107, 4627123, 4661090, 4628781, 4627118, 4627069, 4628775, 4628780, 4627122, 4628787, 4627081, 4627065, 4661079, 4627074, 4627110, 4627127, 4627142, 4627130, 4627111, 4627095, 4627102, 4661096, 4627047, 4627050, 4634821, 4627113, 4627134, 4627112, 4627063, 4627079, 4628778, 4627039, 4627104, 4628786, 4627108, 4627124, 4627067, 4628769, 4627099, 4661093, 2665883, 2546873, 3670719, 3670584, 3670718, 4121947, 4257056, 4439110, 4439145, 4216348, 4108456, 4280859, 4509440, 4142001, 4142005, 2920003, 4165426, 4020702, 4439097, 4232365, 4322394, 4242422, 4108460, 4049185, 4066130, 4149757, 4005404, 4507312, 4128068, 4298093, 4266702, 2979118, 4175266, 3961561, 4322385, 3832796, 4266707, 2812238, 4216343, 4334362, 4242445, 3673141, 4439065, 4505724, 4228069, 3823213, 4507313, 4509459, 3823235, 4334368, 3985526, 2920028, 4322389, 3932409, 4322387, 3673228, 4128073, 4193492, 4334405, 4185974, 4238990, 4509446, 4142060, 4257054, 3953198, 4379748, 4224170, 3823758, 4108454, 4507311, 4298088, 4334359, 3876297, 4439079, 4193482, 3670651, 2812250, 4505708, 3850714, 4322426, 4298089, 4439112, 2770148, 4322381, 4075926, 4334360, 4439072, 4257053, 4505777, 4334370, 4439109, 4298104, 4505778, 3673218, 3832832, 4266681, 4075928, 4266686, 3985529, 4141997, 4266673, 3776225, 4266714, 3832821, 4242420, 4379752, 4505794, 4334413, 4121958, 3985530, 4509460, 4238988, 3756686, 3953201, 4505731, 4507294, 3610576, 4298103, 3724456, 4507305, 4121957, 2689263, 4505797, 4216326, 4505782, 3876290, 4063673, 4439139, 4121942, 4439064, 4216327, 4242421, 4280856, 4193470, 3823756, 4439073, 3876296, 4505716, 4334351, 4509452, 4238970, 4266684, 3953178, 4142081, 4257052, 4439151, 4509442, 4505779, 4505796, 4020700, 3961560, 3611293, 4322413, 4216328, 4334420, 4379753, 4507297, 4334404, 4266674, 4507301, 4505775, 3953177, 3985531, 4005411, 4108451, 4509444, 3823255, 4142066, 4509447, 4439104, 4122643, 4159923, 4257045, 4505730, 3876256, 4505710, 4507317, 2689262, 4228065, 4238984, 4439105, 4175280, 3961568, 4193472, 4334399, 4439102, 3750150, 3673140, 3750152, 4934158, 5121497, 4627060, 4768185, 4934159, 4934139, 4795008, 4628777, 4721841, 4916585, 4661124, 4721834, 4661149, 4794992, 4729874, 4836698, 4768148, 4794997, 4898398, 4795001, 4627094, 4729891, 4990303, 4724883, 4627139, 4826820, 4729865, 4810504, 4956335, 4627144, 4795009, 4889609, 4956341, 4898370, 4889620, 4826844, 5030272, 4916594, 4628770, 4661098, 4661108, 4729849, 4729862, 4836720, 4836712, 4634816, 4729858, 4724887, 4627098, 4916603, 4810494, 4768151, 4661111, 5094438, 4977018, 4795007, 4826840, 4889618, 4810485, 4627042, 5030277, 4627046, 4810487, 4826827, 4661134, 4634817, 4729883, 4724884, 4755080, 4627149, 4810505, 4990315, 4916604, 5121481, 4856675, 5163013, 4898397, 4836718, 4634824, 5030280, 4661097, 4768189, 4836721, 4889607, 4729860, 4729889, 4729884, 4721840, 4729888, 5040012, 4627087, 4627126, 4755082, 4956338, 4627083, 4729863, 4627097, 5040016, 4729892, 4916591, 4898376, 4916595, 4729867, 5087153, 4729885, 4810477, 5039991, 4898389, 4628765, 4627160, 4661109, 4795015, 4729882, 4862012, 4916592, 4977036, 5030273, 4990324, 4916576, 5030290, 4661118, 4729890, 4977023, 4627147, 4634809, 4661122, 4729897, 4889598, 4628796, 4627053, 5121499, 4628772, 4627061, 4661163, 4916577, 4898385, 4956332, 4836709, 4898374, 4661107, 4627135, 4724885, 4729853, 4768154, 4768174, 4826828, 4795005, 4934143, 4990319, 4794989, 4627106, 4661128, 4661161, 4729877, 4826825, 4990312, 4768179, 4934160, 4628774, 4721827, 4661154, 4856651, 4898383, 4856656, 4768170, 4826819, 4862020, 4990323, 5040004, 5152509, 4721836, 4889600, 4898373, 4721830, 4661144, 4862016, 4661085, 4836713, 4628779, 4721853, 4729859, 4836703, 4627138, 4729876, 4956325, 4898399, 4627140, 4810481, 4856673, 4934147, 4916578, 4768187, 4889597, 4627048, 4794999, 5163015, 4627089, 4729894, 4889615, 4898387, 5087144, 5094436, 4856665, 4627073, 4795002, 4634812, 4628782, 4810493, 4721848, 4856669, 4810496, 4956339, 4862015, 4934161, 4836699, 4956337, 4634813, 4628792, 4810484, 5121491, 4634810, 4934152, 5040003, 5152506, 4721846, 4627070, 4634825, 4755089, 4729851, 4661112.4661158, 4729875, 5121472, 4795003, 4661101, 5030293, 5152498, 4729864, 4628764, 4768190, 4856649, 4661076, 4628752, 4627150, 4661155, 4661080, 4768153, 4990302, 5087138, 4627115, 4856646, 4627161, 4826829, 4934149, 4628799, 4627040, 4856647, 4661100, 5040015, 4729887, 4755087, 4661095, 4729880, 4721839, 4898388, 4862006, 4856674, 4627137, 4977027, 4990316, 4956328, 5094445, 4661137, 4627085, 4627154, 4729881, 4634815, 4755104, 4755094, 4826843, 4889595, 4898386, 4916602, 4836717, 4627076, 4661094, 4729872, 4627064, 4768193, 4898377, 4826815, 4729895, 4889599, 5087140, 5030269, 4627159, 4934146, 5121492, 4916587, 4627057, 4721851, 4934145, 4628771, 4627084, 4826841, 4661082, 4661140, 4721833, 4661162, 4661089, 4755084, 4768171, 4755086, 4768152, 4836706, 4810489, 4810501, 4856671, 4836694, 4977015, 5094450, 5087159, 4768145, 4889602, 4898393, 4862030, 