Date Initiated by Firm |
January 31, 2024 |
Date Posted |
April 17, 2024 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1593-2024 |
Recall Event ID |
94138 |
Product Classification |
Calibrator, secondary - Product Code JIT
|
Product |
Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567
|
Code Information |
GTIN: 05055273204520
(1) Batch/Lot: 619282 / 1094MA-1098MA |
Recalling Firm/ Manufacturer |
Randox Laboratories Ltd. 55 Diamond Road Crumlin (North) Ireland
|
Manufacturer Reason for Recall |
Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Randox Laboratories (Manufacturer) issued Urgent Field Safety Notice on 31 Jan 24
Complaint Reference: REC716 letter via email to Distributors. Distributors will contact the end-users. Letter states reason for recall, health risk and action to take:
We have reassigned
the target values in the above lot numbers in line with the ERM reference material. Please
refer to the table below for the updated calibrator targets. You may experience a shift in
Quality Control and patient sample recovery of up to +12%. Please discard all copies of the calibrator Instructions For Use (IFU)s and download the updated IFUs from www.randox.com .
If further information is required, please contact technical.services@randox.com .
In line with the restandardisation to the Microalbumin Calibrator Series, we are updating the
target and range for the Microalbumin Liquid control, MA1361, lots 1102MA and 1103MA
which are present in batch 631698. The updated targets and ranges can be found in the table below. Please discard any copies of the IFU and download the updated version from www.randox.com.
|
Quantity in Commerce |
8 units |
Distribution |
US Nationwide distribution in the states of GA, MA, WV including PR. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|