| | Class 2 Device Recall iLED 7 surgical lights |  |
| Date Initiated by Firm | June 04, 2024 |
| Date Posted | July 05, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2315-2024 |
| Recall Event ID |
94766 |
| Product Classification |
Lamp, surgical - Product Code FTD
|
| Product | iLED 7 surgical lights, Model Numbers:
a) 4068110 (Ceiling Single Surgical Light);
b) 4068120 (Mobile Surgical Light);
c) 4068140 (Pendant Surgical Light);
d) 4068210 (Ceiling Duo Surgical Lights);
e) 4068310 (Ceiling Trio Surgical Lights);
f) 4068410 (Ceiling Quad Surgical Lights) |
| Code Information |
All Serial Numbers:
a) UDI/DI 00887761968325;
b) UDI/DI 00887761968318;
c) UDI/DI 00887761968301;
d) UDI/DI 00887761968295;
e) UDI/DI 00887761968288;
f) UDI/DI 00887761968271
|
Recalling Firm/ Manufacturer |
Baxter Healthcare Corporation 1 Baxter Pkwy Deerfield IL 60015-4625
|
| For Additional Information Contact | Center for One Baxter 800-422-9837 |
Manufacturer Reason for Recall | Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction due to customer reports of thermal skin injuries when using the iLED 7 surgical light
system products listed below. Specific VidiaPort and Trulight product codes are also impacted as they are equipped with iLED 7 light-heads |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Baxter issued an Urgent Medical Device Correction notice to its consignees on 06/04/2024 via US mail, first class. The notice explained the issue with the devices, potential hazards involved, and requested the following:
"Actions to be Taken by Customers:
1. Operators may continue to use the iLED 7 surgical light systems by following
the current instructions for use (IFU) cautions and warnings while considering
the additional information below.
. Utilize the lowest possible illumination level suitable for the procedure,
especially in certain neurological or intestinal procedures on delicate,
thin, dry, or abnormal tissue.
. Avoid overlapping light fields that are set at high intensity (80% or higher).
- If the light intensity of one light-head is set at 80% or higher the
second light-head, if overlapping, should be set at 50% or lower.
- If the light intensity of two or more light heads is set at 80% or
higher, to minimize risk do not overlap them.
. If a very high-intensity setting is temporarily required, reduce intensity as
soon as the need passes.
. Ensure the Adaptive Light Control (ALC) Plus sensor is activated and
indicated on the wall or mobile control panel. If activation of the ALC Plus
is not possible, please contact a technical service representative for
further instructions.
. If the ALC Plus is switched off, the light-head needs to be positioned at
100 centimeters (39.37 inches) from the surgical field. At other distances
the illuminance and irradiance limits can be exceeded, which may result
in thermal skin injuries.
2. Additional safety information reviewing the potential risks of using an overhead
surgical light, including educational safety content is available through Pfiedler
education (a division of AORN) at the URL listed below. This website requires
a one-time registration.
https://www.pfiedlereducation.com/diweb/gateway/f/https*3A*2F*2Fwww.pfi
edlereducation.com*2Fdiweb*2Fcatalog*2Fitem*2Feid*2F1552-2024. "
The notice also requested tha |
| Quantity in Commerce | 6243 units |
| Distribution | Worldwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|