5165433, 4661120, 4627052, 4661083, 4810498, 4627049, 4755083, 5030282, 4634823, 4627114, 4661092, 4627131, 4661088, 4729857, 4729870, 4768156, 4627155, 4627054, 4768172, 4862004, 4627153, 5094444, 4627044, 4661132, 4634814, 4628758, 4724876, 4627092, 4826824, 4889605, 5030265, 5121486, 4810483, 4628776, 4768191, 4768188, 5121483, 4661104, 4755091, 4810476, 4795013, 4810488, 4898395, 4836722, 4934165, 4836723, 4977012, 4661143, 4768173, 4634811, 4721854, 4810491, 4628790, 4724869, 4724874, 4724881, 4856654, 4990310, 4898392, 4627093, 4724882, 4661145, 4768175, 4768186, 4934144, 4889622, 4768169, 4661116, 4661150, 4729893, 4956326, 4794995, 5039995, 4956330, 4826836, 4661130, 5030288, 4634820, 5121490, 5094433, 4627096, 5094448, 4990298, 4627100, 4977024, 4934151, 5094430, 4794990, 4634818, 5152495, 4794998, 4916597, 4661114, 4724877, 4755098, 4627068, 4768177, 4795017, 4627105, 4889619, 5040000, 4627078, 4661113, 4628767, 4627146, 4721828, 4661126, 4627101, 4661153, 4724878, 4628757, 4634807, 4889617, 4794988, 5121496, 4856659, 4661117, 4634826, 4628754, 4627059, 4916582, 4768157, 4862007, 4768178, 5087141, 4768192, 4627117, 4627148, 4628760, 4661136, 4661147, 4661152, 4661156, 4729852, 4755092, 4628798, 4721843, 5040008, 4724888, 4661165, 4721832, 4990296, 4826837, 4810478, 4627133, 4627141, 4724868, 4862026, 4661151, 4627158, 4724866, 4628753, 4721847, 4628761, 5087136, 4956334, 4977010, 4627056, 4661129, 4755088, 4628759, 4755101, 4826816, 4810495, 4956346, 4810482, 5094431, 4977011, 4990308, 5152494, 4826818, 4755100, 4794993, 4627157, 4768150, 4661081, 4634808, 4724880, 4729866, 4889613, 4627072, 4628795, 4729856, 4661142, 4990321, 4634822, 4755102, 4661121, 4661115, 4862018, 4627103, 4661148, 4724864, 4862014, 4889594, 4977016, 4768162, 4755097, 4856648, 4856666, 4889603, 4889608, 5094434, 4628755, 4856657, 4768166, 4724872, 4862031, 4439149, 4266692, 4266680, 4232367, 3750145, 4439101, 4193491, 3577274, 4298092, 4439082, 4505717, 4103644, 4439054, 4439136, 4175265, 4439144, 4439133, 4242433, 4149765, 4379747, 3670567, 4108463, 4142056, 4224337, 4507315, 4334357, 4507306, 3823246, 4032023, 2897364, 4041863, 3953206, 4507303, 4439067, 4242435, 4224336, 3564192, 4175290, 4507308, 4266679, 2951921, 4238972, 4439138, 3850698, 4228059, 4141990, 3823248, 3850701, 4242441, 4505722, 4108457, 3776235, 4228067, 2770442, 4439121, 4322375, 4228076, 4108449, 4322369, 3823210, 3850715, 4509453, 4334398, 4224166, 3893529, 4224168, 3611295, 4175288, 4856668, 5414788, 4990304, 5521096, 5256705, 5461774, 4721826, 5236863, 5414794, 4836705, 5311904, 5328382, 4755103, 4627055, 5448304, 5121485, 4661099, 5216645, 5087148, 5254397, 5254374, 4627120, 4856650, 5210052, 5040013, 5087157, 5233254, 5163023, 5163024, 5165448, 4810499, 5328514, 5233258, 5087165, 4794996, 4836711, 5163011, 5176705, 5152511, 5163006, 5236853, 4956344, 5359856, 4956347, 5256723, 5345090, 5152499, 5233264, 5359850, 5216652, 5087152, 5030271, 5087158, 5328381, 5359851, 5433381, 5152514, 4768176, 4836701, 5233259, 5256712, 5311896, 5176725, 5216641, 5152504, 5210056, 5414810, 4856660, 5328519, 4768160, 4628788, 4661133, 5414804, 4836714, 4826817, 5176719, 5163016, 4977039, 4956327, 4990309, 4956342, 5414792, 5328513, 5163021, 4795012, 5216642, 5359832, 5461797, 4721831, 5233267, 4627086, 5233240, 5256731, 5254404, 5345099, 5256746, 5433375, 5414815, 5254370, 5094447, 5319312, 4862011, 5165438, 5176720, 5254372, 5359860, 5121487, 5163005, 4862024, 4934163, 4889614, 4977021, 4862021, 5216637, 5256726, 5254384, 5254400, 5121477, 5328500, 4990297, 5030286, 5176716, 4661138, 4836700, 4810480, 5328380, 4836715, 4661131, 4977026, 5256722, 5345120, 5256711, 5319304, 5152512, 5448298, 5256734, 5152513, 5165449, 5233255, 5328496, 5448310, 5165461, 5521081, 4934157, 5176724, 4627090, 4862029, 5233252, 4916584, 5254371, 5176721, 5328378, 5359829, 5414798, 5521068, 5359858, 5165459, 5254411, 5216653, 5233268, 4916596, 5087147, 5328401, 5414796, 5345109, 5461790, 5236878, 5521115, 5236868, 4862017, 5162997, 5328377, 5254398, 5448300, 5433365, 5521071, 5165451, 5311910, 5345104, 4934156, 5311907, 4721844, 5216655, 5448287, 5319323, 4628797, 5121489, 4934164, 5087154, 5256725, 5414811, 5328495, 5165457, 5254392, 5328506, 4916581, 4898371, 5176692, 5176706, 4977025, 5210047, 5163007, 4862023, 5233256, 4628793, 5210068, 4934148, 4755085, 5094437, 5165460, 5040002, 5311903, 5256748, 5345105, 4826830, 4934153, 4990320, 5254403, 5163019, 5210060, 5236859, 4795006, 5152500, 5461796, 5256735, 5521109, 5328396, 4836696, 5448294, 5165456, 5319308, 5359827, 5165447, 5328391, 5345088, 5236861, 5254407, 5165436, 4729896, 4628763, 5461781, 5319325, 5521090, 4810503, 5176703, 5345114, 5521113, 5414805, 4661157, 5319306, 5521103, 5311891, 5233249, 4862022, 5359847, 5163022, 5254389, 5210057, 4627080, 4990313, 5256738, 5152508, 4755090, 4898390, 4862009, 5341787, 4977017, 5176702, 5341764, 5210074, 4768155, 5433374, 5433379, 5254399, 5121495, 5165454, 5094453, 5256707, 5319313, 4661164, 4856662, 4956329, 5152491, 5121480, 4627077, 5216648, 5236851, 5328518, 5414784, 4724865, 4856663, 5152492, 4661146, 4628751, 5256714, 5030294, 5311900, 5311911, 5087145, 5414787, 5414797, 5311892, 5176697, 5345112, 5165440, 5521095, 5254412, 5359839, 5087137, 5311917, 5254391, 5433372, 5210048, 5311897, 5165458, 5165432, 5236864, 4934154, 5521086, 4826832, 5341780, 5039989, 5311902, 5359843, 4729855, 5328386, 5087150, 5319330, 5254402, 5341765, 5040018, 5254405, 5176704, 5256749, 5254416, 5256727, 5254375, 5359859, 5341766, 5121478, 4916586, 4627128, 4755081, 4724875, 5216646, 5210051, 5236860, 5433376, 4898384, 5256736, 5256718, 5256743, 4795011, 5319319, 5094442, 4627041, 4729848, 5461785, 5216661, 5233260, 5254396, 4889616, 5163018, 4634819, 4916593, 5176693, 5210059, 5233245, 5040017, 5256747, 5448291, 5328397, 5461789, 5256706, 5162998, 5152489, 5030279, 5328393, 5328387, 5461791, 4729868, 5210071, 5345124, 5176712, 4810500, 5328376, 4627143, 5359845, 4721849, 5210066, 5319314, 5162996, 5094432, 5341759, 5094440, 5433386, 5094428, 5433361, 4934141, 5521110, 4956350, 4862003, 5521098, 5121473, 4889621, 5236873, 4990301, 5216663, 5216660, 5328375, 5414795, 4956349, 5254383, 5359852, 5345123, 5521106, 5521073, 5521111, 4956331, 5433377, 5236856, 5163000, 5328383, 5521087, 5163010, 5359854, 4729869, 5311906, 5521099, 5256709, 4934140, 5121471, 5176718, 5256708, 5236875, 5328400, 5414802, 5254393, 4956352, 5163002, 4956348, 5152493, 5094441, 5341775, 5254386, 5328388, 5341783, 5328511, 5216658, 4628762, 5152488, 5176695, 5152490, 5176696, 5233244, 5176727, 5121482, 5210070, 5254395, 4836710, 4977038, 5176710, 4628768, 5163012, 5040014, 5254377, 5176715, 4956340, 5319301, 5448312, 4826831, 4768184, 5040010, 4768182, 5328402, 5414800, 5328505, 5319307, 4990314, 5030289, 4795004, 5521104, 4916580, 5094425, 5210061, 5345107, 4977013, 4826823, 4956345, 4889604, 4990306, 5121500, 5030291, 5094439, 4627075, 5165443, 5236869, 5210069, 5216643, 5254382, 5256713, 5359849, 5319324, 5311914, 5433383, 5521085, 5319310, 5448285, 5359836, 5094426, 4934150, 5094443, 5254378, 5461773, 5094452, 4810486, 5030287, 4627066, 5236872, 5256745, 5319322, 5328499, 4729873, 4661110, 4916598, 5094451, 4934142, 4721838, 5521093, 4889612, 5239532, 4755095, 5319315, 5521107, 5039990, 5328389, 5165444, 5233251, 5210067, 5414789, 5121484, 4898394, 5163004, 5341788, 5345102, 5087149, 4856664, 5448309, 5328508, 4768194, 5162999, 5165450, 5448295, 5341773, 4794991, 5163001, 5087143, 5254413, 4768168, 4795000, 5165437, 4628750, 5030278, 4768161, 4661105, 5039994, 4627062, 5152515, 5319326, 5433371, 5311905, 5319309, 5328385, 5461776, 5256741, 5087163, 4628766, 5236880, 5087164, 5345113, 4977022, 4898378, 5121498, 5030276, 4916579, 5030285, 5433357, 4916583, 5328502, 5030274, 5569333, 5644409, 5579204, 5569345, 5569338, 5602794, 5569329, 5602791, 5573400, 5602781, 5573412, 5579209, 5569332, 5579194, 5579192, 5569328, 5604871, 5579216, 5579199, 5569344, 5644418, 5602775, 5569350, 5569323, 5602785, 5604877, 5573399, 5569324, 5579206, 5644413, 5644433, 5604873, 5644414, 5569342, 5579212, 5604869, 5602783, 5644416, 5581220, 5644421, 5569337, 5573397, 4439061, 4322424, 4507310, 4238983, 4266694, 4509439, 3850703, 4103645, 4505713, 4505786, 4020710, 3823234, 4266709, 4142079, 4334372, 4224331, 4165428, 4142004, 4322405, 4005400, 4266691, 4257055, 4322392, 4334409, 4142076, 4049187, 4505709, 3823227, 4439062, 3823867, 3823223, 3893514, 3670581, 4661077, 5210053, 5341758, 5319311, 5311908, 4934155, 4661084, 5319318, 5414807, 5414791, 5094424, 5233248, 5233262, 5345117, 4768159, 4836707, 4856658, 5233266, 5039996, 5256717, 5254409, 4810497, 5345100, 4768163, 5328521, 5256710, 5341771, 4862028, 5039999, 5359855, 4661091, 4724873, 4956351, 5359826, 5328379, 5521114, 5328395, 5319328, 5216638, 5165435, 5448297, 5256729, 4990299, 5162995, 5341786, 5328384, 5311909, 5039993, 5341768, 5152510, 5216659, 5233241, 5176723, 5216639, 4889610, 5152505, 4990300, 4795016, 5448308, 5341777, 4721835, 4661159, 4898375, 5448289, 5176729, 5341785, 5521060, 5254390, 4990317, 5359830, 5254385, 5433368, 5433370, 4977019, 4627109, 5256732, 5039997, 5414790, 5414808, 5414812, 5328398, 5328503, 4628773, 5345095, 4977032, 5254410, 5121493, 5521094, 4795014, 4628791, 5345111, 5152497, 5414801, 4898382, 5165453, 5165441, 4856653, 5345092, 5030266, 5254381, 5040005, 5448305, 5341789, 5521057, 5210073, 5328512, 5341782, 5152507, 4862013, 5328510, 5236866, 4627071, 5254394, 5176713, 5433358, 4661141, 5319329, 5176726, 5341763, 5163009, 5176700, 5433385, 5461784, 5341761, 5433373, 5216635, 5165445, 5521105, 5087151, 5254379, 5328509, 4661086, 4836719, 5256728, 5210050, 5176708, 5359840, 5521092, 5328392, 5311894, 5433364, 5040009, 4898380, 4627116, 5341762, 5319327, 4826842, 4856670, 5359844, 5433384, 4755096, 5345098, 5256742, 4724879, 5163020, 5521089, 5216664, 5236870, 5254414, 5414799, 5359834, 5448288, 5448293, 4627129, 4721837, 5521078, 5152502, 5233246, 5210065, 4721842, 5341794, 5094449, 4768180, 4826821, 4836695, 4810492, 5236858, 5236867, 5216651, 4724867, 5461778, 5311901, 4810490, 5233247, 4934138, 4990325, 5359837, 5311898, 5256716, 5256740, 5094435, 5311915, 5521076, 5448284, 5345121, 4916599, 4990307, 4795010, 5256715, 4721829, 5040001, 5521082, 5345119, 5254408, 5461780, 5236877, 5461788, 5039992, 4916589, 5121488, 5521080, 5087146, 5341790, 5328507, 5461795, 5433367, 5359853, 5359842, 4862008, 4768181, 5152501, 5359857, 5233242, 5254373, 5216662, 5254388, 5165452, 4729850, 4916575, 5341776, 4889601, 4768149, 4898381, 4729871, 4661160, 5176699, 5414813, 5433366, 4916600, 5216650, 4661087, 5341793, 5165434, 5345096, 4862010, 4956343, 5345091, 5163008, 5256744, 5341791, 5345118, 5956420, 5604882, 5845265, 5579215, 5956419, 5956410, 5956433, 5998753, 6003067, 5573392, 5809382, 5845237, 6003087, 6178171, 5604891, 5604883, 5573387, 6080925, 5579197, 6088019, 6151798, 5579205, 5602779, 5581204, 5845235, 6003137, 6003152, 6003098, 6094922, 5998749, 5845264, 5602382, 6110987, 6269477, 5998750, 5569334, 5998864, 5644410, 5804569, 5913048, 5998756, 6003079, 6003147, 6087977, 5573395, 5998745, 6079051, 5956417, 5604870, 5602376, 6156669, 5809394, 5951180, 6151794, 5998741, 5845242, 5845257, 5998767, 6079049, 5644426, 6045154, 5809372, 5604868, 5573404, 5604878, 5845227, 6151789, 6151777, 6003104, 6003102, 5804567, 5804555, 5845249, 5956432, 5579202, 5644412, 5581227, 5604867, 6003105, 5845250, 5602378, 5569351, 5602370, 5581209, 5956444, 5804564, 6045171, 5956443, 6003111, 5804560, 5644424, 6003150, 5998744, 5998752, 5913042, 5809377, 5956408, 5956437, 5998755, 6003117, 6087991, 6079041, 6238017, 6178176, 6003134, 5602381, 6003146, 6003103, 5604876, 5845239, 6003153, 5569331, 6080897, 5951175, 6087988, 5998746, 6094935, 6079039, 6094915, 6003155, 6254860, 5951183, 5956427, 5573411, 6003127, 5581200, 5581226, 5804557, 5604893, 5897766, 5845262, 6087997, 6238016, 5604874, 5845233, 5604892, 5644408, 5602390, 5809378, 6079042, 6079063, 5809391, 5644420, 5804561, 6003068, 5573394, 5644431, 6227812, 6151800, 5569343, 6080921, 5581203, 5845241, 5581202, 5804554, 5602379, 5569348, 5804563, 6227811, 5913045, 6151802, 5573403, 5581210, 5581214, 5602786, 5569336, 5845258, 6178152, 6003110, 5579211, 5804542, 5602787, 5581229, 5569347, 6003142, 6087998, 5845260, 5602793, 5956416, 6045151, 6156656, 5998740, 5951178, 5602767, 5804568, 5602792, 5809371, 5845253, 5913046, 5956439, 5956422, 6094929, 5951172, 5809373, 5998748, 5998738, 5998870, 6003095, 6003107, 5998765, 6003081, 6003072, 6079037, 6088000, 6094934, 6003122, 5602367, 5845240, 5602386, 6079058, 6251523, 5579208, 6088011, 5579193, 6178170, 5998754, 5602778, 5809374, 5897768, 5804546, 5951176, 5913047, 6110973, 6110971, 6003065, 6178164, 6227809, 6238043, 5579219, 5602365, 5845247, 5998872, 5897769, 6087981, 5602784, 6094940, 5845252, 6003077, 5604887, 6178156, 5573396, 5602385, 5804556, 5581217, 5809375, 5956426, 6110992, 5845244, 5602769, 5804550, 6151781, 5573390, 6045163, 5845266, 5956414, 6003090, 5998862, 5913043, 6094919, 6003093, 5579201, 5804549, 5998857, 6045170, 6079052, 6094924, 5602384, 6156682, 6227819, 6178175, 6254858, 5809390, 5602773, 6003069, 6156659, 6079046, 6238037, 5644411, 5604885, 5602362, 6045148, 5804559, 5956415, 5998742, 6088010, 6094914, 6094916, 6003076, 5579195, 5998845, 6003123, 5804553, 5809383, 6003078, 6003085, 5602776, 5573393, 5602371, 6254864, 5602364, 5581221, 5602363, 5602374, 5569327, 6079038, 5581208, 5581207, 5845228, 6003071, 6003126, 6079034, 5573401, 5602380, 5604880, 5569340, 5573402, 5581223, 5581225, 6079059, 5602373, 6156676, 5897765, 5804570, 5569339, 5644427, 6045156, 5602383, 5581201, 5644425, 5602766, 5845263, 5804565, 5804545, 5998849, 6003084, 6254862, 6003066, 6003109, 6003091, 5604895, 5956428, 5845254, 5845231, 5804547, 5579218, 5602391, 6003151, 5913044, 5804558, 6094920, 6045169, 6094927, 5602780, 6110991, 5845255, 5845256, 5998847, 5998860, 5998843, 5998854, 6003128, 6080908, 6080913, 6156667, 6156674, 6269474, 5845232, 6178177, 6178180, 6227805, 5809380, 5809393, 6003140, 6110988, 5573413, 6003074, 5845259, 5644405, 5573391, 5809396, 5581218, 4505718, 4439089, 4242428, 4439084, 4266689, 4322411, 3961575, 4142061, 4862019, 5311920, 5254401, 5521056, 5328399, 5414806, 5414814, 4836716, 5236865, 5176701, 4755093, 5040011, 5521058, 5521065, 5319320, 5176728, 4729878, 5039998, 5345108, 5256737, 4826826, 5177696, 5521072, 5210075, 5254380, 5433378, 5319305, 5030281, 5319303, 4862025, 5030267, 4916588, 5521061, 5121479, 5521088, 5521067, 5433359, 5433360, 5521074, 4628756, 5359831, 4856655, 5414816, 4729879, 4856672, 4627043, 5341781, 5233257, 4721850, 5165439, 5176711, 5341784, 5359848, 5345122, 5311893, 5448306, 5448311, 4627088, 5433388, 5311895, 5121476, 5094446, 5216644, 5433382, 4977030, 5521091, 5030283, 5448302, 5319302, 4826835, 5328515, 5461786, 5521097, 4721845, 5040007, 4724886, 4768164, 5216636, 5433362, 5256704, 5254387, 5163017, 4768146, 5328394, 4724871, 5341792, 5448301, 4862032, 4862005, 6088006, 6238015, 6352917, 6079067, 6227808, 6339334, 6553707, 6468276, 6079040, 6352926, 6178167, 6178154, 6301599, 6352920, 6301593, 6003121, 6110983, 6238033, 6553706, 6254835, 6088020, 6254839, 6388553, 6079061, 6254849, 6301585, 6438042, 6301607, 6110995, 6254855, 6339335, 6251509, 6094930, 6388560, 6227832, 6045159, 5581215, 6238024, 5581211, 6079043, 6227830, 5956445, 6156660, 6080909, 6156658, 6238019, 6151806, 6339325, 6269463, 6388582, 6269478, 6388549, 5998758, 6088017, 6094936, 6339326, 5956438, 6238031, 6388551, 6156663, 5569335, 6079054, 5956436, 6251516, 6291906, 6003144, 5602389, 6301588, 6254838, 6301608, 6388571, 6388575, 6437434, 5581213, 5998747, 6227828, 6291929, 6088018, 5951182, 6291928, 6079050, 6428588, 6438040, 5569330, 5845230, 6079065, 6437444, 6156657, 6325834, 5956409, 6227818, 5998763, 6468252, 6388586, 6352914, 6094938, 6325855, 6468261, 6553722, 6339341, 6151796, 6045147, 6437451, 6254842, 5998764, 6428598, 6291923, 6151783, 6254847, 6238022, 6003106, 6251526, 5998743, 6254859, 6325852, 6251514, 6291907, 6301587, 6151786, 6003125, 6151787, 6251532, 5573407, 6291917, 6003094, 6291926, 6325840, 6087982, 6339329, 6301605, 6468257, 6339320, 6251517, 6238040, 6269462, 6301595, 5998737, 5845261, 6388561, 6468254, 6553702, 6428616, 6227807, 6339322, 6388555, 6291930, 6339333, 6553708, 5897767, 6094928, 6110994, 6110978, 6110993, 6251536, 5998739, 6251525, 6428595, 6151791, 6352918, 6254863, 6088021, 5809376, 6553728, 5573405, 6428601, 6388552, 5998766, 6151784, 6388590, 6352919, 5845245, 6227823, 6254856, 6560895, 6156662, 6388558, 6388576, 6339342, 6325831, 6437453, 5573410, 5569326, 6251530, 6553737, 6437443, 6238027, 6178165, 5956429, 6428593, 6301594, 6388564, 6428611, 6080904, 6269482, 6388581, 6238014, 6388587, 5913049, 6094932, 6269486, 6094917, 6238021, 6301586, 6178168, 6269469, 6269480, 5951185, 6251508, 6269483, 6151797, 6301604, 6156678, 6269466, 6352916, 6339327, 6352908, 6388574, 6388580, 6238035, 6553710, 5602790, 6560868, 6388548, 5951171, 6254853, 6325848, 6269490, 6227822, 6291927, 6388554, 5644406, 6269472, 6553709, 5573414, 6088004, 5956441, 6003138, 6003141, 6238025, 6251515, 6291921, 6437455, 6269470, 6088014, 6238034, 6269487, 6003092, 6227814, 6079062, 6269485, 6251527, 6156666, 6003073, 6178172, 6291909, 6428589, 6325827, 6087985, 6003108, 6352909, 6553730, 6110989, 5604894, 6087983, 6428610, 6094942, 6254848, 6178173, 6339347, 6251534, 5804552, 5604866, 6428608, 6003063, 6251535, 6254852, 6238023, 6352904, 6238038, 6269471, 6325837, 6291924, 6110999, 5804543, 6468273, 6003124, 6301584, 6339338, 6227821, 6468250, 6291918, 6291915, 5602788, 6352921, 6156670, 6254837, 5604884, 6227829, 6238029, 6251520, 6045152, 6080912, 6178174, 6254851, 6560857, 5951173, 5573388, 6238030, 6388585, 6468274, 6251529, 6339321, 6254857, 6352899, 5569325, 6325830, 6178178, 6339323, 5644404, 6553723, 6325842, 6438029, 6438048, 6438051, 6227824, 5956431, 5998850, 6110998, 6087999, 6080898, 6151782, 6227810, 5956425, 5956413, 5809385, 6003118, 6352933, 5604875, 5581222, 5602772, 5581228, 4507316, 5256739, 5040006, 5521108, 5414785, 4889611, 4889606, 5433369, 5461787, 5121494, 5210054, 5448307, 5521069, 5319316, 5236854, 5345106, 5448296, 5461793, 5087155, 5414786, 5345093, 5521062, 5433380, 5328390, 4898379, 5521112, 5461772, 5359846, 5345116, 5216656, 5256730, 5319317, 5087162, 5216657, 4898396, 5521064, 4661106, 4977031, 5311912, 5165455, 5311918, 5414793, 4627125, 5311919, 5176709, 4836702, 4977014, 5121474, 5521059, 5359838, 5152496, 5236855, 5233243, 4956324, 5461777, 4990311, 5345115, 5328404, 5644422, 6291919, 6560898, 6560903, 6080902, 5804551, 5644428, 6553725, 6779712, 6560889, 6468256, 6428615, 6779728, 6848830, 6428592, 6110977, 6388538, 6560884, 6845825, 6845844, 6094925, 6560874, 6776807, 6291920, 6291914, 6560850, 6779719, 5602768, 6227834, 6325829, 6227815, 6828042, 6428613, 6428597, 6560906, 6352925, 6560888, 6779716, 6352907, 6776840, 6848859, 6045162, 6437436, 6848840, 6087994, 6560881, 6388568, 6560893, 6339328, 6776808, 6428590, 5604881, 5845238, 6352910, 6301589, 5602774, 6339346, 6828029, 6388579, 6352931, 6553704, 6560851, 6779729, 6437431, 6553700, 6776796, 6826040, 6269479, 6553732, 6745545, 6291903, 6745563, 6560891, 6776833, 6269481, 6437442, 6560859, 6080920, 6080900, 6080926, 6776805, 6438054, 5998751, 6438053, 6848858, 6087996, 6388546, 6560875, 6352913, 6553694, 6428600, 6178153, 5602377, 6087979, 6560869, 6080901, 6779733, 6438039, 6553738, 6468262, 6848871, 6438041, 6845835, 6325832, 6110979, 6560882, 6156680, 6845843, 6325836, 6848832, 6776824, 6438057, 6779725, 6388593, 5569349, 6325854, 6845832, 6779751, 6553682, 6045149, 6251522, 6437460, 6325846, 6428605, 6779703, 6745549, 6560909, 6560894, 6428603, 6438028, 6848854, 6828028, 6003114, 6776848, 6437454, 6291916, 6291922, 6388540, 6437459, 6848866, 6251531, 6428594, 6776799, 6828040, 6110975, 6560877, 6437433, 6553718, 6388545, 6745547, 6776828, 6779715, 6779704, 6848877, 6325843, 6438049, 6079066, 6468251, 6779741, 5604888, 6110982, 6553679, 6437456, 5951177, 6553711, 6553683, 6428591, 6779714, 6438052, 6845824, 6776801, 6080907, 6468272, 5951174, 6388556, 6560899, 6438047, 6088009, 6553739, 5602795, 6776812, 5573406, 6553691, 6560866, 6438031, 6553687, 6003143, 6826041, 6339345, 6560856, 6560864, 6087984, 6845838, 6553684, 6437439, 6151780, 6776831, 6468265, 6828035, 6848878, 6110996, 6178166, 6339330, 6553733, 6269467, 6045146, 6826042, 6178160, 5579220, 6560853, 6178162, 6845827, 6560865, 6826045, 6776800, 6352934, 6325828, 6339324, 6848872, 6776794, 5602366, 5804566, 6745565, 6776818, 6826039, 6745550, 6848853, 6845853, 6776816, 6437445, 6553729, 6553724, 6301612, 6828027, 6227820, 6468248, 6828032, 6325835, 6291908, 6560871, 5956421, 6151792, 5804562, 6826046, 6079064, 6045166, 6848862, 6428604, 6238042, 6560901, 6828043, 6745560, 6826034, 6745556, 6553727, 6151799, 6553714, 6094933, 6437440, 6848863, 6848882, 6428617, 6428612, 6745554, 6745544, 6779730, 6553721, 6251513, 6776849, 6553686, 6776842, 6560862, 6779723, 6227833, 5602368, 6080924, 6776821, 6560892, 5956440, 6352930, 6468258, 6553705, 6745553, 6079056, 6437438, 6776839, 6779731, 6779700, 6560878, 6553697, 6828037, 6251512, 6352912, 5581212, 6388584, 6848876, 6560900, 6094918, 6776813, 6848880, 6848829, 6080916, 6776809, 6560876, 5644415, 6388578, 6848855, 6553696, 6776843, 6560855, 6151803, 6151805, 6269475, 6779718, 6269468, 6776802, 6848839, 6779707, 6251518, 6301610, 6178161, 6779726, 6553688, 6111000, 6553701, 6848860, 6437437, 6468266, 6388592, 6776851, 6087978, 6110986, 5569346, 5604886, 6088007, 6553698, 6779744, 6848864, 6003088, 6238039, 6325838, 6553726, 5581205, 6553692, 6254841, 6560861, 6339344, 6438033, 6079035, 6003149, 6745546, 6779740, 5998762, 6045161, 6156665, 6845850, 6848849, 6238036, 6094926, 6468270, 6388588, 5998851, 6088003, 6088016, 5809387, 5809381, 5809388, 6094937, 6779711, 6003064, 5956435, 6352898, 5956407, 6178179, 6251521, 5956446, 6325844, 6468268, 6438037, 7034199, 6941296, 6905133, 7043815, 7002841, 6909021, 7034210, 6905148, 7043825, 6905147, 6905101, 6905119, 6909014, 6977754, 6905140, 6941252, 6977764, 7043842, 6977761, 6941270, 6941292, 6977773, 6977770, 7043823, 6905146, 7043809, 7034209, 6941307, 6941258, 6941282, 6941255, 6977756, 6977780, 6977778, 6905127, 6941262, 6941271, 7002852, 6909032, 6941305, 6905122, 6909019, 6905103, 6905154, 6905144, 6970714, 6941298, 6941251, 6941297, 6941264, 6905100, 6905102, 6909037, 6977763, 7043845, 6909002, 7002857, 6905151, 6909004, 6909034, 6977755, 6909018, 6909020, 6909023, 6970711, 6970712, 6970715, 6970716, 6970721, 6970722, 6977772, 4990322, 5256721, 5521084, 5341767, 5087142, 5521066, 5328403, 5341779, 5233250, 5311916, 5448292, 5152503, 5341772, 5345089, 5121475, 5341769, 4956323, 5210072, 5448313, 5328520, 5359841, 5210062, 5176714, 5461775, 5030275, 6045165, 6227806, 6080918, 6227825, 6776825, 6178169, 6776810, 6468259, 6254850, 6079055, 6352929, 6269476, 5998853, 6776814, 6779717, 6779732, 6080914, 6848867, 6845826, 6388559, 6776830, 6110984, 6745562, 5956430, 6826038, 6779705, 6269489, 6437450, 6826028, 6339348, 5956412, 5998852, 6151778, 6428602, 6087990, 6468247, 6437452, 6560873, 6156673, 6553690, 6560908, 6848842, 6776827, 6079053, 6828038, 6352901, 6388596, 5951184, 6848845, 6325850, 6110974, 6845841, 6828039, 6828031, 6388541, 6553713, 6438056, 6468263, 6845836, 6438038, 6301611, 5579217, 6776837, 6553680, 5998868, 6003129, 6779747, 6848834, 6845833, 6045168, 5602770, 6088008, 6291905, 6352915, 6779735, 6553731, 6560907, 6156664, 6745551, 6291902, 6156677, 6776804, 6388539, 6560858, 6828034, 6745557, 6045158, 6848857, 6291910, 6352911, 6352928, 6428606, 6003135, 6776834, 6339336, 6745555, 6156683, 6779743, 6438050, 6003086, 5602777, 6388591, 5951179, 6779713, 5998761, 6080919, 6178158, 5998866, 6745552, 6151788, 6468249, 6776846, 6553695, 6080905, 5809392, 6560896, 6388577, 6388566, 6845847, 6438035, 6428609, 6301600, 6826044, 6003113, 6437449, 6227816, 6003089, 6560886, 6388563, 6848846, 6468275, 7002842, 7043839, 7043841, 6970719, 6977782, 7043814, 6909031, 6905150, 7034197, 6977787, 7034212, 7002849, 7034202, 6941268, 7002826, 6977760, 7002838, 7043835, 6941275, 7043844, 7034213, 6909030, 7002843, 6970720, 7034201, 7043805, 6905125, 6941306, 7043833, 6905106, 7002827, 7034203, 6977785, 6941290, 6905145, 7034214, 7043834, 7043816, 7043848, 7043852, 6909009, 6905137, 6909010, 7002822, 7043807, 7002853, 7002828, 6970710, 7002829, 7034207, 7002855, 7002848, 7002830, 7002835, 6905113, 6905135, 6905129, 7043838, 6905109, 7043810, 6941261, 7034208, 7002825, 6905112, 6977781, 6909006, 7043843, 6977762, 7043826, 7043818, 7043849, 6977751, 6905117, 7043817, 6941299, 6977779, 6941302, 6977776, 7002850, 6970708, 7043857, 6941304, 6977783, 6909027, 5521077, 5521070, 5521063, 5236879, 6110990, 6776847, 6553693, 6468271, 6776817, 6003099, 6003080, 6848837, 6087986, 6845840, 6745558, 6438032, 6388550, 6776803, 6269484, 6325845, 6848879, 6776838, 5579221, 5573389, 6560860, 5998757, 6779736, 6352924, 6238026, 6339332, 6553719, 6553716, 6352923, 6848869, 6468264, 6848841, 6094921, 5644429, 6776820, 6437432, 6779708, 5809398, 5809389, 5956423, 5998856, 5809386, 6848850, 6560870, 6301590, 6845834, 6254854, 6977750, 6909015, 7002839, 6977774, 7043846, 7002833, 6941273, 7043824, 6977769, 7043856, 6977757, 7043819, 6905152, 6909003, 7043836, 6970713, 7043812, 6941280, 6941265, 7043830, 6905114, 6905118, 7043811, 7043820, 6970717, 6905105, 6941287, 6941256, 6905149, 6941260, 6909026, 7043847, 6776822, 6079047, 6301601, 6088013, 6745561, 6848844, 6779724, 6339339, 6560852, 6437435, 6227831, 6909008, 6941294, 7002859;
c) Item Number 314-13-15, UDI/DI 10885862172723, Serial Numbers: 2336366, 2336360, 2336372, 2336354, 2204845, 2204849, 2331043, 2338609, 2331048, 2338607, 2336349, 2336358, 2504082, 2336355, 2204835, 2204842, 2204841, 2331045, 2204832, 2331047, 2204840, 2336364, 2204843, 2331044, 2336350, 2336371, 2479386, 2336365, 2336361, 2522082, 2622516, 2622514, 2336362, 2338622, 2338619, 2504071, 2460117, 2336370, 2460111, 2544647, 2336347, 2504075, 2522092, 2479387, 2622512, 2622517, 2460102, 2331050, 2336348, 2544653, 2336374, 2338617, 2622503, 2504084, 2504085, 2504080, 2544646, 2504076, 2331046, 2460118, 2622502, 2622520, 2622532, 2479385, 2689095, 2689115, 2812056, 2812061, 2812066, 2331475, 2544652, 2544677, 2336351, 2204847, 2544669, 2336373, 2338612, 2204839, 2460113, 2460116, 2460114, 2544666, 2544667, 2331051, 2544658, 2504078, 2504077, 2338611, 2338615, 2622526, 2460108, 2622513, 2204834, 2331474, 2544661, 2336346, 2479402, 2460100, 2544674, 2336357, 2544659, 2622527, 2544672, 2479397, 2331477, 2700861, 2812064, 2777499, 2689106, 2770505, 2951878, 2777493, 2951866, 3577238, 2812057, 2777498, 2700858, 3577240, 2973848, 3588563, 3577230, 2770513, 2700857, 3588562, 2770499, 2770500, 2777497, 2770501, 2951877, 2689113, 2700864, 2689096, 3588566, 2770507, 2689107, 2689109, 2812071, 2973841, 2973821, 2973863, 2973849, 2973836, 3594077, 3588568, 2777489, 2973831, 2770502, 3594087, 3594088, 2522081, 2544665, 2479396, 2479393, 2338621, 2544668, 2622519, 2544656, 2622509, 2622521, 2544648, 2544655, 2479390, 2460105, 2951862, 2951873, 2777492, 3588555, 3594069, 2812063, 2689102, 3594080, 3594082, 2770519, 3897173, 3577237, 2689112, 3672115, 3672119, 3897156, 3897177, 3577223, 3588559, 3917657, 3917644, 3672127, 3672184, 3577234, 2973855, 2951869, 3594092, 3897141, 3897163, 3594081, 3843774, 2973860, 3594096, 3577226, 2973825, 3917650, 3897137, 4150442, 2973854, 2951880, 3594102, 2812049, 2951860, 3672126, 3672130, 2812060, 3594089, 2951872, 3588547, 3588560, 3588558, 3843757, 2770520, 3594079, 2973827, 3672120, 3594078, 3672113, 3672117, 3672191, 3715460, 3715462, 3715466, 2544650, 2479395, 2504073, 2522083, 2544679, 2504081, 2622500, 2460109, 2544649, 2338614, 2338610, 3594099, 3672194, 3588573, 2700853, 4150444, 3594090, 2973865, 2973840, 4142087, 2812059, 4150445, 4339654, 2973857, 4150451, 2777485, 2812067, 3588569, 4257029, 3897174, 3897167, 3917651, 3577236, 3897145, 3672196, 4150466, 4429053, 4216286, 4242455, 4224314, 3897144, 4216284, 4339648, 4182960, 3897165, 3917653, 4257030, 4266593, 3672123, 3672187, 2973852, 3594065, 4193505, 3897152, 4339633, 3672128, 4150453, 4242452, 4224313, 2689103, 4242456, 4242459, 4339645, 3843769, 4429052, 4429046, 4429047, 4142091, 4266596, 3917665, 4257028, 2951863, 3672190, 3672116, 4142089, 3672192, 3925835, 4122648, 2812050, 4429039, 2973843, 3594075, 4150460, 4150462, 2689098, 2973823, 4339655, 4266610, 3897142, 4224177, 2622522, 2522091, 2622508, 2544657, 2331052, 4193500, 2812052, 4266588, 3577233, 4150463, 2812055, 3577231, 2689101, 3672193, 3594064, 4469353, 3588577, 4469351, 4339647, 4429050, 3843765, 4193503, 3672183, 2973828, 4339656, 2973853, 2973861, 4224315, 2951861, 2770506, 4469342, 3588557, 3594094, 4339631, 4150443, 2777494, 2973835, 3843756, 4339650, 2770510, 4142095, 3917659, 3594076, 2951859, 3897158, 4469346, 2770512, 3843760, 2689100, 2973850, 4266604, 2812062, 4122645, 4257027, 4193502, 4142086, 4429049, 3594072, 3594066, 4224318, 4266612, 3588549, 4122647, 3865805, 3925837, 4150459, 2951858, 3917646, 4150456, 4150446, 2973844, 4266580, 3897170, 3588561, 4266613, 3897164, 4429051, 4975634, 4666483, 4645417, 4645407, 4666474, 5061033, 4645437, 5133367, 4889572, 4645431, 4645430, 4889563, 4666479, 4836154, 5033584, 4645449, 4889579, 4645409, 4645453, 4645420, 4924865, 4810065, 5064267, 4645412, 4810074, 4645446, 4666471, 4861709, 4795194, 4924839, 4836163, 4924866, 4924849, 4975630, 4889587, 5160525, 4645447, 4924858, 4810070, 3588571, 2700866, 3577225, 2973832, 4266578, 4266602, 4339639, 3588551, 3917655, 3594074, 3594068, 3588553, 4339657, 4142096, 4142085, 4142093, 4266590, 2689114, 4257026, 3594086, 4469335, 4469341, 4339659, 4429062, 4469329, 3588548, 3577229, 4224171, 4142090, 3897160, 2973846, 3672200, 4469338, 4242460, 4469352, 4339635, 4339636, 3917654, 3897166, 2700860, 3843763, 2973826, 4224312, 3917656, 4924846, 4666488, 4645454, 5133359, 5160529, 5290578, 4645438, 5284319, 5133376, 4810059, 5303674, 5133356, 5303651, 4836160, 5331210, 5033585, 4889582, 4666472, 4924855, 5160517, 4645432, 5284324, 4924844, 4889586, 4950849, 5160531, 5160533, 5290574, 5325806, 5049392, 5160527, 5485392, 5458491, 5458506, 5331944, 5331221, 5303655, 5331215, 4836149, 5049385, 5290589, 4836158, 5290584, 5284323, 5458500, 4645421, 4645442, 4645450, 4950853, 5303661, 5284318, 4810068, 5303667, 5331202, 5458507, 5458505, 4889565, 5576128, 5576124, 5615376, 5615367, 5615362, 5576137, 3865803, 4469345, 3672199, 4216285, 4469350, 3588575, 3917666, 4469340, 4142100, 3843772, 4224172, 4224179, 4150454, 3925834, 4339658, 4242454, 4429040, 4224310, 4242458, 3843767, 3672131, 3917652, 4889576, 4889585, 5133355, 4889569, 4924841, 4889590, 5284307, 4645427, 4645416, 5457301, 5331209, 4645429, 4666485, 5133360, 5160523, 4666480, 5331193, 4836153, 4666481, 5457300, 4645439, 4645456, 5331218, 5303647, 4924840, 5485382, 4924864, 4924857, 4795197, 5325801, 5033583, 4645460, 5485389, 4645459, 4924863.5133371, 4645436, 5458498, 5033578, 5457295, 5049386, 5325805, 5615387, 5583872, 5615377, 5621801, 5583868, 5943283, 6039340, 5583886, 5959811, 5583885, 5615369, 5576143, 5583880, 5576147, 5615370, 5621798, 6051658, 6081635, 6039336, 6081622, 6051639, 5583873, 5576146, 5576131, 6039332, 5583887, 5615361, 5576135, 5615386, 5576130, 6051665, 5576134, 5583894, 6051652, 6051661, 6081641, 6081631, 5621797, 5621799, 5621819, 6238007, 6238012, 6087946, 4339638, 4339643, 4429060, 4429058, 4339646, 4266600, 4266585, 5331935, 5485399, 4645433, 5457304, 5325816, 5303672, 5331945, 5303654, 5284299, 5133363, 5160511, 5331927, 4795192, 4889568, 5325796, 4924838, 5033577, 5331207, 5303669, 5331947, 5160526, 5331925, 5331204, 4924843, 5049384, 5325803, 6388515, 6251169, 6388510, 5576136, 6087943, 6087938, 6087948, 6516137, 5576142, 6238008, 6087949, 6251167, 6356272, 5576145, 6081629, 6291932, 6307750, 6238005, 6291938, 6339003, 6356261, 6356266, 6356269, 6388514, 5160504, 5331933, 5458488, 4924854, 4950844, 5303648, 4889570, 5290597, 4975631, 5457307, 5331198, 5133354, 5303657, 5457310, 5331931, 4645402, 4645452, 4924860, 4810073, 4645404, 4645445, 5133375, 6356259, 5583882, 6087947, 6388531, 6829919, 5583884, 6388534, 5576149, 6554922, 6554926, 5576123, 6081632, 6238011, 6323511, 6051645, 6454936, 6452636, 6554918, 6516144, 6388526, 6516152, 6356267, 5621807, 6972998, 6937063, 6936644, 6937058, 6937054, 6937041, 7004286, 6936664, 6972984, 6972985, 6972986, 4924850, 5458509, 5331192, 5325795, 5485388, 5133364, 5458508, 5331943, 5457297, 5064273, 5290579, 4889567, 5331937, 5331219, 5331949, 5049393, 5331948, 5331199, 5303653, 5325809, 6452618, 6087958, 6307749, 5959805, 6554920, 6291943, 6087935, 6087940, 6070747, 6554875, 6452628, 6356264, 6291931, 6070746, 6238002, 6554911, 6039335, 7018518, 6936652, 6936638, 6936648, 6937046, 7004282, 6936656, 6937059, 7004273, 6936661, 5303663, 5943276, 6338998, 6554882, 6081626, 6087969, 6081615, 6454929, 7018509, 7004284, 6972996 |
Recalling Firm/ Manufacturer |
Exactech, Inc. 2320 Nw 66th Ct Gainesville FL 32653-1630
|
For Additional Information Contact | Thien Doan 352-377-1140 |
Manufacturer Reason for Recall | The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH). |
FDA Determined Cause 2 | Nonconforming Material/Component |
Action | Beginning 03/06/2024, Exactech issued an URGENT MEDICAL DEVICE RECALL notice and an URGENT DEAR HEALTHCARE PROFESSIONAL COMMUNICATION to its consignees via email. The recall notice explained the issue, clinical impact, and requested the following actions be taken:
"" CAREFULLY REVIEW THIS RECALL NOTIFICATION to ensure that you fully understand the issue identified, the recall strategy, and all actions required.
" IMMEDIATELY IDENTIFY AND QUARANTINE any of the subject devices in your inventory and/or customer s inventory listed Product Scope Listing.
" EXTEND THE DESCRIPTION OF ISSUE AND CLINICAL IMPACT as described in the recall notification to your accounts that may have this product in their possession.
" COMPLETE AND RETURN the attached Recall Inventory Response Form to Exactech via email at recalls@exac.com.
The communication to healthcare providers referenced the FDA Safety Communication issued on 01/16/2024, explained the issue, potential health risks, and suggested patient monitoring for those with implanted affected devices for potential device wear, failure, or bone loss and should consider performing X-rays to further evaluate the patient and their implanted device if you suspect a failed device.
Contact your local Exactech Representative if you have any questions regarding this communication. |
Quantity in Commerce | 12643 units |
Distribution | Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database | 510(K)s with Product Code = KWS
|
|
|
|